Scale Indicator
ABS / INOX
USER´S MANUAL
Ref: 49DMI610EN08
Contents
1.1 KEYBOARD AND DISPLAY........................................................................................................................ 2
1.2 ACCESSORIES .......................................................................................................................................... 3
1.2.1 Accesories Model ABS ................................................................................................................... 3
1.2.2 Accessories Model INOX ............................................................................................................... 3
1.3 FUNCTIONS OF THE KEYS......................................................................................................................... 4
1.4 PLATFORM CONNECTION ......................................................................................................................... 4
1.5 CHARACTERISTICS................................................................................................................................... 5
1.5.1 Functional characteristics ............................................................................................................. 5
1.5.2 Metrological characteristics. ......................................................................................................... 5
2. SWITCH ON ................................................................................................................................................. 5
1
DESCRIPTION OF THE INSTRUMENT
Rear
casing
Front
casing
Display
Keyboard
2
1.2 Accessories
1.2.1 Accesories Model ABS
The adhesives Ref: BL-P3624 must be placed in the holes located at the back side of the indicator (2
in model stainless steel and 1 in model ABS) when the installation of the indicator doesn´t require
metrologic control (without M ).
3
1.3 Functions of the keys
Below are the functions of the keys:
On /Off
Delete the value of the digit being edited. Keep pressed to quit.
4
1.5 Characteristics
1.5.1 Functional characteristics
2. Switch on
Press for two seconds to switch on the device.
Once the instrument has been plugged in, press for two seconds.
U 1 . 0 1 5
This version can be changed as further functions are added to the display.
P 1. 0 0 1
The OIML indicator displays to indicate that it is working with OIML parameters.
- O I M L -
5
Then the display does the initial auto zero in line with the following criteria:
c E r 0
1. The weight must be stable (the stable indicator displays )
2. The value of the weight is not less than -5% the calibration zero value.
3. The value of the weight is not greater than 15% of the zero calibration value.
3. Operating procedure
The possible operations are as follows
To do a tare operation, place the mass to be tared on the tray and press .
The scale displays the word Net.
To do successive tare operations, place the new mass to be tared on the tray and press again.
To see Net weight again, press again. The scale displays the Net indication.
t A r. b L 0
To remove the tare, press →0← for two seconds, and the following displays:
t A r. S t L
6
3.3.5 Manual tare
To enter a manual tare value using the keyboard, keep pressed. Use the keys described
below to enter the new value.
Cal. PMU
PMU
CHECK
bY 10
ACUMUL
t. CodE
n tiC
Date
Time
ICount
DoSiFi
LiMite
Weight
7
3.4.1 Cal PMU – CALCULATING UNIT WEIGHT PIECECOUNTING
With this function you can calculate the unit weight of the pieces. To calculate the average unit
weight, place a number of known items on platform and press PRINT . The screen displays as
follows
The indicator displays the text “WEiUni” and then the unit weight of the piece for two seconds. Then
the indicator returns to weight mode and displays the number of pieces on the scale.
To return to the weight display mode press the key MODE , go to the parameter “weight” and
press PRINT .
This function allows to enter the known weight value of the pieces and from that weight calculate
the number of pieces. To enter the unit weight, press PRINT . The screen displays,
P. x x x x
Use the keys to enter the value of the unit weight.
On terminating the process, the device automatically displays the number of pieces on the scale. To
return to the weight display mode press the key MODE , go to the parameter “weight” and press
PRINT .
8
3.4.3 CHECK
The display is ready to automatically manage up to 3 limits using weight records for each one of
them.
These records are called Lower limit (L.Inf), Limit (OK) and Upper limit (L.Sup).
The indicators light signal (LED) will be enabled provided the weight on the scale is within these two
limits.
signal
When accessing the checkweigher programming menu, the following menu displays:
START WEIGHT
TARGET TARGET WEIGHT
Ttype LIMIT TYPE
LOW L LOWER WEIGHT
HIGH. L UPPER WEIGHT
ALARM ALARM
BTYPE TYPE OF SOUND ALARM
BMODE MODE OF SOUND ALARM
CHU. ON PROGRAM ACTIVATED
t A r G E t - It programs the value for the target weight , when it reaches this value
the signal OK is enabled, green Led (t xx.xxx)
- P C 0 0 0 - This parameter is for programming the lower limit in % of the target weight.
Below the target weigh the yellow led “- “will be activated (-PC.xxx)
9
Type of limits using weight values
L O U L - This parameter is for programming the lower limit weight. . Below this
weigh the yellow led “- “will be activated (L xx.xxx).
