Autoclave Statim 2000 Error Codes
Autoclave Statim 2000 Error Codes
There are 2 different sets of error codes: Original Error Codes are prefaced by "Check Cassette" followed by
a number &s; New Error Codes are prefaced by "Cycle Fault" followed by a number.
The Statim Sterilizer should never leak steam or have water dripping under the machine. If either of these
conditions occur, check your cassette and seal. All leaks should be corrected before proceeding with
troubleshooting procedures.
Error
Problem Corrective Action
Message
Cycle The cassette temperature failed to reach 95°C No power to boiler, Check Thermal Fuse.
Fault 1 within a time-out period
Replace if necessary Part #SCF006-377
Cycle The cassette temperature failed to increase from An extremely large steam leak. Replace
Fault 2 95°C to 100°C within a time-out period. seal or repair cassette as needed.
Cycle The cassette has failed to pressurize and achieve If visible steam leaks appear
Fault 3 a temperature of 110°C within a time-out period. replace cassette seal, lid or tray as
needed.
Cycle The software causes the pump to activate at If visible steam leaks appear
Fault 4 predetermined times replace cassette seal, lid or tray as
needed.
Cycle Boiler went dry prematurely. If a request to pump water occurs outside
Fault 5 of the predetermined time Cycle Fault 5
occurs. Check pump using Pump Tester
Bottle Part #SCK024-384.
Cycle The cassette temperature has dropped 4°C below If visible steam leaks appear
Fault 7 set point replace cassette seal, lid or tray as
needed.
Cycle The cassette temperature has failed to drop to Check for kinks in the exhaust tubing.
Fault 10 115°C during the Unwrapped or Wrapped Cycle
or the temperature has failed to drop to 110°C Check for a clogged duct in the rear of the
during the Rubber and Plastics Cycle in the cassette tray. Blow out with compressed
purge conditioning stage air.
Cycle The cassette temperature has failed to drop to Check for kinks in the exhaust tubing.
Fault 11 102°C within 60 seconds of the end of a cycle
during venting Check for a clogged duct in the rear of the
cassette tray. Blow out with compressed
air.
Cycle This indicates a problem with the temperature Check thermocouples, they should read
Fault 12 measuring system approximately 10 ohms at room
temperature.
Cycle The steam generator (boiler) temperature is Check pump using Pump Tester
Fault 14 above 171°C during the Sterilization phase of a Bottle Part #SCK024-384
cycle.
Cycle The cassette temperature is 3°C or more above Check for kinks in the exhaust tubing.
Fault 15 set point during the Sterilization phase of a cycle
Check for a clogged duct in the rear of the
cassette tray. Blow out with Compressed
air.
Cycle The steam generator (boiler) temperature went Check pump using Pump Tester
Fault 16 above 171°C during the heat up phase of a cycle. Bottle Part #SCK024-384.
Cycle The steam generator (boiler) calibration is Calibrate the boiler. Refer to Boiler
Fault 19 invalid. A new calibration is required. Calibration Section
Cycle The pump has failed to pump water into the Check pump using Pump Tester
Fault 20 steam generator (boiler) during a pre-vent pump Bottle Part #SCK024-384.
time-out.
Cycle The software has failed to detect a need to pump No power to boiler, Check Thermal Fuse.
Fault 25 water in 90 seconds
Replace if necessary Part #SCF006-377.
Cycle The sterilization phase has failed to start within 3 Calibrate boiler. Refer to Boiler
Fault 26 minutes of the cassette reaching the sterilization Calibration Section
temperature.
Cycle The internal temperature of the steam generator Check pump using Pump Tester
Fault 27 (boiler) has exceeded 150°C for 25 seconds Bottle Part #SCK024-384