Preoot eoeusen? Service Request Type: Corrective Repair
wavnegeheatthcare.com Service Request Number: 1-7931525186
@ GE Healthcare Service Report
Customer Name: TBNU SINA, RSI Location: Padang-Indonesia
Pos: Contact Phone: +6281363754493
Contact Name: Mr -Syahrom Email Address:
Initial Problem Description: Suspected detector issue
Eames
system 0 roszmotss itl auipment status: 1-5ystem Doan
Ser Number Enlements N
Produc Description nova 2000 16TAL
sei sens sett cos Date
fet Type Feld supper Engineer Name: Rasen Topon
Tube Se Reading
mroblem Desrpn trot Conmuniestn AC wth Oster
Detector performance was degraded /not really good
‘After UP / Running the system, when user test to expose, any error “power supply failure’ ond
system down
Cause/ Diagnosis: Performed reseat fiber optic cable in detector and RTAC, ond check fiber optic conition Igood)
Check communication RTAC from DL (OK)
Performed Detector diagnostic, there are error in detector voltage and ARC Anclog
check ond inspect to machine fcom, PDB ponel was OFF ond OMOI Trip
Performed OMDI reset
Resolution/ Service Performed: June 19th, 2020
Performed test exposed use cine and fluoro (OK!
Performed bad pixel there are error in segement density ond bad pixel density ond pike gain calibration
(ox)
Performed 0ST \some result any Passed ond no data)
take some photo for error log and results of cabration and diagnostics
tune 20th, 2020:
check ond inspect al fuse (0K)
check ond inspect continuity, ony shortin Rt [pin 13 -pin 14)
check and inspect input and output voltage Contactor Cl, in LI-L2 or L2-L3 only 190-220 VAC but if refer
to schematic diagrams should be 400 VAC
check ond inspect input and output voltage Trafo TRI and TR2, but output TRL ond TR2 are not voltage
Recommendation
tune 19th, 2020
‘According to me, the detector performand have been degraded or have decreased.
dune 20th, 2020
Suspect pat for next troubleshooting in below
= Trafo TRL ond TR2
“Contactor C1
orb
Poge No.1of2PT. GE Operations indonesia
South Quarter Tower 8, Lanta 18-19,
JLR.A Kortin Kav 8, Gilandak Borat Glendok
‘okarta Selotan Oki, Jakarta 12430
+6221 5730500
Preoot eoeusen? Service Request Type: Corrective Repair
‘snya.geheattncare.com Service Request Number: 1-7931525186
@ GE Healthcare Service Report
Ens
Date Time Hours Labour Type
June 20, 2020 0300-1100 200 Onsite Repair
June 19, 2020 0630-1300 650 Zone Travel
June 19,2020 13:00-2100 8.00 Onsite Repair
Rens
Part Number oy Description ‘Asset Serial No
ERC
Expense Date Type ty Comments
Sh ommesnnnme IS
20-Jun-2020 11:20 20-Jun-2020 11:23
RADEN TOPAN. SYAHRON
Poge No.20f2
Engineer Signature: