You are on page 1of 15

International Journal of Laboratory Hematology

The Official journal of the International Society for Laboratory Hematology

ORIGINAL ARTICLE INTERNAT IONAL JOURNAL OF LABORATO RY HEMATO LOGY

ICSH guidelines for the evaluation of blood cell analysers


including those used for differential leucocyte and reticulocyte
counting
INTERNATIONAL COUNCIL FOR STANDARDIZATION IN HAEMATOLOGY, WRITING GROUP: C. BRIGGS*,
N. CULP † , B. DAVIS † , G. D’ONOFRIO ‡ , G. ZINI ‡ , S. J. MACHIN § , ON BEHALF OF THE INTERNATIONAL
COUNCIL FOR STANDARDIZATION OF HAEMATOLOGY

*Department of Haematology, S U M M A RY
University College London
Hospitals, London, UK This revision is intended to update the 1994 ICSH guidelines. It is

Trillium Diagnostics, LLC, based on those guidelines but is updated to include new methods,
Brewer, ME, USA

Department of Hematology, such as digital image analysis for blood cells, a flow cytometric
Catholic University, Rome, Italy method intended to replace the reference manual 400 cell differen-
§
Haemostasis Research Unit, tial, and numerous new cell indices not identified morphologically
University College London,
are introduced. Haematology analysers are becoming increasingly
London, UK
complex and with technological advancements in instrumentation
Correspondence: with more and more quantitative parameters are being reported in
Carol Briggs, Department of the complete blood count. It is imperative therefore that before an
Haematology, 60 Whitfield
instrument is used for testing patient samples, it must undergo an
Street, London W1T 4EU, UK.
Tel.: +44 2034479882; evaluation by an organization or laboratory independent of the
E-mail: carolbriggs@hotmail. manufacturer. The evaluation should demonstrate the perfor-
com mance, advantages and limitations of instruments and methods.
These evaluations may be performed by an accredited haematology
doi:10.1111/ijlh.12201
laboratory where the results are published in a peer-reviewed jour-
nal and compared with the validations performed by the manufac-
Received 29 November 2013;
accepted for publication 20 turer. A less extensive validation/transference of the equipment or
January 2014 method should be performed by the local laboratory on instru-
ments prior to reporting of results.
Keywords
Haematology analyser,
evaluation/validation, digital
imaging, flow cytometry,
regulatory science

Automated analysers speed up the workflow in the


INTRODUCTION
laboratory and improve precision as more cells are
Since the previously published ICSH guidelines in counted and cell classifications are based on more mea-
1994, haematology analysers have evolved greatly. sured objective properties (light scatter, fluorescence,

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627 613
614 C. BRIGGS ET AL. | EVALUATION OF BLOOD CELL ANALYSERS

digital imagine, etc.). For leucocyte, differentials in a peer-reviewed journal. Such a focused validation
provide clinicians with more reliable data for patient should provide evidence that the analyser can meet
treatment as compared to manual microscopic meth- specific requirements within the test site. Certainly, a
ods. Data obtained in one laboratory on one instru- normal or reference range and other abbreviated vali-
ment should give comparable results to that provided dations must be performed by each laboratory before
from another laboratory using a different analyser. the analyser is used for patient testing. It should be
Several parameters have been introduced to the noted that claims by the manufacturer about through-
complete blood count (CBC), such as nucleated red put may be overestimated due to testing in optimal
blood cells (NRBC), immature granulocytes (IG), and conditions using relatively normal samples in nonclin-
cell indices such as immature reticulocyte fraction, ical environments. Therefore, the performance attri-
immature platelet fraction and red cell fragments, as butes of the device should be obtained in the
well as new red cell parameters for detection of func- environment where the instrument will be sited and
tional iron deficiency. New parameters related to cell by the routine staff that will be operating the instru-
size such as mean platelet volume (MPV) and red cell ment. The scope of any validation should depend
distribution width (RDW) have also been established. upon the availability of independent evaluation data,
Digital imaging analysis for all blood cells has also been the range of CBC parameters reported by the labora-
shown to be equivalent to manual microscopy with tory, the range of samples available for the validation
improved traceability as all significant images can be and regional laboratory regulations.
re-reviewed and stored for future reference [1], but
these too must be evaluated for both normal and
L E V E L S O F E VA L UAT I O N
pathological samples before introduction into clinical
laboratory practice. Manufacturers must conduct validation of their
The proposed reference for an extended flow cyto- instruments in accordance with European and Inter-
metric differential [2] to replace the CLSI manual national Conference on Harmonization (ICH) Guide-
counting method [3], while not intended for routine lines [4, 5], the Chinese State Food and Drug
laboratory use, but for manufacturers testing proto- Administration [6], the American National Standards
type and preproduction models, more accurate value Institute [7, 8] and U.S. FDA guidelines [9], Health
assignments to calibrator and control products and use Canada, Ministry of Health and Welfare of Japan and
at select sites performing a full regulatory evaluation . any other regional/national requirements.
A defined immunophenotypic reference method for In the USA, the FDA clearance process, which
high-accuracy nucleated cell identification will further encompasses all diagnostic devices, requires a descrip-
diminish subjectivity on cell classification and statisti- tion of the study design and the results of the studies.
cal imprecision, a problem for both microscope and These should be conducted to demonstrate that the
digital imaging systems. device shows an insignificant risk of yielding errone-
Advances in instrumentation have resulted in some ous results in the hands of the intended user [10]. A
parameters, previously only indicated by abnormal laboratory in the USA buying an instrument cleared
cell flags, are now being quantitated, such as NRBC, by the FDA should have confidence that the product
fragmented RBCs, IG and some new reticulocyte and meets FDA requirements; however, these instruments
platelet maturation parameters. Choosing a new ana- may have different requirements when sold outside
lyser is an important decision; research should begin the USA. Europe has its own in vitro diagnostic
with government or peer-reviewed reports on com- devices directive, which is currently under rapid tran-
parative evaluations, as well as information from sition. In the European community, a CE mark is
instrument manufacturers. A laboratory purchasing required indicating that the manufacturer, its autho-
an evaluated instrument may perform an abbreviated rized representative or the person placing the product
assessment/validation appraising aspects of the equip- on the market or putting it into service asserts that
ment in its intended location. Verification/validation the item meets all the essential requirements of the
is a conformation of the evaluation performed by a relevant European Directive(s) [11], but an indepen-
manufacturer or other centre of excellence published dent evaluation is still necessary. The Australian

