You are on page 1of 7

Received: 1 October 2019 Accepted: 26 November 2019

DOI: 10.1111/cid.12915

ORIGINAL ARTICLE

Comparison of histomorphometry and microradiography


of different implant designs to assess primary implant stability

Georgios Romanos DDS, PhD1,2 | Michael Damouras DDS2 |


Alexander A. Veis DDS, PhD3† | Pablo Hess DDS2 | Frank Schwarz DDS, PhD2 |
Silvia Brandt DDS, PhD4

1
Department of Periodontology, Stony Brook
School of Dental Medicine, Stony Brook, Abstract
New York Objectives: To contribute toward optimizing the long-term stability of dental
2
Department of Oral Surgery and Implant
implants. Our working hypothesis was that the degrees of immediate implant-bone
Dentistry, Center for Dentistry and Oral
Medicine, Johann Wolfgang Goethe contact, and hence of primary stability, would demonstrably differ between implant
University, Frankfurt am Main, Germany
systems due to their different external geometries and thread designs (macro-design).
3
Department of Dento-alveolar Surgery,
Implantology and Oral Radiology, School of This demonstration was provided in a bovine model (ex vivo) by employing and com-
Dentistry, Aristotle University of Thessaloniki, paring histomorphometry and microradiography as evaluation methods.
Thessaloniki, Greece
4
Materials and methods: A total of 120 implants, representing six different implant
Department of Prosthodontics, Center for
Dentistry and Oral Medicine, Johann thread designs, were inserted following the recommended surgical protocol in ribs of
Wolfgang Goethe University, Frankfurt am
freshly slaughtered cattle. Twenty specimens of implants with surrounding bone
Main, Germany
were prepared per system and were divided into two equally sized groups of 60 spec-
Correspondence
imens for analysis by either histomorphometry or micro-computed tomography. Data
Georgios Romanos, DDS, PhD, Department of
Periodontology, Stony Brook University, were analyzed by Mann-Whitney U test (P ≤ .05).
School of Dental Medicine, 106 Rockland Hall,
Results: One of the implant systems, featuring a slight tapered external geometry
Stony Brook, NY 11794.
Email: georgios.romanos@ and a progressive thread design, consistently revealed the most favorable bone-
stonybrookmedicine.edu
implant contacts in both histomorphometric and microradiographic evaluations.
Overall, consistently higher values of bone-implant contact were obtained with the
microradiographic than the histomorphometric approach, and this difference reached
statistical significance in three of the six implant systems tested.
Conclusions: Progressive threads offering a bone-condensing effect can significantly
help to maximize implant-bone contact percentages. Compared to his-
tomorphometry, microradiography is likewise a suitable method to evaluate bone-
implant contact, offering the additional benefits of being noninvasive and less time
consuming.

KEYWORDS

bone-implant contacts, dental implants, primary stability, x-ray microtomography

1 | INTRODUCTION

Dental implants are a well-established treatment modality to replace



Died November 21, 2016. missing teeth. Implants represent a treatment option in oral

Clin Implant Dent Relat Res. 2020;1–7. wileyonlinelibrary.com/journal/cid © 2020 Wiley Periodicals LLC 1
2 ROMANOS ET AL.

rehabilitation to secure different kinds of prostheses in partially or 2 | M A T E R I A L S A N D M ET H O D S


