You are on page 1of 10

Anal Bioanal Chem (2004) 378 : 1952–1961

DOI 10.1007/s00216-004-2516-2

PA P E R I N F O R E F R O N T

Charles R. Evans · James W. Jorgenson

Multidimensional LC-LC and LC-CE


for high-resolution separations of biological molecules

Received: 23 December 2003 / Revised: 22 January 2004 / Accepted: 27 January 2004 / Published online: 12 February 2004
© Springer-Verlag 2004

Abstract In multidimensional separations, two or more of resolution. The need for multidimensional separations
independent separation methods are coupled in an effort arises from the inability of one-dimensional (1D) separa-
to resolve complex mixtures. The displacement mecha- tion methods to adequately resolve highly complex sam-
nisms of each method should be orthogonal, such that lit- ples. One field of research that highlights the need for the
tle correlation exists between the retention of compounds enhanced separation power of multidimensional tech-
in each dimension. When multiple orthogonal separation niques is proteomics, which involves the characterization
methods are coupled such that all sample components are of all proteins expressed by an organism. For example, the
subjected to complete analysis on all dimensions, the simple single-celled organism E. coli produces a total
method is considered “comprehensive”. The primary ad- complement of over 4,000 proteins [1]. The most power-
vantage of comprehensive multidimensional separations ful 1D column chromatography techniques typically have
over their one-dimensional counterparts is the potential for a peak capacity of only a few hundred [2], where peak ca-
dramatically enhanced resolution. High resolving power pacity is defined as the maximum number of peaks that
can be achieved because the peak capacity of a compre- can fit into the accessible separation space side by side
hensive multidimensional separation is roughly equal to with a resolution of 1.0 from each neighboring peak [3].
the product of the individual peak capacities of each di- Moreover, due to the usually random distribution of peaks
mension. In this review, the theory and instrumentation of and the statistical probability of peak overlap, the actual
two-dimensional liquid chromatography (LC-LC) and liq- number of detectable components that can be expected to
uid chromatography-capillary electrophoresis (LC-CE) sep- be fully resolved is substantially smaller than the peak ca-
arations are discussed. Some applications of these tech- pacity [4, 5, 6]. The disparity between the capabilities of
niques to the separation of biological molecules are high- traditional 1D separations and the complexity of real-
lighted. Future directions for the development of multidi- world samples has driven researchers to push toward mul-
mensional separations are also considered. tidimensional separations, due to the greatly enhanced re-
solving power they offer.
Keywords Multidimensional separations · Liquid
chromatography · Capillary electrophoresis · LC-LC ·
LC-CE · Biomolecules Multidimensional separation theory

In a multidimensional separation, a sample is first sub-


Introduction jected to separation via one method, and then the sepa-
rated components are further separated by at least one ad-
A multidimensional separation involves the coupling of ditional independent method. Although there is no inher-
two or more separation mechanisms in order to carry out ent limitation to the number of independent separation
a single analysis. Although simple in concept, multidimen- methods that can be coupled, practical constraints have
sional separations can be exceedingly powerful in terms limited the vast majority of the multidimensional separa-
tions reported to date to two dimensions. Two-dimen-
sional (2D) separations provide enhanced resolution and
peak capacity as compared to 1D separation systems. In
C. R. Evans · J. W. Jorgenson (✉) an ideal case, the total peak capacity of a 2D separation
Venable and Kenan Laboratories, Department of Chemistry,
University of North Carolina at Chapel Hill,
system is approximately equal to the product of the peak
Chapel Hill, NC 27599, USA capacities of both dimensions [4]. Hypothetically, if two
e-mail: jj@unc.edu techniques with a peak capacity of 100 each were com-
1953

