Professional Documents
Culture Documents
Manual Zeus A5 LR
Manual Zeus A5 LR
Integrator Manual
TABLE OF CONTENTS
Table of Contents
1 Introduction ................................................................................ 6
1.1 Intended Use of ZEUS ..................................................................... 6
1.2 About this Manual ........................................................................... 6
1.2 About this Manual Use adequate enclosure of ZEUS to ensure appropriate user protection.
This Manual is designed to support the integration engineer during integration, operation For safe and correct use of ZEUS, it is essential that both operating and service personnel follow
and qualification of ZEUS. You should read through the entire Manual before beginning to install generally accepted safety procedures as well as the safety instructions given in this document,
and operate the pipetting module. It describes the features of ZEUS and its integration in liquid the ZEUS Manual.
handling devices. Both the hardware and the communication between the pipetting module and Any collisions of the pipetting module with other
the liquid handling platform are detailed in this manual. After reading this manual the integration components have to be ruled out. ATTENTION! Teaching of
engineer should be capable of operating ZEUS. ZEUS and accurate definition
Cleaning, dismantling, and maintenance should only
of the Container Geometry and
Throughout this manual, following conventions are used to highlight various kinds of information: be performed by personnel trained in such work,
Deck Geometry parameters is
and who are aware of the possible hazards involved.
required prior operation.
Only certified technicians are authorized to perform
ATTENTION! Essential information for avoiding personal injury or damage mechanical maintenance on ZEUS.
to equipment. For repair or shipment, all mechanical parts must be put in their rest positions. Any ZEUS sent
away for repair must always be decontaminated (see Chapter 9.4). The pipetting module must
be sent back in the original reusable packaging box. There must not be any labware including
tip on the pipetting module during transportation.
NOTE: Important instructions or interesting information.
Only original ZEUS-approved parts, tools and accessories (e.g. pipetting tips) may be used with
the pipetting module (see Chapter 11). The use of any other parts, any alterations or modifica-
tions to the system may be dangerous and will void the warranty.
used in the analytical work. In case of incidental spillage, carefully wipe with a dry cloth, taking 2.4 Chemical, Radioactive or Biological Hazard Precautions
into account the nature of the spilled liquid and the necessary safety precautions.
Selection of the appropriate safety level and implementation of the required safety measures for
During operations, stand clear of any moving parts or the working deck of the instrument. working with ZEUS is the sole responsibility of the user.
Do not attempt to disable any security measure.
If ZEUS becomes contaminated with biohazardous, radioactive or chemical material, it should
The instrument developer is responsible for the use of the appropriate probe, materials (e.g. pipet- be cleaned in accordance with the maintenance procedures given in the Chapter 6. Always de-
ting tips, containers), and parameters (e.g. liquid class, container geometry, tip type) consistent
contaminate your instrument and replace the samples on the instrument in case a tip detaches,
with the validated application. Ensure the use of tips of the appropriate size. Failure to do so may
to avoid cross-contamination and pipetting errors. Failure to observe and carry out the mainte-
lead to pipetting error, contamination or damages to ZEUS. Make sure that the pipetted samples
nance procedures may impair the reliability and correct functioning of the pipetting module.
are consistent with the samples used for performance validation. Generally, it is recommended to
centrifuge the samples to avoid foam, bubbles or clot formation. If working with hazardous samples observe and carry out the maintenance procedures, pay-
Pipetting accuracy and precision may not be ensured in case of process interruption. As a conse- ing particular attention to cleaning and decontamination. Use filter tips and do not reuse tips
quence it is recommended to discard the pipetted sample and repeat the required pipetting steps. to avoid cross-contamination. Design your waste container to avoid spillage during tip discard.
Wear gloves during the cleaning process. Any surfaces on which liquid is spilled must be
decontaminated.
2.2 Environmental Influence Do not use disinfecting materials which contain hypochlorite (Javelle water, Chlorox) or
The accuracy and the precision of pipetting with ZEUS are highly influenced by the laboratory envi- bleaching fluids.
ronmental conditions. Temperature, humidity and air pressure of the laboratory should be controlled
to achieve specified accuracy and precision. When working with samples, which will be used in
particularly sensitive tests, take the evaporation and condensation that may occur into account.
3 Product Description
During operation, ZEUS should be shielded from direct sunlight and intense artificial light.
Liquid
Sensor board
Sample tube
Figure 2:
Innovative pipetting head technology.
Main board
3.3 Automated Pipetting Process LLD only works with new and empty tips during the aspiration phase. This process is slightly
slower than capacitive LLD but is the only way to detect non-conductive liquids. The sensitivity of
both cLLD and pLLD may be adjusted to the vessel size, tip type and the physical properties of
3.3.1 Pipetting Process the liquid to be detected.
The automated pipetting with the ZEUS follows five successive steps (Figure 4).
3.3.3 Qualitative Pipetting Monitoring (QPM)
1 Zeus picks up a pipetting tip and ZEUS is equipped with a process monitoring system called Qualitative Pipette Monitoring (QPM),
automatically verifies its presence. which ensures proper pipetting. If QPM is activated, the air pressure in the pipette tip is monitored
1
pick-up during the entire aspiration and dispensing cycle and is instantly compared to predefined
tip 2 Zeus supports two modes of Liquid Level Detection (LLD), capacitance thresholds. When ZEUS aspirates or dispenses foam, clots, or air due to insufficient reagent, the
LLD for conductive liquids and pressure LLD for non-conductive liquids. system immediately identifies the error. Adequate error-handling procedures may be associated
2 The integrated z-motion is used to detect the liquid level and to follow the to these errors and automatically performed to ensure proper pipetting. The foam detection is
Detect liquid liquid during aspiration and dispense. application dependent.
level
The working principle of QPM is illustrated in Figure 5 for the case of tip clogging during aspira-
3 ZEUS aspirates the volume of liquid defined by the user in the range of
tion. The pressure is monitored during the whole pipetting cycle. Normal behavior is shown in
3 1 μl - 1000 μl into the pipetting tip. The complete pipetting cycle is
Aspirate red. If the tip clogs, e.g. during clot aspiration (blue), the pressure drops below a predefined
automatically verified with the Qualitative Pipette Monitoring (QPM).
liquid threshold (dashed line). The system identifies an error. A more detailed description of all error
QPM detects the aspiration of foam, clots or air and follows predefined error
cases is provided in Chapter 6.
handling procedures to ensure precise and accurate pipetting.
4
Dispense
4 The liquid is dispensed in one or more aliquots and the process is
monitored by QPM.
5
Discard 5 Zeus discharges the used pipetting tip.
tip Successful
Aspiration
Pressure
Figure 4: ZEUS automated pipetting process.
