You are on page 1of 10

Anal. Chem.

1997, 69, 587-596

Solid-Phase Microextraction for the Analysis of


Human Breath
Christoph Grote† and Janusz Pawliszyn*

The Guelph-Waterloo Center for Graduate Work in Chemistry, Department of Chemistry, University of Waterloo,
Waterloo, Ontario, Canada N2L 3G1

Solid-phase microextraction (SPME) has been applied to polyunsaturated fatty acids.3 Several organic compounds can be
the quantitative determination of ethanol, acetone, and associated, for example, with lung cancer,4 liver disease,5,6 and
isoprene in human breath. The method involves extrac- myocardial infarction.7 Increased concentrations (>50 nmol/L)
tion and preconcentration with a fused silica fiber coated of acetone have been detected in the breath of patients suffering
with a polymeric stationary phase, desorption at 200 °C, from diabetes, a chronic disorder in which either the pancreas is
and assay by gas chromatography/mass spectrometry. unable to produce insulin or the body is unable to effectively use
Three different fiber coatings have been evaluated with insulin.8-10 Moreover, breath acetone can be used to monitor
regard to sensitivity, linear range, precision, and detection patients on diets. A weight reduction of about 225 g/week is
limits. Typical RSD values in the range 2%-6% could be indicated by an acetone concentration of 500 nmol/L. Eating
obtained, depending on the fiber coating and the com- meals with a high carbohydrate content significantly decreases
pound investigated. The calibration curves for the com- this concentration.10 Breath acetone analysis can, therefore, be
pounds are reproducible and linear over the concentration used as a motivational tool during weight-loss programs. Inhala-
ranges found in human breath samples. The method is tion toxicology is another area of interest. Exposure of workers
capable of detecting concentrations of acetone and iso- to environmental pollutants can be monitored by breath analysis.11
prene reported for healthy subjects. The influence of Routine breath analysis is applied for the determination of ethanol
temperature and humidity on the extraction process has
because its concentration in breath is directly related to the
been studied in detail. A linear relationship between log
concentration in blood. A breath-to-blood ratio of 1:2100 is used
K versus 1/T allows the calibration of the method for any
when breath alcohol measurement devices are calibrated to reflect
given temperature. The device is portable, economical,
blood alcohol concentrations.12 Values displayed by breathalyzers
and easy to use in patient sampling.
show the blood concentration of alcohol that would be assumed
by the breath alcohol concentration.
In recent years there has been increased interest in the Breath compounds are present in nanomolar or even lower
determination of breath compounds in medicine and clinical
quantities.1 To improve the sensitivity and precision of the
toxicology. More than 100 compounds have been identified in
determination of these compounds, the sample has to be concen-
normal human breath by gas chromatography and mass spec-
trated before assay with gas chromatography/mass spectrometry
trometry.1 These voliatile organic compounds (VOCs) are pro-
(GC/MS). The three main methods currently utilized for pre-
duced by metabolic processes and partition from the blood stream
concentration are chemical interaction, adsorptive binding, and
via the alveolar pulmonary membrane into the alveolar air. This
cold trapping. For example, cold trapping utilizes solid adsorbents
implies that the concentration measured in breath is related to
cooled to temperatures of liquid nitrogen or solid carbon
the concentration in blood. Breath analysis can be used as a
dioxide.13-15 Trapping systems operated at ambient temperatures
diagnostic tool because increased or decreased concentrations of
some compounds have been associated with various diseases or
(3) Van Gossum, A.; Decuyper, J. Eur. Respir. J. 1989, 2, 287-291.
altered metabolism.2 The major advantage of breath analysis is
(4) Gordon, S. M.; Szidon, J. P.; Krotoszynski, B. K.; Gibbons, R. D.; O’Neill,
that it is a noninvasive procedure. It is well accepted by patients H. J. Clin. Chem. 1985, 31 (8), 1278-1282.
because it does not incur the inconvenience of the collection of (5) Kaji, H.; Hisamura, M.; Saito, N.; Murao, M. Clin. Chim. Acta 1978, 85,
279-284.
blood or urine samples. Although many diseases have not yet (6) Chen, S.; Mahadevan, V.; Zieve, L. J. Lab. Clin. Med. 1970, 75, 622-627.
been examined in the context of using breath analysis as a (7) Weitz, Z. W.; Birnbaum, A. J.; Sobotka, P. A.; Zarling, E. J.; Skosey, J. L.
diagnostic tool, the usefulness has already been demonstrated by Lancet 1991, 337, 933-935.
(8) Levey, S.; Balchum, O. J.; Medrano, V.; Jung, R. J. Lab. Clin. Med. 1964,
the investigation of various clinical conditions. Volatile hydro- 63, 574-584.
carbons, especially ethane and pentane, have been used as (9) Rooth, G.; Ostenson, S. The Lancet 1966, ii, 1102-1105.
(10) Crofford, O. B.; Mallard, R. E.; Winton, R. E. Trans. Am. Clin. Climatol.
markers of lipid peroxidation, the oxidative degradation of
Assoc. 1977, 88, 128-139.
(11) Raymer, J. H.; Pellizzari, E. D.; Thomas, K. W.; Cooper, S. D. J. Exposure
† Present address: Department of Analytical Chemistry, Fraunhofer Institute
Anal. Environ. Epidemiol. 1991a, 1 (4), 439-451.
of Toxicology and Aerosol Research, Nikolai-Fuchs-Strasse 1, D-30625 Hannover, (12) Van Berkom, L. C. Alcohol Drugs Driving 1991, 7, 229-234.
Germany. (13) Van Gossum, A.; Shariff, R.; Lemoyne, M.; Kurian, R.; Jeejeebhoy, K. N.
(1) Krotoszynski, B. K.; Gabriel, G.; O’Neill, H. J. Chromatogr. Sci. 1977, 15, Am. J. Clin. Nutr. 1988, 48, 1394-1399.
239-244. (14) Roberts, R. J.; Rendak, J.; Butcher, J. Dev. Pharmacol. Ther. 1983, 6, 170-
(2) Manolis, A. Clin Chem. 1983, 29 (1), 5-15. 178.

