You are on page 1of 5

Designation: D 7483 – 08

Standard Test Method for


Determination of Dynamic Viscosity and Derived Kinematic
Viscosity of Liquids by Oscillating Piston Viscometer1
This standard is issued under the fixed designation D 7483; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

1. Scope D 4177 Practice for Automatic Sampling of Petroleum and


1.1 This test method covers the measurement of dynamic Petroleum Products
viscosity and derivation of kinematic viscosity of liquids, such D 5967 Test Method for Evaluation of Diesel Engine Oils in
as new and in-service lubricating oils, by means of an T-8 Diesel Engine
oscillating piston viscometer. D 6792 Practice for Quality System in Petroleum Products
1.2 This test method is applicable to Newtonian and non- and Lubricants Testing Laboratories
Newtonian liquids; however the precision statement was de- 2.2 ISO Standards:3
veloped using Newtonian liquids. ISO/EC 17025 General Requirements for the Competence
1.3 The range of dynamic viscosity covered by this test of Testing and Calibration Laboratories
method is from 0.2 mPa·s to 20 000 mPa·s (which is approxi- 2.3 NIST Standard:4
mately the kinematic viscosity range of 0.2 mm2/s to 22 000 NIST Technical Note 1297 Guideline for Evaluating and
mm2/s for new oils) in the temperature range between –40 to Expressing the Uncertainty of NIST Measurement Results
190°C; however the precision has been determined only for
3. Terminology
new and used oils in the range of 1.434 mPa·s to 154.4 mPa·s
at temperatures of 40 and 100°C (as stated in the precision 3.1 Definitions:
section). 3.1.1 dynamic viscosity (h), n—the ratio between the ap-
1.4 The values stated in SI units are to be regarded as plied shear stress and rate of shear of a liquid.
standard. No other units of measurement are included in this 3.1.1.1 Discussion—It is sometimes called the coefficient of
standard. dynamic viscosity or, simply, viscosity. Thus, dynamic viscos-
1.5 This standard does not purport to address all of the ity is a measure of the resistance to flow or to deformation of
safety concerns, if any, associated with its use. It is the a liquid under external shear forces.
responsibility of the user of this standard to establish appro- 3.1.1.2 Discussion—The term dynamic viscosity can also
priate safety and health practices and determine the applica- be used in a different context to denote a frequency-dependant
bility of regulatory limitations prior to use. quantity in which shear stress and shear rate have a sinusoidal
time dependence.
2. Referenced Documents 3.1.2 kinematic viscosity (n), n—the ratio of the dynamic
2.1 ASTM Standards:2 viscosity (h) to the density (r) of a liquid.
D 445 Test Method for Kinematic Viscosity of Transparent 3.1.2.1 Discussion—For gravity flow under a given hydro-
and Opaque Liquids (and Calculation of Dynamic Viscos- static head, the pressure head of a liquid is proportional to its
ity) density, (r). Therefore the kinematic viscosity, (n), is a measure
D 2162 Practice for Basic Calibration of Master Viscom- of the resistance to flow of a liquid under gravity.
eters and Viscosity Oil Standards 3.1.3 rate of shear (shear rate), n— in liquid flow, the
D 4057 Practice for Manual Sampling of Petroleum and velocity gradient across the liquid.
Petroleum Products 3.1.4 shear stress, n—the force per unit area in the direction
of the flow.
3.1.4.1 Discussion—The SI unit for shear stress is the
1
This test method is under the jurisdiction of ASTM Committee D02 on pascal (Pa).
Petroleum Products and Lubricants and is the direct responsibility of Subcommittee
D02.07 on Flow Properties.
Current edition approved Dec. 15, 2008. Published February 2009.
2 3
For referenced ASTM standards, visit the ASTM website, www.astm.org, or Available from International Organization for Standardization (ISO), 1, ch. de
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM la Voie-Creuse, Case postale 56, CH-1211, Geneva 20, Switzerland, http://
Standards volume information, refer to the standard’s Document Summary page on www.iso.ch.
4
the ASTM website. Available from http://physics.nist.gov/ccu/Uncertainty/index.html.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.