fully edentulous patients or as a single tooth replacement in the
esthetic zone. The only way to ensure adequate retention of any fixed 2.1 | Implant systems and implant insertion
or removable implant-supported restoration is by establishing reliable
anchorage with direct bone-implant contact. The term A total of 120 implants were included, representing six common implant
osseointegration was introduced in reference to the functional direct systems, which featured various shapes, thread designs, surface textures,
1,2
integration of load-bearing implants with living bone. implant-abutment connections, and implant dimensions. Table 1 gives an
Osseointegration has been defined as successful whenever rigid overview of these systems and these properties. In the following manu-
anchorage of an implant inside its endosseous bed is maintained, and script, the systems investigated are divided into groups in alphabetic
remains clinically asymptomatic, even under functional loading.3 order: group 1: Ankylos, group 2: Astra TiOblast, group 3: Frialit-2, group
Branemark first described the shapes and materials of dental implants. 4: Replace select tapered, group 5: Straumann, group 6: Xive.
Nowadays, the focus on the implant geometry shifted to the Ten bovine rib bones (only the middle part of the ribs was selected in
osseoinductive potential of the implant surface.4 order to have similar bone quality) from freshly slaughtered cattle were
Achieving proper osseointegration is very much a function of pri- obtained for implant insertion, which was performed in accordance with the
5
mary implant stability whose level, and the related degree of bone- surgical protocol recommended by the manufacturer for each specific sys-
implant contact, depends on a number of factors. Associations of pri- tem, using a standardized drilling unit (947/948 Intra K Boxes; KaVo,
mary stability have repeatedly been reported with implant macro- Biberach, Germany) and an appropriate 1:5 contra-angled handpiece (W&H
geometry and with the surgical implant procedure.6-8 A major role has Dentalwerk, Bürmoos, Austria). All implant beds were prepared with low
9
also been ascribed to the surface characteristics of the implant, and pressure, using irrigation with saline for cooling, at a speed of 800 rpm.
patient-specific factors, such as bone density and bone-condensation None of the implant systems required pretapping. All implants of each sys-
methods are known to affect the bone-implant contact.10 In this latter tem were inserted by six experienced oral surgeons having advanced train-
context, a classification system for bone quality (type I-IV) is available ing for the individual implant systems (each system placed by one clinician).
11
to support decisions in the perioperative stages.
Once the implant has been inserted with primary stability, the fur-
ther course of osseointegration is characterized by the development 2.2 | Specimen preparation
of secondary stability, which includes the formation of new tissue sur-
rounding the implant surface and depends on the degree of primary Specimens from the implants and their surrounding bone were created
stability achieved.12 The biomechanical process of secondary stability directly after the insertion procedure. Figure 1 illustrates how the bone
is determined as the structural and functional connection between was cut into square pieces, resulting in a total of 20 specimens for each
newly formed bone and implant surface, namely osseointegration.13 implant system. Subsequently the specimens were divided into two
Hence it is fair to state that primary stability is a fundamental prereq- equally sized groups of 60 specimens each. One group of specimens was
uisite for osseointegration and for the long-term stability of dental evaluated histologically and histomorphometrically, while the other one
implants. Various methods have been used to determine primary was analyzed by radiographic imaging in the form of micro-CT scans.
implant stability, including the measurement of insertion or removal
torque6,8,14) or the Periotest device.15 This technique is, however,
highly user-dependent and therefore offers little sensitivity. 2.3 | Histological and histomorphometric
Much more objective and reproducible approaches to determin- assessment
ing implant stability are available in the form of histological, notably
histomorphometric, and microradiographic techniques. The latter take Following fixation in 10% formalin for 48 hours and subsequent dehy-
the form of micro-computed tomography (micro-CT) scans16,17 and dration with alcohol, the specimens were embedded in methacrylate-
offer additional reliable insights into alveolar bone quantity and qual- based resin (Technovit 7200; Heraeus Kulzer, Wehrheim, Germany)
ity, which can pertinently be used to evaluate implant sites.18,19 and cut into thin sections (60-80 μm) using a high-speed microtome
Another benefit is their noninvasive nature, even though, as with any (Accutom 2; Struers, Copenhagen, Denmark) and polishing machine
radiological technique, radiation exposure is naturally a factor requir- (DAP-V; Struers). Then the ground sections were stained with tolui-
ing attention. dine blue 1% and pyronine G and analyzed using a light microscope
We designed the present study in an effort to contribute toward (Axiostar Plus; Zeiss, Göttingen, Germany) with an integrated color
optimizing the long-term stability of dental implants. Our working video camera (DC88AP; Sony, Tokyo Japan) (Figure 2). For the subse-
hypothesis was that the degree of immediate implant-bone contact, and quent histomorphometric investigation, a microscope (Zeiss;
hence of primary stability, would demonstrably differ between implant Oberkochen, Germany) and a computer setup with appropriate soft-
systems due to the different external geometries and thread designs ware (Videoplan; Kontron, Eching, Germany) were used. The BIC was
involved. This demonstration was to be accomplished in a bovine ex vivo measured with a 2.5× objective lens. A 10× lens was used to calculate
model, including microradiographic and histomorphometric techniques the area of the peri-implant mineralized bone. For calculation, 10 pre-
and also comparison between these two evaluation methods. viously defined fields of measurement were used.
ROMANOS ET AL. 3