S0003-2700(96)00749-4 CCC: $14.00 © 1997 American Chemical Society Analytical Chemistry, Vol. 69, No. 4, February 15, 1997 587
Figure 1. SPME device modified for breath analysis.

have also been used.16 Several published reviews discuss the contaminants in a hot GC injector. The SPME technique coupled
currently used sampling techniques.2,17 These methods require with GC has been used to analyze volatile and semivolatile air
complex devices, and each method has its particular problems. compounds.21 The fiber can be exposed directly to the air or to
The methods based on adsorptive binding or cold trapping are a sample collected in a gas sampling bulb. This paper reports on
susceptible to contamination and require an additional water trap the first application of solid-phase microextraction for the analysis
to remove the large amount of water vapor present in human of human breath. The validation of the method was based on
breath. Otherwise this water would saturate the traps, and after ethanol, acetone, and the nonpolar compound isoprene, which was
thermal desorption it could damage the GC column. Gas sampling found to be the main endogenous hydrocarbon of human breath.1
bags, often used to transport a definite volume of breath to a place
where the concentration and analysis are carried out, can be the EXPERIMENTAL SECTION
source of contamination and may lead to selective adsorption at Chemicals and Materials. Chemicals. Acetone (99.9%+,
the inner wall. Current sampling devices used to preconcentrate HPLC grade) was purchased from Sigma-Aldrich (Mississauga,
breath compounds at the time of sample collection normally ON, Canada) and isoprene (99%, inhibited with 100 ppm p-tert-
consist of a complex arrangement of tubing and valves which often butylcatechol) from Aldrich (Milwaukee, WI). Undernatured
lead to a buildup of contaminants and loss of analytes by leakage. ethanol (99.9%+, H2O < 0.1%) was obtained from a local pharmacy.
The difficulty in quantitative delivery of the sample to a gas Carbon disulfide was purchased from BDH Inc. (Toronto, ON,
chromatograph is another problem of the collection methods used Canada). Dimethyldichlorosilane (Supelco, Canada) was used to
currently. deactivate the glassware. For the Varian GC/MS system, helium
A technique is needed that does not concentrate H2O or CO2 was purchased from Praxair (Waterloo, ON, Canada).
and that can be applied directly to the mouth to preconcentrate Materials. Gas sampling bulbs (0.5 and 1.0 L) were used for
the compounds so that possible contamination due to tubing standard gas preparation (Supelco). Prior to the first use, all bulbs
material is avoided. were silanized to deactivate the interior glass surface using a 10%
Solid-phase microextraction (SPME), developed in recent years (v/v) solution of dimethyldichlorosilane in toluene. The solution
by Pawliszyn et al.,18-20 is a rapid, inexpensive, and efficient was kept in the bulbs for 12 h. After the bulbs were rinsed with
technique for sampling solid, liquid, or gaseous samples. It toluene and methanol several times, they were dried at 100 °C.
requires no solvents or complicated apparatus and provides linear After use, the gas bulbs were cleaned with deionized water,
results over a wide range of analyte concentrations. Analysis of followed by rinsing with methanol and drying at 100 °C. The bulbs
the extracts is performed using GC, GC/MS, or HPLC.18-20 In were then purged with purified nitrogen for 30 min to remove
the SPME technique, a fused silica fiber coated with a polymeric any contaminants.
stationary phase is contained in a specially designed syringe whose The SPME fibers and the holder were purchased from Supelco.
needle protects the fiber when septa are pierced. The fiber is Prior to their first use, new fibers have been conditioned under a
directly exposed to a liquid or gaseous sample to extract and helium stream to clean the fiber from any contaminants present
concentrate the analytes. After the absorption equilibration is in the coating. Conditioning was performed in a normal GC
attained, the fiber is withdrawn into the needle and introduced injection port following the instructions of the manufacturer.
into an injector of a gas chromatograph, where the extracted SPME Device Modified for Breath Analysis. The SPME
compounds are thermally desorbed and analyzed. SPME can be fiber can be directly exposed into the mouth of a subject. To
performed manually or by an autosampler. The method is avoid the subject’s tongue coming into contact with the fiber
economical, because one single fiber can be used repeatedly. during sampling, the fiber has to be protected. This is achieved
Unlike in conventional methods for analysis of gaseous samples, simply by using a small piece of inert tubing which fits onto the
modified equipment like a complex valve injection system, a depth guide of the SPME device. The tubing also serves as a
thermal desorption device, or a cooling trap is not required by mouthpiece, which can be easily replaced for hygienic reasons.
SPME. The SPME fiber is easily cleaned by desorbing any One little aperture in the tubing allows the subject to replace the
(15) Snider, M. T.; Balke, P. O.; Oerter, K. E.; Francalancia, N. A.; Pasko, K. A.;
dead volume in the tubing and to exhale during the sampling
Robbins, M. E.; Gerhard, G. S.; Richard, R. B. Life Chem. Rep. 1985, 3, procedure. Figure 1 shows the modified SPME device. The
168-173. breath sampling procedure is described below.
(16) Lawrence, G. D.; Cohen, G. Anal. Biochem. 1982, 122, 283-290.
Preparation of Standards. Preparation of Standards for the
(17) Schaeffer, H. J. J. High Resolut. Chromatogr. 1989, 12, 69-81.
(18) Chen, J.; Pawliszyn, J. Anal. Chem. 1995, 76, 2530-2533. Instrument Calibration. The standards used to calibrate the mass
(19) Zhang, Z.; Yang, M. J.; Pawliszyn, J. Anal. Chem. 1994, 66, 844A-853A.
(20) Boyd-Boland, A. A.; Pawliszyn, J. Anal. Chem. 1996, 68, 1521-1529.
(21) Chai, M.; Pawliszyn, J. Environ. Sci. Technol. 1995, 29, 693-701.

