You are on page 1of 130
MANUAL PART NUMBER MEP77234-4 ID NUMBERS This manual applies directly to the products with 1D numbers suffixed 122, For additional information on ID numbers, refer to paragraph 1-2 ID NUMBER in Section 1 AUDIO ANALYZER VP-7723A Matsushita Communication Industrial Co., Lid. 4-3-1, Tsunashima-Higashi, Kohoko-ku Yokohama, Japan Tel: (045) 531-1231 Printed in Japan SAFETY PRECAUTIONS GENERAL This instrument has been designed and tested to ensure reasonable personal protection and protection of the surrounding area against damage, and has been supplied in a safe con- dition. ‘The following precautions must be ob- served by the user to ensure safe operation and to retain the instrument in a safe condi- tion BEFORE SUPPLYING POWER Verify that the instrument is set to suit the available mains voltage and that the correct fuse is installed. PROTECTIVE EARTH The protective earth of the instrument must be connected to the earth before connecting the instrument to the product input or output terminals. A protective ground connection by way of the grounding conductor in the power cable is essential for safe operation. The plug shall only .be inserted into a socket outlet provided with a protective earth con- tact. SAFETY SYMBOLS A Instruction manual symbol : The instrument willbe marked with this symbol when it is necessary for the user to refer to the instruc tion manual for safety. 4 High voltage symbol : Dangerous voltage exceed- ing 1kV is indicated by this symbol The WARNING sign is a head- WARNING || ing of the requirement(s) that should be observed to avoid personal or fire hazards. (CAUTION } The CAUTION sign leads the 1 precaution(s) that should be observed to avoid damage or destruction of the instrument. POWER SOURCE This instrument is intended to operate from a mains supply that will not be more than 250 volts rms. For suitable voltage selection, see the INSTAL- LATION paragraph in this manual. HAZARD ARISING FROM LOSS OF GROUND The protective action must not be negated by the use of an extension cord without protec- tive conductor. If this instrument is to be energized via an au- totransformer for voltage reduction make sure the common terminal is connected to the earth terminal of the power source. DAMAGE IN TRANSPORT OR STORAGE Whenever it is likely that protection has been impaired, for example as a result of damage caused by abnormal stresses in transport. or storage, the instrument shall be made inoper- ative and be secured against any unintended operation USE OF PROPER FUSE Use only the fuse of correct type, voltage rat- ing and current rating as specified in the IN- STALLATION paragraph in this manual. REMOVAL OF COVERS Removal of covers is likely to expose live parts although reasonable precautions have been taken in the design of the instrument to shield such parts. The instrument shall be discon- nected from the supply before carrying out any adjustment, replacement or maintenance and repair during which the instrument shall be opened. if any adjustment, maintenance or repair under voltage is inevitable it shall only be carried out by a qualified personnel who is aware of the hazard involved DO NOT OPERATE IN EXPLOSIVE AT- MOSPHERES To avoid explosion, do not operate the instru- ment in an explosive atmosphere. TABLE OF CONTENTS SECTION | GENERAL 1-1 INTRODUCTION 1-2 ID NUMBER 1-3. DESCRIPTION cee dooondd 1-4 SIGNAL GENERATOR SECTION 1-5 MEASUREMENT SECTION ... 1-6 FREQUENCY MEASUREMENT ce 1-7 AC LEVEL MEASUREMENT... 1-8 TOTAL DISTORTION MEASUREMENT 1-9 S/N MEASUREMENT 4-10 DC LEVEL MEASUREMENT ... cece 4-11 WOW & FLUTTER MEASUREMENT (OPTION 01, 02) 1-12 MEASUREMENT FILTERS . 1-13 ADDITIVE FUNCTIONS . 7 1-14 PRESET FUNCTION ... 1-15. LIMIT JUDGMENT FUNCTION 116 EXT CONTROL 1/0 FUNCTION .........cceceeeeseeeee 1-17 MEMORY SYNC AND MEMORY COPY FUNCTIONS. 4-18 FLOATING CONNECTION AND BALANCED INPUT SECTION Il SPECIFICATIONS 2-1 ELECTRICAL PERFORMANCE 2-2. ENVIRONMENTAL CONDITIONS 2-3. MECHANICAL PERFORMANCE ........ 2-4 ACCESSORIES 2-5 OPTION . SECTION III INSTALLATION 3-1 POWER REQUIREMENTS 3-2. MAINS VOLTAGE SELECTION 3:3 FUSE. - 34 POWER CABLE ceseeee ceeeeeeees 35 INTERCONNECTIONS WITH OTHER EQUIPMENT 36 MOUNTING ON THE BENCH . 3-7 BACK MOUNTING 3-8 WEIGHTING FILTERS AVAILABLE AS OPTIONAL ACCESSORIES . 3-9 WOW & FLUTTER MEASUREMENT FUNCTION (OPTION 01, 02) 3-10 BATTERY ..... ae 3-11 OTHERS SECTION IV OPERATION 44 42 43 aa 45 46 47 48 49 4-10 ant 412 4.13 414 415 416 4.17 418 419 4.20 42 422 4:23 424 SECTION V SECTION VI -if- 5-1 5.2 53 54 55 5-6 57 58 et 6-2 63 6-4 65 6-6 67 68 GENERAL, SPECIFIC FUNCTIONS AND TERMS FRONT PANEL REAR PANEL ce SIGNAL SOURCE FREQUENCY ..... SIGNAL SOURCE OUTPUT AMPLITUDE SIGNAL SOURCE OUTPUT ON/OFF SELECTION OF MEASUREMENT FUNCTIONS FREQUENCY MEASUREMENT DISTORTION MEASUREMENT DC LEVEL MEASUREMENT AC LEVEL MEASUREMENT ..... RELATIVE LEVEL INDICATION S/N MEASUREMENT ....... WATT INDICATION ....-2.602cc0seeeeeeeee WOW & FLUTTER MEASUREMENT (OPTION 01, 02) SELECTING INDICATION RESPONSE INDICATION UNITS ........2.5 MEASUREMENT FILTERS BALANCED INPUT AND FLOATING CONNECTION. AUTOMATIC / MANUAL MEASUREMENT UMIT JUDGMENT FUNCTION ASSORTED PRESET MEMORY : AUTO SEQUENCE FOR ASSORTED PRESET MEMORY GP-18 OVERVIEW GENERAL : DEVICES IN THE SYSTEM. MESSAGE PATHS AND BUS STRUCTURE DATA LINES (DATA BUS) . HANDSHAKE LINES (DATA BYTE TRANSFER CONTROL US) CONTROL LINES (INTERFACE MANAGEMENT BUS) ....... MAJOR SPECIFICATIONS OF THE IEEE488, GP-IB CODE ASSIGNMENT OF COMMAND INFORMATION GP-IB INTERFACE GENERAL GP-18 INTERFACE CAPABILITY GPIB ADDRESS SELECTION DEVICE CLEAR FUNCTION FUNCTIONS REMOTELY UNCONTROLLABLE REMOTE / LOCAL SELECTION DEVICE TRIGGER FUNCTION . RESPONSE TO COMMANDS 69 6-10 611 SECTION Vil TA 72 73 74 75 76 17 TB 7-9 7-10 7 712 PROGRAM CODE INPUT FORMAT PROGRAM CODE OUTPUT FORMAT . MEMORY SYNC FUNCTION, MEMORY COPY FUNCTION EXT CONTROL 1/0 GENERAL CONNECTOR PIN ARRANGEMENT AND FUNCTION OF EACH PIN MODE SELECTION OF EXTERNAL CONTROL INTERFACE COMMON ITEMS FOR EXTERNAL CONTROL INTERFACE OPERATIONS REMOTE SEQUENTIAL RECALL REMOTE MODIFY REMOTE DIRECT RECALL LIMIT JUDGMENT OUTPUT CONTROL OUTPUT PRINTING OUT OF MEMORY (LIST OUTPUT) DATA READ DATA PRINT FUNCTION SECTION VIII OPTION AND OPTIONAL ACCESSORIES Bt 82 83 SECTION IX ot 92 93 94 95 96 GENERAL, WOW & FLUTTER MEASUREMENT OPTION. FILTERS AS OPTIONAL ACCESSORIES MAINTENANCE CLEANING JUDGMENT FOR MEMORY BACKUP CALIBRATION OR SERVICE DAILY MAINTENANCE TRANSPORTATION AND STORAGE REMOVAL OF MEMORY BACKUP BATTERY 9-1 9-1 9-1 9-1 9-1 9-1 SECTION I GENERAL 1-1 INTRODUCTION This Instruction manual is composed of the following section (1) Section | GENERAL This section outlines VP-7723A Audio Analyzer. (2) Section I SPECIFICATIONS This section shows the specifications of this instrument in a list form. (3) Section Ill INSTALLATION This section describes electrical and mechanical preparations to be made to use the VP-723A and safety precautions. Be sure to read this section before operating the instrument. (4) Section IV. OPERATION This section describes in detail the func- tions and operating procedures of this in- strument. (5) Section V_ GPIB OVERVIEW This section describes the GP-IB standard. (6) Section VI_ GPIB INTERFACE This section describes in detail the operat ing procedures of this instrument using ‘the GP-B interface. (7) Section Vil EXT CONTROL 1/0 INTERFACE This section describes in detail the functions and operating procedures of the external control 1/0 interface specific to this instrument. (8) Section Vill OPTIONAL FUNCTIONS This section provides information about options. Included are specifications, install- ing procedures, and operation procedures of the optional accessories to be installed in this instrument. (9) Section IX MAINTENANCE This section describes the usual mainten- ance procedures. 1-2. 1D NUMBER This instrument has a ten-character 1D number. The first seven characters are assigned uniquely for each product. The last three figures comprise the ID suffix which is the same for all identical products and changes only when a change is made. All. correspondence with the factory or representatives concerning this instrument should include the complete ten-character ID number. The contents of this manual apply directly to products numbered with the same ID number suffix(es) as noted under the title of 1 NUMBERS on the first page of this manual. 1-3, DESCRIPTION The VP-7723A Audio Analyzer is a measur- ing instrument with functions to measure seven items including AC level, DC level, distortion, and S/N ratio, and has a measuring signal source as shown in Figure 1-1. An audio mea- suring system featuring low noise, a high accu- racy, and an excellent measuring efficiency can be built by combining the signal source and the various measuring functions. As Figure 1-1 shows, this instrument extensively incorporates digital control technique. The VP-723A is de- signed for fully-automatic measurement such as automatic ranging, automatic tuning, and out- put of measured data. The audio analyzer has versatile functions as follows : a preset’ function to preset up to 100 sets of measuring conditions, a limit judg- ment function for PASS/ FAIL judgment of mea- sured results, an auto-sequence function to automatically recall preset memory in sequence, @ printer output function to output measured results to a printer. It also equips the GP-18, and EXT CONTROL 1/0 interfaces as standard facilities Therefore, the VP-7723A Audio Analyzer is used in research and development, as well as in production and inspection lines of audio equip- ment, and also widely applied as a component in automatic measuring system GENERAL EXT CONTROL 1/0 ts Operation and Display Section, Control Section, Memory, Interfaces U Wow’ and Flutter Frequency Counter] | Measurement Section (Option co | Extremely low-distortion noe IRC Oscillator NOTCH HPF LF psoRHO a rons 7AVG VW J “™\ im /\ = | a | 6 6 © © © OSC OUT ACINPUT OC INPUT MONITOR OUTPUT 0¢ OUTPUT Figure 1-1 Configuration of VP-7723A Measurement Section ‘AC level measurement HH frreaueney Frccsrement j [DC level measurement Relative level indication |i $/N measurement UY i n functi i ac oxilator : [WATT indication function |} i [Total distortion }——1—Frequency measurement |} Penton measurement ; OV rms max. JAC level measurement 001% {frequency measurement 'Wow and flutter {measurement (Option) Figure 1-2 Configuration of measurement functions of VP-7723A 1-4 SIGNAL GENERATOR SECTION The instrument provides a built-in extremely low-distortion programmable RC oscillator as signal source for measurement. The oscillator adopts a bridged-T type oscillation system with a frequency range from SHz to 110kHz. The maximum output of this signal Generator is 5Vims across 2 600.2 load (10V rms at an open end). The output is adjustable in 0.1-48 steps by an output attenuator of total 99.9 d8. The output can be set in dBV "1 or dom *2. Either of these output units can be selected. The distortion ratio is 0.001% or less (80 kHz BW) in the range of 20Hz to 1SkHz. 4-5. MEASUREMENT SECTION The measurement section of this instrument has the following basic measurement functions 1) Frequency measurement 2) AC level measurement 3) Total distortion measurement 4) Signal-to-noise ratio (S/N) measurement 5) DC level measurement 6) Wow & flutter measurement (option) The following paragraphs 1-6 to 1-11 de- scribe the general features of each measure- ment function. 1-6 FREQUENCY MEASUREMENT This instrument contains a reciprocal frequency counter for high-speed and high GENERAL resolution measurement of low frequency signals. Input signal periods are measured with an accuracy of 5x 10-5 using the reference time base of 8 MHz (1251s), and are displayed in frequency after performing division calculation by a microprocessor. Frequencies ranging from 5 Hz to 110 kHz can be measured in all the measurement modes except the DC level measurement 4-7, AC LEVEL MEASUREMENT The instrument is used as a high-sensi y that permits a selection of average-responding or rms-responding charac- teristics *3 AC voltmeter The measurement range is divided into seven sub-ranges as follows: 0.316 mV, 3.16 mV, 31.6 mV, 316 mV, 3.16V, 31.6V, and 100V fr. d. (f.r.d, =full range display). At least 10% of over-range is provided for each range except the 100.0V range. The measurement indication units that can be selected are VimV), dBV, and dam. Since internal residual noise is 10 pV or less, the measurable range of this instrument is approximately 30 nV to 100 Vrms (~90 to 40 dBV or 88 to 42 dBm) *4 The range selection can be automatically or manually. made 1 Od8V=1Vims across 600.2 load. “dBV" is read merely as “dB” on the front panel marking of the instrument, "2 0d8m= 1 mW into 6002 load. The units of “dBV" and “dBm” are used to distinguish the voltage and power level from the generally used “d8" unit to express relative ratio such as S/N and distortion. *3. The response characteristics can be selected in AC level, distortion, and S/N measurement functions. The response characteristics of input signal level measurement in distortion and wow & flutter measurements are fixed to rms-responding. Response character ics of wow & flutter measurement are quasi-peak-responding for option 01 and rms-responding for option 02. *4 The specification of internal residual noise is 10 pV or less with 500kHz BW or 4 yV with BOkHz BW. Thus, if the 80kHz LPF contained in this instrument is turned on, the AC level measureable range will be extended down to 10 nV. GENERAL Two additional functions for AC level mea- surement are the relative level measurement and the WATT indication functions. The rela- tive level measurement function indicates a rel- ative level to a reference level in d8. The mea- surable range of this function is £130 dB *1 This function is useful for measurement of fre quency characteristics, level ratio, and S/N ra- tio, The WATT indication function calculates wattage with the following expression using AC level measured value and an assumptive load resistance R.*2, WATT=(AC level measured value)2/R. (1-1) 4-8 TOTAL DISTORTION MEASUREMENT This instrument measures distortion with the fundamental frequency range of 5Hz to 110 kHz as defined in the following = DISTN = Vagivegis Foyt e276, 100%] (1-2) or DISTN = 20log (Vega egi4 .