USER MANUAL
/i
Copyright
Copyright © 2011, by Petroleum Analyzer Company, L.P.
All rights reserved.
All text, graphics, design, and other works contained herein are the copyrighted
works of Petroleum Analyzer Company, L.P. (hereinafter referred to as PAC).
Any redistribution or reproduction of any materials contained herein is strictly
prohibited without the express written permission of the copyright holder.
The PAC trademark, the PAC service mark, PAC’s logo trademark, PAC division’s
trademarks, PAC product’s trademarks, and PAC’s distinctive colour trademark as
used on its instruments and in these materials, are trademarks or service marks or
registered trademarks or registered service marks of PAC in the U.S. and other
countries and are its sole and exclusive property.
MS Windows, MS Windows XP, and MS Windows Professional are registered
trademarks of Microsoft Corp.
Other trademarks contained within these materials are the property of their
respective owners.
Information
The information contained within these materials is subject to change without
notice. PAC makes no warranties or guarantees, either expressed or implied in
these materials, including the warranty merchantability or fitness for a particular
purpose. At any time, PAC may modify these materials, its instruments or its
programs without notice and subsequent versions of these materials may contain
different information.
These materials could contain technical inaccuracies and/or typographical errors.
PAC does not assume responsibility for the accuracy of any translation of these
materials.
About PAC
PAC is an international manufacturing and service organization with a portfolio
that spans petroleum, petrochemical, biofuels, environmental, food and beverage,
pharmaceutical and industrial analysis solutions.
PAC provides advanced scientific apparatus and testing equipment for laboratory
process, on-line and field use from small to the largest enterprise businesses.
PAC offers analytical solutions for a wide variety of applications, including:
chromatographic systems and detectors, elemental, laboratory, and on-line
process instruments, software applications, and spectroscopy.
Many automatic analyzers bearing the PAC brand are recognized for increasing
test precision, boosting laboratory efficiency and reducing costs.
UM_2.00A_EN iii
OptiDist™
iv UM_2.00A_EN
OptiDist™
Table of contents
1 General information
1.1 Policy statement ..................................................................................................................... 2
1.2 Standard limited warranty of PAC .......................................................................................... 3
1.2.1 Limited warranty ......................................................................................................... 3
1.2.2 Limitations on PAC liability ......................................................................................... 7
1.2.3 Compliance ................................................................................................................ 8
1.3 Software license ..................................................................................................................... 9
1.3.1 Grant of license .......................................................................................................... 9
1.3.2 Description of other rights and limitations .................................................................. 9
1.3.3 Upgrades.................................................................................................................. 10
1.3.4 Copyright .................................................................................................................. 10
1.3.5 Dual-media software ................................................................................................ 10
1.3.6 U.S. Government restricted rights ............................................................................ 11
1.4 Assumptions ......................................................................................................................... 11
1.4.1 Good Laboratory Practice ........................................................................................ 11
1.4.2 The standards of laboratory safety........................................................................... 11
1.4.3 Electronic device ...................................................................................................... 11
1.5 Return policy......................................................................................................................... 12
1.5.1 Returned product warranty determination ................................................................ 12
1.5.2 On-site repair............................................................................................................ 12
1.5.3 Inspection ................................................................................................................. 12
1.5.4 Information ............................................................................................................... 13
1.6 Disposal................................................................................................................................ 14
1.7 Documentation ..................................................................................................................... 15
1.7.1 Document validity ..................................................................................................... 15
1.7.2 Document storage .................................................................................................... 15
1.7.3 Language ................................................................................................................. 15
1.7.4 Typographical and authoring conventions................................................................ 15
2 Safety
2.1 Certifications......................................................................................................................... 18
2.2 Warning pictograms.............................................................................................................. 19
2.3 Caution pictograms............................................................................................................... 20
2.4 Note pictograms ................................................................................................................... 21
2.5 Tip pictograms ...................................................................................................................... 21
2.6 Environmental protection...................................................................................................... 21
3 Introduction
3.1 Intended use......................................................................................................................... 24
3.2 Intended audience ................................................................................................................ 24
3.3 Product overview .................................................................................................................. 24
3.4 Applicable standard methods ............................................................................................... 25
3.5 Benefits and features............................................................................................................ 26
3.6 General Scope...................................................................................................................... 27
3.7 Terminology.......................................................................................................................... 27
3.8 List of abbreviations.............................................................................................................. 28
UM_2.00A_EN v
OptiDist™
vi UM_2.00A_EN
OptiDist™
5 Installation
5.1 Technical specifications and requirements........................................................................... 70
5.2 Care in Unpacking ................................................................................................................ 70
5.3 Installation of the OptiDist™ ................................................................................................. 71
5.4 Main power connection......................................................................................................... 71
5.5 CO2 gas connection ............................................................................................................. 72
5.6 OptiDist™/ PC Connection ................................................................................................... 72
5.6.1 Optional external PC control .................................................................................... 72
5.6.2 Connection with RS 232C serial link ........................................................................ 72
5.7 Flask stopper holder ............................................................................................................. 72
5.8 Temperature probe connection ............................................................................................ 73
5.8.1 Pt100 vapor temperature probe ............................................................................... 73
5.8.2 Ambient temperature probe...................................................................................... 73
5.9 125 ml and 200 ml automatic dry point kit and holder .......................................................... 73
5.9.1 Installation of the automatic dry point kit and holder ................................................ 73
5.9.2 Adjustment of the vapor probe in the dry point stopper............................................ 74
5.10 Installation of the status indicator beacon ............................................................................ 75
5.11 VOC Control Option Connection .......................................................................................... 75
5.12 Connection of an optional external printer............................................................................ 75
5.13 Connection of the optional bar code reader ......................................................................... 75
5.14 Power-On procedure ............................................................................................................ 76
5.14.1 Switch on the OptiDist™ .......................................................................................... 76
5.14.2 PAC addresses screen............................................................................................. 76
5.14.3 Heat exchanging liquid (level checking and recommendations) .............................. 76
5.14.4 Self diagnostic screen .............................................................................................. 76
5.14.5 OptiDist™ calibration................................................................................................ 76
6 Operation
6.1 Preparation ........................................................................................................................... 80
6.1.1 Basic preconditions .................................................................................................. 80
6.1.2 Safety ....................................................................................................................... 80
6.1.3 Sampling, storage and sample conditioning according to ASTM D86 ..................... 81
6.1.4 Clean the devices..................................................................................................... 84
6.1.5 Preparation of the OptiDist™ and accessories ........................................................ 86
6.1.6 Set-up of the distillation flask and the receiving cylinder.......................................... 88
6.1.7 Set up the receiving chamber temperature .............................................................. 90
6.1.8 Scan the volume in the receiver cylinder.................................................................. 91
6.2 Measurements...................................................................................................................... 93
6.2.1 Start modes .............................................................................................................. 93
6.2.2 Start procedure - run a distillation test.................................................................... 108
6.2.3 Residue measurement ........................................................................................... 110
6.2.4 Displaying and printing of results ........................................................................... 111
7 Database management
7.1 The Results database......................................................................................................... 114
7.1.1 Displaying of results ............................................................................................... 114
7.1.2 Printing of results.................................................................................................... 117
7.1.3 Transmitting of results ............................................................................................ 117
7.1.4 Modifying of results ................................................................................................ 119
7.1.5 Creating of a new optimized product...................................................................... 121
7.1.6 Removing of results................................................................................................ 122
UM_2.00A_EN vii
OptiDist™
viii UM_2.00A_EN
OptiDist™
UM_2.00A_EN ix
OptiDist™
x UM_2.00A_EN
General information OptiDist™
1 General information
UM_2.00A_EN 1
OptiDist™ General information
This apparatus should be used only for its intended operation; alterations will void
the warranty.
2 UM_2.00A_EN
General information OptiDist™
1.2.1.1 Warranty
For each product sold, PAC offers to the original owner ("Buyer") a limited
warranty against failure to conform to the product specifications or any defects in
material and workmanship for a period of
i twelve (12) months from the date of installation of the product or
ii eighteen (18) months from date of invoice,
During the Initial Warranty period, PAC offers a limited warranty on each part or
product repaired or replaced by a PAC service person for a period ending the later
of
a the remaining term of the Initial Warranty of the product or
b ninety (90) days from the date of repair or replacement,
whichever is longer. After expiration of the Initial Warranty period, PAC offers a
ninety (90) day limited warranty on each part or product repaired or replaced by a
PAC service person. PAC further warrants that the products and parts it sells will
conform to PAC's written specifications therefor. The foregoing limited warranties
cover parts and labor only and PAC does not warrant and will not reimburse the
Buyer for any other costs relating repairing or replacing the product at issue. The
foregoing limited warranties apply only to the repair or replacement of defective
parts and/or products and such determination will be in the sole discretion of PAC.
The limited warranties of this Section 1.2.1.1. Warranty are further subject to those
warranty exclusions set forth in Section 1.2.1.3. Limited warranty exclusions.
UM_2.00A_EN 3
OptiDist™ General information
1.2.1.2 Aftermarket
The use of third party parts in the operation or maintenance of the product or
repairs or servicing by unauthorized service personnel immediately voids all
further warranty obligations of PAC. PAC genuine consumables are sold on an as-
is basis and have no warranty beyond being shipped in good working order.
4 UM_2.00A_EN
General information OptiDist™
To the extent a PAC product is used in connection with the operation of a nuclear
power facility, Buyer agrees to indemnify and hold PAC harmless from any and all
actions, claims, suits, damages and expenses arising from such use. PAC
provides no warranty on the oral representations made by its personnel while they
are attempting to assist Buyer in the operation of a product. This Standard Limited
Warranty does not apply to items consumed by the products during their ordinary
use, including but not limited to fuses, batteries, paper, septa, fittings, screws,
fuses, pyrolysis, dryer or scrubber tubes, sample boats, furnaces or lamps.
This warranty is valid only if genuine PAC parts, consumables and standards are
used in PAC products.
UM_2.00A_EN 5
OptiDist™ General information
6 UM_2.00A_EN
General information OptiDist™
1.2.2.1 General
Buyer agrees that in no case shall PAC be liable for any special, incidental,
consequential, or punitive damages based upon any legal theory whether or not
such damages are foreseeable. Such damages include, but are not limited to, loss
of profits, loss of savings or revenue, loss of use of the product or any associated
equipment, cost of capital, cost of any substitute equipment, facilities or services,
downtime, the claims of third parties including customers, injury to property and,
unless precluded under applicable state law, bodily and personal injury. PAC's
total liability for any and all losses and damages arising out of any and all causes
whatsoever including, without limitation, defects in the product(s), services,
software, or documentation supplied or breach of this agreement, shall in no event
exceed the purchase price of the applicable product(s). Buyer agrees these
limitations on PAC’s liability are reasonable and reflected in the amounts charged
by PAC for its products.
UM_2.00A_EN 7
OptiDist™ General information
1.2.3 Compliance
1.2.3.1 General
These Warranty Terms are subject to change without notice. PAC also retains the
right to modify these warranty terms in order to comply with policy or laws
governing warranty issues in states or countries having specific remedies differing
or additional to those described within this document.
1.2.3.2 Severability
If any one or more of the provisions or subjects contained in the Agreement shall
for any reason be held invalid, illegal, or unenforceable, it shall not affect the
validity and enforceability of any other provisions or subjects.
8 UM_2.00A_EN
General information OptiDist™
UM_2.00A_EN 9
OptiDist™ General information
• Single COMPUTER
The SOFTWARE PRODUCT is licensed with the COMPUTER as a single
integrated product. The SOFTWARE PRODUCT may only be used with the
COMPUTER.
• Rental
You may not rent or release the SOFTWARE PRODUCT.
• Software Transfer
You may permanently transfer all of your rights under this EULA only as part
of a sale or transfer of the COMPUTER, provided you retain no copies, you
transfer all of the SOFTWARE PRODUCT (including all component parts, the
media and printed materials, any upgrades, this EULA and, if applicable, the
Certificate(s) of Authenticity), and the recipient agrees to the terms of this
EULA. If the SOFTWARE PRODUCT is an upgrade, any transfer must include
all prior versions of the SOFTWARE PRODUCT.
• Termination
Without prejudice to any other rights, PAC may terminate this EULA if you fail
to comply with the terms and conditions of this EULA. In such event, you must
destroy all copies of the SOFTWARE PRODUCT and all of its component
parts.
1.3.3 Upgrades
If the SOFTWARE PRODUCT is an upgrade of a component of a package of
software programs which you licensed as a single product, the SOFTWARE
PRODUCT may be used and transferred only as part of that single product
package and may not be separated for use on more than one computer.
1.3.4 Copyright
All title and copyrights in and to the SOFTWARE PRODUCT (including but not
limited to any images, photographs, animations, video, audio, music, text and
“applets,” incorporated into the SOFTWARE PRODUCT), the accompanying
printed materials, and any copies of the SOFTWARE PRODUCT, are owned by
PAC or its suppliers. The SOFTWARE PRODUCT is protected by copyright laws
and international treaty provisions. You may not copy the printed materials
accompanying the SOFTWARE PRODUCT.
10 UM_2.00A_EN
General information OptiDist™
1.4 Assumptions
UM_2.00A_EN 11
OptiDist™ General information
1.5.3 Inspection
This system was thoroughly inspected and carefully packed before leaving our
factory. Responsibility for its safe delivery was assumed by the carrier upon
acceptance of the shipment. Claims for loss or damage sustained in transit must
be made upon the carrier by the recipient as follows.
12 UM_2.00A_EN
General information OptiDist™
CAUTION
DO NOT - for any reason - return the instrument without first
obtaining authorization.
1.5.4 Information
In any correspondence to PAC, please supply the data from the nameplate,
including catalogue number and serial number.
/i
Catalog number:
Serial number:
Purchase date:
UM_2.00A_EN 13
OptiDist™ General information
1.6 Disposal
Correct disposal of Waste Electrical and Electronic Equipment in
the end-of life
Applicable in the European Union and other European countries with
separate collection systems.
The above WEEE-symbol is the official marking for equipment under the WEEE-
scope. In some EC-Member states "pure B2B equipment" is not necessarily
marked with the waste bin-symbol. To provide a homogenous EC-wide procedure,
PAC however uses the marking in all EC-Member states.
14 UM_2.00A_EN
General information OptiDist™
1.7 Documentation
1.7.3 Language
The source language of this manual is English.
As a translation may be inaccurate or imprecise, the original source text will be
legally binding and shall prevail in the case of any difference between the source
text and any translation.
UM_2.00A_EN 15
OptiDist™ General information
• Throughout this manual, PAC automatic laboratory equipment (or device, unit,
or apparatus) is referred to as the ‘instrument’.
• Throughout this manual, the complete analysis system is referred to as ‘the
system’.
16 UM_2.00A_EN
Safety OptiDist™
2 Safety
UM_2.00A_EN 17
OptiDist™ Safety
2.1 Certifications
The system and its components have been designed with safety in mind. The
design applies to the use, the conditions and the instructions described in the
documentation. The system
• Is CSA/cCSAus approved.
• Has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC rules.
• Meets the requirements of the European Directives concerning health,
hygiene and safety. The standards applied are listed in the “CE Declaration of
Conformity”.
If the user makes changes or modifications that are not expressly approved by
PAC, PAC reserves the right to void the user’s authority to operate the equipment.
WARNING
Failure to follow the instructions and procedures contained in this
manual could result in death or serious injury.
The instructions and procedures must be performed by qualified
and trained personnel only.
Do not perform installation or maintenance operations or proce-
dures outside the scope of this manual.
Please refer to the corresponding Material Safety Data Sheet (MSDS) for each
PAC-supplied product and also reference the corresponding MSDS that is vendor-
supplied for each sample and standard utilized with the product for the established
safety requirements and government safety and disposal standards.
18 UM_2.00A_EN
Safety OptiDist™
WARNING
General warning
WARNING
Danger of high voltages: shock hazard, electrocution hazard
WARNING
Fire hazard
WARNING
Explosion danger: release of pressure hazard
WARNING
Hot surfaces present: burn hazard
WARNING
Cold surfaces present: freeze hazard
WARNING
Danger to jam hands or feet
WARNING
Pinch danger
UM_2.00A_EN 19
OptiDist™ Safety
WARNING
Wedge present
WARNING
Sharp edges present
WARNING
Dangerous fluids and gases
CAUTION
General caution
CAUTION
Biohazard
CAUTION
Recycling
CAUTION
Wear a gas mask!
CAUTION
Wear protective gloves!
