IVD - Workflow

COBAS® AmpliPrep Instrument - COBAS® TaqMan® Analyzer


COBAS® AmpliPrep Instrument - COBAS® TaqMan® 48 Analyzer
Objectives

Upon completing this section, the trainee should be able to:

• Know the configuration and content of the Test Kits.

• Know and understand different workflow configurations using the COBAS®


AmpliPrep Instrument and the COBAS® TaqMan® / COBAS® TaqMan® Analyzers.

• Perform the complete workflow including the process of generating results.

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 2
Generic IVD Kits
Contents

Sample Preparation Reagents


• Bead Reagent Cassette (CS1)
• Lysis Reagent Cassette (CS2)
• Multi Reagent Cassette (CS3)
• Test specific Cassette (CS4)
Control Reagents
• 4 or 6 Negative control (- Negative)
• 4 or 6 Positive controls (# High positive)
• 4 or 6 Positive controls (+ Low positive)
Control Barcode Clips
• 4 or 6 for each control
Documents
• Package Insert, Blue Card
Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 3
Generic IVD Kits
Storage and Working temperature

• Store at 2-8°C

• Insert immediately from


refrigerator into COBAS®
AmpliPrep Instrument and
equilibrate in the instrument for
30 min

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 4
Generic IVD Kits
Cassettes contents (1)

BD
Cassette 1: Beads
– MGP in Isopropanol

Cassette 2: Lysis reagent


LYS
– Guanidine Thiocyanate

P Cassette 3: Multi-Reagent Cassette


– Pase: 8% Proteinase
– EB: Tris Buffer with 0.2% Methylparaben
EB
Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 5
Generic IVD Kits
Cassettes contents (2)

Cassette 4: Test specific Reagent Cassette


QS – QS
• Non-Infectious transcribed RNA (HCV, HIV),
armored with MS2-phage coat protein
• non-infectious DNA construct (HBV)
Mn
– Mn2+
• Mn-Acetate
– MMX
MMX
• Z05 DNA Polymerase
• Primers
• Fluorescent-labeled probes
• dNTPs
• AmpErase (Uracil-N-Glycosylase) enzyme
• Aptamer
Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 6
Easy begins here.
CAP/CTM System configuration

Software: control of the process

Samples Nucleic Acids + Master Mix Results

Integration : instruments, consumables, reagents, processes.

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 7
Easy begins here.
CAP/CTM Docked System configuration

Software: control of
the process

Samples
Results

Integration : instruments, consumables, reagents, processes.

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 8
Easy begins here.
CAP/CTM 48 System configuration

Software: control of the process

Samples Nucleic Acids + Master Mix Results

Integration : instruments, consumables, reagents, processes.

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 9
1. Instrument Startup

Log On to Datastation and AMPLILINK Software

• It is recommended to shut down and restart the datastation weekly.

• Log on by entering your assigned AMPLILINK user ID and password.

Inspecting and maintenance of the COBAS® AmpliPrep and COBAS®


TaqMan® / COBAS® AmpliPrep and COBAS® TaqMan® 48 Instruments

• Perform daily maintenance (if necessary also periodic maintenance)

• Check for wash reagent supply on the COBAS® AmpliPrep Instrument

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 10
2. Handle Reagents

• Remove reagents from storage


• Load immediately
(before condensation can occur Cassettes 2 – 4
and reagents need to equilibrate
for at least 30 minutes before
use)

Cassette 1: MGP

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 11
2. Insertion of Reagents Racks

Reagents Rack 1 insert


Cassette 1: Beads

Reagents Rack 2 insert


Cassettes 2 – 4

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 12
3. Sample handling
Remove Samples from Storage & prepare S-tubes

Storage recommendation:

• 1100 – 1200 µl-Aliquots of Plasma


(store according to the recommendation in the Package Insert)

• Minimal (maximal) S-tube Input Volumes:

HBV v1.0 HBV v2.0


HCV v1.0
HIV v1.0/v.20
Input 1000 (1050) µL 650 (1050) µL
Volume
Processing 850 µL 500 µL
Volume

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 13
4. Load Consumables (I)

K-tips (full rack)


(GMMI#: 03287343 001)

• K-tubes
– CAP-CTM 48 on sample rack (performed
by operator)
– CAP-CTM via OT-Platform (full rack)

(GMMI#: 03137082 001)

• Flapless SPUs (loaded onto the SPU-rack)


(GMMI#: 03755525 001)

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 14
4. Load Consumables (II)
SPU-loading

• Load SPUs correctly onto the SPU-


rack

• The SPU-rack can also be partially


filled
Instrument will count the SPUs via
S-tip Light-barrier

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 15
4. Load Consumables (III)
K-carrier