1
D 7483 – 08

FIG. 1 Cross Sectional View of Measurement Chamber

3.1.5 density (r), n—mass per unit volume. The instrument requires a sample volume of less than 5 mL and
3.2 Definitions of Terms Specific to This Standard: typical solvent volume of less than 10 mL which minimizes
3.2.1 oscillating piston viscometer, n—a device that mea- cleanup effort and waste.
sures the travel time of a piston driven electromagnetically into
stationary oscillating motion through a liquid at a controlled 6. Apparatus
force in order to determine the dynamic viscosity of the liquid. 6.1 Oscillating Piston Viscometer:56
6.1.1 The oscillating piston viscometer (see Fig. 2) com-
4. Summary of Test Method prises a measurement chamber and calibrated piston capable of
4.1 A specimen of sample is placed in the thermally measuring the dynamic viscosity within the limits of precision
controlled measurement chamber where the piston resides. The given in Section 16.
piston is driven into oscillatory motion within the measurement 6.1.2 Piston—Free moving, magnetically driven body
chamber by a controlled magnetic field. Once the sample is at within a Oscillating Piston Viscometer which is used for
the test temperature, as determined by the temperature detector, measuring the viscosity of liquids. Individual pistons are sized
the piston is propelled repeatedly through the liquid (by the to measure specific viscosity ranges by varying the sensor
magnetic field). A shear stress (ranging from 5 Pa to 750 Pa) is annulus. See Table 1 for the selection of the piston according
imposed on the liquid under test due to the piston travel. The to the viscosity range.
dynamic viscosity is determined by measuring the average 6.1.3 Measurement Chamber—Location within Oscillating
travel time of the piston. The kinematic viscosity is derived by Piston Viscometer where piston motion (through the liquid
additionally measuring the ratio between the up and down under test) occurs due to an imposed electromagnetic field. See
travel times. This information is then applied to a calibration Fig. 1.
curve using liquids of known viscosity to calculate the dynamic 6.1.4 Electronics—Capable of controlling the electromag-
viscosity and kinematic viscosity of the liquid. See Fig. 1. netic field to propel and detect the travel time of the piston with
a discrimination of 0.01 s or better and uncertainty within
5. Significance and Use 60.07 %. The travel time is calibrated to be between 0.4 s and
5.1 Many petroleum products, as well as non-petroleum 60 s, at a distance of 5 mm.
materials, are used as lubricants for bearings, gears, compres-
sor cylinders, hydraulic equipment, etc. Proper operation of
this equipment depends upon the viscosity of these liquids. 5
The Oscillating Piston Viscometer is covered by a patent. Interested parties are
5.2 Oscillating piston viscometers allow viscosity measure- invited to submit information regarding the identification of an alternative to this
ment of a broad range of materials including transparent, patented item to the ASTM International headquarters. Your comments will receive
careful consideration at a meeting of the responsible technical committee, which
translucent and opaque liquids. The measurement principle and you may attend.
stainless steel construction makes the Oscillating Piston Vis- 6
The sole sources of supply for the apparatus known to the committee at this
cometer resistant to damage and suitable for portable opera- time is Cambridge Viscosity Inc., 101 Station Landing, Medford, MA 02155
(www.cambridgeviscosity.com). If you are aware of alternative suppliers, please
tions. The measurement itself is automatic and does not require
provide this information to ASTM International Headquarters. Your comments will
an operator to time the oscillation of the piston. The electro- receive careful consideration at a meeting of the responsible technical committee,
magnetically driven piston mixes the sample while under test. which you may attend.