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627
C. BRIGGS ET AL. | EVALUATION OF BLOOD CELL ANALYSERS 615

Therapeutic Goods Administration is seeking to har- The instrument manufacturer should be responsible
monize its decision-making with that of overseas for instrument installation, set-up and initial calibra-
authorities, including the FDA. This may offer an tion using the method specified by the manufacturer.
opportunity to harmonize diagnostic device approval A written report on the calibration and performance of
systems internationally. In Japan, for international the instrument using control material and/or patient
standardization on reference measurement procedures samples should be provided to the evaluating or vali-
of haematology, manufacturers are required to har- dating laboratory.
monize with the internationally recognized method
shown by the Japanese Society for Laboratory Haema- Training. The supplier should provide training for the
tology (JSLH). intended evaluators. This should encompass the
In some countries, a national evaluation may be principles of the methods of measurement, operation
carried out by an official organization at an approved and maintenance of the instrument, and
centre and performed in accordance with the proto- troubleshooting. The instruction manual supplied by
col for the evaluation of blood analysers produced by the manufacturer should cover these topics in more
ICSH [12] and in accordance with this updated detail: the evaluation should include assessment of
guideline; however, these organizations are increas- the instruction manual quality and completeness. It is
ingly being reduced. Where a national evaluation is vital that manufacturer’s instructions are carefully
not available, an evaluation published in a peer- followed.
reviewed journal should be sought. The local labora-
tory user may wish to perform a less extensive Blood samples. Fresh, human whole blood samples
assessment/validation that appraises aspects of the used in the evaluation should be anticoagulated using
equipment and user-dependent steps in its intended K ethylene diamine tetraacetic acid (K2EDTA or
location or focused on any unique aspect of the med- K3EDTA as specified by the manufacturer), and its
ical practice at that institution. The recommendations concentration recorded. Advice should be sought from
for evaluation of coagulation analysers [13] provide the manufacturer on the minimum volume of blood
some general advice relevant to other haematology needed for testing in both automatic and manual
analysers. modes, including dead space. Blood samples should be
processed within 4–8 h of venesection, with the
exception of those samples used for the assessment of
S TAG E S O F P E R F O R M I N G A F U L L
sample stability. Transport and storage of samples
E VA L UAT I O N
should satisfy the conditions for appropriate
international or national codes of safe practice [14]. All
Preliminary information required from the manufacturer
blood specimens should be surplus after all other
and planning the technical evaluation
laboratory testing is complete and would otherwise
have been discarded. No donor identity should be
Instrument installation
recorded other than a numerical specimen coding
Methodology and principles of operation should be system that allows comparison between instruments
assessed and validated with the support of existing lit- and methods. Clinical information is permitted as long
erature. as it does not make the patient identifiable, and even
Planning of the evaluation should include an esti- necessary for the study of pathological samples.
mate of the availability of internal resources of the Planning is needed to ensure that samples are available
evaluating organization. As a preliminary step, a selec- with a wide variety of quantitative and qualitative
tion should be carried out of adequate reference abnormalities. The range of samples tested should
methods, the evaluation of staff, statistical tools and cover the entire clinical reportable range and include
experienced morphologists: their predictable working the most severe abnormalities encountered by the
time and availability should be verified in advance. laboratory. Samples should be included with possible
Establish the range of instrument parameters that interfering substances, such as lipid, high bilirubin
have to be evaluated and should be selected. concentration, haemolysis or the presence of

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627
616 C. BRIGGS ET AL. | EVALUATION OF BLOOD CELL ANALYSERS

Table 1. Abnormal samples and potential interfering substances that should be included in the evaluation of the
haematology analyser

Interfering
WBC RBC Platelets substances

Extreme leucocytosis Sickle cells Giant platelets Haemolysis


Extreme leucopenia Target cells Platelet clumps Cryoglobulins
Neutrophilia
Lymphocytosis
Monoctyosis
Eosinophilia
Basophilia
Blast cells Fragmented cells Immature Platelets Paraproteins
Atypical lymphocytes Microcytic cells CD61 labelled platelets High bilirubin
Smear/smudge cells Macrocytic cells Lipaemia
Immature granulocytes Spherocytes
Left shift/band neutrophils Extreme polycythaemia
CD3/CD4/CD8 Lymphocytes Extreme anaemia
Nucleated red blood cells
Reticulocytosis
IRF
Low Retic Hb Conc/Content
Howell–Jolly bodies
Heinz bodies
Pappenheimer bodies
Malarial parasites

WBC, white blood cell count; RBC, red blood cell count; IRF, immature reticulocyte fraction; Hb, haemoglobin; Retic,
reticulocytes; Conc, concentration.

cryoglobulins. Examples of diseases and interfering computer spreadsheets. It is also mandatory to keep the
substances to be included are listed in Table 1. Not all original instrument printouts of results for future
parameters are available on all instruments. Samples reference in some regions. Records should be kept
with visible clots should be excluded, but those with of instrument downtime, with reasons for any
platelet clumps and/or red cell agglutinates and breakdown, service response times and maintenance
detected microscopically should be included, as this schedules. An operator log with the name and
will allow assessment of the platelet clump flag. One- professional level of the operator should be used to
third to half the total number of samples should be record any instrument problems encountered, as well
from normal or nondiseased individuals. Comparability as reagent and control usage, batch numbers and
results can be unreliable if the proportion of normal expiry dates.
samples included in the evaluation is too high or too
low.
Preliminary assessment
Blood films. At least two blood films should be Advice on the safety of the instrument should be
prepared for each sample. These should be made sought from the manufacturer and is usually included
using the laboratory’s standard protocol. For digital within the instrument operator manual. Before any
image analysis systems, the stain may need to be evaluation can proceed, an assessment of the safety of
optimized according to the manufacturer’s instruction. the instrument should be made. Health and safety,
legal and insurance implications, data storage and
Records. Accurate records of all results, including security, efficiency and ease of operation should be
quality control data, should be kept in worksheets or assessed and recorded. The staff skill level intended to

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627
C. BRIGGS ET AL. | EVALUATION OF BLOOD CELL ANALYSERS 617

use the instrument should be taken into consideration evaluation, so should the clarity of data and graphics,
as well as cost-effectiveness. Failure of any of the fol- validation process, quality control programmes and
lowing categories, electrical, mechanical or chemical, data storage and retrieval.
should result in the suspension of the evaluation.

Performance assessment
Electrical. Electrical specifications should comply with
appropriate national or international standards, such Sample mode. Where the instrument has multiple
as the Gepr€ ufte Sicherheit (GS) mark, American sampling modes utilizing separate sampling pathways
National Standards Institute (ANSI), Conformit e (such as automatic from a closed tube and manual
Europeene (CE) mark, Underwriters Laboratories (UL) from an open tube or prediluted sample),
listing, Japanese Industrial Standards (JSA) and performance should be tested for all methods.
Canadian Standards association (CSA) mark. In Precision, carry-over and linearity (or parameters as
addition, the manufacturer may also cite various stated by the manufacturer) should be assessed in all
international standards with which the instrument modes and comparability between modes on a limited
complies. number of samples (at least 30).