bined, the resulting 2D system would have a maximum


theoretical peak capacity of 10,000. Achieving peak ca-
pacities this high using 2D separations is not a trivial task,
however, due to the inherent difficulties in coupling two
dissimilar techniques.
Due to the large number of different types of 2D sepa-
rations reported in the literature, it is important to make
clear distinctions between the different classes of 2D sep-
arations. One important concept is the difference between
separations that are 2D-in-space versus 2D-in-time. A sep-
aration which is 2D-in-space results in a separation of an-
alytes over a two-dimensional planar surface, such as a Fig. 1a,b Hypothetical orthogonal (a) and nonorthogonal (b) 2D
gel. Techniques that are 2D-in-time are carried out by first separations (Adapted from Ref. [35])
performing a 1D separation, often on a LC column, and
then sequentially subjecting individual fractions from this
first-dimensional separation to a second 1D separation. other dimension [8]. Hypothetical examples of orthogonal
Among techniques that are 2D-in-time, another distinc- and nonorthogonal 2D separations are shown in Fig. 1.
tion must be made between off-line and on-line multidi- Giddings’ second criterion states that no resolution
mensional separations. Off-line methods involve collect- gained in the first dimension of separation may be lost in
ing fractions from the first dimension, then later subject- any subsequent dimension. For on-line, 2D-in-time sepa-
ing them to separation on the second dimension. On-line rations, this requirement is satisfied by designing instru-
techniques employ switching valves or other instrumenta- mentation that transfers fractions from the first dimension
tion that allow the fractions from the first dimension to be to the second dimension at a rate sufficient to ensure the
transferred directly to the second dimension, usually while resolution of the first dimension is not substantially di-
the first-dimensional separation continues simultaneously. minished. Therefore, the second dimension must be capa-
Both off-line and on-line methods have notable advan- ble of completing separations very rapidly compared to
tages and disadvantages. Off-line fraction collection al- the first dimension. In order to achieve optimal resolution,
lows for easier optimization of the separation in both di- each peak in the first dimension should be sampled at
mensions and permits sample manipulations, such as con- least three times by the second dimension [9]. The result-
centration or reconstitution, between dimensions. The abil- ing need for high-speed analyses in the second dimension
ity to optimize each step of the separation independently represents an experimental challenge in carrying out 2D
causes the entire analysis to take a substantial period of separations. In spite of this and other difficulties, 2D sep-
time, which can become prohibitive when collecting large arations that satisfy Giddings’ criteria and produce good
numbers of fractions unless automated sample-handling separations of complex mixtures have been demonstrated.
methods are used. On-line techniques tend to be substan-
tially faster because all the second-dimension separations
are run within the time that the first-dimension separation 2D gel electrophoresis versus LC-LC and LC-CE
takes to complete. However, run times and mobile phase
compositions tend to be more limited than they are with Before discussing 2D LC-LC and LC-CE, which are both
off-line 2D separations. Instead, compromises in both di- 2D-in-time techniques, it is worthwhile to consider a 2D-
mensions must usually be made to allow successful on- in-space technique known as two-dimensional polyacry-
line coupling. Although both off-line and on-line techniques lamide gel electrophoresis (2D-PAGE). At present, most
have merit, the majority of recent efforts to develop 2D separations of complex mixtures of proteins are carried
separations have been focused on on-line methods, which out using 2D-PAGE, which was first reported in 1975 by
are the primary subject of this review. O’Farrell [10]. In 2D-PAGE, proteins are resolved into in-
Certain theoretical considerations apply to all multidi- dividual spots on a flat gel using two different separation
mensional separations, regardless of whether a technique mechanisms. Proteins are separated first according to
is 2D in space or time, off-line or on-line. Giddings estab- charge via isoelectric focusing (IEF), and then according to
lished two such fundamental requirements for ideal multi- molecular size via sodium dodecyl sulfate polyacrylamide
dimensional separations [4, 7]. First, the separation mech- gel electrophoresis (SDS-PAGE). Due to the high resolu-
anisms must be orthogonal. This means that the chemical tion and orthogonal nature of both separation mechanisms,
or physical property by which analytes are separated must at least 2,000 proteins can be resolved in a single sample
be different for each dimension. For example, reversed- [11]. 2D-PAGE also satisfies Giddings’ criteria for multi-
phase liquid chromatography (RPLC), which separates based dimensional separations. In spite of its power, 2D-PAGE
on hydrophobicity, is orthogonal to ion-exchange chro- has certain fundamental limitations. It is useful only for
matography (IEC), which separates according to coulom- proteins and is a labor-intensive analysis that usually re-
bic interactions. If a 2D system is truly orthogonal, the dis- quires at least two days to complete. It cannot be coupled
tribution of components (peaks) in one dimension should in an on-line fashion to mass spectrometry (MS), thereby
not correlate with the distribution of components in the requiring further manual intervention or complex robotics
1954