-- ¥enT+ O27) [AB] 1-3) Where ein: rms value of input signal level ey: tms value of the Nth harmoni N= 2,3, @n: rms value of noise contained in the signal This instrument measures frequencies of input signals and automatically tunes the center frequency of the fundamental-rejection filter. Wide and sharp characteristics are obtained due to a multi-stage filter circuit with low noise and low distortion. It enables to measure signals with some frequency fluctuations while providing the measurement capability of distortion factors of 0.001% (-100dB, 80 kHz BW) or less. The measurement is made in five ranges from 0.01% to 31.6% f.1.d. where the automatic ranging is performed. Though the input signal level range in normal distortion measurement is 0.1 to 100 Vrms, the high- sensitive input range of 3.16 mV f.1.d. is also provided to measure the dynamic range of digital audio equipment. For distortion measurement, both input sig nal level and noise / distortion signal level are converted into DC voltage by each detection circuit, and then into digital data by an A/D converter. These two data are ratio-calculated by a microprocessor, to obtain a distortion fac- tor. The reference level setting operation often used in conventional distortion meters is not re- quired. The input signal level obtained in the detection circuit is read on the display together with the measurement result of the distortion, The detection circuit of the input signal level_has rms-responding characteristics while the detection circuit of the noise and distortion components can select either rms- or average- responding characteristics. The bandwidth of the measuring system is 5Hz to 500 kHz. In distortion measurements, input signals containing a large amount of noise are handied, disabling frequency measurement and making automatic ranging operation unstable. The tuning frequency of ‘the fundamental-rejection filter, input signal level range, and distortion factor measurement range can be fixed independently to avoid these effects Figure 1-3 typical distor characteristics when the signal generator and distortion measuring section of this instrument are directly connected. sometimes shows ‘1 The input voltage range that can be handled in the relative level measurement function is also approximately 30 pV to 100 Vrms (-90 to 40dBV or ~88 to 42.d8m). range of the relative level depends on the spe The measurement ied value of the reference level. For example, if the reference level is 10 Vrms (+20dBV), the measurable range of the relative level is varied between +20 and ~ 11048. *2 The assumptive load resistance R. does not mean that a load resistor is actually contained within this instrument. 1-4 The Ri value is specified merely for calculation. GENERAL () ox input rage 3.16V Input level 3.16 UNBAL 0.01 Distortio 0.001 THD +N 1 =A feet 80 kHz BW 0.0001 10 100 1k 10k 100k Frequency (Hz) Figure 1-3 Typical overall distortion characteristics 1-9 S/N MEASUREMENT The S/N ratio can be obtained by calculat ing the ratio between the signal (5) and noise (N) components: The S component is obtained by applying a signal to the object to be mea- sured to measure its output level, and the N component is measured by disconnecting the applied signal and terminating the input termi nal of the object to be measured by a charac- teristic impedance. In the S/N measurement function of this in- strument, ON and OFF of the signal generator output and measurements of the Sand N com- ponent signal levels are automatically synchro- nized so that the measurement can be per- formed by merely pressing the S/N measure- ment key. As measurement results, the $ com- ponent signal level and the S signal frequency are displayed together with S/N ratio simulta neously. The S/N measurement is a measurement of both S and N components in AC level measure- ment and a calculation of the S/N ratio. Thus the $ and N components that can be measured are approximately 30 pV to 100 Vrms as in AC level measurement. (However, S component 2 N component.) Residual noise is 10 pV or less as in AC level measurement. The range of the $/N ratio that can be measured: depends on the S component signal level. For example, S/N ratio of 130 dB or greater can be measured for the S compo- nent signal level of 31.6Vrms. The S/N mea surement range decreases 10 dB for every 10dB reduced in this $ component signal level. The range constructions of $ and N component sig- nal level measurements are the same as that of ‘AC level measurement. The automatic and manual ranging can be performed for both S and N component measurements. The S/N measurement is shown in Figure 1-4 conducted as GENERAL S component N component SLOW/ FAST key 215/36) (RESPONSE block) | t Signal source Signal source Sov | one : 7 slow 3 seconds output ON output OFF 2 “ Measurement starts Figure 1-4 S/N measurement 4-10. DC LEVEL MEASUREMENT This instrument allows to measure the DC voltage. The measurement range is divided into four ranges as follows: 316.0 mV, 3.160 V, 31.60.V, and 100.0V f.r.d. At least 10% of over-range is provided for each range except the 100.0V range. The range selection can be made automatically or manually. 4-11 WOW & FLUTTER MEASUREMENT (OPTION 01, 02) Option 01 has the wow & flutter measure- ment function with the quasi-peak responding characteristics conforming to the IEC, DIN, and EIAJ standards. Option 02 has the function with the rms-responding characteristics conform- ing to JIS. Two measurement center frequen- cies of both option 1 and option 2 are avail- able at 3 and 3.15kHz. Turning on or off pso- phometric weighting is selectable. ‘As measurement results, the frequency and the input level are displayed with the measured value simultaneously The input signal level is 0.1 to 100 Vrms, and the measurement range is divided into three: 10.00%, 1.000%, and 0.1000 %. 4-12 MEASUREMENT FILTERS The eight types of measurement filters are internally provided to insert into the measure- ment system in the AC level, distortion, and $/ N ratio measurements. They are two types of high-pass filters, three types of low-pass filters, and three types of noise evaluation filters. In addition to them, one type of each low-pass fil- ter and noise evaluation filter are available’ as optional accessories. Figure 1-5 shows the char- acteristics of the internally provided filters. GENERAL 8 8), ° HPF roo ae LPF ae 80 ke 200 Hz i 15 ktiz_| Il Ler -20 20 kite -30 40 / =50 10 100 1k 10k 100k Frequency (Ha) (@) Low-pass and high-pass filters (ce) 10 Co 0 Nt aa 7" ri -20 30 40 \ 50 \ 10 100 1k 10k 100k Frequency (he) (b) Noise evaluation filters Figure 1-5 Characteristics of internally provided filters GENERAL 1-13, ADDITIVE FUNCTIONS The following three functions are provided for more convenient use of this instrument in addition to the basic measuring functions : 1) Preset function 2) Limit judgment function 3) EXT CONTROL 1/0 function Each of these functions is outlined below. 4-14 PRESET FUNCTION This function will be useful if measurement setups have been determined. Oscillator fre- quency, output amplitude, measuring function selection, filter selection, and other setups of this instrument can be stored in registers as one set of data. By recalling these registers as required, the setups can be reproduced at a time. One hundred setups can be stored in to- tal ‘The auto sequence function is also provided to automatically recall the memory in sequence at any desired time interval 4-15. LIMIT JUDGMENT FUNCTION In some applications such as production pro- cesses, it is often required to perform PASS / FAIL judgment for certain limit range. Limit judgment function sets upper and lower limit values for each measuring function, and gives warning by turning on “OVER" or "UNDER" LED when measured values exceed the specified limit values. This function can be used more effectively when combined with the preset function, 4-16 EXT CONTROL 1/0 FUNCTION The EXT CONTROL 1/0 connector on the rear panel of this instrument offers the follow- ing functions, 1) Remote sequence recall Remote controls the memory sequence re- call function externally. 2) Remote modify Remote controls the modifications of signal source frequency and the signal source out- put amplitude by an external rotary encod- er. 1-8 3) Remote direct recall Remote controls the memory direct recall function externally. 4) Limit judgment output Gives external LED lighting output to repre- sent the limit judgment result. 5) Control output Gives the TTL output signal of 8 bits x 2 ports to control external equipment. 6) Printing output for memory (list output) Outputs the contents of preset memory to a printer. 7) Data read Reads externally applied &-bit TTL input sig- nal with the GP-IB controller. 8) Data print Outputs measured values to a printer. 4-17 MEMORY SYNC AND MEMORY COPY FUNCTIONS The memory sync function is provided to re- call preset memory contents of several instru- ments simultaneously in the talk-only / listen- only mode of the GP-IB interface. The memory copy function allows this in- strument to transfer the contents of memory to another VP-7723A. 1-18 FLOATING CONNECTION AND BALANCED INPUT The common lines of the signal source and the measuring section can be separately floated from the frame of this instrument to avoid the problem of ground-loop caused by interconnec- tion of devices during measurement of low lev- el signals or low distortion. nes are marked as 7 and YY on the panel to distinguish from the frame ground (L). The common line v of the measuring section can be connected directly to the frame ground with a switch The balanced input is also available to match objects to be measured with floating output terminals such as BTL amplifiers. These common [i- ExT GPIB CONTROLI/O OC = MONITOR Control section toa} S100) Optoisolator Measuring LN} section osc BAL GENERAL ATT HOA att H—©) 8 i pc INPUT Figure 1-6 Common line configurs 1-9 SECTION II SPECIFICATIONS NOTE1) The specifications have been given in the condition of the automatic operation or properly set manual operation. NOTE2) The dB unit to express a voltage level (0 dB=1 Vrms) is described dBV throughout this instruction manual 2-41 ELECTRICAL PERFORMANCE SIGNAL SOURCE Designation Performance Conditions or remarks QUENCY, Display 4 digits Frequency range and resolution 5Hz to 10kHz in 4 ranges 5.0 to 200.9Hz — [0.201 to 2.009Hz |2.01 to 20.09kHz |20.1 to 110.0kHz O.1Hz resolution |1Hz resolution —_|10Hz resolution | 100 Hz resolution Accuracy 3% of reading SHz to 110kHz 2% of reading 0.201 to 20.09 kHz QUTPUT AMPLITUDE Display 3digits and a minus sign Units d8m (0dBm : 1 mW into 600.2) ABV (0dBV: 1Vrms) Range and resolution | Variable from 16.2 d8m to -83.7 |The voltage in dBV is defined as dBm (14.0 dBV to ~85.9dBV) in 0.1 |the matched ‘output (terminated dB steps. by 2 6002 load). Accuracy +£0.58 of reading at 1 kHz > -37.10d8V +£0.8d8 of reading at 1kHz 5 -37.2d8V Flatness With reference to 1 kH2; With a load resistor of 6000 £0308 SH2 to 110kHz 40.05 48 20.0Hz to 20.09 kHz Noise voltage with OSC switch OFF 5 10pVems DISTORTION 5 0.01% (—80d8) 5Hz to 110kHz 0.001 % (~ 10048) 20Hz to 1SkHz, 80 kHz BW SPECIFICATIONS MEASUREMENT Designation Performance Conditions or remarks ‘Measurement Functions Frequency measurement AC level measurement including : Relative level measurement function, and WATT indication function DC level measurement S/N measurement Total distortion measurement Wow & flutter measurement Option 01 FREQUENCY MEASUREMENT Designation Performance Conditions or remarks Measurement range Display resolution Input level range Accuracy Measuring system SHz to 110 kHz Frequency = 100 Hz: S-digit display Frequency <100 Hz: 0.01 Hz resolution 30 mV to 100Vrms 1 mV to 3.16 mV rms £5x10- £1 digit Reciprocal counter Operates in AC level, distortion and wow & flutter measurement modes. At the input range of 3.16 mV in distortion measurement AC LEVEL: MEASUREMENT range display input; 210% 25% Designation Performance Conditions or remarks Full range display T ranges Unit: (mv | Unitzd8V | Unit:d8m 100.0 40.0d8V | 42.2d8m |Note: 31.60 30.0dev | 32.2dam | At least 10% of over-range 3.160 wooaav | 122d8m | is provided for 3160mv | -100dBv | -7.8dam | each range ex- 31.60mv | -30.0dav | -27.8dam | ePt the 100.0 V range 3.160mv | -s0.0dav | -478dam 03160mv | -70.0dav | -67.8d8m Accuracy £2% of full range display tka Flatness With reference to 1 kHz and full SHz to 110kHz 20 Hz to 20kHz SPECIFICATIONS AC LEVEL MEASUREMENT (Cont'd) Designation Performance Conditions or remarks Residual noise Relative level measurement range Response WATT indication function Measuring method Display resolution <10nVems <4yVims £1308 or more ‘Average or RMS responding Calculates wattage with measured AC level and specified assumptive load resistance value (Rv. 5 digits at most, 500 kHz BW 80 kHz BW The range will be limited de- pending on the reference value selected. No actual load is contained. o.o1w OLAX watts DC LEVEL MEASUREMENT Designation Performance Conditions or remarks Full range display 4 ranges At least 10% of over-range is 100.0 provided for each range except 31.60 the 100.0 range. 