Professional Documents
Culture Documents
only in the late l950s and early l960s were antibodies used for
direct monitoring of immuno reactions, provides opportunity to
quantification of an antigen [1, 2]. Today, immunoassay is the
gain new insight into antigen-antibody reaction kinetics and to
predominant analytical technique for quantitative measure- create rapid assay devices having wide applications.
ments, being used over a wide concentration range, in many
different biological matrices, and in a range of delivery formats. Biosensor Technology
The unique feature of immunoassay that provides the desired A biosensor has two major components: a biological detector or
specificity is the complementary reaction (both in chemical sensor molecule, and a signal transducer that provides an
binding and spatial orientation of reactive groups) between indication (or signal) that the ligand has bound to the receptor
antigen and antibody. (sensor) molecule (Fig. 1). A variety of receptor molecules can be
Apart from the obvious unique contribution of an antibody used, including enzyme, antibody, affinity ligand (e.g., lectin),
receptor, peptide, oligonucleotide, organelle, organism, or tis-
sue slice [12].
Department of Clinical Biochemistry, St. Bartholomew’s and Royal London The transducer detects a change in one or more physico-
School of Medicine and Dentistry, Turner St., London El 2AD, UK. chemical properties as a ligandbinds to its receptor; the
Author for correspondence. Fax 44-171-377-1544, e-mail c.price@lhmc.
ac.uk.
interaction may result in a change of pH, electron transfer,
Received August 18, 1995; accepted October 23, 1995. refractive index (RI), heat transfer, or uptake or release of gases
193
194 Morgan et al.: Immunosensors review
SAMPLE
Table 1. Types of blosensors.
3 biologically
sensitive
Transducer system
Ion-selective
electrodes (ISE)
Measurement
Potentiometric
Applications
Ions (in biological media),
enzyme electrodes, and
coating immunosensors
SENSOR Gas-sensing Potentiometric Gases, enzyme/organelle/
electrodes cell/tissue electrodes for
substrates and inhibitors,
enzyme immunoelectrodes
Field-effect Potentiometric Ions, gases, enzyme
transistors (FET) substrates, immunological
analytes
Enzyme electrodes Amperometric Enzyme substrates,
or specific ions.’ Amplification and processing of the signal metric, amperometric, or conductimetric), mass (piezoelectric
change provides the opportunity for quantification of the or acoustic wave), heat (calorimetric), or optical [luminescent,
ligand-receptor binding. There are many examples of sensors in fluorescent, reflective, ellipsometric, surface plasmon resonance
routine use, including ion-selective and gas electrodes; however, (SPR), and waveguide] properties (Table 1). The principles of
fewer kinds of biosensors are in routine use, although a large these transduction systems will be described in broad terms and
number of approaches for doing so have been described. The then as applied to the design of immunosensors.
most successful biosensor to date is probably that using an
enzyme as the biological detector molecule with a range of Immunosensors
transducers; several such biosensors have been described for Biosensors concerned with monitoring solely antibody-antigen
measurement of glucose [13]. Fewer examples of immunosen- interactions can also be termed immunosensors [14]. Similarly
sors, both in principle or application, have been described, but to conventional immunoassays, these devices are based on the
the burgeoning literature in this field clearly marks it as a fertile principles of solid-phase immunoassay, with either antibody or
area for development. antigen immobilized at the sensor surface. An extensive range of
The term biosensor has been used here fairly literally and analytes can be detected and measured by immunosensors, e.g.,
includes what could be described as direct and indirect sensors. medical diagnostic markers such as hormones (steroids and
For the purpose of this review, a direct biosensor is one in which pituitary hormones), drugs (therapeutic and abused), and bacte-
there is direct detection of the antigen-antibody reaction in real ria, and environmental pollutants such as pesticides. However,
time, with or without the use of a label; this is a true biosensor. the advantages of an absence of labeling requirements and the
An indirect biosensor is one in which a preliminary “biological” ability to investigate the reaction dynamics of antibody-antigen
reaction takes place in a reaction tube and the products of that binding have given these devices the potential to revolutionize
reaction are detected with a sensor; this is not a biosensor in our conventional immunoassay techniques. Of course, there are
view and will not be considered in detail in this review. differences between conventional solid-phase immunoassays and
Four main types of transducer have been used in biosensor immunosensors. In the former, reagents are generally used only
technology; these exploit changes in electrochemical (potentio- once (albeit in small quantities), whereas immunosensors can
facilitate the regeneration of the immobilized component; by
thus using the reversibility of the antibody-antigen reaction,
Nonstandard abbreviations: RI, refractive index; ISE, ion-selective electrode; immunosensors can maximize reagent usage. However, a par-
SPR, surface plasmon resonance; FET, field effect transistor; ISFET, lon-sensinve
ticularly high affinity constant and a labile immobilized ligand
FET; hCG, human chorionic gonadotropin; FCFD, fluorescence capillary fill
device; IRS, internal reflection spectroscopy; 10, integrated optics; TE, electrical may make regeneration of the surface difficult.