systems in order to implement MS analyses. Most signifi-


cantly, 2D-PAGE is only semiquantitative and may not
separate all the components present in a given sample;
proteins that are extremely large or hydrophobic may not
enter the gel, very acidic or basic proteins may be poorly
resolved, and low abundance proteins may be below the
detection limit of the silver-staining techniques used in
2D-PAGE [2]. Fig. 2 Schematic diagram of a directly coupled-column 2D LC in-
Many of the limitations of 2D-PAGE are not shared by strument. Columns 1 and 2 represent two distinct, orthogonal sep-
2D separation methods based on liquid chromatography aration modes
(LC) and/or capillary electrophoresis (CE). When appro-
priate separation modes are selected, LC and CE can sep-
arate a broader range of biomolecules than 2D-PAGE, in- Directly coupled-column multidimensional LC
cluding those that are large, hydrophobic, acidic or basic.
Depending on the detection technique used, LC and CE The simplest means of performing comprehensive 2D sep-
can have greater sensitivity to analytes present in small arations is to directly couple two or more columns with
amounts than 2D-PAGE. Perhaps most importantly, sepa- orthogonal separation mechanisms in series, as shown in
ration methods such as LC and CE are instrumental tech- Fig. 2. If such a system were run using isocratic elution,
niques which are capable of full automation of sample in- this arrangement would be in violation of Giddings’ sec-
jection and interfacing between dimensions, and can be ond criterion for multidimensional separations, since ma-
coupled with a variety of detection methods such as UV terials separated on the first column could recombine on
absorbance, laser-induced fluorescence or MS [2]. Given the second dimension [4]. Furthermore, it is unlikely that
these advantages, it is not surprising that researchers have the same mobile phase would provide effective separa-
looked to 2D LC and related techniques to provide an al- tions on both columns. However, useful 2D separations
ternative to 2D-PAGE. The ongoing challenge for separa- can be carried out on certain arrangements of directly cou-
tion scientists has been to devise 2D separation methods pled columns when both dimensions are operated in a
and instrumentation that can equal or surpass 2D-PAGE in carefully controlled gradient elution mode [2]. Specifi-
resolving power, speed, and reliability. cally, fractions from the first column are eluted onto the
second column by using a stepwise gradient of increasing
eluent strength with respect to the first column. After a
On-line multidimensional liquid chromatography fraction is eluted from the first column, a continuous gra-
(LC-LC) dient appropriate to the second dimension is run to sepa-
rate and elute the components of the fraction. Then an-
Heart-cutting multidimensional LC other first-dimension gradient step is performed and the
process is repeated.
One multidimensional method that has been used com- The most prominent recent method that follows the
monly is an on-line, 2D-in-time technique called heart- scheme of directly coupled-column multidimensional LC is
cutting two-dimensional liquid chromatography [12]. In a technique known as multidimensional protein identifica-
this technique, the first-dimension separation is carried tion technology (MudPIT), developed by Yates and colleagues
out like a standard 1D analysis on a chromatography col- [14, 15, 16, 17, 18, 19, 20]. The method was developed
umn. A single desired segment of the first-column efflu- for proteomics research as an alternative to 2D-PAGE.
ent is then transferred, typically via a switching valve, to This technique follows the general scheme of 2D directly
a second column for further separation via a different coupled-column separations just described, although in-
method. Heart-cutting is useful when extra resolution is stead of two separate columns in tandem, a single micro-
needed to examine a small segment of peaks in a complex capillary column packed with two different stationary phases
chromatogram, but it requires foreknowledge of the sam- is used. The microcapillary column is a packed fused-sil-
ple’s composition. Since heart-cutting does not subject the ica capillary with a pulled tip at the outlet end for direct
entire sample to 2D separation, it is of little use when coupling to MS instrumentation via electrospray ionization.
every component of a sample is an analyte of interest, as The end of the capillary nearest the pulled tip is packed
in proteomics. Heart-cutting multidimensional chromato- with several centimeters of reversed-phase material, fol-
graphy has been reviewed elsewhere in the literature [12]; lowed by several centimeters of strong cation-exchange
this method of 2D separations is not considered further in (SCX) material. The column is connected to a PEEK mi-
this report. The alternative to heart-cutting is known as crocross which splits the flow from a conventional quater-
“comprehensive” multidimensional separations, because nary LC pump, reducing it to 0.15–0.25 µL/min, and also
all sample components are subjected to displacement on provides a connection where the electrospray voltage is
both dimensions [13]. Comprehensive multidimensional applied, as shown in Fig. 3. The 2D separation is carried
separations are more powerful than heart-cutting separa- out using a multistep gradient elution profile, an example
tions because they provide greater peak capacity. of which is shown in Fig. 4. To elute fractions of the sam-
ple from the ion-exchange portion of the column onto the
1955