TABLE 1 Overview of the six implant systems investigated in alphabetical order

I-A
System Manufacturer Surface pattern Shapes and thread designs connections Dimensions
Ankylos Dentsply Sirona Roughened by ablative processing Progressive thread depth and Conical Ø: 3.5 mm
(Group 1) (Mannheim, polished collar (2 mm) Length:
Germany) 11 mm
Astra Astra Tech OsseoSpeed (fluoride-modified Cylindrical Conical Ø: 3.5 mm
TiOblast (Gothenburg, roughened surface) Length:
(Group 2) Sweden) 11 mm
Frialit-2 Dentsply Sirona Trimodal morphology Stepped screw design Internal Ø: 3.5 mm
(Group 3) (Mannheim, hexagon Length:
Germany) 11 mm
Replace Nobel Biocare TiUnite (surface characterized by Tapered with non-self-tapping Internal Ø: 4.3 mm
select (Gothenburg, antioxidative modification) design tri-lobe Length:
tapered Sweden) 13 mm
(Group 4)
Straumann Straumann SLA (sandblasted, large-grit, Cylindrical in the middle and apical Internal Ø: 3.5 mm
(Group 5) (Waldenburg, acid-etched surface) and tapered in the cervical area octagon Length:
Switzerland) (tissue level) 12 mm
Xive Dentsply Sirona Sandblasted featuring a Advanced cylindrical design with a Internal Ø: 3.5 mm
(Group 6) (Mannheim, high-temperature-etched screw-type thread hexagon Length:
Germany) microstructure 11 mm

Note: I-A connection = implant-abutment connection.

Stanford, California) using the Wilcoxon-Mann-Whitney U test, a 95%


confidence interval, and P < .05 level of significance.

3 | RE SU LT S

3.1 | Histological findings

All implants had primary contacts with the bone surrounding them.
Some sites along the threads exhibited area-type bone contacts. Area-
type lingui-form bone contacts were typically seen along the coronal
and middle thirds of the implants. Different thread designs did not
involve differences in peri-implant bone quality.

3.2 | Histomorphometric findings


F I G U R E 1 Specimen preparation. Implants were inserted into
bone of uniformly high quality from freshly slaughtered cattle. Then,
the bone was cut into pieces incorporating one implant each Table 2 summarizes the percentages of bone-implant contact emerging from
the histomorphometric assessments based on both the total bone structures
2.4 | Microradiographic assessment and statistical and their cortical and cancellous compartments. A statistically significant dif-
analysis ference was found between the six implant systems (Mann-Whitney U test/
t test: P < .05). The results for total bone-implant contact percentages were
The radiographic evaluations were performed using a cone-beam micro- significantly more favorable with group 1 than group 3, group 4 and group
CT desktop unit (μCT 40; Scanco Medical, Bassersdorf, Switzerland) with 6 implants, and they were significantly better with the group 6 than with the
an x-ray tube featuring a focal point of 7 μm. Voxel size was 30 μm. For group 4 implant system (all P < .001; also see Table 4).
each implant, 450 layers were scanned from coronal to apical, the scan
height being 13.5 mm, and the tube voltage 70 kV. A complementary
evaluation script allowed us to calculate the percentage values of contact 3.3 | Microradiographic findings
area between each implant and its surrounding bone (Figure 3). Descrip-
tive calculations were followed by analysis with statistical software Table 3 summarizes the bone-implant contact percentages as evaluated
(SPSS; Statistical Product and Service Solutions, Stanford University, by micro-CT scanning. Again, the most favorable results were obtained
4 ROMANOS ET AL.