3.160V 316.0 mV Accuracy + (0.3% of full range display +0.75 % of reading) S/N MEASUREMENT Designation Performance Conditions or remarks 5 and N levels S/N measurement range 30 pV to 100 Vrms for either $ or N 0 to 13008 The following table shows the limitation for S/N measurement range depending on the $ component level. S>N S: signal, N: Noise $ component level Limitation for S/N (frequency $ 10 kHz) measurement 2 31.6V (30dBV) 130.48 = 3.16V (10dBV) 110d 2 316mv (- 10d8V) 90 d8 2 31.6 mv (-30¢d8V) 70d8 2 3.16mv (-50d8V) 5008 2 0.316 mv (~70¢BV) 3048 SPECIFICATIONS. S/N MEASUREMENT (Cont'd) 5 component level accuracy SIN accuracy Response SIN delay time £2% of full range display 2108 ‘Average- or RMS-responding Approximately 1.55 for fast response or 35 for slow response Designation Performance Conditions or remarks Unit S component level | (m)V, dB, d8m SIN ratio oe tkHe Range composition is the same as that in AC level measurement function TOTAL DISTORTION MEASUREMENT Designation Performance Conditions or remarks Fundamental frequency range Full range display and Display resolution Display of measurement units Response Second harmonic accuracy Residual noise and distortion 5.0 Hz to 110.0 kHz 5 ranges 31.60 % (~ 10.00.48) 10.00% (20.00 da) 1.000% (40.00 d8) 0.1000 % (~ 60.00 dB) 0.01000 % (~ 80.00 d8) Selection of unit key |v, % dB Input signal level | mv, V [dB*, dam dB :0dB=1Vrms Distortion % dae }+* dB : Distortion (ratio) Input signal level : RMS-responding Distortion signal level : RMS- or average-responding £308 £108 SHz to 110kHz 20 Hz to 20.09 kHz Input range 100, 31.6, 10, 3.16, 1 o3i6v Detection Input tevel [Full range display input] 173 of full range Full range display input] 173 of ful range | bandwidth display input display input Hz to 20kHz fs 10008, 0.001% 9008, 0.0032% |< -9 3,0002% |= -e5ae, 0.006% [a0kHz ew Hato okiil< -e0g8, 001% [x -eods,o0r% |e -7 soo16% [x 740, 002% — fookna a 2-4 SPECIFICATIONS TOTAL DISTORTION MEASUREMENT (Cont'd) Designation Performance Conditions or remarks Input signal level Range Full range display Input range 3.16 mV: S ~45d8, 0.56% for 2mVrms input 0.1 to 100V rms 1 to 3.16Vims 7 ranges 100.0 (40.0 dBV, 42.2 dBm) 31.6 V (30.0 dBV, 32.2d8m) 10.0V (20.0dBV, 22.2d8m) 3.16 V (10.0 dBV, 12.2 48m) 1.00 (0dBV, 2.2d8m) 0.316 V (~ 10.0d8V, ~7.8d8m) 3.16 mV (~50.0dBV, ~ 47.8 d8m) Fundamental frequency range: 20Hz to 10kHz Detection bandwidth : 20 kHz 8W As shown in residual noise and distortion specifications, the dis- tortion measurement range will be limited depending on the in- put signal level Bandwidth of 3.16 mV range is 20kHz BW. it is not available for automatic ranging function, Accuracy £2% of full range display tkHe Flatness With reference to 1kHz and with full range display input; 5% from SHz to 110kHz Excluding 3.16 mV and 100V ranges £10% from 10Hz to 20kHz At 3.16 mV range WOW & FLUTTER MEASUREMENT (OPTION) Designation Performance Conditions or remarks Measurement center frequencies Full range display and Display resolution Response Flatness. WTD UNwTD Accuracy Input signal level range 3kHz, 3.15kHz +200Hz 3 ranges 10.00%, 1.000%, 0.1000 % Quasi-peak responding (OPT 01), Rms-responding (OPT 02) Psophometric weighting characteris. tics in accordance with DIN 45507 O.SHz to 300Hz £5% of full range display Same as input signal level range of distortion measurement SPECIFICATIONS COMMON FOR ALL MEASUREMENT FUNCTIONS Designation Performance Conditions or remarks input impedance Maximum allowable input voltage Filters 100 Hz HPF 200 Hz HPF 15kHz LPF 20 kHz LPF 80 kHz LPF PSOPHO A PSOPHO CCIR ARM PSOPHO AUDIO 100 kQ shunted by 200 pF or less ima Fig. 2-1 shows the maximum allowa- ble input voltage for AC component only. 10V to 100V ranges: 150 (DC + AC peak) 3.16 or lower ranges: 150\V (DC + AC peak) for AC com- ponent below 17 kHz. See Fig. 2-1 for higher frequencies. 348 cutoff frequency 75 + 1SHz Rolloff > 40d8 at 25 Hz ~3.dB cutoff frequency 180 + 25Hz Rolloft 608 / decade Pass band £18 in S15kHz Attenuation >30dB at 2 19kHz Pass band ‘£148 in S20kHz Attenuation >30dB at 224.1 kHz ~38 cutoff frequency 80 + 10kHz Rolloff 60dB/decade A characteristics of IEC recommenda- tion ARM characteristics of CCIR A or 8 terminal to common line DC terminal to common line vov - ‘and higher ranges Ss 3.16V “Sand lower ranges + Input vottage (peek) > Frequeney (#2) Fig.2-1 Maximum allowable input voltage for AC component aniy For rejection of pilot component in measuring AM stereo. Ninth Chebyshev filter Ninth Chebyshev filter AUDIO characteristics of DIN 45405 See Fig. 1-6. SPECIFICATIONS COMMON FOR ALL MEASUREMENT FUNCTIONS (Cont'd) Designation Performance Conditions or remarks Monitor output and DC output [Monitor output voltage] OC output voltage | Output Measurement | Monitor output onan Aone poo ‘AC LEVEL [Provides scaled |Approx. 0.2 Vrms, levellApprox. —25 V, level presentation oflindicator at a fullfndicator at a full input signal. |range display. ange display. SIN Noise compo-|Approx. 0.2 V rms, NlApprox. -25V, N level nent onl level at a full rangelat a full range display. a aisplay, a 19° GPIY. | a a5 04 DISTN Noise and dis-JApprox. 0.2V rms, bothlApprox. -2.5 V, both tortion compo-|input level and distor-fnput level and distor- nents. hion signal level ation signal level at| leach full range display. leach full range display. oc tevet [None lapprox. £25 V, level indicator at a full range display. WOW & FLUTTER [None Approx. 25 V, level in- ———— flicator ata full range lisptay. OTHERS Designation Performance Conditions or remarks PRESET function Number of memory registers Items to be stored ata time 100 1) 2 Frequency of the signal source Output amplitude of the signal source (including OFF) Measurement mode Input range Measurement range Frequency of fundamental rejec tion filter Selection of filter 3) 4) 5) 6) 7?) 8) 9) Indication response assignment Units for two measurement dis- plays, % or V/dB 10) Operation mode, automatic or manual 11) Upper and/or lower limits for the LIMIT judgment function SPECIFICATIONS OTHERS (Cont'd) Designation Performance Conditions or remarks MODIFY function LIMIT function INTERFACE Remote control 12) BAL or UNBAL input selection 13) Data of output ports P1 and P2 in EXT CONTROL 1/0 14) Reference data in relative level measurement 15) Load resistance (Rt) data in WATT indication function. 16) Wow & flutter measurement center frequency 1) Modifies signal source frequency and output amplitude. 2) In manual operation - Modifies input level range + Modifies measurement range OVER and UNDER lights indicate the comparison result of the measured value to the predetermined upper and Jor lower limits GP-I8, EXT CONTROL 1/0. GP-IB: SH1, AHI, T7, 13, SRO, RLI, PPO, DCI, DTI, CO + Copy function of preset data in talk only or listen only mode, + Memory syne function in talk only or listen only mode. EXT CONTROL 1/0: + Sequence memory recall opera- ion Direct memory recall operation + Remote control of modify function + External control output Bbits x 2 ports = Reading of external data Bbits x 1 ports + Printing out of preset memory contents and measured value. SPECIFICATIONS OTHERS (Cont'd) Designation Performance Conditions or remarks Power requirements Voltage 100V: 90V to 112 120V: 106V to 132V 220V: 196V to 24a 20V: 214V to 250 Frequency 50 or 60Hz Power consumption | 60VA or less 2-2_ ENVIRONMENTAL CONDITIONS Designation Performance Conditions or remarks Ambient temperature and humidity 0° to 40°C, RH 20 to 90% 10° to 35°C, RH 20 to 85% =20* to 70°C, RH 20 to 90% Limit range of operation Limit range for guaranteed per- formance Storage and transportation 2:3. MECHANICAL PERFORMANCE Designation Performance Conditions or remarks Dimensions Mass 426 mm wide, 132mm high, and 400 mm deep Approx. 13kg Excluding knob, connectors, han- dle and feet 2-4 ACCESSORIES Designation Performance Conditions or remarks Power cable a Spare fuse eee 1 GP-IB connector cap... 1 Instruction manual ..... 1 2-5 OPTION Designation Performance Conditions or remarks Option 01 Wow & flutter measurement func- | Ordering model number tion (quasi-peak-responding) VP-723A01 Option 02 Wow & flutter measurement func- | Ordering model number tion (rms-responding) VP-7723A02 Optional accessories Measurement filters CCITT P53 TEL MESSAGE 1 kHz BPF 3kHz BPF IEC-C See Section Vill for details. SECTION III INSTALLATION 3-1 POWER REQUIREMENTS The model VP-7723A can be operated from any power source supplying 100 V, 120V, 220V and 240V (nominal values), 50 or 60Hz. Pow- er consumption is less than 60 VA. WARNING Before connecting AC power to the instru: ment, be sure it is set for the proper mains voltage and is properly fused as indicated on the rear panel. (See Figure 3-1) 3-2 MAINS VOLTAGE SELECTION Refer to Figure 3-1. Set the NOMINAL VOLTAGE switches to the setting (100 V, 120V, 220V or 240.V) that corresponds to the mains voltage to be used. The voltage must be with- in the range noted above the fuse holder. 3:3 FUSE Verify the proper fuse is installed in the fuse holder. Ratings of the fuse are noted on the rear panel and listed below, Voltage selection switches (shown set for 220V) NOMINAL VOLTAGE /[ioov 120 Nominal voltage Fuse s00V 250 120 1A(T) 220 250 240 0.54 (T) z20v }/)- 240V NOMINAL 100 RANGE 90-112 120 106-132 220v_ | 196-2aav 240 214-250 Figure 3-1 Voltage selection [warn Make sure that only fuses with the re- quired rated current and of the specified type are used for replacement. The use of make-shift fuses and short-circuiting of fuse holders are prohibited. INSTALLATION 3-4 POWER CABLE The VP-7723A is equipped with a detach- able power cable assembly. The type of the plug shipped with each instrument depends on the country of destination. Figure 3-2 illustrates four types of power cables available. To order a power cable, include the instru- ment model number, instrument ID number, and the cable type shown in Figure 3-2. Ad- dress the order to the dealer or representative from which you purchased the instrument. 125 250V OPERATION OPERATION A-3P type C-3P type 250 250 OPERATION OPERATION AS-3P type BS-3P type Figure 3-2. Power cables 3-5 INTERCONNECTIONS WITH OTHER EQUIPMENT Plug the power cable into a properly wired receptacle before connecting the instrument to another equipment. The interconnections are made with input / output coaxial connectors and measurement earth terminal on the front panel; the output coaxial connectors of MONITOR and -DC OUTPUT, GP-IB connector, and EXT CONTROL 1/ © connector on the rear panel. The common line of the signal generator section (outer metal shell of the OSC OUTPUT connector indicated with VV) is floated from the frame ground. The common line of the measurement section (outer metal shells of three INPUT connectors) can be either connected directly to the frame ground or floated from the frame with the switch on the panel. The floated common line indicated with Y is isolated from the common line of the signal generator section WY. The common lines of MONITOR and DC OUTPUT are separately floated and each of which is connected to the measurement section common line W/ through a 109 resistor. This is for avoiding problems caused by common ground loop. They cannot be used in floating connection and should be regarded as directly grounded line in actual applications. The common lines of GP-IB connector and EXT CONTROL 1/0 connector are directly con- nected to the frame ground. These connectors have no dangerous terminal to touch. Connect them only to the control devices meeting the specifications of this instrument. See Sections V through VII of this manual. When connecting EXT CONTROL 1/0 con- nector to a printer in the memory list output function, use the dedicated cable VQ-023H10. Otherwise it may cause malfunctions or trou- bles. 3-6 MOUNTING ON THE BENCH The instrument cabinet has plastic feet and a fold-away tilt stand. The tilt stand raises the INSTALLATION front of the instrument for easier operation of the front panel controls Stacking with other instruments may be al- lowed only when it dose not cause degradation of the performance due to interference such as vibration, electromagnetic induction or tem- perature rise. The instrument is equipped with @ cooling fan located on the rear panel. When operating the instrument, choose a location that provides adequate air flow. 3-7 RACK MOUNTING The VP-7723A may be rack mounted by using the Rack Mount Kit VQ-069H31. It will fit 480 mm wide rack conforming to JIS C6010. 3-8 WEIGHTING FILTERS AVAILABLE AS OPTIONAL ACCESSORIES Up to two optional filters for psophometric measurement and noise suppression can be installed in the audio analyzer. For any detail on the characteristics of each filter, see Section Vill. If the user desires to have it installed, contact the dealer or repre- sentative from which you purchased the instru- ment. 3-9 WOW & FLUTTER MEASUREMENT FUNCTION (Option 01, Option 02) Installation of wow & flutter measurement function is specified as “Option 01” or “Option 02", The model number for ordering is VP- 7723001 or VP-7723A02. Option 01 is of quasi-peak-responding type conforming to DIN, CCIR, IEEE, and EIAJ. Op- tion 02 is of rms-responding type conforming to JIS (apanese Industrial Standard). 3-10 BATTERY The memory back-up battery built in this in- strument is a rechargeable nickel-cadmium cell and is always being recharged when the instru- ment is turned on. Normally charged battery will back-up the memory for several weeks, however, a long term of turning off may cause insufficient back-up due to gradual discharge of the battery. 3-3 INSTALLATION Keep the instrument turned on for more than eight hours when the instrument is used for the first time after unpacking or after a long term of storage period The battery should be treated as follows (1) The battery life is five years or longer un- der normal operating environment. After the life is over, replacement will be needed to avoid insufficient back-up. Contact the dealer or representative from which you purchased the product in such case. (2) Do not remove or short-circuit the battery and never throw it into a fire. | ATTENTION 1. iin the USA only) + The product contains a rechargeable battery. The battery is recyclable. At the end of its useful life, under various state and local laws, it may be illegal to dispose of this battery into the municipal waste stream, Check with your local solid waste officials for details in your area for recycling options or proper disposal. If removal of the battery is required, contact the dealer or representative from which you have purchased the product. Removal of the battery should be performed only by a qualified person. 3-11 OTHERS (1) Ambient temperature The instrument can be operated within the temperature range of 0°C to 40°C. For entirely-quaranteed performance, use the in- strument in the range of 10°C to 35°C. 2) Warm-up Allow a warm-up period of at least fif- teen minutes before using the instrument for measurement. SECTION IV OPERATION 4-1 GENERAL This section describes basic operations performed on the front and rear panels of the VP-7723A. In addition to internal signal source and various measurement operations, this instrument has the Preset memory function. To interface with external devices, the GP-I8 and EXT CONTROL 1/0 interfaces are provided. First, this section outlines the functions and terms specific to this instrument. Next, the section describes the overall operation panels briefly, and then gives detailed explanation for each operation in the following part. The GP-IB program codes associated with each operation are also explained in the corresponding part. 4-5. Signal source frequency 4-6 Signal source output amplitude 4-7 Signal source output on/off 4-8 Selection of measurement func 4-9 Frequency measurement 4-10 Distortion measurement 4-11 DC level measurement 4-12 AC level measurement 4-13 Relative level measurement 4-14 S/N measurement 4-15 WATT indication 4-16 Wow & flutter measurement (option 01) 4-17 Indication response 4-18 Indication units 4-19 Measurement filters 4-20 Balanced input and floating connection 4-21 Automatic/manual measurement 4-22 Limit judgment function 4-23 Assorted preset memory 4-24 Auto sequence for assorted preset memory The GP-IB interface is described in SECTION V and VI, and the EXT CONTROL 1/0 interface in SECTION VII The list of GP-IB program codes is attached to the end of tl manual. 