H i G H L - This parameter is for programming the upper limit weight. Above this
weigh the red led “+ “will be activated (H xx.xxx)
A L A R M - This parameter is for programming the weight which above this value will
activate the signal alarm, 3 Leds on. (A xx.xxx)
2. A sound is emitted when the weight is outside the limits and is greater
than 20 divisions of the lowest range used.
When this function is selected, the device displays one more decimal in the weight. After 10
seconds, the function deactivates automatically. Press PRINT Once enabled, the following
message displays for 1 second:
b Y - 1 0
Once the function is enabled the display is enabled. The smallest figure of the display is enabled
and indicates the value by 1/10 of the verification division.
10
THE DECIMAL IS MOVED, A DIGIT IS NOT ADDED TO THE DISPLAY
8 8 8 8 8 8 8
When the device is switched on, the weight data display does not respect the metrological
indicate. As a result all communication with the serial channel is inhibited.
The display is active for approximately 10 seconds. Then the terminal automatically displays the
normal weight.
3.4.5 ACUMUL
This parameter is used when there is a printer connected to the indicator. Is also required to select
the protocol “Printer” in the communication parameters.
With this function you can enter a 6 digits code which is added to the printed receipt. Use the keys
described below to enter the code in order to enter the new value.
With this function you can print on the receipt a receipt number which automatically increases on
completing the weighing operation (delivery or receipt number). If this field is set as equal to 0, it is
not printed and does not increase. Use the keys described below to enter the new value.
11
3.4.8 Date – DATE
This function is for updating the date. Use the keys described below to enter the new value.
This function is for updating the time. Use the keys described below to enter the new value..
Use this function to see the converter’s internal counts. The following are the possible options:
To return to the weight display mode enter in the “Weight” parameter which is explained below or
disable the internal readings function
When the dispenser mode is activated, the indicator will work with the latest configuration
programmed, the settings programmed will be not reset when the indicator is turned off.
When the parameter “DOSIFI”, is selected the following menu will appear:
12
RESET YES / NO
PROD1 FAST FILLING WEIGHT
WEIGHT TIMER FAST FILLING
PROD2 SLOW FILLING TIME
WEIGHT TIMER SLOW FILLING
Z. DESCARGA REMAINDER WEIGHT
MAXIMUN TIMER REMAINDER WEIGHT
ACTIVA YES / NO
R E S E T - Resets the values , puts ceros in all the settings of the dispenser menu.
P r o d 1 - Its programmed the value of the fast filling weight, when this value is
reached will activate the output 2 (U xx.xxx), following to program is the
time is seconds the output 2 will be activated , (S xxxx.x), once the time
goes over the time programmed, and the weight doesn’t change, the alarm
relay will be activated (output 4).
P r o d 2 - Its programmed the value of slow filling weight, when this value is
reached will activate the output 3(U xx.xxx), following to program is the
time in seconds the output 3 will be activated. (S xxxx.x), once the time
goes over the time programmed and the weight doesn’t change the alarm
relay will be activated (output 4).
- 0 - E r r
A C t i v A - Asks to activate the disperser mode, in the display will show - DOSI
ON / OFF. The option ON, will activate output 1.
The 4 inputs that the relay board has emulate the keyboard when the dispenser mode is
activated, the function that do these inputs are the following:
13
The input / output board consist on 4 inputs opto-accopled and 4 independent digital outputs
they are potential free pins with four relays, which is possible to operate lights, signals to PLCs.
14
3.4.12 LIMITS (APPLIES ONLY WITH INPUT/OUTPUT BOARD)
With this parameter when the weight is betwwen the values of each limit programmed , the
corresponding output programmed will be activated, and will remain active until the weight reaches
the values of another limit.
When the parameter “LIMIT”, is selected the following menu will appear:
RESET YES / NO
INICIA (START) START WEIGHT
L. INF LOWER LIMIT WEIGHT
L SUP HIGHER LIMIT WEIGHT
ALARM WEIGHT ALARM
ACTIVA YES / NO
R E S E T - Resets the values of this parameter, put ceros in all the setting of the
limits menu.
i n i C i A - When the weight reaches the value programmed in this parameter the
signal “Inicia” will be activated. (S xx.xxx)
L i n f - Its programmed the weight value L INF , when the weight reaches or
goes over this value will activated the output 2 , (L xxxx.x).
L S u P - Its programmed the weight value L SUP , when the weigh is between L
INF and L SUP will activated output 3 , (Hxxxx.x).
A L A r M A - Its programmed the value above of which the relay alarm will be
activated output 4 (A xx.xxx)
When the weight is higher than cero the mode of limits activate automatically.
15
This is the description of the connections:
Use this function to return from any working mode to the weighing mode.