gle sample. Washburn and colleagues applied MudPIT to


analysis of the proteome of the yeast S. cerevisiae and
identified 5,440 peptides using MS, which were assigned via
database searching to 1,484 unique proteins [15]. Among
these were a substantial number of low-abundance and
trans-membrane proteins, which suggests that the tech-
nique is largely unbiased and gives a representative sam-
pling of the yeast proteome. Lack of bias represents a sig-
nificant advantage over 2D-PAGE, which has limited dy-
namic range and typically performs poorly with trans-
membrane proteins since they are hydrophobic and do not
easily enter into the gel.
Fig. 3 Diagram of MudPIT column and electrospray interface
(Adapted from Ref. [16]) Another report indicates that the number of protein
identifications can be further improved by using a three-
phase MudPIT column in which an additional section of
reversed-phase material is added at the inlet of the column
to provide on-line removal of salts from the sample [18].
MudPIT has also been used for quantitative studies of
proteomic samples via the use of 15N isotope labeling
[17]. An assessment of quantitative MudPIT suggests that
the technique can provide useful data about relative levels
of protein expression, but that experimental reproducibil-
ity could be improved by enhancing resolution in order to
increase the number of peptide hits per protein [20].
One of the greatest advantages of MudPIT over other
2D techniques is the relative simplicity of the instrumen-
tation needed to carry out 2D separations. Only a single
quaternary gradient LC pump, an appropriately packed
microcapillary column, and minimal interfacing compo-
Fig. 4 An example five-step MudPIT gradient profile. The bal- nents are required. However, the simplicity of this arrange-
ance of the mobile phase at any time is made up of buffer A. ment also limits its flexibility. The two separation modes
Buffer A is 5% acetonitrile (ACN), 95% H2O, and 0.02% hepta- coupled in tandem must both utilize gradient elution. Fur-
fluorobutyric acid (HFBA). Buffer B is 80% ACN, 20% H2O, and ther, the gradient in the first dimension must always be
0.02% HFBA. Buffer C is 250 mM ammonium acetate/5% ACN,
95% H2O, and 0.02% HFBA. Buffer D is 400 mM ammonium ac- run in a stepwise fashion in order to elute concise frac-
etate/5% ACN, 95% H2O and 0.02% HFBA (Adapted from Ref. tions of sample onto the second column. The resolution
[15]) contributed by the first dimension is limited by the num-
ber of fractions transferred, which is equal to the number
of steps used in the first-dimension gradient. Each step re-
reversed-phase material, a step-gradient of buffer with in- quires substantial time, typically 100 min, which includes
creasing ionic strength is used. In between each of these periods for reequilibration and washing of salts off the col-
incremental steps, a linear gradient of increasing acetoni- umn. A maximum of 15 steps in the first-dimension salt
trile content is used to separate the eluted fraction on the gradient have been reported [15], which may suggest that
reversed-phase segment of the column. The outlet of the the first dimension is undersampled in spite of long run
column is interfaced with an electrospray mass spectrom- times. Notably, the only combination of separation modes
eter. Typically the mass spectrometers used with MudPIT that has been used thus far in MudPIT is strong cation ex-
have tandem mass spectrometry capabilities to enable the change–reversed phase, perhaps because this is the only
identification of peptides via shotgun sequence analysis. arrangement where suitable gradients can be applied alter-
A database searching algorithm is then used to determine nately to two different stationary phases in a single col-
the proteins present in the sample based on sequence umn in order to bring about an effective, orthogonal 2D
matches with the peptides detected by MS. separation.
The capabilities of MudPIT as a separation method for
the purpose of proteomics have been assessed by several
studies. The peak capacity of the 2D separation carried out Column-switching multidimensional LC
in one 15-fraction MudPIT analysis was estimated as ap-
proximately 3,200 [16]. When the inherent peak capacity of Unlike directly coupled-column LC-LC, column-switch-
the mass spectrometer was included, the estimated overall ing multidimensional LC uses valves to interface the first
peak capacity increased to 23,000, which compares quite column with the second and any subsequent columns.
favorably with 2D-PAGE techniques. In practice, MudPIT This allows greater flexibility than directly coupled 2D
is capable of identifying well over 1,000 proteins in a sin- systems since, in principle, any two separation modes can
1956
Fig. 5 Schematic diagram of a
2D IEC-SEC instrument
(Adapted from Ref. [13])

be combined in any order. In general, the trade-off for this mental setup of this 2D LC system is provided in Fig. 5.
flexibility is greater instrument complexity [2]. Most The effluent of the cation-exchange column is directed to
comprehensive 2D techniques require the use of multiple a storage loop. When the valve is switched, the isocratic
LC pumps, automated switching valves, and computer- LC pump forces the contents of the loop onto the size-ex-
controlled instrumentation. In spite of the extra complex- clusion column where they are further separated. Mean-
ity, the majority of the true comprehensive multidimen- while the effluent of the first dimension is directed to the
sional separations reported to date have been based on second storage loop. The flow rates on the two columns
column-switching multidimensional LC. are selected such that the analysis time of the second di-
A wide variety of combinations of liquid chromato- mension exactly matches the amount of time it takes for
graphic modes have been used in column-switching 2D the first dimension to fill the storage loop. This allows all
LC separations, including ion exchange-reversed phase the effluent of the first column to be transferred to the sec-
(IEC-RPLC) [21, 22, 23, 24, 25, 26, 27], size exclusion- ond dimension. A computer is used to control the switch-
reversed phase (SEC-RPLC) [28, 29, 30, 31, 32], reversed ing of the valve and the acquisition of data from a UV ab-
phase-size exclusion (RPLC-SEC) [9], ion exchange-size sorbance detector at the end of the second column. In this
exclusion (IEC-SEC) [13], and normal phase-reversed manner the SEC column can analyze a large number of
phase (NPC-RPLC) [33]. Although many different instru- fractions from the ion-exchange column during the course
mental setups for 2D LC have been designed, most of the of a run lasting a total of 2.5 to 6 h. The chromatographic
components used are similar, and include an injector, two
isocratic or gradient LC pumps, a single column for the
first dimension, one or more columns for the second di-
mension, one or more computer-controlled switching valves,
and an appropriate detection system.
The first report of a 2D separation carried out using a
method resembling the comprehensive column-switching
approach was published in 1978 by Erni and Frei [28].
Their system, which was used to separate plant extracts,
consists of a size-exclusion column in the first dimension
coupled to a reversed-phase column in the second dimen-
sion. An eight-port switching valve is used to interface the
two dimensions. This configuration is very similar to
those presently used for comprehensive 2D separations,
but in this report only a total of seven fractions from the
first dimension were transferred to the second. This se-
verely reduces the amount of resolution that can be con-
tributed by the first dimension, thereby limiting the total
peak capacity of the system.
The first truly comprehensive system for 2D LC sepa-
ration was reported in 1990 by Bushey and Jorgenson
[13]. This system was used to analyze a mixture of protein
standards and a sample of human serum. A cation-ex-
change column, operated in gradient elution mode for the
first dimension, is coupled to a size-exclusion column Fig. 6 2D IEC-SEC chromatogram of protein sample. Peak iden-
used for the second dimension. As in the report by Erni tities: A glucose oxidase, B ovalbumin, C β-lactoglobulin, D tryp-
sinogen, E α-lactoglobulin, F conalbumin, G ribonuclease A,
and Frei, an eight-port switching valve is used to interface H hemoglobin; M exclusion volume “pressure” ridge, N inclusion
the two columns and allows on-line fraction transfer from volume “salt” ridge. Valve actuated every 6 min. UV detection at
the first column to the second. A diagram of the instru- 215 nm (Reproduced with permission from Ref. [13])
1957