2
D 7483 – 08

FIG. 2 Viscometer with Electronics

TABLE 1 Viscosity Ranges of Oscillating Viscometer Pistons


Minimum Viscosity (mPa·s ) Maximum Viscosity (mPa·s ) Piston Designation Nominal Piston Diameter (mm) Recommended Sample Volume (mL)
0.02 2 SP20 7.87 3.2 -5
0.25 5 SP50 7.83 3.2 -5
0.5 10 SP11 7.81 3.2 -5
1 20 SP21 7.76 3.5 -5
2.5 50 SP51 7.68 3.5 -5
5 100 SP12 7.62 3.5 -5
10 200 SP22 7.54 3.5 -5
25 500 SP52 7.34 3.5 -5
50 1000 SP13 7.21 4.0 -5
100 2000 SP23 6.96 4.0 -5
250 5000 SP53 6.27 4.0 –5
500 10000 SP14 6.05 4.0 -5
1000 20000 SP24 5.72 4.0 -5

6.1.5 Temperature Controlled Jacket—Sufficient for main- 7.2.1 The certified viscosity reference should have a pub-
taining measurement chamber temperature within 60.06°C. lished viscosity in accordance with Test Method D 445 or
6.1.6 Temperature Measuring Device—Industrial platinum equivalent means that is close to that of the liquids being tested
resistance thermometer (IPRT) or equivalent sensor with a at the test temperature. For example, if intended measurements
maximum permissible error of 60.02°C. It is recommended, are to be made from 5-25 mPa·s at 100°C, then a reference oil
that the temperature measuring device be verified with an viscosity of 15 mPa·s at 100°C would be appropriate.
independent, calibrated temperature probe at the test tempera- 7.3 Cleaning solvents miscible with the sample and chemi-
ture. cally compatible with the wetted viscometer components (such
6.2 Temperature Regulation System: as alcohols, toluene, etc.). These wetted components are
6.2.1 Any liquid bath or thermoelectric means for regulating typically 316L and 430 Stainless Steel.
the jacket temperature. 7.4 Quality control (QC) liquid similar to 7.1, but with
6.2.2 The temperature control must be such that the tem- viscosity values internally certified as noted in 12.2.
perature of the measurement chamber is held within 60.06°C
of the desired measurement temperature. 8. 8. Sampling, Samples, and Test Units
6.3 Sample Introduction Mechanism—A syringe, micropi- 8.1 Ensure that the sample is homogenous. Engine sampling
pette, or flow-through adapter for introducing between 3.2 mL is generally specified in the test method, for example Test
and 5 mL, inclusive by pressure, into the measurement Method D 5967. When applicable, refer to Practice D 4057
chamber. (manual) or Practice D 4177 (automatic) for proper sampling
7. Reagents and Materials techniques.
7.1 Certified viscosity reference standards shall be certified
by a laboratory that has been shown to meet the requirements 9. Preparation of Apparatus
of ISO/EC 17025 by independent assessment. Viscosity stan- 9.1 Place the viscometer on a stable surface.
dards shall be traceable to master viscometer procedures 9.2 Operate the unit according to the procedure in Section
described in Practice D 2162. 13.
7.2 The uncertainty of the certified viscosity reference 9.3 Verify calibration accuracy by testing a reference stan-
standard shall be stated for each certified value (k = 2, 95 % dard or QC liquid at the test temperature. Follow the procedure
confidence). See ISO/EC 17025 or NIST TN 1297. in Section 13.

3
D 7483 – 08
10. Calibration and Standardization 13.5.1 Allow the sample to equilibrate in the measurement
10.1 Calibrate according to manufacturer’s instructions to chamber for at least 2 min while the piston is oscillating.
obtain a calibration curve (using two test liquids with refer- 13.5.2 Ensure that temperature stability is within the crite-
enced viscosity values near, but within, the extremes of the rion set in 6.2.
piston range being used). 13.5.3 Measure the upward and downward piston travel
10.2 Certified Viscosity Standards may be used as confir- times for each cycle and compute viscosity until the standard
matory checks on the procedure in the laboratory. This proce- deviation as percent of mean, over the previous 20 dynamic
dure is outlined in Section 13. If the dynamic viscosity result, viscosity computations, is less than 0.6%.
at the calibration test point, does not agree with the certified 13.6 Record the average result from the last 20 computa-
value within the limits of precision in Section 16, each step in tions.
the procedure should be rechecked, as well as the temperature
measuring device and viscometer calibration, to locate the 14. Calculation and Interpretation
source of error. If the source is not detected, consult the
manufacturer. 14.1 The calculation of dynamic viscosity and kinematic
viscosity are computed and displayed automatically by the
11. Sample Conditioning apparatus.
11.1 Shake all new and used oil samples using the following
procedure. 15. Report
11.1.1 Ensure cap is tight on the container.
15.1 Dynamic Viscosity result in mPa·s to three significant
11.1.2 Shake vigorously by hand for 30 s. Wait 10 s, or
figures, as displayed by the apparatus.
longer if needed, for air bubbles to dissipate.
11.1.3 A specimen of the sample shall be taken by pipette, 15.2 Kinematic Viscosity result in mm2/s, to three signifi-
pouring or pumping. Suspected nonhomogeneous samples cant figures, as displayed by the apparatus.
must be conveyed for analysis promptly following the shaking 15.3 Temperature in degrees Celsius, to the second decimal
and dissipation procedure of step 11.1.2. place, as displayed by the apparatus.