Clot Threshold
3.3.2 Liquid Level Detection Tip Clogged
ZEUS contains both capacitive and pressure liquid level detection (LLD). Capacitive LLD (cLLD)
may only be used with conductive liquids and requires conductive tips. The pipetting channel
Time
modulates the tip with a high frequency AC signal. As soon as the tip touches the surface of the
liquid, the capacitance of the system changes, which causes a change in the AC signal: the liquid
level is detected. Figure 5: The working principle of QPM, illustrated for tip clogging during aspiration.
The pressure LLD (pLLD) can be used with both conductive and non-conductive liquids. The z-
drive moves down slowly while the plunger aspirates (i.e. moves up). When the tip hits the surface
of the liquid the pressure sensor measures a change in pressure indicating that the surface has
been found. This detection method cannot be used if liquid is inside the tip. Therefore, pressure
4 Getting Started
4.1 Unpacking
Remove the transportation strap and pull the foam packaging carefully out of the cardboard
box. Enclosed you will find the ZEUS pipetting module and a Declaration of Quality with the Spindle
results of the final inspection (Figure 7). Please take the pipetting module carefully out of the box
and inspect it for shipping damages or missing parts.
Reference surfaces
M4 threads (6 pieces)
ATTENTION! Keep the original packing material in case the module has to
be returned to Hamilton. Observe standard handling precautions to avoid Figure 8: Position of the seven M4 threads required for proper mounting of ZEUS.
electrostatic discharge.
4.3.2 Example
Below is a sample of the protocol. This example is for requesting of the firmware-version.
10 9 8 7 6 5 4 3 2 1 0
0-Data CAN node of sender CAN node of receiver
1-Kick
Control byte
Number of bytes Bit 6 Bit 5 Bit 4
0 0 0 0
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
1 0 0 1
Byte 1 > data
2 0 1 0
Byte 2 > data
3 0 1 1
Byte 3 > data
4 1 0 0
Byte 4 > data
5 1 0 1
Byte 5 > data
6 1 1 0
Byte 6 > data
7 1 1 1
Byte 7 > data
EOM Number of data bytes counter Byte 8 > control
Data (Byte 1…7) These 4 bits are used as a counter for the different messages. As one message can consist of
more than 7 data bytes, this counter is used to index the frames. The counter always starts on
In the Hamilton CAN protocol 7 bytes of data can be transmitted in one frame. The 8th byte is 0001b with every new frame. If 1111b is reached in one message, it must be started over at 0001b.
used for protocol information (EOM, number of data bytes, counter). If the message is longer
than 7 bytes, it must be sent in more than one CAN frame. The data field contains the com- To detect duplicate objects, the counter should always be checked. This it can be decided if
mand string to ZEUS or the response string from ZEUS. The possible commands and respons- the object should be used or ignored.
es are detailed in the following chapter.
The CAN node and the Baud Rate must be assigned using the DIP-switches on the Main
Board (Figure 11). The address of ZEUS is already preset (default: 1 ON). It must be adjusted if
additional units are used.
DIP SWITCH After execution of a command, or in the event of an error, the module returns a confirmation of
CAN
node the command (symbolized by in the command tables below). This consists of the command
6 5 4 3 2 1
itself, the returned id parameter and the error parameter (er) which contains the error number. If
0 0 0 0 1 1 no id parameter is sent along with the command string, the id parameter last sent or the default
Baud Rate 0 0 0 1 0 2 value is used for the response. The possible errors are listed in the error table (Chapter 5.5).
(0 = 500 kBaud,
... ... ... ... ... ... An example command string is shown in Figure 12.
1 = 125 kBaud)
1 1 1 1 1 31
Figure 11: Assignment of CAN node and baud rate. Command string sent
<GA><ai10000ge01go01lq01gq0lb0cf1000> -- Command
-- Associated parameters
NOTE: the address 00 is reserved to the master and not recommended for the
ZEUS module.
Response
<GA><id0001><er00> -- Command
-- Default id value
-- response
5 Programming Basic Functions Figure 12: Example command string and the associated response.
NOTE: Only one command can be sent per message. Parameters are not sent to
a defined address as in standard open CAN protocols. Request commands may be Example response
sent anytime. <RF><id0001><rf><1.00.000(xx)> -- Command
-- Default id value
-- Parameter name
Parameters are initialized with default values after power on and software reset. These values re- -- response firmware version
main valid until new values are defined. All required parameters must be sent with a command.
If a parameter is missing the value last sent or the default value will be used. Figure 13: Command string required to request a firmware version and example response.
Parameter tables have been created in order to define and store the values of certain pre-defined
parameters, such as the tip parameters, the geometry of the container, the geometry of the
instrument or the liquid class of the solutions.
ZI Initialize z-Drive
This command moves Zeus to the
manufacturer preset z-origin, which gy
corresponds to the highest possible
position.
id#### 0000..9999 0 Identification Number
ZIid####er## See Error Messages gy
2800 mm 2800 mm
Figure 15: Definition of the z-position of ZEUS with and without tip.
Command Parameter Parameter Default Name and Description Command Parameter Parameter Default Name and Description
values value values value
GZ Move z-Drive GT Pick-up Tip
This command moves Zeus along the Following steps are performed:
z-axis to a defined position relative to the 1) Zeus drives up to the Minimal Start
preset z-origin and at the choosen speed. Traverse Height defined in the Deck
Geometry Table (th), only if currently
id#### 0000..9999 0 Identification Number
at a lower position.
gy# 0000..2800 0 Position of the z-Drive 2) Zeus drives to the Beginning of Tip
The position of the z-drive is set in Picking Position defined in the Deck
increments of 0.1 mm and is relative to Geometry Table (tm).
the preset z-origin. 3) Zeus drives further down until a tip
gw 0..1 0 Speed of the z-Drive has been picked up, as recognized
Two speeds can be set: by the sensor detecting tip presence.
