You are on page 1of 4

PHYSICS

Physics of ultrasound Learning objectives


Alexander EJ Powles After reading this article, you should be able to:
David J Martin C describe the characteristics of sound waves
C explain how ultrasound waves are produced, how they interact
Irving TP Wells
with the tissues they encounter to produce echoes and how
Christopher R Goodwin they are then interpreted to form a two-dimensional image
C identify common ultrasound imaging artefacts
C explain the Doppler effect and its use in ultrasound imaging
Abstract
Ultrasound is a form of non-ionizing radiation that uses high-frequency
sound waves to image the body. It is a real-time investigation which
Ultrasound refers to sound waves that have too high a frequency
allows assessment of moving structures and also facilitates measure-
for the human ear to detect, that is above 20 kilohertz (kHz);
ment of velocity and directionality of blood flow within a vessel. It can
however, the frequencies used in medical imaging are far greater,
be used for a variety of purposes in the intensive care setting, for
typically 2e10 megahertz (MHz).
example to aid central venous catheter and pleural drain insertion.
When using this imaging modality it is vital to understand the relevant
Wavelength is the distance travelled by sound in one cycle. It is
physical principles and how the images are created. This article will
inversely proportional to the frequency, i.e. the shorter the
explain these principles, including the use of Doppler ultrasound and
wavelength the higher the frequency (Figure 2). Shorter wave-
the interpretation of common artefacts.
lengths result in higher resolution images but less penetration
Keywords Doppler effect; echoes; imaging artefacts; sound into the soft tissues. Therefore, in medical imaging higher fre-
attenuation; ultrasound safety; ultrasound waves quency probes (5e10 MHz) are used for superficial structures,
(e.g. peripheral vessel visualization) and lower frequency probes
Royal College of Anaesthetists CPD Matrix: 1A03
(2e5 MHz) for imaging deeper structures (e.g. the abdominal
and pelvic organs).

Ultrasound image production


Wave characteristics Production of echoes
Ultrasound waves are produced by briefly passing an electrical
Sound is a mechanical wave that causes disturbance in a me-
current through a piezoelectric crystal within the ultrasound
dium, transferring energy from one point to another. As the
probe. The resulting pulse of ultrasound waves is delivered to the
sound wave travels through a medium, the energy is propagated
tissues. The crystal then waits for the rebounding echoes to be
through collisions of adjacent particles which oscillate around
received before transmitting the next pulse. The pulse duration is
their resting position without net displacement. Sound waves can
typically 1 ms repeated at 1 ms intervals. Each crystal therefore
be longitudinal or transverse depending on the direction of their
emits ultrasound waves 0.1% of the time and receives the
oscillations in relation to the direction of the energy travelling
returning echoes for the remaining 99.9% of the time. In practice,
through the medium. Only solids can propagate sound waves
the probe contains a phased array of many piezoelectric crystals
transversely, whereas all materials can support a longitudinal
which are stimulated sequentially by electronic pulses which
wave, and this is therefore how ultrasound is transmitted
sweep from one side of the probe to the other.
through the soft tissues and fluid within the body (Figure 1).
As the ultrasound wave passes through the patient it en-
The frequency of a sound wave (measured in hertz, Hz) is the
counters interfaces between different tissues. At these interfaces
number of oscillations (or cycles) per second. If a particle com-
a proportion of the wave’s energy is reflected and the remainder
pletes one full oscillation per second, its frequency is 1 Hz.
is transmitted. If the angle between the interface and transducer
is greater than around 60 degrees then some of the reflected
echoes will return to the transducer. The piezoelectric crystal
Alexander EJ Powles BA BMBCh is an ST3 Specialty Registrar in the converts the reflected sound waves into electrical pulses, and
South Wales Radiology Training Scheme, UK. Conflict of interest: these are interpreted into a two-dimensional (2D) image. The
None declared.
more energy in the returning echoes, the brighter the image
David J Martin MBBCh MRCP FRCR is a Consultant Radiologist at displayed.
Morriston and Singleton Hospitals, Swansea, Wales, UK. Conflict of The amount of energy reflected at an interface between tissues
interest: None declared. depends on the difference in acoustic impedance of those tis-
Irving TP Wells MBBS MRCP FRCR is a Consultant Radiologist at sues. The acoustic impedance (Z) of a tissue is the product of its
Morriston and Singleton Hospitals, Swansea, Wales, UK. Conflict of density and the velocity that sound travels through it. Air has a
interest: None declared. much lower density than water or soft tissue, which in turn have
Christopher R Goodwin MB ChB MRCS (Eng) FRCR is a Consultant a much lower density than bone. The larger the difference in
Radiologist at Princess of Wales Hospital, Bridgend, Wales, UK. acoustic impedance between two materials at an interface, the
Conflict of interest: None declared. more energy will be reflected, and the brighter the resulting