Mechanical. Checks should be made for any hazards Precision. Precision may be defined as the closeness of
that may cause injury, such as exposed moving parts agreement between test results when a sample is run
and sharp edges. repeatedly. It depends on the distribution of random
errors and is not a measure of accuracy. Imprecision
Chemical. Any reagents that may be corrosive, rises with lower cell concentrations because fewer
carcinogenic or toxic should be considered, for cells are counted. Imprecision can be reported as the
example cyanide reagents used in the measurement of standard deviation (SD) or the percentage coefficient
haemoglobin. All reagents used must have undergone of variation (CV%), which is the SD expressed as a
complete control of substances hazardous to health percentage of the mean value of the replicate
(COSHH) assessments and should be supplied with the measurements on the same sample within a short
material safety data sheet information. Different period of time. An increasing SD or CV% indicates
countries have different labelling requirements, but increasing imprecision. Where practical, precision
these should conform to local requirements. Any should be established for the full reportable range of
reagents that are premarketed should be labelled as for each measurand, this includes low counts especially
performance evaluation only or for research use only. for haemoglobin and platelets at the transfusion
threshold.
Microbiologic. Ideally specimen analysis should be by
closed-vial sampling; however, all instruments should Within-run precision. This is the repeatability and
be tested for infective aerosols. The manufacturer usually consists of a single run of 10 measurements
should perform a microbiologic assessment by testing on the same sample, with all reported parameters
for aerosolization and surface contamination using a analysed. Normal, abnormal low and abnormal high
fluorescent bacterial tracer [15]. All procedures should samples for white blood cell count (WBC),
conform to the appropriate legislation [16–18]. haemoglobin concentration (Hb) and platelets should
Protocols must also be available for the disinfection be sought. Low and high values should be outside the
and decontamination of equipment, containment of reference limits as low and high as samples seen in
spillages and disposal of waste and samples. the laboratory, for example chemotherapy patients,
polycythaemic patients and patients with untreated
leukaemia. If the analyser reports reticulocytes or
Sample identification and handling
NRBC, samples with values around clinical decision
Sample identification either achieved by barcode or by points for these parameters should also be tested [19].
manual input should be assessed. The reliability of Data need to be collected for all parameters reported
barcode readers should be monitored throughout the by the instrument not just those listed above. The

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627
618 C. BRIGGS ET AL. | EVALUATION OF BLOOD CELL ANALYSERS

results for mean, SD and CV% should be calculated include the entire analytical measurement range, from
[12]. the highest counts to the lowest. In laboratories where
very low WBC and platelet are encountered, it is
Between-batch precision. Between-batch precision advisable to examine linearity in the low range sepa-
may be affected by calibration or drift. A single rately. For example, a sample with a platelet count of
measurement on the same sample repeated each day 50 9 109/L could be serially diluted down to a count
for a period 20–30 days is used to measure the total of 5 9 109/L or a WBC of 2.0 9 109/L to 0.2 9 109/L.
between day (batch) precision for all parameters. As Replicate tests should be performed to give results at
for within-run precision, abnormal low and abnormal evenly spaced concentrations, for example reducing in
high samples for WBC, Hb, platelets, reticulocytes and increments of 10% from 100% to 0%. Group AB
NRBCs, if appropriate, should be included. Samples serum or the diluent reagent for the analyser may be
are required for a long period of time, and fixed blood used as the diluent; however, certain haematological
may be required; it may be convenient to use quality parameters, for example, the red cell indices or per-
control material supplied by the manufacturer for this centage reported results, will not be affected by dilu-
purpose where all parameters available on the tion of the sample. Some commercial companies
instrument will be assessed. For some parameters and provide products that can be used for linearity checks
from some manufacturers, this is not the case and for validation of reportable ranges if no patient sam-
laboratories should treat the reporting of these ples are available, such as Streck, international@streck.
parameters with caution without specific internal com, R&D Systems, CustomerService@RnDSystems.
quality control [19, 20]. As these samples are com. As part of the evaluation, the analytical measure-
analysed at any time during the daily workload, the ment interval (AMI) and the clinically reportable
effect of carry-over from high to low specimens needs interval (CRI) should be assessed.
to be considered and should be assessed in separate The AMI is the range of analyte values that a
studies see below. method can directly measure on a specimen without
any dilution, concentration or other pretreatment not
Carry-over. Carry-over is defined as the
part of the usual assay process. In some instances,
contamination of a sample by the sample analysed
AMI may be considered synonymous with linearity.
immediately preceding it [21]. The clinical concern is
Verification of AMI may be accomplished by evalua-
that results from a high sample analysed routinely
tion of known samples of abnormal high and abnor-
may elevate the results in a cytopaenic or anaemic
mal low values or by the use of high and low
sample analysed subsequently. Carry-over from a high
calibration materials with known values. The CRI of
sample to a low sample should be assessed by running
patient test result is the range of analyte values that a
sample A (high sample) three times, A1, A2, A3,
method can measure, allowing for specimen dilution,
followed by sample B (low sample) three times, B1,
concentration or other pretreatment used to extend
B2, B3. This should be performed at least three times
the direct analytical measurement range. The CRI is
for WBC, Hb, platelets, reticulocytes and NRBC.
based on diagnostic needs and is a clinical decision by
Percentage carry-over is calculated by:
the laboratory authorities, which is related to dilution
B1 B3 and concentration protocols used in the laboratory
 100 [22–24]. The verification of reportable range may be
A3 B3
accomplished by evaluation of known samples of
Linearity. Regulatory Affairs agencies require that a abnormal high or abnormal low values or with low
laboratory validates an instrument for the reportable and high calibration materials with known values.
range before patient testing. Assuming no constant Precision of results from the analyser will affect linear-
bias, this is the ability to provide results that are ity results so this needs to be taken into consideration
directly proportional to the concentration of cells. and may be best tested before linearity. Limit of blank
There should be a linear relationship for the param- (LoB) and limit of detection (LoD) are measurements
eter measured at various dilutions over as large a determined from precision and linearity procedures
range as possible. Dilutions should be chosen to and biases of the instrument. This may be more

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627
C. BRIGGS ET AL. | EVALUATION OF BLOOD CELL ANALYSERS 619

important for very low cell counts of WBC, red blood flow from the vein to avoid haemoconcentration and
cell (RBC) and platelets and for body fluid counting. clotting. If appropriate, ranges for neonates and
A regression graph should be plotted, with concen- children of different ages should be tested. Mean, SD
tration on the x-axis and cell count or concentration and 95% confidence ranges should be calculated for
on the y-axis. The regression line should pass through each group. The reference interval for each
the origin and the R-value (correlation coefficient) measurand having a normal distribution (not skewed)
should be as close to 1.0 as possible, but cells in low within a group is defined from the lower bound
numbers will always show higher variability between (mean 2 SD) to the upper bound (mean + 2 SD).
methods due to counting statistics. Mann–Whitney U-test for percentiles and confidence
limits is applied, if the population results are not
Sample stability. Sample stability may be defined as normally distributed. Residual samples from the
the ability of a sample to retain a consistent value for laboratory may be used if all testing is complete; but if
a measured quantity over a defined time period and volunteers are bled, informed consent should be
within specific limits when stored under defined sought in line with regional ethical guidelines.
conditions [25]. A change in the measured quantity of
various components of the CBC over time following Accuracy. Accuracy is defined as the closeness of
venesection is a well-known phenomenon, and there agreement between the result of a measurement and
may be a higher or lower change in the direction of a known true value. The concept of a true value for
results. To determine changes on different cell counts many components of the CBC is often not applicable
or CBC parameters, blood should be taken from as the true value, which is obtained by a definitive
five normal individuals and five patients with reference method and may not be available. The only
abnormalities of different cell lines. Analysis is relevant parameters that can be estimated correctly
performed at time zero (or as close as possible), the are Hb [28], packed cell volume (PCV) [29, 30], RBC
blood sample should then be divided into two sets of count and WBC [31], platelet count [32], reticulocyte
six aliquots, one set stored at room temperature count [33, 34] and differential leucocyte count [3,
(which should be recorded) and the other at 4 °C. 19]. The haemiglobincyanide (cyanmethaemoglobin)
Subsequent testing should then be performed after 4, method is the internationally recommended method
8, 12, 24, 48 and 72 h. Samples stored at 4 °C should for determining the haemoglobin concentration of
be allowed to come to room temperature before blood. The basis of the method is dilution of blood in
analysis. The effect of storage time and temperature a solution containing potassium cyanide and
assessed by plotting differences from the initial, or potassium ferricyanide. Haemoglobin and HbCO, but
time zero, results against times of testing. The effect of not haemoglobin sulphate, are converted to
precision of results also needs to be considered and haemoglobincyanide. The absorbance of the solution
should be assessed in separate studies. is then measured in a spectrometer at a wavelength of
540 nm or a photoelectric colorimeter with a yellow
Reference intervals. To establish the clinical utility of green filter [35]. Most automated counters measure
an instrument for diagnosis, screening and monitoring haemoglobin by a modification of the manual method
of disease, it is necessary for the laboratory to with cyanide reagent or with a nonhazardous
establish a reference range. Reference ranges specific chemical such as sodium lauryl sulphate, which
to the instrument, for all components of the CBC, avoids possible environmental hazards from disposal
should be calculated during the instrument evaluation of large volumes of cyanide-containing waste.
[26, 27]. Ideally at least 120 samples, from apparently Whatever method is used for Hb measurement, the
healthy individuals, all normal values for CBC and haemoglobincyanide method is the only accepted
differential parameters, of each sex, 60 from each, are reference method.
tested within 4 h of venesection. Sex and ethnic For reticulocytes, a flow cytometric reference
groups should be examined where possible. Special method has been proposed, which has shown to be
care should be used to ensure adequate venipuncture more precise than the manual method using new
technique, time of tourniquet application and blood methylene blue [36–38]; however, the manual

