YumizenCR
Clinical-Chemistry Reagents
C-Reactive Protein (CRP)
Turbidimetric Immunoassay
INTENDED USE
This reagent is intended for in vitro quantitative determination of C-
reactive protein (CRP) inhuman serum.
CLINICAL SIGNIFICANCE
CC-Reactive Protein (CRP) is an acute phase protein produced by the
Iverin response to lftammation, infection and tissue injury, Increased
CRP concentrations occur much earler than other acute phase
reactants and this rapid resporse to trauma or infection Is the
{istingushing feature of CRP. In alton, CRP tevels return to normal
Quickly a the end ofan acute ep'sode making CRP useful for both the
Aotection of acute episodes as wellasin treatment monitoring
PRINCIPLE
Latex particles coated with specific rabbit anti-human CRP are
Sggiutnated when mixed with samples cortsining CRP. The
‘agglutination causes an absorbance change, dependent upon the CRP
Coren of the patient sample that can be quantified by comparison
{rom calibrators oFknawn CRP concentrations.
KIT COMPONENTS
R1(Divent): Tis butter
2. Ro{Latex Reagent): Latex particles costed with specific rabbt anti
human CRP
3.Caliocator Lyophilized (Cone. as mentioned on vial label)
Reconstitute with + micistlied
‘STORAGE & STABILITY
+ R1QR2& Calibrator
The sealed reagents are stable fr 24 months from manufacture date,
wen stored at 2-8°C,
+ Working Reagent
Stable during 30 days at2-8°C. Shake gant the vial btore use, Products
‘must natbe stored at:oom tomperatueforlengerthan 3Dhours during use
PRECAUTIONS.
1. Ifthe reagents became turbid or the absorbance of blank reagent is.
higher than 1.0000, it means that he reagentis vad and you shoul
iscardit.
2, Donat freeze; rozen Latex or Dluent could change the functionality
ofthetest.
3. Components from human ongin have been tested and found to be
negative for te presence of HBsAg, HCV, and antivody to HIV (14). It
isrocommended handle with eauton.
CALIBRATION
Use GRP Calbrators, The calibration in automated analyzer is stable
for 2 weeks, after which @ new curve must be generatee, Re-calisrate
When contol results are out of spectied tolerances, when using
dlfferentlotofreagentand when theinstrument's adjusted,
PREPARATION
Working reagent: Swit the latex vial genty before use. Prepare the
necessary volume as flow:
BmRI-2miR2
CRP Calibrator (Lyophilized): Reconsttute with 1 mlcistiled water
Seeiente
MATERIAL REQUIRED BUT NOT PROVIDED
‘Thermostatic bath at 37°C.
Spectrophotometar or photometer thermostatable at 37°C witha 640+
2onmiter.
Sample
Fresh Serum (Do nol use lipaemic or haemolysod sample). Stable for 7
days at 2-81. Samples with presence of fin should be centrfuged
betoretesting
PROCEDURE
+ system Parameters
Meda “we Pointed Time
Reaction ‘Ascenaing
Wavelength ‘540820
Blank Distiied water
Sample Volume Opt
Reagent Volume: O00 aL ROO VL RTS BOO ERA)
Delay Time, 10 See
Read Time 720 Sec
Calbraior lated on the val
Normal range Up S mai
Linearity ime Ome
Unit ma
TEST PROCEDURE
1. Bring the working reagent and the photometer (cuvette holder) to
arc.
2, Assay conltons
Wavelength 540 nm (520-560)
Temperature :37°C
Cuvettelight path 1m
3. Adjustthe nstumentto zero wit distiled water.
4. Pipette intoacavette
\Werking Reagent (a) (4R1:1R2): 1000
Calbratorsor sample (ui): 10
5, Mix and read the absorbance immediatly (1) and after 2 mites
(2) ofthe sample addition
‘CALCULATION
‘One point Calibration:
CORP (gil) =
+ Calibrator Concentration
‘QUALITY CONTROL
It is recommended to use Quality Controls to verify the performance of
the assay. Each Laboralory has to establish its own intemal qualty
‘control scheme and procedures fr corective action if controls do not
recover withinthe acceptable tolerance
REFERENCE VALUE
It is recommended that each laboratory should establish its own
reference values. The following value may be used. as. guideline
‘Serum: Normal values up to 8 mg/l. Results obtained for patient
Samples are to be correlated with clinical findings of patient for
interpretation and diagnosis,
batePERFORMANCE
4. Lineaniy limit: Upto 110 mgIL, under tne described assay conditions.
IF the concentration is greater than linearity (110 mgiL) dite
the sample with normal saline ané repeat the assay. Muliply the
resuitwith dation factor.
2. Detection limit: Values less than 2 mg/L give non-reproducile
results,
3.Pro-zone effec: Nopro.zone effect was detected upon 1000 mgt
4 Sonsiviy:A42mA mg.
5 Precision
Intra Run Inter Run
Contral tow [Hien [tow | High
a 20 20 20 | 20
Meanime/t) | 52 | 601 | 53 | 605
sD ou [| o9 | 016 | 10
evaom [27% | 15% | 30% [ 17%
6. Accuracy: Results obtained using this reagent (y) wore compared to
those obtained using 2 commercial roagant (x) with similar
characteristics, 50 samples of diferent concentrations of CRP were
‘assayed. The corelation coefficient (°) was 0.99 and the regression
‘equation was y = 0.968x + 1.197. The results ofthe performance
characterises depend the analyzer used
INTERFERENCE
Blirubin (20 mg/dl), lipemia (10g/L) and rheumatoid factors (300
|UimL) do not interfere. Hemoglobin (> § gil), interferes. Other
substances may interfere,
NoTES.
1. The reagent systemis erin vito use only
21 The volume of reagents ana sample can be adjusted according to
diffrent instruments, While the ratio ofthe reagensample shall be
Manufactured By:
HORIBA India Private Limited
(subsidiary of HORUBA Limited Japan)
Plot No,26, Sector-7, I.E, SIDCUL,
Haridwar-249403, Uttarakhand, India
Toll Free No.: 1800 103 4470
SCRE HORIBA