20 UM_2.00A_EN
Safety OptiDist™
CAUTION
Wear safety glasses!
NOTE
General note
TIP
General tip
TIP
Read the Service Manual
UM_2.00A_EN 21
OptiDist™ Safety
22 UM_2.00A_EN
Introduction OptiDist™
3 Introduction
UM_2.00A_EN 23
OptiDist™ Introduction
Any other use or modification of the apparatus without written agreement from
PAC will be regarded as noncompliant: PAC will not be liable for any possible
resulting damage or injury, and the warranty will be voided.
The apparatus may only be operated and serviced by trained qualified authorized
persons.
The operator must refrain from any operation that can affect the safety of the
system and of persons. The operator is always required to ensure that the
apparatus is in proper condition when it is operating.
The content of this document must be read, understood and observed in all points
and by each person responsible for the preparation, the transport, the storage, the
installation, the operation, and the maintenance of the instrument, before starting
to work with and on the instrument.
Utilizing advanced technology and a high quality design, the OptiDist™ allows
laboratory analysts of all skill levels to perform automatic atmospheric pressure
distillations, which comply with consensus group standard test methods, on a wide
variety of liquid hydrocarbon materials. Typically, these distillations are used to
characterize the tendency of a hydrocarbon fuel to vaporize. Vaporization
characteristics are a critical measure of overall fuel safety and performance.
The instrument offers several modes of operation that employ the power of
automation to significantly increase test productivity (sample throughput) while
also improving repeatability and reproducibility. In all modes, a large graphical
24 UM_2.00A_EN
Introduction OptiDist™
For fast and reliable quality control, analyzed samples can be instantaneously
compared to specifications programmable in the pre-stored product for in spec /
out-spec control. Up to 200 products can be preprogrammed and saved. Thus,
analysts can instantaneously see the information on how obtained results match
to pre-programmed typical specifications, i.e. typical temperature vs. volume or
volume vs. temperature, for fast product quality evaluation. The instrument can
automatically generate a Pass/Fail quality statement providing a detailed report.
UM_2.00A_EN 25
OptiDist™ Introduction
26 UM_2.00A_EN
Introduction OptiDist™
• Enhanced optics
• 24-bit high-resolution A to D converter
3.7 Terminology
/i
Dry Point (DP): Thermometer reading observed when the last drop of liquid evapo-
rates from the lowest point in the distillation flask. Drops or films of liq-
uid on the side of the distillation flask are disregarded.
Final Boiling Point (FBP) The maximum thermometer reading obtained during the test. Usually
occurs after the evaporation of all liquid from the bottom of the distilla-
tion flask. Also known as Maximum Temperature.
Initial Boiling Point (IBP) The temperature at the instant the first drop falls from the lower end of
the condenser tube.
Per Cent Residue % volume of residue left in the distillation flask (measured in accord-
ance with the standard in a 5 ml graduated cylinder).
Corrected temperature Temperature reading as describe above corrected for barometric pres-
reading sure at 101,3 kPa.
Per Cent recovery The volume measured at the end of the distillation if after a 2-minute
interval two successive observations agree (measured in accordance
with standard method).
Percent total Recovery The combined percent recovery and residue in the distillation flask. To
get percent loss deduct percent total recovery from 100.
Thermometer reading The temperature of the vapor measured in the neck of the distillation
flask below the vapor tube.
UM_2.00A_EN 27
OptiDist™ Introduction
°C Degrees Celsius
°F Degrees Fahrenheit
AN Application Note
CI: Cetan index
DI: Driveability (or distillation) Index
DP Dry Point
FBP Final Boiling Point
hPa hecto Pascal (1 hPa = 1 mbar)
IBP Initial Boiling Point
ICM Installation and Configuration Manual
ID Identity - Identification
IM Interface Manual
K Degrees Kelvin
kPA kilo Pascal (1 kPa = 10 mbar)
LCD Liquid Crystal Display
LED Light Emitting Diode
LIMS Lab Information Management System
mA milliamps
MM Maintenance Manual
mmHG millimeters of mercury column
MMI Man Machine Interface
OM Operator Manual
P/N Part Number
PAC Petroleum Analyzer Company, L.P.
PM Process Manual (System Description)
QC-DB Quality Control Database
SM Supervisor Manual
SPM Site preparation Manual
SS System Specification
Temp. Temperature
TFT Thin Film Transistor
VOC Volatile Organic Components
3.9 Nameplate
The nameplate is located on the rear of the OptiDist™ housing.
28 UM_2.00A_EN
Introduction OptiDist™
3.10 Specifications
/i
UM_2.00A_EN 29
OptiDist™ Introduction
Measurements
Vapor Temperature • Range : 0 °C to 450 °C (32 °F to 842 °F)
• Accuracy: Pt 100 IEC 751 probe Class A
• Built in calibration memory with 10 point calibration table and auto-
matic probe ID detection; calibration history; optional calibration certif-
icate
Sample Volume • Optical measuring system compatible with samples producing smoke
in the receiver
• Range: 0% to 103% of charge volume
• Resolution: 0,03 ml (internal)
• Accuracy: ± 0,1 ml
Ambient Pressure • Built-in pressure sensor
• Range: 70 to 110 kPa (500 to 800 mmHg)
• Calibration: Single point against reference barometer
Condenser Bath Tem- • Measuring instrument: Incorporated IC temperature sensor (Inte-
perature grated Circuit)
• Range: 0 °C to 65 °C (32 °F to 149 °F)
• Accuracy: ± 0,5 °C at 25 °C (± 0,9 °F at 77 °F)
Receiving Chamber • Measuring instrument: Incorporated IC temperature sensor (Inte-
Temperature grated Circuit)
• Range: 0 °C to 40 °C (32 °F to 104 °F)
• Accuracy: ± 0,5 °C (± 0,9 °F)
Heating element tem- • Thermocouple K
perature
Optional equipments • Built-in printer
• External printer
• Barcode reader
• External status indicator
• Ambient temperature sensor
• Automatic dry point kit for 200 ml
• Automatic dry point kit for 125 ml
• Option for reduced VOC emission
• Crude oil testing capability
• 200 ml test kit for 10% bottom residue
• Doctor Box
• CRM reference materials
30 UM_2.00A_EN
Description OptiDist™
4 Description
UM_2.00A_EN 31
OptiDist™ Description
A
B
C
D
K
E
J
F G
32 UM_2.00A_EN
Description OptiDist™
E
D
B C
UM_2.00A_EN 33
OptiDist™ Description
J
A
I
B
C
H
D
E
F
At the front of the OptiDist™, above the distillation flask heater platform (from top
to bottom)
• 8-pin DIN socket connector: Connection of the sample vapor temperature
probe
• 5-pin DIN socket connector: Connection of the optional automatic dry point
detection kit
34 UM_2.00A_EN
Description OptiDist™
At the rear of the OptiDist™, on the left top (from top to bottom)
• RS 232C serial port: Connection to a LIMS with user-customizable
communication protocol (CTS, ENQ, ACK, XON, XOFF)
• RJ45 Ethernet connector: Connection of the unit to the OptiDist™ network
At the rear of the OptiDist™, on the left top (from top to bottom)
• Hose nozzle: Fire extinguisher gas source connection
UM_2.00A_EN 35
OptiDist™ Description
During a distillation run, the Central Processing Unit (CPU) of the OptiDist™
controls and monitors the operation of the overall system by means of a user
designated distillation product program and data received from the heating block,
the condenser bath and the receiving chamber.
36 UM_2.00A_EN
Description OptiDist™
4.3.1 Overview
B J
C I
D H
E G
The distillation flask heater assembly, located to the left of the touch-screen,
breaks down as follows:
UM_2.00A_EN 37
OptiDist™ Description
• Distillation flask heater assembly enclosure door (G) for protection of the
operator from accidents and against air drafts draughts.
The distillation flask enclosure has a transparent glass door, at the front for
easy access to and inspection of the distillation flask. The transparent door
has been specially selected for its resistance to heat and shocks.
• A distillation flask stopper attachment device (A) assures that the distillation
flask is well positioned in the distillation flask heater platform (C).
• Two fire extinguisher injection slots (B) (one above and one under the heating
element). CO2 gas injection spray is triggered by two flame detectors (J)
situated on the internal right side of the distillation flask heater assembly
enclosure, above and under the heater platform.
• A fan (E) at the bottom of the distillation flask heater assembly is positioned
for rapid cooling of the distillation flask. For increase distillation flask cooling
efficiency the distillation flask heater platform (C) is automatically released in
low position at the end of test run for a better air distribution. This is performed
by energizing an electromagnet at the rear of the distillation flask heater
platform releasing the automatic positioning device of the heating platform.
The lower part of the distillation flask heater platform is hidden by housing,
while the upper part is easily accessible by opening the transparent door (G)
for placing the distillation flask.
The manually activated lever of the automatic positioning device of the heater
platform is used to lower or to raise the heater platform:
• Lower the flask heater platform positioning lever and the flask heater platform
comes in contact with the flask bottom (high position).
• Pull the flask heater platform positioning lever towards the top to release the
flask heater platform (low position) for flask insertion or removal.
A base plate or flask support board with centered hole, diameter 32 mm (1 1/4 in.),
38 mm (1 1/2 in.) or 50 mm (2 in.) is placed on the heater platform. The heater
platform is equipped with a base plate identification sensor that allows
confirmation that the right base plate is used according to the product group
selected and that the flask and base plate are well positioned.
38 UM_2.00A_EN
Description OptiDist™
preliminary trials and manual adjustments, the operator selects the test method,
enter the samples distillation group and starts the distillation by just pressing the
Start button.
Heater temperature monitoring and advanced mechanical design of the oven
assures excellent repeatability of distillation conditions in complete safety. Any
dependence on the heating element aging or on flask conditions are eliminated.
The optimized heating protects flask material from critical overheating improving
flask life-time and preventing against potential fire.
UM_2.00A_EN 39
OptiDist™ Description
C B
Distillation flask
A Flask neck
B Vapor tube sealing ring
C Vapor tube
The distillation flask provided is made of heat resistant glass and meets standard
specifications. Its capacity depends on the used standard method.
The distillation flask is equipped with a sealing ring on the distillation flask vapor
tube to ensure an airtight connection to the condenser.
Use only original PAC distillation flasks as replacement parts.
You have to use different sizes of the distillation flask according to the different
distillation standards. See the spare parts list.
40 UM_2.00A_EN
Description OptiDist™
UM_2.00A_EN 41
OptiDist™ Description
The vapor temperature probe is a Pt100 IEC 751 platinum probe class A with
built-in memory chip including probe correction values. This means the OptiDist™
automatically recognizes the probe and imports data at plug-in avoiding the need
to enter manually probe correction values and probe ID. Up to 5 calibration points
can be memorized (2 calibration points as standard).
Insert the probe wiring through the wire support atop the distillation flask heater
assembly and install the vapor temperature probe terminal plug into the connector
socket located on the top of the OptiDist™ above the distillation flask heater
platform. When not in use the temperature probe should be placed in the probe
holder atop the unit.
Before the distillation test, the vapor temperature probe must be inserted into the
distillation flask so that the probe is centered into the distillation flask: this is
ensured by the stopper assembly; and the depth of the vapor temperature probe
in the distillation flask must be precisely adjusted. To help the operator to position
the vapor temperature probe with a perfect repeatability, the stopper holder has
been specially designed as a gauge.
42 UM_2.00A_EN
Description OptiDist™
4.4.1 Overview
A O
B N
C M
L
K
J
E
F I
G H
UM_2.00A_EN 43
OptiDist™ Description
44 UM_2.00A_EN
Description OptiDist™
Receiver cylinder
A Receiver cylinder
The receiver cylinder provided with the OptiDist™, also called receiver or
graduated cylinder, meets standard test method specifications. The receiver
cylinder is used to
• measure the volume of the material to be distilled (exactly 100% by volume)
for the sample preparation
• for distillate recovery.
After the material is poured into the distillation flask, the receiver must be
positioned in the receiving chamber. The drop plate allowing the distillate to run
down the wall of the receiver is fixed directly to the end of the condenser tube
assuring it is not forgotten before test starting and it is well positioned. The
receiver stand positioning at the bottom of the receiving chamber assures the right
position of the receiver so that the drip tip automatically touches the wall of the
receiver.
If the glass cylinder is replaced after damage, take care that the glass cylinder is
completely on the base to ensure that the 100% volume mark is positioned
properly inside the cylinder (= 180 mm).
NOTE
A 200 ml receiver cylinder is available as optional accessories for
ASTM D189 tests. No re-calibration of the system is needed
between tests with 100 ml and 200 ml receiver cylinders; the stand-
ard method selected in the distillation product determines the cali-
bration scale.
UM_2.00A_EN 45
OptiDist™ Description
46 UM_2.00A_EN
Description OptiDist™
NOTE
The heat exchanging liquid level in the cooling system may have to
be serviced about every two years. The system will be alerted by
the level sensor of the reservoir and the failure alarm “Low cooling
liquid level” will be triggered.
The procedure for filling the cooling system is explained in the maintenance
manual.
The heat exchanging liquid supplied by PAC has been specially studied and
developed. Its thermal capacity, its viscosity at the working temperature, its anti-
corrosive and anti-foaming properties are particularly important for the correct
operation and preservation of the circuits.
NOTE
The heat exchanging liquid must be ordered at PAC (see the spare
parts list). It is important to use coolant provided by PAC for correct
operation of your OptiDist™, as its properties are particularly com-
patible with the application.
In case of problem in the cooling circuit, the circuit must be drained before
replacing any part (call for a qualified technician).
UM_2.00A_EN 47
OptiDist™ Description
4.5.3 Alarms
No parts of the OptiDist™ at dangerous voltages are accessible by the operator
(EN 61010-1).
Alarms are accompanied by a discontinuous audio alarm,. They remain visible as
long as the default persists; even if the alarm has been acknowledged.
Alarms warn the operator in the event of a malfunction whether the OptiDist™ is in
test or idle:
• Detection of a default during the test run:
- Opening of the receiving chamber door for example
• Component failure detection
• Detection of a safety state (either absolute or fixed by the user)
WARNING
Generally when the alarm buzzer is triggered, the operator must
display the cause and ensure that safety features are working cor-
rectly.
When a test is stopped the following occurs (if a power supply cutoff has not been
occurred before):
• The distillation flask heater is switched off
• The level follower goes to the upper or standby position
• The distillation flask cooling starts up
• The receiver chamber temperature is regulated to the standby value
(configured in the service parameters)
48 UM_2.00A_EN
Description OptiDist™
CAUTION
A fire extinguisher must always be available in the laboratory. If a
flame has been extinguished with CO2 gas, we strongly advise to
open the windows to air the laboratory.
When a fire is detected, the system injects extinguisher gas into the distillation
flask heater assembly, switches the only power consuming devices off and emits a
continuous loud alarm. Faults can be displayed on the screen by activating the
Message main menu: “Fire Detection” will appear. Switch off the unit to stop the
alarm.
NOTE
A serial contact is provided to connect the fire emergency signal to
an external “fire” alarm device.
NOTE
The gas supply is connected externally at the rear of the Opti-
Dist™.
NOTE
In order to make sure of the good working order of the system, it is
advised regularly testing the flame detector and the control sole-
noid valve of the CO2 inlet.
Time between controls depends on the OptiDist™ environment. The user and the
service personal will determine the ideal interval according to their observations
and their experience.
UM_2.00A_EN 49
OptiDist™ Description
CAUTION
If a flame has been detected and extinguished, the system must be
tested again.
WARNING
Repairs should only be carried out by a qualified technician.
WARNING
If one of the failures does not trigger the power supply off automati-
cally, switch off the main power and disconnect the OptiDist™ from
the mains.
50 UM_2.00A_EN
Description OptiDist™
4.6.2 Optional 125 ml and 200 ml automatic dry point detection kit
The dry point sensor must be installed in the distillation flask so that the point of
the sensor rests exactly at the center of the lowest point at the bottom of the
distillation flask. This is ensured by the specific stopper.
The automatic dry point sensor (DP sensor) consists of a long stainless steel
jacket with a thermocouple at the lower end to detect temperature differences.
NOTE
The type of dry point sensor (125 or 200 ml) is automatically recog-
nized at plug in by the OptiDist™.
NOTE
Boiling stones should not be used with the automatic dry point
detection kit.
4.6.2.1 Adjust the vapor temperature probe in the dry point stopper
Vapor temperature probe location and fixing
procedure in the dry point stopper: A
1. Loosen the threaded probe-sealing
B
device (compression nut) (B).
