Commands

8.2 Command lists

Command code Command Description


12 Load standard parameters Like "Load factory settings" (command code 11), but inter-
face settings for Ethernet and Modbus RTU are not reset to
the factory setting.
31 Reserve
51 Reserve
60 Calibration point 0 valid Calibration point 0 valid / save values for calibration point 0.
61 Calibration point 1 valid Calibration point 1 valid / save values for calibration point 1.
62 Calibration point 2 valid Calibration point 2 valid / save values for calibration point 2.
81 Characteristic shift Move calibration characteristic. The command defines the
current weight of the scale as the new zero point (0 kg) and
shifts the complete characteristic without changing the gra-
dient. The command can be used, for example, in order to
compensate parts used for mounting calibration weights on
the scale at the end of the calibration.
82 Perform automatic calibration Calculating the scale characteristic curve with reference to
the load cell from data record 10. The calculated character-
istic curve is entered directly in data record 3 and 4, and
thus activated immediately after executing the command.
The scale must be empty when the command is executed.
83 Perform calibration check The command calculates the theoretical digital values
in•relation to the calibration weights using the load cell pa-
rameters from data record 10 and the adjustment weights 0,
1 and 2 from data record 3. The output of these theoretical
digits is made in data record 4. The function can be used to
check the plausibility of adjustment digits in data record 3,
which have been determined in a legal-for-trade calibration.
907 Apply load cell impedance The currently measured impedance value (DS31) is trans-
ferred to the DS 10.

Table 8- 2 Commands 450 ... 499: Trace commands

Command code Command Description


451 Trace RAM on Start permanent tracing
452 Trace RAM off Stop permanent tracing
454 Delete trace RAM Delete tracing memory.

Table 8- 3 Commands 700 to 899: HMI display switching

Command Command Description


code
701 Increased resolution Increased resolution on the main display
705 Display current tare weight Current tare weight on the main display
710 Activate standard display Activate standard display gross/net (main dis-
play)
714 N Process value Show net process weight on the main display

SIWAREX WP521/WP522
134 Manual, 04/2016, A5E36695225A