CALIBRATION CERTIFICATE

The humidity sensor identified in this certificate is calibrated according to the ASTM E104-02
standard practice.

Relative humidity calibration method

Calibration was made utilizing constant humidity of saturated aqueous salt solution.
Used salts Magnesium chloride (MgCl2) 32.8 % RH ±0.2 at +25.0 °C
Sodium chloride (NaCl) 75.3 % RH ±0.2 at +25.0 °C

Calibrated probe

Type THPS
Serial (MAC) 70B3D5C39B001BC2
Manufacturer MeshWorks Wireless Ltd
Measuring range 0...100 % RH
Accuracy ±3.0 % RH. Annual drift max ± 0.5 % RH

Calibration result

Calibration temperature 22.2 °C


Reference value 33.0 / 75.4 % RH
Value of calibrated probe 33.9 / 76.4 % RH
Result ACCEPTED
Calibration date 2018-01-29
Valid until 2020-01-29
Location Tampere
Company MeshWorks Wireless Ltd
Calibration operator Samuel Niemi