2. lnsert the vapor temperature probe (A)
through the threaded nut opening
3. Position the adjustment gauge (C) C
supplied against the dry point stopper.
4. Adjust the depth of the vapor temperature D
probe (A) so that the probe tip just
touches the bottom of the corresponding
notch.
5. Tighten the threaded compression nut until the vapor temperature probe (A) is
securely fixed at the proper depth.
6. Confirm that the vapor temperature probe (A) is properly sealed in place by
grasping the metal sheath at the probe base and applying a gentle twisting
motion.
UM_2.00A_EN 51
OptiDist™ Description
The vapor temperature probe should be securely fixed in position and not be
easily moved.
lf the vapor temperature probe is not held securely, repeat the adjustment
procedure.
52 UM_2.00A_EN
Description OptiDist™
4.6.3 Optional kit for “10 % (V/V) Distillation Residue” with 200 ml
The optional kit available to get the 10% Distillation Residue for Carbon Residue
measurement with 200 ml consists of
• a 200 ml receiver cylinder
• a drip tip expansion
• a 250 ml distillation flask
NOTE
The Alan® and HLIS-32® kits are optional hardware and software
packages (see the spare parts list).
Refer to the corresponding software Installation and Getting
Started Manual for requirements and instructions.
These two software are multitasking software suitable for use with 32-bit system
such as Windows©. They allow several OptiDist™ to be connected to the same
PC. Besides multi-OptiDist™s results retrieving and database management (and
according to the OptiDist™ type: run control, alarms displaying, internal
parameters management, etc.), ALAN® and HLIS-32® software allow transmitting
results directly to a LIMS (Laboratory Information Management System).
Each program comprises several modules. The main program operates as the
control center. This component of the software makes the functions available with
all types of respective brand units in the same way. The OptiDist™ specific
program module can thus be added to the main program if already installed in
your laboratory.
UM_2.00A_EN 53
OptiDist™ Description
PDS™ collects test results from your PAC instruments and stores this data in an
SQL database.
With the integrated Online Manager you can define your operation, tests and
instruments. You can also specify users, access privileges and test result
parameters.
The Workbench, a further component of PDS™, communicates with the
instruments: it controls the execution of tests and gathers and analyzes real-time
data.
Each PAC instrument has a PAC instrument plug-in associated with it.
Instrument plug-ins extend the funcionality of the Workbench by providing features
that are specific to the corresponding instrument.
PDS™ comes in two versions:
• a stand-alone version
• a Client-Server version
The status indicator beacon has to be mounted at the rear of the OptiDist™.
54 UM_2.00A_EN
Description OptiDist™
NOTE
The printer must support PCL5 emulation language (or above such
as PCL6 or variants such as PCL5e).
A bar code reader is available as optional accessory (see the spare part list). The
bar code reader allows to enter the sample name or number during test
configuration without spelling mistakes. When connected, the bar code reader can
be activated after selecting the sample name/number field.
Doctor Box 1
UM_2.00A_EN 55
OptiDist™ Description
The Doctor Box allows a rapid diagnostic of the OptiDist™ components and
electronic boards without need for OptiDist™ disassembly.
F F1 F2 30A
U BIP 24V FLASK HT
S VOLTMETER
E F3 F4 F5 F6 OUTPUT
S 24V
12V 12V
5V
T
R
5V 3.3V
A RED COND. HT BOOST HT REC . FAN LED
F
F
I 3.3V
C COOL . UNITT COOL . SV
ORANGE
L ADA DOCTOR BOX
I
G DOWN SOL. COND. REC.
H
T
GREEN
FLASK FAN REC . HT VOC PUMP
Doctor Box 2
The Doctor Box can be connected to the 25-pin connector on the right side of the
OptiDist™.
WARNING
Do not apply vacuum to the VOC outlet.
56 UM_2.00A_EN
Description OptiDist™
NOTE
When the instrument is shipped, the default passwords are:
• for operator (Level 0): no password
• for supervisor (Level 1): PAC-1
• for maintenance (Level 2): PAC-2
NOTE
To change the password, you need to know the previous password.
UM_2.00A_EN 57
OptiDist™ Description
NOTE
The system administrator should set-up these passwords during
installation to prevent unauthorized persons from changing the set-
tings of the instrument. In this way, access to important operations
can be restricted to a limited number of persons.
Set-up of the passwords of access levels is covered in the supervisor manual.
58 UM_2.00A_EN
Description OptiDist™
4.8.1 Touch-screen
The preloaded software provides a LCD touch-screen Man Machine Interface
(MMI) with a large (640 X 480) high-resolution graphic color display .
A number of large area function keys allow short cuts to most frequently used
functions for comfortable operation. User / machine dialogue is carried out via a
virtual keyboard, a virtual keypad and location sensitive graphics located on the
touch-screen.
The system is protected from unauthorized access and entry by various security
levels that allows changes to be made to the system settings only after a valid
password has been entered.
UM_2.00A_EN 59
OptiDist™ Description
60 UM_2.00A_EN
Description OptiDist™
Esc 7 8 9 Bcks
Del 4 5 6 Ins
123 1 2 3 ENT
<_ 0 +- . _>
The keypad screen is typical for all numerical entry. It appears each time a
numerical entry field is activated to enter a new value or to change settings by
entering the desired numbers by pressing the 0-9 number keys.
• ESC: Quit numerical entry and return to the previous screen
• BCKSP: Removal of the character positioned on the left of the cursor
• DEL: Removal of the character positioned above the cursor
• INS: Insertion of a character on the left of the cursor / replacement of the value
above the cursor
• ENT: Registration of the new value into the corresponding field
• < > Displacement of the cursor to the left / to the right
• + - Commutation of the new value as a negative value
• 123/789: Modification of the numerical keys order
UM_2.00A_EN 61
OptiDist™ Description
The Welcome screen provides menu buttons at the bottom of the screen allowing
to access to the following screens (from the left to the right):
• The PAC addresses screen to contact us.
• This display remains active until the EXIT button on bottom right is pressed:
Doing this displays the OptiDist™ Run /Start & Display screen allowing
starting a test immediately.
NOTE
The touch screen is factory calibrated but if it is necessary to cali-
brate it again, press five times on the welcome screen (not on the
buttons) to display the calibration screen of the touch screen.
62 UM_2.00A_EN
Description OptiDist™
UM_2.00A_EN 63
OptiDist™ Description
/i
/i
The top of the Run / Start & Display screen shows the status of the unit and is always visible
Status field The Status field indicates the state of the unit:
• Green / READY: The OptiDist™ is ready to start a new test
• Yellow / VOLUME SCAN: The OptiDist™ is waiting for a receiver cylin-
der with sample to measure the charge volume of the sample in the
receiver cylinder
• Yellow / START: Checking of test starting conditions
• Yellow / INIT HEAT: Initial heating – The OptiDist™ is searching the IBP
• Yellow / DISTILLATION: The distillation is in progress
• Yellow / FINAL HEAT: The OptiDist™ is waiting for the FBP (from 5 ml
residual liquid in the flask to the FBP)
• Yellow / RECOVERY: After the FBP, the OptiDist™ is waiting for the vol-
ume in the receiver stabilizes
• Yellow / RESIDUE: Expectation of the input of the recovery
PC link: State indicator of the OptiDist™ / Network link
• 9 (green): Network link established
• X (red): Network link not established
Time display Displays the current time
Unit No.: Name or number of the OptiDist™ unit
/i
The centre left of the Run / Start & Display screen is the operation screen
Operator: Selection of the operator name in the operator name list
Note: The operator name is not mandatory for test starting
Sample No.: Entry field of the sample identifying name or number / Selection of a sample
name in the prestored list of sample names
Product: Selection of a test product in the prestored list of test products
Note: Addition of a comment if wanted
/i
The centre right of the Run / Start & Display screen shows the test monitoring diagram
On the left of the • Actual vapor temperature
diagram • Actual temperature of the heating element
In the centre of the • Actual temperature of the condenser tube
diagram
On the right of the • Volume of condensate in the receiver
diagram • Actual temperature of the receiving chamber
64 UM_2.00A_EN
Description OptiDist™
/i
The bottom of the Run / Start & Display screen is the menu bar with main menu buttons
New Test Set-up of a new test
Results Access to the result database
Products Access to the product database
QC Access to the Quality Control database
Messages Access to the alarm messages database
More Access to more main menu options
Second screen with main menu options of the Run /Start & Display screen
/i
More main menu options of the Run / Start & Display screen
Param. Access to the OptiDist™ internal parameter database
Reports Access to the report profile database
Operator Access to the operator name list
Standard Access to the standard method database
More Access to more main menu options
UM_2.00A_EN 65
OptiDist™ Description
Third screen with main menu options of the Run /Start & Display screen
/i
More main menu options of the Run / Start & Display screen
Calib. Access to the calibration main menu
Diag. Access to the diagnostic main menu
About Access to the information screen on firmware version, serial numbers of elec-
tronic boards and MMI
Backup Access to the data download and upload main menu
More Returns to the first screen with main menu options
66 UM_2.00A_EN
Description OptiDist™
The Run / Start & Display screen with state indicators of starting conditions
The Run / Start & Display screen is reset and the New Test button turns into
START button allowing a new test to be started when it will be set up and when all
starting conditions will be fulfilled.
The starting conditions with state indicators are displayed on the left bottom above
the Start button on the Run / Start & Display screen.
/i
If a condition is not fulfilled, this will be indicated with the following icon “X” in red
next to the condition in question.
UM_2.00A_EN 67
OptiDist™ Description
68 UM_2.00A_EN
Installation OptiDist™
5 Installation
UM_2.00A_EN 69
OptiDist™ Installation
Space requirements:
• Physical dimensions: 440 mm (W) x 570 mm (D) x 650 mm (H) / (17
x 20.1 x 26 in)
• Weight: 68 kg (188lbs)
• Place the OptiDist™ on a level and flat laboratory bench, free of
vibrations, away from drafts and close to electric mains connections,
CO2 gas supply and communication connections. In any case, the
OptiDist™ location should be draught free and preferably near a
fume hood.
• Leave sufficient space behind the OptiDist™ for cabling access and
for clear ventilation: Allow a minimum clearance of 200 mm (8 in)
around the OptiDist™ for circulation of air for the cooling unit.
• Distance between two units: 200 mm (8 in)
• Protect the OptiDist™ from water and aqueous mists and spray (see
LCD handling precautions further on).
• Protect the OptiDist™ from direct sunlight
Gas requirements • Fire extinguisher gas: CO2 (CO2 inlet at rear)
• Gas pressure at inlet: 7 bars (+/-0.5bars) / 700 kPa (+/-50 kPa)
• Maxi. safe pressure: 10 bars / 1000 kPa
• Flow of gas during flame quenching: 30 to 35 liter/s
• Average volume used for flame extinguishing: 150 liters
• Extinction takes between 2 and 4 seconds
• Fire extinguisher connection: hose nozzle with a compression fer-
rule with inside diameter of 8 mm
NOTE
Testing with the OptiDist™ must be carried out at constant ambient
pressure. Care must therefore be taken to avoid placing the Opti-
Dist™ under an air vent or under a fume cupboard functioning dur-
ing a test.
70 UM_2.00A_EN
Installation OptiDist™
NOTE
Do not return the OptiDist™ or any part to the factory without fac-
tory authorization.
After unpacking, verify the unit and its accessories as well as any possible
damage sustained in transit, which must immediately brought to the attention of
the carrier so that a statement of damage can be made.
The various parts of the OptiDist™ are carefully verified and tested before
shipping. Nevertheless, it is worth verifying that the equipment received
corresponds to the packing list enclosed.
On taking delivery of the OptiDist™, unpack all these parts.
CAUTION
Do not put the OptiDist™ in the direct sunlight.
CAUTION
Do not cover the air intake slots of the unit. If you want to set up
many units in rows, be sure that you keep minimum 20 cm distance
(8 inches) between units. Failure to allow sufficient air ventilation
flow can cause instrument damage.
Put the OptiDist™ on a level, flat workbench and not subject to vibrations, near
electrical sockets and near connections to fire extinguisher gas supply. Provide
enough space so that the OptiDist™ can be operated conveniently and for access
to the rear connectors.
Leave sufficient space behind the OptiDist™ for cabling access and for clear
ventilation.
Place the external PC close to the instrument (the length of the Ethernet cable
provided to connect the PC with the OptiDist™ is about 3 m but it is possible to
use a longer one).
UM_2.00A_EN 71
OptiDist™ Installation
NOTE
The CO2 connection system is composed of a hose nozzle with a
compression ferrule with inside diameter of 8 mm to connect a rigid
metallic pipe. Adapt a hose nipple on the hose nozzle to connect a
flexible hose if it is necessary.
2. Connect the CO2 gas source to the connector at the rear of the OptiDist™.
It is possible to connect an external fire alarm device to the OptiDist™. Call for a
qualified technician to perform the connection. For connection instructions refer to
the Maintenance Manual.
NOTE
The RS 232C communication mode needs to be configured.
Parameters are accessible from the setup menu of the internal
parameters.
RS232 software and hardware is provided as standard.
72 UM_2.00A_EN
Installation OptiDist™
1. Connect the vapor temperatur probe to the connector on the front of the
OptiDist™.
2. Insert and adjust the vapor probe in the flask stopper.
3. Pass the vapor probe wire through the wire support.
5.9 125 ml and 200 ml automatic dry point kit and holder
The automatic dry point kit and holder is an optional accessory that has to be
ordered separately.
The dry point sensor must be installed in the distillation flask so that the point of
the sensor rests exactly at the center of the lowest point at the bottom of the
distillation flask. This is ensured by the specific dry point stopper.
The automatic dry point sensor (DP sensor) consists of a long stainless steel
jacket with a thermocouple at the lower end to detect temperature differences.
A Wire support
B Dry point sensor connector C A
C Flask stopper holder
D Dry point stopper
E Dry point stopper holder
UM_2.00A_EN 73
OptiDist™ Installation
5. Insert the vapor temperatur probe and adjust the probe length in the dry point
stopper, see 5.9.2. Adjustment of the vapor probe in the dry point stopper.
6. Install the dry point stopper (D) to the dry point stopper holder (E).
7. Connect the dry point sensor to the dry point sensor connector (B) and pass
the wire through the wire support (A).
NOTE
The type of dry point sensor (125 ml or 200 ml) is automatically rec-
ognized at plug in by the OptiDist™.
NOTE
Boiling stones should not be used with the automatic dry point
detection kit.
74 UM_2.00A_EN
Installation OptiDist™
NOTE
When connected, the external printer needs to be configured in the
OptiDist™ internal parameter settings.
See printer documentation provided for explanations on the opera-
tion.
UM_2.00A_EN 75
OptiDist™ Installation
CAUTION
Once the OptiDist™ has been installed, before switching on, allow
the OptiDist™ to adjust to the ambient temperature in the labora-
tory for 24 hours (especially if the OptiDist™ was stored at low tem-
perature).
1. Push the Main Power switch on the right side of the OptiDist™ to the up or ON
position.
The small green indicator lamp on the front side lights up indicating the
OptiDist™ is powered on and the following operations are performed:
• A self diagnostic test is started.
• The condenser bath, receiving chamber and cooling system temperatures
are regulated to default values.
• After the self dignostic test has completed, the welcome screen appears,
see 4.8.5. The Welcome screen.
1. By pressing the PAC button at the bottom of the welcome screen, the PAC
addresses screen appears giving the list of PAC contacts.
2. Press the EXIT button to return to the welcome screen.
76 UM_2.00A_EN
Installation OptiDist™
NOTE
The touch screen is factory calibrated but if it is necessary to cali-
brate it again, press five times on the welcome screen (not on the
buttons) to display the calibration screen of the touch screen.
UM_2.00A_EN 77
OptiDist™ Installation
78 UM_2.00A_EN
Operation OptiDist™
6 Operation
UM_2.00A_EN 79
OptiDist™ Operation
6.1 Preparation
CAUTION
Wear safety glasses when handling dangerous chemical such as
solvents.
CAUTION
Wear protective gloves when handling dangerous chemical such
as solvents.
NOTE
Clean the distillation flask, the receiving cylinder and the vapor
temperature probe regularly.
Otherwise the boiling range characteristics are influenced.
6.1.2 Safety
This OptiDist™ has been carefully designed, manufactured and inspected for
quality. It has been equipped with a number of safety features.
However, the use of this OptiDist™ may involve the handling of solvents,
chemicals, and other potentially dangerous flammable, toxic, etc.) materials.
