Professional Documents
Culture Documents
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1
D 7483 – 08
3.1.5 density (r), n—mass per unit volume. The instrument requires a sample volume of less than 5 mL and
3.2 Definitions of Terms Specific to This Standard: typical solvent volume of less than 10 mL which minimizes
3.2.1 oscillating piston viscometer, n—a device that mea- cleanup effort and waste.
sures the travel time of a piston driven electromagnetically into
stationary oscillating motion through a liquid at a controlled 6. Apparatus
force in order to determine the dynamic viscosity of the liquid. 6.1 Oscillating Piston Viscometer:56
6.1.1 The oscillating piston viscometer (see Fig. 2) com-
4. Summary of Test Method prises a measurement chamber and calibrated piston capable of
4.1 A specimen of sample is placed in the thermally measuring the dynamic viscosity within the limits of precision
controlled measurement chamber where the piston resides. The given in Section 16.
piston is driven into oscillatory motion within the measurement 6.1.2 Piston—Free moving, magnetically driven body
chamber by a controlled magnetic field. Once the sample is at within a Oscillating Piston Viscometer which is used for
the test temperature, as determined by the temperature detector, measuring the viscosity of liquids. Individual pistons are sized
the piston is propelled repeatedly through the liquid (by the to measure specific viscosity ranges by varying the sensor
magnetic field). A shear stress (ranging from 5 Pa to 750 Pa) is annulus. See Table 1 for the selection of the piston according
imposed on the liquid under test due to the piston travel. The to the viscosity range.
dynamic viscosity is determined by measuring the average 6.1.3 Measurement Chamber—Location within Oscillating
travel time of the piston. The kinematic viscosity is derived by Piston Viscometer where piston motion (through the liquid
additionally measuring the ratio between the up and down under test) occurs due to an imposed electromagnetic field. See
travel times. This information is then applied to a calibration Fig. 1.
curve using liquids of known viscosity to calculate the dynamic 6.1.4 Electronics—Capable of controlling the electromag-
viscosity and kinematic viscosity of the liquid. See Fig. 1. netic field to propel and detect the travel time of the piston with
a discrimination of 0.01 s or better and uncertainty within
5. Significance and Use 60.07 %. The travel time is calibrated to be between 0.4 s and
5.1 Many petroleum products, as well as non-petroleum 60 s, at a distance of 5 mm.
materials, are used as lubricants for bearings, gears, compres-
sor cylinders, hydraulic equipment, etc. Proper operation of
this equipment depends upon the viscosity of these liquids. 5
The Oscillating Piston Viscometer is covered by a patent. Interested parties are
5.2 Oscillating piston viscometers allow viscosity measure- invited to submit information regarding the identification of an alternative to this
ment of a broad range of materials including transparent, patented item to the ASTM International headquarters. Your comments will receive
careful consideration at a meeting of the responsible technical committee, which
translucent and opaque liquids. The measurement principle and you may attend.
stainless steel construction makes the Oscillating Piston Vis- 6
The sole sources of supply for the apparatus known to the committee at this
cometer resistant to damage and suitable for portable opera- time is Cambridge Viscosity Inc., 101 Station Landing, Medford, MA 02155
(www.cambridgeviscosity.com). If you are aware of alternative suppliers, please
tions. The measurement itself is automatic and does not require
provide this information to ASTM International Headquarters. Your comments will
an operator to time the oscillation of the piston. The electro- receive careful consideration at a meeting of the responsible technical committee,
magnetically driven piston mixes the sample while under test. which you may attend.
2
D 7483 – 08
6.1.5 Temperature Controlled Jacket—Sufficient for main- 7.2.1 The certified viscosity reference should have a pub-
taining measurement chamber temperature within 60.06°C. lished viscosity in accordance with Test Method D 445 or
6.1.6 Temperature Measuring Device—Industrial platinum equivalent means that is close to that of the liquids being tested
resistance thermometer (IPRT) or equivalent sensor with a at the test temperature. For example, if intended measurements
maximum permissible error of 60.02°C. It is recommended, are to be made from 5-25 mPa·s at 100°C, then a reference oil
that the temperature measuring device be verified with an viscosity of 15 mPa·s at 100°C would be appropriate.
independent, calibrated temperature probe at the test tempera- 7.3 Cleaning solvents miscible with the sample and chemi-
ture. cally compatible with the wetted viscometer components (such
6.2 Temperature Regulation System: as alcohols, toluene, etc.). These wetted components are
6.2.1 Any liquid bath or thermoelectric means for regulating typically 316L and 430 Stainless Steel.
the jacket temperature. 7.4 Quality control (QC) liquid similar to 7.1, but with
6.2.2 The temperature control must be such that the tem- viscosity values internally certified as noted in 12.2.
perature of the measurement chamber is held within 60.06°C
of the desired measurement temperature. 8. 8. Sampling, Samples, and Test Units
6.3 Sample Introduction Mechanism—A syringe, micropi- 8.1 Ensure that the sample is homogenous. Engine sampling
pette, or flow-through adapter for introducing between 3.2 mL is generally specified in the test method, for example Test
and 5 mL, inclusive by pressure, into the measurement Method D 5967. When applicable, refer to Practice D 4057
chamber. (manual) or Practice D 4177 (automatic) for proper sampling
7. Reagents and Materials techniques.
7.1 Certified viscosity reference standards shall be certified
by a laboratory that has been shown to meet the requirements 9. Preparation of Apparatus
of ISO/EC 17025 by independent assessment. Viscosity stan- 9.1 Place the viscometer on a stable surface.
dards shall be traceable to master viscometer procedures 9.2 Operate the unit according to the procedure in Section
described in Practice D 2162. 13.