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627
620 C. BRIGGS ET AL. | EVALUATION OF BLOOD CELL ANALYSERS

method, which is subjective and imprecise, currently analysed using the paired t-test (when results are
remains as an accepted reference method. CLSI normally distributed) as linear regression analysis may
H26-A2 [19] states that the reference is the flow reti- show good correlation even when there is a bias
culocytes method with manual reticulocytes as an between results from the two instruments. For non-
alternative as this not always available in routine lab- Gaussian data, the Wilcoxon rank sum test or Mann–
oratories. Whitney U-test should be used for paired data. A
In practice, many laboratories would compare the P-value of <0.05 is usually considered as statistically
instrument under evaluation to results from the significant for all the listed tests. Any samples with
instrument in routine use, with the exception of the extreme results where the reason can be explained
differential leucocyte count, where it is recommended should not be merged with other results; this would
that the results are compared with the reference 400- influence the statistical analysis. When discrepant
cell manual differential [3]. However, for a national results are found between the current instruments
evaluation, the available reference methods, listed and those under evaluation, the samples should,
above, may be used if results between systems are sig- where possible, be measured using the reference
nificantly different. methods previously described. Blood films should be
made on all samples analysed and examined
Comparability. A comparison of the results from the microscopically.
evaluation instrument with those obtained by the Some parameters are only available on a single or
current routine procedures should be made for as limited number of instrument types, such as the per-
many normal and abnormal samples as possible, centage hypochromic red cells, reticulocyte haemoglo-
together with samples with interfering substances; bin content/concentration or the immature platelet
normal samples should be half to one-third of the fraction. In these cases, it may not be possible to com-
total. Total number should be at least 250–300 for a pare the results to another method; in such circum-
full validation. Table 1 lists the samples that should stances, the results should be assessed to determine
be included in the evaluation. Samples should be whether they are consistent and appropriate with the
measured over a period of time, at least a week or diagnosis and clinical condition of the patient. Where
more, to assess variability of the instrument and daily routine haematology analysers use monoclonal anti-
differences in the patient population in the bodies and flow cytometry methods for the measure-
laboratory. Results should be presented graphically ment of some cells such as platelets labelled with
showing the difference between the result from the anti-CD61 and lymphocyte subsets with anti-CD4 and
instrument under evaluation (y-axis) and the routine anti-CD8, the results should be compared with the
instrument (x-axis). Regression analysis, correlation results from a validated flow cytometer using guide-
and in particular Bland–Altman [39] plots should be lines for performing CD4+ T cell populations [42] or
used to assess agreement [40]. The correlation the current routine laboratory flow cytometric
coefficient is a measure of how well the data fall on a method for the measurement of CD4+ and CD8+ lym-
straight line; the closer the correlation is to 1.0, then phocyte subsets.
the more linear. Slope, intercept, correlation or bias
assessments are all important information. It is Reference leucocyte differential including NRBC and
important to know the medical diagnosis of outliers IG. It is recommended that the reference 2 9 200-cell
when comparing methods, as it is not possible for leucocyte differential [3], including NRBC and IG
instruments to correctly identify all cells, for example counts, is performed by two experienced examiners on
leucemic blast cells. It is important to look at mean all samples and a third person if the first two results
difference or Bland Altman plots in leukopenic, disagree. Different automated counting techniques
anaemic or thrombocytopenic samples, as differences deal with abnormal cells differently; some analysers
in low values may be masked when analysing the now enumerate NRBC and IG where previously their
whole data set. Total analytical error (based on presence was indicated only by an abnormal cell flag.
analyses of differences) should be calculated [40, 41]. Blast cells are indicated by a flag and should be
Paired results from the same sample should also be correlated with morphological findings. A similar

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627
C. BRIGGS ET AL. | EVALUATION OF BLOOD CELL ANALYSERS 621

approach should be used for the red cell fragment flag adequate feather edge to find cells. Automated blood
or research count available on some instruments. For a films are usually more consistently made and stained.
full evaluation or for manufacturers, a flow cytometric The systems are designed to work with different stain-
differential will in future be recommended [2]. The ing protocols, and the manufacturers may recommend
use of digital imaging systems for leucocytes may be specific settings or provide an instrument protocol to
used instead of the microscope, but these will have to assess the individual laboratory’s current slide mak-
be evaluated/validated before introduction to the ing/staining protocol.
evaluating or routine laboratory. This should be Quality control on these systems allows the labora-
performed using the 2 9 200-cell manual leucocyte tory to assess stain and smear quality within day or
differential or the flow cytometric method using day-to-day using in-laboratory prepared blood films as
both normal and abnormal samples and include IQC slides.
morphological abnormalities of all cell lines. Evaluation of these digital cell locator/preclassify
systems needs to encompass performance characteris-
Digital-image-based haematology systems. Digital tics such as reproducibility/precision, accuracy, compa-
imaging systems have a long history, starting in the late rability and clinical sensitivity and specificity. The CLSI
1960s and the early 1970s [43, 44]. These systems H20-A2 [3] is considered the reference method for the
performed a five-part WBC differential and morphologic cell locator/preclassify system if a previously cleared
examination on wedge smears. The cost of the systems, cell locator/preclassify system is available, that can be
continued maintenance, system performance limitations used instead of a protocol based on H20-A2 [3].
(based on wedge smears, cell distribution on the slide Because blood samples are not aspirated by the cell
and staining consistency) followed by the availability of locator/preclassify systems, carry-over, linearity and
three-part and then five-part automated differentials on LoB/detection/quantification are not applicable to
routine haematology cell counting analysers contributed these systems. Sample stability also does not strictly
to the demise of these original systems. apply in the cell locator/preclassify system even though
With the improvement of computer systems, sample stability is important to determine whether
sophisticated processing and graphics software and the cells can be identified in older samples. Interferences
advent of artificial neural networks, image-based sys- with cell locator/preclassify systems are minimal (rou-
tems have returned to routine haematology laborato- leaux, RBC agglutination and platelet clumps) as these
ries. The systems available are two types: typical interferences can be visually seen if present.
• Cell locator systems with preclassification of cells for Individual laboratory reference ranges established
verification of cells by a skilled operator. for manual or automated differentials should be veri-
• Image-based haematology systems that perform a fied with the installation of any system.
CBC, five-part normal differential and, in some instru-
ments, a reticulocyte count. If any abnormal cells are Digital-image-based CBC/differential/reticulocyte
detected, the sample is flagged by the instrument for haematology analysers. For the image-based systems
further review by a skilled operator. that perform a CBC and identify normal WBCs and
reticulocytes, a full evaluation of all performance
Digital cell locator/preclassify systems. Currently two characteristics must be performed as they are
digital cell locator/preclassify systems are cleared by applicable to these instruments. (Currently no system
the FDA as a ‘Cell Locator with preclassification of is FDA cleared at this time). See Table 2 for a
cells for verification by skilled operator’ with final comparison of performance characteristic testing that
classification of all cells (normal and abnormal) should be examined for each type of digital imaging
performed by the operator. Other systems are system.
available outside of the United States. Testing such as carry-over, linearity and LoD/quan-
The digital cell locator/preclassify systems are titation may have to be modified from the more com-
highly dependent on the quality of the blood films mon methods to accommodate differences in the
and the stain quality for accurate identification of image-based systems vs. ‘traditional’ impedance or
cells. Manually made blood films must have an optical methods used my most haematology analysers.

