eCL8000 maintenance package

Unit Total
No. Item Part Number Name Picture Unit
Price($) Price
Quarterly Silicon tube(Pinch
1 27-0040000-00078 4 10 40
maintenance valve hose)

Total Price: 40

Annual
1 04-I05A000-00023 Measuring cell 1 860 860
maintenance

Total price: 860

Optional Accessories
Cleaning
1 solution(alcohol or 1
water)
2 lubricating oil 1

Needle (To clean the


3 blockage inside 1
probe)

P.S.: Recommend to check the machine with this maintenance package every
3 months

Method to using them:


1. Quarterly maintenance:
① Replace silicon tube and reset count of silicon tube;
② Doing measuring cell maintenance in service interface, this procedure will take about 15
minutes
2. Annual maintenance:
Replace Measuring cell and reference electrode every year or over 20000 tests; After replace,
remember to reset the count

Optional Accessories:
1. Cleaning solution: Used to clean the outer wall of the probe and machine surface.
2. Lubricating oil: Used to lubricate the 2D arms and 3D arms
3. Needle: This needle can use to clean the blockage inside probe and other position.
P.S.:Every engineer is recommended to prepare one set of these accessories, if it's enough, you
can choose not to include these.