0=slow, 1=normal. 4) Zeus drives up to the End Traverse
Standard pipetting applications require Height (te) defined in the Deck
the normal speed. The slow speed is Geometry Table.
essentially used when teaching the drive. id#### 0000..9999 0 Identification Number
GZid####er## See Error Messages tt## 0..30 0 Tip Type Table Index
See Tip Type Table (Chapter 5.12.3)
go## 00..99 0 Deck Geometry Table Index
See Deck Geometry Table (Chapter 5.12.2)
GTid####er## See Error Messages
id#### 0000..9999 0 Identification Number ge## 00..99 0 Container Geometry Table Index
ai##### 0000..10000 0 Aspiration Volume go## 00..99 0 Deck Geometry Table Index
The volume aspirated is set in gq# 0..1 0 QPM on/off, 0=off;1=on
increments of 0.1 μL lq## 00..99 0 Liquid Class Table Index
ge## 00..99 0 Container Geometry Table Index lb# 0..1 0 LLD on/off, 0=off;1=on
go## 00..99 0 Deck Geometry Table Index Pressure LLD is not available during
dispensing.
lq## 00..99 0 Liquid Class Table Index
zp#### 0000..2800 0 LLD Search Position
gq# 0..1 0 QPM on/off If the LLD mode is on, this is the
0=off;1=on z-position (in increments of 0.1 mm)
from which the LLD detection will start
lb# 0..1 0 LLD on/off
looking for liquid. zp must always be
0=off;1=on higher than the real liquid level.
Command Parameter Parameter Default Name and Description Command Parameter Parameter Default Name and Description
values value values value
cf#### 0000..2800 0 Liquid Surface AV Switch Off Command
If both the LLD mode and the Search Must be used before switching off the
Bottom Mode are off, this is the z- position
unit. Lifetime counter and cycle
(in increments of 0.1 mm) at which the
liquid will be dispensed. It should be counters are saved on eeprom prior
defined with a tolerance of ± 0.5 mm. to switching off.
zm# 0..1 0 Search Bottom Mode id#### 0000..9999 0 Identification Number
This mode is not possible with 50 µL AVid####er## See Error Messages
and 25 µL Axygen tips.
0=off, 1=on
ma##### 00000..10000 0 Mix Volume
This is the volume (set in increments of
0.1 µL) required to mix the probe.
5.9 Volume Check Commands
mb##### 00000..25000 0 Mix Flow Rate The Volume Check commands are used to calculate the volume present in the container before
This is the flow rate (set in increments or after pipetting. The calculation relies on the accurate definition of the container geometry and
of 0.1 µL/s) at which the probe is mixed. the use of the proper container geometry table.
dn## 00..99 0 Mix Cycles
Number of mixing cycles.
GDid####yl####er## See Error Messages Command Parameter Parameter Default Name and Description
yl#### is the position of the liquid values value
surface after the command in 0.1 mm
(values between 0000-2800). The values GJ Calculate Container Volume
will not be updated if an error occurred. Calculate the volume present in the
container based on the container
geometry and the liquid level. The liquid
level is detected by LLD if the LLD mode
is on or defined.
id#### 0000..9999 0 Identification Number
5.8 Switch Off Commands ge## 00..99 0 Container Geometry Table Index
It is recommended to perform the following procedure prior to switching off the instrument: go## 00..99 0 Deck Geometry Table Index
Command Parameter Parameter Default Name and Description Command Parameter Parameter Default Name and Description
values value values value
GJid####yl####aw######er## See Error Messages RE Request error code
yl#### is the position of the liquid The last saved error messages can be
surface after the command in 0.1 mm retrieved. The error buffer is automatically
(values between 0000-2800). The values voided when a new command is started.
will not be updated if an error occurred.
id#### 0000..9999 0 Identification Number
aw###### is the calculated volume in
the container in 0.1 µL. REid####er## See Error Messages
Command Parameter Parameter Default Name and Description Command Parameter Parameter Default Name and Description
values value values value
RT Request Tip Presence Status RQ Request Instrument status
id#### 0000..9999 0 Identification Number id#### 0000..9999 0 Identification Number
RTid####rt# rt# is the tip presence status RQid####rq&&&& qq&&&& is the returned instrument status
(0 = no tip, 1 = tip present).
1st character
3rd character
RZ Request Absolute z Position
id#### 0000..9999 0 Identification Number Bit 3 Bit 2 Bit 1 Bit 0
RZid####gy#### gy#### is the returned height of the
z-drive in increments of 0.1 mm. 4th character
bg#### 0000..1800 0 Height of the Container Bottom GCid####er## See Error Messages
Measured from “ce” in 0.1mm.
Command Parameter Parameter Default Name and Description Command Parameter Parameter Default Name and Description
values value values value
GB Request Container Geometry GO Set Deck Geometry Parameters
Parameters id#### 0000..9999 0 Identification Number
id#### 0000..9999 0 Identification Number go## 00..99 0 Index of the Deck Geometry Table
ge## 00..99 0 Index of the Container Geometry Table te#### 0000..2800 0 End Traverse Position
GBid####ge##ca#cb###cc###cd###bg####gx#####ce####ch####ci#### Traverse Position at the end of
a command in increments of 0.1 mm.
This is also the traverse height at
which the x- and y-movements will be
performed.
tm#### 0000..1800 0 Beginning of Tip Picking Position
This is the z-position (in increments of
0.1 mm) at which the tip picking process
5.12.2 Deck Geometry Table
starts. A control is implemented to
detect whether a tip is already present
All parameters required to describe the geometry of the instrument deck (Figure 18) on which or not. This position should be set at
ZEUS is integrated are summarized in the Deck Geometry Table. The deck geometry table least 5 mm above the tip.
includes the traverse height, the tip pick up, and the discard height. Several Deck Geometry tr#### 0000..2800 0 Position of Tip Deposit Process
Tables may be defined within one application, for example if different Traverse Heights are This z-position (in increments of 0.1mm)
should be set to limit splashes of
required with or without tips, if different tubes heights are used or several tip storage positions. remaining liquid and cross
Each deck is described in its respective table, which is identified by the Index of the Deck contamination on your instrument.
Geometry Table. GOid####er## See Error Messages
Current
height
tm
5 mm
End of tip
picking
Disposable tip
te
tr
Waste
Figure 18: Description of the Deck Geometry parameters during tip picking and deposit.