588 Analytical Chemistry, Vol. 69, No. 4, February 15, 1997


spectrometer were prepared in carbon disulfide. Three stock avoided because the analytes would partition into the aqueous
solutions containing 100 µL of acetone, 200 µL of ethanol, and 20 phase.
µL of isoprene, respectively, were prepared in 50.0 mL volumetric Selection of SPME Coating. Four different types of coatings,
flasks. Eight standards were then prepared by taking appropriate a 100 µm polydimethylsiloxane (PDMS) fiber, an 85 µm polyacry-
aliquots of the stock solutions and diluting them with carbon late fiber, a 65 µm polydimethylsiloxane/divinylbenzene fiber, and
disulfide in 25.0 mL volumetric flasks. The solutions were mixed a 65 µm Carbowax/divinylbenzene fiber, were assessed by
in an ultrasonic bath for 15 min and refrigerated at 4 °C. For comparing the extraction time profile and sensitivity of SPME
analysis, 0.5 µL aliquots of the solutions were injected into the analysis with these fibers. The PDMS fibers with thinner films
instrument. Manual injections were performed using a 10 µL were not investigated because, although they equilibrate faster,
syringe (Hamilton, Reno, NV). they have a lower capacity which is not useful for analyzing highly
Preparation of Gaseous Standards. The stock solutions of volatile trace level compounds.
acetone and ethanol were prepared in 50.0 mL volumetric flasks Analytical Instrument Parameters. Instrument Calibration.
using deionized water as a solvent (NANOpure, ultrapure water A Varian 3400 GC coupled with a Varian Saturn ion trap mass
system, Barnstead/Thermolyne, Dubuque, IA). The solutions spectrometer (Varian Chromatography Systems, Walnut Creek,
were mixed in an ultrasonic bath for 15 min. The temperature of CA) was used to carry out the experimental work. Data acquisi-
human breath is about 35 °C. Moreover, it is almost saturated tion was achieved using the Varian Saturn software. A 30 m SPB-5
with water vapor. It has been shown that both temperature and column with 1.0 µm film thickness was connected to a 30 m SPB-5
humidity affect the extraction of analytes on the fiber.21 These column with 0.25 µm film thickness using a glass seal connector
results require the setup of a calibration curve with gas standards (Supelco). Removing this thick film column was easily possible
kept at 35 °C and spiked with the appropriate amount of water to without shutting down the mass spectrometer.
give a saturated atmosphere. Carbon disulfide was chosen as a solvent because it elutes after
the target compounds. The best chromatographic results were
A water bath big enough to accommodate four gas sampling
obtained by using a precolumn and starting the column temper-
bulbs at the same time was used, which made the procedure quite
ature at 45 °C, slightly below the boiling point of the solvent.
practical. A thermostat with a circulator was installed to keep
During injection, the septum-equipped programmable injector
the temperature at 35 °C. A second thermostat was used to heat
(SPI) was kept at the same low temperature of 45 °C and then
water flushing through a copper coil at the bottom of the water
ramped to 200 °C with a rate of 200 °C/min. The acquisition was
bath to ensure a more accurate temperature control. The
stopped before the large amount of solvent entered the ion trap
temperature of this additional thermostat was controlled by a
to protect the electron multiplier. The column temperature was
contact thermometer. Manual adjustment was not necessary, and
then ramped to 120 °C with a rate of 30 °C/min to remove all the
the temperature could be kept constant in a small range of
solvent. Since the peaks were narrow, the highest possible scan
approximately (0.2 °C. The temperature was measured using a
rate was applied (0.14 s/scan).
commercial digital oral thermometer. Its accuracy is reported to
Different characteristic masses for the three analytes have been
be (0.1 °C in the range between 33 and 43 °C.
checked for the quantification. At higher concentrations of
The amount of water which is necessary to give a saturated
ethanol, the mass spectrum changes due to secondary reactions
atmosphere was calculated by the ideal gas law using the vapor
occurring in the ion trap. A steady decrease in detector response
pressure of water at 35 °C. The 1.0 L bulbs had to be spiked
factor with increasing concentration was observed when only m/z
with 39.7 µL of water to give a saturated atmosphere. This water
) 45 was used for the quantification of ethanol. In contrast, the
evaporates to about 55 mL of vapor, which causes a significant response increased when m/z ) 47 was used. Due to reactions
pressure increase in the gas bulb. Therefore, the bulb was first in the trap, the fragment M - 1 (m/z ) 45) turned into M + 1
spiked with 34 µL of pure water. After the water had been (m/z ) 47). Therefore, the sum of m/z ) 45 and m/z ) 47 was
evaporated at 35 °C, the pressure was decreased to atmospheric used to quantify ethanol. As for acetone, both m/z ) 43 (base
pressure by quickly turning the stopcock of the bulb. Then 5 µL peak) and m/z ) 58 (molecular ion) can be used for quantification.
of water containing the appropriate amount of ethanol and acetone Although the fragment m/z ) 58 is more significant for acetone
was spiked into the gas sampling bulb through a half-hole septum than the common fragment m/z ) 43, the latter one was used
using a microsyringe (Hamilton). The total amount of water during the study, because of the much higher intensity of the
injected gave a relative humidity of approximately 100%. Standards peak. The quantitation of isoprene gave good results when m/z
with higher concentrations of ethanol were prepared by spiking ) 53, m/z ) 67, or the sum of both characteristic masses was
neat ethanol into the bulb using a 1.0 µL syringe with no dead taken.
volume (SGE, Australia). Since isoprene is not miscible with Instrument Parameters for the Analysis of Gaseous Standards
water, the procedure described above could not be used for this and Breath Samples. Several column temperatures have been
analyte. Therefore, isoprene was spiked into the bulb using a investigated. Since the time required for the thermal desorption
gas-tight syringe (Hamilton). The appropriate aliquots were taken of the volatile analytes in the hot injector port is relatively short
from gaseous stock solutions obtained by spiking a 500 mL gas (<15 s), an initial temperature of 15 °C was sufficient to focus
sampling bulb with 10 and 20 µL of pure isoprene, respectively. the analytes at the beginning of the column. This cryofocusing
The gas sampling bulbs were kept at 35 °C for at least 30 min was achieved using CO2 as the cooling agent. All tests were
prior to analysis to guarantee a homogeneous mixture. The carried out with standard gas mixtures. Examples include
isoprene stock solutions were kept in the dark at room temper- investigation of different fiber coatings, measurement of the
ature. Using this procedure, no visible condensation occurred extraction and desorption time profiles, and determination of the
on the inner wall of the sampling bulb. Condensation must be reliability of the method. Unless otherwise stated, all the breath

Analytical Chemistry, Vol. 69, No. 4, February 15, 1997 589


Table 1. Instrument Parameters

calibration breath samples


GC
injector SPI, inlet sleeve, 54 mm × 4.6 mm, 0.75 mm i.d. SPI, inlet sleeve, 54 mm × 4.6 mm, 0.75 mm i.d.
column (1) precolumn, 0.8 m deactivated fused silica (1) precolumn, 0.8 m deactivated fused silica
(2) 30 m SPB-5, 0.25 mm i.d., 1.0 µm film (2) 30 m SPB-5, 0.25 mm i.d., 1.0 µm film
(3) 30 m SPB-5, 0.25 mm i.d., 0.25 µm film (3) 30 m SPB-5, 0.25 mm i.d., 0.25 µm film
carrier gas helium helium
head pressure 24 psi 24 psi
linear velocity 34 cm/s 36 cm/s
column temp 45 °C, 5 min, 10 °C/min, 75 °C, 30 °C/min, 120 °C 15 °C, 0.25 min, 5 °C/min, 45 °C, 20 °C/min, 100 °C
injector temp 45 °C, 200 °C/min, 200 °C 200 °C
MS
detector ion trap ion trap
ionization mode EI EI
transfer line temp 220 °C 220 °C
manifold temp 220 °C 220 °C
scan rate 0.14 s/scan 0.14 s/scan
mass range 35-150 amu 35-150 amu
auto ion control on on