field; and TM, magnetic field. The sensitivity and specificity of an immunosensor are
Clinical Chemistry 42, No. 2, 1996 195
determined by the same characteristics as in other solid-phase developed for the detection of gases such as carbon dioxide; this
immunoassays, namely, the affinity and specificity of the binding consists of a gas-permeable membrane and a hydrogen-sensitive
agent and the background noise of the detection system (trans- pH electrode, separated by a layer of electrolyte solution.
ducer). In some cases, the use of labels has been incorporated Dissolved gas diffuses through the membrane to the electrolyte,
into immunosensor design to enhance sensitivity, although these altering its pH. The pH electrode monitors these changes, and
indirect-sensing devices can no longer be said to monitor the the gas concentration is thus determined.
antibody-antigen reaction directly. A great deal of work in the The incorporation of ISEs, pH electrodes, or gas-sensing
development and design of immunosensors has been expended electrodes into potentiometric immunosensors to improve sen-
over the past decade, making use of different transducer systems sitivity has been extensively investigated by Rechnitz and co-
as well as exploring the benefits of miniaturization and large- workers. Examples include the use of carbon dioxide gas-sensing
scale fabrication to ensure their commercial success/viability. electrodes for immunochemical measurements of antibiotics
The basic principles of immunosensors with regard to the [17, 18], human IgG [19], and digoxin [20]. In this approach a
---_______________
monitoring of electrolytes. The ISFET has since been devel-
+ +-+
oped for a wider range of applications [30], including penicillin
determination [3!]. ISFET technology has also been adapted for
charge
immunosensing; one novel example detects the changes in
MEMBRANE
charge densities and isoelectric points that take place upon
at interface
+ + + formation of an antibody-antigen complex [32] with a detection
valinomycin limit in the range 1-10 X lO mol/L.
A different approach to the construction of potentiometric
Fig. 2. Principle of potentiometric sensing, illustrated with the potas-
immunosensors, in which an electro-catalytically active biocat-
sium SE, in which the antibiotic valinomycin selectively binds potas-
sium ions within the membrane, leaving behind the co-ions in the alyst is used as a label, has been more recently described [33]. In
sample; the resulting charge separation, which is established across this indirect detection technique, insulin is immobilized at the
the sample/membrane interface, generates a potential that is related electrode, to which monoclonal anti-insulin antibodies labeled
to the concentration of potassium ions in the sample. with the enzyme lactase may bind. Lactase catalyzes the elec-
196 Morgan et al.: Immunosensors review
concentration
and large changes in
does
I _______
LL] sensing
electrode
reference
electrode potential facilitate kinetic analysis of the reaction to be
performed. electrode
Despite the developments described, however, potentiomet-
T
electrolyte
nc immunosensing has experienced several problems. The
signal-to-noise ratio is low because the charge density on most electrode
biomolecules is low compared with background interferences
body
(e.g., ions), and there is a marked dependence of signal response
CONDUCTIMETRIC TRANSDUCERS
PIEZOELECTRIC CRYSTALS
Conductimetric measurements are widely applicable to chemical
The phenomenon of piezoelectnicity was discovered in 1880 [53]:
systems, given that many chemical reactions produce or con-
Electric dipoles generated in anisotropic natural crystals (with no
sume ionic species and therefore alter the overall electrical
center of symmetry) were subjected to mechanical stress, thus
conductivity of a solution. Biosensors can be designed by
causing them to oscillate at frequencies between 9 and 14 MHz.