4. CONNECTION TO PRINTER
To connect a printer to the indicator, besides using the protocol “PRINT” in the
communication menu, the following configuration has to be set:
• Baudrate: 9600
• Bits: 8
• Parity: none
• Stop bit: 2
16
4.1 SELECTION OF WEIGHT UNITS
In OIML mode: The weights units printed are selected in the weight adjustment (kg or lb)
IN NO OIML mode: It is possible to change the weight units , See the technical manual for more
details . NO OIML-CAL_PA->VISOR.
Formats 1-2-5-6-8-7: Weight units printed are the ones shown at the moment of the weight.
Formats 3-4-5-6: Weight units are the ones chosen when the weight was adjusted.
To Choose the printing format select: Protocols -> Printer -> PrnF
Note: The header lines are enter using the protocol PROTCOM by TCP
17
4.1.2 Printing – with accumulation
The accumulation is made in weight mode, to select this mode enter in: MODE -> ACUMUL.
There are two working modes:
• Auto-accumulation: in this mode it has to be programmed the number of
accumulations we want to do. Once it has been set it will accumulate by stable
weight.
• Accumulation manual: To accumulate each weigh in this mode, press the key
PRINT .
Every time a weight is accumulated it will display -- A C C
Once the accumulations have finished, press the key to print the total ticket.
The accumulation mode has to match with the printing mode, the accumulation mode manual
goes together with the printing mode manual and the auto-accumulation mode goes together
with stable weight.
=========================
TOTAL
=========================
hh:mm dd/mm/aa
****************************
Weights: #XXXXX - #XXXXX
Number of Weights: XXXXX
=========================
Tot Gross: XX.XXX Kg
------------------------------------------
Tot Net: XX.XXX Kg
------------------------------------------
Tot Tare: XX.XXX Kg
==========================
18
5. Menú - Configuration (SET-UP)
To enter in this menu press for two second the key MODE . Then enter the PIN code ( 4 numerical
figures , by default the code is 0000)
Press the key PRINT to access , and the display will show the following message:
C A L C L o - Calibration close. Allows to set up indicators configuration as the
OIML regulation.
C A L O P e - Calibration open. Allows to set up the indicators configuration in
NO OIML mode. Look at the technical manual for more information
In this menu the available functions are:
PArAME
ConFiG
TESt
5.1 PArAME – PARAMETERS
In this Menu are the communications settings of the indicator. The indicator has two RS232
independent serial ports, Com1 and Com2 ; Com1 is powered (5VDC). There is also the option of
having an RS422/485 port and an Ethernet (TCP or UDP) communications.
The settings to program are the followings:
5.1.3 Baudrate
Once selected will display for a second the world BAUD and It enables to define the
communication speed used in the communications ports. The possible choices are:
300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600 115200 baud.
Use the keys described below to select the required speed.
19
5.1.4 Protocols
Use →0← to select the protocol you wish to work with. The device currently works with
different communication protocols to be used through the RS232 ports. The following are the
protocols:
T I s a • Tisa Protocol
V D - 0 • VD-0 Protocol
V D - 3 0 0 • VD-300 Protocol
U s e r d f • Userdf Protocol
S S C a R • SSCARr Protocol
C S C O M P • CsComp Protocol
d O L A r • Dolar Protocol
r D • Remote RD Protocol
N O n e • No Protocol
• Printer Protocol
P r i n T
• SPI2 Protocol
S P I 2
• Tol DS Protocol
T O l D S
• Epsa Protocol
E P s a
• M-T Protocol
M - T
• F-501 Protocol
F 5 0 1
• SAIE Protocol
S A i e
• Multipunto 2000 Protocol
M U l t i P
• Seur Protocol
S E U R
Press PRINT to move onto the next parameter.
5.1.5 Ind add - ADDRESS DISPLAY
Displays the indicators address as Add XX , where XX can go from 00 to 99. It is used to identify
the displays in the same communications network. Use the keys described below to enter the
address.
20
5.1.6 CoM on – TRANSMISSION MODE
It allows you to set up the terminal for sending data to the PC through the RS-232 com1 and com 2
serial channel in one of the following ways:
Continuous: The display sends the data string, depending on the protocol, continuously.
Manual: The display sends the data string, depending on the protocol, on pressing PRINT
Pc req.: The display sends the data string, depending on the protocol, once the order has been
received from the PC
Stable: The display sends the data string, depending on the protocol, once a stable weight value
has been obtained
Interval: The display sends the data string, on each interval of time (ms) , this time is programmed
once this mode is selected. Press PRINT to quit the communications menu.
