Professional Documents
Culture Documents
www.aerogen.com
Contents
Introduction 2
System Warnings 5
Nebulisation 16
Functional Test 19
Troubleshooting 26
Warranty 28
Life Of Product 28
Specifications 29
Performance 30
Symbols 31
Appendix 1 32
Aerogen® Pro is intended for hospital use only. It is designed to operate in-
line with standard ventilator circuits and mechanical ventilators. It operates
without changing patient ventilator parameters and can be refilled without
interrupting ventilation.
The controller operates from the AC/DC adapter and can be operated on
its internal rechargeable battery for up to 45 minutes when fully charged.
The product operates without compressed gas, making it suitable for
portable applications.
The Aerogen Pro System is a portable medical device for multiple patient
use that is intended to aerosolise physician-prescribed solutions for
inhalation to patients on and off ventilation or other positive pressure
breathing assistance. The Aerogen Pro System is suitable for use in adult,
paediatric and neonate patients.
2 Aerogen®
Aerogen Pro System
1 2 3 4
5 6
4. Controller Cable
5. AC/DC Adapter
2. The T-piece securely connects the nebuliser into the breathing circuit
and can be easily removed for cleaning. The T-piece connections are
standard male and female 22 mm ISO conical ports and connect to
standard patient breathing circuits.
3, 4, 5.
The controller can operate from the AC/DC Adapter or the internal
rechargeable battery. The controller includes an On/Off power button
and sockets for the controller cable and the AC/DC Adapter. The
controller also includes indicators for nebulisation cycle selection (15
or 30 minutes), battery charge status and fault conditions.
4 Aerogen®
System Warnings
Read and study all instructions before using the Aerogen Pro System
and accessories. Only trained medical personnel should operate the
device.
Clean, sterilise, assemble and perform a functional test (page 19) according
to the instructions in this manual before first use and between patients.
•• Do not apply undue pressure to the domed aperture plate in the centre
of the nebuliser.
6 Aerogen®
Aerogen Pro Controller
On/Off Power
• 15 Min. - Press & Release
• 30 Min. - Press & hold for 3 Sec.
Error Indicator
30 Minute Mode
Indicator light
15 Minute Mode
Indicator light
Battery Status
Indicator
9V DC Input
Controller
Cable Input
To recharge the battery, connect the AC/DC Adapter to the controller and
connect to AC power source. The battery status indicator is amber while
charging and green when fully charged.
Allow a minimum of four hours for the internal battery to fully recharge.
8 Aerogen®
Assembly & Installation
Aerogen Pro System Set-Up
1 Perform a functional test of Aerogen Pro before use and between
patients as described in the Funtional Test section of this manual
(see page 19).
2 Insert the filler cap into the opening on the nebuliser.
3 Connect the nebuliser to the T-piece by pushing the nebuliser firmly
onto the T-piece (Figure 3).
Figure 4. Connecting controller and nebuliser
10 Aerogen®
6 The Aerogen Pro can be battery-operated for portable applications.
The rechargeable battery can power the System for up to 45 minutes
when fully charged. In the case of AC power failure the controller will
automatically switch to battery operation.
Note: Allow a minimum of four hours for the internal battery to fully
recharge.
12 Aerogen®
Figure 8. Connecting to a neonatal breathing circuit using neonate T-piece
Warnings
• Always maintain the nebuliser in a vertical orientation (with the filler cap
uppermost) while in the patient circuit (Figures 6, 7, & 8). This orientation
prevents condensate from blocking the nebuliser and ensures proper
nebulisation. Always visually inspect the nebuliser prior to placing
in the ventilator circuit to assure that no secretions are blocking the
Aerogen Vibronic® aerosol generator.
• When removing the nebuliser from the patient circuit always replace
the T-piece plug to maintain circuit pressure.
• Always connect a bacteria filter to the expiratory inlet of the ventilator.
Otherwise the function of the expiratory channel may be degraded.
• Do not use a filter or heat-moisture exchanger (HME) between the
nebuliser and patient airway.
14 Aerogen®
Adding Medication
•• Open the filler cap tab on the nebuliser.