F I G U R E 2 Micrograph of histological
ground sections, displaying inserted
implants and the surrounding bone

F I G U R E 3 Microradiographs (micro-CT views) of an inserted implant and its surrounding bone. Red lines indicate the presence of direct
implant-bone-contact; green lines indicate gaps along the interface

TABLE 2 Bone-implant contacts (%) as evaluated by histomorphometry

Total bone-implant contact (%) Cortical bone contact (%) Cancellous bone contact (%)

Implant system Mean ± SD Min Max Mean ± SD Min Max Mean ± SD Min Max
Group 1 64.80 ± 9.51 42 74 92.50 ± 12.86 58 100 37.1 ± 13.08 19 54
Group 2 52.15 ± 6.0 40.5 62 82.30 ± 8.70 64 92 22.00 ± 6.03 16 33
Group 3 56.05 ± 6.3 40 63 98.80 ± 7.19 79 100 18.90 ± 8.62 1 28
Group 4 49.15 ± 11.11 25 61 72.70 ± 20.19 31 91 25.60 ± 7.56 16 38
Group 5 51.15 ± 8.87 39 67 79.80 ± 16.91 55 100 22.50 ± 4.74 17 34
Group 6 58.25 ± 7.40 47.5 70.5 94.81 ± 9.65 70 100 28.60 ± 8.12 14 41

Note: Data are presented as mean values, SD, and ranges (minimum and maximum values).

with group 1, and good interlocking with the surrounding bone was also 3.4 | Histomorphometry vs microradiography
seen with groups 4, 5, and 6 geometries. Less favorable results were
obtained for groups 2 and 3. Intergroup comparisons did reveal statisti- Table 3 also includes a method comparison. Overall, the microradio-
cally significant (P < .001) differences between group 1 and group graphic evaluations yielded markedly higher values of bone-implant
3, group 1 and group 6 and group 1/6 and group 4, group 6 and group contact than the histomorphometric evaluations. These differences
4, group 6 and group 3, as well as between group 4 and group 3 (also were not statistically significant for groups 4, 5, and 6 (Mann-Whitney
see Table 4). U test/t test: P = .08; also, see Table 4). The method-specific
ROMANOS ET AL. 5

TABLE 3 Bone-implant contacts (%) as evaluated by microradiography (micro-CT scanning) in comparison with histomorphometry

Microradiographic evaluation (%) Histomorphometric evaluation (%)

Implant system Mean ± SD Min Max Mean ± SD Min Max Method of comparison
Group 1 70.90 ± 3.68 64.95 73.83 64.80 ± 9.51 42 74 n.s.
Group 2 49.20 ± 6.53 38 55 52.15 ± 6.0 40.5 62 n.s.
Group 3 53.18 ± 10.46 43 70.28 56.05 ± 6.30 40 63 n.s.
Group 4 65.87 ± 8.30 55.67 74.39 49.15 ± 11.11 25 61 P = .08
Group 5 57.36 ± 3.15 55.15 63 51.15 ± 8.87 39 67 P = .08
Group-6 66.18 ± 2.33 62.29 68.01 58.25 ± 7.40 47.5 70.5 P = .08

Note: Data are presented as mean values, SD, and ranges (minimum and maximum values). In this table, the six implant systems are listed in the order of
mean values obtained by micro-CT. See Table 2 for the order emerging from the histomorphometric results (n.s. = not statistically significant).