samples were analyzed using the method described in Table 1. have to be reached as long as the same extraction time is applied
Using this method, the overall run time was about 10 min. Total for both the breath and the standards, a very short sampling time
ion current was monitored for all samples; however, quantitation is possible. This is especially true for the Carbowax/DVB fiber,
was performed by selecting characteristic masses. This provided for which a sampling time of only 10 s is possible. Although the
better sensitivity and a linear calibration curve for standard equilibrium for ethanol is not reached after 10 s, the decrease in
injections. Table 1 summarizes the optimized instrumental sensitivity can be tolerated, considering that ethanol reaches much
parameters for both calibration and analysis of breath samples. higher concentrations in human breath than acetone and isoprene.
Unless otherwise stated, the analytes were thermally desorbed
RESULTS AND DISCUSSION from the fiber for 15 s at 200 °C throughout the study. The
Extraction Time Profiles. The first step in the development determination of the appropriate desorption time is important
of an SPME method is the determination of the time needed for because all of the analytes should be desorbed from the fiber to
the analyte to reach equilibrium between the sample and the fiber. reach the highest possible sensitivity and to avoid carryover. Less
The extraction time profiles of the four differently coated fibers than 1% carryover was found even when standards with high
were established by plotting the detector response versus the concentrations of ethanol (3.5 µmol/L) were extracted. Carryover
extraction time. The equilibration time is reached when a further was determined by analyzing the fiber blank directly after the
increase of the extraction time does not result in a significant initial injection and expressed as percentage of the initial peak
increase in the detector response. Figure 2 shows the extraction area. Fibers thermally desorbed for 30 s at 250 °C did not show
time profiles for 100 µm PDMS, 85 µm PA, 65 µm Carbowax/ any detectable carryover. Therefore, each fiber was predesorbed
DVB, and 65 µm PDMS/DVB fibers. All times were investigated in an extra injector port for 30 s at 250 °C prior to each extraction.
by analyzing duplicates. This procedure also ensured that no analytes were extracted from
The 100 µm PDMS fiber reached equilibrium after about 30 s the ambient air due to passive sampling.
for all three compounds investigated in this study. Acetone and Selection of SPME Coating. The choice of an appropriate
isoprene equilibrated even faster when the Carbowax/DVB fiber coating is essential for the SPME method. Depending on the
was used. Only 10 s was necessary for these compounds, whereas molecular weight and the polarity of the analytes to be extracted,
ethanol required about 1 min to reach equilibrium. The new the sensitivity of each fiber is different. Four types of coatings
PDMS/DVB fiber also showed fast equilibration times. Isoprene were investigated. A standard sample of ethanol, acetone, and
equilibrated in about 15 s, acetone in 30 s, and ethanol in 1 min. isoprene with a relative humidity of 99% was analyzed four times
The more polar the compound, the longer it took to reach
with each fiber. The extraction time was 1 min for the PDMS,
equilibrium. As can be seen in Figure 2b), the equilibration times
Carbowax/DVB, and PDMS/DVB fibers, and 4 min for the PA
for the PA fiber for all three compounds were much longer. This
fiber. The desorption time was always 15 s except for the PA
fiber probably requires a longer equilibration time because it is a
fiber, which was desorbed for 20 s. The temperature of the water
more crystalline polymer than the PDMS coating. Therefore,
bath was kept at 35 ( 0.1 °C. Figure 3a shows the absolute
diffusion into the polymer does take longer for PA than for
amount of each analyte extracted by each type of fiber obtained
PDMS.23
by comparing the detector response with the calibration curves
Since the breath sampling device will be applied directly into
generated with liquid standards. For a better comparison, the
the subject’s mouth, a short equilibration time is required.
amount of each analyte extracted by the PDMS/DVB fiber was
Therefore, the appropriate fibers are the PDMS, the Carbowax/
set at 100%. The results are illustrated in Figure 3b. As can be
DVB, and the PDMS/DVB fibers. Since equilibrium does not
seen from Figure 3b, the PDMS/DVB fiber, recommended for
(22) Supelco chromatography products catalog, 1996, Supelco Inc., Canada. polar volatiles, exhibited a higher sensitivity for the analytes of
(23) Louch, D.; Motlagh, S.; Pawliszyn, J. Anal. Chem. 1992, 64, 1187-1199. interest in comparison to the other fibers. This was especially