4-2. SPECIFIC FUNCTIONS AND TERMS (1) Assorted preset memory The desired setup of this instrument can be stored in memory as one set of data, and recalled at a time as required. Changes can be made freely after recalling each setup. One hundred setups can be stored in total. See paragraph 4-23 for details. Q) Auto sequence This function is used to automatically recall assorted preset memory in sequence at any desired time interval. See paragraph 4-24 for details. OPERATION (3) Automatic mode/ manual mode This instrument permits both automatic and manual measurement. In the automatic mode, selection of suitable range and tuning of the fundamental-rejection filter will be made automatically if the signal to be measured is applied to this instrument. Then the measurement result will be displayed. In the manual mode, the range and the fundamental-rejection filter can be fixed during measurement. Detailed information is given in each measurement function. NOTE This instrument is battery backed up and the setup conditions are retained even after the power is turned off. 4-3. FRONT PANEL The panel illustration is located on a foldout page at the rear of this manual. Parts on the panels that are related to the operation of this instrument are numbered from (#) to @@, and the numbers are also used in the following in describing the operation of the panels, The names of parts on the front panel and outline of their functions are described below. @ POWER switch A pushbutton switch to turn mains on and off. @ ADRESS readout Normally, the address of assorted preset memory is displayed in 00 to 99, but the GP-IB address is dis- played when specifying the GP-IB address. LIMIT indicator The limit judgment result is displayed. FREQUENCY / AMPLITUDE readout Normally, the frequency measured value is displayed. When specifying the signal source frequency, signal source output amplitude, 1/0 MODE, and so on, the corresponding specified value is displayed. © INPUT LEVEL readout Normally, the input signal level, reference value for relative level measurement, or S$ component level in S 7/N measurement is displayed. When specifying the in- ©oO put range, the reference value and others in the man- ual mode, the corresponding specified value is dis- played. @© MEASUREMENT readout Normally, the resultant measured value is displayed, When specifying limit values in the limit judgment function and the measurement range or others in the manual mode, the corresponding specified value is dis- played. @ BAL/UNBAL key The key to select balanced or unbalanced connection of the AC INPUT terminals. @® AC INPUT connectors The input terminals for all signals to be measured ex- cept DC level measurement. @® DC INPUT connector The input terminal for the signal to be measured in DC level measurement. @ Floating/ grounding switch The switch to select floating or frame-ground connec tion of the common line of the input terminals. 6 6868 86 es @ @ @ 88 8 8 @eee8 UNIT key RESPONSE block Filter block MEASUREMENT block MODIFY block ENTER block AUTO key DATA block Setting function selection block MEMORY block SHIFT key LOCAL key REMOTE light OSC ON/OFF key Earth terminal OSC OUTPUT connector REAR PANEL Filter unit mounting section W&F unit mounting section EXT CONTROL 1/0 connector GP-IB connector NOMINAL VOLTAGE selector MAINS INPUT connector Fuse holder OPERATION The key to select the dB system (log) or the V, % sys- ‘tem (linear) as indication unit. The keys to select the detection response. The keys to specify the measurement filter. The keys to select the measuring function. The rotary knob to modify each specified value and the key to select the digit to be modified. The keys to register each specified value into memory. The value can be registered only when the light is flashing. The key to select the automatic or manual mode. The numerical keys to enter each specified value. Combining with the SHIFT key @, these keys can se- lect setting function. The keys to select setting function among signal sour- ce frequency, signal source output amplitude, storing / recalling of preset memory, limit values in limit judg ment function, interval time, 1/0 MODE, and others Operational keys for assorted preset memory The key to switch the dual-function keys from normal function to “shifted” function, The key to switch the instrument from the GP-IB re- mote control state to the local state. The light turned on when the instrument is in the GP- 1B remote control state. The key to select to turn on or off the signal source output signal. Used with SHIFT key @, the key causes FREQUENCY / AMPLITUDE readout ® to display set val- ves of signal source frequency and output level. ‘A measuring earth terminal directly connected to the instrument frame. Signal source output terminal Up to two optional filters can be installed here. The standard type of the analyzer has a plate covering the section. The optional W & F measuring unit can be installed. The standard type of the analyzer has a plate covering the section. ‘A 36-pin connector used for input / output of external control signal, remote control of preset memory or MODIFY knob, and other operations. A 24-pin connector for GP-IB connection. The mains power voltage selector switches. An inlet socket to connect the power cable. ‘A fuse holder to insert the power fuse. OPERATION ® 0¢ OUTPUT connector The output terminal for DC signal proportional to the output level of the MONITOR connector ®. 2.5V or =25V at full-range display. ® MONITOR connector The AC output of 0.2 Vrms at full-range display can be obtained in proportion to the input signal level in the AC level measurement, to the N component in $/ N measurement, and to the noise and distortion com- ponent in the distortion measurement. No signal can be obtained in neither DC [evel nor wow & flutter measurement. ® Auxito4 Auxiliary connector installation section. 4-5. SIGNAL SOURCE FREQUENCY (1) General The frequency of the signal source contained in this instrument can be directly specified by DATA keys and modified by the rotary knob. The setting ranges and resolutions of the signal source frequency are as follows: 5.0 to 200.9Hz/0.1 Hz 0.201 to 2.009 kHz/1 Hz 2.01 to 20.09 kHz / 10 Hz 20.1 to 110.0 kHz/ 100 Hz (2) Direct setting by DATA keys Press FREQ key of Setting function selection block @, and FREQUENCY / AMPLITUDE (called FREQ/AMPTD) readout @ shows the frequency specified up to now. Enter the desired frequen- cy with any keys of DATA block @. ENTER block keys ® start flashing. While they are flash- ing, press kHz or Hz key to set the frequency. Example 4-1) Setting signal source frequency to 1.234 kHz ENTER FREQ 1 2 3 4 © kHz After setting operation, FREQ/AMPTD readout @ will show the newly specified value for two seconds. Then the readout will return to the state before FREQ key was pressed, To retain any set value in FREQ/AMPTD readout @, press SHIFT key @ and then OSC ON/OFF key @. To set the readout back to show frequency measurement values, press the keys similarly once again. (3) Modification by the rotary knob Press FREQ key of Setting function selection block @ and FREQ/AMPTD readout @ will show the frequency specified up to now. Press DIGIT SELECTOR key of MODIFY block ®. One digit of the readout starts flashing. Shift the flashing to the place to be modified with DIGIT SELECTOR key. The numeral on the place can be increased or decreased by turning the rotary knob clockwise or counterclockwise Example 4-2) Modifying signal source frequency DIGIT SELECTOR DIGIT SELECTOR ,~. FREQ oe O After the rotary knob operation, FREQ/AMPTD readout ® will show the newly specified value for two seconds. Then the readout will return to the state before FREQ key was pressed, OPERATION (4) GP-1B program code The signal source frequency can be controlled by GP-IB. Table 4-1 GP-18 program code for signal source frequency setting Header code | Data code Unit code Description FR 5.0 to 110000 Hz, Sets frequency at 5.0Hz [0.0050 to 110.0 Kz. to 110.0 kHz 4-6 SIGNAL SOURCE OUTPUT AMPLITUDE (1) General The output amplitude of the signal source contained in this instrument can be directly specified by DATA keys and modified by the rotary knob. The setting ranges and resolutions of the signal source output amplitude are as follows : ~ 85.9 to 14.04BV/0.1d8 — dBV: OdBV=1V ims across 600. load. Display on the panel is dB. ~ 83.7 to 16.2dBm/0.1.d8m dBm: 0d8m=1mW into 6002 load. (2) Direct setting by DATA keys Press AMPTD key of Setting function selection block ® and FREQ/AMPTD readout @ will show the output amplitude specified up to now. Enter the desired amplitude with any keys of DATA block ®. ENTER block keys ® start flashing. While they are flashing, press d8 or dm key to set the output amplitude. Example 4-3) Setting signal source output amplitude to ~5.6 dBV. ENTER a) Ge) Bs : oe ce After setting operation, FREQ/AMPTD readout ® will show the newly specified value for ‘two seconds. Then the readout will return to the state before AMPTD key was pressed. To retain any set value in FREQ/AMPTD readout @, press SHIFT key @ and then ON OSC/ OFF key @. To set the readout back to show frequency measurement values, press the keys similarly once again. Modification by the rotary knob Press AMPTD key of Setting function selection block @ and FREQ/AMPTD readout ® will show the output amplitude specified up to now. Press DIGIT SELECTOR key of MODIFY block @. One digit of the readout starts flashing. Shift the flashing to the place to be modified with DIGIT SELECTOR key. The numeral on the place can be increased or decreased by turning the rotary knob clockwise or counterclockwise. Example 4-4) Modifying signal source output amplitude @ DIGIT SELECTOR DIGIT SELECTOR a we 6 After the rotary knob operation, FREQ/AMPTD readout ® will show the newly specified val- ue for two seconds. Then the readout will return to the state before AMPTD key was pressed. OPERATION (4) GP-18 program code The signal source output amplitude can be controlled by GP-IB. Table 4-2 GP-IB program code for signal source output amplitude setting Header code | Data code Unit code Description AP - 85.9 to 14.0 ba Setting output amplitude at -85.9 to 14.0dBV 83.7 to 16.2 DM Setting output amplitude at - 83.7 to 16.2d8m 4-7, SIGNAL SOURCE OUTPUT ON/OFF (1) General The signal source contained in this instrument can be turned on or off by OSC ON/OFF key The residual noise with signal source output OFF is - 110 dB or less with reference to 14 BV at 110kHz. Figure 4-1 shows the configuration of the signal source. 6009 \M\ 600.2 ATT ‘ON osc ouTPUT Figure 4-1 Configuration of the signal source (2) ON/OFF operation Pressing OSC ON/OFF key @ allows to select ON (light on) or OFF (light off) alternately. Example 4-5) ON/OFF operation of signal source output osc + ON #0FF ° (3) GP-1B program code Signal source output ON/OFF can be controlled by GP-IB. Table 4-3 GP-IB program code for signal source output ON/OFF operation Header code | Data code | Unit code Description ap oN Signal source output ON OFF Signal source output OFF OPERATION 4-8 SELECTION OF MEASUREMENT FUNCTIONS (1) Operation Eight types of measurement functions are available for this instrument and they can be se- lected by MEASUREMENT block keys @. Table 4-4 shows the relationship between the status of each key and the measurement function. Table 4-4 Selection of measurement functions Status of each MEASUREMENT block key puneony g [OISTN/]DC LEVEL] AC LEVEL | RELATIVE LEVEL/ SEM | WATE | AEF Distortion -| 0 - = - DC level S\| 5 ° = a AC level -| - - ° = Relative level a - ° ° SyN o| ° = - a WATT o| - ° = - Wow-flutter without psophometric weighting }O | — = ° - Wow-flutter with psophumetic weighting | O] = = ° o (OQ: ON, ~: OFF) (2) GP-IB program code The selection of measurement functions can be controlled by GP-IB. Table 4-5 GP-IB program code for function selection Header code| Data code | Unit code Description MM. 1 Distortion 2 OC level 3 AC level 4 SIN 5 WATT indication in AC level measurement 6 Wow & flutter RR 0 Relative level indication OFF in AC level measurement 1 Relative level indication ON in AC level measurement wr 0 Psophometric weighting OFF in W&F measurement 1 Psophometric weighting ON in W & F measurement OPERATION. 4-9 FREQUENCY MEASUREMENT The analyzer has a built-in reciprocal counter to measure input frequency. The measurement is always performed except in DC level measurement mode, and the measurement result is displayed on FREQ/ AMPTD readout(@. Measurable input signal level ranges from 30mV rms to 100 Vrms or 1 to 3.16mV rms (only in distortion measurement, input range of 3.16 mV). Resolution is 0.01 Hz for frequency of 5.00 to 99.99 Hz, and five digits for that of 100.00 Hz to 110.00 kHz. When setting signal source frequency, signal source output amplitude, 1/0 mode, and others, FREQ/AMPTD readout @) shows the corresponding specified value. After each se the display will return to the state displaying the input frequency measured 1g operation 4-10. DISTORTION MEASUREMENT (1) General This instrument can measure the total distortion factor of an input signal by lighting only DISTN key of MEASUREMENT block @. The definitions are as follows. DISTN= (Ve? + eg? +. en? + Gn?/ein) x 100 [96] an or DISTN= 2010g( Vez? + @2 +. -ey? + @ Pein) (C8) (4-2) Where ein: Input signal level en: Nth harmonic level, N en: Contained noise level ‘The fundamental frequency range is 5.0Hz to 110.0kHz. The input level ranges are 1 mV rms (~60.dBV, ~57.8d8m) to 3.16 mV rms (—50dBV, ~47.8.d8m) and 0.1 Vrms (~20dBV, - 17.8 d8m) to 100 Vrms (40 dBV, 42.2. d8m). The measurement range is divided into five ranges as follows: 31.6% (~10d8), 10% (-20 48), 1% (~ 40.48), 0.1% (~60dB), and 0.01 % (~80¢B). (2) Display and unit ‘The measurement value in distortion measurement is read as follows : FREQ/ AMPTD readout @) . Measured frequency value INPUT LEVEL readout@ ............ Measured input level MEASUREMENT readout@ .. Measured distortion factor The indication units for the input level and the distortion are-determined by the combina tion of the setting of UNIT key @ and the setting unit of the signal source output amplitude. iy Table 4-6 Unit of input level and distortion values UNIT key | Signal source Indication unit setting setting unit | Input level | Distortion v.% av v, mv % 68m sav av 8 4B 8m dam Note that d8V is indicated on the panel as d8. 4-8 OPERATION The input level is measured with rms-responding characteristics. The response characteristics Of distortion measurement can be selected by AVG/ RMS key of RESPONSE block (to average- responding (AVG) or rms-responding (RMS). When the input signal is below 100 Hz, set SLOW/ FAST key of RESPONSE block @ to SLOW to minimize measurement error. (3) Automatic measurement When an input signal within the measurement range is applied to AC INPUT connectors @) with AUTO key @ set to AUTO (light on), the suitable input level range, fundamental-rejection filter, and measurement range are automatically selected, and then measured value appears on the readout. In the input signal level range of 1 to 3.16mV rms, however, only the manual measurement described in (4) is available. (4) Manual measurement In the manual mode in the total distortion measurement, the input level range, measure- ment range, and the frequency of the fundamental-rejection filter can be selected manually. (@) Fixing input level range Set AUTO key @ to MANUAL (light off) and press INPUT RANGE key (SHIFT key @) and then “1" key of DATA block @). ENTER keys @ start flashing and INPUT LEVEL readout G) shows the current input level range. Enter any range code given in Table 4-7 by DATA keys @ while ENTER keys @ are flashing. Press either of ENTER keys @ to fix the selected input level range. Table 4-7 Input level range code in distortion measurement Input level range Code v, mv auv aim | dBV:incicated as ¢8 on the panel 0 Automatic ranging (no indication) 1 100V 40 42.2 2 31.6V 30 32.2 3 10v 20 22.2 4 3.16V 10 122 5 iv ° 22 6 0.316V -10 =7.8_| Note that only the manual measure- ment is available in the 3.16 mV 7 3.16 mv =50 -478 | eve Example 4-6) Fixing input level range to 3.16V AUTO SHIFT. INPUT ENTER ° ° 7 a 3° (light off) OPERATION If either of ENTER keys @ is directly pressed without any range code, the current range is fixed. After fixing input level range, INPUT LEVEL readout @ will show the newly fixed input evel range for two seconds. Then it will return to the state displaying the measured value of the input level. Alternatively, select the input level range with the rotary knob after pressing INPUT RANGE key. Then press either of ENTER keys @ to fix the selected input level range. Example 4-7) Fixing input level range by the rotary knob AUTO SHIFT. INPUT ENTER ° ° 1 oO ° (light off) Rotate to indicate desired range. (b) Fixing measurement range Set AUTO key @ to MANUAL (light off) and press MEAS RANGE key (SHIFT key @) and then "2" key of DATA block @). ENTER keys @ start flashing and MEASUREMENT readout © shows the current measurement range. Enter any range code given in Table 48 by DATA keys @ while ENTER keys @ are flashing. Press either of ENTER keys @ to fix the selected measurement range. Table 4-8 Measurement range code in distortion measurement Measurement range Code % dB 0 Automatic ranging (No indication) 1 316 =10 2 10 -20 3 1 -40 4 O41 -60 5 0.01 ~80 Example 4-8) Setting measurement range to 0.1% AuTO SHIFT MEAS ENTER ° ° 2 4 ° (ight off) If either of ENTER keys @ is directly pressed without any range code, the current range is fixed. After fixing measurement range, MEASUREMENT readout @) will show the newly fixed measurement range for two seconds. Then it will return to the state displaying the mea- sured value of the distortion factor. Alternatively, select the measurement range with the rotary knob after pressing MEAS RANGE key. Press either of ENTER keys @ to fix the selected measurement range. 4-10 OPERATION Example 4-9) Setting measurement range with the rotary knob AUTO SHIFT = MEAS ee ENTER oul ea cou! & (ual) 1@ 4 linc (ight off) Rotate to indicate desired range. {c) Fixing fundamental-rejection filter frequency Set AUTO key @ to MANUAL (light off ) and press NOTCH key (SHIFT key @ and then “0" key of DATA block @@). ENTER keys @ start flashing and FREQ/AMPTD readout @) shows the current fundamental-rejection filter tuning frequency. Enter desired tuning fre- quency given in Table 4-9 by DATA keys @ while ENTER keys (@ are flashing. Press either of ENTER keys @ to fix the newly specified frequency. Table 4-9 Fundamental-rejection filter tuning frequency in distortion measurement Data Unit for ENTER key Tuning frequency 0| kHz or Hz Automatic tuning (No indication) 5.0 to 200.9 | Hz 5.0 to 200.9 Hz 0.201 to 2.009 | kHz 0.201 to 2.009 kHz 2.01 to 20.09 2:01 to 20,09 kHz, 20.1 to 110.0 20.1 to 110.0 kHz Example 4-10) Setting fundamental-rejection filter tuning frequency to 1.234 kHz AUTO SHIFT NOTCH ENTER ° [2 0 1 : 2 3 4 © | kHz (light off If either of ENTER keys @ is directly pressed without any frequency data, the current tuning frequency is fixed. After setting the fundamental-rejection filter frequency, FREQ/AMPTD readout @ will show the newly fixed tuning frequency for two seconds. Then it will return to the state displaying the measured frequency value. The rotary knob cannot be used to set the tuning frequency. (5) GP-1B program code The GP-IB controlled operations in the distortion measurement are selection of the measure- ment function, and setting of the input level range, the measurement range, and the fundamental-rejection filter frequency in the manual measurement. OPERATION Table 4-10 GP-IB program code in the distortion measurement Header code] Data code _|unit code Description vm |i Iselects distortion measurement function mo oo elects automatic tuning for fundamental-rejection filter 0.5.0 to 0.110000 | Hz [sets fundamental-rejection filter at 5.0 Hz to 110.0 KHz 10.0.0050 to 0.11.0] KZ _fsame as above 1.0 lselects automatic ranging for input level range 1 sets input level range at 100 (408, 42.2. 48m) 12 sets input level range at 31.6V (30dB, 32.2d8m) 13 sets input level range at 10V (2048, 22.2d8m) 14 lsets input level range at 3.16 (10d8, 12.2d8m) 1s fsets input level range at 1 (0d8, 2.2 dam) 1.6 lsets input level range at 0.316V (-10d8, -7.8d8m) Ww ets input level range at 3.16mV (-50d8, ~47.8d8m) 2.0 selects automatic ranging for measurement range 24 Jsets measurement range at 31.6% (- 1048) 22 Isets measurement range at 10% (-20¢8) 23 jets measurement range at 1% (~40¢d8) 2A sets measurement range at 0.1% (- 60d) 25 [sets measurement range at 0.01% (~80d8) 4-11 DC LEVEL MEASUREMENT (1) General By lighting only DC LEVEL key of MEASUREMENT block @, this instrument can measure the DC level of the input signal supplied to DC INPUT connector(3. The measurement ranges and the resolutions are as follows: 100V/0.1V 31.6V/ 10 mv 3.A6V/T mv 316mV/0.1 mv (2) Display and unit The measured value in the DC level measurement appears on MEASUREMENT readout G) The indication units are only V and mV. Note that the frequency measurement is not available in this function. (3) Automatic measurement Set AUTO key @ to AUTO (light on). If a signal within the measurement range is applied to the DC INPUT connector(9, the suitable range is automatically selected and the measurement value appears on the readout. (4) Manual measurement The measurement range can be fixed in the manual DC level measurement. These oper- ational procedures are shown below. 4-12 OPERATION (2) Fixing the measurement range Set AUTO key @ to MANUAL (light off) and press MEAS RANGE key (SHIFT key @) and then “2" key of DATA block @). ENTER keys @ start flashing and MEASUREMENT read- ‘out @ shows the current measurement range. Enter any range code given in Table 4-11 by DATA keys @) while ENTER keys @ are flashing. Press either of ENTER keys @@ to fix the selected measurement range. Table 4-11 Measurement range code in DC level measurement Code Measurement range 0 Auto ranging (No indication) 1 i00V 2 31.6V 3 316 4 316 mV Example 4-11) Fixing measurement range to 3.16V AUTO SHIFT = MEAS ENTER ° ° 2 3 ° (light off If either of ENTER keys @ is directly pressed without any range code, the current range is fixed, After setting of measurement range, MEASUREMENT readout @) will show the newly fixed measurement range for two seconds. Then it will return to the state displaying the measured value of the DC level Alternatively, select the measurement range with the rotary knob after pressing MEAS RANGE key. Press either of ENTER keys @ to fix the selected measurement range. Example 4-12) Setting measurement range by the rotary knob AUTO SHIFT = MEAS yy ENTER eo] bl bP] O fe (ight off) Rotate to indicate desired range (5) GP-IB program code The GP-I8 controlled operations in the DC level measurement are selection of the measure- ment function and setting of the measurement range in the manual measurement. OPERATION Table 4-12 GP-IB program code in DC level measurement Header code | Data code Unit code Description mm 2 Selects DC level measurement function wp 2.0 Selects auto ranging for measurement range 24 Sets measurement range at 100V 22 Sets measurement range at 31.6V 23 Sets measurement range at 3.16V 2.4 Sets measurement range at 316 mV 4-12 AC LEVEL MEASUREMENT (1) General By lighting only AC LEVEL key of MEASUREMENT block ®, this instrument can measure the AC level of the input signal supplied to AC INPUT connectors ©. The measurement frequency range is 5.0Hz to 110.0kHz, and the input level range is ap- proximately 30 Vrms (~90dBV, -88d8m) to 100V rms (40 dBV, 428m) *1 (2) Display and unit The measurement values obtained in the AC level measurement can be displayed as follows: FREQ/AMPTD readout @ : Measured frequency MEASUREMENT readout © : AC level The indication unit can be selected by the combination of setting of UNIT key @ and the specified unit of the signal source output amplitude as shown in Table 4-13. Table 4-13 Indication unit in the AC level measurement Setting of UNIT key | Specified unit of signal source Indication unit of AC level v% dav Vv, mv 8m dB dev dav dam dam Note that dBV is displayed as dB on the panel. The response characteristics of AC level measurement can be selected by AVG/ RMS key of RESPONSE block ® to average-responding (AVG) or rms-responding (RMS). When the input sig- nal is below 100Hz, set SLOW/FAST key of RESPONSE block ® to SLOW to minimize measure- ment error. (3) Automatic measurement When a signal within the measurement range is applied to AC INPUT terminals ® with AUTO key ® set to AUTO (light on), the suitable range is automatically selected and the mea- sured value appears on the readout. ‘*TrInput level range at SO0kHz BW. In case of 80 kHz BW with 80 kHz LPF ON, the input level range changes to approximately 10 zV to 100 Vrms since the internal residual noise becomes 4 pV or less. OPERATION (4) Manual measurement The measurement range can be fixed in the manual AC level measurement. These opera- tion procedures are shown below. (2) Fixing measurement range Set AUTO key @ to MANUAL (light off) and press MEAS RANGE key (SHIFT key @) and then "2" key of DATA block @). ENTER keys © start flashing and MEASUREMENT read- ‘out © shows the current measurement range. Enter any range code given in Table 4-14 by DATA keys @ while ENTER keys © are flashing. Press either of ENTER keys @ to fix the selected measurement range. Table 4-14 Measurement range code in AC level measurement Measurement range Code ve mv dgv dam 0 Automatic ranging (No indication) 1 100 40 42.2 2 31.6V 30 322 3 3.16V 10 12.2 4 316 mv -10 -78 5 31.6 mv ~30 -278 6 3.16 mV ~50 -47.8 7 0.316mv -70 -678 Example 4-13) Fixing measurement range to 3.16V AUTO SHIFT. = MEAS ENTER ° ° 2 3 ° (light off) Hf either of ENTER keys @ is directly pressed without any range code, the current range is fixed After setting of measurement range, MEASUREMENT readout ©) will show the newly fixed measurement range for two seconds. Then it will return to the state displaying the measured value of the AC level Alternatively, select the measurement range with the rotary knob after pressing MEAS RANGE key. Press either of ENTER keys © to fix the selected measurement range. Example 4-14) Setting measurement range by the rotary knob auto. SHIFT MEAS eTeR ol fb] 2] O fe tight ot Rotate to indicate desired range. OPERATION (5) GP-1B program code The GP-IB controlled operations in the AC level measurement are selection of the measure- ment function and setting of the measurement range in the manual measurement. Table 4-15 GP-IB program code in AC level measurement eader code] Data code | Unit code | Description MM 3 |ssetects AC level measurement function Mo 20 Selects auto ranging for measurement range 24 Sets measurement range at 100 V (4048, 42.2d8m) 22 Sets measurement range at 31.6V (30d8, 32.2d8m) 23 Sets measurement range at 3.16V (1048, 12.2 dm) 24 Sets measurement range at 316 mV (~10dB, -7.8d8m) 25 Sets measurement range at 31.6mV (~30d8, ~27.8d8m) 26 Sets measurement range at 3.16 mV (~50d8, -47.8d8m) 27 Sets measurement range at 0.316 mV (~70d8, -67.8d8m) 4-13. RELATIVE LEVEL INDICATION (1) General In the AC level measurement (only AC LEVEL key of MEASUREMENT block @ is turned on), the analyzer can indicate the relative input level against the reference value by operating RELA TIVE LEVEL key of MEASUREMENT block ®. The indication range/ resolution is #130d8/0.01 dB. The reference value ranges from 0.01 mV rms (~99.99 d8V, -97.77 d8m) to 100 Vrms (40 dBV, 42.22 dBm). (2) Display and unit The measured values obtained in the relative level measurement are indicated as follows. FREQ/AMPTD readout ®: Measured frequency INPUT LEVEL readout @: Reference value MEASUREMENT readout © : Relative value The indication units of the reference and relative values in the relative level measurement can be selected by the combination of setting of UNIT key @ and the specified