TABLE 4 Overview of significant


Group 1 Group 2 Group 3 Group 4 Group 5 Group 6
differences in bone to implant contacts
between the implant systems evaluated Group 1 n.s. n.s. *,** *,** n.s. *,**
by either histomorphometry or Group 2 n.s. n.s. n.s. n.s. n.s. n.s.
microradiography and between the two Group 3 *,** n.s. n.s. ** n.s. **
evaluation methods (ie,
Group 4 *,** n.s. ** *** n.s. *,**
histomorphometry vs microradiography)
for any of the implant systems Group 5 n.s. n.s. n.s. n.s. *** n.s.
Group 6 *,** n.s. ** *,** n.s. ***

*P < .001 between any two implant systems for (total) bone-implant contact as evaluated by histomorphometry.
**P < .001 between any two implant systems for (total) bone-implant contact as evaluated by microradiography.
***P = .08 between histomorphometry vs microradiography for any implant system. (n.s. = not significant).

differences seen with groups 1, 2, and 3 showed lack of statistically period.21-23 As apparent from Table 2, we extended our
significant difference (Mann-Whitney U test/t test; P ≥ .05). Regard- histomorphometric analysis of bone-implant contact from the total
less of whether histomorphometry or microradiography was used, the bone structures to their cortical and cancellous compartments. Note
most favorable bone-implant contacts were consistently observed that an osteo-protective effect on cortical bone can be achieved via
with group 1 and group 6 (Figure 4). increased tension in the cancellous areas and that group 1 revealed
the highest degrees of bone-implant contact even within the confines
of only this compartment, possibly as a result of its progressive and
4 | DISCUSSION non-self-tapping threads24 and load transmission being optimized by
the combination of these progressive threads with a rough surface.
All bovine rib bones used as experimental implant beds in the present Better primary stability can be achieved with tapered than with
study were of uniformly high quality, so that any influence of the cylindrical implants.25,26 Tapered implants have also been shown to
implant beds on the bone-implant contacts may be regarded as negli- offer more favorable load transmission to the bone.20 A slightly
gible. Differences in bone quality between the median and the distal tapered macro-design might, on the other hand, even account for our
end of the ribs were considered in the methodology of this experi- especially poor results of cortical bone contact with group 4 (see
mental protocol. Also, some differences in implant length between the Table 2). Further support for external implant geometry being a modi-
investigated systems was considered and therefore bone-implant con- fier of bone contact is provided by the adverse combination of results
tact percentages were defined in relative terms and implant length we obtained with the cylindrical group 3, which actually showed both
seems not to affect stress transmission to bone.20 Hence, it was rea- the highest cortical and the lowest cancellous bone contact of all six
sonable to attribute the different degrees of bone contact obtained implant systems investigated (see Table 2). Overall, the six implant
for the various implant systems exclusively to their different macro- systems were invariably found to show considerably better cortical
and microscopic design features. than cancellous bone-to-implant contacts.
Our histomorphometric evaluations yielded the highest values of In our microradiographic evaluations, the highest values of bone-
total bone contact for group 1 and group 6. Animal models have previ- implant contact percentages were recorded for groups 1, 4, and
ously revealed that high degrees of bone-implant contact percentages, 6. Hence, this evaluation method supports the notion of external
and hence of primary (mechanical) implant stability, will set the stage macro-geometries like those of groups 1 and 4 being conducive to the
for uneventful bone healing and osseointegration by minimizing establishment of primary bone contacts. All three significant differ-
micromovements inside the bone-implant-interface during the healing ences we found between both evaluation methods (ie, in conjunction
6 ROMANOS ET AL.

AUTHOR CONTRIBU TIONS


Georgios Romanos developed the concept, obtained the funding,
placed implants, and edited the manuscript; Michael Damouras per-
formed the histomorphometric analysis and collected completely the
data; Alexander A. Veis and Frank Schwarz did the histological prepa-
rations, Pablo Hess placed implants and calibrated the other surgeons;
Silvia Brandt prepared the manuscript.