590 Analytical Chemistry, Vol. 69, No. 4, February 15, 1997


to Carbowax/DVB, 5 times higher compared to PDMS, and almost
15 times higher compared to PA. In addition to the poor
sensitivity for isoprene, the PA fiber also required a much longer
extraction time to reach equilibrium. Therefore, the PA fiber was
not further investigated in this study. If only the sensitivity is
considered, the PDMS/DVB fiber would be the fiber of choice,
but characteristics like precision and linear range are also
important and will be discussed later in this study.
Linear Range. The linearity of the method was investigated
by determining calibration curves over the concentration ranges
of interest. The concentration of ethanol was about 3 orders of
magnitude higher than the concentrations of acetone and isoprene.
Therefore, ethanol was spiked into the gas sampling bulbs as a
neat liquid using a 1.0 µL syringe (SGE). Acetone and isoprene
were added as described earlier. Duplicate samples of gas
standards were extracted with three different fiber types and
analyzed using GC/MS. The line of best fit for the relationship
between the average peak area and the concentration of analyte
in the sample was determined by linear regression. The line was
obtained by integrating the selected m/z chromatograms. The
results are presented in Table 2, together with the results obtained
for precision and detection limits.
Precision. Eight replicate measurements were performed by
extracting standards with the same concentration. Each standard
was prepared fresh, and four different bulbs were used. The
extraction was performed 30 min after the analytes had been
spiked into the bulb. The temperature was kept at 35 °C, and
the relative humidity was set at 98% prior to spiking of target
analytes. The precision was evaluated by calculating the mean,
standard deviation, and relative standard deviation (%RSD) of the
observed values. Since only one extraction was performed out
of each gas sampling bulb, the calculated precision include not
only the precision of the analytical method itself (SPME fiber plus
GC/MS system) but also the repeatability of the preparation of
the gaseous standards. The results are shown in Table 2. Relative
standard deviations of retention times for the target compounds
were lower than 0.2% RSD (n ) 8), indicating good repeatability,
which is crucial for a reliable chromatographic method.
Accuracy. Quantification of Acetone and Isoprene. There are
no standard reference materials available for human breath.
Reference methods were also not applicable for acetone and
isoprene because they require additional custom-made equipment.
Figure 2. Extraction time profiles. Concentrations of analytes in Therefore, the accuracy of the method for acetone and isoprene
standard: ethanol ([), 136 nmol/L; acetone (9), 109 nmol/L; and was investigated by creating a calibration curve in the range of
isoprene (2), 100 nmol/L. Relative humidity, 98%; extraction tem- interest for both acetone and isoprene. The analysis was
perature, 35 °C; desorption temperature, 200 °C; desorption time,
performed using the 65 µm PDMS/DVB fiber, because of its
15 s. The lines only connect the data points to show the general
trend. They do not convey mathematical relation. highest sensitivity toward isoprene compared to the other fibers
investigated.
For acetone, a five-point calibration curve was set up in the
true for the less polar compounds, acetone and isoprene, inves- range 6-56 nmol/L (y ) 44447x + 21546, r2 ) 0.994). A new
tigated in this study. Ethanol can best be extracted with fiber standard was then prepared following the same procedure and
coatings recommended for polar analytes. Compared to the containing 20.44 nmol/L acetone. Triplicate analysis was per-
nonpolar PDMS coating, the sensitivity could be increased by a formed. Using the above calibration curve, a concentration of 21
factor of 3 when the PA fiber was used, by a factor of ap- ( 1.2 nmol/L at the 95% confidence level was found (2.3% RSD).
proximately 4 when the Carbowax/DVB fiber was used, and by a For isoprene, a five-point calibration curve was set up in the range
factor of 5 when the PDMS/DVB fiber was used. Whereas 1-12 nmol/L (y ) 3608x - 960). A standard containing 3.00
PDMS, Carbowax/DVB, and PA basically have the same sensitiv- nmol/L isoprene was then analyzed three times and quantified
ity for acetone, the amount extracted was 3 times higher when with the calibration curve to be 2.9 ( 0.15 nmol/L at the 95%
the PDMS/DVB fiber was used. For isoprene, the new PDMS/ confidence level (2.1% RSD). The studies reported to date were
DVB fiber exhibited about 3 times higher sensitivity compared performed using always the same fiber. If several fibers have to

Analytical Chemistry, Vol. 69, No. 4, February 15, 1997 591


Figure 3. (a) Amounts extracted by different SPME fiber coatings. (b) Sensitivity relative to PDMS/DVB. Concentrations of analytes in the
standard: ethanol, 136 nmol/L; acetone, 109 nmol/L; and isoprene, 100 nmol/L. Relative humidity, 99%; extraction temperature, 35 °C; extraction
time, 4 min for PA, 1 min for PDMS, Carbowax/DVB, and PDMS/DVB; desorption temperature, 200 °C Desorption time, 15 s for PDMS, Carbowax/
DVB, and PDMS/DVB, 20 s for PA.

Table 2. Linear Range, Precision, and Limits of


Detectiona the fiber at equilibrium, Vf is the volume of the fiber coating, and
c, is the concentration remaining in the sample at equilibrium.23
SPME fiber
The problem in determining the K value is that cs is the
100 µm 65 µm 65 µm
compound PDMS Carbowax/DVB PDMS/DVB concentration at equilibrium (after the extraction took place) and
not the initial concentration of the sample, cs°. In certain cases,
Linear Range (r2 > 0.99)
ethanol 0-34 µmol/L 0-25 µmol/L 0-8 µmol/L cs in eq 1 can be replaced by the initial concentration cs°. This
acetone 0-545 nmol/L 0-136 nmol/L 0-545 nmol/L simplification is possible when the sample volume Vs is at least 2
isoprene 0-100 nmol/L 0-100 nmol/L 0-100 nmol/L orders of magnitude larger than the product KVf.24 This assump-
Precision, %RSD (n ) 8) tion is true when 1.0 L gas sampling bulbs are used, considering
ethanol 5.7 8.3 1.7
acetone 2.2 4.2 4.8
a constant fiber volume of the PDMS fiber of Vf ) 6.91 × 10-4
isoprene 3.9 12.8 3.2 cm3 and relatively small K values (less than 1000) for the highly
Limits of Detection (nmol/L) volatile compounds of interest. The determination of K values is
ethanol 20 4.6 5.8 then possible by using the equation
acetone 5.0 5.2 1.8
isoprene 2.1 1.1 0.3 K ) nf/cs°Vf (2)
aConditions: relative humidity, 98%; extraction temperature, 35 °C;
extraction time, 1 min; desorption time, 15 s; desorption temperature,
200 °C. This equation can only be applied for the 100 µm PDMS fiber
because this coating can be treated as a “normal” liquid. We
believe that, due to presence of solid polymer particules (DVB)
be used, the fiber length has to be taken into account as well, in the cross-linked structure of the Carbowax/DVB and PDMS/
because a variability in the length of (1 mm would affect the DVB, these fibers extract the analytes primarily by adsorption
results by (10%. rather than by absorption. Therefore, the equilibrium theory on
Effect of Temperature on the Extraction. The temperature effect which eq 2 is based cannot be applied. For these fibers, a different
on the extraction process was studied by analyzing a standard K value can be defined, taking into account the surface area of
with 99% relative humidity. Since the effect of temperature on the fibers. Since this surface area is not known, the K values
the extraction of volatile compounds has already been investi- have been determined only for the 100 µm PDMS fiber.
gated,21,24 only three temperatures (26, 35, and 45 °C) have been To determine the K values, a standard containing 136 nmol/L
studied. At each temperature, four measurements were performed ethanol, 109 nmol/L acetone, and 100 nmol/L isoprene was
with each type of fiber. Since the instrument was calibrated prior prepared. Three different temperatures were investigated. The
to the analysis, this study was also used to determine the partition relative humidity was set at 99% in all cases. Four replicate
coefficients (K values) at different temperatures. extractions were performed at each temperature. The average
The partition coefficient K between the fiber coating and the detector responses were converted into the absolute amount of
gaseous sample matrix can be expressed as analytes extracted by the fiber, nf, using a calibration curve
generated prior to the study. The K values for ethanol, acetone,
K ) nf/csVf (1) and isoprene were 88, 101, and 59 at 26 °C, 53, 72, and 45 at 35
°C, and 35, 43, and 40 at 45 °C, respectively, for the 100 µm PDMS
fiber. The relatively low K values are due to the low boiling points
where nf is the amount of analyte extracted from a gas sample by
and high vapor pressures of the analytes. Several extractions can
(24) Martos, P. A.; Pawliszyn, J. Anal. Chem., in press. be performed using one bulb without a significant depletion in

