You are on page 1of 8

This article has been accepted for publication in a future issue of this journal, but has not been

fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JSEN.2018.2883716, IEEE Sensors
Journal

Force-Sensing-Based Unobtrusive System for


Awakening and Respiration Rate Analysis during
Sleep
Sooji Park, Student Member, IEEE, Hyong-Ho Cho, Hyung Chae Yang, Jin-Sang Yoon and Hangsik Shin,
Member, IEEE

Abstract — In this study, we designed an unobtrusive strip-type sensors attached to them, and it is only possible to perform PSG
sleep monitoring system including a force sensor and algorithms sleep monitoring in a designated laboratory, as opposed to a
for the detection of awakening and respiration rate. The developed normal sleeping environment. In addition, the monitoring pro-
system is installed on a mattress across the user's chest in order to
cess is expensive, requires a skilled technician, and is limited to
measure the trunk movement induced by respiration or body mo-
tion. Thirty subjects’ clinical data, which contain signals that were one-time monitoring. Considering that sleep is a repetitive ac-
measured simultaneously using the developed system and a poly- tivity that occurs daily, PSG is difficult to apply to daily sleep
somnograph (PSG), are used to validate the accuracy of the devel- health management.
oped system. The awakening detection, accuracy, sensitivity, spec- There have been several attempts to simplify PSG. Wrist
ificity, and Kappa coefficient of the designed system were 79.4%, actigraphy is a representative alternative for daily sleep moni-
48.3%, 86.9%, and 0.6, respectively, and the average sleep effi-
toring [1-3]. It measures the amount of activity that occurs in
ciency was 82.3% in the developed system and 84.0% in the PSG
with root-mean-square- error of 17.7%. In terms of respiration one’s wrist during sleep and uses this to analyze the person’s
rate detection, it was shown that the respiration rate measured sleep patterns and awakening during sleep. Among the wrist-
with the developed system can trace the PSG-measured respira- type sleep monitoring devices, Actiwatch (Philips Respironics,
tion rate. The accuracy to detect respiration rates within one, two, Oregon, USA) has been approved by the US Food and Drug
and three breaths per minute error were 76.4%, 97.2%, and Administration (FDA) and is used to clinically analyze the sleep
99.4%, respectively.
patterns of insomnia patients [4]. Wearable devices can monitor
Index Terms — Awakening state detection, PVDF (polyvinyli-
dene fluoride), Respiration rate detection, Sleep monitoring, Un- sleep status more easily than PSG. Therefore, they are much
obtrusive measuremnt more suitable for sleep health management in daily life; how-
ever, they constrain a user’s body because they have to be worn
I. INTRODUCTION at all times during sleep. It is also difficult to use for continuous
monitoring because it must be actively managed by the user.
P olysomnography (PSG) is the gold standard sleep monitor-
ing method that is used clinically to diagnose sleep disor-
ders. While PSG has the advantage of being able to accu-
In recent years, studies have been conducted for both remote
and unobtrusive measurement methods to improve conven-
rately diagnose sleep disease, it also has several limitations: it tional constrained-type sleep monitoring devices. Remote
is necessary for the subject to sleep with a large number of measurement methods monitor the human body without the
need for contact. The methods typically used to remotely meas-
ure body signals include radiometry based on radar signals [5],
Submit Date :
* This research was supported by the MSIP(Ministry of Science, ICT and
microphones measuring sounds generated by bio-activity or
Future Planning), Korea, under the ITRC(Information Technology Research snoring during sleep [6, 7], thermal cameras for measuring tem-
Center) support program (IITP-2018-2014-1-00720) supervised by the perature changes of the face during sleep [8], fiber optics for
IITP(Institute for Information & communications Technology Promotion) and
by the Bio & Medical Technology Development Program of the NRF funded
measuring body movement monitoring [9], millimeter wave for
by the Korean government, MSIP (NRF-2016M3A9F1941328). heart rate and respiration rate detection [10], infrared camera
S. Park is with the Department of Biomedical Engineering, Chonnam Na- and motion sensor for sleep monitoring [11], ultrasonic piezo
tional University, 50 Daehak-ro, Yeosu, South Korea (e-mail: transducer for capturing ultrasound reflections from heart and
susie.soo718@gmail.com)
H.-H. Cho is with the Department of Otolaryngology-Head and Neck Sur- surrounding organ motions due to the breathing [12], and Dop-
gery, Chonnam National University Medical School and Chonnam National pler rader for respiration monitoring [13]. Remote methods
University Hospital, Gwangju, South Korea. (e-mail: victocho@jnu.ac.kr) have the advantage of the user being free from the measuring
H. C. Yang is with Department of Otolaryngology-Head and Neck Surgery,
Chonnam National University Medical School and Chonnam National Univer- device. However, they are limited in that the results can be eas-
sity Hospital, Gwangju, South Korea. (e-mail: dr.yang@hanmail.net) ily distorted by exceptional environmental changes, such as ra-
J.-S. Yoon is with the Department of Psychiatry, Chonnam National Univer- dio signals, ambient noise, room temperature changes, and so
sity Medical School, Gwangju, South Korea (e-mail: jsyoon@jnu.ac.kr)
H. Shin is with the Department of Biomedical Engineering, Chonnam Na- on.
tional University, 50 Daehak-ro, Yeosu, South Korea (e-mail: hang- The unobtrusive method measures physiological signals with
sik.shin@jnu.ac.kr).

