Professional Documents
Culture Documents
Service Manual
Copyright
© 2006 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.
Release, amendment, reproduction, distribution, rent, adaption and translation of this manual
in any manner whatsoever without the written permission of Mindray is strictly forbidden.
All information contained in this manual is believed to be correct. Mindray shall not be liable
for errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for safety, reliability and performance of this product only in the
condition that:
all installation operations, expansions, changes, modifications and repairs of this product
are conducted by Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable national and
local requirements;
I
z This equipment must be operated by skilled/trained medical professionals.
z Be sure to operate the analyzer under the situation specified in this manual;
otherwise, the analyzer will not work normally and the analysis results will
be unreliable, which would damage the analyzer components and cause
personal injury.
II
Warranty
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting from
the improper use or application of the product or the use of parts or accessories not approved
by Mindray or repairs by people other than Mindray authorized personnel.
any Mindray product which has been subjected to misuse, negligence or accident;
any Mindray product from which Mindray's original serial number tag or product
identification markings have been altered or removed;
Return Policy
Return Procedure
In the event that it becomes necessary to return this product or part of this product to Mindray,
the following procedure should be followed:
1. Obtain return authorization: Contact the Mindray Service Department and obtain a
Customer Service Authorization (Mindray) number. The Mindray number must appear on
the outside of the shipping container. Returned shipments will not be accepted if the
Mindray number is not clearly visible. Please provide the model number, serial number, and
a brief description of the reason for return;
2. Freight policy: The customer is responsible for freight charges when this product is
shipped to Mindray for service (this includes customs charges);
3. Return address: Please send the part(s) or equipment to the address offered by
Customer Service department.
III
Company Contact
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial Park,
Nanshan,ShenZhen 518057, P.R.China,
Phone: +86 755 26582479 26582888
IV
Product name: Auto Hematology Analyzer
For this Service Manual, the issued Date is 2006-08 (Version: 1.0).
Table of Contents
1
Table of Contents
5 Troubleshooting....................................................................................... 5-1
5.1 Error Codes............................................................................................ 5-1
5.2 Troubleshooting ..................................................................................... 5-2
5.2.1 Pressure errors ........................................................................... 5-2
5.2.2 Reagent errors ............................................................................ 5-3
2
Table of Contents
7 Appendices ..............................................................................................A-1
A Fluidic Diagram ......................................................................................A-1
B Electric Connection Diagram .................................................................B-1
C Test Point of Hardware Board ................................................................C-1
C.1 CPU board test point...................................................................C-1
C.2 Analog test point .........................................................................C-3
C.3 Drive board test point ..................................................................C-4
D Valve Test ...............................................................................................D-1
3
1 Using This Manual
1.1 Introduction
This chapter explains how to use your BC-3200 service manual, which provides the reference
information and procedures needed for servicing your analyzer. Read this manual carefully
before servicing your analyzer and service your analyzer strictly as instructed in this manual.
This manual is to be used in conjunction with the BC-3200 operation manual and does not
contain information and procedures already covered in the operation manual.
z Be sure to service your analyzer strictly as instructed in this manual and the
operation manual.
1-1
Using This Manual
have the ability to use basic mechanical tools and understand related terminology.
have the ability to read electronic and fluidic schematics and understand related
terminology.
1-2
Using This Manual
This service manual comprises 6 chapters and 4 appendices. Refer to the table below to find
the information you need.
learn about the sub-systems of the BC-3200 analyzer and how Chapter 2 System Discription
they function.
learn about how to install the BC-3200 and update the Chapter 3 Installation and
Updating
software.
learn about the procedures to repair the BC-3200. Chapter 4 Repair Procedures
learn about the main parts of the BC-3200. Chapter 6 Part List
learn about the jumpers and testing points of the BC-3200. Appendix C
1-3
Using This Manual
This manual uses certain typographical conventions to clarify meaning in the text:
All capital letters enclosed in [ ] indicate a key name (either on the built-in keypad or the
external keyboard), such as [ENTER].
All capital, bold and italic letters indicate a special operation defined in the following
section, such as SELECT.
Bold letters included in “ ” indicate text you can find on the screen, such as “Prepare to
ship”.
Bold letters indicate defined screen areas/fields, such as System Status area, or
chapter titles, such as Chapter 1 Using This Manual.
All illustrations in this manual are provided as examples only. They may not necessarily
reflect your analyzer setup or data displayed.
1-4
Using This Manual
z This analyzer adopts a fixed decimal point. You can enter the digits without
bothering to look for the [.] on the external keyboard.
1-5
Using This Manual
You may find the following symbols on the analyzer or the reagents.
BIOLOGICAL RISK
HIGH VOLTAGE
ALTERNATING CURRENT
1-6
Using This Manual
BATCH CODE
USE BY
SERIAL NUMBER
DATE OF MANUFACTURE
TEMPERATURE LIMITATION
MANUFACTURER
IRRITATING SUBSTANCE
Safety precautions:
Please pay attention to the following precautions to insure the personal safety.
z Power off the device when servicing it. Otherwise electric shock or damage
to electronic parts may occur.
z Connect the device to the socket which has a separate fuse and protection
switch. Otherwise the device may be tripped in case of failure, over current
or impact current at startup.
z To avoid personal injury, be sure to keep your clothes, hair and hand away
from such moving parts as the sample probe.
z The warning parts will move mechanically and may cause injury during
normal operation or disassembly or servicing.
1-7
Using This Manual
z The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
z If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
z When maintaining the electronic parts of the analyzer, be sure to remove the
metallic decorations you are wearing so as to avoid damage to you and the
analyzer.
z Close the sample compartment when opening the front door, otherwise you
may destroy them.
z The probe tip is sharp and may contain biohazardous materials. Exercise
caution to avoid contact with the probe when working around it.
z All the analyzer components and surfaces are potentially infectious, take
proper protective measures for operation or maintenance.
1-8
2 System Discription
Hardware
The hardware system controls/drives the valves, pumps and motors, converts the detected
blood cell signals into input signals as required, counts the number of cells, and provides
control/drive and monitoring signals.
Interface
The BC-3200 aspirates blood sample in an enclosed way. Press the [OPEN] key, as shown in
Figure 2-1, to open the sample compartment and close it after putting in the sample tube.
Press the [ASPIRATE] key. Then the device starts analyzing the sample and displays the
results (including graphs/tables and data) on the screen. The recorder prints analysis reports
and interested information in a couple of formats. Moreover, the keypad is used to input
operating instructions and parameters.
There are several external interfaces on the rear panel of the device, as shown in Figure 2-2,
which include two RS-232 serial ports, a parallel port, a keypad interface and a power
interface for floppy disk drive. The two RS-232 serial ports are used for communication with
an external PC and a bar code reader respectively. The parallel port is to connect a printer or
a floppy disk drive when updating the system.
2-1
System Discription
2-2
System Discription
The analyzer is equipped with a 32Bit Motorola ColdFire as its primary processor. The BIOS
and FPGA data and the boot program of the system software are stored in a 1MB NOR Flash.
The system software and sample information are stored in the Disk-on-Drive (DOM). During
startup of the analyzer, the hardware automatically loads the boot program which runs in the
flash and configures the FPGA, drives the keypad, turns on the LCD backlight, loads the
system software and finally transfers the CPU control to the system software. The system
software runs in a 16MB SDRAM, of which 1M is for the software code and the other 15M is
used as a memory during running.
2-3
System Discription
Fluidic System
The fluidic system is used for reagents, samples and air to flow inside the device. The fluidic
components are interconnected according to certain relationship and controlled by the
software and hardware based on the given timing, so that the distribution and flow direction of
the liquid are controlled.
2-4
System Discription
V1 V4 V5 V6 V3
RB
C WBC
Bat Bath 2.5mL 10mL
h Syringe Syringe
50uL DILUENT
Syringe
LYSE
Sample is dispensed in two modes: whole blood and predilute. 13µL venous blood is
aspirated in whole blood mode and 0.3mL prediluted capillary blood in predilute mode.
In whole blood mode, the sample is aspirated by the 50µL syringe and then stored in the
probe.
In predilute mode, the sample is aspirated by the 10mL syringe. The probe lowers down to
the 1.5mL centrifugal tube which holds the prediluted sample. V3 and V4 are opened. The
10mL syringe then starts aspirating the sample, which enters the probe and then the tube
connected to it under the action of pressure. When enough sample volume is aspirated, the
syringe stops and V4 is closed. The aspirated sample is stored in the probe and the tube
connected to it. To prevent the sample from entering V4 and the syringe, a tube with larger
capacity than the aspiration volume can be used.
The sample is dispensed from the probe to the bath. Blood is viscous and may adhere to
interior of the probe and tube. The sample needs to be diluted with defined volume of diluent,
which is dispensed through the probe to rinse the sample remained on the probe/tube wall
while the sample is dispensed. The 10mL syringe is used for diluent dispensing. Since
resistance may occur due to smaller inner diameter of the probe, the syringe dispenses fluid
slowly, thus preventing the tube from disconnecting owing to excessive compressed air and
reducing bubbles.