592 Analytical Chemistry, Vol. 69, No. 4, February 15, 1997


Table 3. Effect of Temperature on the Extraction
Process of the 100 µm PDMS Fibera

log K ) a(1/T) + b
compound a b r2
ethanol 2014 -4.8 0.9935
acetone 1866 -4.2 0.9887
isoprene 852 -1.1 0.9425
a Concentrations of analytes in the standard: ethanol, 136 nmol/L;
acetone, 109 nmol/L; and isoprene, 100 nmol/L. Relative humidity,
99%; extraction temperatures investigated, 26, 35, and 44 °C; extraction
time, 1 min; desorption time, 15 s.

the concentration because of the small amount of analyte extracted


during one sampling process. As can be seen, the K values
strongly depend on the temperature. Once the relationship
between K and T is determined, the plot of log K vs 1/T gives a
straight line.24 Using the equation of that line allows the
determination of K at any given temperature. It has to be
mentioned that the K values are calculated for a relative humidity
of 99%. Table 3 shows the data obtained when log K is plotted
versus 1/T for the 100 µm PDMS fiber. As predicted by the
theory, a straight line was obtained.
Since Carbowax/DVB and PDMS/DVB coatings cannot be
treated as “normal” liquids, a partition coefficient has to be defined
which takes into account the adsorption process. It is assumed
for the linear portion of the adsorption isotherm that the amount
extracted by the fiber is proportional to the concentration in the
sample. Therefore, a similar relationship between K or the mass
of analyte extracted and the temperature should be obtained as
well. Since the volume and the surface of these fibers are not
known, a new K value was not defined. The log of the mass versus
1/T was plotted instead, giving straight lines similar to those
shown in Table 3.
Changing the temperature had a significant effect on the
amount of analyte extracted by the fiber. The positive slope in
the figures illustrates that the amount of analyte extracted by the
fiber increases with decreasing temperature, indicating that the
extraction is an exothermic process. The study demonstrates that
the temperature had to be carefully monitored. However, the
temperature of the standards does not necessarily have to be
monitored at the same temperature as human breath, as long as Figure 4. Effect of relative humidity on the extraction. Relative
humidities investigated: 12.6%, 90.7%, 100%. Extraction time, 1 min;
the saturated humidity is carefully adjusted. The equations of
extraction temperature, 35 °C; desorption time, 15 s; desorption
the straight lines obtained as described above allow the calculation temperature, 200 °C; ethanol, 136 nmol/L, quantitation ions m/z )
of K at any given temperature. Knowing K and the amount of 45 + m/z ) 47; acetone, 109 nmol/L, quantitation ion m/z ) 43;
analyte extracted by the fiber, the concentration of the sample isoprene, 100 nmol/L, quantitation ions m/z ) 53 + m/z ) 67.
can be calculated.
Effect of Relative Humidity on the Extraction. The effect of the bars were defined as mean (2SD and calculated individually for
relative humidity on the amount of analyte extracted by each fiber triplicate measurements of each point. The results are illustrated
type has been investigated for three different humidities (12.6%, in Figure 4. It can be seen from Figure 4a that the extraction
90.7%, and 100%). Each standard was prepared freshly. The 5 efficiency of the 100 µm PDMS fiber is not significantly affected
µL of water containing the appropriate amount of acetone and by the amount of water present in the bulb. Considering the error
ethanol which was always spiked into the gas sampling bulbs in triplicate measurements, there is no significant difference in
determined the lower limit of the humidity range. Not more than the amount of ethanol and acetone extracted with varying relative
three extractions were performed out of the same bulb. Since humidity. The amount of isoprene extracted increases with
the extraction efficiency largely depends on the temperature, the increasing humidity. About 21% less is extracted at a relative
water bath was carefully kept at 35 °C. To exclude an instrumental humidity of 12.6% compared to a relative humidity of 100%. The
drift in response, the analysis of the standard with a relative effect observed for isoprene follows a trend opposite to what would
humidity of 90.7% was repeated after the analysis of the standard be expected. One reason for this phenomenon might be a slightly
with 100% relative humidity. No difference was observed. Error higher concentration of isoprene in the gas phase at saturated