1558-1748 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JSEN.2018.2883716, IEEE Sensors
Journal

minimal contact and without needing to attach the sensor di-


rectly to the body. In representative unobtrusive sleep monitor-
ing studies, sleep efficiency has been derived based on measur-
ing movement using a piezoelectric sensor under the subject’s
back [14]; a load cell on a bed leg [15]; and a pressure sensor
and an air mattress [16]. These unobtrusive methods can meas-
ure biological signals without interfering with the subject’s 
sleep. Moreover, they are advantageous in that continuous and
Fig. 1. Developed strip-type force sensor and the pattern of electrode
long-term monitoring can be performed after the initial instal-
lation of the monitoring devices.
The purpose of this study is to develop an unobtrusive sleep
health management system that overcomes the limitations of
typical sleep monitoring devices, such as the burden of exami-
nation, constraining the subject’s body, and requiring user man-
agement, and to develop algorithms for distinguishing awaken-
ing and for detecting respiration rates, which is an important
indication in sleep health management. Awakening detection is
the basis for calculating sleep efficiency, and respiration is one
of the most important physiological activities, so it should be
monitored at every moment in daily life. Moreover, respiration
rate analysis can be a first step toward analyzing representative
sleep disorders, such as sleep apnea or snoring caused by ab-
normal breathing during sleep.
There are several unobtrusive onbed type sleep tracking de- 
vices that have different structures and schemes. Representative Fig. 2. The sensor interface for analog front-end. (a) interface’s side structure,
examples include Beddit(Apple inc., CA, USA), Nokia (b) interface’s top structure, and (c) appearance of sensor interface
Sleep(Nokia, Espoo, Finland), and EarlySense(EarlySense inc.,
MA, USA). However, they have different structures and are thickness to prevent the user from recognizing the presence of
constantly being updated, so optimization and clinical valida- the sensor and to allow the user's chest to be positioned on the
tion are still required. In the process of this development, our sensor even when the user changes their sleeping position. In
research has meaning as an another attempt involving its own order to measure the electrical response of the PVDF film, elec-
structure and algorithm. In particular, clinical studies on the trodes were screen-printed using silver paste (Silveray-JM,
onbed type devices included in our study can provide useful im- Exax inc., Cheonan, South Korea) on both sides of the film,
plications, because clinically validated results are hard to find then heat curing the film for 30 minutes at a 60.The pattern
in other related studies. of the electrodes was staggered so as to prevent the generation
of electromagnetic (EM) noise from electrostatic induction
II. UNOBTRUSIVE FORCE SENSING SYSTEM when the bilateral electrodes were superposed.
A. Strip-type force sensor The fabricated line force sensor has a size of 12 mm × 720
mm × 0.13 mm (width × length × thickness), and a under 4 ms
We developed a flexible ultra-thin sensor using Polyvinyli- response time. Two rows of line sensors were packed in parallel
dene fluoride (PVDF) film (Fils inc., Sungnam, South Korea) with a 0.05 mm thick adhesive polyethylene terephthalate (PET)
to measure the force generated from body movement. PVDF film, except for 20 mm from the edge for interfacing. Electro-
film has previously been used for unobtrusive monitoring be- magnetic field shielding fabrics(WPD-330-GG, Ajin electron,
cause it is thin and flexible and generates electricity in response Busan, South Korea) was used as a cover for the sensor in order
to mechanical stress [17]. Table I shows the general properties to reduce external EM noise. The developed two-channel sensor
of the PVDF film [18]. size was ultimately 40 mm × 750 mm × 0.25 mm (width ×
The developed strip-type sensor has sufficient width and length × thickness) with a 10 mm gap between line sensors. The
TABLE I interface between the sensor and the system was fabricated us-
PROPERTIES OF PVDF FILM ing a ring terminal, an eyelet, and insulating tape for robust fix-
Mechanical property Unit ation. Fig. 1 and Fig. 2 show the developed two-channel strip-
Tensile strength (@Break) 35-50 MPa type force sensor and the pattern of the electrode, as well as the
Elongation (@Break) 100-250 % sensor interface for the analog front end, respectively.
Tensile modulus 2500 MPa
Electrical property Unit
B. Analog front-end
Dielectric strength, 1 mil 160 kV/mm An analog front-end was designed to preprocess the signal
Dielectric constant, 1 kHz 8.2-10 - measured with the strip-type force sensor, which includes a
Dissipation factor, 1 kHz 0.005-0.02 -