The probe is a moving part and can move vertically and horizontally. The tube connected to it
also changes in shape (bending or straightening) while it moves. The shape change may
2-5
System Discription
influence the tube capacity and even the sample aspiration. To minish the tube shape change
without increasing the flow resistance, a rigid EVA tube is used between the probe and the
50uL syringe and a relatively flexible Teflon tube is used between the 50uL syringe and valve
4.
Usually, the diluent volume is 0.5mL - 6mL and aspirated by the 10mL syringe through the
diluent inlet. As shown in Figure 2-3, since only V3 is connected between the 10mL syringe
and the diluent inlet, small resistance may occur and diluent can be aspirated in not-too-fast
speed to avoid bubbles. To better decrease the resistance, a tube with large inner diameter
(R-3603 ID1/4”) can be used to connect to the diluent inlet. The diluent may be dispensed to
the probe, probe wipe, bath or tubes. When the 10mL syringe aspirates enough diluent,
V1/V3 are opened, then the syringe dispenses certain diluent volume to the probe wipe to
wash the exterior of the probe. When the probe rotates to a bath, V3/V4 are opened while the
syringe is dispensing, then the diluent is dispensed to the bath. During this process, the
exterior of the probe is washed. Moreover, by opening only the V3 or both V3 and V4, the
diluent can be dispensed to a bath through the tube connected to it.
The lyse volume is usually 0.3-1mL and aspirated/dispensed through the cooperation of the
2.5mL syringe and V6. When certain volume of lyse is aspirated through the lyse inlet, V6 is
opened and the aspirated lyse is dispensed to the WBC bath through the lyse tube connected
to the bath.
The probe/E-Z cleanser volume is 1-2mL and aspirated through the probe and dispensed to
the bath through the 10mL syringe. Cleanser is aspirated/dispensed in the same way as
sample in predilute mode. The tube connected to the probe should have a larger capacity
than the aspiration volume, so that the cleanser will not be absorbed into valves and syringes.
2-6
System Discription
Pump
Waste
The fluid flow of the BC-3200 is implemented by not only the syringes, but also drive from the
air and compressed air. To establish the air, vacuum is taken by using the vacuum chamber
along with a pump and V10. The compressed air is established by using the pressure
chamber and the pneumatic pump. The above-mentioned two chambers are connected
respectively to a pressure sensor, which detects the vacuum and pressure status, as shown
in Figure 2-4.
The pneumatic pump is used to establish compressed air. The air flows from atmosphere to
the pressure chamber. When V10 and the pump are opened, the air flows from vacuum
chamber to the waste outlet, and then the air is established. When compressed air needs to
be stored in the vacuum chamber, the normally-closed terminal of V9 is opened after
compressed air’s establishment in the pressure chamber, and then the air flows from the
pressure chamber to the vacuum chamber.
2-7
System Discription
When the syringe starts dispensing diluent to the wipe, the pump and V2 are opened to
discharge the waste of the probe wipe. The probe, including its tip, is washed by relative
movement of the probe and the probe wipe. When diluent dispensing is stopped, the waste
outlet is still taking out air so that the exterior of the probe is dried, as shown in Figure 2-5.
To wash the front bath, when the 10mL syringe aspirates enough diluent, V3 is opened alone
2-8
System Discription
to dispense the diluent to the WBC front bath; or both V3 and V5 are opened to drive the
syringe to dispense the diluent to the RBC front bath. The front bath should be emptied
before dispensing diluent to it. The diluent volume should be greater than the maximum
sample volume in the bath during analysis. The front bath is connected to a pump through an
isolation chamber and a valve, which is closed when diluent is dispensed in the front bath,
and opened when the waste in the bath is being discharged by the pump, as shown in Figure
2-6.
The back bath is washed with rinse and has an inlet and outlet. The inlet is connected to the
rinse inlet through V13/V14, while the outlet shares a tube with the analyzing channel, as
shown in Figure 2-6. When V8 and V18 are opened simultaneously, the rinse inlet is
interconnected to the vacuum chamber, and the rinse enters the vacuum chamber under the
action of compressed air to wash the back bath.
Since the outlet of the front bath needs to be isolated to avoid interference signals, an
isolation chamber is added at the outlet.
V7
V8
V17
V18
RBC WBC
Bath Bath
V14
Rinse
Vacumm
Chamber
Prob Isolation Isolation V13
e Chambe Chambe
r r
Probe Wipe
Housing
Waste
Pump
The waste discharging module deals with discharging the waste from the probe wipe, baths
and the vacuum chamber.
The waste of the front/back baths and the vacuum chamber is discharged through a pump.
To discharge the waste of the WBC/RBC front bath, V11 or V12 is opened while the pump is
opened. It is relatively complicated to wash the back bath. Certain vacuum should first be
2-9
System Discription
established in the vacuum chamber, and then V13/V14/V8/V18 are opened. The rinse enters
in the vacuum chamber through the back bath under the action of pressure. The back bath is
rinsed. During this process, the pump and V10 are opened to discharge the waste of the
vacuum chamber. To discharge waste in the probe wipe, V2 and the pump are opened.
The waste discharging module is shown in Figure 2-7.
As shown in Figure 2-8, in normal status, tube “a” is for liquid (rinse) and tube “b” is for air.
Liquid must not enter V7/V17 and the filters connected to them. Tube “c” is filled with liquid
(rinse) in non-measuring status. Before analysis begins, V7 and V17 are opened to discharge
the liquid in tube “c” by taking in air. Meanwhile, the photo-coupler on the metering tube is
isolated. When analysis begins, V8 and V18 are opened, and liquid flows towards the
vacuum chamber through tube “a” and tube “c”. Corresponding photo-couplers are opened
when the liquid flows to them.
2-10
System Discription
2. To clean the sample probe: The probe is cleaned when it goes up and down. V1 valve
and V3 valve are opened and the 10mL syringe dispenses fluid. V2 valve is opened and
the waste pump starts working. Usually the washing is finished in 2s.
3. To establish vacuum: To establish vacuum, V10 valve is opened and the waste pump
starts working. The vacuum pressure value can be controlled in two ways: 1) To be
controlled by the software (vac: start time/end time, upper/lower limit of the vacuum); 2) To
be controlled directly with V10 valve opened and the waste pump starts working. When
counting, the vacuum pressure must be -24kPa.
4. To establish pressure: To establish pressure, the pressure pump starts working. The
pressure value is controlled by software (press: start time/end time, upper/lower limit of the
pressure). It is used for two purposes: 1) To pump bubbles into baths for mixing the sample
(controlled by press); 2) To increase the pressure in vacuum chamber to flush the aperture,
which is implemented by starting the pressure pump and opening V9 valve.
6. To dispense diluent: With the 10mL syringe and V3 valve opened, diluent is dispensed
into the WBC bath, RBC bath or to the probe wipe.
7. To dispense diluent to the probe wipe: When both V3 valve and V1 valve are opened
simultaneously, diluent is dispensed to the probe wipe.
8. To dispense diluent to the sample probe: When V3 valve, V1 valve and V4 valve are
opened simultaneously, diluent is dispensed into the sample probe.
9. To dispense diluent to the RBC bath: When V3 valve, V1 valve, V4 valve and V5 valve
are opened simultaneously, diluent is dispensed into the RBC bath.
11. To dispense lyse: With the 2.5mL syringe and V6 valve opened, lyse is dispensed into
the WBC bath.
12. To empty the RBC bath: With the waste pump started and V11 valve opened, the RBC
bath is emptied. Usually it finished in 2s.
13. To empty the WBC bath: With the waste pump started and V12 valve opened, the WBC
bath is emptied. Usually it finished in 2s.
14. To pump air into the RBC bath: Certain pressure is first established in the pressure
chamber through press command and then V15 valve is opened, thus the air in the
pressure chamber is pumped into the RBC bath.
15. To pump air into the WBC bath: Certain pressure is first established in the pressure
chamber through press command and then V16 valve is opened, thus the air in the
pressure chamber is pumped into the WBC bath.
16. To aspirate the sample: Sample is aspirated via the 50µL syringe.
17. To dispense the sample: With the 10mL syringe and V4 valve opened, sample is
dispensed from the sample probe. Meanwhile, a little diluent is also dispensed to rinse the
interior of the sample probe.
2-11
System Discription
18. To empty the RBC metering tube: After vacuum is established, V17 valve is opened to
empty the RBC metering tube.
19. RBC counting: After the RBC metering tube is emptied and vacuum is established, V18
valve is opened to start counting.
20. To empty the WBC metering tube: After vacuum is established, V7 valve is opened to
empty the WBC metering tube.
21. WBC counting: After the WBC metering tube is emptied and vacuum is established, V8
valve is opened to start counting.
22. To clean the RBC back bath and the RBC metering tube: After vacuum is established,
V13 valve and V18 valve are opened so that rinse can pass through the RBC back bath
and the RBC metering tube and then to the vacuum chamber.
23. To clean the WBC back bath and the WBC metering tube: After vacuum is established,
V14 valve and V8 valve are opened so that rinse can pass through the WBC back bath and
the WBC metering tube and then to the vacuum chamber.
2-12
System Discription
2-13
System Discription
2-14
System Discription
The sample probe is raised and the exterior of the probe is The exterior of the probe is
cleaned with 1.6mL of diluent. cleaned with 1.6mL of diluent
for 2s and dried for 0.7s.