OR CID
Georgios Romanos https://orcid.org/0000-0002-5952-4752
Frank Schwarz https://orcid.org/0000-0002-5873-9903
Silvia Brandt https://orcid.org/0000-0001-5530-0412

RE FE RE NCE S
1. Branemark PI, Hansson BO, Adell R, et al. Osseointegrated implants
in the treatment of the edentulous jaw. Experience from a 10-year
period. Scand J Plast Reconstr Surg Suppl. 1977;16:1-132.
2. Albrektsson TO, Johansson CB, Sennerby L. Biological aspects of
implant dentistry: osseointegration. Periodontol 2000. 1994;4:58-73.
3. Zarb GA, Albrektsson T. Osseointegration—a requiem for the peri-
F I G U R E 4 Box-and-whisker diagram of total bone-implant
odontal ligament?—an editorial. Int J Periodont Restorative Dent. 1991;
contacts (%) as evaluated by microradiography (R) and
11:93.
histomorphometry (H)
4. Smeets R, Stadlinger B, Schwarz F, et al. Impact of dental implant sur-
face modifications on osseointegration. Biomed Res Int. 2016;2016:
6285620.
with group 6, group 4, and group 5) reflected higher values of bone- 5. Meredith N. Assessment of implant stability as a prognostic determi-
nant. Int J Prosthodont. 1998;11:491-501.
to-implant-contact generated by microradiography than by his-
6. Toyoshima T, Tanaka H, Ayukawa Y, et al. Primary stability of a hybrid
tomorphometry (P = .08). Despite these differences, microradiography
implant compared with tapered and cylindrical implants in an ex vivo
remains a comparable and, above all, noninvasive alternative for model. Clin Implant Dent Relat Res. 2015;17:950-956.
determining both existing and potential bone-implant contacts.27 7. Valente ML, de Castro DT, Shimano AC, Lepri CP, dos Reis AC. Ana-
What is more, micro-CT has emerged as a valid alternative for evalu- lyzing the influence of a new dental implant design on primary stabil-
ity. Clin Implant Dent Relat Res. 2016;18:168-173.
ating bone-implant contacts and osseointegration even in direct com-
8. Valente M, Castro DT, Shimano AC, Reis ACD. Influence of an alter-
parison with histological assessment.28 For the purpose of in vitro native implant design and surgical protocol on primary stability. Braz
studies, histomorphometric methods do offer additional insights not Dent J. 2019;30:47-51.
accessible by radiography. 9. Le Guehennec L, Soueidan A, Layrolle P, Amouriq Y. Surface treat-
ments of titanium dental implants for rapid osseointegration. Dent
In summary, the bone-to-implant contacts we observed in the
Mater. 2007;23:844-854.
present study varied due to differences in the macro-design (macro- 10. Turkyilmaz I, Tozum TF, Tumer C. Bone density assessments of oral
geometry) and thread designs of different implant systems. Compared implant sites using computerized tomography. J Oral Rehabil. 2007;
to other thread geometries, progressive threads seem to offer a bone- 34:267-272.
11. Lekholm U, Zarb GA. Patient selection and preparation. In:
condensation effect to the surrounding bone during insertion and can
Branemark PI, Zarb GA, Albrektsson T, eds. Tissue Integrated Prosthe-
significantly help to ensure increased primary stability by maximizing ses: Osseointegration in Clinical Dentistry. Chicago, IL: Quintessence
the bone-implant contacts. This can be evaluated noninvasively by Publishing Company; 1985:199-209.
microradiographic (micro-CT) imaging. The present study concludes 12. Grognard N, Verleye G, Mavreas D, Vande-Vannet B. Updated sec-
ondary implant stability data of two dental implant systems. A retro-
also that the results obtained from histomorphometry are similar to
spective cohort study. J Clin Exp Dent. 2017;9:e1121-e1128.
the less time-consuming microradiography. 13. Albrektsson T, Jacobsson M. Bone-metal interface in
osseointegration. J Prosthet Dent. 1987;57:597-607.
ACKNOWLEDGMENTS 14. Johansson P, Strid KG. Assessment of bone quality from placement
resistance during implant surgery. Int J Oral Maxillofac Implants. 1994;
The authors would like to thank Drs Argyris Samiotis, and Frank
9:279-288.
Spiegelberg for the implant placement of the individual systems 15. Olive J, Aparicio C. Periotest method as a measure of osseointegrated
(Astra, Replace Select, respectively) as well as Mr Bruno Koller oral implant stability. Int J Oral Maxillofac Implants. 1990;5:390-400.
und Mr Markus Burkhart (Scanco Inc.) for the professional 16. Proussaefs P, Lozada J. Use of titanium mesh for staged localized alveo-
microradiographic analyses and the presentation of the photo- lar ridge augmentation: clinical and histologic-histomorphometric evalua-
tion. J Oral Implantol. 2006;32:237-247.
graphic material. Also, special thanks to Mrs Nina Krymchanska
17. Turkyilmaz I, McGlumphy EA. Influence of bone density on implant
(Dental School, Frankfurt) for the statistical analysis of the stability parameters and implant success: a retrospective clinical
collected data. study. BMC Oral Health. 2008;8:32.
ROMANOS ET AL. 7