immobilizing a suitable enzyme over a set of electrodes made of
There are -20 naturally abundant piezoelectnic crystals, including
noble metal (e.g., gold, silver, copper, nickel, or chromium) and
quartz (Si02), lithium niobate (LiNbO), zinc oxide (ZnO), gal-
measuring the change in conductance of a solution containing
lium arsenide (GaAs), cadmium sulfide (CdS), lithium tantalate
the analyte of interest when an electric field is applied. For (LiTaO3), tourmaline, and Rochelle salt [54]; some manmade
example, when urea is converted to its ionic product NH by
ceramics and polymers also have piezoelectric properties. Quartz is
the enzyme urease, the increase in solution conductance mea- the most commonly used piezoelectnic material because of its
sured is proportional to urea concentration [47]. With use of the chemical stability in aqueous solutions and resistance to high
appropriate enzymes, the concentrations of acetaminophen temperatures without loss of piezoelectnic properties. In piezoelec-
(paracetamol), creatinine, L-asparagine, penicillin G, and glu- tric biosensors, the crystals are coated with an adsorbent that
cose, among others, can be determined, sometimes with multi- selectively interacts with the analyte of interest; subsequent binding
analyte detecting devices [48]. In another example, a conductive increases the mass of the coated crystal and alters its basic fre-
polymer-based immunosensor, composed of an ELISA with an quency of oscillation (Fig. 4 [55J). Monitoring the oscillation
electroconductive solid support, has been developed for pesti- frequency allows determination of the change in mass, which is
cide detection [49]. Variations in ionic strength and buffer proportional to analyte concentration. This technique has had
capacity of measured samples have caused problems with this much success in the analysis of gaseous environmental pollutants
type of biosensor in the past, but these drawbacks have been [56-58]. Piezoelectric crystal sensors have also been investigated
overcome with more recent designs [SO]. Conductimetric sen- for monitoring antibody-antigen interactions; in this procedure,
sors may also have problems with nonspecificity of measure- the crystals are coated with antibody and the amount of antigen
ments, because the resistance of a solution is determined by the bound is determined, or vice versa. One example demonstrated a
migration of all ions present. To date, the development of detection limit for IgG of 43 mg/L [59]. Other published examples
immunosensors based on conductance has hardly been ex- include the detection of herbicides in drinking water [60, 61],
198 Morgan et al.: Immunosensors review
Ellipsoinetry. In ellipsometry, the change in the state of polariza- sensor surface). The resonance angle is determined by the
tion of light caused by reflection from a planar layered structure wavelength and polarization state of the incident light, as well as
is monitored [94, 122, 123]. This method generated little inter- by the refractive indices of the prism, metal, and sample layers of
est for many years after the first observations, and miniaturiza- the system. When all other factors are kept constant, the
tion and low-cost fabrication were thought to be unfeasible for resonance angle measures the RI at the surface of the SPR
this technology [14]. However, ellipsometry has become more system. Thus, changes in RI that take place when hiomolecules
popular, particularly for studying the binding of proteins to become attached to their specific ligands at the sensor surface
surfaces [124]. Once reflected from a mirrored surface, the alter the angle at which the drop in light intensity occurs.
polarized light is interrogated so as to produce two readings; Continuous monitoring of the reflected light intensity and the
these are subsequently converted into changes in amplitude and resonance angle therefore provides a direct profile of RI changes
in phase of light, respectively [125]. These properties are altered at the sensor surface and achieves real-time analysis of the
when a molecule is adsorbed to the surface, and the new binding events involved in the reaction.
between different mirrors inside the device. One interacts with INPUT GRATING COUPLER:
the sample and undergoes a change in phase so that, when the sample
two polarizations are recombined at the output, the interference
pattern alters.