ANAESTHESIA AND INTENSIVE CARE MEDICINE --:- 1 Ó 2018 Published by Elsevier Ltd.

Please cite this article in press as: Powles Alexander EJ., et al., Physics of ultrasound, Anaesthesia and intensive care medicine (2018), https://
doi.org/10.1016/j.mpaic.2018.01.005
PHYSICS

Rearranging, the tissue interface depth is therefore:

d ¼ ct/2
At rest
Sound attenuation and compensation
As a sound wave passes through the body it gradually loses its
energy in a process called attenuation. The causes of this are:
Transverse
reflection, refraction, scatter and absorption (Figure 3).
propagation
Reflection and refraction occur at the interfaces between tissues.
Reflection, as already described, is responsible for the produc-
tion of the required echoes. Refraction causes a transmitted
wave to be deflected from its original direction where it passes
Longitudinal
through an interface between tissues having differing wave
propagation speed. Scatter describes the scattering of the wave in all di-
rections which occurs when a wave encounters a structure much
smaller than its wavelength, typically occurring on interaction
with red blood cells. The majority of attenuation, however, oc-
Figure 1 Diagrammatic representation of the propagation of a me-
chanical force transversely as can occur only in solids or longitudinally curs due to absorption. The energy of the sound wave is con-
as occurs with sound in the tissues of the body. verted into friction between oscillating tissue particles and is lost
in the form of heat.
The combination of general attenuation through all tissue
types, and the fact that only a small proportion of the wave’s
High-frequency wave energy is reflected at many of the tissue interfaces (which is
then attenuated further as it travels back towards the probe),
means that the ultrasound transducer will only receive a very
Time small amount of energy from the returning echoes, particularly
from deeper structures. To compensate for this loss of energy
the ultrasound machine uses a process called time gain
compensation. This gives greater amplification to those echoes
Wavelength
which take longer to return to the transducer, producing a more
even image.
Low-frequency wave
Image resolution
Time Spatial resolution is the ability of an imaging system to distin-
guish two points as separate in space.