Please exercise caution when handling these materials while operating the
OptiDist™.
• read the manual
• wear proper protective clothing
• perform all suggested service procedures
• use care to prevent accidents
The manufacturer accepts no responsibility for any damage or liability arising from
the use of the OptiDist™.
Use of Non-PAC Products and Accessories. Defects or damage that result from
the use of non-PAC branded or certified products, accessories, software or other
peripheral equipment are excluded from warranty.
NOTE
Use only standard OptiDist™ PAC receiving cylinders. If it is not the
case, the OptiDist™ does not detect the receiving cylinder and the
alarm “No receiver detected” is triggered.
80 UM_2.00A_EN
Operation OptiDist™
NOTE
Use duran glass distillation flasks for a maximum heating tempera-
ture of 700 °C to 870 °C.
NOTE
Use quartz glass distillation flasks for a maximum heating temper-
ature of 870 °C to 1000 °C.
WARNING
Handle all solvents, chemicals and other potentially dangerous
materials with care.
The operator needs to have training about potential hazards,
before operating the system.
NOTE
To obtain accurate and repeatable results, prepare the samples
and the OptiDist™ with particular care as described in the following
sections.
1. Clean and dry all instruments and accessories for use before sampling. See
6.1.4. Clean the devices.
2. Determine the group characteristics of the sample according table 2 of the
ASTM D86 test method.
/i
UM_2.00A_EN 81
OptiDist™ Operation
TIP
Referral: Consult table 3 of the ASTM D86 standard test method.
/i
AUnder certain circumstances, samples can also be stored at temperatures below 20 °C (68 °F).
See also 7.3.2 and 7.3.3.
B If sample is to be immediately tested and is already at the temperature prescribed in table 1, see
7.4.1.1.
C
If sample is (semi)-solid at ambient temperature, see also 10.3.1.1.
D
If sample is known to be wet, resampling may be omitted. Dry sample in accordance with 7.5.2
and 7.5.3.
82 UM_2.00A_EN
Operation OptiDist™
NOTE
For group 3 and 4, if the sample has solidified during storage, heat
the sample to a temperature of 9 to 21 °C above its pour point prior
to analysis. Then mix it gently prior to opening the sample con-
tainer to ensure homogeneity.
CAUTION
Samples of materials that visibly contain water are not suitable for
testing.
If the sample is not dry, obtain another sample that is free from sus-
pended water.
If such a sample cannot be obtained, remove the suspended water
by adding anhydrous sodium sulfate and then separating it from the
drying agent by decanting (see standard method). Note in the
report that the sample has been dried by the addition of a
desiccant.
UM_2.00A_EN 83
OptiDist™ Operation
CAUTION
Wear safety glasses when handling dangerous chemical such as
solvents.
CAUTION
Wear protective gloves when handling dangerous chemical such
as solvents.
NOTE
Clean the distillation flask, the receiving cylinder and the vapor
temperature probe regularly.
Otherwise the boiling range characteristics are influenced.
84 UM_2.00A_EN
Operation OptiDist™
UM_2.00A_EN 85
OptiDist™ Operation
1. Choose the appropriate base plate with centered opening and diameter as
directed by the standard for the indicated group.
TIP
Referral: Consult table 1 of the ASTM D86 standard test method.
2. Install the base plate on the distillation flask heater platform: position the cut
corner to the right and to the back of the distillation flask heater platform when
you are facing the OptiDist™.
NOTE
When the distillation flask heater is installed and the distillation
flask heater platform will be in elevated or "run" position, the open-
ing diameter of the base plate will be indicated in front of the base
plate state indicator in the starting condition list of the Run / Start &
Display screen.
86 UM_2.00A_EN
Operation OptiDist™
6.1.5.3 Adjust the vapor temperature probe in the distillation flask stopper
NOTE
The distillation flask stopper assembly is specially designed for the
precise adjustment of the vapor temperature probe positioning in
the stopper (see below) and so, in the distillation flask.
NOTE
When the vapor temperature probe is connected to the OptiDist™,
the "Probe" state indicator in the list of starting conditions of the
Run /Start & Display screen becomes green. In the contrary case, a
test cannot start.
NOTE
If a dry point kit is used, refer to manual for the adjustment of the
dry point probe.
UM_2.00A_EN 87
OptiDist™ Operation
NOTE
Be sure that the sample, the receiver, and the distillation flask are at
the temperature specified by the standard test method. Otherwise,
the recovery determination will give incorrect results.
1. Put a few boiling stones in the distillation flask to prevent a boiling retardation.
2. Pour the content of the receiver cylinder into the distillation flask.
Be sure that the vapor tube of the flask is pointed upward, otherwise material
could run out through the vapor tube.
NOTE
• Do not use any boiling stones for dry point tests.
• Do not clean the receiver cylinder after pouring the sample
into the flask, or the result will be false!
NOTE
Be sure that the vapor temperature probe has been adjusted
with the stopper assembly holder.
7. Pull the flask heater platform positioning lever to release the flask heater
platform in low position.
8. With your left hand, open the door of the flask heater enclosure door.
88 UM_2.00A_EN
Operation OptiDist™
9. With the right hand, position the flask with the stopper in a vertical position in
the flask heater platform:
a. Insert the vapor tube of the flask in the condenser tube opening.
b. Insert the flask stopper attachment device into the attachment device of
the heater assembly.
NOTE
Tightness between the flask vapor tube and the condenser tube
opening is ensured by the vapor tube sealing ring, which prevents
any escape of vapor during the distillation.
NOTE
The diameter of the hole in the base plate and the presence of a
flask in the flask heater platform are parts of the starting conditions
list displayed in the Run /Start & Display screen. These are tested
when the heater platform is in high position. If the diameter of the
hole in the base plate corresponds to the requirements of the
standard test method selected for the sample group to be tested,
the two corresponding fields becomes green and the diameter of
the hole is automatically displayed in front of the “Base plate:” state
indicator.
NOTE
Do not clean the receiver cylinder after pouring the sample into the
flask, or the result will be false!
1. Press the door latch to open the door of the receiving chamber.
2. Place the receiver in the receiving chamber:
a. First insert the lower end of the condenser tube into the receiver cylinder;
b. Then position the bottom of the receiver cylinder on the receiver stand. It
is automatically centered when the receiver chamber door is closed.
3. Make sure that the tip of the drop plate touches the front inner wall of the
cylinder so that the drops run down the wall.
If the drops fall freely into the liquid in the cylinder, the volume measurement
will not work properly.
4. Check that the rubber cap has well gone down onto the top of the receiver. If
this is not the case, push down the rubber cap to cover the top of the receiver.
UM_2.00A_EN 89
OptiDist™ Operation
NOTE
The OptiDist™ detects the presence of the receiver in the receiving
chamber: the “Receiver:” state indicator in the starting condition list
of the Run /Start & Display screen becomes green indicating the
receiver is well positioned. In the contrary case, a test cannot start.
NOTE
The manual adjustment of the receiving chamber temperature
before test starting is part of OptiDist™ internal parameters and
can be activated with a level 1-access authorization (Supervisor).
lf the parameter "Adjust receiver at start" is set on "Yes", the operator can set-up
another receiving chamber temperature thän the one set in the product before
starting the test.
In this case, when the New Test button is pressed, a yellow dialog box appears
allowing:
• To enter the desired receiver temperature
• To select the current ambient temperature as receiver temperature if an
ambient temperature probe is connected to the OptiDist™
• To cancel the adjustment procedure by pushing the CANCEL button
1. Enter the desired receiver temperature value in the field right to the text
2. Press the ENTER button to validate.
3. Press the OK button in the dialog box to validate.
1. Or: Press the CANCEL button to use the receiving chamber temperature set
in the product.
2. Or: If an ambient temperature probe is connected to the OptiDist™, press on
the OK button in front of the ambient temperature displayed to use the
ambient temperature as eceiver temperature.
3. Prepare and install the distillation flask with sample and the receiver.
90 UM_2.00A_EN
Operation OptiDist™
NOTE
The measurement of the charge volume by the OptiDist™ before
test starting is part of OptiDist™ internal parameters and can be
activated with a level 1-access authorization (Supervisor).
If the OptiDist™ internal parameter "Scan charge volume" is set on "Yes", the
OptiDist™ measures the charqe volume in the receiving cylinder and calculates it
to 100% volume for more precise results.
In this case, when the operator presses the New Test button of the Run / Start &
Display screen or after the adjustment of the receiving chamber temperature (if
activated), a yellow dialog box appears indicating the operator can place the
receiving cylinder with sample in the receiving chamber to measure the exact
volume of sample in the receiving cylinder.
NOTE
The New Test button turns into Stop button allowing the operator
aborting the scan.
The Stop button will then turn into Start button allowing starting the
test.
NOTE
When the volume of sample is correct or by aborting the procedure,
the Stop button turns into Start button allowing the operator to start
the test.
5. Prepare and install the distillation flask with sample and the receiving cylinder.
UM_2.00A_EN 91
OptiDist™ Operation
NOTE
When the vapor temperature probe is connected to the OptiDist™,
the "Probe" state indicator in the list of starting conditions of the
Run /Start & Display screen becomes green indicating this starting
condition is fulfilled. ln the contrary case, a test cannot start.
92 UM_2.00A_EN
Operation OptiDist™
6.2 Measurements
This guide allows to understand rapidly the various functionalities offered by the
OptiDist™ interface.
NOTE
The use of the Start Guide is factory activated and can be disabled
in the OptiDist™ internal parameters with a level 1-access authori-
zation (Supervisor).
1. Press the New Test button on the Run / Start & Display screen.
UM_2.00A_EN 93
OptiDist™ Operation
NOTE
According to the OptiDist™ internal settings, a dialogue box
appears when the New Test button is pressed for the adjustment of
the receiver chamber temperature during test or for the measure-
ment of the sample charge volume.
NOTE
The Start Guide can be deactivated at any time by pressing the Exit
start guide button but data entered are saved.
94 UM_2.00A_EN
Operation OptiDist™
NOTE
The next screens of the Start Guide mode contain the additional
Back button.
For access to the previous screen of the Start Guide mode press
the Back button.
First, the operator must select an operator name in the list prestored in the unit.
This list also contains the operator names of the last tests performed with the unit
(up to 100).
1. Press on the field corresponding to the operator name desired to select it.
2. Select an operator name in the list prestored in the unit.
3. Press the Continue start procedure button to continue.
NOTE
lf the operator presses the Continue start procedure button without
entering or selecting a name, the operator name of the last test
performed is entered by default.
UM_2.00A_EN 95
OptiDist™ Operation
NOTE
It is possible to modify the sample name selected: Press on the
Sample No. entry field, the virtual keyboard appears allowing a
manual modification.
.
96 UM_2.00A_EN
Operation OptiDist™
NOTE
Passing to the next screen of the Start Guide is only possible if a
sample name or number is entered.
NOTE
lf the operator presses the Continue start procedure button without
entering or selecting a name, the sample name of the last test per-
formed is entered by default.
NOTE
If (in the previous step) the sample name has been selected in the
prestored list, the corresponding product is proposed by default.
1. Press on the Get product list button to display the list of the distillation
products stored in the unit.
2. Select a distillation product in the list prestored in the OptiDist™ (If needed or
if it is the first test since the OptiDist™ switching on).
UM_2.00A_EN 97
OptiDist™ Operation
3. Press on the distillation product desired to select it: the product name is
automatically entered in the “Product” entry field.
4. Press on the Continue start procedure button to continue.
Data entered previously appear on the left of the screen. Only the state indicators
of starting conditions not fulfilled indicated with the following icon “X ” in red are
displayed to make the reading of the screen easier for the operator.
They indicate that the test cannot be started now.
NOTE
The operator can still prepare the accessories and the OptiDist™
as indicated in the right of the screen.
98 UM_2.00A_EN
Operation OptiDist™
NOTE
The diameter of the base plate installed can only be detected when
the distillation flask heater platform is in high position with the distil-
lation flask installed in the heater assembly and a product is
selected.
The distillation flask filled with sample and the receiver cylinder can now be set up
in the OptiDist™.
The operator will be able to start the test when all starting conditions will be
validated: the Start button will appear in the left bottom of the screen.
1. Press the Start button to start the test.
UM_2.00A_EN 99
OptiDist™ Operation
NOTE
The test-starting mode is part of OptiDist™ internal parameters and
can be modified with a level 1-access authorization (Supervisor).
1. Press the New Test button in the left bottom of the Run / Start & Display
Screen.
When the New Test button is pressed, the operator name, the sample name and
the product used for running the previous test are automatically recalled in the
dedicated field and the New Test button turns into Start button allowing a new test
to be started immediately. The product list is immediately displayed.
If needed, the user can modify all parameters in the same way that in “Normal”
mode.
The OptiDist™ can also be set to increment automatically the sample number. For
this, the “Auto increment sample no.” internal parameter must be set on “Yes”.
100 UM_2.00A_EN
Operation OptiDist™
NOTE
If it is the first test since the OptiDist™ is switched on, a product
must be selected.
NOTE
According to the OptiDist™ internal settings, a dialogue box can
appear when the New Test button is pressed for the adjustment of
the receiver chamber temperature during test or for the measure-
ment of the sample charge volume. These cases are handled in the
section “Setup of the receiver chamber”.
If cleaning and preparation operations have been already done and correctly
performed, the following starting conditions are validated:
/i
If a condition is not fulfilled, this will be indicated with the following icon “X” in red
next to the condition in question.
NOTE
The operator can still prepare the accessories and the OptiDist™
during/after the test setup.
NOTE
Diameter of the base plate installed can only be detected when the
distillation flask heater platform is in high position with the distilla-
tion flask installed in the heater assembly and a product is
selected.
NOTE
The operator can still prepare the accessories and the OptiDist™.
UM_2.00A_EN 101
OptiDist™ Operation
NOTE
The Start button of the Run / Start & Display screen, or the New
Test button if the test is initiated from the Start Guide, turns into
Stop button allowing the operator to stop the test at any time.
102 UM_2.00A_EN
Operation OptiDist™
NOTE
The test-starting mode is part of OptiDist™ internal parameters and
can be modified with a level 1-access authorization (Supervisor).
1. Press the New Test button in the left bottom of the Run / Start & Display
screen.
The Run / Start & Display screen is reset and the New Test button turns into
START button allowing a new test to be started when it will be set up and
when all starting conditions will be fulfilled.
NOTE
According to the OptiDist™ internal settings, a dialogue box can
appear when the New Test button is pressed for the adjustment of
the receiver chamber temperature during test or for the measure-
ment of the sample charge volume.
If cleaning and preparation operations have been already done and correctly
performed, the following starting conditions are validated:
UM_2.00A_EN 103
OptiDist™ Operation
/i
If a condition is not fulfilled, this will be indicated with the following icon “X” in red
next to the condition in question.
NOTE
The operator can still prepare the accessories and the OptiDist™
during/after the test setup.
104 UM_2.00A_EN
Operation OptiDist™
NOTE
It is possible to modify a sample name selected in the prestored
list: Press on the “Sample No.” entry field, the virtual keyboard
appears allowing a manual modification.
UM_2.00A_EN 105
OptiDist™ Operation
106 UM_2.00A_EN
Operation OptiDist™
NOTE
The diameter of the base plate installed can only be detected when
the distillation flask heater platform is in high position with the distil-
lation flask installed in the heater assembly and a product is
selected.
The operator will be able to start the test when all starting conditions will be
validated.
2. Press the Start button to start the test.
UM_2.00A_EN 107
OptiDist™ Operation
NOTE
When the distillation flask will be installed and the distillation flask
heater platform will be in elevated or "run" position, the opening
diameter of the base plate will be indicated in front of the "Base
plate" state indicator in the starting condition list of the Run /Start &
Display screen.
1. Press the New Test button in the left bottom of the OptiDist™ Run /Start &
Display screen then set up the test.
NOTE
According to the configuration of the OptiDist™ internal parame-
ters, the test starting will be done either with the Start Guide or
directly in the Run /Start & Display screen in Normal mode or in
Quick mode.
NOTE
The test can only start if
• all starting conditions, like cleaning operations and proper
installation of appropriate accessories, are fulfilled.
• the receiver cylinder positioned in the receiving chamber is
empty.
NOTE
The test can not start if the condenser tube is not cleaned.
Cleaning is a part of the OptiDist™ internal parameters: parameter
“Show condenser cleaning”.
If another swab than the one provided with the unit is used, test
start refusal can be disabled.
