You are on page 1of 35

Cardiac Rhythm

Device
Management

‘PBL STOP’ – Your acronym for a standardized follow-up


What’s in it for you?
• What do you need to feel comfortable
with a pacemaker/ICD follow-up?

UC200702005 EN
Medtronic, Inc. USA
2 September 2006
CORE OBJECTIVE

• Provide a technique for you to complete a


successful device follow-up

UC200702005 EN
Medtronic, Inc. USA
3 September 2006
“PBL STOP”
P Presenting Rhythm and Rate
B Battery Status
L Lead Status

S Sensing
T Threshold
O Observation, Data, and Events
P Program & Print UC200702005 EN
Medtronic, Inc. USA
4 September 2006
“P”
Presenting Rhythm and Rate

UC200702005 EN
Medtronic, Inc. USA
5 September 2006
Rhythm Recognition

Atrial Fibrillation: No Pacing VVI / 60

Atrial Flutter: No Pacing DDDR / 60

UC200702005 EN
Medtronic, Inc. USA
6 September 2006
Paced Rhythm Recognition

Single Chamber Device: Ventricular Pacing

Single Chamber Device: Atrial Pacing (AAI)

UC200702005 EN
Medtronic, Inc. USA
7 September 2006
Paced Rhythm Recognition

Sensing in the Atrium & Sensing in the Ventricle (AS-VS)

Pacing in the Atrium & Sensing in the Ventricle (AP-VS)

UC200702005 EN
Medtronic, Inc. USA
8 September 2006
Paced Rhythm Recognition

Sensing in the Atrium & Pacing in the Ventricle (AS-VP)

Pacing in the Atrium & Ventricle (AP-VP)

UC200702005 EN
Medtronic, Inc. USA
9 September 2006
Troubleshooting

VVI / 60

Loss of Capture

UC200702005 EN
Medtronic, Inc. USA
10 September 2006
Troubleshooting

Loss of Atrial Sensing

UC200702005 EN
Medtronic, Inc. USA
11 September 2006
“B”
Battery Status
(Volts and Longevity)

UC200702005 EN
Medtronic, Inc. USA
12 September 2006
Assessing Battery Status

UC200702005 EN
Medtronic, Inc. USA
13 September 2006
Assessing Battery Status

UC200702005 EN
Medtronic, Inc. USA
14 September 2006
“L”

Lead Impedance (ohms)

UC200702005 EN
Medtronic, Inc. USA
15 September 2006
Lead Impedance Trends

UC200702005 EN
Medtronic, Inc. USA
16 September 2006
Assessing Lead Status

UC200702005 EN
Medtronic, Inc. USA
17 September 2006
“S”

Sensing (millivolts mV)

The ability of the pacemaker to see


Intrinsic Heart Beats

UC200702005 EN
Medtronic, Inc. USA
18 September 2006
Normal Sensing Values

• Acute Phase (within 3 months of


implant)
– P-waves = > 1.5mV
– R-waves = > 6mV
• Chronic Phase (>3 months post-implant)
– P-waves = > 1.0mV
– R-waves = > 5mV

UC200702005 EN
Medtronic, Inc. USA
19 September 2006
Sensitivity: Not Sensitive Enough
As You Lower The Number,
You Make The Pacemaker More Sensitive

5 mV
Amplitude (mV)

5.0 2 mV

2.5

1mV
1.25

Time UC200702005 EN
Medtronic, Inc. USA
20 September 2006
Sensitivity: Too Sensitive

5 mV

2 mV

1mV
Amplitude (mV)

5.0

2.5

1.25

UC200702005 EN
Time Medtronic, Inc. USA
21 September 2006
Sensitivity: Just Right

5 mV

2 mV

1mV
Amplitude (mV)

5.0

2.5 Why is sensing so


1.25 important?

UC200702005 EN
Time Medtronic, Inc. USA
22 September 2006
Oversensing vs. Undersensing

Scheduled pace
Intrinsic beat delivered
not sensed

Marker Channel™ ...though no


shows intrinsic activity is present
activity...

UC200702005 EN
Medtronic, Inc. USA
23 September 2006
“T”

Threshold

The least amount of energy it takes


to “capture” the heart

UC200702005 EN
Medtronic, Inc. USA
24 September 2006
During Pacing, the Impulse:

Travels via the myocardium,


not the native cardiac
conduction system,
therefore, the QRS is wide

Normal QRS Paced QRS


UC200702005 EN
Medtronic, Inc. USA
25 September 2006
Capture

UC200702005 EN
Medtronic, Inc. USA
26 September 2006
2 Settings Used to Ensure Capture

GOAL  Find the minimum amount of energy needed to


consistently capture the heart (stimulation threshold).

1. Pulse amplitude – the amount of voltage


delivered to the heart by the pacemaker
to ensure capture of the heart (strength
of an output pulse)
2. Pulse width – time (duration) of the
pacing pulse

UC200702005 EN
Medtronic, Inc. USA
27 September 2006
Amplitude Threshold Testing

Keeping pulse width stable, decrement voltage


until loss of capture

@ 0.5ms

1.5V 1.0V .5V

UC200702005 EN
Medtronic, Inc. USA
28 September 2006
“O”
Observations, Data, and Events

UC200702005 EN
Medtronic, Inc. USA
29 September 2006
Diagnostic Information

UC200702005 EN
Medtronic, Inc. USA
30 September 2006
Cardiac Compass®
•Up to 14 months of data
•Annotations to mark Programming
(P), Interrogation (I) or
Remote Interrogations ( _ )
•Rhythm-control assessment with
Treated VT/VF episodes per day and
AT/AF total hours/day
•Rate-control assessment with V.
Rate during AT/AF and V. Rate
during VT/VF
•%Pacing/day to assess reduction of
unnecessary right ventricular pacing
•Three HF Status Indicator Trends
–Average V. Day/Night heart
rates
–Patient Activity
–HR variability

UC200702005 EN
Medtronic, Inc. USA
31 September 2006
“P”
Programming &
Printing Final Reports

UC200702005 EN
Medtronic, Inc. USA
32 September 2006
Programmed Parameters

UC200702005 EN
Medtronic, Inc. USA
33 September 2006
Print Reports

• Print Final Reports


– Sensing test results
– Threshold test results
– Key observations
– Any changes this session noted on final
report

UC200702005 EN
Medtronic, Inc. USA
34 September 2006
“PBL STOP”
P Presenting Rhythm and Rate
B Battery Status
L Lead Status

S Sensing
T Threshold
O Observation, Data, and Events

35
P Program & Print UC200702005 EN
Medtronic, Inc. USA
September 2006

You might also like