Index (tt) Tip Description Tip Part Number 00 Water (Hamilton Verification Water) Jet empty 50 µL cLLD
00 10 µL CO-RE tip, conductive, non-filtered 235900-ZEUS 01 Water (Hamilton Verification Water) Jet empty 300 µL cLLD
01 10 µL CO-RE tip, conductive, filtered 235901-ZEUS 02 Water (Hamilton Verification Water) Jet empty 1000 µL cLLD
02 50 µL CO-RE tip, conductive, non-filtered 235966-ZEUS 03 Water (Hamilton Verification Water) Jet part 50 µL cLLD
03 50 µL CO-RE tip, conductive, filtered 235948-ZEUS 04 Water (Hamilton Verification Water) Jet part 300 µL cLLD
04 300 µL CO-RE tip, conductive, non-filtered 235902-ZEUS 05 Water (Hamilton Verification Water) Jet part 1000 µL cLLD
05 300 µL CO-RE tip, conductive, filtered 235903-ZEUS 06 Serum Jet empty 50 µL cLLD
06 1000 µL CO-RE tip, conductive, non-filtered 235904-ZEUS 07 Serum Jet empty 300 µL cLLD
07 1000 µL CO-RE tip, conductive, filtered 235905-ZEUS 08 Serum Jet empty 1000 µL cLLD
08 300 µL CO-RE slim tip, conductive, non-filtered 235806-ZEUS 09 Serum Jet part 50 µL cLLD
09 300 µL CO-RE slim tip, conductive, filtered 235647-ZEUS 10 Serum Jet part 300 µL cLLD
11 Serum Jet part 1000 µL cLLD
12 Ethanol (100%) Jet empty 50 µL pLLD
13 Ethanol (100%) Jet empty 300 µL pLLD
The choice of tip should be adjusted to the type of container (Figure 19). 14 Ethanol (100%) Jet empty 1000 µL pLLD
15 Glycerin (80%) Jet empty 50 µL cLLD
Container Diameter Example of Container Required x-y- Type of Tip 16 Glycerin (80%) Jet empty 300 µL cLLD
Positioning Accuracy 17 Glycerin (80%) Jet empty 1000 µL cLLD
d ≥ 5.5 mm 96-well microtiter plate 0.5 mm All
3.0 mm ≤ d ≤ 5.5 mm 384-well microtiter plate 0.5 mm 10, 50 and 300 µL tips
d ≤ 1.5 mm 1536-well microtiter plate 0.1 mm 10 µL tips
need to be teached
Command Command
Error Description Possible Action Error Description Possible Action
35 Voltage outside of the Make sure that the voltage is 48V+-5%. 55 Volume check failed ai > tip size
permitted range ai < 1st base point of the correction curve
36 Emergency stop is active or Switch off the emergency stop. ai > 8th base point of the correction curve
was sent during action. tip size < 8th base point of the correction curve
ma > tip size
40 Parallel processes not permited Send only one command at a time.
56 Conductivity check failed The choosen liquid class requires the use of
51 Pipetting drive not initialized Initialize the pipetting drive. conductive tips. Make sure that the liquid class and
52 Movement error on pipetting drive Hardware is not working. Please contact Hamilton. the tip type index parameters are consistent.
61 z-Drive is not initialized Initialize the z-drive. 57 Filter check failed Make sure that the liquid class and the tip type
index are consistent.
62 Movement error on the z-drive Hardware is not working. Please contact Hamilton.
61 z-Drive is not initialized Initialize the z-drive.
65 z-position not possible The z-drive could not reach te. Check parameters.
62 Movement error on the z-drive Hardware is not working. Please contact Hamilton.
69 z-position not possible The z-drive could not reach tr. Check parameters.
65 z-position not possible The z-drive could not reach te. Check parameters.
75 No tip present
66 z-position not possible The z-drive could not reach ce. Check parameters.
77 Tip not discarded
67 z-position not possible The z-drive could not reach cf. Check parameters.
68 z-position not possible The z-drive could not reach zp. Check parameters.
GA
70 Liquid level not detected The liquid level was not detected between zp and
00 No error during the command ce (pLLD and cLLD). Check paramaters.
30 Undefined command Command unknown. Refer to the command tables. 71 Not enough liquid present ZEUS calculated that the desired volume ai is
31 Undefined parameter Unknown parameter. Refer to the command tables. not available between ce and the found LLD position
with the defined container geometry. Check
32 Parameter out of range A value of a sent parameter is out of range. parameters.
Refer to the command tables.
72 Auto calibration of the Hardware is not working. Please contact Hamilton.
35 Voltage outside of the Make sure that the voltage is 48V+-5%. pressure sensor not possible
permitted range
74 Early liquid level detection The liquid level was there before reaching zp.
36 Emergency stop is active or Switch off the emergency stop. Check parameters.
was sent during action.
75 No tip picked up There is not tip on the cone. Pick up tip first.
38 Empty liquid class The liquid class defined in the parameter lq
was empty. Choose another one. 80 Clot detected during aspiration The pressure curve reached the clot-threshold.
Error handling to empty the tip may be defined.
40 Parallel processes not permited Send only one command at a time.
81 Empty tube detected The tube was empty or has been emptied during
51 Pipetting drive not initialized Initialize the pipetting drive. during aspiration aspiration.
52 Movement error on pipetting drive Hardware is not working. Please contact Hamilton. 82 Foam detected during aspiration Foam has been detected. Centrifuge tubes prior
53 Maximum volume of the There is not enough space in the tip for pipetting to avoid foam formation.
tip reached desired volume ai. Correct ai or change tip. 85 No communication to the Hardware is not working. Please contact Hamilton.
54 Maximum volume in pipetting digital potentiometer
drive reached
Command Command
Error Description Possible Action Error Description Possible Action
70 Liquid level not detected The liquid level was not detected between zp and
ce (pLLD and cLLD). Check paramaters.
GD
72 Auto calibration of the pressure Hardware is not working. Please contact Hamilton.
00 No error during the command
sensor not possible
30 Undefined command Command unknown. Refer to the command tables.
74 Early liquid level detection The liquid level was there before reaching zp.
31 Undefined parameter Unknown parameter. Refer to the command tables. Check parameters.
32 Parameter out of range A value of a sent parameter is out of range. 75 No tip picked up There is not tip on the cone. Pick up tip first.
Refer to the command tables.
83 Clot detected during dispensing The pressure curve reached the clot-threshold.
35 Voltage outside of the Make sure that the voltage is 48V+-5%. Error handling to empty the tip may be defined.
permitted range
84 Foam detected during Foam has been detected. Centrifuge tubes prior
36 Emergency stop is active or Switch off the emergency stop. dispensing pipetting to avoid foam formation.
was sent during action.
85 No communication to the Hardware is not working. Please contact Hamilton.
38 Empty liquid class The liquid class defined in the parameter lq digital potentiometer
was empty. Choose another one.
40 Parallel processes not permited Send only one command at a time.
GJ
51 Pipetting drive not initialized Initialize the pipetting drive.
00 No error during the command
52 Movement error on pipetting drive Hardware is not working. Please contact Hamilton.
30 Undefined command Command unknown. Refer to the command tables.
54 Maximum volume in pipetting
31 Undefined parameter Unknown parameter. Refer to the command tables.
drive reached
32 Parameter out of range A value of a sent parameter is out of range.
55 Volume check failed di < 1st base point of the correction curve
Refer to the command tables.
ai > 8tht base point of the correction curve
ma > tip size 35 Voltage outside of the Make sure that the voltage is 48V+-5%.
permitted range
57 Filter check failed Make sure that the liquid class and the tip type
index are consistent. 36 Emergency stop is active or Switch off the emergency stop.
was sent during action.