1558-1748 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JSEN.2018.2883716, IEEE Sensors
Journal

    
     TABLE II

 SPECIFICATION OF THE DEVELOPED DATA ACQUISITION MODULE
Property Description Unit


 Number of ADC channel 24 -
ADC bit depth 14 bit
 ADC range 0-1.45 V

ADC resolution (about) 88.5 

 kHz/Chan-
Maximum sampling frequency 1
nel

 Connectivity WiFi, Bluetooth 4.0 -


Fig. 3. Analog front-end circuit of a single channel

charge-amplifier and a voltage amplifier circuit (Gain = 11). A


reference voltage-generating circuit was added to measure a
negative signal (Vref = 670 mV). Fig. 3 shows the analog front-
end circuit. Because the electrical specifications of the
fabricated sensor and the charge amp’s gain are not obvious, we
set the value of the resistor or the capacitor empirically.
C. Data acquisition module
The data acquisition module was designed to receive up to
24-channel analog signals, and it consists of a micro-controller Fig. 4. Internal block diagram of data acquisition module
unit (MCU), a Bluetooth module, a microSD card slot, a real-
time clock (RTC), and an internal battery. MSP432P401R
(Texas Instruments, Dallas, USA), which supports low-power
operation and a 14-bit analog-to-digital converter (ADC), was
used for data conversion and transmission. The ADC range of
the data acquisition module is 0 to 1.45 V with 88.5 uV resolu-
tion, and the sampling frequency is 1 kHz. The measured data
is transmitted through a serial communication method via USB
interface or ESP-32s (AI-Thinker Co. Ltd, Shenzhen, China), a
Bluetooth and WiFi integration module. An analog front-end
and data acquisition module was designed as an assembly type 
for possible future expansion. Table II presents the specification Fig. 5. Unobtrusive force measurement system. (a) Overall system, (b) Assem-
of the developed data acquisition module. Fig. 4 and Fig. 5 bly of analog front-end circuit, data acquisition module, and case, (c) PCB of
analog front-end circuit and data acquisition module.
show an internal block diagram of the data acquisition module
and the developed unobtrusive force measurement system, re- and a part of awakening detection precedes the respiration rate
spectively. detection. The overall algorithm flow is shown in Fig. 6.
A. Awakening detection
III. ALGORITHM DEVELOPMENT When some movement occurs during sleep, a large pressure
This research consists of multiple algorithms to analyze awak- change is induced between the human body and the force meas-
ening and respiration rate. These algorithms are interconnected, urement system. In this study, abrupt changes in the amplitude
of the measured signal are regarded as movement during sleep

Fig. 6. Algorithms for awakening and respiration rate detection using movement signals.