1.6mL of diluent is added into The sample probe is lowered. 1.6mL of diluent is added into
the RBC bath. the bath to avoid spills and
bubbles when the probe is
dispensing sample.
1mL of prediluted 0.5mL of lyse is -22.5/ -24kPa Diluent and sample are
sample is dispensed into the vacuum is dispensed into the RBC bath
dispensed into the WBC bath and 10 established. through the sample probe at
WBC bath via the bubbles are level 2. The interior of the
sample probe pumped into the probe is cleaned at the same
moving. bath. time. Bubbles pumped into
the bath help to well mix the
sample and established
vacuum for counting.
2-15
System Discription
2-16
System Discription
2-17
System Discription
Counting Sequence
Sequence of post-cleaning
2-18
System Discription
2-19
System Discription
2-20
System Discription
aspirated. level 5.
4mL of diluent is 4 bubbles are WBC bath Diluent is added into the WBC bath
added into the WBC pumped into is emptied. through dispensing tube. Bubbles
bath. the WBC pumped helps to clean the WBC
bath. bath.
4mL of diluent is aspirated. 10mL syringe aspirates diluent at
level 5
4mL of diluent is added into the RBC bath. Diluent is added into the RBC bath
through dispensing tube.
The back bath is rinsed. Rinse pumped by the fluidic pump
flushes the back bath.
5mL of diluent is 0.5mL lyse is WBC bath 10mL syringe aspirates diluent at
aspirated. aspirated. is emptied. level 5. 50µL syringe aspirates lyse
quickly.
5mL of diluent is added into the WBC bath. Diluent is added into the WBC bath
through dispensing tube.
5mL of diluent is aspirated. WBC bath 10mL syringe aspirates diluent at
is emptied. level 5.
5mL of diluent is added into the WBC bath. Diluent is added into the WBC bath
through dispensing tube.
0.5mL of lyse is added 4 bubbles are RBC bath is 50µL syringe adds lyse quickly into
into the WBC bath. pumped into emptied. the WBC bath through the dispensing
the WBC tube. Bubbles pumped helps to mix
bath. diluent and lyse.
5mL of diluent is WBC bath is emptied. 10mL syringe aspirates diluent at
aspirated. level 5.
5mL of diluent is Vacuum chamber is Diluent is added into the WBC bath
added into the WBC emptied. through dispensing tube.
bath.
5mL of diluent is aspirated. Pressure in 10mL syringe aspirates diluent at
the vacuum level 5.
5mL diluent is added The exterior chamber is Diluent is added into the RBC bath
into the RBC bath. of the sample resumed. through dispensing tube. 50µL
50µL syringe probe is syringe re-aspirates 2µL of diluent to
re-aspirates 2µL of dried. The avoid liquid accumulating at the
diluent. aperture is probe hole and dripping. 110V
zapped. voltage is added at ends of the
aperture to zap the aperture. The
vacuum chamber and pressure
chamber are connected to release
the vacuum.
2-21
System Discription
2-22
System Discription
According to the mechanical design, the BC-3200 is divided into circuit unit, volumetric unit,
power supply and panel assemblies.
Analog board: Converts the signals output from each sensor into required digital signals;
provides control/drive and monitoring signals.
Drive board: Drives the valves, pumps and motors of the system; and carries out the
instructions sent by the CPU.
CPU Board
Working Principles
CPU, FPGA and Super I/O are the major parts on this board. The CPU carries out
instructions and functions as the core of the board. The FPGA acts as a logical interface
between the CPU and the Super I/O. The super I/O includes various interfaces that can be
accessed by the CPU through the LPC of FPGA. SDRAM is the system memory. FLASH is
where the boot software and FPGA configuration files are stored. The DOM (Disk-on-Module)
holds the system software and test data. The RTC is a real-time clock. The system
configurations are stored in the EEPROM. VRAM is the memory for video display. See Figure
2-9.
2-23
System Discription
Basic Functions
To monitor such system parameters as the +56V, +12V and -12V supplies of the analog
board, the +3.3V and +12V supplies of the CPU board itself and the temperature of the
analyzer.
To receive the keypad signal and control the keypad buzzer and LCD backlight.
To generate control signals to control the valves, pumps, aperture zapping, HGB LED, current
source and digital pot.
Analog Board
Working Principles
The analog board converts the signals output from each sensor into required digital signals;
provides control/drive and monitoring signals.
Basic Functions
The analog board includes 6 functional units: interface unit, power supply unit (DC-DC),
power supply monitoring unit, volumetric signal unit, HGB unit and vacuum/pressure unit.
2-24
System Discription
Drive Board
Working Principles
The drive board drives the valves, pumps and motors. It carries out the following instructions
sent by the CPU: to open/close the pumps and solenoid valves; to control the syringe motors;
to control the movement of the probe; and to hold the motor in the standby mode.
Basic Functions
The drive board drives the valves, pumps and motors. It carries out the following instructions
sent by the CPU: to open/close the pumps and solenoid valves; to control the syringe motors;
to control the movement of the probe; and to hold the motor in the standby mode.
Power block
The power block includes the 5V, 12V and 30V DC supplies. The 12V powers the valve/pump
and rotation motor drive circuits (max. current: 2A). The 30V powers the drive circuits of
elevator motors and 10mL&50uL&2.5mL syringe motors (max. current: 4A). The 5V powers
the drive chips of valve/pump and rotation motors (max. current: 50mA) and the chip control
system (max. current: 200mA).
The switch control block consists of the photo-coupling circuit and valve/pump drive circuit, as
shown in Figure 2-10.
Photo-coupling Circuit
The photo-coupling circuit mainly consists of the photo-coupler and resistors. It provides 20
TTL outputs to the valves and pumps. The photo-coupler, TLP521-2, isolates the digital
ground from the driving ground.
The drive circuit controls the valves and pumps by using the digital signals from the main
control board. The technical parameters include the following:
Number of channels: 20 (2 pumps and 18 valves);
Output: TTL, active low.
Power: 12V/1A (max.)
The input is isolated from the output (the digital ground is isolated from the power ground).
2-25
System Discription
According to the drive voltage and max drive current for pumps/valves, the ULN2068B chip,
which can drive 1.5A current, is applied to drive the pumps and valves.
The motor control block is composed of the serial communication circuit, probe control/drive
circuit, syringe motor control/drive circuit, position sensor drive/detection circuit.
Since the CPU board requires a 3.3V power supply while the drive board requires a 5V power
supply, a photo-coupler (H11L1) is needed for the purposes of conversion and isolation.
The probe control/drive circuit includes the elevator/rotation motor control/drive circuits. The
control system of the probe motor consists of an AT89S52 MCU and ADM705 WDT. The
AT89S52 also detects signals coming from the position sensors and manages the serial
ports.
Working Principles
The volumetric unit measures the volume of the samples during counting and outputs the
start and end signals of counting. The principle diagram of the volumetric unit is shown in
Figure 2-11.
2-26
System Discription
The volumetric unit consists of a sensor, consistent-current circuit and output circuit.
Filter
Power board
Equipotentiality
terminal
Indicator board
Keypad board
2-27
System Discription
LCD Inverter
LCD adapter
Recorder
The serial signal lines, +5V and +12V power cords, the 5V ground line and the power ground
are directed from the same connector of the CPU board. They are connected to the front
panel by one cable and then split respectively to the recorder and the keypad. The LCD
signal line is isolated. The inverter, powered by keypad power supply, drives the backlight of
LCD. The backlight brightness can be adjusted via the keypad. An LCD adapter is added
here.
Keypad
Working Principles
The keypad adapter mainly consists of an MCU, keypad matrix, backlight/power indicator
control and buzzer, as shown in Figure 2-15.
Basic Functions
To scan the keypad: The keypad adapter scans the keypad and reports the scanned key
2-28
System Discription
To control the LCD brightness: The keypad adapter receives instructions from the main
control board to turn on/off the backlight and power indicators of the LCD and to adjust the
brightness of the backlight.
To control the buzzer: The keypad adapter receives instructions from the main control board
to turn on/off the buzzer.
LCD Adapter
Basic Functions
The LCD adapter connects the LCD to the CPU board, as shown in Figure 2-16.
2-29
System Discription
The software system includes the boot software and system software. The boot software,
loaded automatically by the hardware, runs in the FLASH. It configures the FPGA, drives the
keypad, turns on the LCD backlight, loads the system software, and finally transfers the CPU
control to the system software. The system software, loaded by the boot software, runs in the
SDRAM. It gets resource data from the external DOM, and provides such functions as
counting, system setup, quality control, maintenance, etc. The system software sends
instructions to corresponding fluidic unit at given time according to the basic timing.
2-30
System Discription
2.4.2 Password
There are three levels of password: user, administrator and service engineer. Authorities for
2-31
System Discription
2-32
System Discription
2-33
System Discription
2-34
3 Updating the Software
2. There are two classes of files for updating: information file (UPDATE.CFG) and data file.
They can be put in one or more floppy disks;
3. Only one module can be upgraded at one time. Update disks of different modules must
not be used simultaneously, since some files in the update disks have the same name
which would lead to update failure, i.e. all the information files use the same filename
“Update.cfg”.
4. Find the subfolder where the desired upgrading file locates. Copy all information files and
data files in the subfolder to the floppy disk, and then cancel the Read Only property of all
files.