61 z-Drive is not initialized Initialize the z-drive.
38 Empty liquid class The liquid class defined in the paremeter lq
62 Movement error on the z-drive Hardware is not working. Please contact Hamilton.
was empty. Choose another one.
63 Container bottom search failed No resistance on the z-drive detected between
40 Parallel processes not permited Send only one command at a time.
ch and ce. Check parameters.
51 Pipetting drive not initialized Initialize the pipetting drive.
65 z-position not possible The z-drive could not reach te. Check parameters.
52 Movement error on pipetting drive Hardware is not working. Please contact Hamilton.
66 z-position not possible The z-drive could not reach ce. Check parameters.
56 Conductivity check failed The choosen liquid class requires the use of
67 z-position not possible The z-drive could not reach cf. Check parameters.
conductive tips. Make sure that the liquid class and
68 z-position not possible The z-drive could not reach zp. Check parameters. the tip type index parameters are consistent.
Command Command
Error Description Possible Action Error Description Possible Action
57 Filter check failed Make sure that the liquid class and the tip type
index are consistent.
GK
61 z-Drive is not initialized Initialize the z-drive.
00 No error during the command
62 Movement error on the z-drive Hardware is not working. Please contact Hamilton.
30 Undefined command Command unknown. Refer to the command tables.
65 z-position not possible The z-drive could not reach te. Check parameters.
31 Undefined parameter Unknown parameter. Refer to the command tables.
66 z-position not possible The z-drive could not reach ce. Check parameters.
32 Parameter out of range A value of a sent parameter is out of range.
67 z-position not possible The z-drive could not reach cf. Check parameters. Refer to the command tables.
68 z-position not possible The z-drive could not reach zp. Check parameters. 35 Voltage outside of the Make sure that the voltage is 48V+-5%.
70 Liquid level not detected The liquid level was not detected between zp and permitted range
ce (pLLD and cLLD). Check paramaters. 51 Pipetting drive not initialized Initialize the pipetting drive.
72 Auto calibration of the pressure Hardware is not working. Please contact Hamilton. 52 Movement error on pipetting drive Hardware is not working. Please contact Hamilton.
sensor not possible
54 Maximum volume in pipetting
74 Early liquid level detection The liquid level was there before reaching zp. drive reached
Check parameters.
85 No communication to the Hardware is not working. Please contact Hamilton.
GC
digital potentiometer
00 No error during the command
20 No communication to EEPROM Hardware is not working. Please contact Hamilton.
AB
30 Undefined command Command unknown. Refer to the command tables.
00 No error during the command
31 Undefined parameter Unknown parameter. Refer to the command tables.
30 Undefined command Command unknown. Refer to the command tables.
32 Parameter out of range A value of a sent parameter is out of range.
Refer to the command tables.
AW
00 No error during the command
GO
30 Undefined command Command unknown. Refer to the command tables.
00 No error during the command
20 No communication to EEPROM Hardware is not working. Please contact Hamilton.
XA
30 Undefined command Command unknown. Refer to the command tables.
00 No error during the command
31 Undefined parameter Unknown parameter. Refer to the command tables.
20 No communication to EEPROM Hardware is not working. Please contact Hamilton.
32 Parameter out of range A value of a sent parameter is out of range.
30 Undefined command Command unknown. Refer to the command tables.
Refer to the command tables.
31 Undefined parameter Unknown parameter. Refer to the command tables.
32 Parameter out of range A value of a sent parameter is out of range.
GB
Refer to the command tables.
00 No error during the command
20 No communication to EEPROM Hardware is not working. Please contact Hamilton.
Command Command
Error Description Possible Action Error Description Possible Action
30 Undefined command Command unknown. Refer to the command tables. 30 Undefined command Command unknown. Refer to the command tables.
31 Undefined parameter Unknown parameter. Refer to the command tables. 31 Undefined parameter Unknown parameter. Refer to the command tables.
32 Parameter out of range A value of a sent parameter is out of range. 32 Parameter out of range A value of a sent parameter is out of range.
Refer to the command tables. Refer to the command tables.
GR GS
00 No error during the command 00 No error during the command
20 No communication to EEPROM Hardware is not working. Please contact Hamilton. 20 No communication to EEPROM Hardware is not working. Please contact Hamilton.
30 Undefined command Command unknown. Refer to the command tables. 30 Undefined command Command unknown. Refer to the command tables.
31 Undefined parameter Unknown parameter. Refer to the command tables. 31 Undefined parameter Unknown parameter. Refer to the command tables.
32 Parameter out of range A value of a sent parameter is out of range. 32 Parameter out of range A value of a sent parameter is out of range.
Refer to the command tables. Refer to the command tables.
GL GV
00 No error during the command 00 No error during the command
20 No communication to EEPROM Hardware is not working. Please contact Hamilton. 20 No communication to EEPROM Hardware is not working. Please contact Hamilton.
30 Undefined command Command unknown. Refer to the command tables. 30 Undefined command Command unknown. Refer to the command tables.
31 Undefined parameter Unknown parameter. Refer to the command tables. 31 Undefined parameter Unknown parameter. Refer to the command tables.
32 Parameter out of range A value of a sent parameter is out of range. 32 Parameter out of range A value of a sent parameter is out of range.
Refer to the command tables. Refer to the command tables.
39 Liquid class write protected Index 00-20 can only be saved by Hamilton.
GW
GM 00 No error during the command
00 No error during the command 20 No communication to EEPROM Hardware is not working. Please contact Hamilton.
20 No communication to EEPROM Hardware is not working. Please contact Hamilton. 30 Undefined command Command unknown. Refer to the command tables.
30 Undefined command Command unknown. Refer to the command tables. 31 Undefined parameter Unknown parameter. Refer to the command tables.
31 Undefined parameter Unknown parameter. Refer to the command tables. 32 Parameter out of range A value of a sent parameter is out of range.
Refer to the command tables.
32 Parameter out of range A value of a sent parameter is out of range.
Refer to the command tables.
GG
GQ 00 No error during the command
00 No error during the command 20 No communication to EEPROM Hardware is not working. Please contact Hamilton.
20 No communication to EEPROM Hardware is not working. Please contact Hamilton.
DC Control Download
Calculate the check sums and verify
that the program has been correctly
loaded.
Command Parameter Parameter Default Name and Description Command Parameter Parameter Default Name and Description
values value values value
RQ Request Instrument status AP Switch to loaded Program
After successful download this command
id#### 0000..9999 0 Identification Number
switches to the new program. This
RQid####rq&&&& rq&&&& is the returned instrument status command is not acknowledged.