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627
622 C. BRIGGS ET AL. | EVALUATION OF BLOOD CELL ANALYSERS

Carry-over assessment could require the use of different haematolymphoid lineage markers of
modified samples (such as a serum blank or manufac- leucocytes [2]. This improves precision as many more
tured sample with low RBC count and depleted of cells are counted, and additional cells not identified
WBCs and platelets) as the ‘low’ sample analysed after microscopically can be further classified. Any
‘high’ samples to detect carryover in WBCs, RBCs, subjectivity in classifying cells microscopically is
haemoglobin and platelets with some of these sys- markedly decreased. A preliminary study has been
tems. Eliminating WBCs and platelets from a low performed [2] using three different six colour
sample could be difficult to achieve [45]. It is impor- combinations of monoclonal antibodies and
tant that consideration is given to carryover of abnor- fluorochromes. This preliminary study performed
mal cells. under the auspices of ICSH indicated that an eight or
Interfering substances that impact impedance or more colour panel was required to achieve the desired
optical-based haematology analysers (lipemia, biliru- specifications for a new leucocyte differential method.
binemia, paraproteins, etc., see Table 1) will not The specification for the immunophenotypic method is
impact image-based determination of CBC/Diff/retic that with a single analysis clear identification can be
parameters. Extreme situations for rouleaux or cold afforded to lymphocytes (defining various subsets),
agglutinins/RBC agglutination may affect CBC/Diff/ monocytes (activated, resting and immature),
Retic results requiring the preanalytical treatment of granulocytes (mature and immature), plasma cells and
these types of samples. Morphologic abnormalities (as account for ‘blast’ cells and nucleated red cells. The
outlined in Table 1) should be included in the evalua- final proposed reference method for leucocyte
tion. Both the image-based systems’ haematology counting, intended as a more robust replacement of the
analysers and cell/preclassify systems include recogni- CLSI H20-A2 guideline, currently is under further
tion or counting of these abnormalities in the analyser evaluation and validation by an ICSH working group
software. using the most recent ICSH/ICCS Guidelines for
Validation of Cell Based Fluorescent Assays [46]. The
Flow cytometric immunophenotypic counting antibody combinations used in the study are
methods. An alternative method to the manual summarized in Table 3.
microscopic method of leucocyte counting, expected to
improve precision and accuracy, is a flow cytometric Abnormal cell flags. As well as cell counts, the
method using specific monoclonal antibodies to morphology of the cells needs to be assessed to

Table 2. Performance characteristic testing comparison for image-based haematology systems

Digital image
Performance validation testing cell locator Digital image CBC/Diff/Retic

Precision/repeatability YES YES


Within run YES YES
Between run (day to day) YES YES
Between techs YES YES
Accuracy YES YES (Reference H20-A2)
Comparability YES YES (compare to H20-A2 and predicate haematology analyser)
Sensitivity/Specificity YES YES
Sample stability NO YES
Carryover NO YES/NO (if cartridge system)
Linearity NO YES
Limit of blank (LoB)/Limit of detection NO YES – May be modified or adapted to analyser methodology
(LoD)/Limit of quantitation (LoQ)
Reference intervals NO YES

CBC, complete blood count.

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627
C. BRIGGS ET AL. | EVALUATION OF BLOOD CELL ANALYSERS 623

compare the efficiency of the suspect abnormal cell limits are calculated and returned to the participating
flags generated by the instrument. Most instruments laboratory. All reportable results should have an
will generate flags in the presence of abnormal white assigned value for internal quality control (stabilized
cells, red cells and platelets. In other circumstances, a blood supplied by the manufacturer). This is currently
flag may be generated to indicate that the instrument not true for some parameters on some instruments.
results are unreliable or corrupted, and should not be This brings into question whether these parameters
used. The ideal is that the instrument identifies should be used for clinical decision-making, at least in
abnormal cells for further study in the laboratory. The the minds of laboratory regulator. Ideally, this is true
sensitivity, specificity, positive predictive value, for EQA but this is not always possible for instrument-
negative predictive value and overall efficiency should specific or new parameters. It should be noted that for
be calculated for each individual flag that relates to the some extended parameters and even some established
morphology or presence of abnormal cells according to parameters such as NRBC, IG, mean platelet volume,
Galen and Gambino [47]. Blood films need to be immature platelet fraction, immature reticulocyte
examined on all samples. Sensitivity is the ability to fraction and red cell distribution width, there is no
produce positive results in the presence of abnormality EQA scheme available in many countries [20, 45].
and specificity, the ability of the instrument to show
negative results when an abnormality is not present. Local validation/transference of a haematology
The method for the classification of results is shown in analyser. In all cases, the performance of a new system
Table 4. should be compared with the current in-house system
using a range of samples designed to test for sensitivity,
Quality assurance. It may be useful to register the specificity, reproducibility and results robustness.
instrument into an accredited external quality
assurance (EQA) scheme that will give an indication of
the instrument’s performance in comparison with
Table 4. Calculations for the assessment of efficiency
others. Partially or fully stabilized blood is usually used
of morphologic classification of cells (a) Definition of
for EQA schemes (CAP, UK NEQAS or others) and sensitivity, specificity, predictive value and overall
may not perform in the same way as fresh blood. efficiency and (b) Definition of true/false positive/
Results are often instrument specific, especially for negative
leucocyte differentials, platelet counts and reticulocyte
Efficiency measure Calculation
counts. Instrument groups from multiple laboratories
are compared against each other rather than across all (a)
Sensitivity (positivity in TP  100
groups of instruments. Target values and acceptable TP + FN
abnormal samples)
Specificity (negativity in TN  100
TN + FP
normal samples)
Table 3. Monoclonal antibody combination protocols, Predictive value of a positive TP  100
along with DNA dye Syto16, protocols under TP + FP
result
evaluation by ICSH study group as candidate method Predictive value of a negative TN  100
for immunophenotypic leucocyte differential TN + FN
result
reference method Overall efficiency TP + TN  100
TP + FP + TN + FN
Antibody specificity Fluorochrome
Results of test method
CD16 Pacific Blue
Reference Positive
CD45 Krome Orange
method (abnormal) Negative (normal)
Syto16 FITC
CD7 PE (b)
CD123 PerCP-Cy5.5 Positive TP (true positive) FN (false negative)
CD14 PE-Cy7 (abnormal)
CD3+ APC Negative FP (false positive) TN (true negative)
CD11b APC-A750 (normal)