unit of the sig- nal source output amplitude as shown in Table 4-16. Table 4-16 Indication unit for Relative level Setting of Specified unit of Indication unit UNIT key signal source | Reference value | Relative value Vv % dav Vv, mv eB dam ae av dav dam dam Note that dBV is displayed as dB on the panel OPERATION The response characteristics of the relative level measurement can be selected by AVG/ RMS key of RESPONSE block @ to average-responding (AVG) or rms-responding (RMS). When the in- put signal is below 100 Hz, set SLOW/FAST key of RESPONSE block @® to SLOW to minimize measurement error. (3) Setting the reference level The following two methods are av measurement. ble for setting the reference level in the relative level + Defines the measured AC level as the reference level ¥ Sets the desired reference level in numerals These setting procedures are described below. (2) Defining the measured AC level as the reference level Press RELATIVE LEVEL key of MEASUREMENT block @ under AC level measurement con- dition, and the measured AC level appears on INPUT LEVEL readout @. The level is re- tained as the reference value, The relative level of the input signal with respect to the ref- erence value will be displayed on MEASUREMENT readout ©. (©) Setting the desired reference value in numerals Under relative level measurement condition, set AUTO key ® to MANUAL (light off) and press REF LEVEL key (SHIFT key @ and then "3" key of DATA block @®). ENTER keys ® start flashing. Enter any value by DATA keys ® while ENTER keys ® are flashing. Press ei- ‘ther of ENTER keys ® to fix the newly specified reference level. Example 4-15) Setting the reference level to 1.234V in relative level measurement AUTO SHIFT LEVEL ENTER ©] Ce] C3 eS (light off) The specified unit of the reference level depends on the status of UNIT key @. When UNIT key @ is V % (light off), the unit of the reference level is either V or mY. When Unit key © is dB (light on), on the other hand, the unit is either dV or d3m. The reference level specified in the manual measurement still remains even if AUTO key © is set to AUTO. (4) Automatic measurement When a signal within the measurement range is applied to AC INPUT terminals ® with AUTO key @ set to AUTO (light on), the suitable range is automatically selected and the rela- tive level appears on the readout. (5) Manual measurement In the manual mode in the relative level indication, the measurement range can be fixed. The measurement range fixing procedure is the same as that for fixing the measurement range in the AC level measurement. See paragraph 4-12 (4) (a). However, the measurement range should be selected according to the AC level of the input signal, independent of the ref- erence value or the relative value. OPERATION (6) Relationship between preset memory and relative level indication (a) General When the relative level indication function is used in the preset memory operation dis- cussed in paragraph 4-23, the operation of this instrument after recalling varies with the condition in storing. The relationship between the condition in storing and the operation after recalling is explained in the following examples. (b) Relative level indication with the predetermined reference level Table 4-17 Example of relative level indication in preset memory operation (1) Memory] Storing condition address | signal source Indication after recalling Measuring section 00 | 1kHz, -40dBV [AC level measurement | Output amplitude of signal to be measured 01 | 1kHz, -40dBV [Relative level indication | Level difference from 2V lReference level 2V fixed 02 | 20Hz, -40dBV [same as above same as above 03 [20kHz, ~40dBV|same as above Same as above a ~ Storing condition : Indication after recalling : The reference level is set in numerals (as discussed in the preceding paragraph (3) (b)). ‘The AC lovel of the input signal is measured and indicated as the relati /e level with the reference level specified in storing. (c) Relative level indication with the reference level given by measurement. Table 4-18 Example of relative level indication in preset memory operation (2) =a Storing condition address Signal source Indication after recalling Measuring section 00 |1kHz, ~40dBV {AC level measurement | Output amplitude of signal to be measured 01 | 1kHz, ~40.dBV |Relative level indication Measured value at address 00 as reference Reference level : automatic |level 02 |20Hz, -40.d8v same as above Level difference from the reference level (measured value at address 00) 03 )20kHz, -40dBV|same as above Same as above Storing condition : Indication after recalling : Automatic condition to define the measured AC level as the refer- ence level (as discussed in the preceding paragraph (3) (a)). Note that it is required to select AC level measurement function in the address one step prior to the starting of the relative level indica- tion. ‘The latest AC level measured value is defined as the reference lev- el. After that, the relative level with this reference level is indi- cated. OPERATION (7) GPIB program code The GP-I8 controlled operations in the relative level indication are selection of measurement function, setting of the reference value, and fixing of the measurement range in the manual measurement. (As for measurement range fixing pracedures, see Table 4-15 in paragraph 4-12 (5) Table 4-19 GP-I8 program code in relative level indication Header code Data code Unit code Description RR 0 Releases relative level measurement function 1 Bielects relative level measurement funciton mo 3.0.01 to 3.100000 Mv Sets reference level at 104:V to 100V 3.0.00001 to 3.10.0 v same as above }3.-99.99 to 3.40.00 DB sets reference level at ~99.99 to 40.00d8V 3.-97.77 to 3.42.22 bm sets reference level at ~ 97.77 to 42.22d8m 4-14 S/N MEASUREMENT (1) General By lighting $ and S/N keys of MEASUREMENT block @, this instrument can measure the $/ N ratio while synchronizing the AC level measurement section with the signal source according to the procedure shown in Figure 4-2. The definition for $/N measurement is given in Expression (4-3). S component N component SLOW/ FAST key 513 (8) (RESPONSE block) t Signal source] Signal source FAST 45 seconds output ON output OFF SLOW eieeconae Measurement starts Figure 4-2 $ /N measurement procedure SIN =201og (5 component /N component) [48] (3) The input frequency range is 5.0 Hz to 110.0kHz, and the input level range is approximately 30 pV rms (~90dBV, ~88d8m) to 100 Vrms (40.dBV, 42 d8m) *1 “Ts Input level range at SO0kHz BW. In case of 80kHz BW with 80 kHz LPF ON, the input level range changes to approximately 10 pV to 100 Vrms since the internal residual noise becomes 4 BV or less. The $ component signal level must be larger than the N component signal level. 4-19 OPERATION (2) Display and unit ‘The measurement values obtained in the $/N measurement can be displayed as follows: FREQ/AMPTD readout): $ component signal frequency INPUT LEVEL readout@): _$ component signal level MEASUREMENT readout@: S/N measured value The indication unit for $/N measurement value is dB, while that for the S component signal level can be selected by the combination of setting of UNIT key (@) and the specified unit of ‘the signal source output amplitude as shown in Table 4-20. Table 4-20 Indication unit in the $/N measurement Setting of | Specified unit of UNIT key signal source |5 component] S/N Vv % dev v, mv 8 dam dB dav dav dam dam Note that dBV is displayed as d8 on the panel. ‘The response characteristics of the $ component signal level and S/N measurements can be selected by AVG/ RMS key of RESPONSE block (@ to average-responding (AVG) or rms- responding (RMS). When $ or N signal component extends below 100 Hz or, set SLOW / FAST key of RESPONSE block (2) to SLOW to minimize measurement error. (3) Automatic measurement Set AUTO key @ to AUTO (light on), and start the $/N measurement. When a signal within the measurement range is applied to AC INPUT connectors(®) the suitable $ component signal level range is automatically selected and the S component signal level is measured. Then the signal source output is turned off after a certain period. Then the N component signal level range is automatically selected and the N component signal level is measured. Finally the measured values of the S and N component signal levels are calculated to obtain the S/N measured value. The result appears on MEASUREMENT readout) (4) Manual measurement The measurement ranges of the S and N component signal levels can be fixed in the manual measurement in S/N measurement. These operation procedures are shown below. (a) Fixing measurement range of the $ component signal level Set AUTO key @ to MANUAL (light off) and press INPUT RANGE key (SHIFT key @) and then "1 key of DATA block @). ENTER keys (@) start flashing and INPUT LEVEL readout @shows the current measurement range of the S component signal level. Enter any range code given in Table 4-21 by DATA keys @ while ENTER keys ( are flashing. Press either of ENTER keys ( to fix the selected measurement range of the $ component signal level. OPERATION Table 4-21 Measurement range code of S component in $/N measurement Measurement range Code v, mv dev 8m o Automatic ranging (No indication) 1 100 40 422 2 31.6V 30 322 3 3.16V 10 122 4 316mv -10 -78 5 31.6mv ~30 -278 6 3.16mv 50 -478 7 =a 0316 mv -70 ~678 Example 4-16) Fixing measurement range of the $ component to 3.16V AUTO ‘SHIFT INPUT ENTER, ° ° 1] G 3 (ight off) If either of ENTER keys (@ is directly pressed without any range code, the current range is fixed, After fixing of the measurement range of the $ component signal level, INPUT LEVEL readout © will show the newly fixed measurement range of the $ component signal level for two seconds. Then it will return to the state displaying the measured value of the 5 component signal level. Alternatively, select the measurement range of the $ component signal level with the rotary knob after pressing INPUT RANGE key. Press either of ENTER keys @ to fix the measurement range of the $ component signal level. Example 4-17) Setting the measurement range of the S component signal level by the rotary knob AUTO SHIFT INPUT ENTER eo] Pb] Oo] O Ce (light off) Rotate to indicate desired range (b) Fixing the measurement range of the N component signal level Set AUTO key @ to MANUAL (light off) and press MEAS RANGE key (SHIFT key @) and then "2" key of DATA block @ ). ENTER keys start flashing and MEASUREMENT readout © shows the current measurement range of the N component signal level. Enter any range code given in Table 4-22 by DATA keys @ while ENTER keys @ are flashing, Press either of ENTER keys @ to fix the selected measurement range of the N component signal level. Note that the measurement range of the N component signal level should be 4-24 OPERATION selected according to the N component signal level, independent of the S/N value to be measured. Table 4-22 Measurement range code of N component in $//N measurement Measurement range Code Vv, mv aBv dam 0 Automatic ranging (No indication) 1 100 40 42.2 2 31.6V 30 32.2 3 3.16V 10 122 4 316mv 10 -78 5 31.6mV -30 -278 6 3.16 mV -50 -478 0.316 mV -70 -67.8 Example 4-18) Fixing measurement range of the N component to 3.16mV AUTO SHIFT. = MEAS ENTER ° ° 2 6 ° (ight off) If either of ENTER keys @ is directly pressed without any range code, the current range is fixed. After fixing of the measurement range of the N component signal level, MEASUREMENT readout © will show the newly fixed measurement range of the N component signal level for two seconds. Then it will return to the state displaying the measured S/N value. Alternatively, select the measurement range of the N component signal level with the rotary knob after pressing MEAS RANGE key. Press either of ENTER keys @ to fix the measurement range. Example 4-19) Setting the measurement range of the N component signal level by the rotary knob Mie ° 3 2 © 3 (light off 1 Rotate to indicate desired range. (5) GP-IB program code The GP-I8 controlled operations in the $/N measurement are selection of the measurement function and setting of the measurement ranges of both S and N component signal levels in the manual measurement. OPERATION Table 4-23 GP-IB program code in S/N measurement Header | Data | Unit ‘ti code | code | code eee mM 4 Selects $/N measurement function mo | 1.0 Selects auto ranging for S component measurement range a Sets S:component measurement range at 100V/(40.d8, 42.2d8m) 12 Sets 5 component measurement range at 31.6 (308, 32.2d8m) 13 Sets S component measurement range at 3.16V(10dB, 12.2dBm) 14 Sets S component measurement range at 316mV(~10d8, ~7.8dam) 15 Sets $ component measurement range at 31.6mV(~30d8, ~27.8d8m) 1.6 Sets S component measurement range at 3.16mV(~50d8, ~47.8 dam) 17 Sets S component measurement range at 0.316 mV (-70dB, -67.8d8m) 20 Selects auto ranging for N component measurement range 24 Sets N component measurement range at 100V (40 dB, 42.2 dam) 22 Sets N component measurement range at 31.6V(30d8, 32.2d8m) 23 Sets N component measurement range at 3.16V(10d8, 12.2 dm) 24 Sets N component measurement range at 316mV(~ 1048, ~7.84Bm) 25 Sets N component measurement range at 31.6 mV (~30¢8, ~27.8d8m) 2.6 Sets N component measurement range at 3.16mV(~50d8, -47.8d8m) 27 [Sets N component measurement range at 0.316mV(~70d8, -67.8d8m) 4-15 WATT INDICATION (1) General By lighting the $ and WATT keys of MEASUREMENT block ®, this instrument can perform the WATT indication. The AC level of the input signal supplied to AC INPUT connectors ® is measured, and calculated based on a settable assumptive load resistance to convert to the wattage indication defined in the expression (4-4), WATT = e,2/ Ri {W] (4-4) €in ¢ Input signal level The indication range/ resolution is 0.00 to 999.99 W/0.01W. The setting range/ resolution of the assumptive load resistance is 2 to 5000.0/10. Note that the assumptive load resistance is the value only for calculation use, not an actual load installed in this equipment internally. (2) Display and unit The measurement values obtained in the WATT indication can be displayed as follows: FREQ/AMPTD readout ®: Measured frequency MEASUREMENT readout © : WATT value corresponding to AC level The specified value of the assumptive load resistance appears on FREQ/AMPTD readout ® only in the setting operation discussed in paragraph (4) (b)). The response characteristics of AC level measurement in WATT indication can be selected by 4-23 OPERATION AVG/ RMS key of RESPONSE block @) to average-responding (AVG) or rms-responding (RMS). When the input signal is below 100Hz, set SLOW/FAST key of RESPONSE block @ to SLOW to minimize measurement error. (3) Automatic measurement When a signal within the measurement range is applied to AC INPUT connectors @ with AUTO key @ set to AUTO (light on), the suitable range is automatically selected and the measurement value appears on the readout. (4) Manual measurement In the manual measurement in WATT indication, the AC level measurement range can be fixed and the assumptive load resistance can be set. These operation procedures are shown below. (@) Fixing the AC level measurement range Set AUTO key @ to MANUAL (light off) and press MEAS RANGE key (SHIFT key @) and then "2" key of DATA block @ ). ENTER keys @ start flashing and MEASUREMENT readout @) shows the current measurement range. Enter any range code given in Table 4- 24 by DATA keys @ while ENTER keys @ are flashing. Press either of ENTER keys @ to fix. However, the measurement range of AC level should be selected according to the input level, independent of the WATT value to be indicated Table 4-24 Measurement range code of AC level in WATT indication Measurement range Code v, mv dev dBm 0 ‘Automatic ranging (No indication) 1 100 40 422 2 31.6V 30 322 3 3.16V 10 12.2 4 316mv =10 -78 5 31.6 mV -30 -278 6 3.16 mV -50 -478 7 0.316 mV. -70 -678 Example 4-20) Fixing AC level measurement range to 3.16V AUTO SHIFT MEAS ENTER ° ° 2 3 ° (light off) If either of ENTER keys © is directly pressed without any range code, the current range is fixed, After fixing of the AC level measurement range, MEASUREMENT readout @) will show the newly fixed measurement range for two seconds. Then it will return to the state displaying the WATT value. Alternatively, select the measurement range with the rotary knob after pressing MEAS RANGE key. Press either of ENTER keys @ to fix the measurement range. OPERATION Example 4-21) Setting the AC level measurement range by the rotary knob AUTO SHIFT MEAS ENTER eo] fb] fl O (light off) A Rotate to indicate desired range. (b) Setting assumptive load resistance Set AUTO key @ to MANUAL (light off) and press Ri key (SHIFT key @ and then "4" key of DATA block G). ENTER keys ® start flashing and FREQ/AMPTO readout ® shows ‘the current assumptive load resistance. Enter any resistance value between 2 and S000 by DATA keys @® while ENTER keys @ are flashing. Press either ENTER keys ® to fix the selected value of assumptive load resistance. Example 4-22) Setting assumptive load resistance to 16.2 AUTO SHIFT Re ENTER ° ° 4 1 6 ° (light off) After setting of the assumptive load resistance, FREQ/AMPTD readout @ will show the newly set load resistance for two seconds. Then it will return to the state displaying the measured value of frequency. After setting in the manual measurement, the specified value of the assumptive load resistance will remain even if AUTO key @ is set to AUTO (light on). The rotary knob cannot be used to set the assumptive load resistance. (5) GP-I8 program code The GP-IB controlled operations in the WATT indication are selection of the measurement function, setting of the AC level measurement range in the manual measurement, and setting of the assumptive load resistance. Table 4-25 GP-IB program code in WATT indication Header) pata code | Unit Description code code MM 5 Selects WATT indication function MD 20 Selects auto ranging for AC level measurement range 24 Sets AC level measurement range at 100 (40dB, 42.2 dBm) 22 Sets AC level measurement range at 31.6V(30dB, 32.2dBm) 23 Sets AC level measurement range at 3.16V(10d8, 12.2d8m) 24 Sets AC level measurement range at 316mV(-10d8, ~7.8d8m) 25 Sets AC level measurement range at 31.6mV(~30d8, ~27.8d8m) 26 Sets AC level measurement range at 3.16mV(~50d8, ~47.8d3m) 27 Sets AC level measurement range at 0.316 mV(~70d8, -67.8.d8m) 2 t0 4.000 Sets assumptive load resistance at 2 to 50002 OPERATION Example 4-24) Setting input range by the rotary knob Avro SHIFT = NPUT METER °] fo] [1] © fe (light off) Rotate to indicate desired range. (b) Fixing the measurement range Set AUTO key @ to MANUAL (light off) and press MEAS RANGE key (SHIFT key @ and then "2" key of DATA block @). ENTER keys @ start flashing and MEASUREMENT readout © shows the current measurement range. Enter any range code given in Table 4- 28 by DATA keys @ while ENTER keys @ are flashing. Press either of ENTER keys @ to fix the selected measurement range. Table 4-28 Measurement range code in wow & flutter measurement Code Measurement range (96) 0 ‘Automatic ranging (No indication) 1 10 2 1 3 04 Example 4-25) Fixing measurement range to 0.1% AUTO SHIFT MEAS ENTER ° ° 2 3 ° (light off) If either of ENTER keys @ is directly pressed without any range code, the current range is fixed After fixing measurement range, MEASUREMENT readout © will show the newly fixed measurement range for two seconds. Then it will return to the state displaying the measured value of wow & flutter. Alternatively, select the measurement range with the rotary knob after pressing MEAS RANGE key. Press either of ENTER keys @ to fix the measurement range. Example 4-26) Setting measurement range by the rotary knob auto SHIFT. MEAS ENTER °] [oe 7] O fe (light off) A Rotate to indicate desired range. (0) Selecting the center frequency Set AUTO key @ to MANUAL (light off) and press W &F fc key (SHIFT key @) and then “5” key of DATA block @). The ENTER keys @@ start flashing and FREQ/AMPTD readout @) shows the current center frequency. Enter any frequency code given in Table 4-29 by DATA keys @ while ENTER keys @ are flashing. Press either of ENTER keys @ to fix the selected center frequency. OPERATION Table 4-29 Center frequency selection code in wow & flutter measurement Code Center frequency (kHz) 1 3 2 BAS Example 4-27) Setting center frequency to 3.15 kHz auto shirt WRF ENTER 3 3 5 2] [e (light off) After selection of the center frequency, FREQ/AMPTD readout © will show the newly selected center frequency for two seconds. Then it will return to the state displaying the measured value of the frequency. After selecting, the specified value of the center frequency will remain even if AUTO key @ is set to AUTO (light on). The rotary knob cannot be used to set the center frequency. (6) GP-I8 program code The GPIB controlled operations in the wow & flutter measurement are selection of the measurement function, psophometric weighting ON/OFF, and setting of the measurement and input ranges and that of the center frequency in the manual measurement. Table 4-30 GP-I8 program code in wow & flutter measurement Header code| Data code | Unit code Description vm 6 Selects wow & flutter measurement function wt 0 ‘Turns psophometric weighting off ' ‘Turns psophometric weighting on mo 1.0 Selects auto ranging for input range W Sets input range at 100V (408, 42.2 8m) 12 Sets input range at 31.6V (30dB, 32.2 8m) 13 Sets input range at 10V (20 dB, 22.2 d8m) ia Sets input range at 3.16V (1048, 12.2d8m) 15 Sets input range at 1V (0dB, 2.2d8m) 16 Sets input range at 0.316V (- 1008, ~7.8d8m) fy Sets input range at 3.16 mV (-50d8, - 47.8d8m) 20 Selects auto ranging for measurement range 24 Sets measurement range at 10% 22 Sets measurement range at 1% 23 Sets measurement range at 0.1% SA Sets center frequency at 3 kHz 52 Sets center frequency at 3.15kHz OPERATION 4-17 SELECTING INDICATION RESPONSE (1) General The response characteristics of the AC/DC conversion circuit (detector) of this instrument al- low to select either the rms-responding (RMS) or average-responding (AVG), and also to set the ‘time constant of the indication response to SLOW or FAST. (2) Selecting the rms- or average-responding characteristics If AVG/RMS key of RESPONSE block @ is set to RMS (light off), the indication response characteristics is rms-responding. In case of AVG (light on), the response characteristics is average-responding. Note that some indication response characteristics are always rms or quasi-peak-responding, independent of AVG/RMS key. Table 4-31 shows the relationship be- tween the setting of AVG/RMS key and the response characteristics of each measurement func- tion. Table 4-31 Indication response characteristics for each measurement Indication response characteristics Function Item AVG/RMS key of RESPONSE block @ RMS (light off) AVG (li Distortion Input level rms-responding Measured value | rms-responding _ | average-responding AC level Measured value | rms-responding —_|_average-responding SIN S component rms-responding | _average-responding N component tms-responding | _average-responding WATT indication | AC level tms-responding | _average-responding War Input level rms-responding Measured value | quasi-peak responding (option 01) ims-responding (option 02) (3) Selecting SLOW/ FAST SLOW / FAST key of RESPONSE block @ allows to select the time constant in indication re- sponse. When the input signal frequency. is below 100 Hz, SLOW/FAST key should be set to SLOW (light on) to minimize the variation of the measured value’s indication or a measurement error as well as to ensure the auto-ranging operation. (4) GPIB program code The selection of the indication response characteristics can be controlled by GP-IB. Table 4-32 GP-IB program code regarding indication response characteristics Header code Data code Unit code Description DE 1 Sets to rms-responding 2 Sets to average-responding RS. 1 Sets time constant to FAST 2 Sets time constant to SLOW OPERATION 4-18 INDICATION UNITS (1) General This instrument can select V % (linear) or dB (log) as an indication unit for each measurement function, (2) Selecting indication unit When UNIT key @ is set to V% (light off), the unit is in the linear system such as V, mV, and %, and set to dB (light on), the unit is in the log system such as dB, dBV, and d8m. Note that the indication unit is fixed in some measurement functions, independent of the setting of UNIT key @. The selection between d8V and dBm depends on the specified unit of the signal source output amplitude. Table 4-33 shows the relationship among UNIT key @, the specified unit of the signal source, and the indication unit in each measurement function, Table 4-33 Indication unit in each measurement function Indication unit UNIT key @ setting Function Item v% dB (light on) Make off, Signal source unit av dam Distortion Input level v, mv aBv dBm Measured value % DC level | Measured value Vv, mv AClevel | Measured value} Vv, mv av dam Relative level | Reference value] V, mV. dav dBm Measured value 8 SiN $ component v, mv dav dam Measured value Ey WATT indication | Measured value w Wee Input level Vv, mv av dam Measured value % Ca (3) GP-1B program code The selection of the indication unit can be controlled by GP-IB. Table 4-34 GP-IB program code for indication unit Header code Data code Unit code description uN Sets indication unit to V, % Ls Sets indication unit to dB OPERATION 4-19 MEASUREMENT FILTERS (1) General This instrument can insert various filters in the measuremenit system. The characteristics of each built-in filter will be described in Table 4-35. Table 4-35 Measurement filters Category | Marking Description Note High-pass [100Hz | Cutoff frequency: 75 Hz Pilot signal rejection of AM stereo (HPF) Attenuation: 24048 at 25Hz signal 200Hz —_ | Cutoff frequency: 180 Hz Lower end characteristics of IHF Rolloff : 60.d8/decade BPF Low-pass |1SkHz | Cutoff frequency: 15kHz Pilot signal rejection of FM stereo ) Attenuation: =30dB at 19kHz signal Higher end characteristics 20kHz | Cutoff frequency : 20 kHz For CD use Attenuation: 2308 at 24.1 kHz B0kHz — | Cutoff frequency: 80kHz Rolloff: 60 dB / decade PT Optional filter Psophome- |A 1EC-651 A weighting characteristics trie weighting [AUDIO _ [DIN 45405 AUDIO filter (PsopHo) [ccin-arm |cciR-ARM (DOLBY) weighting OPT JOptional filter As for detailed information of optional filters, see SECTION VII (2) Selecting filter If any key of FILTERS block ® is turned on, the corresponding filter is inserted in the mea- surement system. Each key can independently select filter ON or OFF alternately. As Table 4- 35 shows, the measurement filters are classified into three categories: HPF, LPF, and PSOPHO. Any of the filters in the same category reset one another. (3) GPIB program code Measurement filters can be controlled by GP-I8. Table 4-36 GP-18 program code regarding measuring filters OPERATION Header code | Data code Unit code Description HP 0 HPF OFF 1 100 Hz HPF ON 2 200 Hz HPF ON. Lp 0 LPF OFF 1 1SkHz LPF ON 20 kH2 LPF ON 180 kHz LPF ON Optional filter ON Ps PSOPHO filter OFF 1EC-A filter ON DIN AUDIO filter ON CCIR-ARM filter ON Optional filter ON 4-20 BALANCED INPUT AND FLOATING CONNECIION, (1) General AC INPUT connector ® of this instrument can select either unbalanced input for usual measurement or balanced input for the measurement of a balanced circuit such as BTL amplifier, The common lines of AC INPUT connectors @ and DC INPUT connector ® can be floated or connected to the instrument frame. Figure 4-3 shows the configuration of the input section. UNBAL sO ATT 7 = ‘AC INPUT st AC MEAS iy BAL 8 © AIT a2 oc input ©) AT Ad DC MEAS ¥Y we Figure 4-3 Configuration of the VP-7723A input section x7 OPERATION (2) Selecting balanced unbalanced input When BAL/UNBAL key ()is set to UNBAL (light off), $1 shown in Figure 4-3 is connected to the UNBAL side and the unbalanced input is available only for A-connector. When BAL/ UNBAL key @is set to BAL (light on), $1 is connected to the BAL side and the balanced input is available for both A- and B-connectors. (3) Floating connection When the floating/ grounding switch @ is set to WY side, $2 in Figure 4-3 is opened and the common lines of both AC INPUT connectors @and DC INPUT connector @) are floated from the frame ground. When the switch is set to L side, $2 is short-circuited and the common lines are connected to the frame ground. The selection of floating connection cannot be controlled by GP-IB. (4) GPIB program code The selection of balanced /unbalanced connection can be controlled by GP-IB, Table 4-37 GP-IB program code for balanced input Header code_| Data code Unit code Description aL 0 Sets input terminal to UNBAL 1 Sets input terminal to BAL 