If not all starting conditions are fulfilled, the starting condition(s) not fulfilled will be
indicated in the left bottom of the last screen of the Start Guide.
In case of a normal or a quick test starting, the test starting conditions not fulfilled,
are identified in the left bottom of the Run / Start & Display screen by a cross in a
red box.
108 UM_2.00A_EN
Operation OptiDist™
UM_2.00A_EN 109
OptiDist™ Operation
WARNING
Let the distillation flask cool before removing it. Check the vapor
temperature displayed in the test-monitoring diagram.
The result is saved in the OptiDist™. If the residue is automatically input by the
OptiDist™, the result is immediately displayed according to the parameters of the
report profile set in the product. The result can also be printed and/or sent to the
PC/LIMS. If the residue should be manually input or measured by the OptiDist™,
a dialogue box appears indicating the procedure to follow.
According to the “Residue measurement” parameter set in the product data base
selected to perform the test, the residue can be:
• Predicted by the OptiDist™ using the optimizer.
• Constant (a default residue is entered in the product)
• None (set in the product)
CAUTION
Wait for the distillation flask is cooled before removing it for the res-
idue measurement. Check the vapor temperature displayed in the
test-monitoring diagram.
A yellow box appears under the test-monitoring diagram at the end of test allowing
entering the residue manually using the virtual keypad.
110 UM_2.00A_EN
Operation OptiDist™
WARNING
Wait for the distillation flask is cooled before removing it for the res-
idue measurement. Check the vapor temperature displayed in the
test-monitoring diagram.
1. Spill the distillation flask residue in the receiver cylinder and replace the
receiver cylinder in the receiving chamber.
2. Close the receiving chamber door then press the OK button in the dialogue
box.
To display and print the test result, refer to 7.1. The Results database.
1. Press the Results button on the Run / Start & Display screen.
UM_2.00A_EN 111
OptiDist™ Operation
112 UM_2.00A_EN
Database management OptiDist™
7 Database management
UM_2.00A_EN 113
OptiDist™ Database management
NOTE
• To transmit test results, to modify or to delete them, a level 1
access (Supervisor) or level 2 access (Maintenance) is
required.
• With level 0 access, the operator can only view and print the
test results.
114 UM_2.00A_EN
Database management OptiDist™
/i
View button: Displaying of the list of results stored in the internal memory
Left/right arrow buttons to scroll through the up to 100 pages with results stored in
the internal memory of the OptiDist™
NOTE
Information displayed is presented by default according to the
parameters of the report profile set in the product used to perform
the test.
UM_2.00A_EN 115
OptiDist™ Database management
View list of results with tested sample name and corresponding date:
1. Press the View button in the Results main menu.
The following screen appears:
116 UM_2.00A_EN
Database management OptiDist™
/i
All these parameters are set in the report profiles and results are printed according
to the report profile set in the product used to perform the test.
lt is possible to print test result automatically after the end of each test.
UM_2.00A_EN 117
OptiDist™ Database management
Send results to PC
118 UM_2.00A_EN
Database management OptiDist™
UM_2.00A_EN 119
OptiDist™ Database management
5. Validate by pressing the Store button or quit without validating by pressing the
No Store button.
120 UM_2.00A_EN
Database management OptiDist™
UM_2.00A_EN 121
OptiDist™ Database management
8. Validate by pressing the Store button or quit without validating by pressing the
No Store button.
122 UM_2.00A_EN
Database management OptiDist™
4. Validate by pressing the Yes button or quit without validating by pressing the
No button.
UM_2.00A_EN 123
OptiDist™ Database management
NOTE
• To modify, delete, copy and create products, a level 1 access
(Supervisor) or level 2 access (Maintenance) is required.
• With level 0 access, the operator can only view and print the
parameters of the preset products.
Exception:
• A Supervisor with level 1 access can give authorisation to the
operator in the Parameter main menu by setting the Parameter
“Product modif. by operator: = yes”. The operator can then
modify products with the following limitations:
- Product modifications are only effective for one test
- Product modifications will not be stored in the product data-
base
- The operator cannot create new products to be stored in
the database
- The operator cannot delete prestored products from the
database
124 UM_2.00A_EN
Database management OptiDist™
/i
Copy button: Creation of a new product by making a copy of the one displayed
Standard button: Accessing to the list of prestored standard test methods and
D850/D1078 compounds
Left/right arrow buttons to scroll through the eight pages with all parameter settings
for each product.
Exit button: Return to the Run / Start & Display screen
/i
UM_2.00A_EN 125
OptiDist™ Database management
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further product parameter setting options.
The following screen appears:
126 UM_2.00A_EN
Database management OptiDist™
/i
UM_2.00A_EN 127
OptiDist™ Database management
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further product parameter setting options.
The following screen appears:
128 UM_2.00A_EN
Database management OptiDist™
/i
UM_2.00A_EN 129
OptiDist™ Database management
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further product parameter setting options.
The following screen appears:
/i
130 UM_2.00A_EN
Database management OptiDist™
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further product parameter setting options.
The following screen appears:
NOTE
In the case of the distillation of a QC sample, specifications limits
define the points recorded in the QC sample database.
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further product parameter setting options.
The following screen appears:
UM_2.00A_EN 131
OptiDist™ Database management
/i
Parameters on the fifth and sixth page of the Products main menu display
Spec. limits for defined • Observed: Percent recovered specified
temperatures: • Pressure corrected: Volume specified according to corrected
temperature readings
• Evaporated: Percent evaporated specified
Temperature: Set-up of the defined temperature for accepted minimum and maximum
volume value
Mini vol / Maxi vol: Set-up of the accepted minimum and maximum volume value
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further product parameter setting options.
The following screen appears:
132 UM_2.00A_EN
Database management OptiDist™
/i
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further product parameter setting options.
The following screen appears:
UM_2.00A_EN 133
OptiDist™ Database management
/i
134 UM_2.00A_EN
Database management OptiDist™
NOTE
• To modify the QC database or to modify a quality product, a
level 1 access (Supervisor) or level 2 access (Maintenance) is
required.
• With level 0 access, the operator can only view and print
parameters and information diplayed on the QC main menu
screen.
UM_2.00A_EN 135
OptiDist™ Database management
Left/right arrow buttons to scroll through the two pages of the QC main menu.
136 UM_2.00A_EN
Database management OptiDist™
/i
/i
where {x1, x2, ..., xN} is the sample and x is the mean of the sample
Green/red colours in The colour of the listed lines and of the value indicates if the result is within
the list or outside of the specification limits:
• Complete line is green: All specified values of the QC sample run are
within the specification limits
• Line is red and only the measured value is green: One or more speci-
fied values of the QC sample run is outside of the specification limits but
the specified value displayed is within the specification limits
UM_2.00A_EN 137
OptiDist™ Database management
138 UM_2.00A_EN
Database management OptiDist™
NOTE
When a specification of a Quality Product is modified, if a point is
added, it will be recorded in the next quality records. If a point is
deleted, it will be no more available for quality control.
UM_2.00A_EN 139
OptiDist™ Database management
If an alarm occurs, the alarm review is displayed on the alarm diplay screen. In
this case, the status bar on the centre-top of the screen changes from green to red
and indicates “ERROR”.
• To quit the alarm, press the OK button.
• To return to the Run / Start & Display screen, press the Exit button.
140 UM_2.00A_EN
Database management OptiDist™
The alarm history list can store up to 200 alarm messages including date, time,
alarm code and alarm message.
The last alarm message of the list is the last message saved. When the memory is
full, the OptiDist™ automatically deletes the first message saved (the oldest).
• To print the alarm history list press the Print button.
• To exit the alarm history list and return to the Run / Start & Display screen
press the Exit button.
UM_2.00A_EN 141
OptiDist™ Database management
NOTE
• To modify the parameter settings, a level 1 access (Supervisor)
or level 2 access (Maintenance) is required.
• With level 0 access, the operator can only view and print
parameter settings.
142 UM_2.00A_EN
Database management OptiDist™
/i
Left/right arrow buttons to scroll through the five pages with all internal parameter
settings for the OptiDist™.
/i
Unit parameters on the first page of the Parameter main menu display
Start Mode: Selection of the default test-starting mode
• Guided: Set-up and starting of a new test using the Start Guide.
• Normal: Complete set-up and starting of a new test from the
Run / Start & Display screen with the Normal Start Mode.
• Quick: Starting of a new test with the Quick Start Mode from the
Run / Start & Display screen with the parameters of the last sam-
ple tested.
Ambient sensor connected: Measurement of the ambient temperature with the ambient tempera-
ture probe (available as option). If “Yes” is selected and no probe is
connected, an alarm is released indicating the probe is faulty.
Adjust receiver at start: Request to enter the desired receiver chamber temperature which
should be the same as the charge temperature of the sample
Pressure Unit: Selection of the pressure measurement unit used by default (hPa,
kPa, mBar, Bar, mmHg or psi).
Autoincrement sample no.: Last sample number is always proposed when starting a new test. If
this parameter is set to “Yes”, the sample number is automatically
incremented avoiding to have the same sample number in the result
database. If the sample ID is selected in the list stored in the Opti-
Dist™, the ID will be incremented when selected
Autoclear sample no.: Sample number field will be cleared at test start
Check cond. cleaning: Checking or not of the condenser cleaning before test starting (part of
the test starting conditions)
Scan charge volume: Measuring of the charge volume in the receiver before test starting
Product modif. by operator: • If this parameter is set to “No”, the operator cannot modify pro-
ducts
• If this parameter is set to “Yes”, the operator can modify products
with the following limitations:
Temp. unit: Selection of the temperature measurement unit used by default by the
OptiDist™
UM_2.00A_EN 143
OptiDist™ Database management
Unit parameters on the first page of the Parameter main menu display
Touch beep: Beep when a button is pressed
Lanuage: Selection of the language for the user interface of the OptiDist™ (Eng-
lish, German or French)
Screen saver: Screen saver on or off
S. Saver time: Time before the Screen saver starts
NOTE
From each page of the Parameter main menu it is always possible
to return to the Run / Start & Display screen by pressing the Exit
button.
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further parameter setting options for the OptiDist™.
The following screen appears:
144 UM_2.00A_EN
Database management OptiDist™
/i
Unit parameters on the second page of the Parameter main menu display
Access level reset time: Set-up of the time during which the Supervisor or the Service access
level is active. If the touch screen does not detect any contact up to
the end of this time, the access level is reset to the Operator level (this
time is not taken into account in Diagnostic mode, during a test or dur-
ing calibration). The countdown is reset at each contact.
Startup access level: Selection of the access level activated at the analyzer switching-on.
Supervisor password: Set-up of the password of the Supervisor access level – 5 characters
minimum.
Maintenance password: Set-up of the password of the Maintenance access level – 5 charac-
ters minimum.
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further parameter setting options for the OptiDist™.
The following screen appears:
UM_2.00A_EN 145
OptiDist™ Database management
/i
Unit parameters on the third page of the Parameter main menu display
Unitname: Set-up of the unit name which is shown on start screen and print out.
Serial No.: Displaying of the unit serial number.
Network configuration: Set-up of the network used
• Stand alone unit: Use of the unit as an independent OptiDist™
• PC controlled: Networking of the unit with the new PC software
• ALAN controlled: Networking of the unit with Alan
• HLIS controlled: Networking of the unit with HLIS
Unit Id: Id of the unit in the network
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further parameter setting options for the OptiDist™.
The following screen appears:
Unit parameters on the fourth page of the Parameter main menu display
Printer: Selection of the printer used
• Internal: Printing of data with the built-in printer
• External: Printing of data with an external printer connected to
the USB A connector (located at the rear of the unit).
Lines per page: Set-up of the number of lines per page.
146 UM_2.00A_EN
Database management OptiDist™
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further parameter setting options for the OptiDist™.
The following screen appears:
Unit parameters on the fifth page of the Parameter main menu display
Ethernet settings
Server IP address: Set-up of the Server IP address to which the OptiDist™ is connected.
Contact your network administrator (requires level 2 access)
Unit IP-Address: Set-up of the unit IP address. Contact your network administrator
(requires level 2 access)
automatic / manual: Set-up of the address manually / automatically (provided by the sys-
tem)
RS232 settings
Baudrate: Set-up of the rate of transfer
Data bit: Set-up of the data bit
Parity bid: Set-up of the parity bit
Stop bit: Set-up of the stop bit
Protocol: Set-up of the communication protocol used (xon-xoff, hardware or
none)
UM_2.00A_EN 147
OptiDist™ Database management
Left/right arrow buttons to scroll through the five pages with all internal parameter
settings for the OptiDist™.
2. If not logged in with level 2 access, press on the dedicated field to enter the
maintenance password. The virtual keypad appears.
3. Use the right arrow button to scroll trough the pages until the sixth page
appears:
148 UM_2.00A_EN
Database management OptiDist™
/i
The service screen of the Parameter main menu contains the following menus:
Reset calibration Left and right arrow keys allowing scrolling through the set-up
pages of OptiDist™ internal parameters
Delete all products Return to the Run / Start & Display screen
Delete all reports Removal of pre stored report profile list
Delete all operators Removal of the pre stored operator name list
Delete all standards Removal of the pre stored standard method list
Delete all compounds 850 Removal of all D850 compound list
Delete all compounds 1078 Removal of all D1078 compound list
Delete all results Removal of all results stored in the OptiDist™ memory
Delete all QC results Removal of all results stored in the Quality Control database
Reset supervisor password Removal of the supervisor password entered and replacement
by the factory set password
Reset maintenance password Removal of the maintenance password entered and replacement
by the factory set password
UM_2.00A_EN 149
OptiDist™ Database management
The service screen of the Parameter main menu contains the following menus:
Reset unit parameters Initialization of internal parameters with factory set parameters
Up/Download mode Preparation of the OptiDist™ for the uploading of data (controlled
from the PC) and OptiDist™ software update.
Print Setup of control parameters to be transmitted to the RS 232C
serial port during a test. Code numbers from 1 to 6 are specifying
the range of the control parameters to be transmitted. For details
about the code numbers refer to the Service Manual.
150 UM_2.00A_EN
Database management OptiDist™
NOTE
• To modify and create new report profiles, a level 1 access
(Supervisor) or level 2 access (Maintenance) is required.
• With level 0 access, the operator can only view the report pro-
files.
The OptiDist™ can store up to 40 report profiles displayed in a list located on the
left of the screen:
UM_2.00A_EN 151
OptiDist™ Database management
/i
Copy button: Creation of a new report by making a copy of the one displayed
Left/right arrow buttons to scroll through the six pages with all parameter settings
for each report.
/i
NOTE
From each page of the Reports main menu it is always possible to
return to the Run / Start & Display screen by pressing the Exit but-
ton.
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further report profile settings.
The following screen appears:
152 UM_2.00A_EN
Database management OptiDist™
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further report profile settings.
The following screen appears:
UM_2.00A_EN 153
OptiDist™ Database management
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further report profile settings.
The following screen appears:
154 UM_2.00A_EN
Database management OptiDist™
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further report profile settings.
The following screen appears:
UM_2.00A_EN 155
OptiDist™ Database management
Settings for LIMS transfer on the fifth page of the Reports main menu
Transfer: Selection of the type of transfer to the LIMS:
• Automatic: Automatic transfer of the result at the end of test.
• Transfer of the result on request of the operator.
Protocol name: Keying of the name of the communication protocol (LIMS header).
Separator: Selection of the character dividing each group of data.
End: Selection of the character sent at each end of string:
• CR: Carriage return
• CR + LF: Carriage return + line feed
• LF + CR: Line feed then carriage return
Product: Transfer of the product parameters.
Spec. limits: Transfer of limit values set in the specification pages of the product.
Table: Sending of the spreadsheet following requirements of the desired standard
test method (ASTM D86 only or following the operator needs (every 1%,
5% or 10%).
Boiling range: Transfer of the difference between the IBP and the FBP (in Kelvin).
Stand. limits: Transfer of limit values set in the standard specification page of the product.
Dist. rate: Transfer of the distillation rate.
Heating: Transfer of the heating temperature.
Computation: Transfer of the DI, CI and TVL20 value.
156 UM_2.00A_EN
Database management OptiDist™
1. Use the right arrow button at the bottom of the screen to scroll to the next
page with further report profile settings.
The following screen appears:
Settings for LIMS transfer on the sixth page of the Reports main menu
Temp. unit: Selection of the unit of temperature readings transferred (°C or °F).
Press. Unit: Selection of the unit of atmospheric pressure measures transferred (hPa,
kPa, mBar, Bar, mmHg or psi).
