What Is HPLC Column Conditioning - How Long To Equilibrate
What Is HPLC Column Conditioning - How Long To Equilibrate
Depending on the usage conditions and sample cleanliness, most columns can undergo
potential changes such as chemical contamination, surface changes, and partial Disturbance
of the Silica material during use (Void). Such changes can lead to the deterioration of the
packing materials in the column, Which reduces the separation selectivity of certain
compounds. Such special separation selectivity is often impossible to reproduce. Resulting in
the need to develop methods or perform a lot of unnecessary diagnostic work for instrument
systems, chromatographic columns, and mobile phases in the future.
https://www.pharmaceuticalsky.com/2022/05/what-is-hplc-column-conditioning-how.html 1/5
8/13/22, 9:38 AM What is HPLC Column Conditioning? How Long to Equilibrate?
Clean, stable baselines are critical to good liquid chromatography, and properly conditioning
the HPLC column is a helpful way to ensure a good baseline. Analysts need to flush out the
storage solvent and equilibrate the column to the solvent mixture and may need to use an
intermediate solvent to ensure complete mixing and flushing.
Every reversed-phase HPLC column should be conditioned for the first time use, following
long-term storage, and when the significant changes in the mobile phase. The mobile phase
utilized in the conditioning was tought to be similar to those used in the consequent
chromatography.
https://www.pharmaceuticalsky.com/2022/05/what-is-hplc-column-conditioning-how.html 2/5
8/13/22, 9:38 AM What is HPLC Column Conditioning? How Long to Equilibrate?
Equilibration time is dependent on the flow rate and dimension of the column. In general,
flushing the column with ten up to twenty column volumes is enough for equilibration.
Therefore the flow rate has to be kept in mind. It is possible to increase the flow rate during
equilibration to shorten the equilibration time. If a column should be equilibrated with a
buffered mobile phase or with a mobile phase containing an ion-pair reagent, the
equilibration time should be extended.
When a column has just been installed on a reversed-phase HPLC system then it will
typically require between 10 and 20 column volumes before it is fully equilibrated and ready
to use. However, some applications are likely to require additional column volumes.
Examples are methods that include ion-pairing reagents and chiral HPLC methods. In these
cases, suitable equilibration should be investigated and documented for future use.
A starting point for investigation may be approximately 30 column volumes. When re-
equilibrating after a gradient injection has been run (prior to the next injection) 10 column
volumes are recommended. Although it refers to the column volume, the volume that is
interested in is more correctly called the void volume, Vm. This is the volume of the HPLC
column that is not taken up by the stationary phase. This is typically approximately 70% of
the total column volume.
There are two methods that we can use to calculate the void volume:
Method 1
Use the dimensions of the column and the formula for the volume of a cylinder:
https://www.pharmaceuticalsky.com/2022/05/what-is-hplc-column-conditioning-how.html 3/5
8/13/22, 9:38 AM What is HPLC Column Conditioning? How Long to Equilibrate?
Example
A HPLC column of internal diameter 4.6mm and length 10cm:
To calculate the required equilibration time simply multiply the calculated void volume (in
mL) by the number of required column volumes, e.g. 10, then divide by the flow rate (in
mL/min) to determine the total time required.
For a flow rate of 1 mL/min on the column above, this would result in an equilibration time
of 11.6 minutes for 10 column volumes.
Method 2
Inject an unretained solute to obtain t0, the column dead time (minutes). Then multiply this
by the flow rate to obtain the void volume:
Where F is the flow rate expressed in
mL/min.
Example
The flow rate for the method is 1.0 mL/min.
https://www.pharmaceuticalsky.com/2022/05/what-is-hplc-column-conditioning-how.html 4/5
8/13/22, 9:38 AM What is HPLC Column Conditioning? How Long to Equilibrate?
In fact, the required equilibration time is just t0 multiplied by the number of column
volumes required. In this example, the equilibration time is simply 1.05 x 10 = 10.5 minutes.