Autoclave STATIM 2000 ERROR CODES (NEW ERROR CODES)

Statim 2000 New Error Codes 


What They Mean & How To Correct The Problem

New Error Codes

There are 2 different sets of error codes: Original Error Codes are prefaced by "Check Cassette" followed by
a number &s; New Error Codes are prefaced by "Cycle Fault" followed by a number. 

Be sure to select the right error code for your unit 

The Statim Sterilizer should never leak steam or have water dripping under the machine. If either of these
conditions occur, check your cassette and seal. All leaks should be corrected before proceeding with
troubleshooting procedures. 

Error
Problem Corrective Action
Message

Cycle The cassette temperature failed to reach 95°C No power to boiler, Check Thermal Fuse.
Fault 1 within a time-out period
Replace if necessary Part #SCF006-377 

An extremely large steam leak. Can be


caused by over loading the cassette

Cycle The cassette temperature failed to increase from An extremely large steam leak. Replace
Fault 2 95°C to 100°C within a time-out period. seal or repair cassette as needed. 

Can be caused by overloading the cassette

Cycle The cassette has failed to pressurize and achieve If visible steam leaks appear
Fault 3 a temperature of 110°C within a time-out period. replace cassette seal, lid or tray as
needed. 

If no leaks are visible disassemble


solenoid valve and check for debris. 

Make sure plunger slides smoothly in


plunger tube. 

Refer to Solenoid Valve Section. Check


for leaky pressure relief valve (PRV) or
check valve.

Cycle The software causes the pump to activate at If visible steam leaks appear
Fault 4 predetermined times replace cassette seal, lid or tray as
needed. 

If no leaks are visible disassemble


solenoid valve and check for debris. 

Make sure plunger slides smoothly in


plunger tube. 

Refer to Solenoid Valve Section. Check


for leaking Pressure Relief
Valve or Check Valve.

Cycle Boiler went dry prematurely. If a request to pump water occurs outside
Fault 5 of the predetermined time Cycle Fault 5
occurs. Check pump using Pump Tester
Bottle Part #SCK024-384.

Cycle The software has detected a steam generator Calibrate boiler.


Fault 6 (boiler) temperature 5°C greater than the
chamber, within 7.2seconds after a purge during
the sterilizing phase of a cycle.

Cycle The cassette temperature has dropped 4°C below If visible steam leaks appear
Fault 7 set point replace cassette seal, lid or tray as
needed. 

Disassemble solenoid valve and check for


debris. 

Make sure plunger slides smoothly in


plunger tube. 

Refer to Solenoid Valve Section.

Cycle The software has detected a steam generator Calibrate boiler. 


Fault 8 (boiler) temperature 5'less than the chamber,
within 7.2 seconds after a purge during the
sterilizing phase of a cycle Check for kinks in the exhaust tubing. 

Check to see if condenser bottle is full of


hot water.

Cycle N/A N/A


Fault 9

Cycle The cassette temperature has failed to drop to Check for kinks in the exhaust tubing. 
Fault 10 115°C during the Unwrapped or Wrapped Cycle
or the temperature has failed to drop to 110°C Check for a clogged duct in the rear of the
during the Rubber and Plastics Cycle in the cassette tray. Blow out with compressed
purge conditioning stage air. 

Check solenoid valve for proper


operation.

Cycle The cassette temperature has failed to drop to Check for kinks in the exhaust tubing. 
Fault 11 102°C within 60 seconds of the end of a cycle
during venting Check for a clogged duct in the rear of the
cassette tray. Blow out with compressed
air. 

Check the solenoid valve for proper


operation.

Cycle This indicates a problem with the temperature Check thermocouples, they should read
Fault 12 measuring system approximately 10 ohms at room
temperature.

Cycle N/A N/A


Fault 13

Cycle The steam generator (boiler) temperature is Check pump using Pump Tester
Fault 14 above 171°C during the Sterilization phase of a Bottle Part #SCK024-384
cycle.

Cycle The cassette temperature is 3°C or more above Check for kinks in the exhaust tubing. 
Fault 15 set point during the Sterilization phase of a cycle
Check for a clogged duct in the rear of the
cassette tray. Blow out with Compressed
air. 

Check solenoid for proper operation.

Cycle The steam generator (boiler) temperature went Check pump using Pump Tester
Fault 16 above 171°C during the heat up phase of a cycle. Bottle Part #SCK024-384.

Cycle N/A N/A


Fault 17

Cycle N/A N/A


Fault 18

Cycle The steam generator (boiler) calibration is Calibrate the boiler. Refer to Boiler
Fault 19 invalid. A new calibration is required. Calibration Section

Cycle The pump has failed to pump water into the Check pump using Pump Tester
Fault 20 steam generator (boiler) during a pre-vent pump Bottle Part #SCK024-384.
time-out. 

The steam generator (boiler) temperature was


greater than 140°C for 3.6 seconds after the
pump was activated to pump water to cool the
steam generator (boiler).

Cycle N/A N/A


Fault 21

Cycle N/A N/A


Fault 22

Cycle N/A N/A


Fault 23

Cycle N/A N/A


Fault 24

Cycle The software has failed to detect a need to pump No power to boiler, Check Thermal Fuse.
Fault 25 water in 90 seconds
Replace if necessary Part #SCF006-377.

Cycle The sterilization phase has failed to start within 3 Calibrate boiler. Refer to Boiler
Fault 26 minutes of the cassette reaching the sterilization Calibration Section
temperature.

Cycle The internal temperature of the steam generator Check pump using Pump Tester
Fault 27 (boiler) has exceeded 150°C for 25 seconds Bottle Part #SCK024-384

"Statim Autoclave Repair Made Simple™"