Analytical Chemistry, Vol. 69, No. 4, February 15, 1997 593


humidity. Although no visible condensation of water was observed tions as used for the calibration curves were prepared. The
on the inner glass wall of the sampling bulbs, there is a certain relative humidity was set at 98%. Thirty minutes after acetone
amount of water adsorbed at the active sites of the glass. Due to and isoprene had been spiked into the bulb, two extractions were
its polarity, water has a higher affinity to these sites than isoprene, performed using the 100 µm PDMS fiber. Ethanol was then
which might also be adsorbed to some degree. At higher relative spiked into the same bulb, and 30 min later the bulbs were
humidity, more water is present, which can replace some of the extracted again in duplicate.
adsorbed isoprene, resulting in a higher concentration in the gas Using this procedure, six different concentrations were ana-
phase. It is not the purpose of this study to carefully investigate lyzed. It was found that high ethanol concentrations (i.e., 17.0
the mechanism by which humidity affects the extraction process. µmol/L ethanol compared to 136 nmol/L acetone and 20.0 nmol/L
It should rather demonstrate that the standards have to be isoprene) do not significantly affect the extraction of acetone and
prepared at 100% relative humidity because this also affects the isoprene when the 100 µm PDMS fiber is used. This allows
accuracy of this analytical method. reliable quantification of acetone and isoprene, even if large
The importance of humidity is even more obvious when the amounts of ethanol are present in the breath.
Carbowax/DVB and PDMS/DVB fibers were investigated. The A short study has also been performed to check the effect of
hypothesis that these fibers extract the analytes by adsorption high ethanol concentrations on the extraction of acetone and
rather than by absorption is consistent with the observations. The isoprene when the Carbowax/DVB and the PDMS/DVB fibers
affinity to the coating material now plays a major role, and were used. Two standards containing 136 nmol/L ethanol, 109
replacement of analytes by compounds with higher affinity is very nmol/L acetone, and 100 nmol/L isoprene were prepared with a
likely to occur. Whereas for all three compounds the difference relative humidity of 98%. The extraction was performed at a
in extraction efficiency between a relative humidity of 90.7% and temperature of 35 °C. Each standard was analyzed three times
100% is not significant, the response differs significantly at a with the same type of fiber. Then, 0.2 µL of pure ethanol was
relative humidity of 12.6%. When the relative humidity is spiked into the bulb, giving a new ethanol concentration of 3.55
increased from approximately 12% to 100%, the amount of acetone µmol/L, which is approximately 26 times greater than before. This
extracted by the fiber drops by 13% and that of isoprene by 33%. standard which still has the same acetone and isoprene concentra-
This can be explained by replacement of these compounds by tion was then analyzed again three times. All six extractions in
water. Ethanol shows the opposite trend. When the relative total were performed out of the same bulb. It was found that the
humidity increases from about 12% to 100%, the response increases effect of high ethanol concentrations on the extraction of acetone
by a factor of 2. Hence, the more water that is present causes and isoprene depends on the type of fiber used for extraction.
more ethanol to be extracted. This is likely caused by a change The Carbowax/DVB fiber indicates no difference in sensitivity to
of the properties of the sorbent material, depending on the amount acetone and isoprene when the ethanol concentration is signifi-
of sorbed water at different humidities. A hypothesis that cantly increased. In this case, the concentrations of the com-
adequately explains the data may be that water will be included pounds are probably too small, so that the active sites of the fiber
in the porous structure of the polymeric Carbowax/DVB coating. are not completely occupied. The results obtained for the 65 µm
This water, now available as a kind of stationary phase, will lead PDMS/DVB fiber indicate that ethanol replaces acetone and
to a higher extraction efficiency for polar ethanol molecules. isoprene from the fiber coating. A decrease in sensitivity by about
These molecules have high affinity to water due to possible 10% is observed, demonstrating that, whenever high ethanol
formation of hydrogen bonds. Whereas the nonpolar isoprene concentrations are present in human breath, the uptake of acetone
molecule has no affinity to water, the polar acetone provides less and isoprene by the PDMS/DVB coating will be affected.
hydrogen bonding interaction. Thus, acetone and isoprene do Limit of Detection. The sensitivity of the SPME technique
not show the same effect. was considered in terms of limit of detection (LOD). The limits
The extraction efficiency for the 65 µm PDMS/DVB fiber were estimated on the basis of the signal-to-noise ratios obtained
strongly depends on the relative humidity as well (Figure 4c). with standards containing the compounds of interest at low
The amount of analytes extracted by the fiber coating decreases concentration levels. Since the sensitivity of the PDMS/DVB fiber
by about 50% for all the analytes of interest when the relative is much higher compared to those of the PDMS and Carbowax/
humidity increases from 12% to 100%. This result indicates that DVB fibers, two standards with different concentrations were
water replaces some of the analytes from the fiber due to the prepared in such a way that the signal-to-noise ratios for the
higher affinity to the coating. analytes were less than 20 when the characteristic masses were
All findings obtained for the three types of fibers investigated monitored. The LOD was defined as that concentration of an
in this study clearly show that it is crucial to set up the standards analyte which produced a signal 3 times greater than the baseline
at 100% relative humidity to obtain accurate results with the SPME noise with GC/MS detection. The average signal-to-noise ratio
method. of four measurements was used to calculate the LOD. The fiber
Effect of Other Matrix Compounds on the Extraction. Extraction blanks always contained coeluting impurities at the same position
efficiency for trace level compounds might be affected by other as acetone with the same characteristic fragment m/z ) 43.
compounds present in much higher concentrations in the sample. Therefore, eight blank samples at a relative humidity of 98% were
In breath this can occur when the subject has consumed alcoholic analyzed, and the average peak are of these measurements was
beverages. Even after one drink, the concentration of ethanol is considered as noise. The signal-to-noise was then calculated
much higher than the concentrations of the other compounds of manually, and the limit of detection was determined in the same
interest. To investigate the effect of high ethanol concentrations way as described above. The results are shown in Table 2. The
on the extraction of acetone and isoprene, the following experi- results indicate that all three types of fiber coatings are suitable
ment has been performed. Standards with the same concentra- for detecting low-level concentrations of ethanol, acetone, and

