LIS Data format:
• Connect COM 1 to the PC using RS232
• Test name | Patient ID | Result | Unit | reference range | Prozone |
• Test Name = HbA1c
• Patient ID = 0000000001 ( 10 digit )
• Result = <3.00
• Unit = %
• Flag = Result < reference limit minimum ( Low) = 0
• Flag = Result > reference limit maximum ( High) = 2
• Flag = Result = between reference limit minimum and maximum ( Normal )= 1
Prozone Flag:
• Prozone detected = 1
• Prozone not detected = 0
Connection Details:
• RS232 Female cross connector
• Baud rate : 9600, Parity : None, Data Size :8 bit
• Connection type : Serial line (Cross connection 2-3, 3-2, 5-5)
• Serial line : COM X ( X- COM port of LIS connected PC )