data are presented as a three-dimensional view of a 2D


chromatogram as shown in Fig. 6. The system was esti-
mated to have a total peak capacity of approximately 130,
which is the product of the peak capacities visually esti-
mated for each dimension. Other examples of column-
switching 2D LC separations using dual storage loops
have since been reported in the literature, some with
greater peak capacity [8, 9, 21, 22]. A 2D anion-exchange
reversed-phase system that was used to analyze a tryptic
digest of reduced porcine thyroglobulin gave a peak ca-
pacity over 2,000 [25].
A different approach, in which the effluent from the
first column is not captured by storage loops but instead is
transferred directly to the head of one of two parallel sec-
ond-dimension columns, was reported in 1997 by Opiteck,
Jorgenson, and Anderegg [29]. This system was used to
analyze the fragments produced by tryptic digests of the
proteins ovalbumin and bovine serum albumin. The first
dimension of their system consisted of six size-exclusion
columns connected in series. These were coupled via a
pair of four-port valves to two parallel reversed-phase LC
columns, which serve as the second dimension. This study
was also one of the first reports of on-line MS detection
used to analyze a 2D separation. A diagram of the entire
setup is shown in Fig. 7. As sample elutes from the series
of SEC columns, it is routed to RPLC column alpha. Fig. 7 Schematic diagram of a 2D SEC-RPLC instrument with
on-line MS detection. The two four-port valves are switched si-
Since the aqueous buffers used for SEC are weak eluents multaneously (Adapted from Ref. [29])
for RPLC, the sample material is concentrated in a narrow
zone at the head of the column. When the two four-port
valves are switched simultaneously, a second LC pump ceding the first dimension for on-line sample cleanup and
starts a gradient to elute the sample from RPLC column size-selective fractionation. The theoretical peak capacity
alpha, while the effluent from the first dimension is of the entire system is estimated to be as high as 3,000.
loaded onto RPLC column beta. The effluent from the Although this method is a powerful example of a fully au-
RPLC column being eluted is directed to a UV absor- tomated technique for proteomics separations, it involves
bance detector followed by a 10:1 flow splitter and an complex instrumentation with many switching valves and
electrospray mass spectrometer. The total peak capacity of LC pumps. An additional disadvantage is that on-line
this system was estimated to be about 500. Other 2D LC- coupling of the second-dimension RPLC columns to a
LC techniques using parallel columns in the second di- single electrospray mass spectrometer is not possible since
mension have been reported, many of which replace the the effluent from two columns must be monitored simul-
two four-port switching valves with a single ten-port valve taneously.
that accomplishes the same purpose [23, 24, 27, 31, 34].
Wagner and colleagues have described a novel com-
prehensive 2D anion-exchange reversed-phase system tar- Multidimensional liquid chromatography-capillary
geted at analysis of peptides and proteins of molecular electrophoresis (LC-CE)
weight below 20 kDa [26]. The most significant new fea-
ture of their system is the use of four parallel reversed- As has already been emphasized, the best multidimen-
phase columns in the second dimension. In order to inter- sional separations are achieved when techniques with very
face all four second-dimension columns to the single first- different separation mechanisms are coupled. Liquid chro-
dimension column, a set of three ten-port valves is used. matography-capillary electrophoresis (LC-CE) is there-
At any time during an analysis, effluent from the first di- fore a logical candidate for a 2D separation method, since
mension is being deposited onto one of the RPLC columns, separations in CE are based on electrophoretic mobility, a
two of the RPLC columns are being eluted, and one is be- mechanism nearly completely orthogonal to LC tech-
ing regenerated. This parallel elution allows the gradient niques. CE is a relatively fast separation method com-
time in the second dimension to be slowed down com- pared to LC, which makes it well suited as the second di-
pared to a system using only two reversed-phase columns mension in a 2D system because of its ability to sample
[24], which substantially enhances the peak capacity of the first dimension at a relatively high frequency. Al-
the second dimension. Two parallel UV detectors are used though LC and CE are complementary separation techniques,
to sample the system. This system also uses a novel re- their dissimilarity also makes coupling them somewhat
stricted access material (RAM) column positioned pre- complicated. Elution volumes of conventional LC columns
1958