4-21, AUTOMATIC/ MANUAL MEASUREMENT (1) General This instrument provides automatic measurement functions, such as auto ranging and auto tuning, which allow te measure various data automatically only by applying a signal Where required, manual measurement for range setting is provided. See paragraphs 4-10 to 4-16 for detailed information on automatic and manual operation for each measurement function This paragraph describes the all-hold function. (2) All-hold operation Set AUTO key @ to MANU (light off). Press ALL HOLD key (SHIFT key @) and then “6" key of DATA block (@®)) so that the items of each measurement function shown in Table 4-37 are fixed to the currently set condition. Table 4-38 Items fixed by the all-hold function Measurement function tems fixed Distortion Tuning frequency of fundamental-rejection filter, input level range, measurement range DC level Measurement range AC level Measurement range Relative value AC level measurement range SIN S component measurement range N_component measurement range WATT indication AC level measurement range Wow & flutter Input level range, measurement range OPERATION (3) GP-18 program code The selection of automatic measurement or the all-hold function in the manual measurement can be controlled by GP-IB. Table 4-39 GP-IB program code regarding automatic/ manual measurement Header code Data code Unit code Description AU Selects automatic measurement mo 6. Selects all-hold function 4-22 LIMIT JUDGMENT FUNCTION (1) General This instrument specifies an upper limit, lower limit, or both, judges whether the measured values are within a range, and indicates the judgment result. Upper and lower limits can be specified for each measurement function, and also stored in the assorted preset memory. Each light of OVER, PASS, and UNDER of LIMIT indication(@) means the following : OVER light on Measured value is over the upper limit. PASS light on Measured value is within the range. UNDER light on Measured value is under the lower limit. (2) Setting limit values First, specify the required items for each measurement function. Press UPPER LIMIT key (SHIFT key @) and then FREQ key of Setting function selection block @) or LOWER LIMIT key (SHIFT key @ and then AMPTD key of Setting function selection block @ ). The current specified upper or lower limit value appears on MEASUREMENT readout @) and ENTER keys @ start flashing. Enter any limit value within the range given in Table 4-40 by DATA keys while ENTER keys @ are flashing. Press either of ENTER keys @ to fix the upper or lower limit value. Table 4-40 Setting range for limit judgment function Measurement function Unit Setting range DIsTN % 0.00010 % to 31.6% a8 = 120.00 d8 to ~ 10.008 Dc LeveL, v £1.0mV to £100.0V AC LEVEL vimv _|0.0010mv to 100.0v 8 = 120.00 d8 to 40.0048 dBm 117.78 dB to 42.22 dBm RELATIVE LEVEL dB = 160.00 dB to 160.00 dB SIN dB 0.0.48 to 160.048 WATT w 0.01 W to 999.99 W W &F UNWTD % 0.0010 % to 10.00% W &F WTD OPERATION Example 4-28) Setting upper limit to 0.05 % in distortion measurement UNIT a8 DISTN ove ° ° Selection of measurement function and indication unit light on light off SHIFT UPPER LIMIT ENTER ° FREQ 0 0 5 ° if either of ENTER keys (@) is directly pressed without any limit value, the limit value is cleared, After setting of the limit value, MEASUREMENT readout © will show the newly specified limit value for two seconds. Then it will return to the state displaying the measured value. The rotary knob cannot be used to specify the limit value (3) GPB program code The setting of limit values in limit judgment function can be controlled by GP-1B. Table 4-41 GP-IB program code for limit function Header code Data code unit code Description ue 0.00010 to 31.6 Pc Sets the upper limit for limit function of 0.000010 to 100.0 v the current measurement 0.0010 to 100000 mv 0.01 to 999.99 w = 160.00 to 160.00 DB = 117.78 to 42.22 bm [Same as dbove| Cancels upper limit (no data code) uw Same data codes|Same as above] Sets the lower limit for limit function of jas in UL the current measurement same as above| Cancels lower limit (no data code) Since the setting range of data code depends on the measurement function, see Table 4-40 for details. In addition to panel indication, the GP-I8 output data, LED lighting output of EXT CONTROL 170, output data in data print function are also provided as the limit judgment result. See Paragraphs 6-10, 7-8 and 7-12 for details. OPERATION 4-23 ASSORTED PRESET MEMORY The assorted preset memory is used to store up to 100 setups such as the signal source and measurement conditions specified according to each operation procedure, and to recall any sets of them at a time. (1) Presettable contents in one setup Table 4-42 shows the contents to be stored in the assorted preset memory. Table 4-42 Contents stored in assorted preset memory Item Contents ignal source Output ON/OFF Frequency 5.0 Hz to 110.0 kHz Output amplitude ~ 85.9 to 14.0 dBV/ - 83,7 to 16.2d8m fueasurement function | DISTN, OC LEVEL, AC LEVEL, RELATIVE LEVEL, $/N, WATT, W&F Distortion AUTO/ MANUAL JAUTO/ MANU : INPUT RANGE, MEAS RANGE, NOTCH FREQ Joc tevet AUTO/ MANUAL, JAUTO/ MANU : MEAS RANGE JAC level, AUTO/ MANUAL JAUTO/ MANU : MEAS RANGE [Relative level Reference value 0.01 mV to 100V/ ~ 99.99 dBV to 40 dBV/ ~97.78dBm to 42.22 d8m AUTO / MANUAL AUTO/ MANU: MEAS RANGE iN AUTO/ MANUAL JAUTO/ MANU :S RANGE, N RANGE /ATT indication Assumptive resistance |2 to 50009 AUTO/ MANUAL JAUTO/ MANU : MEAS RANGE &F Center frequency 3kHz/3.15 kHz Psophometric weighting] UNWTD/WTD AUTO/ MANUAL |AUTO/ MANU : INPUT RANGE, MEAS RANGE Response characteristics. |AVG/RMS, SLOW/ FAST Indication unit Vv % 1B leasurement filter HPF 100 Hz /200 Hz / OFF LPF 15 kHz /20 kHz /80 kHz / OPT/ OFF PSOPHO AJ AUDIO / CCIR-ARM / OPT / OFF Input BAL/UNBAL it judgment function |UPPER LIMIT, LOWER LIMIT lexternal control output Port 1 0 to 255 Port 2 0 to 255 OPERATION (2) Memory address The 100 setups of preset memory are given memory addresses of 00 to 99. The memory address appears on ADDRESS readout. (3) Storing procedure Operate the front panel to obtain the required setup. Press STO key of Setting function selection block @ so that ENTER keys @ start flashing. This means the system is ready for entering of memory address to be stored. Enter any memory address by DATA keys @ while ENTER keys @ are flashing. Press either of ENTER keys @ to store the setup at that memory address in the preset memory. Example 4-29) Storing the current setup into memory address 12 ENTER STO i 2 ° If either of ENTER keys @ is directly pressed without any address, the setup will be stored into the currently displayed address. (4) Direct recall operation Press RCL key of Setting function selection block @ so that ENTER keys @ start flashing, The system is ready for entering of memory address to be recalled. Enter any two-digit memory address by DATA keys (@ to recall a stored preset data while ENTER keys are flashing. Example 4-30) Recalling from memory address 12 RCL 1 2 The preset memory at address 00 to 09 can be also recalled by entering the single-digit address as below. Example 4-31) Recalling from memory address 1 ENTER RCL i ° If either of ENTER keys @ is directly pressed without any address, the setup stored at the currently displayed address is recalled. (5) Sequential recall operation (a) General All addresses between specified start and end addresses can be sequentially recalled with @ single key. This paragraph deals with the setting operation of start and end addresses and the sequential recalling procedure. (b) Setting start/end address Press STO key of Setting function selection block @ so that ENTER keys @ start flashing. The system is ready for entering of memory address. To set the start and end addresses, press the point ( .) key, two-digit start address, point (.) key, two-digit end address, and either of ENTER keys @ in turn while ENTER keys are flashing Example 4-32) Setting start address to 01 and end address to 23 ENTER STO o 1 2 3 ° 4-38 OPERATION When the start/end addresses are set, the decimal point at the rightmost digit of the address number on ADDRESS readout @ is turned on. Example 4-33 indicates the keystroke to release the start/end addresses, which causes the same result as setting start address to 00 and end address to 99. Example 4-33) Releasing start/end address ENTER STO E ° NOTE Ih setting start/end address as shown in Example 4-32, the system always regards the smaller address number as the start address. For example, in the following keystroke of sequential recalling operation, the sequential recalling starts at address 01 and ends at address 23. ENTER NOIRIRMIC 7] [7] Le (0) Sequential recalling procedure Press t or | key of MEMORY block @, and CLR key to recall the preset memory sequentially. When fis pressed, the address following the currently displayed memory address is recalled. Pressing this key when the end address is displayed recalls the start address. When | is pressed, the address preceding the currently displayed address is recalled Pressing this key when the start address is displayed recalls the end address. When CLR is pressed, the start address is recalled. Pressing this key when the start and end addresses are released recalls address 00, (6) Grouping of sequential recall operation (a) General Preset memory can be divided into up to 10 groups, any one of which can be selected to execute the sequential recall operation. This paragraph describes the procedures to divide the memory into groups, to select a group for sequential recall, and to release groups and sequential recalling. (b) Grouping Press STO key of Setting function section block ® so that ENTER keys ® start flashing, The system is ready for entering of memory address. While ENTER keys are flashing, press the point (.) key, two-digit start address, point (.) key, two-digit end address, point (.) key, single-digit group number, and either of ENTER keys ® in turn to specify the start and end addresses and the group number. Example 4-34) Setting start address to 01, end address to 23, and group to 4 ENTER STO é o | 1 2 3 4 ° ‘Two or more groups can share the same addresses. OPERATION Example 4-35) Division by three groups sharing same address ENTER ‘STO : 0 0 1 9 5 0 ° ENTER STO 1 0 : 2 9 : 1 ° ENTER STO 2 0 3 9 2 ° (0) Group designation for sequential recall Press RCL key of Setting function selection block @ so that ENTER keys @ start flashing. The system is ready for entering of memory address. While ENTER keys @ are flashing, press the point (.) key, single-digit group number, and either of ENTER keys @ in turn to designate the group number where sequential recall is performed. Example 4-36) Designating group 0 ENTER RCL 0 ° When a group number is designated, the decimal point at the rightmost digit on MEMORY ADDRESS readout@)is turned on. (d) Releasing grouped sequential recalling operation Example 4-37 indicates the keystroke to release the start and end addresses. It causes the same result as setting start address to 00 and end address to 99 without group designation, The grouping is still effective even after this operation. Example 4-37) Releasing grouped sequential recalling operation ENTER STO i : ° (6) GP-18 program code Storing of preset memory and direct recall operation can be controlled by GP-IB. Table 4-43 GP-IB program for preset memory Header code Data code Description RC 00099 | Recalls data from preset memory address 00 to 99 st 00 t099 | stores data into preset memory address 00 to 99 4-24 AUTO SEQUENCE FOR ASSORTED PRESET MEMORY (1) General This instrument can perform the auto sequence function which automatically recalls assorted preset memory (explained in the previous paragraph 4-23) in sequence at desired time interval. The basic operations of the auto sequence function are mode and interval time setting for auto sequence, and start/stop of auto sequence operation. (2) Mode setting for auto sequence Four operational modes are provided for auto sequence as follows : Mode Nbr. ° 1 2 3 Mode REPEAT UP SINGLE UP REPEAT DOWN SINGLE DOWN (a) Indication The auto sequence mode is indicated with other 1/0 modes on FREQ/AMPTD readout (@) only in setting and verifying operations. OPERATION Executes auto sequence from start to end repeatedly. Executes auto sequence from start to end only once. Executes auto sequence from end to start repeatedly. Executes auto sequence from end to start only once. Press 1/0 MODE key (SHIFT key @ and then STO key of Setting function selection block @), so that 1/0 MODE light of ADDRESS readout @) is turned on and FREQ/AMPTD readout (@) changes as shown in Example 4-38. Example 4-38) Mode verification for auto sequence keystroke SHIFT 1/0 MODE ° STO FREQ/AMPTD readout 4 TL| P1{p2]as | PR On this readout, AS indicates the mode of auto sequence. If any key other than those used in the succeeding paragraph (b) is pressed after the verifying operation in Example 4-38, 1/O MODE light of ADDRESS readout @) is turned off and FREQ/AMPTD readout @) will return to the state displaying the measured value of the frequen (6) Setting ey. operation Press 1/0 MODE key (SHIFT key @) and then STO key of Setting function selection block ®), so that ENTER keys ©@ start flashing. The system is ready for 1/0 MODE settiny, While ENTER keys @ are flashing, make AS digit flashing by pressing DIGIT SELECTOR key of MODIFY block @, and enter desired mode number by DATA keys (3). Press either of ENTER keys @ to set the mode for auto sequence. Example 4-39) Selecting SINGLE UP mode for auto sequence step Keystroke FREQ/ AMPTD readout @ o ‘PEELE «2 SHIFT 1/0 MODE @ ° STO DIGIT SELECTOR DIGIT SELECTOR ° an 0 0 ENTER ® 7 ° ® one (3) Interval time NOTE Current measured value of frequency. Current specified value of 1 10 mode. Make AS digit flashing, Set 1 for the mode. Dis- played for about 2 seconds. Current measured value of frequency. In the auto sequence operation, the time interval between two memory recalling operations ified. In the $/N measurement, the $ component measurement time is added to Interval time can be changed for each memory address. can be spe the speci 1d interval time. (2) Indication The interval time is indicated in second on FREQ/AMPTO readout @ along with print setting for data print function (See paragraph 7-11) only in setting and verifying operations 4-41

You might also like