1558-1748 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JSEN.2018.2883716, IEEE Sensors
Journal


 Fig. 8. The process of threshold decision for awakening state detection

Fig. 7. An example of preprocessing to detect motion signal. (a) Original signal


by the strip-type force sensor signal, (b) Signal after applying the IIR notch fil- awakening (, 1 ) if the result was greater than or
ter, and (c) Signal after applying IIR BPF equal to 0.5. Otherwise, it was classified as sleep (, 1
). This can be expressed as (2), and
, is the total number of
and are classified as awakening. The awakening detection algo- instances of (, in a 30-second period.
rithm consists of preprocessing, threshold determination, and

awakening detection stages. ! ,4(, 5 3  45
Preprocessing enhances the movement signal from the meas- , 1 7 /.0 (2)
! ,4(, 5 2  4 5
ured signals. In the preprocessing stage, a second-order IIR /.0
notch filter is applied so as to remove 60 Hz powerline interfer-
ence. Next, a third-order IIR band pass filter (BPF) with a pass-
B. Respiratory rate detection
band of 0.4 to 1 Hz is applied in order to enhance the movement
component. Fig. 7 shows an example preprocessing result. Noise removal precedes extracting the respiratory component
In order to determine the proper decision criteria for awaken- from the force signal using the proposed system or the ab-
ing, a histogram of the standard deviation of the measured sig- dominal movement signal using the PSG equipment. Except for
nal is used. (see Fig. 8). First, we repeatedly calculate the stand- the fact that the proposed system uses a second-order IIR notch
ard deviation of a seven-second time duration with overlapping filter to remove the 60 Hz powerline noise, all procedures are
( ,&%)$' ). We then draw the histogram of ,&%)$' and calcu- the same as the signals obtained by the developed system and
late a Gaussian curve using bins less than the maximum value the abdominal respiratory movement signal of PSG; excluding
of histogram. Then, the 5σ on the positive side of the histogram the awakening section ((, 1 ) which includes a mo-
is determined as a criterion for awakening state classification. tion artifact; interpolating the removed interval with neighbor
In this process, we empirically selected a window length of data; applying third-order IIR BPF with a pass band of 0.1 to
seven seconds for calculating standard deviation and a 0.5 Hz to extract the respiration component; and applying a
coefficient threshold of 5σ. moving average filter (MAF) of 2 seconds length for smoothing.
The awakening is determined in two steps. (see Fig. 6) In the Spectrograms are useful for analyzing frequency components
first step, if the ,&%)$' is greater than or equal to the threshold, over time. In this study, we used a spectrogram to find the res-
piration rate for the force signal of the developed system and
it is classified as awakening ((, 1 ); in contrast, if the
the abdomen movement of PSG. We assume that the respiratory
value is less than the threshold, it is classified as sleep
frequency component is the most prevalent in the filtered sig-
((, 1 ). This can be expressed as (1)
nals, and that the maximum value of the power spectral density
 ,&%)$' 3  45 could reflect the respiration rate. In spectrogram analysis, we
(, 1  6 (1) set a window size of 120 seconds and an overlap of 60 seconds.
,&%)$' 2  4 5
Next, we find the frequency that has the maximum power spec-
tral density for each window, and then calculate the respiration
The notation, ‘5ms’, of the awakening state means that values rate per minute.
were sampled every 5 ms because force signals were recorded
with 1 kHz sampling frequency, but it was down-sampled to IV. ALGORITHM EVALUATION
200 Hz to match the sampling frequency of the PSG equipment.
In the second step, we determine awakening every 30 seconds. A. Dataset
The 30-second period is the reference epoch of PSG’s sleep di- An experiment was conducted in order to evaluate the perfor-
agnosis. By averaging the previously obtained awakening mance of the measurement system in comparison with PSG.
((, 1 ) of 30-second duration, it was classified as The experiment was conducted at Chonnam National

1558-1748 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JSEN.2018.2883716, IEEE Sensors
Journal

