90 - PDFsam - 03 - 1261 - Technical Description 3460aroMA AA83346-07
90 - PDFsam - 03 - 1261 - Technical Description 3460aroMA AA83346-07
UC-E
c h
600-800V
d f g
e
t
Figure 3.15
In order to observe these signals, 100x probes must again be used in differential mode, the potential of the
cathode rising regularly to 400 V or 450 V. In order to measure the current, it is preferable to use a current
probe, but the measurement can also be made, in differential, at the terminals of R26 on the Power Supply
and IGBT’s Gate Circuit board.
Currents gate
0.1 A
0A
15 V
Voltage gate
0V
Figure 3.16
We can measure this current by using a current probe, or by the intermediary of the 1 Ω limitation resistor
which is in series with the coil.
In order to avoid excessive heating of the coil, the current must not exceed 6 A peak. The only way to reduce
the current is to diminish t1.