3-1
Updating the Software
4. At the “Count” screen, press [↑] (up arrow). The following prompt is displayed;
5. Insert the floppy disk with information files into the floppy drive, press [ENTER].The
following prompt is displayed;
6. After the analyzer reads and analyzes the upgrading files, the following prompt is
displayed;
7. Insert the floppy disk with data files into the floppy drive, press [ENTER]. The system
writes the data files to the temporary folder of the DOM;
8. When the system reads the files successfully, the following prompt is displayed.
9. After the files are written to the working folder of the DOM and the temporary folder is
cleared, the following prompt is displayed;
10. Few seconds later, the analyzer is restarted. The update is finished.
3.4 Precautions
1. During update, the operator must not go away and should watch each prompt displayed;
3. During update, if “Printing. Try again later.” is displayed, don’t operate until the printing
is finished;
“Insert disk of config, then press [ENTER] key.” or “Insert disk of data, then press
[ENTER] key.”
5. Check if you insert the correct disk. If the disk reading fails 3 times, the analyzer returns
to the “Count” screen;
6. If “Updating failed” is displayed, the analyzer will return to the “Count” screen in few
seconds. If the system software can not be normally loaded in later use, you can only
restore it by reinstalling the DOM.
3-2
4 Repair Procedures
4.1 Introduction
Refer to section 4.2 - 4.13 for detailed procedures. The service engineer must verify the
repair strictly as instructed by the enclosed steps, if any.
4-1
Repair Procedures
The left door needs to be opened when you replace lyse or service the analyzer.
Opening
1. Shut down the analyzer and disconnect it from the power supply;
3. While pushing the latch, rotate the left door in the direction indicated in Figure 4-1.
Installation
Install the left door according to the steps above in reverse order.
Verification
4-2
Repair Procedures
Purpose
The right door needs to be opened when you replace the fluidic components or open the front
door.
Opening
1. Shut down the analyzer and disconnect it from the power supply;
3. While pushing the latch, rotate the right door in the direction indicated in Figure 4-2.
Installation
Install the right door according to the steps above in reverse order.
Verification
The front door needs to be opened when you replace the probe, bath, LCD and recorder.
Opening
1. Shut down the analyzer and disconnect it from the power supply;
4-3
Repair Procedures
4. Pull upwards the latch on the front sheet metal, as shown in Figure 4-3;
z Close the sample compartment when opening the front door, otherwise you
may destroy them.
Installation
4-4
Repair Procedures
Verification
The top cover needs to be removed when you replace the analog board, drive board, power
supply assembly, connecting cables and tubes, etc.
Tool
2# cross screwdriver.
Removal
1. Shut down the analyzer and disconnect it from the power supply;
2. Remove the three M4X8 screws with the screwdriver, as instructed by Figure 4-4;
3. Pull the top cover backwards and then remove it, as instructed by Figure 4-4.
4-5
Repair Procedures
Installation
Install the top cover according to the steps above in reverse order.
Verification
The back cover needs to be removed when you replace the boards, cables and fluidic
components.
Tool
2# cross screwdriver
Removal
1. Start up the system and empty the tubes. Shut down it and disconnect it from the power
supply;
5. Unscrew the five M3X4 screws fixing the lyse support and then remove the support, as
shown in Figure 4-5;
6. Unplug the tubes from the diluent inlet, rinse inlet and waste outlet. Unplug the diluent
and rinse sensor connectors;
7. Unscrew the eight M4X8 screws and remove the back cover (together with the left and
right doors, as shown in Figure 4-5).
4-6
Repair Procedures
z The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
z If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
Installation
Install the back cover according to the steps above in reverse order.
4-7
Repair Procedures
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware Æ
fluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization.
4-8
Repair Procedures
z When maintaining the electronic parts of the analyzer, be sure to remove the
metallic decorations you are wearing so as to avoid damage to you and the
analyzer.
The analog board, CPU board and drive board are fixed on the mounting plate, as shown in
Figure 4-6.
The analog board collects original signals and provides control/drive and monitoring signals.
It needs to be replaced when going wrong.
4-9
Repair Procedures
Tool
2# cross screwdriver.
Removal
1. Shut down the analyzer and disconnect it from the power supply;
3. Unscrew the #6-32X6 knurled screw fixing the shielding box and then remove the box
cover, as shown in Figure 4-7;
4. Unplug the J1-J5 connectors and the two tubes from the analog board;
5. Unscrew the four M3X8 screws and remove the analog board, as shown in Figure 4-7.
Installation
4-10
Repair Procedures
Install the analog board according to the steps above in reverse order.
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware Æ
fluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization;
3. Re-adjust the gains and conduct repeatability test as instructed by section 4.13.
The CPU board, the key part of the analyzer, receives signals and monitors/drives peripheral
devices. It needs to be replaced when going wrong.
Tool
2# cross screwdriver.
Removal
1. Shut down the analyzer and disconnect it from the power supply;
3. Unplug the J1, J3, J4, J5, J6, J7, J8, J10, J11 and J16 connectors from the CPU board,
as shown in Figure 4-8;
4. Remove the DOM from the CPU board, as shown in Figure 4-8;
5. Unscrew the four M3X8 screws and remove the CPU board, as shown in Figure 4-8.
4-11
Repair Procedures
Installation
Install the CPU board according to the steps above in reverse order.
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware Æ
fluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization.
The drive board drives the valves, pumps and motors. It needs to be replaced when going
wrong.
4-12
Repair Procedures
Tool
2# cross screwdriver.
Removal
1. Shut down the analyzer and disconnect it from the power supply;
3. Unplug the J1-J10 connectors from the drive board, as shown in Figure 4-9;
4. Unscrew the four M3X8 screws and remove the drive board, as shown in Figure 4-9.
Installation
Install the drive board according to the steps above in reverse order.
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware Æ
fluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization;
4-13
Repair Procedures
4.3.4 Keypad
Purpose
Tool
2# cross screwdriver.
Removal
1. Shut down the analyzer and disconnect it from the power supply;
3. Unscrew the six M3X8 screws fixing the LCD, as shown in Figure 4-10;
4. Unplug the J2, J3 connectors from the keypad and remove the LCD;
5. Unscrew the seven M3X8 screws fixing the keypad and then remove the keypad, as
shown in Figure 4-10.
4-14
Repair Procedures
Installation
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware Æ
4-15
Repair Procedures
fluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization.
Tool
2# cross screwdriver.
Removal
1. Shut down the analyzer and disconnect it from the power supply;
3. Unplug the J1, J2 connectors from the indicator board, as shown in Figure 4-11;
4. Unscrew the four M3X8 screws and then remove the indicator board, as shown in Figure
4-11;
5. Remove the key caps from the keys on the indicator board.
4-16
Repair Procedures
Installation
Install the indicator board according to the steps above in reverse order.
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware Æ
fluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization.
4-17
Repair Procedures
The power supply assembly needs to be replaced when it goes wrong or when you remove
the back cover.
Tool
2# cross screwdriver.
Removal
1. Shut down the analyzer and disconnect it from the power supply;
3. Unplug the J1, J11 and J7 from the analog board, CPU board and drive board
respectively, as shown in Figure 4-12;
4. Unscrew the four M4X8 screws and then remove the power supply assembly in the
direction indicated in Figure 4-12.
4-18
Repair Procedures
Installation
Install the power supply assembly according to the steps above in reverse order.
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware Æ
fluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization.
4-19
Repair Procedures
Tool
Removal
1. Shut down the analyzer and disconnect it from the power supply;
5. Unplug the J8 connector from the CPU board, and unscrew the M3X8 screw fixing the
LCD ground wire, as shown in Figure 4-13;
6. Unscrew the six M3X8 screws fixing the LCD to the front door, as shown in Figure 4-13;
7. Unplug the J2, J3 connectors from the keypad and then remove the LCD in the direction
indicated in Figure 4-13.
4-20
Repair Procedures
Procedure 2: To further disassemble the LCD, as Figure 4-14 Disassembling the LCD shows.
4-21
Repair Procedures
Installation
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware Æ
fluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization.
4-22
Repair Procedures
Tool
Removal
1. Shut down the analyzer and disconnect it from the power supply;
4. Unscrew the two M3X12 screws fixing the recorder to the shielding box;
6. Unscrew the four M3X8 screws fixing the shielding box to the front door;
7. Unplug the JP1 and JP3 connectors from the recorder drive board;
8. Squeeze the bulges on the left/right sides of the recorder and then remove the recorder
in the direction indicated in Figure 4-15.
4-23
Repair Procedures
1. Toggle the bulge on the recorder box, remove the drive board and unplug the circuit
board connected to it. Replace the drive board if it goes wrong;
4. Toggle upwards the paper tension lever on the thermal head. Remove the paper, if any;
5. Unscrew the two PT2X6 screws fixing the thermal head and then remove the thermal
head.
4-24
Repair Procedures
Installation
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware Æ
fluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization.
4-25
Repair Procedures
The fluidic components, which perform tasks according to the given sequencing, include:
solenoid valve, syringe, volumetric metering tube, pump, vacuum chamber and pressure
chamber, as shown in Figure 4-17. They need to be maintained or replaced when going
wrong.