Axial or depth resolution is the ability to distinguish between


Wavelength two structures in the direction parallel to the beam, that is, along
the same scan line. A structure can only be visualized if it is
larger than several wavelengths of the emitted ultrasound.
Figure 2 Displays the relationship between wavelength and frequency.
Remembering that wavelength is inversely proportional to fre-
quency, the higher the frequency (shorter the wavelength) the
image. At the interface of tissues of similar densities, such as better the axial resolution. Hence a high-frequency probe is
liver and kidney, less than 1% of the wave’s energy is reflected. required to image small structures. However, as discussed above,
However at the interface between soft tissue and air or bone, higher frequency waves are attenuated more rapidly, and there-
nearly all of the wave’s energy is reflected. No energy is trans- fore will only be able to visualize superficial structures.
mitted, and hence no information can be gained about tissues
which lie deeper than this point. This explains why ultrasound is Lateral resolution is the ability to distinguish two structures
generally not useful for assessment of bone, bowel, or lung. It lying side by side at the same depth. This is dependent on the
also explains why a coupling gel is required between the probe beam width being narrower than the distance between the two
and patient’s skin and why air bubbles must be avoided to structures and is also improved by focussing. By exploiting the
minimize any reflection at the skin/probe interface. phased array of the probe, if a delay is introduced energizing the
To create a 2D image, the depth (d) of the tissue interface outermost elements first with a short time interval to the inner
must be calculated. Ultrasound travels at an average speed (c) of elements and finally central element, there will be a point at
1540 ms1 through soft tissue. The time (t) taken for the ultra- which the pulses all arrive together and thus reinforce allowing
sound pulse to travel the distance (d) to the interface, and for the the beam to be focussed. Practically it is important that the
reflected wave to return to the transducer, is: operator correctly positions the focal zone when scanning to
ensure the best lateral resolution for assessment of a given
t ¼ 2d/c structure.

ANAESTHESIA AND INTENSIVE CARE MEDICINE --:- 2 Ó 2018 Published by Elsevier Ltd.

Please cite this article in press as: Powles Alexander EJ., et al., Physics of ultrasound, Anaesthesia and intensive care medicine (2018), https://
doi.org/10.1016/j.mpaic.2018.01.005
PHYSICS

Artefacts Posterior acoustic shadowing (Figure 4d) occurs when the ul-
trasound beam encounters a strongly attenuating or highly
Imaging artefacts can lead to misinterpretation but can also aid
reflective structure or surface. There is little or no through
the operator in certain circumstances.
transmission and this creates a dark area beyond the surface.
Reverberation artefact (Figure 4a) occurs when sound waves
bounce back and forth between two reflective surfaces before
returning to the transducer. These returning echoes will arrive
later than the original echo in a sequential manner creating
numerous evenly spaced lines, as the processor believes these
echoes to be returning from deeper structures.

Ring-down artefact (Figure 4b) occurs when sound waves hit a


small structure, such as a small air bubble, that then resonates
at the same frequency as ultrasound and emits sound. The
sound is produced after the original wave has been reflected
back to the transducer. It will therefore be interpreted as a
second echo originating from a deeper structure and is dis-
played as a bright line deep to the original target. Practically this
artefact can be utilized to aid identification of air within the
biliary tree.

Mirror artefact (Figure 4c) occurs at highly reflective surfaces.


The original beam is reflected from a surface and then en-
counters an adjacent structure, at which point the original echo
is reflected back towards the highly reflective surface, before
being reflected back again to the transducer. This is interpreted
as returning from a deeper structure which appears to be on the
opposite side of the reflective surface. This is classically seen at
the diaphragm.

a b

c d

Figure 4 Ultrasound artefacts. (a) Reverberation artefact e evenly


spaced lines in the bladder. (b) Ring-down artefact e bright white line
caused by air within the biliary tree. (c) Mirror artefact e liver ‘reflected’
Figure 3 (a) Absorption; (b) reflection; (c) scatter; and (d) refraction on the other side of the diaphragm. (d) Posterior acoustic shadowing e
all contribute to the overall attenuation of the ultrasound wave seen caused by two gallstones in the gallbladder. (e) Posterior acoustic
in (e). enhancement e due to simple liver cyst.

ANAESTHESIA AND INTENSIVE CARE MEDICINE --:- 3 Ó 2018 Published by Elsevier Ltd.