1st character
2nd character
5.14.5 Error Messages
1) Set the LC parameters for your specific application (either based on a predefined LC or by
creating a new set of parameters). The right choice of parameters ensures high precision. “Jet empty” dispense mode is used for dispensing all the liquid in a jet without touching the
2) Perform gravimetric calibration for the new LC. This step ensures high accuracy. surface to avoid cross-contamination. “Jet part” mode dispenses only a part of the liquid in a
jet. If “Surface empty” mode is active, the tip touches the surface and dispenses all the liquid.
3) Specify the associated settings for Qualitative Pipetting Monitoring (QPM). “Surface part” mode leaves a residual volume in the tip. Surface modes are recommended for
dispensing small volumes (<20µl).
NOTE: If LC parameters have been changed (step 1), then the gravimetric
calibration and QPM parameters have to be adjusted as well to ensure proper
pipetting (steps 2 and 3).
Next, the relevant background parameters for a pipetting cycle have to be set. All liquid class
parameters are described in the table below. To start with, choose a predefined liquid class Command Parameter Parameter Default Name and Description
which is similar to the liquid you would like to handle. After that, run a test and visually inspect range value
the quality of the pipetting process (e.g. using a high-speed camera). If the pipetting behavior is #### 0000..9999 0 Over-Aspirated Volume in 0.1 µL
fine, volume calibration can be done as described in the next section. Otherwise you might ad- After aspirating the required volume,
just some of the liquid class parameters to optimize the pipetting properties. Several examples this additional volume is aspirated and
of situations and possible solutions are listed below. dispensed again immediately to pre-wet
the tip (in simple mode). The effect is
that the meniscus is balanced. It is
Situation Possible solution used for accurate pipetting of small
volumes (<10 µL).
Unwanted air inside the tip after aspiration Reduce flow rate; increase settling time
Drop formation at the tip orifice Pre-wet tip by adding over-aspirate volume; ##### 00000..10000 0 Transport Volume in 0.1 µL
add transport volume; add stop back volume; Air is aspirated at the end of the
activate ADC aspiration or dispensing and is
automatically dispensed again as an
Liquid sticks to the outside of the tip Reduce swap speed
extra volume during the following
Residual liquid volume inside the tip Reduce flow rate; add blowout volume; dispense step. It is influenced by tip
after dispensing reduce stop flow rate geometry and liquid vapor pressure.
Liquid splashes during dispensing Reduce flow rate Prevents droplet formation during
transport. This parameter is helpful
when pipetting volatile substances.
##### 00000..10000 0 Blowout Air Volume in 0.1 µL
Command Parameter Parameter Default Name and Description This air volume is taken up first during
range value aspiration. It is blown out at the end
of dispensing (empty mode only). It is
GL Set Liquid Class Parameter influenced by liquid viscosity and vapor
id#### 0000..9999 0 Identification Number pressure. The blowout volume helps to
completely empty the tip.
lq## 00..99 0 Liquid Class Table Index
#### 0000..3000 0 Swap Speed in 0.1 mm/s
uu# 0..1 0 Liquid Class for Filter Tips?
Determines how fast the tip is
(0 = no, 1 = yes)
withdrawn from the liquid after
Aspiration aspiration or dispense. This parameter
# 0..2 0 Aspiration mode (0 = simple; is chosen based on the liquid viscosity
1 = empty cup; 2 = consecutive) and vapor pressure. Using a slow speed
prevents a viscous liquid from sticking
##### 00000..25000 0 Flow Rate in 0.1 µl/s to the outside of the tip.
This parameter corresponds to the
plunger speed during aspiration or
dispense. The flow rate is the most
critical parameter, influenced by tip
geometry, liquid viscosity, density and
vapor pressure:
small tip opening -> slow flow rate,
high viscosity -> slow flow rate,
high density -> slow flow rate,
high vapor pressure -> fast flow rate
Command Parameter Parameter Default Name and Description Command Parameter Parameter Default Name and Description
range value range value
### 000..999 0 Settling Time in 0.1 s ##### 00000..25000 0 Flow Rate in 0.1 µL/s
This parameter sets how long the tip Dispensing speed of the plunger. See
remains in liquid after aspiration or description of aspiration Flow Rate.
dispense before moving out of the ##### 00000..25000 0 Stop Flow Rate in 0.1 µL/s
liquid. The tip geometry and liquid Dispensing speed of the plunger at
viscosity are considered when setting which the dispense stops. If equal
this parameter. If settling time is too to “Flow Rate”, then the plunger stops
short the liquid has not enough time to abruptly without slowing down. If set to
fill the tip and unwanted air is aspirated minimum, the plunger movement slows
as soon as the tip is lifted out of the liquid. down gradually until it stops. When
# 0..1 0 Liquid Level Detection (LLD) mode pipetting a viscous liquid such as
(0 = cLLD, 1 = pLLD) glycerin, it is recommended to have a
Use pressure LLD (pLLD) with gradual stop (slow Stop Flow Rate) for
non-conductive liquids and tips. pLLD complete dispensing.
only works with new and empty tips ### 000..325 0 Stop Back Volume in 0.1 µL
during aspiration. For conductive This air volume is aspirated again
liquids, choose capacitive LLD (cLLD) in immediately after dispensing to have an
combination with conductive tips. abrupt cut-off of the liquid flow (jet part
# 1..4 1 cLLD Sensitivity (1 = very high, 2 = high, mode only). Prevents droplet formation
3 = medium, 4 = low) when pipetting volatile liquids such as
The sensitivity may be adjusted to the ethanol.
vessel size and the liquid volume to be ##### 00000..10000 0 Transport Volume in 0.1 µL. See
detected (small volume -> high sensitivity). description of aspiration Transport Volume.
# 1..4 1 pLLD Sensitivity (1 = very high, 2 = high, ### 000..200 0 Acceleration in 0.1 µL/s 2
3 = medium, 4 = low) Sets the acceleration of the plunger
Mainly determined by tip geometry. Use during dispensing. Required for
high sensitivity with large tip orifices. controlled flow cut-off. Prevents droplet
# 0..1 0 Anti-Droplet Control (ADC; 0 = off, 1 = on) formation at the tip orifice.
Use ADC when pipetting liquids having Recommended value is 40.
low viscosity and high vapor pressure, #### 0000..3000 0 Swap Speed in 0.1 mm/s. Surface
such as ethanol. If the pressure reaches dispensing modes only. See description
a critical value where dripping is likely of aspiration Swap Speed.
to occur, the plunger is withdrawn as
much as needed to reduce the pressure ### 000..999 0 Settling Time in 0.1 s. Surface
and to prevent droplet formation. dispensing modes only. See description
of aspiration Settling Time.