12. Quality Control/Quality Assurance (QC/QA) 16. Precision and Bias 7

12.1 Confirm proper performance of the instrument and the


16.1 The precision and bias of this test method as deter-
test procedure by analyzing reference oil as QC sample.
mined by the statistical examination of test results is as
12.2 If suitable reference oil is not available, prepare a QC
follows:
sample by replicate analyses of a batch of oil sample. Then
statistically analyze the data to assign a mean value and 16.1.1 Repeatability—The difference between successive
uncertainty limit to the sample. test results, obtained by the same operator with the same
12.3 When QC/QA protocols are already established in the apparatus under constant operating conditions on identical test
testing facility, these may be used to confirm the reliability of material would, in the long run, and in the normal and correct
the test result. operation of the test method, exceed the following values only
12.4 When there is no QC/QA protocol established in the in one case in twenty:
testing facility, guidance may be obtained from Practice Percent of Mean
Dynamic viscosity 2.9
D 6792. Kinematic viscosity 4.2

13. Procedure NOTE 1—This provisional repeatability was obtained from a sample set
13.1 Verify or set the temperature control settings, as tested of ten used and ten fresh oils measured by seven laboratories at 40 and
with the control standard, so the viscometer temperature reads 100°C. Certified Reference Standards were included among the samples
set.
the desired set point 60.06°C while the piston is in motion.
13.2 Remove the piston and clean the specimen from the 16.1.2 Reproducibility—The reproducibility of this test
measurement chamber as described in the viscometer manual. method is being determined and will be available on or before
13.3 Load the measurement chamber with sample using the December 14, 2009.
volume listed in Table 1 related to the piston size being used 16.2 Bias—The average results for the standard liquid used
for the viscosity range anticipated. To minimize contamination, to develop the repeatability statement agreed with the reference
and if sample volume allows, pre-wet the chamber and piston value within statistical error (see Table 2).
with the sample material and dry wipe with a lint free cloth.
13.4 Load the measurement chamber with a clean piston. 17. Keywords
The piston size should be selected such that the measured
viscosity is between the minimum and maximum viscosity 17.1 dynamic viscosity; kinematic viscosity; oscillating pis-
values listed in Table 1. If the reported result is outside of this ton; oscillating piston viscometer; viscosity
range, the measurement shall be repeated using the appropriate
piston size.
13.5 Start the reporting software in accordance with manu- 7
Supporting data have been filed at ASTM International Headquarters and may
facturer’s instructions, which in turn will: be obtained by requesting Research Report RR: D02–1657.

4
D 7483 – 08
TABLE 2 Measurement Bias
Viscosity in mP•s
ILS Average Value in mP•s 1.420 3.077 4.693 9.661 15.36 85.19 155.9
Certified Reference Value 1.434 3.098 4.643 9.671 15.12 86.05 154.4
Viscosity in mm2/s
2
ILS Average Value in mm /s 1.856 3.869 5.747 11.71 18.16 18.88 99.76 184.8
Certified Reference Value 1.853 3.901 5.706 11.83 18.00 18.05 100.8 180.0

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).

You might also like