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627
624 C. BRIGGS ET AL. | EVALUATION OF BLOOD CELL ANALYSERS

If a complete evaluation, as described in this guide- presentation and compatibility with other systems
line, has not been performed, a literature search should also be considered. New instruments should
should be undertaken to find independent peer- generate results that are comparable both in the units
reviewed evaluations of the equipment. Validation or reported and reference ranges to those already in
transference may be defined as establishing docu- place in the laboratory including other instruments in
mented evidence that provides a high degree of assur- the laboratory and point-of-care (POC) instruments
ance that a specific process will consistently produce within the same institution. The instrument should be
results meeting its predetermined specifications and able to process those sample tubes most commonly
quality attributes. Instrument precision, comparability received by the laboratory.
to reference methods, linearity, carry-over and drift A plan, with a realistic time-scale should include
should have been assessed during any formal national the time to obtain the relevant specimens, maintain
evaluation, but still need a brief validation and refer- records, perform analysis of results and statistical pro-
ence range study. In general, the routine laboratory cess to be used, and write a final report. This is partic-
should perform an abbreviated validation based on ularly important when the instrument is loaned or
the evaluation described above in this document. It is leased. The quantities of reagents and consumables
necessary for the laboratory to determine whether or required for the evaluation must be estimated.
not the reference ranges are the same as with existing Arrangements must be in place for service and main-
instruments, at least 120. If the laboratory tests neo- tenance of the instrument during the evaluation per-
natal or paediatric samples, these should also be iod and, if training is available, this must take place
examined separately. If paediatric samples are tested before the evaluation begins.
in open mode, this should also be assessed. On many A signed statement from the manufacturer or sup-
analysers, capillary blood can be analysed using plier that the instrument installed has undergone test-
diluted blood in the analyser-specific diluent. Dilution ing and calibration and are all within predetermined
of blood to diluent will vary, but usually the instru- limits. Appropriate grades of staff should be selected
ment will calculate the whole blood results if run in to complete the evaluation and their availability
capillary mode. If this is used in the routine labora- ensured. Staff with skills similar to those of the poten-
tory, then this should be assessed with at least 30 tial users of the equipment should be used for the
samples compared with the whole blood mode. evaluation.
Before the start of any evaluation, it is important The purchaser should perform a local assessment of
to ensure that the instrument is compatible with the the instrument and validation of the results. Prelimin-
laboratory and the service for which it is intended. ary information regarding the instrument installation,
The evaluator should obtain the following informa- training and suitable blood samples is required and
tion: – name, manufacturer and distributor of the sought from the manufacturer before planning the
instrument, list price including options for rental or technical evaluation. A written protocol should be
leasing, reagent and consumable costs, and terms of established that specifies how validation will be con-
service contracts. The overall dimensions of the ducted. Firstly, the laboratory should establish that
instrument, power requirements, drainage, opera- performance specifications are the same, or better, as
tional environment (temperature range) and heat pro- those demonstrated by the manufacturer and on par
duced by the instrument should be investigated. with previous published reports. Comparability of
Information on consumables (reagents, controls and results, reference ranges, reliability and acceptability of
calibrators), formulation and shelf life of reagents, the instrument are also considerations in a local evalu-
quantity and number of tests possible, and storage ation and should be analysed as for a national evalua-
requirements would be part of any complete instru- tion. Fewer samples are needed to validate compared
ment assessment. The ability of the instrument to be with the full evaluation but at least 50 should be used
interfaced with the laboratory/hospital information for comparability and reference ranges.
system should also be understood at this stage. In the future, more core/centralized laboratories
The repertoire of parameters available, measure- will be implemented with samples travelling some dis-
ment principles, minimum sample volume, data tances and over time for testing, up to 48 h. The

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627
C. BRIGGS ET AL. | EVALUATION OF BLOOD CELL ANALYSERS 625

results for these samples need to be validated to be assessed, together with the process of result
ensure that results given to clinicians are correct and validation, data storage capacity, speed of retrieval of
clinically safe. results and the quality control programs available.
Some guideline documents, such as the Clinical
Laboratory Standard Institute (CLSI) H20-A2: Refer- Training. The quality of training given by the
ence Leukocyte (WBC) Differential, state that at least manufacturer should be described, as well as the ease
200 samples should be included in the study (1/2 nor- of use and clarity of the operator’s manual.
mal, 1/2 abnormal) [3, 19, 45]. As a minimum, good Laboratories should be on the manufacturer’s contact
laboratory practice should dictate that at least 50–100 list for any laboratory upgrades, recall notifications or
samples (1/3–1/2 normal and the remainder abnor- technical updates.
mal) should be compared between the manual differ-
ential and the automated differential. Reliability. The length of time that an instrument was
unusable due to breakdown should be recorded as
well as the response time for the manufacturer’s
Efficiency
repair.
Throughput. The throughput of samples per hour,
including control material, should be recorded, taking Cost. The cost per test should be determined,
into consideration any tests that require more time including costs for controls and the staff time needed
than the standard CBC and leucocyte differential, for to maintain and operate the instrument and process
example, reticulocyte counts, and on some analysers, samples to final interpretation.
there is a reflex option to another channel, extended
counting or different method for abnormal samples Acceptability. Staff opinions and preferences should be
The time needed for start-up, shutdown and routine taken into consideration. An assessment of the level
maintenance of the instrument should also be of expertise required for the operation of the
determined. The number of samples that need instrument should be examined, as well as the impact
repeating, for any reason, should also be documented. of the instrument on the workflow and organization
of the laboratory. Any modifications to the laboratory
Sample identification. Sample identification through design should be considered.
the use of a barcode reader is now almost universal and Record keeping of the evaluation should be per-
offers significant advantages for laboratories, but the formed as previously described in these guidelines.
manual input of patient identification and tests should The manufacturer’s claims for within- and between-
be available for flexibility. If barcodes are used for batch precision should be confirmed. The suitability of
sample identification during the evaluation, barcode equipment and comparability with current methods
read errors by the instrument should also be assessed must be studied using samples including abnormal
and documented as they will also impact instrument samples and samples with interfering substances as
efficiency. Information should be sought regarding possible. The number of abnormal cell flags generated
information technology (IT) options, availability of and any failure of the analyser to provide a result
middleware (software communication from the should be documented and compared with existing
analyser to the laboratory information system) and methods. Precision, comparability of results, reference
expert rule-based systems for autovalidation of results ranges, reliability and acceptability of the instrument
or automatic ordering of blood films [48]. are also considerations in a local evaluation and
should be analysed as for a national evaluation.
IT, presentation and storage of results. The Once the equipment has been established within a
compatibility of the instrument with the laboratory few months, an audit should be undertaken to ensure
and hospital information systems, particularly that all manufacturers’ claims are being met and that
bidirectional interfacing, should be established. The parameters being reported are those needed by the
quality, format of data presentation and graphics, laboratory as well as any problems with middleware
including the display of quality control results, should or the laboratory information system.