7.2 The uncertainty of the certified viscosity reference 9.3 Verify calibration accuracy by testing a reference stan-
standard shall be stated for each certified value (k = 2, 95 % dard or QC liquid at the test temperature. Follow the procedure
confidence). See ISO/EC 17025 or NIST TN 1297. in Section 13.
3
D 7483 – 08
10. Calibration and Standardization 13.5.1 Allow the sample to equilibrate in the measurement
10.1 Calibrate according to manufacturer’s instructions to chamber for at least 2 min while the piston is oscillating.
obtain a calibration curve (using two test liquids with refer- 13.5.2 Ensure that temperature stability is within the crite-
enced viscosity values near, but within, the extremes of the rion set in 6.2.
piston range being used). 13.5.3 Measure the upward and downward piston travel
10.2 Certified Viscosity Standards may be used as confir- times for each cycle and compute viscosity until the standard
matory checks on the procedure in the laboratory. This proce- deviation as percent of mean, over the previous 20 dynamic
dure is outlined in Section 13. If the dynamic viscosity result, viscosity computations, is less than 0.6%.
at the calibration test point, does not agree with the certified 13.6 Record the average result from the last 20 computa-
value within the limits of precision in Section 16, each step in tions.
the procedure should be rechecked, as well as the temperature
measuring device and viscometer calibration, to locate the 14. Calculation and Interpretation
source of error. If the source is not detected, consult the
manufacturer. 14.1 The calculation of dynamic viscosity and kinematic
viscosity are computed and displayed automatically by the
11. Sample Conditioning apparatus.
11.1 Shake all new and used oil samples using the following
procedure. 15. Report
11.1.1 Ensure cap is tight on the container.
15.1 Dynamic Viscosity result in mPa·s to three significant
11.1.2 Shake vigorously by hand for 30 s. Wait 10 s, or
figures, as displayed by the apparatus.
longer if needed, for air bubbles to dissipate.
11.1.3 A specimen of the sample shall be taken by pipette, 15.2 Kinematic Viscosity result in mm2/s, to three signifi-
pouring or pumping. Suspected nonhomogeneous samples cant figures, as displayed by the apparatus.
must be conveyed for analysis promptly following the shaking 15.3 Temperature in degrees Celsius, to the second decimal
and dissipation procedure of step 11.1.2. place, as displayed by the apparatus.
13. Procedure NOTE 1—This provisional repeatability was obtained from a sample set
13.1 Verify or set the temperature control settings, as tested of ten used and ten fresh oils measured by seven laboratories at 40 and
with the control standard, so the viscometer temperature reads 100°C. Certified Reference Standards were included among the samples
set.
the desired set point 60.06°C while the piston is in motion.
13.2 Remove the piston and clean the specimen from the 16.1.2 Reproducibility—The reproducibility of this test
measurement chamber as described in the viscometer manual. method is being determined and will be available on or before
13.3 Load the measurement chamber with sample using the December 14, 2009.
volume listed in Table 1 related to the piston size being used 16.2 Bias—The average results for the standard liquid used
for the viscosity range anticipated. To minimize contamination, to develop the repeatability statement agreed with the reference
and if sample volume allows, pre-wet the chamber and piston value within statistical error (see Table 2).
with the sample material and dry wipe with a lint free cloth.
13.4 Load the measurement chamber with a clean piston. 17. Keywords
The piston size should be selected such that the measured
viscosity is between the minimum and maximum viscosity 17.1 dynamic viscosity; kinematic viscosity; oscillating pis-
values listed in Table 1. If the reported result is outside of this ton; oscillating piston viscometer; viscosity
range, the measurement shall be repeated using the appropriate
piston size.
13.5 Start the reporting software in accordance with manu- 7
Supporting data have been filed at ASTM International Headquarters and may
facturer’s instructions, which in turn will: be obtained by requesting Research Report RR: D02–1657.
4
D 7483 – 08
TABLE 2 Measurement Bias
Viscosity in mP•s
ILS Average Value in mP•s 1.420 3.077 4.693 9.661 15.36 85.19 155.9
Certified Reference Value 1.434 3.098 4.643 9.671 15.12 86.05 154.4
Viscosity in mm2/s
2
ILS Average Value in mm /s 1.856 3.869 5.747 11.71 18.16 18.88 99.76 184.8
Certified Reference Value 1.853 3.901 5.706 11.83 18.00 18.05 100.8 180.0
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).