Dispensing
##### 00000..25000 0 Flow Rate Transport Volume in 0.1 µL/s
# 0..3 0 Dispensing mode Plunger speed during aspiration of
(0 = jet empty, 1 = jet part, transport volume. Should be
2 = surface empty, 3 = surface part) synchronized with the movement of
z-drive. Recommended value is 325.
Command Parameter Parameter Default Name and Description Command Parameter Parameter Default Name and Description
range value range value
GG Set Calibration Curve for Aspiration GH Set Calibration Curve for Dispensing
id#### 0000..9999 0 Identification Number id#### 0000..9999 0 Identification Number
gg## 00..99 0 Calibration Curve Table Index gh## 00..99 0 Calibration Curve Table Index
ck##### 00000..11000 0 1. Target Volume (in units of 0.1 µL) cl##### 00000..11000 0 1. Target Volume (in units of 0.1 µL)
##### 00000..11000 0 1. Average Aspirated Volume ##### 00000..11000 0 1. Average Dispensed Volume
(in units of 0.1 µL) (in units of 0.1 µL)
##### 00000..11000 0 2. Target Volume ##### 00000..11000 0 2. Target Volume
##### 00000..11000 0 2. Average Aspirated Volume ##### 00000..11000 0 2. Average Dispensed Volume
##### 00000..11000 0 3. Target Volume ##### 00000..11000 0 3. Target Volume
##### 00000..11000 0 3. Average Aspirated Volume ##### 00000..11000 0 3. Average Dispensed Volume
##### 00000..11000 0 4. Target Volume ##### 00000..11000 0 4. Target Volume
##### 00000..11000 0 4. Average Aspirated Volume ##### 00000..11000 0 4. Average Dispensed Volume
##### 00000..11000 0 5. Target Volume ##### 00000..11000 0 5. Target Volume
##### 00000..11000 0 5. Average Aspirated Volume ##### 00000..11000 0 5. Average Dispensed Volume
##### 00000..11000 0 6. Target Volume ##### 00000..11000 0 6. Target Volume
##### 00000..11000 0 6. Average Aspirated Volume ##### 00000..11000 0 6. Average Dispensed Volume
##### 00000..11000 0 7. Target Volume ##### 00000..11000 0 7. Target Volume
##### 00000..11000 0 7. Average Aspirated Volume ##### 00000..11000 0 7. Average Dispensed Volume
##### 00000..11000 0 8. Target Volume ##### 00000..11000 0 8. Target Volume
##### 00000..11000 0 8. Average Aspirated Volume ##### 00000..11000 0 8. Average Dispensed Volume
GGid####er## See error messages GHid####er## See error messages
Pressure
to separate these events by continuously monitoring the pressure. If the measured pressure
exceeds predefined threshold values, the system identifies an error. This section explains how
to set these thresholds. The commands and parameters for QPM are listed in the tables at the Clot Threshold
end of this section (aspiration and dispense). Specific thresholds have to be defined for each Tip Clogged Figure 24: Schematics of a typical pres-
of the 8 calibration points (see previous section). sure curve for successful aspiration (red)
and a clogged tip (blue). The threshold
Time for clot detection is set to a value larger
than pmax (e.g. 2pmax ).
6.4.1 Successful Aspiration and Dispensing
A successful pipetting cycle is sketched in Figure 23. The pressure in the tip has a certain value 6.4.2 Tip Clogging
at the beginning, and decreases below atmospheric when the plunger starts to move up. The
liquid enters the tip to compensate for the pressure difference. The pressure increases again Some samples can contain solid objects such as blood clots. If these are aspirated they might
when the plunger movement stops. However, the pressure at the end of the aspiration cycle (partially) clog the tip resulting in an inaccurate pipetting volume. In Figure 24, a successful aspi-
is lower than the initial value, because gravity acts on the liquid inside the tip. During dispense ration cycle (red) is compared to a clogged tip (blue). The pressure drops rapidly if a blood clot
the pressure increases above atmospheric while the plunger moves down. If such pressure is aspirated clogging the tip. The clot threshold (black dashed line) is set, for example, to twice
curve shape is measured, no error message is generated indicating a successful aspiration and the maximum pressure difference of a successful cycle 2pmax.
dispense. Deviations from this curve shape are recognized by the QPM and discussed in the
following for the aspiration cycle.
6.4.3 Aspiration of Air
Air can be aspirated if not enough liquid is present in the container, as illustrated in Figure 25
(empty tube, blue). If air enters the tip, the pressure increases and saturates at a certain level close
to the initial value. The empty tube threshold is set to half of the pressure difference between start
and end value of a successful cycle p/2. In the case of dispensing, the empty tube threshold is
not defined, because this situation is already recognized and handled during aspiration.
Pressure
Successful Successful
Aspiration Dispense
Empty Tube
Empty Tube
Time Pressure Threshold
Foam
B the pressure. # 0..1 0 3. Pressure measurement range
Command Parameter Parameter Default Name and Description Command Parameter Parameter Default Name and Description
range value range value
#### 00º00..1023 0000 6. Foam detection threshold #### 0000..1023 1023 2. Clot detection threshold
# 0..1 0 6. Pressure measurement range #### 0000..1023 0000 2. Foam detection threshold
#### 0000..1023 0000 7. Clot detection threshold # 0..1 0 2. Pressure measurement range
#### 0000..1023 0000 7. Empty tube detection threshold #### 0000..1023 1023 3. Clot detection threshold
#### 0000..1023 0000 7. Foam detection threshold #### 0000..1023 0000 3. Foam detection threshold
# 0..1 0 7. Pressure measurement range # 0..1 0 3. Pressure measurement range
#### 0000..1023 0000 8. Clot detection threshold #### 0000..1023 1023 4. Clot detection threshold
#### 0000..1023 0000 8. Empty tube detection threshold #### 0000..1023 0000 4. Foam detection threshold
#### 0000..1023 0000 8. Foam detection threshold # 0..1 0 4. Pressure measurement range
# 0..1 0 8. Pressure measurement range #### 0000..1023 1023 5. Clot detection threshold
#### 0000..1023 0000 0. Clot detection threshold. If the #### 0000..1023 0000 5. Foam detection threshold
pressure measurement range is
# 0..1 0 5. Pressure measurement range
switched, then the pressure curve
at the switching value is measured #### 0000..1023 1023 6. Clot detection threshold
twice, using both ranges. This is the #### 0000..1023 0000 6. Foam detection threshold
first threshold value set for the range “1”.