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627
626 C. BRIGGS ET AL. | EVALUATION OF BLOOD CELL ANALYSERS

Special considerations for the evaluation of POC The performance of the device should be tested in
testing analysers. Point-of-care testing (POCT) can be the environment where the instrument will be sited
defined as any analytical test performed for a patient and, importantly, by the staff that will be operating
by a healthcare worker outside the laboratory the instrument, so as to ensure that ease of operation
environment [49, 50]. POCT is designed to move by nonlaboratory staff can be confirmed. It has been
laboratory testing closer to the patient providing a previously demonstrated that experiences of skilled vs.
more rapid service than can be achieved from the unskilled users can be different, and usually, the qual-
central laboratory. POCT devices can vary from hand- ity of testing by the POCT user is poorer than that by
held devices, designed to be used at the patient experienced laboratory staff [52].
bedside, intensive care units or in operating theatres,
to small bench top analysers, but all should generate
CONFLICT OF INTEREST
results with reference ranges that are comparable with
those of the main haematology analysers. Portable Carol Briggs: The department has received an unre-
haemoglobinometers, some of which now measure stricted grant from Sysmex Europe. Naomi Culp:
WBC and platelets, which use capillary blood and Employment by Trillium Diagnostics, LLC; consult to
blood gas analysers that test for haemoglobin alone, Beckman Coulter, Horiba Medical, and Abbott Diag-
should also be evaluated in the same way. nostics. Bruce Davis: Employment by Trillium Diag-
Ideally, where several POCT instruments are nostics, LLC; consult to Beckman Coulter, Horiba
required at different sites within a single institution, Medical, Abbott Diagnostics, and BD Biosciences.
only one instrument type should be selected so that Giuseppe d’Onofrio, Gina Zina: None. Samual J
results and reference ranges are the same wherever a Machin: The department has received an unrestricted
patient is tested. This also simplifies training, ordering grant from Sysmex Europe. Contributors. Sue Mead:
and storage of reagents, as well as servicing and main- Siemens; Nigel Llwelyn Smith: Abbott; Elena Sukha-
tenance contracts. cheva: Beckman Coulter; Jolanta Kunickaj: Sysmex;
Three extensive reviews of POCT validation have Yutaka Nagai: Nihon Khoden.
been previously published [49–51].

REFERENCES of in vitro diagnostic medical devices. EN 11. MHRA. The CE Mark, Bulletin No. 2. Lon-
13612. 2002. don: Medicines and Healthcare Products Reg-
1. Briggs C, Longair I, Slavik M, Thwaite K, Mills 6. State Food and Drug Administration. Regu- ulatory Agency; 2006. Available from: http://
R, Thavaraja V, Foster A, Romanin D, Machin lations for the supervision and administra- www.mhra.gov.uk. Accessed 6 June 2011.
SJ. Can automated blood film analysis tion of medical devices. People’s Republic 12. ICSH. Guidelines for evaluation of blood
replace the manual differential? An evaluation of China. Available from: http://eng.sfda. cell analysers including those used for dif-
of the CellaVision DM96 automated image gov.cn/cmsweb/webportal/W45649038/A48 ferential leucocyte and reticulocyte count-
analysis system. Int J Lab Hematol 2009;31: 335998.html. ing and cell marker applications. Clin Lab
48–60. 7. American National Standards Institute. Haematol 1994;16:157–74.
2. Roussel M, Davis BH, Fest T, Wood BL; on Clinical investigation of medical devices for 13. Gardiner C, Kitchen S, Dauer RJ, Kottke-
behalf of the International Council for human subjects-Part 1. General require- Marchant K, Adcock DM. Recommenda-
Standardization in Hematology (ICSH). ments. ANSI/AAMI/ISO 14155-1. 2003. tions for evaluation of coagulation analyz-
Toward a reference method for leukocyte 8. American National Standards Institute. ers. Lab Hematol 2006;12:32–8.
differential counts in blood: comparison of Clinical investigation of medical devices for 14. CLSI. Handling, Transport and Storage of
three flow cytometric candidate methods. human subjects-Part 2. Clinical investiga- Specimens, Quick Guide H18-A3. Wayne,
Cytometry A 2012;81A:973–82. tion plans. ANSI/AAMI/ISO 14155-2. 2003. PA: Clinical and Laboratory Standards Insti-
3. CLSI. Reference Leukocyte (WBC) Differ- 9. US Department of Health and Human Ser- tute; 2007.
ential Count (Proportional) and Evaluation vices. Food and Drug Administration, Office 15. Kennedy DA, Stevens JF, Horn AN. Clinical
of Instrumental Methods. H20-A2. Wayne, of the Commissioner. Computerized Sys- laboratory environmental contamination:
PA: Clinical and Laboratory Standards Insti- tems Used in Clinical Investigations. Wash- use of fluorescence/bacterial tracer. J Clin
tute; 2007. ington DC: FDA; 2007. Pathol 1998;41:1229–32.
4. ICH. International conference on harmoni- 10. USA Department of Health and Human 16. Advisory Committee on Dangerous Patho-
sation harmonised triparite guideline. Services. Medicine, Medicaid and CLIA gens (ACDP). Infection at Work: Control-
Guideline for good clinical practice (E6R1). programs; regulations and implementing ling the Risk. London: Department of
1996. the Clinical Laboratory Improvement Health; 2003.
5. European Committee for Standardisation. Amendments of 1998 (CLIA). Final rule. 17. Health Services Advisory Committee. Safe
European Standard. Performance evaluation Fed Regist 1992;57:7002–186. Working and the Prevention of Infection in