4-26
Repair Procedures
21---SV13
4.7.1 Valves
Purpose
The solenoid valves control the fluid flow or direct it to different tubes. Many kinds of fluid may
flow through the valves during analysis. The valves will probably go wrong when getting
blocked or eroded.
Tool
2# cross screwdriver.
Removal
1. Start up the system and empty the tubes. Shut down it and disconnect it from the power
supply;
4. Unplug the connector on the back of the SV5, as shown in Figure 4-18;
5. Unplug the tubes connecting the SV5 to the SV4 and bath, as shown in Figure 4-18;
6. Unscrew the two M3X8 screws fixing SV5 and then remove the SV5, as shown in Figure
4-18.
4-27
Repair Procedures
Installation
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware
Æfluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization;
3. Click [MENU] on the “Count” screen. The system menu is displayed. SELECT
“ServiceÆValve Test” to enter the “Valve Test” screen. SELECT a valve to check the
valve you’ve selected.
4-28
Repair Procedures
4.7.2 Pumps
Purpose
The pumps, consisting of the pneumatic pump, fluid pump and supports, drive the whole
system during analysis. They need to be maintained or replaced when going wrong.
Tool
Removal
1. Start up the system and empty the tubes. Shut down it and disconnect it from the power
supply;
3. Unscrew the five M3X4 screws fixing the lyse support and then remove the support, as
shown in Figure 4-19;
4. As shown in Figure 4-19, unplug the tubes connected to the air vent of the pneumatic
pump and to the inlet/outlet of the fluid pump. Unplug the connectors from the two pumps.
Remove the two 82-32UNC-2B nuts;
5. Unscrew the M4X8 screws fixing the pumps and then remove the pumps, as shown in
Figure 4-19.
4-29
Repair Procedures
Unscrew the four M3X8 screws fixing the fluid pump and then remove the pump, as shown in
Figure 4-20
4-30
Repair Procedures
Cut the belts binding the pneumatic pump to the support and then remove the pump, as
shown in Figure 4-21.
Installation
4-31
Repair Procedures
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware Æ
fluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization;
3. The status of the vacuum chamber and the pressure chamber can indicate how the two
pumps are going;
4. Press [MENU] at the “Count” screen. The system menu is displayed. SELECT “Service
Æ System Test” to enter the “System Test” screen. SELECT “Vacuum” to check the fluid
pump; or SELECT “PC Pressure” to check the pneumatic pump.
4.7.3 Syringes
Purpose
Syringes aspirate specified volume of fluid. They often need to be replaced when going
wrong.
Tool
Removal
1. Start up the system and empty the tubes. Shut down it and disconnect it from the power
supply;
4. Unplug the connectors of position sensors, motors and cables, as shown in Figure 4-22.
5. Unplug the tubes connecting the syringes to the SV6, SV4, SV3 and the probe, as shown
in Figure 4-22;
6. Unscrew the four M4X8 screws fixing the syringes to the support and then remove the
syringes, as shown in Figure 4-22.
4-32
Repair Procedures
2. The two syringes are removed in completely the same way. The 50µL one is taken for
example;
3. Unscrew the three M3X12 screws and then remove the presser bar;
4. Unscrew the screws fixing the plunger rod and then remove the 50µL syringe.
4-33
Repair Procedures
1. Unscrew the two M3X12 screws and then remove the presser bar;
2. Unscrew the screw fixing the plunger rod and then remove the 10mL syringe.
4-34
Repair Procedures
2. Unscrew the two M3X8 screws fixing the motor and then remove the motor.
Unscrew the two M3X8 screws fixing the photo-coupler switch and then remove the sensors,
as shown in Figure 4-26.
4-35
Repair Procedures
Installation
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware Æ
fluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization;
3. If hardware test is finished without error, fluidic system test starts. If the syringes go
normally, the analyzer will run normally, otherwise the fluidic system test is skipped and the
analyzer directly enters the “Count” screen.
The vacuum chamber stores compressed air to drive the fluid flow and also stores the air
when the analyzer flushes the aperture, therefore it may be broken in some cases because it
always suffers from external pressure.
Tool
2# cross screwdriver.
Removal
4-36
Repair Procedures
1. Start up the system and empty the tubes. Shut down it and disconnect it from the power
supply;
3. Unplug the tubes connecting the vacuum chamber to the metering tube and SV10, as
shown in Figure 4-27;
4. Unscrew the two M3X8 screws fixing the vacuum chamber and then remove the vacuum
chamber, as shown in Figure 4-27.
Installation
Install the vacuum chamber according to the steps above in reverse order.
Verification
3. Press [MENU] at the “Count” screen. The system menu is displayed. SELECT “Service”
Æ”System Test” to enter the “System Test” screen. SELECT “Vacuum” to check the fluid
4-37
Repair Procedures
pump.
The pressure chamber stores the air to drive the air flow inside the analyzer. It may be broken
in some cases because it always suffers from external pressure.
Tool
2# cross screwdriver.
Removal
1. Start up the system and empty the tubes. Shut down it and disconnect it from the power
supply;
3. Unscrew the five M3X4 screws fixing the lyse support and then remove the support, as
shown in Figure 4-28;
4. Unplug the tubes connecting the pressure chamber to the pneumatic pump, the analog
board and the bath, as shown in Figure 4-28;
5. Unscrew the M3X8 screws fixing the pressure chamber and then remove the pressure
chamber, as shown in Figure 4-28.
4-38
Repair Procedures
Installation
Install the pressure chamber according to the steps above in reverse order.
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware
Æfluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization;
3. Press [MENU] at the “Count” screen. The system menu is displayed. SELECT “Service
Æ System Test” to enter the “System Test” screen. SELECT “PC Pressure” to check the
pneumatic pump.
The metering tube measures the volume of the samples during counting. It may bring
erroneous results if foreign matters exist in it. The failure of the photo-coupler sensor may
also affect the analysis.
Tool
2# cross screwdriver.
Removal
1. Start up the system and empty the tubes. Shut down it and disconnect it from the power
supply;
5. Unscrew the two M3X8 screws fixing the shielding box and then remove the box;
6. Unplug the tubes connecting the metering tube to the vacuum chamber, the SV7, SV8,
SV17 and SV18;
7. Unscrew the two M3X8 screws and the two M3X12 screws and then remove the
4-39
Repair Procedures
metering tube.
1. Separate the volumetric unit, the metering tube and other parts;
2. Remove the metering tube with rubber sleeves from the holder;
4-40
Repair Procedures
Installation
Install the metering tube according to the steps above in reverse order.
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware Æ
fluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization;
3. After the analyzer enters the “Count” screen, remove the shielding box over the
metering tube and start background counting. Check the fluid flow in the tube and the count
time. The analyzer returns to the “Count” screen after counting if the photo-coupler works
normally.
4-41
Repair Procedures
The bath assembly consists of the WBC/RBC units, isolation chamber and shielding box. The
parts to be replaced include the HGB unit/aperture of the WBC unit and aperture of the RBC
unit.
The HGB unit, installed on the WBC bath, measures the Hemoglobin Concentration of
samples. It may lead to erroneous results when installed incorrectly.
The aperture is the key part of analyzer. It may often be blocked due to its small inner
diameter.
Tool
Removal
The HGB unit and aperture of the WBC unit are removed in three procedures.
Procedure 1: To remove the WBC unit from the analyzer, as shown in Figure 4-31.
1. Start up the system and empty the tubes. Shut down it and disconnect it from the power
supply;
5. Unplug the WBC connector (J5) from inside the shielding box of the analog board as
instructed by section 4.3.1;
6. Unscrew the two M3X6 screws fixing the outer WBC shielding box and then remove it in
the direction indicated in Figure 4-31;
7. Loosen the two M3X6 screws fixing the WBC shielding box and then remove it in the
direction indicated in Figure 4-31;
8. Unplug the tubes connecting the WBC unit to the SV8, the shielding plate and the
isolation chamber;
4-42
Repair Procedures
Procedure 2: To remove the HGB unit from the WBC unit, as shown in Figure 4-32
Unscrew the M2X10 screws and then remove the HGB unit from the font/back bath.
4-43
Repair Procedures
Procedure 3: To remove the aperture from the WBC unit, as shown in Figure 4-33.
Unscrew the two M3X10 screws and then remove the aperture from the front/back bath.
Installation
Install the HGB unit and aperture according to the steps above in reverse order.
4-44
Repair Procedures
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware Æ
fluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization;
3. When the analyzer enters the “Count” screen, open the front door and check if blue light
exists in the WBC bath. If it does, the HGB unit is properly connected;
4. Start background counting on the “Count” screen. If the aperture works normally, the
WBC counting time is within the normal range, otherwise bubbles will appear or the
aperture will be blocked.
The aperture is the key part of analyzer. It may often be blocked due to its small inner
diameter.
Tool
Removal
1. Start up the system and empty the tubes. Shut down it and disconnect it from the power
supply;
5. Unplug the RBC connector (J4) from inside the shielding box of the analog board as
instructed by section 4.3.1;
6. Unscrew the two M3X6 screws fixing the outer RBC shielding box and then remove it in
the direction indicated in Figure 4-34;
7. Loosen the two M3X6 screws fixing the RBC shielding box and then remove it in the
direction indicated in Figure 4-34;
8. Unplug the tubes connecting the RBC unit to the SV8, the shielding plate and the
isolation chamber. Unscrew the M3X8 screw fixing the ground wire of the RBC unit;
4-45
Repair Procedures
Procedure 2: To remove the aperture from the RBC unit, as shown in Figure 4-35
Unscrew the two M3X10 screws and then remove the aperture from the front/back baths.