# 0..1 0 6. Pressure measurement range
#### 0000..1023 0000 0. Empty tube detection threshold
#### 0000..1023 1023 7. Clot detection threshold
#### 0000..1023 0000 0. Foam detection threshold
#### 0000..1023 0000 7. Foam detection threshold
GQid####er## See error messages
# 0..1 0 7. Pressure measurement range
#### 0000..1023 1023 8. Clot detection threshold
GS Request QPM Parameter for Aspiration
#### 0000..1023 0000 8. Foam detection threshold
id#### 0000..9999 0
# 0..1 0 8. Pressure measurement range
gv## 00..99 0
#### 0000..1023 1023 0. Clot detection threshold
GSid####gv##vv#### #### #### # #### #### #### # #### #### #### #
(cf. table for aspiration)
#### #### #### # #### #### #### # #### #### #### # #### #### #### #
#### #### #### # #### #### #### #### 0000..1023 0000 0. Foam detection threshold
GVid####er## See error messages
GV Set QPM Parameter for Dispensing
id#### 0..9999 0 Identification Number GW Request QPM Parameter for Dispensing
gp## 00..99 0 QPM Parameter Table Index id#### 0..9999 0
ww#### 0000..1023 1023 1. Clot detection threshold gp## 00..99 0
#### 0000..1023 0000 1. Foam detection threshold => GWid####gp##ww#### #### # #### #### # #### #### # #### #### # ####
#### # #### #### # #### #### # #### #### # #### ####
# 0..1 0 1. Pressure measurement range
(0=0-20 mbar, 1=0-200 mbar).
7 Maintenance 5) Verify that the tip ejection sleeve is clean. Change the tip ejection sleeve if this is not the case.
6) Clean the naked barrel of the ZEUS pipetting module using a lint-free cloth soaked with
Microlab™ Detergent & Disinfectant (Figure 27).
Periodic maintenance routines need to be run in order to ensure safe and reliable operation of
the ZEUS pipetting module.
ATTENTION! Take care that no liquid gets inside the pipetting barrel.
4) Remove the tip ejection sleeve. To do so loosen the fixation clip carefully, and strip down
the tip ejection sleeve (Figure 27).
Perform the daily maintenance at the beginning of each working day. The following tasks belong
to the daily maintenance:
1) Perform visual check of the pipetting module to identify spillage and damages.
Fixation clip 2) If the pipetting head needs to be cleaned, perform the cleaning procedure (see Chapter 7.1)
Barrel
Tip ejection
sleeve
NOTE: The Lifetime Counter indicates the lifetime of the plunger seal. It is mea-
sured as the distance run by the seal. It is recommended to change the plunger 8.2 Recommended Qualification Procedures
seal before the Lifetime Counter reaches 12’000’000’000. This value is equivalent to Qualification of the ZEUS module in your instrument may be required for:
1 million full stroke pipetting cycles (aspiration and dispensation).
• Factory testing after instrument production
• Installation and operation qualification at the customer site
1) Perform visual check of the pipetting channel to identify spillage and damages. • Instrument verification during yearly maintenance
• Functional validation after service
2) Perform the cleaning procedure (see Chapter 7.1). • The table below describes the recommended qualification procedures.
3) Change the tip ejection sleeve.
4) Change the CO-RE tip adaptor after 500’000 tip picking cycles.
5) Control the lifetime counter. Return the ZEUS module to Hamilton before it exceeds Installation Operation Performance
Qualification Qualification Qualification
12’000’000’000 for replacement of the plunger seal.
Incoming goods
n.a. n.a. n.a.
6) Perform installation and operation qualification (Chapter 8). inspection
Factory testing Recommended Recommended The instrument manufacturer carries
(see Chapter 8.3) (see Chapter 8.4) the responsibility for the particular
process. This exact procedure will
strongly depend on the specific
applications that ZEUS as a
component is used for. Hamilton
can offer support related to the
liquid handling.
Verify that the environmental conditions are within the specifications. The operation qualifi-
Installation Operation Performance cation has to be performed with a well-defined solution such as Hamilton Verification Solution
Qualification Qualification Qualification (Chapter 11).
Installation at Recommended Recommended The instrument manufacturer carries
customer site (see Chapter 8.3) (see Chapter 8.4) the responsibility for the particular 1) Install and calibrate the balance according to the manufacturer’s instructions.
process. This exact procedure will
2) Pick up the disposable tip (Chapter 5.6).
strongly depend on the specific
applications that ZEUS as a 3) Aspirate the test volume (Chapter 5.7).
component is used for. Hamilton
can offer support related to the 4) Read weight.
liquid handling.
Yearly Recommended Recommended The instrument manufacturer carries 5) Dispense in jet mode the test volume into a weighing vessel (Chapter 5.7).
Maintenance (see Chapter 8.3) (see Chapter 8.4) the responsibility for the particular
process. This exact procedure will 6) Measure the dispensed weight (stable balance value).
strongly depend on the specific
applications that ZEUS as a
7) Eject tip to waste (Chapter 5.6).
component is used for. Hamilton 8) Repeat steps 1) – 6) ten times per volume and channel. Use new tip for each pipetting.
can offer support related to the
liquid handling. 9) Calculate the average volumes using the corresponding liquid density (Chapter 6.3).
Service Recommended Recommended The instrument manufacturer carries
(see Chapter 8.3) (see Chapter 8.4) the responsibility for the particular 10) Calculate the accuracy R and precision CV (Chapter 6.3) and compare with field
process. This exact procedure will verification specifications (Figure 28). The verification is passed if the obtained values
strongly depend on the specific are lower than the specified. The field verification specifications are broader than the design
applications that ZEUS as a
component is used for. Hamilton specifications. They allow the use of a transportable balance and measurements under
can offer support related to liquid broader environmental conditions, typically found in laboratories.
handling.
9 Troubleshooting Example:
Example: ATTENTION! Hamilton reserves the right to refuse a return shipment of any
Hamilton product that could be hazardous to Hamilton employees.
12 Appendices
11.2 Disposable CO-RE Tips* 12.1 Appendix A: ASCII Chart
Hamilton Americas & Pacific Rim Hamilton Europe, Asia & Africa
4970 Energy Way Via Crusch 8
Reno, Nevada 89502 USA CH-7402 Bonaduz, Switzerland
Tel: +1-775-858-3000 Tel: +41-58-610-10-10
Fax: +1-775-856-7259 Fax: +41-58-610-00-10
Web: www.hamiltoncompany.com sales@hamiltoncompany.com contact.pa.ch@hamilton.ch
USA: 800-648-5950
Europe: +41-58-610-10-10 To find a representative in your area, please visit www.hamiltoncompany.com.