Fig. 8 Schematic diagram of a LC-CE instrument based on the and high resolution of CE, LC-CE offers substantial promise
transverse flow-gated interface (Adapted from Ref. [44]) in terms of separation power. The peak capacity of one
RPLC-CE system was estimated as 20,000 for a 5-h sepa-
ration [35].
are much greater than sample volumes of CE capillaries.
Therefore only a small fraction of the LC effluent can be
transferred to the CE dimension. Alternatively, the flow LC-CE instrumentation
from the LC dimension can be reduced by using capillary
LC columns [35]. Even when flow rates are relatively The first automated comprehensive LC-CE system was
well matched, the transfer of fractions from an LC column developed by Bushey and Jorgenson in 1990 [39]. It was
to a CE capillary is usually more complex than in LC-LC based on the coupling of a commercial microbore RPLC
methods. As in LC-LC, off-line coupling is a simple op- column with a fused-silica capillary using an electrically
tion that has been used frequently [36, 37], but it is slower actuated six-port valve with an external sample loop. This
than on-line methods and usually sacrifices resolution system was used to separate mixtures of peptide standards
from the first dimension. In order to take full advantage of and tryptic digests of ovalbumin with fluorescence detec-
both dimensions, on-line interfacing techniques must be tion. The peak capacity of the original system was esti-
employed. In spite of the difficulties, on-line LC-CE sep- mated as 420, which was an order of magnitude improve-
arations have been successfully performed and the tech- ment over either of the two dimensions alone. An improved
nique has been shown to be a powerful multidimensional version of this system was also devised which, among
separation method. other modifications, used a CE capillary with a smaller in-
Since LC-CE is not yet commonly used, the number of ner diameter to allow higher electric field strengths to be
variations of the technique and the range of its applica- applied, thereby resulting in faster separations [40].
tions are somewhat limited compared to LC-LC. To the Many improvements in LC-CE instrumentation were
present, most work with comprehensive LC-CE systems made during the 1990s. Enhanced separation efficiency was
has focused on the separation of complex biological mix- achieved by using packed fused-silica capillaries for the
tures, particularly peptides and proteins, although some LC dimension as opposed to conventional LC columns
work with smaller neutral molecules has been reported [46]. This also facilitated coupling the two dimensions be-
[38]. The two modes of LC that have been coupled most cause flow rates in capillary LC are more similar to those
often with CE are reversed-phase chromatography [39, typically used in CE. Other improvements in instrumenta-
40, 41, 42, 43, 44] and size-exclusion chromatography tion focused on the interface used to couple LC columns
[41, 45, 46]. Some other variants of CE have also been and CE capillaries. The transverse flow-gated interface,
used in 2D LC-CE separations, such as pressurized capil- first reported in 1993 [45] and improved in 1997 [44],
lary electrochromatography (pCEC) [47]. Due to the speed eliminates the need for sample loops that contribute to ex-
1959

Fig. 10 Schematic diagram of an optically gated 2D RPLC-fast


CZE system (Adapted from Ref. [42])

laser beam is focused on the CE capillary at some point


beyond the interface tee; this beam photodegrades the fluo-
Fig. 9 2D chromatoelectropherogram from a micro-RPLC-CZE rescently labeled sample passing through the capillary.
separation of FITC-tagged human urine (Reproduced with permis- When an injection is desired, a shutter is used to momen-
sion from Ref. [44]) tarily block the laser beam, allowing a small amount of un-
degraded sample to be introduced onto the capillary. This
undegraded sample is then separated and detected. Al-
tra-column broadening. In a LC-CE system using this in- though the method is limited to fluorescently labeled sam-
terface, the LC column outlet and CE capillary inlet are ples, it offers the fastest injection method of any LC-CE
positioned directly opposite to one another, separated by a interface.
narrow channel, as shown in Fig. 8. During most of a run, On-line coupling of LC-CE with MS represents a fur-
CE buffer is continuously flushed through the channel, ther enhancement in LC-CE instrumentation [43, 47]. Nu-
which carries the LC column effluent to waste. To inject a merous challenges exist in terms of coupling CE with on-
portion of the LC column effluent onto the CE capillary, line MS, however. One challenge is the selection of a CE
the flow of flush buffer is stopped and some of the LC ef- buffer that gives suitable separation performance and also
fluent is drawn onto the CE capillary by electromigration. performs well when used with electrospray ionization.
Fast, reproducible injections can be performed by using A more fundamental difficulty in interfacing CE and MS
an air-actuated switching valve to control the flow of is the fact that most mass spectrometers are not capable of
flush buffer. A 2D chromatoelectropherogram of a sample sampling a CE capillary as rapidly as would be desired to
of fluorescently labeled human urine that was separated generate true 2D separations. Peak widths in CE separa-
using a transverse flow-gated LC-CE instrument is shown tions are usually of the order of a second or less, which is
in Fig. 9. In this chromatoelectropherogram over 400 peaks an inadequate period of time for most mass spectrometers
are resolved and appear scattered randomly over the sepa- to acquire spectra over a reasonable m/z range [43]. As
ration space. This suggests that the two dimensions are or- more rapid mass spectrometers are produced, this problem
thogonal and that the system has a high total peak capac- will be ameliorated. In spite of the difficulties, the use of
ity. on-line MS with LC-CE still provides much additional in-
An even faster method of performing injections from a formation as compared to other detection methods.
LC column onto a CE capillary is provided by optical gat-
ing. In this “inverse” injection technique, a LC capillary
column and the CE capillary are coupled using a simple Future directions for multidimensional separations
interface tee, as shown in Fig. 10, so that some of the ef-
fluent of the first dimension is always being pulled onto As for any area of research as diverse and wide-reaching
the CE capillary by electromigration [42]. An intense as multidimensional separations, it is difficult to predict
1960