4-46
Repair Procedures
Installation
Verification
2. Power on the analyzer. It starts initialization test in the order of system Æ hardware Æ
fluidic system. If the cables and tubes are connected normally, the analyzer enters the
“Count” screen without alarm after initialization;
3. Start background counting on the “Count” screen. If the aperture works normally, the
RBC counting time is within the normal range, otherwise bubbles will appear or the
aperture will be blocked.
4-47
Repair Procedures
Tool
2# cross screwdriver, 2.5mm hex wrench, 150mm monkey wrench and diagonal pliers.
Removal
1. Shut down the analyzer and disconnect it from the power supply;
2. Remove the top cover as instructed by section 4.2.4. Open the right door and front door
as instructed by section 4.2.2./ 4.2.3;
3. Remove the probe from the probe assembly, as shown in Figure 4-36;
4-48
Repair Procedures
z The probe tip is very sharp. Be sure to treat it carefully to avoid personal
injuries during analysis or maintenance.
z The probe tip is sharp and may contain biohazardous materials. Exercise
caution to avoid contact with the probe when working around it.
4. Remove the probe wipe from the probe assembly, as shown in Figure 4-37;
7. Unscrew the 3 socket head screws fixing the sample compartment and then hold and put
4-49
Repair Procedures
8. Remember the probe tube layout and then remove the belts binding the tube on the
bracket;
9. Unscrew the three M4 screws fixing the probe assembly and then remove it slowly, as
shown in Figure 4-39.
4-50
Repair Procedures
Installation
1. Install the probe assembly to the corresponding place on the front plate;
3. Install the sample compartment on the probe assembly and tighten the three socket
head screws;
4. Connect the motor connectors (MOTOR Y/C34937-A for upper motor and MOTOR
X/C34936-A for lower motor);
a) The external sensor on the top left corner of the probe assembly indicates “XL” and
internal one indicates “XR”;
b) The upper sensor on the right of the probe assembly indicates “YU” and the lower one
indicates “YD”.
6. Install the probe wipe in the aligning hole on the lower rotation plate, insert the clip and
direct the two tubes in the tube clamp;
a) Pull the guide block of the probe assembly upward to the upper position;
b) Insert the probe down into the aligning hole. When the tip reaches to the probe wipe,
rotate the probe to have its tip against the outlet of the probe wipe;
c) Fix the probe with a hexagon screw by using a monkey wrench.
8. Direct the probe tube according to its original layout and fix it with belts. Connect the tube
to the upper end of the probe.
4-51
Repair Procedures
Tool
Removal
1. Shut down the analyzer and disconnect it from the power supply;
2. Remove the top cover as instructed by section 4.2.4. Open the right door and front door
as instructed by section 4.2.2./4.2.3;
3. Unscrew the screw fixing the rotation sensor, as shown in Figure 4-40;
6. Unscrew the screw fixing the to-be-replaced sensor to the mounting plate.
4-52
Repair Procedures
Installation
Tool
Removal
1. Shut down the analyzer and disconnect it from the power supply;
2. Open the right door and front door as instructed by section 4.2.2./4.2.3;
3. Unscrew the two M3 screws fixing the elevator motor, as shown in Figure 4-41;
4. Remove the elevator motor assembly on the front plate from the right of the analyzer;
6. Unscrew the screw fixing the to-be-replaced sensor to the mounting plate.
4-53
Repair Procedures
Installation
You can refer to this section when removing or replacing the entire probe assembly or its
special parts.
Tool
2# cross screwdriver, 2.5mm hex wrench, 1.5mm hex wrench, 150mm monkey wrench,
slotted head screwdriver and diagonal pliers.
Removal
2. Separate the elevator motor assembly from the probe assembly and then remove the
reduction synchronous belt;
3. Unscrew the four M3X6 screws on the elevator motor assembly and then remove the
motor;
4. Loosen the two M3X5 (Figure 4-42) screws between the synchronous belt pulley and the
rotor, and then remove the synchronous belt pulley.
4-54
Repair Procedures
Figure 4-42 Removing the elevator motor and the reduction synchronous belt
2. Press the middle of the belt for several times till sense its tension, as shown in Figure
4-43;
4-55
Repair Procedures
3. Loosen the two M3X5 screws (Figure 4-44) on the lower belt pulley;
4. Unscrew the two M3X8 screws fixing the synchronous belt and then remove it from the
belt pulleys;
5. Take out theφ4slotted washer and then remove the belt roller from the 2 flanged
bearings.
4-56
Repair Procedures
Figure 4-44 Removing the synchronous belt and the lower belt pulley
2. Loosen the two M3X5 screws on the bigger belt pulley and then remove the belt pulley;
3. Loosen the two M3X5 screws on the smaller belt pulley and then remove the belt pulley;
4. Take out the φ4 slotted washer and then remove the belt roller (Figure 4-45) from the 2
flanged bearings.
4-57
Repair Procedures
2. Unscrew the M3X6 screw fixing the tube bracket and then remove the bracket, as shown
in Figure 4-46;
4. Unscrew the 4 socket head screws fixing the rotation motor and then remove the motor,
guide block assembly and the flanged bearing, as shown in Figure 4-46;
4-58
Repair Procedures
Figure 4-46 Removing the guide block assembly and the rotation motor
5. Loosen the M3X5 screw (Figure 4-47) on the control sleeve and then remove the sleeve;
6. Unscrew the M3X6 screw fixing the induction plate and then remove the plate;
7. Hold the guide rod, unscrew the M3X8 screw fixing the rod and then remove the rod;
8. Unscrew the two M3X5 screws fixing the guide rotor and then remove the lower rotation
plate;
4-59
Repair Procedures
10. Unscrew the four M3X16 screws on the guide block assembly.
2. Unscrew the 4 socket head screws fixing the linear guide pair and then remove the guide
pair.
Note: When removing the guide pair, ensure the block on the guide pair does not exceed its
4-60
Repair Procedures
two ends.
Installation
1. Install the guide pair in the corresponding slot on the mounting plate, and then tighten the
four M3X8 screws;
2. Install the upper/lower blocks tightly on the block of the guide pair, and then tighten the
four M3X8 screws.
4-61
Repair Procedures
1. Put the probe support on the screw side of the guide sleeve, with the screw thread side
upward. String them with the linear bearing and then tighten the four M3X16 screws;
2. Insert the guide rotor into the guide block assembly, with the linear bearing side upward;
3. Install the lower rotation plate on the guide rotor. Insert the guide rod through the upper
rotation plate, probe support and then the lower rotation plate. Then install the M3X6 screw
on the upper end of the guide rod;
4. Put the guide block assembly on the corner of a flat table or on a pad, as shown inFigure
4-50, with the guide rod and guide rotor close to the table or pad. Then tighten the two
M3X5 screws to fix the rotor of the rotation motor;
5. Do the remaining installation according to steps 1-6 in section Guide block assembly
and rotation motor.
4-62
Repair Procedures
1. First loosely install the two M3X8 screws to fix the synchronous belt assembly, as shown
in Figure 4-51;
3. Press the middle of the belt in the direction indicated in Figure 4-51 till sense the same
tension as before removing;
6. Install the elevator motor and reduction synchronous belt according to steps 1- 4 in
section Elevator motor and reduction synchronous belt.
4-63
Repair Procedures
z The reduction synchronous belt shall be tensed to the extent that it is just
straightened.
4-64
Repair Procedures
Tool
Removal
1. Shut down the analyzer and disconnect it from the power supply;
2. Remove the top cover as instructed by section 4.2.4. Open the right door and front door
as instructed by section 4.2.2./4.2.3;
3. Unplug the 5 connectors from the top of the analyzer: K6, K5, K1, K3 and K7;
4. Unscrew the 3 socket head screws fixing the sample compartment and then take out the
sample compartment.
z Exercise caution on the cable connectors when taking out the sample
compartment.
4-65
Repair Procedures
Installation
A new sample compartment can be installed according to the above-mentioned steps 1-4.
Note: Connect the upper switch of tube position to K6 and the lower switch to K5.
Position correction
Adjust the compartment door according to the panel opening, as shown in Figure 4-53.
Loosen the retaining screw and adjust the compartment door in the arrow direction. Ensure
the space between one side of the compartment door and the panel is about 1.5mm.
Tool
Removal
1. Shut down the analyzer and disconnect it from the power supply;
2. Remove the top cover as instructed by section 4.2.4. Open the right door and front door
4-66
Repair Procedures
3. Unplug the connector marked “K7” from the top of the analyzer;
4. Unscrew the 3 retaining screws fixing the electromagnet assembly, as shown in Figure
4-54;
5. Take out the φ2 slotted washer with nipper pliers and then remove the spring, as shown
in Figure 4-55;
6. Unscrew the retaining screws to remove the electromagnet, support, paper washer and
magnetic core, as shown in Figure 4-56;
4-67
Repair Procedures
7. Remove other parts such as washer, cotter pin and fitting pin, as shown in Figure 4-57.
Installation
A new electromagnet assembly can be installed according to the above-mentioned steps 1-7.