TABLE III
SUBJECT’S INFORMATION
Gender Age BMI Total Sleep
No. (M/F) (years) (kg/m2) Time* Diagnosis
(minutes)
1 M 38 22.9 427 IS, SN
2 M 47 25.0 484 OSA, SN
3 M 36 28.7 551 OSA
4 M 51 27.5 424 OSA, SN, PLMS
5 M 51 23.8 493 IS
6 M 22 24.5 464 SN
7 M 23 36.8 499 OSA, SN
8 M 51 24.5 467 PLMS 
9 M 52 26.1 442 OSA, SN
Fig. 9. Experimental setup of an unobtrusive force measurement system
10 M 57 23.4 505 IS, SN, PLMS
11 F 56 19.5 497 IS, SN, PLMS (SpO2), chest and abdominal respiratory movement signal, res-
12 M 64 26.6 482 OSA, SN
piratory air flow, sound of snoring, and the position of the body
13 M 31 26.3 514 IS
14 M 55 26.0 505 IS
were all recorded using PSG equipment, Embla N7000 (Embla
15 F 65 27.8 413 IS, OSA
systems, Ontario, Canada) [19].
16 M 63 24.3 511 OSA, SN The developed system was installed between the mattress and
17 M 63 29.7 529 OSA, SN, PLMS the mattress cover, and attached with adhesive tape. It was laid
18 M 74 24.9 292 OSA, SN, PLMS across the subject's chest in order to measure the force changes
19 M 21 22.0 519 OSA, SN caused by the subject's motion during sleep. The signal was
20 M 21 24.7 481 Normal measured at a sampling frequency of 1 kHz. Fig. 9 shows the
21 M 23 30.0 538 OSA, SN experimental setup of the developed system.
22 M 50 24.1 496 OSA, SN
23 M 62 23.0 424 OSA, SN B. Preprocessing
24 M 49 29.4 535 OSA, SN Although the measurements using the developed system and
25 M 52 19.8 517 OSA, SN PSG were performed simultaneously, a timing difference be-
26 M 63 22.5 379 OSA, SN tween the two different measuring instruments is inevitable.
27 F 54 27.2 447 IS, OSA, SN Therefore, post-synchronization is required for accurate evalu-
28 M 54 31.2 483 OSA, SN ation. Cross correlation is used to analyze how similar signals
29 M 51 25.1 493 OSA are over time. In order to compensate for timing difference, we
30 M 24 19.0 494 Normal
find the time lag that maximizes the correlation coefficient, then
SN (N=21), OSA adjust the timing with that time lag. If time synchronization can-
Mean M 27 47.0 25.8 476.8 (N=20), IS (N=8)
(SD) F3 (16.0) (3.8) (53.6) PLMS (N=6), nor- not be achieved through the above method, the time difference
mal (N=2)
is then corrected manually.
*
Total Sleep Time is defined as a time between light-off ans light-on In analysis, only the data between the 'light off', which is the
IS: Insomnia, SN: Snoring, BMI: body mass index, OSA: Obstructive sleep
apnea syndrome, PLMS: Periotic limb movement in sleep, SD: Standard de-
start point of the analysis of PSG, and 'light on' at the end point
viation was used. In addition, the signal measured by the proposed sys-
University Hospital sleep testing laboratory and approved by tem used only the second channel data, having excellent meas-
the Institutional Review Board (IRB) of Chonnam National urement results of the two channels. Matlab R2016a (Math-
University biomedical research ethics committee. (IRB No. works, Inc., Natick, USA) was used for signal preprocessing,
CNUH-2017-274) algorithm development, and evaluation.
As an experiment, diagnostic overnight PSG was performed, C. Awakening detection
and a total of 39 subjects participated. Ultimately, data from 30
The sleep stage, which is one of the results of PSG, was di-
participants were used; nine subjects were excluded because
vided into WAKE, NREM (stage 1, stage 2, stage 3, stage 4),
their data was not properly acquired due to operational error. In
and REM based on the characteristics of EEG and EOG [19].
evaluating the respiration rate detection, only data from 26 sub-
In this study, we divided sleep state into two classes; ‘awaken-
jects were used because four subjects whose reference signals
ing’ and ’sleep’. A WAKE sleep stage was considered an
were not measured were excluded.
‘awakening’ state, while NREM and REM were considered to
The subjects included patients who had sleep disorders, such
correspond to a ‘sleep’ state.
as obstructive sleep apnea (OSA), snoring (SN), periodic limb
In order to evaluate the awakening detection algorithm, we
movement during sleep (PLMS), or insomnia (IS). All subjects
evaluated its accuracy, sensitivity, specificity, Cohen's Kappa
who participated in the experiment were provided with suffi-
coefficient, and sleep efficiency in comparison to PSG. In this
cient information regarding the purpose and method of the
evaluation, 'accuracy' refers to the probability of classifying a
study and informed consent was obtained from all subjects. Ta-
awakening state in PSG as an awakening state in the proposed
ble III shows a summary of the subjects’ information.
system, and classifying a sleep state in PSG as a sleep state in
PSG was performed according to the reference sleep test man-
the proposed system; 'sensitivity' refers to the probability of a
ual. Electrocardiogram (ECG), electrooculogram (EOG), chin
result being classified as an awakening state in the proposed
electromyogram (EMG), saturation of percutaneous oxygen

1558-1748 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JSEN.2018.2883716, IEEE Sensors
Journal

system among the results of classified as awakening state in TABLE V


RESPIRATION RATE DETECTION ACCURACY OF THE UNOBTRUSIVE FORCE
PSG; 'specificity' refers to the probability of a result being clas- MEASUREMENT SYSTEM
sified as sleep state in the proposed system among the results of
Detection error Accuracy (%)
those classified as sleep state in PSG; ‘Cohen's Kappa coeffi-
≤ 1 bpm 76.4
cient’ indicates the degree of agreement between the two test
≤ 2 bpm 97.2
results. It has a value between 0 and 1; 0 indicates no agreement,
≤ 3 bpm 99.4
0–0.20 slight agreement, 0.21–0.40 fair agreement, 0.41–0.60
moderate agreement, 0.61–0.80 substantial agreement, and
0.81–1.00 near perfect agreement [20]. The 'sleep efficiency' is
defined as the ratio of sleep time to total recording time.
The accuracy, sensitivity, specificity, and Kappa coefficients
of the awakening state detection algorithm were 79.4%, 48.3%,
86.9%, and 0.6, respectively. In addition, the average sleep ef-
ficiency values were 82.3% and 84.0% in the developed system
and PSG, respectively, and the mean absolute error and root-
mean-square error were 12.9% and 17.7%, respectively. Table
IV shows the evaluation results of awakening state detection by
subjects.
D. Respiration rate detection
We used abdominal respiratory movement signal, which is one Fig. 10. An example of respiration rate estimation process based on maximum
of the measurands of PSG, as the reference breathing signal. Fig. frequency of spectrogram. (a) Spectrogram and maximum frequency graph by
10 and Fig. 11 respectively represent an example of the respira- the unobtrusive force measurement system, (b) Spectrogram and maximum
frequency graph by the PSG’s abdominal respiratory movement signal
tion rate estimation process based on maximum frequency of
spectrogram, and an example of respiration rate measured by
the developed system and by PSG. In this study, Bland-Altman
analysis was performed between the reference respiration rate
of PSG and the detection error of the estimated respiration rate.
The respiration rate per minute was used as the average respi-
ration rate of the 26 subjects over two minutes. Fig. 12 shows a
Bland-Altman plot indicating the difference against the respira-
tion rate of the developed system and the PSG-measured respi-
ration rate. Table V shows the respiration rate detection accu-
racy of the unobtrusive force measurement system. As can be
seen in the table, 76.4%, 97.2%, and 99.4% of the respiratory
rates per minute were less than one, two, and three breaths per Fig. 11. An example of respiration rate measured by developed system and by
minute (bpm), respectively. PSG. #-+&* is the mean respiration rate of the proposed system, "# is the
mean respiration rate of PSG, and bpm is breaths per minute
V. DISCUSSION
In this study, we developed an unobtrusive sleep health man-
agement system that can overcome the limitations of existing
sleep monitoring devices, such as being physically constraining
and the requirement of user management. In addition,
TABLE IV
PERFORMANCE COMPARISON WITH PREVIOUS METHOD
Mean (SD) (%)
Type of Sen- Subjects
Author Awakening Detection Sleep Efficiency
sor Ndisorder/N
Ac Se Sp proposed PSG
this 79.4 48.3 86.9 82.3 84.0
PVDF film 28/30
study* (11.0) (21.9) (12.7) (13.2) (10.8)
Kushida 77 48 92 72.8 68.4
Actigraphy 100/100a
[2] (11.0) (17.5) (6.9) (15.9) (18.6)
92.0
Actigraphyb - - -
(7.0)
Weiss 92.0 87.0
Actigraphyc 0/30 - - -
[3] (7.0) (10.0) Fig. 12. Bland-Altman plot showing the difference against the respiration rate
94.0 of the developed system and PSG-measured respiration rate with limits of
Actigraphyd - - -
(6.0) agreement (dashed line). #-+&* is the mean respiration rate of the proposed
*Cohen’s Kappa value of awakening detection was 0.6 and mean absolute error and system, "# is the mean respiration rate of PSG, SD is standard deviation,
root-mean-square error were 12.9 ± 11.0% and 17.7% in sleep effiency, respectively.
and bpm is breathd per minute
SD: standard deviation, N: number of subject, Ndisorder: number of sleep disorder pa-
tient, Ac: accuracy, Se: sensitivity, Sp: specificity, SE: sleep efficiency, aOSA,
PLMS, insomnia, other, bSleepwatch (Ambulatory Monitoring, Inc., Ardsley, NY),
c
Actiwatch (Respironics, Pittsburgh, PA), dActical (Respironics, Pittsburgh, PA)