layer
The difference interferometer [165] is based on a two-mode
interferometric thin-film waveguide [166] similar to the Mach- direction
Zehnder set-up. One of its applications is as a direct immunosensor
for label-free molecular-affinity analysis of biomolecular binding
propagation glass
events such as antibody-antigen reactions [167], for which a substrate
limit of 2.0 X iO B
.7,,,
detection mol/L was reported for hepatitis
surface antigen in serum. The molecular recognition component
(ligand) is immobilized on the sensor surface, and a polarized laser
FIg. 10. Structure of the lAsys device: At the resonance angle (0), light
J detector
times,
tion
and
costs
leakage
are also often
particularly
interaction,
phase.
with some electro-
in comparison
chemical devices. However, the extensive range of applications
offered by optical immunosensing, real-time moni-
probably overcomes
Manufacturing/fabrica-
passes through a low-RI coupling layer, then couples into a resonant or outweighs these issues. Perhaps a combination of optical and
layer of higher RI, generating an evanescent wave at the interface with
electrochemical approaches, e.g., SPR at electrochemical inter-
the sample layer that propagates some distance along the surface
before coupling back out of the device. faces or luminescence generated electrochemically, will be in-
Both TE and TM polarizations of light are detected, but unlike SPR there is no
vestigated in the future and developed with success [179].
decrease in the intensity of the reflected light and the angle of incidence is
monitored. TIR, total internal reflection.
Clinical Applications of Immunosensors
Not all of these sensing technologies developed for immuno-
and resemble sine waves (see Fig. 8). As the system approaches logical quantification have achieved practical usage in biological
resonance, there is a brief delay/break in the reflected light, fluids. Table 3 gives an overview of the sensing systems that have
caused by light coupling into the resonant layer. This results in been demonstrated with a sample derived from a biological
a change of 2ir radians (360#{176})
in the phase of light; the resulting matrix, i.e., serum on whole blood. The most sensitive (i.e.,
light beam is still in phase, but the resonance angle (TE phase lowest detection limit) is of the order of 2 x 10- 13 mol/L for
relative to TM) shifts by in radians (180#{176}).
When this occurs, the hepatitis B surface antigen in serum [167], which compares well
resulting polarization of the beam is rotated by 90#{176}
from that of with many conventional immunoassay techniques. Hapten as-
the original beam and is now out of phase. This rotated beam says operate in the low nanomolar region for digoxin and
can be monitored by a suitably orientated polarizer. Both estradiol. There is even
a highly sensitive assay for prostate-
incident polarizations (TE and TM) can undergo resonance and specific antigen blood by FCFD (3.3 x 10 12 mol/L)
in whole
be detected simultaneously, because their respective resonance technology [117], which is comparable with quoted detection
angles are widely separated. However, the TE polarization has a limits for automated heterogeneous immunoassays.
sharper resonance peak than does TM and is more sensitive to Large sums of money have been invested by diagnostic
any changes in the resonance angle and therefore in RI. companies to develop commercially viable immunosensors, as
Accordingly, the system is optimized to track the TE resonance yet with little obvious success. The two widely available immu-
angle but uses the TM resonance angle as a reference. nosensors are both direct optical systems-the BlAcore and the
Changes in RI occurring at or near the sensing surface IAsys-and both have surfaces of carboxylated dextran. These
because of the adsorption of biomolecules are within the region have proved to have very low nonspecific binding in biological
of the evanescent wave, and therefore alter the position of the matrices and achieve good detection limits for a variety of
resonance angle for TE. This angle shift is measured and molecules, but their major impact has been to revolutionize the
recorded throughout the experiment, providing a direct inter- kinetic rate analysis of biomolecular interactions. Surface-effect
pretation of RI and therefore of mass changes occurring at the measurement of kinetic rate constants by real-time, stable
sample surface [1 78]. The lAsys, which provides an accurate and automated immunosensors has received widespread usage in the
continuous representation of the RI changes taking place at the diagnostic, biotechnology, and pharmaceutical industries as well
sensor surface, differs from SPR in two ways: First, there is no as in basic clinical science.