4-68
Repair Procedures
You can refer to this section when removing or replacing the entire sample compartment
assembly or its special parts.
Tool
2# cross screwdriver, 2.5/1.5/3mm hex wrench, nipper pliers, slotted head screwdriver and
diagonal pliers.
Removal
Switching elements
1. Cut the belts on the switch cable support, as shown in Figure 4-58;
4-69
Repair Procedures
2. Loosen the M3X5 screw on the switch inductor and then remove the inductor;
3. Loosen the two M3X5 screws on the gear and then remove the gear and bush;
4. Loosen the two M3X5 screws on the tube holder. Remove the slotted washer on the
electromagnet side and then remove the revolution axis;
4-70
Repair Procedures
4-71
Repair Procedures
2. Unscrew the M4X20 (Figure 4-62) screw in the center of the rotation body, and then
remove the tube compartment from the rotation body;
3. Unscrew the 2 retaining screws fixing the lock plate and then remove the lock plate.
4-72
Repair Procedures
Installation
4-73
Repair Procedures
4-74
Repair Procedures
To completely wash the exterior of the lower probe end, the probe position shall be adjusted
according to the probe wipe. The probe assembly is designed to have a fixed start position.
Since error may exist, the pre-sample position of the probe needs to be calibrated before the
analyzer leaves the factory, and users don’t need to calibrate it again. Service engineers
should readjust the above-mentioned two positions of the probe when replacing the DOM,
probe assembly, sample compartment assembly, drive board and CPU board.
Adjustment
For password level of service engineer or above, press [MENU] to enter the “System Test”
screen, then move the cursor to “Elevator motor”, press [ENTER] and then [ASPIRATE].
The “Adjust sample probe position” window, as shown in Figure 4-65, is displayed. Move
the dotted line frame to “Start Position” and then press [ENTER]. The selected option
appears in black. Press [↑] and [↓] to adjust the probe tip position, ensuring it is level to the
inlet of the probe wipe. When the probe tip is in required position, press [ENTER] to save the
adjustment.
Error may exist when the sample probe is installed. To better aspirate samples, the
pre-sample position of the probe needs to be calibrated before the analyzer leaves the factory,
and users don’t need to calibrate it again. Service engineers should readjust the start position
and pre-sample position of the probe when replacing the DOM, probe assembly, sample
compartment assembly, drive board and CPU board.
4-75
Repair Procedures
Adjustment
Adjustment of the pre-sample position is generally the same with the start position. However,
in the “Adjust sample probe position” window, move the dotted line frame to “Pre-sample
Position” and then press [ENTER]. The selected option appears in black. Press [↑] and [↓] to
adjust the probe tip position, ensuring it is aligned to the surface of the tube holder. When the
probe tip is in required position, press [ENTER] to save the adjustment.
Note: Sometimes 1step forward or backward is ideal for the probe tip. In this case, either of
the 2 steps can be set to be the home position.
Verification
Check whether the sample probe stops in the given position of the probe wipe without alarm.
If it does, the adjustment is successful, otherwise readjust it.
z Be sure to close the compartment door when adjusting the sample probe
position.
z The probe tip is sharp and may contain biohazardous materials. Exercise
caution to avoid contact with the probe when working around it.
4-76
Repair Procedures
Purpose
Due to assembly error, the sample probe probably can not move to the center of each hole.
Therefore, the three positions (right, middle and left) need to be calibrated before the
analyzer leaves the factory, and users don’t need to calibrate them again. Service engineers
should readjust the start position and pre-sample position of the probe when replacing the
DOM, probe assembly, sample compartment assembly, drive board and CPU board.
Adjustment
For password level of service engineer or above, press [MENU] to enter the “System Test”
screen, then move the cursor to “Rotation motor”, press [ENTER] and then [ASPIRATE].
The “Rotation adjustment” window, as shown in Figure 4-66 is displayed. Move the dotted
line frame to “Middle Position” and then press [ENTER]. The selected option appears in
black. Check whether the sample probe is in the center of the tube by selecting “Elevator
motor”. Press [←] and [→] to adjust the probe position. When it is in the center of the tube,
press [ENTER] to save the adjustment.
Enter the “Rotation adjustment” window to move one step forward and backward, save the
result and exit to the “System Test” screen. Check whether alarms occur when the sample
probe is in the two positions. If no alarm occurs, the position is acceptable. Enter the
“Rotation adjustment” window again to reset the home position. The home position and
critical status confirmation procedure of the left/right position of the probe is similar to that of
the middle position, except that the sample probe should be in the middle of the WBC bath or
the RBC bath.
4-77
Repair Procedures
Verification
If no alarm regarding rotation motor error occurs, the adjustment is successful, otherwise
readjust the rotation position.
z The probe tip is sharp and may contain biohazardous materials. Exercise
caution to avoid contact with the probe when working around it.
4-78
Repair Procedures
Purpose
To get accurate test results, the gain shall be adjusted again after replacement of the analog
board, aperture or the entire bath, DOM, and after long use of the analyzer.
Adjustment
1. Enter the service engineer password to access to the “Setup” Æ “Gain” screen, as
shown in Figure 4-67;
2. To set the WBC gain with EDTA-K2 venous blood: Run the anticoagulant venous blood in
whole blood mode. Adjust the WBC (whole blood) gain to make most sample histograms
end at 300-350. At least 5 normal samples shall be used;
3. To set the WBC gain with capillary blood: Run the diluted capillary blood. Adjust the WBC
(predilute) gain to make most sample histograms end at 300-350. At least 5 normal
samples shall be used.
1. Enter the service engineer password to access to the “Setup Æ Gain” screen, as shown
in Figure 4-67;
2. Run a normal control two times on the “Count” screen and take the average of the two
4-79
Repair Procedures
Assume the expected MCV result is 90.0fL and the obtained MCV result is 82.0fL.
Then
ExpectedMCV 90.0
× 100%= × 100% = 109.8%
ActualMCV 82.0
1. Enter the service engineer password to access to the “Setup Æ Gain” screen, as shown
in Figure 4-67;
2. Adjust the HGB gain to make the blank voltage of HGB about 4.5V.
Verification
1. Run several normal samples in whole blood and predilute modes and then check if the
WBC sample histograms end at 300-350;
2. Run controls on the “Count” screen and check if the MCV result is close to the reference
value;
3. Enter the “Service Æ System Status” screen, check if the blank voltage of HGB is about
4.5V.
4-80
5 Troubleshooting
5-1
Troubleshooting
5.2 Troubleshooting
1. According to the indicated errors, find the corresponding error in the table below;
2. Remove the errors according to the probable causes and recommended action.
VC Pressure The pressure inside 1. Air leakage from the fluidic and/or vacuum
Low the vacuum chamber chamber;
does not reach the
2. Poor connection of the vacuum transducer;
expected value within
the given time. 3. Pressure pump error;
5-2
Troubleshooting
5-3
Troubleshooting
photocoupler;
2.5mL & 50µL The motor does not 1. Pressed or blocked tubes;
Syringe Motor reach the expected
2. Poor contact of the drive line;
Error position.
3. Abnormal motor;
5-4
Troubleshooting
WBC Interrupt Something is wrong 1. Enter the “Service → System Test” screen
Error with the A/D part of and check the WBC AD interrupt;
the CPU board.
2. The error will be removed if the test result is
normal.
RBC Interrupt Something is wrong 1. Enter the “Service → System Test” screen
Error with the A/D part of and check the RBC AD interrupt.
the CPU board.
2. The error will be removed if the test result is
normal.
PLT Interrupt Something is wrong 1. Enter the “Service → System Test” screen
error with the A/D part of and check the PLT AD interrupt;
the CPU board.
2. The error will be removed if the test result is
normal.
Deviated Tube The tube position 1. Open the sample compartment door and
Position where a sampling readjust the sample position;
shall run deviates
2. The error will be removed if the adjustment is
from sample position.
successful. Or, readjust again.
5-5
Troubleshooting
5-6
Troubleshooting
HGB Error HGB blank voltage 1. No liquid in the bath during counting;
within 0 V - 3.2 V or
2. Reagents expired or contaminated.
4.9 V - 5 V.
HGB Adjust HGB blank voltage 1. HGB light installed improperly;
within 3.2 V - 3.4 V or
2. HGB light strength set improperly;
4.8 V - 4.9 V.
3. Dirty WBC bath;
WBC Clog WBC counting for too 1. WBC channel reference count time set
long or no counting incorrectly;
WBC Bubbles WBC counting for too 1. WBC channel reference count time set
short incorrectly;
RBC Clog RBC counting for too 1. RBC channel reference count time set
long or no counting incorrectly;
RBC Bubbles RBC counting for too 1. RBC channel reference count time set
short incorrectly;
5-7
Troubleshooting
Recorder Com Something is wrong 1. Poor connection between the recorder and
Error with the the CPU board;
communication
2. Damaged recorder.
between the recorder
and the CPU board.
Printer out of Something is wrong Printer paper running out or not properly installed.
Paper with the installation of
the printer paper.
Recorder out of Something is wrong Something is wrong with the installation of the
Paper with the installation of recorder paper.
the recorder paper.