1558-1748 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JSEN.2018.2883716, IEEE Sensors
Journal

algorithms for awakening state detection and respiration rate distinguishable with a performance similar to those of existing
detection, which are important indicators in sleep health man- motion-based sleep monitoring devices. This will overcome the
agement, have been developed. disadvantages of wearable devices that require direct wearing
The performance of the awakening state detection algorithm of the sensor during sleep, and it will enable periodic and long-
of the proposed system can be compared with the results of pre- term sleep monitoring in an easy and convenient way. However,
vious studies that have focused on actigraphy, which is a repre- in order to apply it to everyday life, optimization processes and
sentative motion-based sleep monitoring method (see Table IV). furthur validations are needed through various cases of clinical
Kushida et al. developed an actigraphy-based awakening state trials, including sleep disorders or simple testing in daily living.
detection algorithm and compared it with PSG across 100 pa- Furthermore, there is a need to expand the parameters for mon-
tients with sleep disorders [2]. Compared to the results of pre- itoring sleep health, such as sleep apnea detection and others.
vious studies, the average accuracy and average sensitivity of
the proposed system are higher, but the average specificity is ACKNOWLEDGMENT
lower. This means that the proposed system achieves perfor- We would like to thank to Ji Yhea Kim and Han Byeon of
mance comparable to that of the actigraphy, but that the mis- Clinical Trial Center, Chonnam National University Hospital,
classification rate of sleep states as awakening states is higher. Geun Ho Lee and Youyang Jin of the sleep testing laboratory,
In addition, the Kappa statistics showed that the awakening Chonnam National University Hospital for assistance in data
state detection results of the proposed system show significant collection and scoring polysomnography.
similarities with the PSG results. The reason for the low speci-
ficity in both the proposed system and Kushida’s study is that REFERENCES
both studies distinguish the awakening state based on move-
[1] A. Kosmadopoulos, C. Sargent, D. Darwent, X. Zhou, and G. D. Roach,
ment during sleep, while PSG distinguishes the awakening state "Alternatives to polysomnography (PSG): a validation of wrist actigraphy
based on EEG and EOG. In other words, in motion-based re- and a partial-PSG system," (in eng), Behav Res Methods, vol. 46, no. 4, pp.
search, sleep is inevitably classified as a case of awakening but 1032-41, Dec 2014.
with less motion. [2] C. A. Kushida, A. Chang, C. Gadkary, C. Guilleminault, O. Carrillo, and
W. C. Dement, "Comparison of actigraphic, polysomnographic, and
In addition, sleep efficiency was overestimated based on the subjective assessment of sleep parameters in sleep-disordered patients," (in
wrist-type actigraphy used by Kushida [2] and Weiss [3], but it eng), Sleep Med, vol. 2, no. 5, pp. 389-96, Sep 2001.
tended to be underestimated in this study. However, it should [3] A. R. Weiss, N. L. Johnson, N. A. Berger, and S. Redline, "Validity of
be considered that variations in the number of subjects, the pres- activity-based devices to estimate sleep," (in eng), J Clin Sleep Med, vol.
6, no. 4, pp. 336-42, Aug 15 2010.
ence of sleep disorders, and sleep efficiency among the three [4] P. Respironics. (Jul 5). Actiwatch 2. Available:
studies can affect the outcome. http://www.actigraphy.com/solutions/actiwatch/actiwatch2.html
The result of using the proposed system to detect the respira- [5] E. M. Staderini, "UWB radars in medicine," IEEE Aerospace and
tion rate during sleep were confirmed to follow the same trend Electronic Systems Magazine, vol. 17, no. 1, pp. 13-18, 2002.
[6] H. Nakano et al., "Monitoring Sound To Quantify Snoring and Sleep
as the respiration rates measured by PSG. The Bland-Altman Apnea Severity Using a Smartphone: Proof of Concept," Journal of
analysis showed that most of the respiration rates detected by Clinical Sleep Medicine : JCSM : Official Publication of the American