5.1.8 Ethernet
To communicate by Ethernet the following parameters must be configured:
IP vis Indicator´s Ip
Mask Subnet Masc
MAC vi Indicator´s Mac address
Gatewa Gateway network
TCP? Yes/No ( select to activated TCP)
Port vi Indicator´s TCP port
Port PC PC´s TCP port
IP-PC PC´s IP connected by TCP
Protoc Protocol to use by TCP
UDP? Yes/No ( select to activated UDP)
Port vi Indicator´s UDP port
Port PC PC´s UDP port
IP-PC PC´s IP connected by UDP
Protoc Protocol to use by UDP
ACTIVA Activate Ethernet (Yes/No)
21
5.1.8.1 Protocols Ethernet
Activated the Ethernet communication is added to the ones already available by the serial port
the following protocol:
P r o C o m • Protocol Protcom
Press PRINT to select the weight filter you wish to apply to weighing operations.
The possible weight filters go from 1 to 6. The higher the value the criterion to obtain a stable
weight is bigger. Use the keys described below to enter the new value.
22
5.2.2 Date
Press PRINT to program the display’s date. The date is not lost on switching off the display.
Use the keys described below to enter the new value.
5.2.3 Time
Press PRINT to program the time on the display. The time is not lost on switching off the
display.
Use the keys described below to enter the new value.
By Pressing PRINT allows to set a timer (seconds) of switching off the indicator if it is not
operated during in this time. Use the keys described below to enter the new value.
5.2.5 Mu – Voltage
By pressing PRINT displays the voltage of the load cell (mV). To escape of this function keep
5.3 Test
Allows to test the COM 1, COM 2, RS485/ RS422, input and output board and the LEDs of the
indicator. Select the test to do and the test will run for 5 seconds. To find out more information
about how to run these test consult the technical manual.
Use the keys below to select the test you require.
23
6. WEIGHT ADJUSTMENT
To start the calibration process, press the calibration button on the inside of the indicator
through the calibration hole.
Access to
calibration button
Access to
calibration button
24
The indicator displays:
O I M L ?
Select the OIML option using the MODE and →0← .keys.
- O I M L -
G r a v i t G . 9 8 0 4
Uni t s
Selects the unit weight kg or lbs in which the indicator will do the calibration , the unit chosen will be
the predetermined from this moment.
PAS CE R
Selects if it is needed or not to pass through zero when send data by stable weight. The options are:
SI : -To send data consecutive the weight should pass throw zero , or NO : To send data
continuously the weight should vary 20 times the step selected during the calibration procedure or to
pass thorough zero.
press PRINT to continue.
M o d E
In order to select the range type, these are the possible options:
Normal= Monorange, MI=Multi-interval, MR=Multirange:
n o r M a l Monorange
- M i - Multi-Interval
-M–r- Multi-Range
25
Press MODE or →0← keys to select the type of range and then press PRINT to
select the number of ranges. The indicator will display the following message for two
seconds:
i n F S
The display then allows you to enter the instrument’s maximum range ( kg ). The keys
with the following functions are used:
F. 0 0 0 1 5
Once the maximum range has been selected, press PRINT .
The next parameters are the weight decimals,
d. P o i n 3
Select the number of decimals with the MODE and →0← keys. Then press PRINT to enter
the interval.
In the case of multi-range or multi-interval instruments, the following is the interval of range 2.
d i v. 0 5
Use the MODE and →0← keys to select the interval. The possible values are: 1,2,5,10,20 and
50. Then press PRINT
Select the type of filter you are going to use in weight adjustment. The possible weight filters go
from 1 to 6. The higher the value is set the higher the criterion to obtain a stable weight.
F i l t E r
Then press PRINT
26
USER’S MANUAL
In the case of multi-range or multi-interval instruments, the display requests the value of the first range
(Max1)
r A n G or 1 F. 0 0 0 0 6
Enter the value of the first range using the edit keys and press PRINT to continue. The
display will than carry out the zero calibration.
- - 0 - -
Check that the platform is empty and stable and press PRINT
Once zero calibration is completed, enter the g value where the calibration has been done. Enter the
value using the edit keys.
G r a v i t G . 9 8 0 4
Once the gravity value is entered, press the PRINT key. The display will show the mass value to
be used for calibration
L o a d P 1 0. 0 0 0
Edit the value with the edition keys.
Press PRINT to validate the selected value
Place the calibration mass on the plate, check that it is stable and press PRINT .
C A L F S
The screen flashes for a few seconds and then the following displays
1 0. 0 0 0
Press PRINT again. The display indicates that calibration has ended.
C A L E n d
27
Once calibration has finished, the indicator displays a request message to close calibration.
Use the MODE and →0← keys to select Calibration Closed or Calibration Open. The display
shows:
C A L o P E C A L C L o
Calibration open is selected if you wish to carry out calibration tests and calibration closed is selected
when the indicator is definitively installed.
The indicator will display a message to indicate if the calibration has been saved:
S A v E d
The calibration process has finished and the indicator will restart itself.
28