18. Parsa A, Ibrahim N, Hassan B, van der Stelt P, Wismeijer D. Bone 25. Valente ML, de Castro DT, Shimano AC, Lepri CP, dos Reis AC.
quality evaluation at dental implant site using multislice CT, micro-CT, Analysis of the influence of implant shape on primary stability
and cone beam CT. Clin Oral Implants Res. 2015;26:e1-e7. using the correlation of multiple methods. Clin Oral Investig. 2015;
19. Van Dessel J, Nicolielo LF, Huang Y, et al. Accuracy and reliability of 19:1861-1866.
different cone beam computed tomography (CBCT) devices for struc- 26. O'Sullivan D, Sennerby L, Meredith N. Measurements comparing the
tural analysis of alveolar bone in comparison with multislice CT and initial stability of five designs of dental implants: a human cadaver
micro-CT. Eur J Oral Implantol. 2017;10:95-105. study. Clin Implant Dent Relat Res. 2000;2:85-92.
20. Gehrke SA, Frugis VL, Shibli JA, et al. Influence of implant design 27. Schicho K, Kastner J, Klingesberger R, et al. Surface area analysis of
(cylindrical and conical) in the load transfer surrounding long (13 mm) dental implants using micro-computed tomography. Clin Oral Implants
and short (7 mm) length implants: a photoelastic analysis. Open Dent Res. 2007;18:459-464.
J. 2016;10:522-530. 28. Becker K, Stauber M, Schwarz F, Beissbarth T. Automated 3D-2D
21. Romanos G, Toh CG, Siar CH, et al. Peri-implant bone reactions to registration of X-ray microcomputed tomography with histologi-
immediately loaded implants. An experimental study in monkeys. cal sections for dental implants in bone using chamfer matching
J Periodontol. 2001;72:506-511. and simulated annealing. Comput Med Imaging Graph. 2015;44:
22. Romanos GE, Toh CG, Siar CH, Swaminathan D. Histologic and 62-68.
histomorphometric evaluation of peri-implant bone subjected to
immediate loading: an experimental study with Macaca fascicularis. Int
J Oral Maxillofac Implants. 2002;17:44-51.
23. Romanos GE, Toh CG, Siar CH, Wicht H, Yacoob H, Nentwig GH. How to cite this article: Romanos G, Damouras M, Veis AA,
Bone-implant interface around titanium implants under different Hess P, Schwarz F, Brandt S. Comparison of
loading conditions: a histomorphometrical analysis in the Macaca histomorphometry and microradiography of different implant
fascicularis monkey. J Periodontol. 2003;74:1483-1490.
designs to assess primary implant stability. Clin Implant Dent
24. Rabel A, Kohler SG, Schmidt-Westhausen AM. Clinical study on the
primary stability of two dental implant systems with resonance fre- Relat Res. 2020;1–7. https://doi.org/10.1111/cid.12915
quency analysis. Clin Oral Investig. 2007;11:257-265.

You might also like