COBAS® AmpliPrep/ COBAS® TaqMan® System COBAS® AmpliPrep/ COBAS® TaqMan® 48 System

K-carrier has to be loaded manually onto the COBAS®


AmpliPrep Instrument
K-carriers are loaded onto the COBAS® TaqMan®
Analyzer Necessary tools:

K-carrier-rack

Bar-coded K-carrier

K-carrier transporter

K-carrier holder
Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 16
Pre-preparation of K-carrier rack (I)
Ordering-Info – Taqman 48

• K-carrier rack and Labels have to be ordered


separately it’s not part of any Start-Up-Kit

– Order-Info: 2) Pins
• K-carrier rack (contains also the pins)
GMMI # 03286436 001
• K-carrier rack label
GMMI #:03287335 001

• The rack has to be prepared before the first


use as following: K-carrier Rack
1) Stick barcode label onto the rack 1) Labels
2) Add pins to the correct position
according to the label number (use the
binary-format)

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 17
Pre-preparation of K-carrier rack (II)
Binary code of K-carrier rack

20 = 1 20 + 21 =
20 = 1 21 = 2 1+2=3
21 = 2
22 = 4
Rack No 1 Rack No 2 Rack No 3
23 = 8
24 = 16
25 = 32
22 = 4 20 + 22 = 21 + 22 =
26 = 64
1+4=5 2+4=6
27 = 128
Rack No 4 Rack No 5 Rack No 6
Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 18
5. Associate K-carrier with K-carrier rack (I)
Only COBAS® AmpliPrep/COBAS® TaqMan® 48 Analyzer

• Place a K-carrier on a K-carrier rack (1)


(nothing else !!!) (1) Pins

• Scan the barcode of K-carrier rack &


within 5 sec scan the barcode of K-
carrier using handheld barcode- K-carrier Rack
scanner Labels

• Note: the association (scanning) of the


K-carrier with the K-carrier rack can 2(2)
only be done under the instrument
system window (2)

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 19
5. Associate K-carrier with K-carrier rack (II)
Only COBAS® AmpliPrep/COBAS® TaqMan® 48 Analyzer

• After scanning the K-carrier rack


barcode a pop-up-window opens:

• Scan the barcode of the K-carrier


within 5 seconds.

Rack No
Binary code
of Rack No

Rack No

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 20
5. Associate K-carrier with K-carrier rack (II)
Only COBAS® AmpliPrep/COBAS® TaqMan® 48 Analyzer

• Load K-carrier rack onto the OT-platform


of the instrument

• The K-carrier rack occupies 2 slot positions


on Inserts M-P

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 21
6. Creating Orders
Using the Sample rack Tab

1. Select „New“

2. Type in respective Sample Rack ID


2 1
3
3. Select desired Test Icon
(Check or uncheck „Docked“) 6
4
4. Assign Controls to desired position
1. HPC
2. LPC
3. NC 5

5. Assign/enter patient samples to


desired position
3
6. Save your order

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 22
7. Sample & Control Handling

• Attach barcode clips for:


– controls (Kit specific)
– and samples (provided via Barcode clip
K-tube
the S-tube input box)

• Place S-tubes input in the S-tube


Sample Rack

• Vortex & pipette samples &


controls into the S-tubes input

• Place K-tubes on Sample rack


(only when using COBAS
AmpliPrep/COBAS® TaqMan® 48
analyzer configuration)

Controls are pipetted into S-tubes


Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 23
8. Load Sample Racks

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 24
8. Check Instrument Status (I)
for missing resources

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 25
9. Check Instrument Status (II)
…Start the Run

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 26
10. Finished Instrument Run
Transfer of the K-tubes

Docked system Undocked system COBAS® TaqMan® 48

•Transfer of K-carrier •Transfer Sample rack • Transfer K-carriers


via Docking-Station •Transfer must be done • Balance K-carriers
within 120 min (if needed)
• Transfer must be done
within 120 min

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 27
Positioning scheme of K-carrier
For all 3 workflows

1 = 1st sample processed from the


sample rack
23 19 11 15 22
2 = 2nd sample processed from the
sample rack
13 5 3 8 18 3 = ……

9 1 2 10
17 7 4 6 14
21 16 12 20 24
Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 28
11. Results (I)
Where to find the results

Access the results using


the results icon/menu.

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 29
11. Results (I)
Review and Accept/Reject Results

• Double click on one


sample to see the report
windows.

• Accept/reject the
results as batch or
individually (marked with
a black asterisk).

• Only accepted/rejected
results can be archived
and purged.

