Professional Documents
Culture Documents
ANAESTHESIA AND INTENSIVE CARE MEDICINE --:- 1 Ó 2018 Published by Elsevier Ltd.
Please cite this article in press as: Powles Alexander EJ., et al., Physics of ultrasound, Anaesthesia and intensive care medicine (2018), https://
doi.org/10.1016/j.mpaic.2018.01.005
PHYSICS
d ¼ ct/2
At rest
Sound attenuation and compensation
As a sound wave passes through the body it gradually loses its
energy in a process called attenuation. The causes of this are:
Transverse
reflection, refraction, scatter and absorption (Figure 3).
propagation
Reflection and refraction occur at the interfaces between tissues.
Reflection, as already described, is responsible for the produc-
tion of the required echoes. Refraction causes a transmitted
wave to be deflected from its original direction where it passes
Longitudinal
through an interface between tissues having differing wave
propagation speed. Scatter describes the scattering of the wave in all di-
rections which occurs when a wave encounters a structure much
smaller than its wavelength, typically occurring on interaction
with red blood cells. The majority of attenuation, however, oc-
Figure 1 Diagrammatic representation of the propagation of a me-
chanical force transversely as can occur only in solids or longitudinally curs due to absorption. The energy of the sound wave is con-
as occurs with sound in the tissues of the body. verted into friction between oscillating tissue particles and is lost
in the form of heat.
The combination of general attenuation through all tissue
types, and the fact that only a small proportion of the wave’s
High-frequency wave energy is reflected at many of the tissue interfaces (which is
then attenuated further as it travels back towards the probe),
means that the ultrasound transducer will only receive a very
Time small amount of energy from the returning echoes, particularly
from deeper structures. To compensate for this loss of energy
the ultrasound machine uses a process called time gain
compensation. This gives greater amplification to those echoes
Wavelength
which take longer to return to the transducer, producing a more
even image.
Low-frequency wave
Image resolution
Time Spatial resolution is the ability of an imaging system to distin-
guish two points as separate in space.
ANAESTHESIA AND INTENSIVE CARE MEDICINE --:- 2 Ó 2018 Published by Elsevier Ltd.
Please cite this article in press as: Powles Alexander EJ., et al., Physics of ultrasound, Anaesthesia and intensive care medicine (2018), https://
doi.org/10.1016/j.mpaic.2018.01.005
PHYSICS
Artefacts Posterior acoustic shadowing (Figure 4d) occurs when the ul-
trasound beam encounters a strongly attenuating or highly
Imaging artefacts can lead to misinterpretation but can also aid
reflective structure or surface. There is little or no through
the operator in certain circumstances.
transmission and this creates a dark area beyond the surface.
Reverberation artefact (Figure 4a) occurs when sound waves
bounce back and forth between two reflective surfaces before
returning to the transducer. These returning echoes will arrive
later than the original echo in a sequential manner creating
numerous evenly spaced lines, as the processor believes these
echoes to be returning from deeper structures.
a b
c d
ANAESTHESIA AND INTENSIVE CARE MEDICINE --:- 3 Ó 2018 Published by Elsevier Ltd.
Please cite this article in press as: Powles Alexander EJ., et al., Physics of ultrasound, Anaesthesia and intensive care medicine (2018), https://
doi.org/10.1016/j.mpaic.2018.01.005
PHYSICS
Visualization of any structure deep to this point is therefore in conjunction with the angle correction function on the ultra-
diminished. Practically this artefact can help identify calcified sound machine.
structures such as gallstones and renal calculi. Doppler imaging can provide information about flow direc-
tionality which is typically displayed using colour mapping, with
Posterior acoustic enhancement (Figure 4e) occurs when the red conventionally representing flow towards the probe and blue
ultrasound beam passes through a weakly attenuating structure, away from it. Alternatively, it can be displayed as a graph of
such as fluid in a simple cyst or urine within the bladder. The velocity against time, i.e. spectral Doppler, which allows the
structures posterior to the cyst will appear brighter, as more operator to obtain measurements of flow velocity.
energy of the original beam will reach them, and hence an echo
of greater energy will be reflected back to the probe. Safety
Ultrasound is safe and does not use ionizing radiation. Known
Doppler
bio-effects fall into two main categories: mechanical and ther-
The Doppler Effect describes the change in frequency of the mal. Mechanical effects are related to tiny bubbles which are
incident wave that occurs when it is reflected at a moving produced within tissues as ultrasound propagates through
interface, and is the reason the pitch of a horn from a moving them. If these bubbles collapse rapidly (microcavitation), en-
vehicle appears to change as it approaches and passes by a sta- ergy is released which can damage tissues. Thermal effects
tionary observer. occur from the heat produced from the friction of oscillating
particles.
Doppler shift frequency is measured as the difference between In routine medical imaging neither of these bio-effects causes
the transmitted and observed frequency (ff1). If the object(s) any significant or lasting damage but it is good practice to keep
being imaged (e.g. blood cells) are moving towards the trans- the examination time to a minimum and to use appropriate set-
ducer, their motion will cause the reflected waves to be com- tings to minimize the power, particularly during foetal
pressed and return back to the transducer more rapidly, scanning. A
shortening the wavelength and increasing the frequency. The
opposite occurs when the blood cells are moving away from the
FURTHER READING
transducer, resulting in a decrease in frequency of the reflected
Aldrich JE. Basic physics of ultrasound imaging. Crit Care Med 2007;
wave. The change in frequency is proportional to the velocity of
35(suppl 5): S131e7, http://ovidsp.uk.ovid.com.abc.cardiff.ac.uk/
the moving interface:
sp-3.12.0b/ovidweb.cgi?&S¼MDEAPDMAJKHFMDMLFNMKA
CDGGHAMAA00&LinkþSet¼jb.search.32%7c2%7csl_10.
change of frequencyðf f1 Þ 2 velocity of the interfaceðvÞ
¼ Feldman MK, Katyal S, Blackwood MS. US artifacts. Radiographics
original frequencyðfÞ velocity of soundðcÞ
2009; 29: 1179e89.
Hangiandreou NJ. AAPM/RSNA physics tutorial for residents. Topics
The Doppler shift frequency is also related to the cosine of the
in US: B-mode US: basic concepts and new technology. Radio-
angle between the probe and the vessel being imaged:
graphics 2003; 23: 1019e33.
(f f1)/f ¼ 2(v/c) cosq Hopkins R, Peden C, Gandhi S. Ultrasound. In: Radiology for anaes-
thesia and intensive care. 2nd edn. Cambridge: Cambridge Uni-
The cosine of 90 degrees is 0, therefore no Doppler signal will be versity Press, 2009.
present when the probe is perpendicular to the vessel. Ideally the Middleton B, Phillips J, Thomas R, Stacey S. Ultrasound. In: Physics in
probe should be positioned such that the flow of the vessel is anaesthesia. 1st edn. Scion Publishing, 2012.
directly toward it. In practice this is difficult, and if measurement Ziskin MC. Fundamental physics of ultrasound and its propagation in
of velocity is required an angle less than 60 degrees can be used tissue. Radiographics 1993; 13: 705e9.
ANAESTHESIA AND INTENSIVE CARE MEDICINE --:- 4 Ó 2018 Published by Elsevier Ltd.
Please cite this article in press as: Powles Alexander EJ., et al., Physics of ultrasound, Anaesthesia and intensive care medicine (2018), https://
doi.org/10.1016/j.mpaic.2018.01.005