Barom. corr.: Transfer of temperature readings corrected for barometric pressure.
Evap. corr.: Transfer of percent evaporated.
Table format: Transfer of volumetric readings following requirements of the desired stand-
ard test method or following the operator needs (every 1%, 5% or 10%).
Rounding: Adjust the precision of nearest vapor temperature.
UM_2.00A_EN 157
OptiDist™ Database management
NOTE
• To add, remove or modify operator names in the list of
prestored operators, a level 1 access (Supervisor) or level 2
access (Maintenance) is required.
• With level 0 access, the operator can only view the list of
prestored operators.
158 UM_2.00A_EN
Database management OptiDist™
/i
1. Press the Add button to add a new operator name to the list.
2. Press on the Operator name field to display the virtual keyboard, enter the
new operator name and then press the ENTER button.
The Operator main menu - add a new name to the operator list
3. Validate by pressing the Store button or quit without validating by pressing the
No Store button.
If the Store button was pressed, the new name appears on the operator list.
UM_2.00A_EN 159
OptiDist™ Database management
The Operator main menu - new name added to the operator list
4. Press the Exit button to return to the Run / Start & Display screen.
160 UM_2.00A_EN
Database management OptiDist™
The Operator main menu - delete a name from the operator list
3. Validate by pressing the Yes button or quit without validating by pressing the
No button.
4. Press the Exit button to return to the Run / Start & Display screen.
NOTE
Should the parameters of an existing standard test method be offi-
cially modified, the software should be updated.
UM_2.00A_EN 161
OptiDist™ Database management
NOTE
• To add, remove or modify standard test methods in the list, a
level 1 access (Supervisor) or level 2 access (Maintenance) is
required.
• With level 0 access, the operator can only view and print the list
of standard test methods.
162 UM_2.00A_EN
Database management OptiDist™
/i
Copy button: Creation of a new standard test method by making a copy of the
one displayed
Print button: Printing of the parameters of the standard test method displayed
Only appears when a D1078 standard test method is selected – Access to the
list and set-up menu of D1078 compounds
Only appears when a D850 standard test method is selected – Access to the list
and setup menu of D850 compounds
Exit button: Return to the Run / Start & Display screen
/i
UM_2.00A_EN 163
OptiDist™ Database management
164 UM_2.00A_EN
Database management OptiDist™
NOTE
• To modify or remove compounds, a level 1 access (Supervisor)
or level 2 access (Maintenance) is required.
• With level 0 access, the operator can only view and print the
parameters of compounds.
UM_2.00A_EN 165
OptiDist™ Database management
166 UM_2.00A_EN
Database management OptiDist™
/i
The setup screens of the prestored compounds lists contain the following buttons:
Removal of the compound displayed / selected
/i
UM_2.00A_EN 167
OptiDist™ Database management
7.10.1 Introduction
The OptiDist™ measurement instrumentation requires regular verification. How
often depends on the user, but it is recommended that vapor temperature and
volume measurement system are verified at least once a year.
This section includes instructions for calibration of measurement circuits and
component parameters of your OptiDist™, which are critical to distillation
accuracy.
These circuits and parameters were calibrated prior to shipment from the factory,
but must be checked at regular intervals (see 8.1. Frequency of measurement
checks) and readjusted if necessary.
Calibration of the OptiDist™ is in great part dedicated to the supervisor and
service staff.
NOTE
• To calibrate the measurement circuits and components of the
OptiDist™, a level 1 access (Supervisor) is required.
• With level 2 access (Maintenance) the following additional
functions are available:
- the Service buttons
- the Temperature Correction Table submenu
• An Operator with level 0 access can only:
- calibrate the cursor sensivity of the touch panel
- calibrate the backlight of the screen
- view and print information which is displayed on the Cali-
bration main menu
168 UM_2.00A_EN
Database management OptiDist™
UM_2.00A_EN 169
OptiDist™ Database management
NOTE
For the calibration of the barometric pressure sensor, a reference
barometer indicating absolute pressure values (not values cor-
rected to sea level) is required.
Only absolute values are acceptable for the barometric pressure
calibration.
1. Press the Barometric Pressure button of the Calibration main menu. The
following screen appears:
2. Press the Modif button to unlock the setting: the Modif button turns into Save
button allowing registering modifications.
3. Select the desired pressure unit by pressing on the entry field corresponding
to Barometric Pressure Unit to scroll through the various possibilities (Pa,
hPa, kPa, mBar, Bar, HG, PSI or Torr).
4. Enter the reference barometer identity by pressing on the entry field
corresponding to Reference Barometer ID. This calls up the virtual keyboard
allowing to enter a name or a serial number.
• Press the ENTER button to validate the entry;
• or press the ESC button to abort the modification and to go back to the
previous screen.
170 UM_2.00A_EN
Database management OptiDist™
5. Enter the actual ambient barometric pressure by pressing on the entry field
corresponding to Reference Read. This calls up the virtual keypad allowing to
enter the barometric pressure reading from the reference barometer.
• Press the ENTER button to validate the new value;
• or press the ESC button to abort the modification and to go back to the
previous screen.
7. Configure the OptiDist™ to refuse or allow a test start when the calibration is
overdue by pressing on the entry field corresponding to Refuse test:
• Yes: the test start is refused
• No: the test can be started
8. Press the Save button to register the new settings or the Exit button to abort
and to go back to the Calibration main menu.
The OptiDist™ corrects this deviation using the "Vapor Probe Calibration Table".
The vapor temperature probe supplied with the OptiDist™ is equipped with a built-
in memory circuit that can store up to 10 calibration points, the probe identity and
its serial number. The probe correction values are signed (+/-) and rounded to the
nearest 0,01 °C.
The OptiDist™ detects the probe at plug-in and automatically downloads these
calibration parameters into the instrument memory. This guarantees the correct
calibration values for each instrument even if the vapor temperature probe is
replaced or exchanged between different units.
NOTE
A PAC probe with memory for calibration data is required to start a
test.
The Vapor Probe menu enables the display of these values and the user can
modify them if needed.
UM_2.00A_EN 171
OptiDist™ Database management
1. Press the Vapor Probe button of the Calibration main menu. The following
screen appears:
2. Press the Modif button to unlock the setting: the Modif button turns into Save
button allowing registering modifications.
NOTE
The Init button allows the initialization of the vapor probe table with
the factory set of correction values (correction values correspond-
ing to the first probe supplied with the OptiDist™).
172 UM_2.00A_EN
Database management OptiDist™
3. Select the entry field to modify the calibration frequency of the probe on the
entry field corresponding to Freq. This calls up the virtual keypad.
• Press the ENTER button to validate the new value;
• or press the ESC button to abort the modification and to go back to the
previous screen.
4. Configure the OptiDist™ to refuse or allow a test start when the calibration of
the probe is overdue by pressing on the entry field corresponding to Refuse:
• Yes: the test start is refused
• No: the test can be started
6. Then press the entry field corresponding to Mv (Measured value). This calls
up again the virtual keypad. Enter the temperature measured by the probe for
the given reference temperature.
UM_2.00A_EN 173
OptiDist™ Database management
You get these values from the calibration certificate delivered with your probe
or from another calibration service. Press the ENTER button to validate it. The
correction is automatically transferred into the memory and displayed in the
vapor probe table.
7. To delete a correction point, press on the entry field corresponding to Item on
the right bottom of the screen. This calls up the virtual keypad. Enter the
number of the line to be deleted and then:
• press the ENTER button to validate the new value;
• or press the ESC button to abort the modification and to go back to the
previous screen.
8. To choose another temperature unit and to convert the values entered in the
vapor probe table, press on the box indicating the temperature unit used
above the table (scroll selection: °C/ °F).
9. Press the Save button to register the new settings or the Exit button to abort
and to go back to the Calibration main menu.
174 UM_2.00A_EN
Database management OptiDist™
3. To keep the present values press the Save button or continue with entering of
new values.
4. Enter the identity of the PS400 probe simulator used by pressing on the entry
field corresponding to "Probe simulator". This calls up the virtual keyboard
allowing to enter a name.
• Press the ENTER button to validate the entry;
• or press the ESC button to abort the modification and to go back to the
previous screen.
5. Enter the serial number of the PS400 probe simulator used by pressing on the
entry field corresponding to S/N. This calls up the virtual keypad.
• Press the ENTER button to validate the new value;
• or press the ESC button to abort the modification and to go back to the
previous screen.
UM_2.00A_EN 175
OptiDist™ Database management
7. Configure the OptiDist™ to refuse or allow a test start when the calibration is
overdue by pressing on the entry field corresponding to Refuse test:
• Yes: the test start is refused
• No: the test can be started
8. Having filled these 4 fields, disconnect the vapor temperature probe on the
front of the OptiDist™ and replace it by the PS400 probe simulator as
described on the screen then press the Continue button to start the
calibration. The following screen appears:
176 UM_2.00A_EN
Database management OptiDist™
15. Enter the resistance value corresponding to this temperature position on your
PS400 probe simulator (you get these values from the Quality Control
Certificate delivered with the PS400 probe simulator).
16. Press the Next button to go to the next step.
17. Press the Save button to register the new settings or the Exit button to abort
and to go back to the Calibration main menu.
NOTE
For access to the service buttons, level 2 access (Maintenance) is
required.
1. If not logged in with level 2 access, press on the dedicated field to enter the
maintenance password. The virtual keypad appears.
2. Press successively the buttons More > More > Calib. in the Run / Start &
Display screen.
The Calibration main menu opens (see 7.10. The Calibration database).
3. On the Calibration main menu screen, press the Service button next to the
Vapor Probe circuit button.The following screen appears:
UM_2.00A_EN 177
OptiDist™ Database management
/i
Entry fields on the service screen of the Vapor Probe Circuit submenu
Freq Calibration frequency of the vapor temperature measurement circuit
Refuse test Refuse or allow a test start when the calibration frequency is overdue
• Yes: the test start is refused
• No: the test can be started
Resistance Information about the resistance value corresponding to 0 °C/32 °F on the
PS400 probe simulator used for last calibration
AD Points Information about calculated AD points corresponding to 0 °C/32 °F on the
PS400 probe simulator used for last calibration
Resistance Information about the resistance value corresponding to 400 °C/752 °F on the
PS400 probe simulator used for last calibration
AD Points Information about calculated AD points corresponding to 400 °C/752 °F on the
PS400 probe simulator used for last calibration
NOTE
Use only standard OptiDist™ PAC receiver cylinders. If it is not the
case, the OptiDist™ does not detect the receiver cylinder and the
alarm “No receiver detected” is triggered.
1. Press the Volume measurement button of the Calibration main menu. The first
screen of the Volume Measurement Calibration submenu appears.
178 UM_2.00A_EN
Database management OptiDist™
UM_2.00A_EN 179
OptiDist™ Database management
3. To keep the present values press the Save button or continue with entering of
new values.
4. Enter the identity of the gauge used: press on the entry field corresponding to
Receiver, this calls up the virtual keyboard allowing keying a name or a
number.
• Press the ENTER button to validate the entry;
• or press the ESC button to abort the modification and to go back to the
previous screen.
5. Enter the serial number of the gauge used: press on the entry field
corresponding to S/N, this calls up the virtual keypad.
• Press the ENTER button to validate the entry;
• or press the ESC button to abort the modification and to go back to the
previous screen.
7. Configure the OptiDist™ to refuse or allow a test start when the calibration/
verification is overdue by pressing on the entry field corresponding to Refuse
test:
• Yes: the test start is refused
• No: the test can be started
8. Press the Continue button to perform the volume calibration or the Exit button
to abort and to go back to the Calibration main menu.
9. If you press the Continue button, the next screen of the Volume Measurement
Calibration menu appears:
180 UM_2.00A_EN
Database management OptiDist™
10. As explained in the display, place a standard OptiDist™ PAC receiver cylinder
containing the volume calibration gauge in the receiving chamber.
NOTE
The slots of the volume calibration gauge need to show to the rear
of the receiving chamber.
11. Close the receiving chamber door, then press the Continue button.
12. If you press the Continue button, the next screen of the Volume Measurement
Calibration menu appears and the OptiDist™ is now calibrating the volume
measurement system by scanning the volume calibration gauge:
UM_2.00A_EN 181
OptiDist™ Database management
182 UM_2.00A_EN
Database management OptiDist™
14. Press the Save button to register the new settings or the Exit button to abort
and to go back to the Calibration main menu.
NOTE
For access to the service buttons, level 2 access (Maintenance) is
required.
1. If not logged in with level 2 access, press on the dedicated field to enter the
maintenance password. The virtual keypad appears.
2. Press successively the buttons More > More > Calib. in the Run / Start &
Display screen.
The Calibration main menu opens (see 7.10. The Calibration database).
3. On the Calibration main menu screen, press the Service button next to the
Volume measurement button.
The following screen appears:
UM_2.00A_EN 183
OptiDist™ Database management
/i
NOTE
For access to the temperature correction table, level 2 access
(Maintenance) is required.
1. If not logged in with level 2 access, press on the dedicated field to enter the
maintenance password. The virtual keypad appears.
2. Press successively the buttons More > More > Calib. in the Run / Start &
Display screen.
The Calibration main menu opens (see 7.10. The Calibration database).
3. On the Calibration main menu screen, press the Tmp Correction Table button.
The following screen appears:
184 UM_2.00A_EN
Database management OptiDist™
4. Press the Modif button to unlock the setting: the Modif button turns into Save
button allowing registering modifications.
UM_2.00A_EN 185
OptiDist™ Database management
NOTE
Pushing the Save button will update the calibration date, even if
nothing was changed.
NOTE
The Init button allows the initialization of the temperature correction
table with the factory set of correction values.
5. Press the Init button in order to delete all values in the Temperature Correction
Table.
6. After the values in the Temperature Correction Table are deleted, press the
Save button to save the changes.
7. Press the Exit button to go back to the Calibration main menu, then press
again the Exit button to go back to the Run / Start & Display screen.
8. Press successively the buttons More > More > Diag. on the Run / Start &
Display screen to open the Diagnostic main menu (see 8.2. Displaying current
readings).
9. Disconnect the vapor temperature probe on the front of the OptiDist™ and
replace it by the PS400 probe simulator.
10. Set the rotary switch of the PS400 probe simulator to the 50 °C (122 °F)
position.
11. Check the measured temperature value displayed next to Vapor Tmp (first
box on top left of the Diagnostic screen).
12. Wait a few seconds until the temperature reading has stabilized and write
down the measured temperature value displayed next to Vapor Tmp.
13. Then set the rotary switch of the PS400 probe simulator successively to the
positions 100 °C (212 °F), 200 °C (392 °F), 300 °C (572 °F) and 400 °C (752
°F) and write down the displayed corresponding measured temperature value
for each of the rotary switch positions.
14. Press the Exit button of the Diagnostic screen to go back to the Run / Start &
Display screen.
15. Press successively the buttons More > More > Calib. in the Run / Start &
Display screen to return to the Calibration main menu.
16. Press the Tmp Correction Table button and then press the Modif button to
unlock the setting: the Modif button turns into Save button allowing registering
modifications.
17. Enter or modify the name of the temperature correction table by pressing on
the entry field corresponding to Table Name. This calls up the virtual keyboard
allowing keying a name or a number.
186 UM_2.00A_EN
Database management OptiDist™
20. Then press on the entry field corresponding to Mv (Measured value). This
calls up again the virtual keypad.
21. Enter the previously written down temperature values measured by the unit
for the given reference temperature.
• Press the ENTER button to validate the entry (the correction is
automatically transferred in internal parameters of the OptiDist™ and
displayed in the Temperature Correction Table);
• or press the ESC button to abort the modification and to go back to the
previous screen.
22. Repeat the above steps until all values in lines 1 to 5 of the Temperature
Correction Table are entered: reference temperature values (Rt) and
corresponding measured values (Mv) according to the PS400 probe simulator
rotary switch positions: 50 °C (122 °F), 100 °C (212 °F), 200 °C (392 °F), 300
°C (572 °F) and 400 °C (752 °F).
23. To delete a correction, press on the entry field corresponding to Item under
Delete a Correction on the right bottom of the screen. This calls up the virtual
keypad. Enter the number of the line to be deleted and then:
• press the ENTER button to validate the entry;
• or press the ESC button to abort the modification and to go back to the
previous screen.
24. To choose another temperature unit and to convert the values entered in the
unit correction table, press on the box indicating the unit used above the table
(scroll selection: °C / °F).
25. After all values are correctly entered, press the Save button to save the
settings or press the Exit button to cancel the action and to return to the
Calibration main menu.