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627
C. BRIGGS ET AL. | EVALUATION OF BLOOD CELL ANALYSERS 627

Clinical Laboratories and Similar Facilities. reference” method for the packed cell vol- CD45 Gating for Persons Infected with
Sudbury, Suffolk: HSE books; 2003. ume. Lab Hematol 2003;9:1–9. Human Immunodeficiency Virus. National
18. CLSI. Protection of Laboratory Workers 30. CLSI EP05-A2. Evaluation of precision per- Center for Infectious Diseases; 2003. Avail-
from Occupationally Acquired Infections. formance of quantitative measurement able from: http://www.cdc.gov/mmwr/pre-
M29-A3. 3rd edn. Wayne, PA: Clinical and methods; approved guideline – second edi- view/mmwrhtml/rr5202a1.htm.
Laboratory Standards Institute; 2005. tion. 2004. 43. Tatsumi N, Pierre RV. Automated image
19. CLSI. Validation and Quality Assurance of 31. ICSH. Reference method for the enumera- processing. Past, present and future of
Automated Hematology Analyzers. tion of erythrocytes and leucocytes. Pre- blood cell morphology identification. Clin
Approved Standard, 2nd edn. H26-A2. pared by the Expert Panel on Cytometry. Lab Med 2002;22:299–315.
Wayne, PA: Clinical and Laboratory Stan- Clin Lab Haematol 1994;16:131–8. 44. Bracco D, Zahniser D, Swinehart W, Braga
dards Institute; 2010. 32. ICSH Expert Panel on Cytometry and ISLH M, Linder J. Assessment of carryover on
20. Briggs C. Quality counts: new parameters Task Force on Platelet Counting. Platelet the Bloodhound integrated hematology
in blood cell counting. Int J Lab Hematol counting by the RBC/platelet ratio method: analyzer. Poster abstract – Intl J Lab Hemal.
2009;3:277–97. a reference method. Am J Clin Pathol Special Issue: Abstracts of the XXVI Inter-
21. Broughton PM, Gowenlock AH, McCor- 2001;115:460–4. national Symposium on Technological
mack JJ, Neill DW. A revised scheme for 33. CLSI. Methods for Reticulocyte Counting Innovations in Laboratory Hematology
the evaluation of automatic instruments for (Automated Blood Cell Counters, Flow 2013;35(Suppl.1):1–139.
use in clinical chemistry. Ann Clin Bio- Cytometry, and Supravital Dyes). Approved 45. CLSI. Validation, Verification, Calibration,
chem 1974;11:207–18. Guideline, 2nd edn. H44-A. Vol. 17, no. and Quality Control of Automated Hema-
22. Ezzelle J, Rodriguez-Chavez IR, Darden 15: Clinical and Laboratory Institute; 2004. tology Analyzers. H26-P2. Wayne, PA:
JM, Stirewalt M, Kunwar N, Hitchcock R, 34. ICSH Expert Panel on Cytometry. Proposed Clinical and Laboratory Standards Institute;
Walter T, D’Souza MP. Guidelines on good reference method for reticulocyte counting 2009.
clinical laboratory practice: bridging opera- based on the determination of the reticulo- 46. Wood B, Jevremovic D, Bene MC, Yan M,
tions between research and clinical cyte to red cell ratio. Clin Lab Haematol Jacobs P, Litwin V; ICSH/ICCS Working
research laboratories. J Pharm Biomed Anal 1998;20:77–9. Group. Validation of cell-based fluorescence
2008;46:18–29. 35. Dacie and Lewis Practical Haematology. assays: practice guidelines from ICSH and
23. Ricos C, Alvarez V, Cava F, Garcia-Lario JV, Chapter 3, Basic haematological tech- ICCS-part V-assay performance criteria.
Hemandez A, Jimenez CV, Minchinela J, niques. In: Dacie and Lewis Practical Hae- Cytometry B Clin Cytom 2013;84:315–23.
Perich C, Simon M. Current databases on bio- matology, 11 edn. Bain BJ, Bates I, Laffan 47. Galen RS, Gambino SR. Beyond Normality.
logic variation: pros, cons and progress. Scand MA, Lewis M (eds). London: Churchill Liv- The Predictive Value and Efficiency of
J Clin Lab Invest 1999;59:491–500. Updated ingstone; 2012. Medical Diagnostics, 19. New York: John
in 2012. Available from: www.westgard.com/ 36. Schimenti KJ, Lacrna K, Maston L, Iaffald- Wiley & Sons; 1975.
intra-inter.htm. ano C, Straight M, Rabinovitch A, Lazarus 48. Barnes PW, McFadden SL, Machin SJ, Sim-
24. Harris EK. Statistical principles underlying HM, Jacobberger JW. Reticulocyte analysis: son E. The international consensus group
analytic goal-setting in clinical chemistry. comparison of flow cytometry, image for hematology review: suggested criteria
Am J Clin Pathol 1979;72(2 Suppl):374–82. analysis and manual counting. Cytometry for action following automated CBC and
25. Guder WG. Preanalytical factors and their 1992;13:853–62. WBC differential analysis. Lab Hematol
influence on analytical quality specifica- 37. Lee LG, Chen CH, Chiu LA. Tiazole orange: 2005;11:83–90.
tions. Scand J Clin Invest 1999;59:545–50. new dye for reticulocyte analysis. Cytome- 49. Briggs C, Carter J, Lee S-H, Sandhaus L,
26. CLSI. Defining, Establishing and Verifying try 1986;7:508–17. Simon-Lopez R, Vives Corrons J-L; for the
Reference Intervals in the Clinical Labora- 38. Serke S, Huhn D. Improved specificity of International Council for Standardization
tory. Approved Guideline, 3rd edn. C28- determination of immature erythrocytes in Haematology (ICSH). ICSH Guideline for
A3. Wayne, PA: Clinical and Laboratory (reticulocytes) by multiparameter flow worldwide Point-of-care testing in haema-
Standards Institute; 2008. cytometry thiazole orange using combined tology with special reference to the com-
27. Zwart A, van Assendelft OW, Bull BS, Eng- staining with monoclonal antibody (anti- plete blood count (CBC). Int J Lab
land JM, Lewis SM, Zijlstra WG. Recom- glycophorin-A). Clin Lab Haematol Hematol. 2008;30:105–16.
mendations for reference method for 1993;15:33–44. 50. British Committee for Standards in Haema-
haemoglobinometry in human blood (ICSH 39. Bland JM, Altman DG. Statistical methods tology (BCSH) General Haematology Task-
standard 1995) and specifications for inter- for assessing agreement between two meth- force, Briggs C, Guthrie D, Hyde K, Mackie
national haemoglobin cyanide standard ods of clinical measurement. Lancet I, Parker N, Popek M, Porter N, Stephens C.
(4th edition). J Clin Pathol 1996;49:271–4. 1986;1:307–10. Guidelines for point-of-care testing: haema-
28. CLSI. Reference and Selected Procedures for 40. CLSI. Method Comparison and Bias Estima- tology. Br J Haematol 2008;142: 904–15.
the Quantitative Determination of Hemoglo- tion Using Patient Samples. EP9-A2. 51. Briggs C, Kimber S, Green L. Where are we
bin in Blood. H15-A3. Wayne, PA: Clinical Wayne, PA: Clinical and Laboratory Stan- at with point-of care testing in haematolo-
and Laboratory Standards Institute; 2000. dards Institute; 2002. gy? Br J Haematol 2012;158:679–90.
29. Bull BS, Fujimoto K, Houwen B, Klee G, 41. CLSI. Estimation of Total Analytical Error 52. Skeie J, Thue G, Nerhus K, Sandberg S.
van Hove L, van Assendelft OW, Bun- for Clinical Laboratory Methods. EP-A. Instruments for self-monitoring of blood
yaratvej A, Buttarello M, Davis B, Koepke Wayne, PA: Clinical and Laboratory Stan- glucose: comparisons of testing quality
JA, Lewis SM, Machin SJ, d’Onofrio G, dards Institute; 2003. achieved by patients and a technician. Clin
Rowan RM, Tatsumi N. International 42. Mandy FF, Nicholson JK, McDougal JS. Chem 2002;48:994–1003.
Council for Standardization in Haematology Guidelines for Performing Single-Platform
(ICSH) recommendations for “surrogate Absolute CD4+ T-Cell Determinations with

© 2014 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2014, 36, 613–627

You might also like