•• Use a pre-filled ampoule or syringe to add medication into the filler port
of the nebuliser (Figure 10).
The maximum capacity of the nebuliser is 10 mL. Do not fill the nebuliser
beyond the maximum fill indication point (Figure 10). The underside of the
filler cap represents maximum fill indication point.
Max. fill
indication
point
3. To stop the nebuliser at any time, press the On/Off power button. The
indicator turns off to indicate that nebulisation has stopped.
Note: When delivering a dose greater than 3 mL, select the 30 Minute
cycle.
16 Aerogen®
Installation for use Off-Ventilator
Use with a Face Mask
Mask kits, which include a vented elbow and mask elbow, are available
separately (visit www.aerogen.com for full parts list). Contact your sales
representative for ordering information.
1. When using a mask, connect the vented elbow, mask elbow and mask
to the nebuliser by firmly pushing the parts together.
2. Rotate the vented elbow to suit the position of the patient (Figure 11).
Vented
Vented FaceFace
MaskMask
Elbow
Elbow Elbow
Elbow
Patient
Patient
Upright
Upright Patient
Patient
Reclined
Reclined
18 Aerogen®
Functional Test
Perform a functional test of the Aerogen Pro System prior to first use, after
each sterilisation before each patient use or at any time to verify proper
operation. Follow these steps:
1. Visually inspect each part of the system for cracks or damage and
replace if any defects are visible.
4. Press and release the blue On/Off power button and verify that the
green 15 Min. indicator light illuminates and that aerosol is visible.
5. Disconnect the nebuliser from the controller. Verify that the amber
Error Indicator lights. Reconnect the nebuliser to the controller.
6. Press the On/Off power button again to turn the system off. Press and
hold the button for at least three seconds. Verify that the green 30 Min.
indicator light illuminates and that aerosol is visible.
7. Disconnect the controller from the AC/DC Adapter and verify that
nebulisation continues and that the battery status indicator turns off.
8. Turn the system off and verify that the 15 Min. and 30 Min. indicators
are off.
•• Neonate Adapters
•• Mounting Bracket*
•• Do not apply undue pressure to the domed aperture plate in the centre
of the nebuliser.
20 Aerogen®
Manual Cleaning
Cleaning nebuliser, T-pieces and Neonate Adapters
3. Clean all parts with warm water and mild liquid detergent.
5. Shake excess water from parts and allow parts to fully air dry.
Disinfection
Aerogen Pro nebuliser, T-pieces and Neonate Adapters with
disinfection agents.
Note: Aerogen approves the following disinfection solutions for use with
its Aerogen Pro nebulisation system regarding material compatibility. With
respect to microbiological effectiveness, please ask the manufacturer.
Refer to the product labelling for specific instructions regarding activation,
safe use and disposal of these solutions.
•• Isopropyl (70%)
•• CIDEX®
•• NU-CIDEX®
•• CIDEX® OPA
•• Hexanios G+R
Method:
3. Clean the components with liquid alkaline cleaner at 55 °C (131 °F) for
10 minutes.
Method:
22 Aerogen®
Sterilisation of the Aerogen Pro Nebuliser
Sterilisation of Aerogen Pro Nebuliser, T-Pieces & Neonate Adapters
1. Disconnect the nebuliser from the controller, and then remove the
nebuliser and adapters from the ventilator circuit, mask or mouthpiece.
4. Clean all parts with warm water and mild liquid detergent in accordance
with current hospital protocols. Rinse thoroughly and air dry.
5. Check for cracks or damage and replace if any defects are visible.
Sterilise Components
Steam sterilisation can be performed using the following three methods:
Note: Sterilisation using the long autoclave cycle (No. 3 above) may cause
some areas of the nebuliser to become discolored. This is not indicative of
the performance of the nebuliser.
1. Check for cracks or damage and replace if any defects are visible.
24 Aerogen®
Cleaning the Aerogen Pro Controller
Cleaning of controller, controller cable & AC/DC adapter
3. Visually inspect for damage and replace the controller if any damage
is observed.
Warnings
• Do not autoclave.
Wipe clean with a damp cloth and mild liquid detergent. Do not use
abrasive or sharp tools.