the future directions of the field in the long term. Never- multidimensional separations can become a routine, reli-
theless, certain trends are evident within recent publica- able technique for analysis of real-world samples.
tions that may indicate where major research effort will be
focused in the immediate future. One possibility is the de-
velopment of separation methods that employ more than References
two dimensions. Although research in this area has been
very limited due to the complexity of the instrumentation 1. Regnier FE, Amini A, Chakraborty A, Geng M, Ji J, Riggs L,
required, three-dimensional separations using a SEC- Sioma C, Wang S, Zhang X (2001) LC GC N Am 19:200–213
2. Wehr T (2002) LC GC N Am 20:954–957
RPLC-CZE system have been reported [48]. As 2D sepa- 3. Giddings JC (1967) Anal Chem 39:1027–1028
rations become faster and the instrumentation more robust, 4. Giddings JC (1987) J High Res Chromatogr 10:319-323
incorporation of additional orthogonal separation methods 5. Davis JM, Giddings JC (1985) Anal Chem 57:2168–2177
to perform n-dimensional separations will result in further 6. Davis JM, Giddings JC (1985) Anal Chem 57:2178–2182
7. Giddings JC (1984) Anal Chem 56:1258A-1270A
enhancement of peak capacity. 8. Liu Z, Lee ML (2000) J Microcolumn Sep 12:241–254
The coupling of multidimensional separations with on- 9. Murphy RE, Schure MR, Foley JP (1998) Anal Chem 70:
line MS will likely become more commonplace in future 1585–1594
years. MS provides more information than any other de- 10. O’Farrell PH (1975) J Biol Chem 250:4007–4021
11. Wehr T (2001) LC GC N Am 19:702–711
tection method available for LC-LC or LC-CE separa- 12. Cortes HJ (1990) Multidimensional chromatography: tech-
tions. Indeed, the ability to obtain mass spectra of separated niques and applications. Marcel Dekker, New York
components is essentially a prerequisite for any 2D sys- 13. Bushey MM, Jorgenson JW (1990) Anal Chem 62:161–167
tem intended for proteomics research. When tandem mass 14. Link AJ, Eng J, Schieltz DM, Carmack E, Mize GJ, Morris
spectrometry (MS-MS) is used, peptides can be fragmented DR, Garvik BM, Yates JR III (1999) Nat Biotechnol 17:676–
682
and sequence information can be obtained. Electrospray 15. Washburn MP, Wolters DA, Yates JR III (2001) Nat Biotech-
ionization (ESI) allows the direct coupling of a LC col- nol 19:242–247
umn to a mass spectrometer. This gives LC-LC a dramatic 16. Wolters DA, Washburn MP, Yates JR III (2001) Anal Chem
speed advantage over techniques such as 2D-PAGE, where 73:5683–5690
17. Washburn MP, Ulaszek R, Deciu C, Schieltz DM, Yates JR III
obtaining mass spectra involves extensive off-line sample (2002) Anal Chem 74:1650–1657
preparation. Given the versatility of MS, it is logical to 18. McDonald WH, Ohi R, Miyamoto DT, Mitchison TJ, Yates JR
expect many future multidimensional separations to in- III (2002) Int J Mass Spectrom 219:245–251
corporate on-line MS as a detection method. 19. Wu CC, MacCoss MJ, Howell KE, Yates JR III (2003) Nat
Biotechnol 21:532–538
Another trend to be anticipated is the miniaturization 20. Washburn MP, Ulaszek RR, Yates JR III (2003) Anal Chem
of multidimensional separation methods. In recent years 75:5054–5061
much attention has been given to the concept of perform- 21. Holland LA, Jorgenson JW (1995) Anal Chem 67:3275–3283
ing chemical analyses on a very small scale using mi- 22. Opiteck GJ, Lewis KC, Jorgenson JW, Anderegg RJ (1997)
crofluidics devices – the so-called lab-on-a-chip [49]. Var- Anal Chem 69:1518–1524
23. Wagner K, Racaityte K, Unger KK, Miliotis T, Edholm LE,
ious separation methods using microchips have been re- Bischoff R, Marko-Varga G (2000) J Chromatogr A 893:
ported, including free-zone electrophoresis [50, 51, 52], 293–305
open-channel electrochromatography (OCEC) [53, 54], 24. Unger KK, Racaityte K, Wagner K, Miliotis T, Edholm LE,
gel electrophoresis [55, 56, 57], and micellar electroki- Bischoff R, Marko-Varga G (2000) J High Res Chromatogr
23:259–265