Clinical Chemistiy 42, No. 2, 1996 205
The potential of immunosensing technologies has been demonstration systems have failed to reach commercial-scale
around for quite a while and new opportunities, e.g., microspot production-presumably because of fabrication difficulties or
technology, are continually developing. However, the main simply cost issues.
impact of these technologies has been (a) the innovative detec-
tion systems that have furthered conventional immunoassay Conclusions
systems and (b) the feasibility studies in both clinical and The diversity of the physical sciences that have been drawn into
environmental testing areas (Table 4). Despite huge innovations the realms of quantitative biochemical analysis is exciting and
and investments, however, immunosensors fill the pages of our has proved to be challenging to both the physical and biological
literature rather than the benches of our laboratories, and scientist. Several techniques clearly are worthy of developinent
in the field of immunosensors, some of which are now beginning
to emerge into the field of biochemical analysis. The require-
Table 4. ApplIcatIons of Immunosensors for clInIcal and ment for highly reproducible fabrication of devices, particularly
eds. Immunochemical assays and biosensor technology for the 34. Scheller F, Kirstein D, Kirstein
L, Schubert F, Wollenberger U,
199Os. Washington, DC: American Society for Microbiology, Olssen B, et al. Enzyme electrodes and their application. Philos
1992:23-35. Trans R Soc Lond B Biol Sci 1987;316:85-94.
10. Price CP, Newman Di, eds. Principles and practice of immuno- 35. Wilson GS. Fundamentals of amperometric sensors. In: Turner
assay. London: Stockton Press, 1991:643pp. APF, Karube I, Wilson GS, eds. Biosensors. Fundamentals and
11. Zuk RE, Ginsberg VK, Houts T, Rabbie J, Merrick H, UlIman EF, et applications. Oxford: Oxford University Press, 1987:165-79.
al. Enzyme immunochromatography-a quantitative immunoas- 36. Clark U, Lyons C. Electrode systems for continuous monitoring
say requiring no instrumentation. Clin Chem 1985;31:1144-5O. in cardiovascular surgery. Ann N Y Acad Sci 1962;1O2:29-45.
12. Turner APF, Karube I, Wilson GS, eds. Biosensors. Fundamentals 37. Cass AEG, Davis G, Francis GD, Hill HAO, Higgins Ii, Plotkin EV,
and applications. Oxford: Oxford University Press, 1987:77Opp. et al. Ferrocene-mediated enzyme electrode for amperometric
13. Schumann W, Schmidt H-L. Amperometric biosensors for sub- determination of glucose. Anal Chem 1984;56:667-71.
strates of oxidases and dehydrogenases. In: Turner APE, ed. 38. Aizawa M, Morioka A, Suzuki 5, Nagamura Y. Enzyme immu-
Advances in biosensors. London: iAl Press, 1992:79-130. nosensor Ill. Amperometric determination of human choriogo-
14. North i. Immunosensors: antibody-based biosensors. Trends
55. King W. Analytical uses of the piezoelectric crystal. Anal Chem devices and their analytical applications. AppI Biochem Bioeng
1964;36:1735-9. 1981:3:97-143.
56. AIder J, McCallum J. Piezo-electric crystals for mass and chemi- 81. Satoh I, Danielsson B, Mosbach K. Triglyceride determination
cal measurements. Analyst 1983;1O8:1169-89. with use of an enzyme thermistor. Anal Chim Acta 1981:131:
57. Guilbault GG, Jordan J. Analytical uses of the piezoelectric crystal 255-62.
detector [Review]. Crit Rev Anal Chem 1988:19:1-28. 82. Mattiasson B, Danielsson B. Calorimetric analysis of sugars and
58. Ngeh-Ngwainbi J, Suleiman A, Guilbault G. Piezoelectric crystals sugar derivatives with aid of an enzyme thermistor. Carbohydr
biosensors. Biosens Bioelectron 1990;5:13-26. Res 1982:12:273-82.
