CTM - Result - Interpretation (Compatibility Mode)
CTM - Result - Interpretation (Compatibility Mode)
3. “Aborted”
Aborted order after processing has started, no flag remark will be displayed
4. “>[TiterMax]”
Elbow in range, but calculated titer is above the range defined for the test.
or
Elbow below range, result is set to >[TiterMax] without calculation.
5. “[Titer]”
Elbow in range and calculated titer within range.
6. “<[TiterMin]”
Elbow in range, but calculated titer below the range defined for the test.
7. “Target Not Detected”
Result is below the detectable limit.
Roche Molecular Diagnostics Version 2.0 Result Interpretation
November 2009 Page 8
Range Checks
ElbowMin ElbowMax
Elbow
TiterMax TiterMin
Titer
June 2014 / Version 1.0 DRAFT Global Customer Support IVD – Portfolio and Clinical Information
Page 10
HIV Clinical Information
HIV pathogenesis: Viral replication drives
disease progression
June 2014 / Version 1.0 DRAFT Global Customer Support IVD – Portfolio and Clinical Information
Page 11
HIV Clinical Information
HIV pathogenesis: Anti-viral drugs allow the
immune system to recover
June 2014 / Version 1.0 DRAFT Global Customer Support IVD – Portfolio and Clinical Information
Page 12
Result Interpretation
(AMPLILINK
display)
Target Not
Target Not Detected Target Not Detected Report results as ‘HCV RNA not detected’
Detected
< 1.5E+01
Titer < Limit of Detection < 1.5E+01 IU/ml Target DNA or RNA detected, below limit of detection
IU/ml
Parameter HBV
No growth curve Target Not Detected
Titer < [TiterMin] < 6 IU/ml
Example for
[TiterMax] > 3.29E+4 IU/ml
Titer> [TiterMin]
Titer> [TiterMax] > 1.10E+8 IU/ml
Lower Limit of
29 IU/ml
Linear Range
Target Elbow Min 9
Target Elbow Max 50
QS Elbow Min 20
QS Elbow Max 50
An invalid result Invalid
1.4 April 2007 Remove the sample rack picture from the “Components: Status Mapping” slide.
DRAFT This is only applicable for the COBAS® TaqMan® /TaqMan48 TC.
1.5 July 2007 Removed COBAS® TaqMan® & TaqMan 48 from title.
DRAFT
„Objectives“: removed „Be familiar with result calculation“
„Results“: added „INVALID“ and „Failed“ to the possible results
„Result Interpretation“: added disclaimer regarding result interpretation between
LOD and LLoLR.
1.5 July Added Slide “Result Interpretation - _S_ ABOVE_RANGE / _S_ BELOW_RANGE”
2007
Added Slide “Result Interpretation -RFITOOLOW Example”
Removed „Result Summary“ Slide
Removed „Detail Flag Display“ Slide
1.6 DEC Added slide „High titer Samples PI note“
2007
Slide 4: only HBV test has results needs users‘ judgement
Inserted Slide „Result Interpretation for High Titer Samples“ for more clarity
1.7 SEP Removed special result cases: “Target Not Detected” with target Ct in range, DRIFT High,
2008 Result interpretation high titer specimens. All these were moved to „Flag Interpretation“
Presentation
Inserted Ct/titer range explanation slide
Completed example slides for all result cases
Replaced 3.1 print-screens with 3.2 print-screens
Rearranged slide order (more logical order)
Inserted two slides with US IVD result interpretation for HPS and CAP CTM
1.8 FEB Added QS Ct range for each test.
2009 Added updated information regarding CAP CTM HIV 2.0 and HBV 2.0
2.0 NOV •Updated content with AMPLILINK 3.3 Software screenshots.
2009 –Controls are flagged if not in range (no titer display anymore)