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 30
12. Remove Instrument Waste…
From the COBAS® AmpliPrep Instrument

COBAS® AmpliPrep Instrument

• Remove reagent racks and store remaining reagents in refrigerator

• Remove used SPU racks

• Remove empty K-tip / K-tube racks

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 31
12. Remove Instrument Waste…
From the COBAS® TaqMan® / TaqMan® 48 Analyzers

COBAS® TaqMan® Analyzer

• Remove Sample Racks

• Remove K-tubes from waste box & discard accordingly

COBAS® TaqMan® 48 Analyzer

• Remove K-carrier from COBAS® TaqMan® 48 Analyzer & discard K-tubes


accordingly

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 32
Summary: COBAS® AmpliPrep Instrument loading configurations

Workflow
C1 C2, C3, C4

Docked Beads Reagents S tubes SPUs K tubes, K tips K carriers

CTM96 Beads Reagents S tubes SPUs K tubes1, K tips

CTM48 Beads Reagents S tubes SPUs K tips,


K tubes2 K carrier rack
introducing: K carriers
TNAI Beads Reagents S tubes SPUs Output S tubes, K tips
1 After sample processing, the used K tubes go to the sample racks and are ready for manual transfer
2 After sample processing, the K tubes are placed in the K carrier on the K carrier
Roche Molecularrack
Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 33
Sampling COBAS TaqMan
Handling Area Analyzers AMPLILINK
COBAS AmpliPrep Software
Instrument

Daily Maintenance Daily Maintenance Start up Datastation

Load reagents (from fridge)

Remove samples from


storage
Load consumables
Sample handling Create order
Attach barcode clips Prepare/print worklist

Sample handling
•Vortex/pipet samples &
controls Associate K-carrier with K-
•Fill and place S-tubes in carrier-Rack
the sample rack
•Place K-tubes on Sample Transfer
rack Check instrument status and
start run.

Review and accept


results.
Remove instrument waste Remove instrument waste

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 34
Features and Benefits of COBAS® AmpliPrep /
COBAS® TaqMan®
• Reduced Hands-On Time, Improved Lab Productivity
– The only fully automated system with sample-in, sample-out capability with no manual intervention required
during a run.

• Greater Flexibility and Rapid Turnaround Time


– Multiple test types can be run simultaneously and runs can be loaded continuously throughout a work shift.

• Simplified Workflow and Inventory Management


– Reagents are packaged in complete kits (stored at fridge temperature) consisting of multiple-use, bar-coded,
ready-to-use cassettes, requiring no preparation before use or manual assay calibration.

• Greater Confidence in Result Integrity


– Fully competitive quantitation standard, AmpErase (Uracil-N-Glycosylase) contamination control and closed-
tube format reduce the risk of contamination and ensure result integrity.

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 35
Features and Benefits of COBAS® AmpliPrep /
COBAS® TaqMan®
• Constant Innovation in Assay Development
– Roche, through its Global Surveillance Program, is a global leader in monitoring viral genome sequence
changes. This allows for continual adaption of assays to the changing nature of viral genomes, ensuring assays
remain reliable and effective.

• Broad Test Portfolio


– HIV-1 qualitative, HIV-1 viral load, HBV viral load, HCV viral load, CMV viral load.

• Easy-to-Use Software
– AMPLILINK software provides a simple and efficient user interface to the custom-built, Windows®- based PC
supplied with the system. The system has the capability for bidirectional interfacing to laboratory information
systems (LIS).

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 36
IVD Workflow
Version History
Version Date Modifications
1.0 March 2005 Original File
1.1 March 2006 • P15: new slide, Contains information on K-carrier, K-carrier Holder
and K-carrier Transporter

• P17: new slide, Contains information that K-carrier rack and labels
have to be ordered separately

1.2 May 2006 • P4: changed from weekly shutdown to shut down and restart the
datastation daily

1.3 December 2007 •Replaced AMPLILINK 3.1 with AMPLILINK 3.2 Series printscreens
(added slide 22)
• P3: Restricted area for loading reagents on CAP
• P3: Added “Print worklist” to point E
•Include MMix components and description
•daily maintenance: removed the statement using the Service Wizard

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 37
IVD Workflow
Version History

Version Date Modifications


•Combined from two previous workflow presentations (versions Workflow_CAP-
CTM48_DRAFT V1.4_Dec_2008.ppt and Workflow_CAP-CTM_DRAFT V1.3_Dec_2008.ppt
2.0 November 2009 •Combined information of COBAS TaqMan and COBAS TaqMan 48 into one presentation.
•Updated content to reflect the AMPLILINK 3.3 software (print-screens)
•Restart of datastation weekly instead of daily

Roche Molecular Diagnostics Version 2.0 Workflow COBAS® AmpliPrep Instrument with
November 2009 COBAS® TaqMan® Analyzers
Page 38
We Innovate Healthcare