59. Roederer JE, Bastiaans GJ. Microgravimetric immunoassay with 83. Urban G, Kamper H, Jachimowicz A, Kohl F, Kuttner H, Olcaytug
piezoelectric crystals. Anal Chem 1983;55:2333-6. F, et al. The construction of microcalorimetric biosensors by use
60. Guilbault GG, Hock B, Schmid R. A piezoelectric immunobiosen- of high resolution thin-film thermistors. Biosens Bioelectron
sor for atrazine in drinking water. Biosens Bioelectron 1992:7: 1991:6:275-80.
411-9. 84. Birnbaum 5, Bulow L, Hardy K, Danielsson B, Mosbach K.
Automated thermometric enzyme immunoassay of human prom-
for the bioluminescent flow determination of ATP or NADH. Anal evaluated by a combined study of ellipsometry and radiotracer
Chim Acta 199O;235:43. techniques. i Colloid Interface Sci 1985:103:360-72.
102. Coulet P. Luciferase-based biosensors. Biosensors 1992:2-9. 125. Mandenius CE, Welin 5, Danielsson B, Lundstr#{246}m I, Mosbach K.
103. Maier CL. US Patent 4 104 029, 1978. The interaction of proteins and ions with affinity ligands co-
104. Starodub N, Arenkov P. Starodub A, Berezin V. Construction and valently coupled to silicon surfaces as monitored by ellipsom-
biomedical application of immunosensors based on fiber optics etry. Anal Biochem 1984:137:106-14.
and enhanced chemiluminescence. Optical Eng 1994:33:2958- 126. Stenberg M, Nygren H. A receptor-ligand reaction studied by a
63. novel analytical tool-the IsoscopeTM ellipsometer. Anal Bio-
105. Starodub N, Arenkov P. Starodub A, Berezin V. Fiber optic chem 1982:127:183-92.
immunosensors based on chemiluminescence and their appli- 127. Ruzgas TA, Razumas VJ, Kulys JJ. Ellipsometnic immunosensors
cation to determine different antigens. Sensors Actuators B for the determination of interferon and human serum albumin.
Chem 1994:18:161-5. Biosens Bioelectron 1992:7:305-8.
106. Robinson GA. Optical immunosensing systems-meeting the 128. Tronin A, Dubrovsky I, Denitti C, Gussoni A, Erokhin V. Nicolini C.
market needs. Biosens Bioelectron 1991:6:183-91.
144. Morgan H, Taylor D. A surface plasmon resonance immunosen- 166. Lukosz W, Stamm C. Integrated optical difference interferometer
sor based on the streptavidin-biotmn complex. Biosens Bioelec- as relative humidity sensor and differential refractometer. Sen-
tron 1992:7:405-10. sors Actuators 1991:25- 27:185-8.
145. Severs AH, Schasfoort RBM, Salden MHL. An immunosensor for 167. Schlatter D, Barnes R, Fattinger C, Huber W, HUbscher I, Hurst
syphilis screening based on surface plasmon resonance. Bio- 1, et al. The difference interferometer: application as a direct
sens Bloelectron 1993:8:185-9. affinity sensor. Biosens Bioelectron 1993:8:109-16.
146. Pollard-Knight D, Hawkins E, Yeung D, Pashby DP, Simpson M, 168. Heideman RG, Kooyman RP, Greve I. Immunoreactivity of ad-
McDougall A, et al. Immunoassays and nucleic acid detection sorbed anti-human chorionic gonadotropin studied with an opti-
with a biosensor based on surface plasmon resonance. Ann Biol cal waveguide interferometric sensor. Biosens Bioelectron
Clin 1990:48:642-6. 1994:9:33-43.
147. Attnidge 1W, Daniels PB, Deacon 1K, Robinson GA, Davidson GP. 169. Tiefenthaler K, Lukosz W. Integrated optical switches and gas
Sensitivity enhancement of optical immunosensors by the use of sensors. Opt Left 1987:9:137-9.
a surface plasmon resonance fluoroimmunoassay. Biosens Bio- 170. Nellen P. Tiefenthaler K, Lukosz W. Integrated optical input
electron 1991:6:201-14.