UM_2.00A_EN 187
OptiDist™ Database management
2. Press successively the buttons More > More > Diag. on the Run / Start &
Display screen to open the Diagnostic main menu (see 8.2. Displaying current
readings).
3. Make sure that the PS400 probe simulator is still connected.
4. Set the rotary switch of the PS400 probe simulator to the 50 °C (122 °F)
position.
5. Wait a few seconds until the temperature reading has stabilized and check the
measured temperature value displayed next to Vapor Tmp (first box on top left
of the Diagnostic screen).
NOTE
In comparison with the stated temperature values in the Quality
Control Certificate (delivered with the PS400 probe simulator), the
deviation of the corresponding measured temperature values must
be less than 0,2 °C (0.36 °F).
6. Then set the rotary switch of the PS400 probe simulator successively to the
positions 100 °C (212 °F), 200 °C (392 °F), 300 °C (572 °F) and 400 °C (752
°F) and check the corresponding measured temperature value for each of the
rotary switch positions.
7. If one of the measured temperature values is out of the allowable tolerance,
repeat the calibration of the temperature correction table.
8. To go back to the Run / Start & Display screen, press the Exit button of the
Diagnostic screen.
188 UM_2.00A_EN
Database management OptiDist™
NOTE
For the calibration of the measurement circuit of the heating ele-
ment temperature a reference thermometer is required.
CAUTION
The flask heater temperature must be calibrated when not ener-
gized.
1. Press the Heater Temperature button of the Calibration main menu. The
following screen appears:
2. Press the Modif button to unlock the setting: the Modif button turns into Save
button allowing registering modifications.
3. Select the temperature unit by pressing on the entry field corresponding to
Temperature Unit to scroll through the various possibilities (scroll selection: °C
or °F).
4. Enter the reference thermometer identity by pressing on the entry field
corresponding to Reference Thermometer ID. This calls up the virtual
keyboard allowing to enter a name or a number.
UM_2.00A_EN 189
OptiDist™ Database management
5. Position the bulb of the reference thermometer in the center of the heating
element next to the temperature measurement thermocouple.
6. Once the temperature reading on the reference thermometer is stable, enter it
by pressing on the entry field corresponding to Reference Read. This calls up
the virtual keypad allowing keying the heating element temperature read on
the reference thermometer.
• Press the ENTER button to validate the new value;
• or press the ESC button to abort the modification and to go back to the
previous screen.
7. Press the Save button to register the settings or the Exit button to abort and to
go back to the Calibration main menu.
NOTE
For access to the service buttons, level 2 access (Maintenance) is
required.
1. If not logged in with level 2 access, press on the dedicated field to enter the
maintenance password. The virtual keypad appears.
2. Press successively the buttons More > More > Calib. in the Run / Start &
Display screen.
The Calibration main menu opens (see 7.10. The Calibration database).
3. On the Calibration main menu screen, press the Service button next to the
Heater Temperature button.
The following screen appears:
190 UM_2.00A_EN
Database management OptiDist™
Entry fields on the service screen of the Heater Temperature Calibration submenu
Reference Tmp: Entry field for the reference temperature measured by a reference thermome-
ter. The bulb of the reference thermometer must be placed in the center of the
heating element next to the temperature measurement thermocouple.
Correction: Correction of the temperature will be calculated automatically.
UM_2.00A_EN 191
OptiDist™ Database management
CAUTION
Use pure Toluene or ISO-Octane and pure Hexadecane
(purity > 95% minimum).
NOTE
The optimizer calibration is not performed from the Calibration
main menu.
If you activate the optimizer menu, a screen appears indicating that
the optimizer calibration is done by distilling two pure samples and
these distillations must be started from the Run / Start & Display
screen.
1. Press the Exit button to go back to the Calibration main menu then press
again the Exit button to go back to the Run / Start & Display screen.
2. Prepare a sample of Toluene and then prepare the OptiDist™.
3. Press the New Test button then parameterize the test: select the product
OPTIMIZER CAL 1 for the distillation of the Toluene or ISO-Octane sample.
NOTE
When the OPTIMIZER CAL product is validated, the Start button
turns into Calibration button indicating the following distillation is a
calibration run.
4. Insert the sample and the receiver cylinder then press the Calibration button
to start the distillation run.
NOTE
For access to the service buttons, level 2 access (Maintenance) is
required.
1. If not logged in with level 2 access, press on the dedicated field to enter the
maintenance password. The virtual keypad appears.
2. Press successively the buttons More > More > Calib. in the Run / Start &
Display screen.
The Calibration main menu opens (see 7.10. The Calibration database).
192 UM_2.00A_EN
Database management OptiDist™
3. On the Calibration main menu screen, press the Service button next to the
Optimizer button.
The following screen appears:
UM_2.00A_EN 193
OptiDist™ Database management
194 UM_2.00A_EN
Database management OptiDist™
3. Press the Save button to save the settings or press the Exit button to cancel
the action and to return to the Calibration main menu.
UM_2.00A_EN 195
OptiDist™ Database management
The laboratory supervisor has the possibility to configure the calibration frequency
of the following measurement circuits and devices:
• The vapor temperature measurement circuit:
• The vapor temperature probe correction values
• The receiving chamber volume measurement circuit
It is possible to configure the OptiDist™ to deny a test start if the date set for the
calibration is passed if the customer’s Quality Assurance Program requires it.
For this, indicate a calibration date in the corresponding calibration menu and set
the field Refus (test refusal) on Yes.
196 UM_2.00A_EN
Database management OptiDist™
The Diagnostic main menu is in great part dedicated to the OptiDist™ service and
maintenance. With this menu, it is possible to:
• Display current readings
• In Diagnostic mode: Control the main parts of the analyzer to check for their
proper operation.
NOTE
• The diagnostic functions of the OptiDist™ are mainly dedicated
to service and maintenance staff. Using of the dignostic func-
tions needs having a level 2 access (Maintenance).
• With level 0 access, the operator can only view the current
readings on the Diagnostic main menu screen.
Open the Diagnostic main menu:
1. Press successively the buttons More > More > Diag. on the Run / Start &
Display screen.
The Diagnostic main menu opens.
UM_2.00A_EN 197
OptiDist™ Database management
/i
Measure - The active mode is the “Measure mode”. In Measure mode, the measurement-
Command parameter screen allows checking all current readings and the corresponding
Analog / Digital conversion counts.
This enables the ability to quickly check for abnormal data, which may be due to
a faulty sensor or a faulty electrical component.
Pressing the Measure button activates the “diagnostic mode”. The Measure but-
ton turns into Command button indicating, that the main parts of the OptiDist™
can be controlled.
Activation of the “diagnostic mode” needs having a level 2 access (Maintenance).
/i
The Diagnostic main menu screen gives access to the following information:
The flask heater assembly • Vapor Tmp:
(first box on top left) Temperature given by the vapor temperature probe with the
resistance value of the temperature probe (Res:) and the corre-
sponding conversion count (A/D)
• Optimizer:
Value given by the Optimizer and the corresponding conversion
count (A/D)
• Dry Point:
Temperature given by the thermocouple (if connected)
• Heating Tmp:
Temperature of the heating element with the corresponding con-
version count (A/D)
The volume measurement • Volume:
(second box on center left) Volume measured with the level follower optical fork
• Follower Back:
Conversion counts of the reception light signal of the level fol-
lower 1
• Follower Central:
Conversion counts of the reception light signal of the level fol-
lower 2
Ambient conditions • Bar. Pressure:
(third box on bottom left) Volume measured with the level follower optical fork
• Ambient Tmp:
Conversion counts of the reception light signal of the level fol-
lower 1
The condenser bath • Condenser:
(first box on top right) Temperature given by the condenser bath temperature probe
The receiving chamber • Receiver:
(second box on center right) Temperature given by the receiving chamber temperature probe
The cooling system • Exchanger:
(third box on bottom right) Temperature of the heat exchanging liquid in the cooling circuit
198 UM_2.00A_EN
Database management OptiDist™
/i
UM_2.00A_EN 199
OptiDist™ Database management
200 UM_2.00A_EN
Database management OptiDist™
NOTE
• To restore data and to update the firmware on the OptiDist™
needs having a level 1 access (Supervisor) or level 2 access
(Maintenance).
• The operator can perform a Backup only, which needs having a
level 0 access (Operator).
UM_2.00A_EN 201
OptiDist™ Database management
/i
Restore button; only active with level 1 access (Supervisor) or level 2 access
(Maintenance)
Exit button, return to Run / Start & Display screen
1. Insert the USB memory stick into the USB A port on the right side of the
OptiDist™ or into the USB A port at the rear.
2. Select the data to be saved or restored from the pull down list in the Backup
field or from the pull down list in the Restore field.
3. Press the Backup button to perform a backup or the Restore button to restore
data from the USB memory stick to the internal memory of the OptiDist™.
202 UM_2.00A_EN
Database management OptiDist™
NOTE
According to the data transferred, the OptiDist™ might need to be
switched off for the new data to be taken into account. If it is
needed, it will be indicated in the status bar. The OptiDist™ takeoff
is automatically done when the USB stick is removed.
NOTE
To update the firmware on the OptiDist™ needs having a level 1
access (Supervisor) or level 2 access (Maintenance).
1. Insert the USB memory stick into the USB A port on the right side of the
OptiDist™ or into the USB A port at the rear.
NOTE
The USB stick is not considered as a valid, if it does not contain all
files needed. In this case an alarm message appears.
2. In the Backup and Restore main menu press the Upload Firmware button to
start the transfer of the new firmware to the internal memory of the OptiDist™.
3. After the data transfer is completed,
a. either remove the USB stick to return to the Run / Start & Display screen,
b. or press the Exit button to return to the Run / Start & Display screen.
NOTE
According to the data transferred, the OptiDist™ might need to be
switched off for the new data to be taken into account. If it is
needed, it will be indicated in the status bar. The OptiDist™ takeoff
is automatically done when the USB stick is removed.
UM_2.00A_EN 203
OptiDist™ Database management
204 UM_2.00A_EN
Diagnostic functions OptiDist™
8 Diagnostic functions
UM_2.00A_EN 205
OptiDist™ Diagnostic functions
This section includes instructions for measurement checks and testing procedures
of the instrument’s major parts using diagnostic functions and the Doctor Box.
NOTE
The diagnostic functions are mainly dedicated to service and main-
tenance of the OptiDist™. It requires having a level 2-access
authorization (maintenance).
206 UM_2.00A_EN
Diagnostic functions OptiDist™
NOTE
For the corresponding calibarations of these circuits refer to section
7.10. The Calibration database.
The Diagnostic main menu is in great part dedicated to the OptiDist™ service and
maintenance. With this menu, it is possible to:
• Display current readings
• In Diagnostic mode: Control the main parts of the analyzer to check for their
proper operation, see chapter 8.4. Various individual component diagnostic
functions.
UM_2.00A_EN 207
OptiDist™ Diagnostic functions
/i
Measure - The active mode is the “Measure mode”. In Measure mode, the measurement-
Command parameter screen allows checking all current readings and the corresponding
Analog / Digital conversion counts.
This enables the ability to quickly check for abnormal data, which may be due to
a faulty sensor or a faulty electrical component.
Pressing the Measure button activates the “diagnostic mode”. The Measure but-
ton turns into Command button indicating, that the main parts of the OptiDist™
can be controlled, see 8.4.1. Activate the diagnostic mode.
Activation of the “diagnostic mode” needs having a level 2 access (Maintenance).
/i
The Diagnostic main menu screen gives access to the following information:
The flask heater assembly • Vapor Tmp:
(first box on top left) Temperature given by the vapor temperature probe with the
resistance value of the temperature probe (Res:) and the corre-
sponding conversion count (A/D)
• Optimizer:
Value given by the Optimizer and the corresponding conversion
count (A/D)
• Dry Point:
Temperature given by the thermocouple (if connected)
• Heating Tmp:
Temperature of the heating element with the corresponding con-
version count (A/D)
The volume measurement • Volume:
(second box on center left) Volume measured with the level follower optical fork
• Follower Back:
Conversion counts of the reception light signal of the level fol-
lower 1
• Follower Central:
Conversion counts of the reception light signal of the level fol-
lower 2
Ambient conditions • Bar. Pressure:
(third box on bottom left) Volume measured with the level follower optical fork
• Ambient Tmp:
Conversion counts of the reception light signal of the level fol-
lower 1
The condenser bath • Condenser:
(first box on top right) Temperature given by the condenser bath temperature probe
The receiving chamber • Receiver:
(second box on center right) Temperature given by the receiving chamber temperature probe
The cooling system • Exchanger:
(third box on bottom right) Temperature of the heat exchanging liquid in the cooling circuit
208 UM_2.00A_EN
Diagnostic functions OptiDist™
If the deviation is greater than ±3 mbar carry out the calibration of the atmospheric
pressure measurement.
If there is an abnormal reading displayed, the atmospheric pressure sensor may
be disconnected or faulty. Call for a qualified technician.
If the OptiDist™ does not indicate the expected value, calibrate the vapor
temperature measurement circuit, see 7.10.4. Vapor probe circuit calibration.
UM_2.00A_EN 209
OptiDist™ Diagnostic functions
CAUTION
The heating element temperature must be checked when the heat-
ing block is at ambient temperature, either before a test or at least
ten minutes after the end of a test and the heater must not be ener-
gized.
If the two readings are not the same, the temperature probe of the condenser bath
can be faulty or disconnected. Call for a qualified technician.
210 UM_2.00A_EN
Diagnostic functions OptiDist™
If the two temperatures are not identical, the receiving chamber temperature
probe can be faulty or disconnected. Call for a qualified technician.
UM_2.00A_EN 211
OptiDist™ Diagnostic functions
NOTE
Be aware that an OptiDist™ that is running a test cannot carry out
any diagnostics and an OptiDist™ in diagnostic mode cannot start
a test.
1. Press successively the buttons More > More > Diag. on the Run / Start &
Display screen.
The diagnostic main menu opens.
2. Press the Measure button in the right bottom of the screen to activate the
diagnostic mode.
212 UM_2.00A_EN
Diagnostic functions OptiDist™
NOTE
The diagnostic function is mainly dedicated to service and mainte-
nance of the OptiDist™. It requires having a level 2-access author-
ization (maintenance).
If entry to a protected level is attempted, a warning message is dis-
played, indicating the current level and the access level requested.
3. Press on the password entry field to display the virtual keyboard and enter the
maintenance password.
The Measure button turns into Command button indicating the OptiDist™ is in
diagnostic mode. From now on, each part can be controlled separately.
UM_2.00A_EN 213
OptiDist™ Diagnostic functions
Commands remain active until they are disabled and as long as the OptiDist™ is
in diagnostic mode.
4. To disable the diagnostic mode, press the Command button or press the Exit
button and confirm that you want to quit the diagnostic mode.
NOTE
When a problem is identified on a specific element, its supply cir-
cuit can be quickly tested with the optional Doctor Box, see 8.5.
Connection and use of the “Doctor Box”.
214 UM_2.00A_EN
Diagnostic functions OptiDist™
UM_2.00A_EN 215
OptiDist™ Diagnostic functions
2. If the level follower does not move, call for a qualified technician to check:
• the condition of the drive belt
• the motor operation
• the motor supply circuit
3. Press on the Command button to quit the diagnostic mode and for the level
follower detector returns to its standby position.
Conversion counts on the optical receptors according conversion count of the optical trans-
mitter:
Conversion count on the transmitter Conversion count on the receptor
0 points > 4000 points (provided the OptiDist™ is away
from direct sunlight)
4095 points (maximum) < 1500 points
NOTE
Power on the pump to flow the heat exchanging liquid in the con-
denser bath (see 8.4.4.4. Condenser bath pump). Thus, the raising
of temperature of the condenser bath will be more rapid and homo-
geneous.
216 UM_2.00A_EN
Diagnostic functions OptiDist™
The condenser bath temperature indicated just above should begin to rise
after few seconds. If not, the condenser heating element or its supply circuit is
failing.
2. Check the supply circuit of the condenser heating element using the Doctor
Box or call for a qualified technician to check the supply circuit and heating
element operation.
3. Press again on the entry field modified and stop the heating command or
press on the Command button to quit the diagnostic mode.
NOTE
Power on the pump to flow the heat exchanging liquid in the con-
denser bath (see 8.4.4.4. Condenser bath pump). Thus, the drop
(assuming the exchanger temperature is < 0 °C) of temperature of
the condenser bath will be more rapid and homogeneous.