the proposed system were within a 95% confidence interval, but Academy of Sleep Medicine, vol. 10, no. 1, pp. 73-78, 2014.
that the error tended to increase linearly with increasing respi- [7] H. Shin and J. Cho, "Unconstrained snoring detection using a smartphone
during ordinary sleep," Biomedical engineering online, vol. 13, no. 1, p.
ration rate. It can be seen that most of the detected respiration 116, 2014.
rates were concentrated in the interval of six to seven bpm, be- [8] A. H. Alkali, "Evaluation of Thresholding Based Noncontact Respiration
cause respiration tends to remain steady during sleep, and the Rate Monitoring using Thermal Imaging.," Evaluation, vol. 6, no. 7, pp.
respiration rate decreases. The proposed system can be used for 50-58, Jul 2017.
[9] I. Sadek, E. Seet, J. Biswas, B. Abdulrazak, and M. Mokhtari,
studies on central sleep apnea, in which a patient has an abnor- "Nonintrusive vital signs monitoring for sleep apnea patients: A
mally low respiration rate during sleep. preliminary study," IEEE Access, vol. 6, pp. 2506-2514, 2018.
In order to improve the reliability of the system and algorithm [10] Z. Yang, P. H. Pathak, Y. Zeng, X. Liran, and P. Mohapatra, "Vital sign
developed in this study, further experiments and evaluations and sleep monitoring using millimeter wave," ACM Transactions on
Sensor Networks (TOSN), vol. 13, no. 2, p. 14, 2017.
should be conducted with more subjects. In addition, a wider [11] F. Deng et al., "Design and Implementation of a Noncontact Sleep
variety of sleep health parameters, such as sleep time measure- Monitoring System Using Infrared Cameras and Motion Sensor," IEEE
ment and sleep-stage classification, should be analyzed using Transactions on Instrumentation and Measurement, vol. 67, no. 7, pp.
the proposed system. This study showed the potential of the 1555-1563, 2018.
[12] A. Shahshahani, S. Bhadra, and Z. Zilic, "A Continuous Respiratory
proposed method for monitoring sleep health during daily life Monitoring System Using Ultrasound Piezo Transducer," in Circuits and
without having to rely on medical devices. The proposed sys- Systems (ISCAS), 2018 IEEE International Symposium on, 2018, pp. 1-4:
tem can be used to monitor the health statuses of elderly people IEEE.
living alone or those living with chronic diseases that require [13] T. Hall et al., "Non-Contact Sensor for Long-Term Continuous Vital Signs
Monitoring: A Review on Intelligent Phased-Array Doppler Sensor
continuous management. It can also be used to collect daily Design," Sensors, vol. 17, no. 11, p. 2632, 2017.
health data for medical big data studies in the future. [14] S. H. Hwang et al., "Unconstrained sleep apnea monitoring using
polyvinylidene fluoride film-based sensor," IEEE Transactions on
VI. CONCLUSION Biomedical Engineering, vol. 61, no. 7, pp. 2125-2134, 2014.
[15] G. S. Chung, B. H. Choi, D. U. Jeong, and K. S. Park, "Noninvasive heart
Using the unobtrusive force measurement system developed rate variability analysis using loadcell-installed bed during sleep," (in eng),
in this study, it is possible to detect the respiration rate per mi- Conf Proc IEEE Eng Med Biol Soc, vol. 2007, pp. 2357-60, 2007.
[16] J. H. Shin, Y. J. Chee, D.-U. Jeong, and K. S. Park, "Nonconstrained sleep
nute during sleep, and the awakening state is expected to be monitoring system and algorithms using air-mattress with balancing tube

1558-1748 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JSEN.2018.2883716, IEEE Sensors
Journal

method," IEEE transactions on information technology in biomedicine, vol. [19] R. B. Berry et al., The AASM Manual for the Scoring of Sleep and
14, no. 1, pp. 147-156, 2010. Associated Events: Rules, Terminology and Technical Specifications :
[17] G. H. Gautschi, Piezoelectric Sensorics : Force, Strain, Pressure, Version 2.3. American Academy of Sleep Medicine, 2015.
Acceleration and Acoustic Emission Sensors, Materials and Amplifiers / G. [20] J. R. Landis and G. G. Koch, "The measurement of observer agreement for
Gautschi. 2002. categorical data," biometrics, pp. 159-174, 1977.
[18] S.-G. P. P. Corporation, "Norton PVDF Fluoropolymer Film," ed, 2002.

1558-1748 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.

You might also like