TITLE: Resetting X/Y step size calibration parameters


Category: XLp series GERBERplotter

Document Number: 2011-GT-001

Supplied by: Ronny Sprangers

Last Modified: 8 February 2011

Summary: How to reset the x/y step size calibration parameters in XLp
Interface

When doing the X/Y step size calibration it can happen that the value’s on the screen
get completely out of range causing the plotter to behave strangely. This usually
happens if one does not follow the correct procedure and keep changing the values
without updating the plotter parameters.

Sometimes these values can be so far out of range that you can not recover easily.

When entering the service level from inside the calibration tab, a reset button will
appear in the bottom right of the calibration screen allowing the X/Y values to be reset
to 1 (factory default)

Entering Service Level

The Service Level


Password is the serial
number of the unit with 1
added to each digit.
In this case: serial
number 2041100805

password 3152211916

In case there is a 9 in the


serial number, this digit
becomes : and not 0