NOTE
Note: This command is only available if the optional condenser
boost heating is installed.
1. Press on the entry field opposite Boost and enter the additional heating power
desired in percent.
NOTE
Power on the pump to flow the heat exchanging liquid in the con-
denser bath (see 8.4.4.4. Condenser bath pump). Thus, the raising
of temperature of the condenser bath will be more rapid and homo-
geneous.
The condenser bath temperature indicated just above should begin to rise
after a few seconds. If not, the optional condenser boost heating element or its
supply circuit is failing.
UM_2.00A_EN 217
OptiDist™ Diagnostic functions
2. Check the supply circuit of the boost heating element using the Doctor Box or
call for a qualified technician to check the supply circuit and boost heating
element operation.
3. Press again on the entry field modified and stop the boost heating command
or press on the Command button to quit the diagnostic mode.
NOTE
To detect an immediate raising of temperature in the receiving
chamber, power on the fan (see 8.4.5.3. Receiving chamber fan).
The receiving chamber temperature indicated just above should begin to rise
after few seconds. If not, the heating element or its supply circuit is failing.
3. Check the supply circuit of the heating element of the receiving chamber using
the Doctor Box or call for a qualified technician to check the supply circuit and
heating element operation.
4. Press again on the entry field modified and stop the heating command or
press on the Command button to quit the diagnostic mode.
218 UM_2.00A_EN
Diagnostic functions OptiDist™
NOTE
To detect an immediate drop of temperature (assuming the
exchanger temperature is < 0 °C) in the receiving chamber, power
on the fan, see 8.4.5.3. Receiving chamber fan.
NOTE
The receiving chamber fan allows the air flowing in the receiving
chamber during heating or cooling. If this fan is not powered on, the
temperature of the receiving chamber will not be homogeneous
and the temperature indicated opposite Receiver will not evolve in
linear way.
2. Check the supply circuit of the receiving chamber fan using the Doctor Box or
call for a qualified technician to check the supply circuit and the fan operation.
3. To stop the command, press on the Fan button again or press on the
Command button to quit the diagnostic mode.
UM_2.00A_EN 219
OptiDist™ Diagnostic functions
NOTE
Note: This command is only available if the VOC control option is
installed.
1. Press on the VOC pump button to start the extraction pump of the volatile
organic components.
The VOC Pump should start up: disconnect the exhaust pipe at the rear of the
OptiDist™, to check the air in the receiving chamber is being moved at a rapid
rate. If not, the VOC pump or its supply circuit is failing.
2. Check the supply circuit of the VOC pump using the Doctor Box or call for a
qualified technician to check the supply circuit and the VOC pump operation.
3. To stop the command, press on the VOC pump button again or press on the
Command button to quit the diagnostic mode.
Conversion counts on the optical receptors according conversion count of the optical trans-
mitter:
Conversion count on the transmitter Conversion count on the receptor
0 points > 4000 points (provided the OptiDist™ is away
from direct sunlight)
4095 points (maximum) < 1500 points
3. Pass an object, e.g. a pencil, between the emitter and the receiver of the first
drop optical detector. The indicator lamp opposite the entry field Transm:
should light.
4. Press again on the entry field and stop the transmitter.
5. If the conversion count of the receptor is out of the boundary, first clean the
drop detection optical fork.
6. Restart the test. If the problem persists, call for a qualified technician to
replace the drop detection optical fork.
220 UM_2.00A_EN
Diagnostic functions OptiDist™
1. Activate the diagnostic mode (see 8.4.1. Activate the diagnostic mode).
2. Press on the CoolSyst button to start the cooling unit of the OptiDist™
(provided the OptiDist™ has been switched on for 2 min)
You should hear the cooling unit starting up and the temperature of the
exchanger of the cooling system should decrease. If not, the cooling unit or its
supply circuit is failing.
3. Check the supply circuit of the cooling unit using the Doctor Box or call for a
qualified technician to check the supply circuit and cooling unit operation.
4. Press on the CoolSyst button again to stop the command or press on the
Command button to quit the diagnostic mode.
NOTE
If the Command button is not pressed, the OptiDist™ remains in
diagnostic mode, all regulations are switched off and a test run can-
not be started.
2. Press the Exit button and confirm that you want to quit the diagnostic mode
and to return to the Run / Start & Display screen.
8.5.1 Introduction
The optional available Doctor Box allows rapid diagnostics of the OptiDist™
components and electronic boards without need for OptiDist™ disassembly.
The Doctor Box connects to the 25-pin connector on the right side of the
OptiDist™.
A F
F F1 F2 30A
U BIP 24V FLASK HT
S VOLTMETER
E F3 F4 F5 F6 OUTPUT
S 24V
12V 12V
5V
T
R
5V 3.3V
A RED COND. HT BOOST HT REC . FAN LED
F
F
I 3.3V
C COOL . UNITT COOL . SV
ORANGE
L ADA DOCTOR BOX
I
G DOWN SOL. COND. REC.
H
T
GREEN
FLASK FAN REC . HT VOC PUMP
B C D E
UM_2.00A_EN 221
OptiDist™ Diagnostic functions
When the Doctor Box is connected, it supplies information on the operation of the
following components (when they are energized):
• The fuses (A)
• The status indicator beacon (B)
• The heating elements (C, D, E)
• The fans (C, E)
• The cooling unit (C)
• The pumps (E)
• The heating block solenoid (C)
• The cooling unit solenoid valve (C)
Fuses are located on the electronic boards. If fuses are operable, the LED
corresponding to each fuse in box (A) of the figure above should light up
(except fuse F1).
NOTE
If installed, the optional built-in printer should be activated to check
the fuse F1.
222 UM_2.00A_EN
Diagnostic functions OptiDist™
/i
• Cooling unit
• Receiver fan
• Condenser pump
• Receiver pump
• Receiver heating element
• VOC pump
30A Heating control Protects the supply circuit of the distillation flask
FLASK HT board heating element
If one fuse is failing, call for a qualified technician to verify the various elements
controlled by the concerned circuit (see table above) before replacing it with a fuse
of the specified type and current ratings.
WARNING
Switch off the OptiDist™ and disconnect it from the mains before
removing the housing.
If the service operations, to be carried out by a qualified technician,
require you to switch on the OptiDist™, switch it off again before
disconnecting or removing any electrically powered components.
UM_2.00A_EN 223
OptiDist™ Diagnostic functions
NOTE
The Doctor Box checks the powered output of electronic boards.
Consequently, the LED corresponding to the supply circuit of a con-
trolled part lights up even if the part is failing.
If a LED does not light up and the fuse corresponding to the concerned supply
circuit is ok, the concerned electronic component(s) on board is(are) failing. The
electronic board has to be replaced.
NOTE
These checked voltages are generated by the control electronic
board.
224 UM_2.00A_EN
Maintenance OptiDist™
9 Maintenance
UM_2.00A_EN 225
OptiDist™ Maintenance
It is best that the maintenance checks and operations explained in this manual
should be performed by the operator. In this way, the operator will better
understand the operation of the OptiDist™ and will be more apt to notice any
anomalies as regards the measurement circuits.
Timing Action
Daily Check the state of the distillation flask
Check the state of the vapor tube
Check the state of the condenser tube - distillation flask silicone tube
Check the state of the stopper seals
Check the state of the stopper tip
Check the state of the receiver rubber cap
Weekly Execute the measurement checks
Carry out a cleaning destillation:
Depending on the nature of the material being distilled and on the frequency of
distillations, a cleaning distillation should be performed at regular intervals (e.g.
weekly or monthly) to remove wax crystals that deposit in the condenser tube.
This is best done with a mixture of 50% toluene and 50% methanol/ethanol
226 UM_2.00A_EN
Maintenance OptiDist™
Timing Action
Once a year Calibrate the measurement circuits and elements
Control the working condition of the flame extinguisher device
Control the heating element coil
Clean all vent holes and protection bars of the extinguisher injection slopes
Clean the condenser of the cooling unit
Control of heat exchanging liquid coolant level in the cooling system
As required Replace the printer paper (see the printer manual).
Replace the printer ink ribbon (see the printer manual)
Replace the condenser tube
Replace the distillation flask silicone tube
Replace the receiver rubber cap
Replace the stopper seals
Replace the stopper tip
Drain the receiving chamber spillage trap
UM_2.00A_EN 227
OptiDist™ Maintenance
WARNING
Switch off the OptiDist™ and disconnect it from the mains before
removing a housing.
If the service operations (carried out by a qualified technician)
require you to switch on the OptiDist™, switch it off again before
disconnecting or removing any electrically powered components.
WARNING
Replace the housing cover before switching on the OptiDist™.
Ascertain that the distillation flask heater assembly housing is prop-
erly seated: Openings provided for the fire sensors and the opti-
mizer must face the sensors).
WARNING
Switch off the OptiDist™ and disconnect it from the mains before
removing a housing.
If the service operations (carried out by a qualified technician)
require you to switch on the OptiDist™, switch it off again before
disconnecting or removing any electrically powered components.
1. Remove the 2 screws (A) on the top of the OptiDist™ main housing.
2. Remove the 3 screws (A) on the left hand side of the OptiDist™ main housing.
3. Remove the left part (B) of the OptiDist™ main housing by pulling it to the rear
and then to the left.
4. Remove the 3 screws (A) on the right hand side of the OptiDist™ housing.
5. Remove the right part (C) of the OptiDist™ main housing by pulling it to the
rear and then to the right.
228 UM_2.00A_EN
Maintenance OptiDist™
A C
B
A
A
WARNING
Switch off the OptiDist™ and disconnect it from the mains before
removing the housing.
1. To control the heating element coil, remove the distillation flask heater
assembly housing.
2. Control, that the coils of the heating element do not touch each other to avoid
a hot spot.
3. If the coils touch each other, separate the coils with a flat-bladed screwdriver.
If this is not sufficient, replace the heating element.
4. Reassemble the base plate and the distillation flask heater assembly housing.
UM_2.00A_EN 229
OptiDist™ Maintenance
WARNING
Switch off the OptiDist™ and disconnect it from the mains before
removing the housing.
230 UM_2.00A_EN
Maintenance OptiDist™
CAUTION
Wear protective cloves if you have to cut off the distillation flask sili-
cone tube to avoid all cut and burn hazards .
1. Pull-off the condenser/distillation flask silicone tube (A) carefully or cut it off
with caution if it is stuck to the condenser tube.
2. Attach a new condenser/distillation flask silicone tube to the condenser tube.
CAUTION
When replacing the rubber cap, wear protective gloves to avoid all
cut and burn hazards .
1. Pull-off the the rubber cap (A) carefully to slide it over the drip tip.
2. Attach a new rubber cap to the condenser tube.
UM_2.00A_EN 231
OptiDist™ Maintenance
NOTE
The distillation flask stopper assembly is specially designed for the
precise adjustment of the probe positioning in the stopper and so,
in the distillation flask.
The vapor temperature probe is equipped with a built-in memory circuit including
up to 10 correction values, the probe identity and its serial number. Thus, no
adjustment is necessary: the OptiDist™ automatically recognizes the probe and
imports data at plug-in avoiding manual entry of probe correction values and
probe ID.
NOTE
PAC probe is required. If another probe is used, a new test will not
be able to start.
To display these correction values and adjust the calibration frequency of the
vapor temperature probe, see 7.10.3. Vapor probe calibration.
To order a new probe, please see spare part list for the item (the probe, wiring and
connector cannot be dismantled).
232 UM_2.00A_EN
Maintenance OptiDist™
WARNING
Take care of hot or acid liquids!
NOTE
The heat exchanging liquid must be ordered at PAC (see the spare
part list). It is important to use coolant provided by PAC for correct
operation of your OptiDist™, as its properties are particularly com-
patible with the application.
CAUTION
Switch off the OptiDist™ and disconnect it from the mains before
removing a housing.
If the service operations (carried out by a qualified technician)
require you to switch on the OptiDist™, switch it off again before
disconnecting or removing any electrically powered components.
UM_2.00A_EN 233
OptiDist™ Maintenance
234 UM_2.00A_EN
Maintenance OptiDist™
NOTE
To obtain accurate and repeatable results, prepare the samples
and OptiDist™ with particular care as described in the following
sections.
CAUTION
Switch off the OptiDist™ and disconnect it from the mains before
removing a housing.
If the service operations (carried out by a qualified technician)
require you to switch on the OptiDist™, switch it off again before
disconnecting or removing any electrically powered components.
1. Shut off the external gas supply and disconnect it from the fire extinguisher
gas inlet at the rear of the OptiDist™.
2. Approach a naked flame (e.g. a cigarette lighter flame) to the vicinity of the
flame detector, located on the right side of the distillation flask heater
assembly.
UM_2.00A_EN 235
OptiDist™ Maintenance
NOTE
The flame detection audio alarm remains active until the OptiDist™
is switched off, even if the fault has been acquitted and the flame
has been extinguished.
CAUTION
If a real fire detection and extinction operation is performed, should
it be after an accident or during testing by setting fire to a sample in
the distillation flask compartment, then the test described in this
section should be used to verify that the flame detector works cor-
rectly.
When a real fire has been extinguished, it is strongly advised to
open windows in order to air the laboratory (extinction gas being
CO2).
The series contact provided for the connection of the user’s external alarm device
functions as an ON/OFF switch for the user’s device: this series contact has 2
connection positions such that when a fire is detected:
• 1-2 normally closed
• 2-3 normally opened
NOTE
To stop the alarm, the OptiDist™ must be switched off
CAUTION
Caution! Switch off the OptiDist™ and disconnect it from the mains
before removing the housing.
236 UM_2.00A_EN
Maintenance OptiDist™
5. Connect the wiring to the J4 connector of the fire safety board (connection
made by screw).
J1 J5
J4 connector
J6
NO Com NC
J2
J3 Fire Safety Board
J4
UM_2.00A_EN 237
OptiDist™ Maintenance
238 UM_2.00A_EN
Alarm treatment OptiDist™
10 Alarm treatment
UM_2.00A_EN 239
OptiDist™ Alarm treatment
CAUTION
Generally when the alarm buzzer is triggered, the operator must
display the cause and ensure that everything is working correctly.
For example, owing to a faulty component an alarm that normally should place the
OptiDist™ in idle mode fails to do so. In that case, the operator must switch off the
OptiDist™.
1. Press the OK button to acquit the alarm and stop the buzzer.
2. Check causes of alarm setting of: see 10.4. List of alarm messages and
troubleshooting.
240 UM_2.00A_EN
Alarm treatment OptiDist™
UM_2.00A_EN 241
OptiDist™ Alarm treatment
242 UM_2.00A_EN
Alarm treatment OptiDist™
UM_2.00A_EN 243
OptiDist™ Alarm treatment
244 UM_2.00A_EN
Alarm treatment OptiDist™
UM_2.00A_EN 245
OptiDist™ Alarm treatment
246 UM_2.00A_EN
Alarm treatment OptiDist™
UM_2.00A_EN 247
OptiDist™ Alarm treatment
248 UM_2.00A_EN
Alarm treatment OptiDist™
UM_2.00A_EN 249
OptiDist™ Alarm treatment
250 UM_2.00A_EN
List of accessories OptiDist™
11 List of accessories
UM_2.00A_EN 251
OptiDist™ List of accessories
252 UM_2.00A_EN
OptiDist™
/i
NOTES
UM_2.00A_EN 253
OptiDist™
PAC Companies
Address: PAC L.P.
PAC Corporate Headquarters
8824 Fallbrook Drive
Houston, Texas 77064
USA
Phone: +1 (0) 281 580 0339
Fax: +1 (0) 281 580 0719
Sales: sales@paclp.com
Service: service@paclp.com
Address: PAC ISL
BP 70285 - Verson
Parc d’Activités de la Mesnillière
14653 Carpiquet Cedex
France
Phone: +33 (0) 231 264 300
Fax: +33 (0) 231 266 293
Sales: salessupport.france@paclp.com
Service: service.france@paclp.com
Address: PAC Herzog
Badstrasse 3 - 5
P.O. Box 1241
DE-97912 Lauda-Königshofen
Germany
Phone: +49 (0) 9343 640 0
Fax: +49 (0) 9343 640 101
Sales: sales.germany@paclp.com
Service: service.germany@paclp.com
Address: PAC Rotterdam
Innsbruckweg 35
3047 AG Rotterdam
The Netherlands
Phone: +31 (0) 10 462 4811
Fax: +31 (0) 10 462 6330
Sales: sales.netherlands2@paclp.com
Service: service.netherlands@paclp.com
254 UM_2.00A_EN