Please cite this article in press as: Powles Alexander EJ., et al., Physics of ultrasound, Anaesthesia and intensive care medicine (2018), https://
doi.org/10.1016/j.mpaic.2018.01.005
PHYSICS

Visualization of any structure deep to this point is therefore in conjunction with the angle correction function on the ultra-
diminished. Practically this artefact can help identify calcified sound machine.
structures such as gallstones and renal calculi. Doppler imaging can provide information about flow direc-
tionality which is typically displayed using colour mapping, with
Posterior acoustic enhancement (Figure 4e) occurs when the red conventionally representing flow towards the probe and blue
ultrasound beam passes through a weakly attenuating structure, away from it. Alternatively, it can be displayed as a graph of
such as fluid in a simple cyst or urine within the bladder. The velocity against time, i.e. spectral Doppler, which allows the
structures posterior to the cyst will appear brighter, as more operator to obtain measurements of flow velocity.
energy of the original beam will reach them, and hence an echo
of greater energy will be reflected back to the probe. Safety
Ultrasound is safe and does not use ionizing radiation. Known
Doppler
bio-effects fall into two main categories: mechanical and ther-
The Doppler Effect describes the change in frequency of the mal. Mechanical effects are related to tiny bubbles which are
incident wave that occurs when it is reflected at a moving produced within tissues as ultrasound propagates through
interface, and is the reason the pitch of a horn from a moving them. If these bubbles collapse rapidly (microcavitation), en-
vehicle appears to change as it approaches and passes by a sta- ergy is released which can damage tissues. Thermal effects
tionary observer. occur from the heat produced from the friction of oscillating
particles.
Doppler shift frequency is measured as the difference between In routine medical imaging neither of these bio-effects causes
the transmitted and observed frequency (ff1). If the object(s) any significant or lasting damage but it is good practice to keep
being imaged (e.g. blood cells) are moving towards the trans- the examination time to a minimum and to use appropriate set-
ducer, their motion will cause the reflected waves to be com- tings to minimize the power, particularly during foetal
pressed and return back to the transducer more rapidly, scanning. A
shortening the wavelength and increasing the frequency. The
opposite occurs when the blood cells are moving away from the
FURTHER READING
transducer, resulting in a decrease in frequency of the reflected
Aldrich JE. Basic physics of ultrasound imaging. Crit Care Med 2007;
wave. The change in frequency is proportional to the velocity of
35(suppl 5): S131e7, http://ovidsp.uk.ovid.com.abc.cardiff.ac.uk/
the moving interface:
sp-3.12.0b/ovidweb.cgi?&S¼MDEAPDMAJKHFMDMLFNMKA
CDGGHAMAA00&LinkþSet¼jb.search.32%7c2%7csl_10.
change of frequencyðf  f1 Þ 2  velocity of the interfaceðvÞ
¼ Feldman MK, Katyal S, Blackwood MS. US artifacts. Radiographics
original frequencyðfÞ velocity of soundðcÞ
2009; 29: 1179e89.
Hangiandreou NJ. AAPM/RSNA physics tutorial for residents. Topics
The Doppler shift frequency is also related to the cosine of the
in US: B-mode US: basic concepts and new technology. Radio-
angle between the probe and the vessel being imaged:
graphics 2003; 23: 1019e33.
(f  f1)/f ¼ 2(v/c) cosq Hopkins R, Peden C, Gandhi S. Ultrasound. In: Radiology for anaes-
thesia and intensive care. 2nd edn. Cambridge: Cambridge Uni-
The cosine of 90 degrees is 0, therefore no Doppler signal will be versity Press, 2009.
present when the probe is perpendicular to the vessel. Ideally the Middleton B, Phillips J, Thomas R, Stacey S. Ultrasound. In: Physics in
probe should be positioned such that the flow of the vessel is anaesthesia. 1st edn. Scion Publishing, 2012.
directly toward it. In practice this is difficult, and if measurement Ziskin MC. Fundamental physics of ultrasound and its propagation in
of velocity is required an angle less than 60 degrees can be used tissue. Radiographics 1993; 13: 705e9.

ANAESTHESIA AND INTENSIVE CARE MEDICINE --:- 4 Ó 2018 Published by Elsevier Ltd.

Please cite this article in press as: Powles Alexander EJ., et al., Physics of ultrasound, Anaesthesia and intensive care medicine (2018), https://
doi.org/10.1016/j.mpaic.2018.01.005

You might also like