netic chromatography (MEKC) [58, 59, 60]. Two-dimen- 25. Holland LA, Jorgenson JW (2000) J Microcolumn Sep 12:
sional separations using microchips have also been dem- 371–377
onstrated. In one system, a reversed-phase capillary LC 26. Wagner K, Miliotis T, Marko-Varga G, Bischoff R, Unger KK
column was coupled to on-chip CE to produce a hybrid (2002) Anal Chem 74:809–820
27. Haefliger OP (2003) Anal Chem 75:371–378
off-chip/on-chip 2D separation method [61]. Ramsey and 28. Erni F, Frei RW (1978) J Chromatogr 149:561–569
colleagues reported another system in which MEKC was 29. Opiteck GJ, Jorgenson JW, Anderegg RJ (1997) Anal Chem
coupled with open-channel electrophoresis on a single mi- 69:2283–2291
crofluidic device for a true on-chip 2D separation [62]. 30. Opiteck GJ, Jorgenson JW, Moseley MA III, Anderegg RJ
Due to their advantages over conventional 2D methods in (1998) J Microcolumn Sep 10:365–375
31. Opiteck GJ, Ramirez SM, Jorgenson JW, Moseley MA III
size and speed, on-chip multidimensional separations will (1998) Anal Biochem 258:349–361
likely gain importance for certain separation tasks. 32. Stroink T, Wiese G, Lingeman H, Bult A, Underberg WJM
Regardless of the specific developments that take place, (2001) Anal Chim Acta 444:193–203
the field of multidimensional separations seems poised for 33. Murphy RE, Schure MR, Foley JP (1998) Anal Chem 70:
4353–4360
substantial growth. There is no fundamental limit to the 34. Murahashi T (2003) Analyst 128:611–615
number or type of separation methods that can be cou- 35. Jeffery DJ, Hooker TF, Jorgenson JW (1997) Two-dimensional
pled, so substantial increases in resolving power are pos- liquid chromatography-capillary electrophoresis. In: Handbook
sible. Although many 2D separation techniques being de- of capillary electrophoresis, 2nd edn. CRC, Boca Raton, pp 765–
790
veloped at present are targeted for use in the field of pro- 36. Issaq HJ, Chan KC, Liu C, Li Q (2001) Electrophoresis 22:
teomics, many other types of samples can be analyzed us- 1133–1135
ing similar methods. In practical terms, there is also much 37. Issaq HJ, Chan KC, Janini GM, Muschik GM (1999) Electro-
room for the improvement of instrumentation such that phoresis 20:1533–1537
1961
38. Zhang X, Hu H-l, Xu S, Yang X, Zhang J (2001) J Sep Sci 24: 52. Jacobson SC, Hergenroder R, Koutny LB, Warmack RJ, Ram-
385–391 sey JM (1994) Anal Chem 66:1107–1113
39. Bushey MM, Jorgenson JW (1990) Anal Chem 62:978–984 53. Jacobson SC, Hergenroder R, Koutny LB, Ramsey JM (1994)
40. Bushey MM, Jorgenson JW (1990) J Microcolumn Sep 2:293– Anal Chem 66:2369–2373
299 54. Kutter JP, Jacobson SC, Matsubara N, Ramsey JM (1998) Anal
41. Larmann JP Jr, Lemmo AV, Moore AW Jr, Jorgenson JW Chem 70:3291–3297
(1993) Electrophoresis 14:439–447 55. Effenhauser CS, Paulus A, Manz A, Widmer HM (1994) Anal
42. Moore AW, Jorgenson JW (1995) Anal Chem 67:3448–3455 Chem 66:2949–2953
43. Lewis KC, Opiteck GJ, Jorgenson JW, Sheeley DM (1997) 56. Woolley AT, Mathies RA (1994) Proc Natl Acad Sci USA
J Am Soc Mass Spectr 8:495–500 91:11348–11352
44. Hooker TF, Jorgenson JW (1997) Anal Chem 69:4124–4142 57. Jacobson SC, Ramsey JM (1996) Anal Chem 68:720–723
45. Lemmo AV, Jorgenson JW (1993) Anal Chem 65:1576–1581 58. Moore AW, Jacobson SC, Ramsey JM (1995) Anal Chem 67:
46. Lemmo AV, Jorgenson JW (1993) J Chromatogr 633:213–220 4184-4189
47. Stahl M, Jakob A, von Brocke A, Nicholson G, Bayer E (2002) 59. Kutter JP, Jacobson SC, Ramsey JM (1997) Anal Chem 69:
Electrophoresis 23:2949–2962 5165–5171
48. Moore AW, Jorgenson JW (1995) Anal Chem 67:3456–3463 60. von Heeren F, Verpoorte E, Manz A, Thormann W (1998)
49. de Mello AJ (2002) Anal Bioanal Chem 372:12–13 Anal Chem 68:2044–2053
50. Effenhauser CS, Manz A, Widmer HM (1993) Anal Chem 65: 61. Yang Z, Zhang X, Li A, Zhu S, Huang Y (2003) Electrophore-
2637–2642 sis 24:1451–1457
51. Harrison DJ, Fluri K, Seiler K, Fan Z, Effenhauser CS, Manz A 62. Rocklin RD, Ramsey RS, Ramsey JM (2000) Anal Chem 72:
(1993) Science 261:895–897 5244–5249

You might also like