Battery will not retain Rechargeable battery may Contact your local Aerogen
initial charge. need to be replaced. sales representative.
26 Aerogen®
Table 2. Aerogen Pro Troubleshooting (Continued)
Note: The rechargeable battery in the Aerogen Pro controller should only
be replaced by Aerogen authorised personnel: contact your Aerogen sales
representative.
Life of Product
As with all active electronic components, the Aerogen Pro nebuliser has a
defined life. In the case of Aerogen Pro controller, the life of the controller
unit has been validated for use for 1460 doses. This is based on a typical
product usage profile over a two year period, including four treatments per
day, 50% of the time.
The life of the Aerogen Pro nebuliser and components have been validated
for use for 730 doses and 26 autoclave treatments based on a typical one
year usage profile of four treatments per day and one sterilisation per
week, where the device is assumed to be in service for 50% of the time.
The user should note that any use in excess of this may result in reduced
life of the product.
28 Aerogen®
Specifications
Table 3. Physical Specifications of the Aerogen Pro System
45 mm H x 50 mm W x 50 mm D
Nebuliser Dimensions
1.8” H x 2.0” W x 2.0” D
Aerogen Pro Controller 33mm H x 75mm W x 131mm D
Dimensions 1.3” H x 2.9” W x 5.2”D
Controller Cable Length 1.8 m (5.9 ft.)
AC/DC Adapter Cable Length 2.1 m (6.7 ft.)
Nebuliser Weight 25 g (0.9 oz) nebuliser and filler cap
Aerogen Pro Controller Weight 230 g (8.1 oz.), including battery and cable
Nebuliser Capacity Maximum 10 mL
* Consult your local representative for the order number extension specific
to your country and for pricing information.
100%
80%
60%
40%
20%
0%
0.1 1 10
Particle Size (µm)
30 Aerogen®
Symbols
Warnings
• Only use the Aerogen Pro nebuliser with components specified in the
Instruction Manual. Use of the Aerogen Pro nebuliser with components
other than those specified in the Instruction Manual may result in
increased emissions or decreased immunity of the Aerogen Pro
nebuliser system.
• Do not use the Aerogen Pro adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the device should
be observed to verify normal operation in this configuration.
• The Aerogen Pro needs special precautions regarding electromagnetic
compatibility (“EMC”) and must be installed and put into service
according to the EMC information provided in the Instruction Manual.
32 Aerogen®
• Portable and mobile radio frequency (“RF”) communication devices
can disrupt medical electrical equipment.
* Consult your local representative for the order number extension specific
to your country and for pricing information.
34 Aerogen®
Recommended separation distances between portable and mobile RF
communication equipment and the Aerogen Pro nebuliser system that is not life
supporting
This Aerogen Pro nebuliser system is intended for use in the electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the user
of the Aerogen Pro nebuliser system can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Aerogen Pro nebuliser system as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum Separation distance according to frequency of transmitter m
output power of
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
transmitter W
d = [1.17] √P d = [1.17] √P d = [2.33] √P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.75
1 1.17 1.17 2.33
10 3.70 3.70 7.36
100 11.70 11.70 23.30
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (w) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
36 Aerogen®
Guidance and manufacturer’s declaration - electromagnetic immunity
This Aerogen Pro nebuliser system is intended for use in the electromagnetic
environment specified below. The customer or the user of the Aerogen Pro nebuliser
system should assure that it is used in such an environment.
IEC/EN 60601 Compliance Electromagnetic
Immunity Test
Test Level Level Environment - Guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the Aerogen Pro nebuliser
system, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a) F
ield strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the Aerogen Pro nebuliser system is used exceeds the applicable
RF compliance level above, the Aerogen Pro nebuliser system should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as
re-orientating or relocating the Aerogen Pro nebuliser system.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m
38 Aerogen®
INTL. +353 91 540 400
e. info@aerogen.com
w. www.aerogen.com
twitter.com/aerogen
Manufacturer
© 2015 Aerogen Ltd. Aerogen Ltd.
Part No. AG-AP1080-UK Galway Business Park,
P/N 30-012 Rev P Dangan, Galway,
Ireland.