594 Analytical Chemistry, Vol. 69, No. 4, February 15, 1997


isoprene in human breath. Jones et al. reported a range of At the 95% confidence level, the data revealed no significant
isoprene concentrations found in human breath of 1.60-10.33 loss of analytes, even after a storage time of 8 h for the Carbowax/
nmol/L.25 A mean acetone concentration in normal subjects of DVB and the PDMS/DVB fibers. The mean values obtained for
23.2 nmol/L was reported by Phillips and Greenberg.26 This the PDMS fiber differed significantly for acetone (increase by 10%
indicates that all fibers are capable of detecting concentrations of after storage of 8 h), whereas for ethanol and isoprene no
acetone and isoprene found in the breath of healthy subjects. significant difference was observed. This phenomenon was not
Stability of Standards. To ensure that the developed method further investigated.
will be practical for everyday use, the gaseous standards should Breath Analysis. Breath Sampling. As mentioned before, the
be stable for at least 1 work day. Once prepared in the morning, amount of analyte extracted by the fiber is independent of the
the standards could then be used during the day to quantify the sample volume if the sample volume is large compared to the
compounds of interest. To study the stability of the standards, fiber volume (Vs . KVf). That is true when the fiber is exposed
four gas sampling bulbs were spiked with the analytes in the same directly into the mouth. Moreover, the analyte uptake of the fiber
way as described earlier (98% relative humidity). The temperature does not depend on the flow rate for analytes with K values lower
of the water bath was carefully kept at 35 °C. The standards were than 5000, which has been shown by Martos and Pawliszyn.24 The
extracted using the 100 µm PDMS fiber. The extraction time was whole breath normally contains about two-thirds alveolar breath.
1 min and the desorption time 15 s. The first extraction was The other third is dead space air of the respiratory tract (mouth,
performed 30 min after the analytes were spiked. The other times nose, pharynx, trachea, and bronchi), with lower concentrations
investigated were 1, 2, 4, 8, 24, and 48 h. During the study, seven of breath compounds.28 To obtain reproducible samples, it is
extractions in total were performed out of each bulb. After the important to use alveolar air for breath analysis. The following
end of the study (48 h), a control sample with the same procedure should gaurantee that the fiber is exposed mainly to
concentration was prepared and analyzed 30 min after the analytes the end exhaled breath with the highest concentration of com-
had been spiked to check for any trend in instrument response pounds of interest. Using the modified SPME device described
during the study. The mean value for four replicate measure- above, the subject has to inhale through the nose and hold the
ments was calculated for each time period investigated. The breath for about 5 s to attain equilibration between the blood
relative standard deviation was then calculated for these mean stream and the alveolar air. Then the subject exhales ap-
values (n ) 7) to evaluate the stability of the standards during 48 proximately the first third of the breath quickly through the mouth
h. The RSD values were 4.4% for ethanol, 3.7% for acetone, and into the specially designed mouthpiece. The end part of the
3.8% for isoprene. Assuming an inherent random error and breath should be exhaled as slowly as possible until the total
comparing the results with those for the control sample, it was extraction time of 30 s is reached. Following this procedure was
found that the standards are stable for at least 2 days. Thus, quite convenient and easy as stated by the subject involved in
several calibration standards could be prepared at one time and this study.
stored in the water bath for use throughout the working day. Quantitation of Acetone and Isoprene. Three breath samples
Storage Time. An analysis normally cannot be carried out from a healthy subject were obtained by the method described
immediately after the collection of the sample. Therefore, it is above (65 µm PDMS/DVB fiber) and assayed for acetone and
essential to know how long a sample can be stored on the fiber. isoprene immediately after the extraction (RSD ) 1.2% for acetone,
The storage time for selected VOCs in air has previously been 5.7% for isoprene). Background samples from the room air and
investigated.21 The best results were obtained by sealing the fiber blanks assayed prior to the collection of the breath samples
needle of the SPME device with a septum and keeping the did not show any peaks eluting at the same time as acetone and
sampled fiber in a box filled with dry ice. Martos improved the isoprene.
storage by retracting the fiber approximately 1.5 cm back into The concentrations of acetone and isoprene in the breath were
the needle.24 Therefore, the fiber was removed from the holder, determined to be 38 ( 1.1 nmol/L for acetone and 10 ( 1.4
retracted into the needle, and embedded in a Thermogreen LB-2 nmol/L for isoprene at the 95% confidence level. These values
septum (Supelco) immediately after extraction. The fiber was kept are within the range found for healthy people.25,26 Figure 5 shows
in a box with dry ice for 8 h. It has been shown that the standards the total ion chromatogram of a breath sample taken 1 h after
are stable during 1 working day. Before and after analyzing the the consumption of approximately 0.5 L of beer and stored for 3
fibers which were stored on dry ice for 8 h, two control samples h on dry ice prior to analysis. The peaks, which were identified
were run to monitor any instrumental drift. Since one single fiber by comparing the retention times and mass spectra with reference
of eacy type was used to study the storage conditions, the whole materials, correspond to air, CO2 (1), acetaldehyde (2), ethanol
procedure was repeated on consecutive days to ge three values (3), acetone (4), isoprene (5), and carbon disulfide (6).
for each fiber type. The standards were prepared fresh daily. The
results have been statistically evaluated by comparing the mean CONCLUSIONS
The present work has shown that solid-phase microextraction
values of the control samples which had been analyzed prior to
is applicable for the collection of breath samples. SPME provides
the stored sample (n ) 6) with the mean values of the stored
an effective method for the quantitative analysis of ethanol,
samples (n ) 3) by applying Student’s t-test at the 95% confidence
acetone, and isoprene. Comparing the SPME method to currently
level.27 Prior to the t-test, an F-test was applied to verify that the
used sampling techniques which are usually cumbersome and
standard deviations did not differ significantly.
costly shows numerous advantages. SPME is a rapid method
because the sampling process takes less than 1 min. Hence,
(25) Jones, A. W.; Lagesson, V.; Tagesson, C. J. Chromatogr. B 1995, 672, 1-6.
(26) Phillips, M.; Greenberg, J. J. Chromatogr. B 1987, 422, 235-238. donating a breath sample does not require much effort from the
(27) Miller, J. C.; Miller, J. N. Statistics for Analytical Chemistry, 2nd ed.; Ellis
Horwood Limited: Chichester, England, 1988. (28) Periago, J. F.; Prado, C.; Ibarra, I. J. Chromatogr. A 1993, 657, 147-153.

Analytical Chemistry, Vol. 69, No. 4, February 15, 1997 595


further improved by adding two one-way valves. This would not
only simplify the sampling procedure, but it could also be used
to provide the subject with a pure air supply for inhalation to
reduce atmospheric contamination. The technique is economical,
and since the device is portable, it is suitable for patient sampling.
Samples can be stored on the fiber for 8 h without any significant
loss of analytes. This storage time can certainly be increased and
should be investigated in further studies. Breath samples can,
therefore, be easily obtained at the patient’s room without the
requirement to carry large equipment. The SPME device is
simple enough for the potential use in a physician’s consulting
room.
To date, the method is limited to compounds with relatively
high concentrations in human breath, but as more coatings
Figure 5. GC/MS chromatogram of a breath sample. Breath become available, the sensitivity and selectivity of the method
sample taken after the consumption of alcohol, 65 µm Carbowax/
could be improved. The initial results of this study justify further
DVB fiber. Peak assignment: air, CO2 (1), acetaldehyde (2), ethanol
(3), acetone (4), isoprene (5), and carbon disulfide (6). Extraction research in this new use of SPME for application to other
time, 30 s; desorption time, 15 s; desorption temperature, 200 °C. compounds of interest in human breath.

ACKNOWLEDGMENT
patient. A complete analysis is performed in less than 10 min, Financial support from the Natural Sciences and Engineering
which allows a much higher sample throughput compared to other Research Council of Canada, Supelco, and Varian is greatly
methods described so far. The technique is sufficiently sensitive appreciated. This work was presented in part at the 21st
to detect 5.8 nmol/L ethanol, 1.8 nmol/L acetone, and 0.3 nmol/L International Symposium on Chromatography, September 15-
isoprene, respectively. The response was linear over the concen- 20, 1996, Stuttgart, Germany.
tration ranges of interest, and the data obtained for accuracy and
precision are satisfactory. However, further evaluation of this new
sampling approach against well-understood blood measurements Received for review July 25, 1996. Accepted November
is necessary before it can be relied upon for clinical measure- 18, 1996.X
ments. AC960749L
The device is easy to handle, and because of its simplicity,
the buildup of contaminants is unlikely. The mouthpiece can be X
Abstract published in Advance ACS Abstracts, January 1, 1997.

596 Analytical Chemistry, Vol. 69, No. 4, February 15, 1997

You might also like