Recorder too Something is wrong 1. The recorder has been running for a too long
Hot with the recorder. period;
5-8
Troubleshooting
3. Damaged recorder.
5-9
Troubleshooting
Abnormal pulse graph will be generated if some parts of the analyzer go wrong. Since the
analyzer can store the original pulse graphs automatically, checking the abnormal pulse
graphs helps to troubleshoot the analyzer.
To check the pulse graphs, you enter the password “3210” and press [ENTER] key to exit.
Pressing the [1], [2] and [3] respectively at the “Count” screen, the original pulse graphs of
WBC, RBC and PLT channels appear. Do not use the channel data length to troubleshoot the
analyzer.
Figure 5-1 WBC pulse graph of normal Figure 5-2 Pulse graph of normal WBC
sample background
Figure 5-3 RBC pulse graph of normal Figure 5-4 Pulse graph of normal RBC
sample background
5-10
Troubleshooting
Figure 5-5 PLT pulse graph of normal Figure 5-6 Pulse graph of normal PLT
sample background
Figure 5-7 Severe interference in WBC Figure 5-8 Severe interference in WBC
channel (Background) channel (Normal sample)
5-11
Troubleshooting
Figure 5-11 Severe interference in PLT Figure 5-12 Severe interference in PLT
channel(Background) channel(Normal sample)
5-12
Troubleshooting
Figure 5-17 Slight interference in RBC Figure 5-18 Sample of too dense
channel(Normal sample) concentration in RBC channel
( Does not occur in normal situation )
Figure 5-19 Slight interference in PLT Figure 5-20 Sample of too dense
channel(Normal sample) concentration in PLT channel
(Does not occur in normal situation)
5-13
Troubleshooting
Figure 5-23 Insufficient liquid in WBC bath Figure 5-24 Interference in RBC
during count channel from tubing
Feature: data length increases, the base
line of signal is not stabile.
Figure 5-25 Insufficient liquid in RBC bath Figure 5-26 Interference in PLT
during count channel from tubing
Feature: data length increases, the base
line of signal is not stabile.
5-14
Troubleshooting
Background abnormal
5-15
Troubleshooting
Clog
Yes
Yes
End
End
No
Replace SV3, clog
Yes End
solved?
No Yes
Yes
End No
End
5-16
Troubleshooting
HGB error
5-17
Troubleshooting
Vacuum low
No
No
No
No
No
No
No
No
No No
No No
5-18
Troubleshooting
Display error
Replace the
No anything displayed? Yes display(screen), Yes End
problem solved?
No
No
No
No
No
No
No
No
No No
5-19
Troubleshooting
Bubbles
Yes
Yes
End
End
No
Replace SV3, clog
solved?
Yes End
No Yes
Yes
End No
5-20
Troubleshooting
Lyse out
py
Replace new solution,
Solution enough? No Yes End
problem solved?
No
Yes
Does the plastic washer on the senor Refix the washer, problem
Yes Yes End
fall off? solved?
No
No
No
Yes
No
5-21
Troubleshooting
48V low
No
Yes
No
Yes
No
5-22
Troubleshooting
Measure more
samples(controls) and
Is WBC value correct? No Yes End
calibrate or check the 50ul
syringe, problem solved?
No
Yes
No
Yes
No
Yes
Yes No
End
Connect with manufacturer
5-23
Troubleshooting
No
Yes
No
No
No
Yes
Yes
No
End
5-24
Troubleshooting
Background noise
When the count data of a container filled with diluent just exceeds the tolerable value of the
parameter shown in the table in ”Background Noise Check” of this chapter, the seven
possible causes of error are:
1. Dirty diluent;
3. Dirty baths;
4. Dirty valves;
7. Tubing noise affecting the counting/calculation circuit: Liquid flow with some remained
electric charges.
z Before storing the instrument for a long time, clean and empty all fluid
tubings in the instrument with distilled water. Otherwise, the diluent salt
adheres to the baths and cannot be removed easily.
Cause (5)
−Securely ground the instrument, including any optional units such as external printer.
Cause (6)
− Keep the instrument away from the vibration source.
Cause (7)
− Securely ground the tubing.
5-25
Troubleshooting
For reproducibility
This subsection describes the cases when the reproducibility for the following measuring
parameter or calculated parameter is poor.
− PLT
− HGB
− WBC
− RBC
− Data other than red cell indexes (MCV, MCH and MCHC)
− RBC and PLT coefficient related parameters
− HCT and MCV
5-26
Troubleshooting
5-27
Troubleshooting
The sample probe position is not Adjust the probe position using localizer.
correct
The 3-way valve (SV11) cannot Replace the 3-way valve (SV11).
drain liquid empty.
The 3-way valve (SV11) is dirty 1. Clean the valves;
The constant current is not stable Change the transformer or analog board.
The diuent syringe exists bubbles Remove the bubbles from the syringe.
Faulty the Analog board Replace the analog board.
Others
Initiation time is too long Disk on module is broken. Replace hard disk (DOM)
Keyboard, some buttons no CPU board or keyboard is Replace the CPU board or
response broken. keyboard.
LCD screen, there is a line on 1. CPU board is broken if 1. Replace the CPU board;
the screen the line is a dot line and the 2. Replace the LCD screen.
position is fixed;
LCD screen is dark 1. The color palette setting 1. Re-set the color palette
is not correct; to 8-color;
5-28
Troubleshooting
No sample inspired and no SV4 valve doesn’t work. Replace the SV4 valve.
diluent injected
No initialization, but exist The CPU board is broken. Replace the CPU board.
backlight and a moment it
becomes black, the indicators
flash like saver
No backlight The inverter is broken. Replace the inverter.
Power-on is normal, but no The CPU board is broken. Replace the CPU board.
display, no initiation, no
response
WBC bath full and liquid The V12 valve doesn’t work Clean or replace the V12
overflows or dirty or doesn’t work well. valve.
WBC differential part is WBC channel gain is not Adjust WBC channel gain.
abnormal, and it’s end point correct.
at 200fL
5-29
6 List of Spare Parts
When you order the parts or components from MINDRAY, please provide part description and
MINDRAY code, because it will help us to satisfy your requirement as soon as possible.
Using parts and components provided by MINDRAY can ensure your analyzer’s optimal
work.
6-1
List of Spare Parts
6-2
List of Spare Parts
6.3 Consumables
Table 6-3 Consumbles list
6-3
List of Spare Parts
6.4 Tubing
6.4.1 Main fluidic tubing connection table
See the Table 6-4 for the connection. For the whole unit fluidic connection graph, refer to
Appendix A.
No. Tube No. Start from Stop at Tube Tube length (mm) Remarks
type
T1 T40 Sample Tube 10# 450 Buffer tubing
1
probe
2 T2 T40 T91 Tube 4# 750
It connects T1
and T2. T1
3 T40 T1 T2 Tube 1# 40 must be at least
12mm deep into
T40.
4 T3 50µL SV4-NC Tube 9# 30 /
Syringe
5 T4 SV4-C SV1-NO Tube 2# 140 /
6 T5 SV3-C 10mL Tube 2# 40 /
Syringe
7 T6 SV3-NC SV1-C Tube 2# 110 /
8 T7 SV4-NO SV5-C Tube 2# 110 /
9 T8 SV5-NO T84 Tube 2# 50 /
10 T9 SV5-NC T85 Tube 2# 50 /
11 T10 SV6-C 2.5mL Tube 2# 45 /
Syringe
12 T11 SV6-NC T86 Tube 2# 40 /
13 T12 SV6-NO Connecter Tube 2# 1100 /
to lyse
container
14 T13 J12 SV17-2 Tube 1# 30 /
15 T14 J21 J12 Tube 1# 20 /
16 T15 J12 SV7-2 Tube 1# 40 /
17 T16 J18 F1 Tube 7# 20 /
18 T17 J18 F3 Tube 7# 20 /
19 T18 Vacuum Below Tube 5# 40 /
chamber WBC
metering
tube
20 T19 Vacuum Below Tube 5# 40 /
6-4
List of Spare Parts
chamber RBC
metering
tube
21 T22 SV8-2 Middle of Tube 2# 380 /
WBC back
bath
22 T23 SV18-2 Middle of Tube 2# 370 /
RBC back
bath
23 T30 J13 SV14-2 Tube 1# 230 /
24 T31 J14 SV13-2 Tube 1# 100 /
25 T32 J5 J22 Tube 1# 650 /
26 T57 J5 SV14-1 Tube 1# 35 /
27 T58 J5 SV13-1 Tube 1# 40 /
28 T35 Beside SV15-1 Tube 1# 430 /
upper part
of RBC
isolated
chamber
29 T36 Beside SV16-1 Tube 1# 470 /
upper part
of WBC
isolated
chamber
30 T37 J4 SV15-2 Tube 1# 40 /
31 T38 J4 SV16-2 Tube 1# 40 /
32 T39 J4 SV9-NO Tube 1# 320 /
33 T41 SV9-C Pressure Tube 1# 620 /
chamber
34 T42 J18 Pressure Tube 5# 40 /
pump inlet
35 T43 Pressure Pressure Tube 5# 110 /
chamber pump
outlet
36 T44 SV9-NC Vacuum Tube 1# 250 /
chamber
37 T45 J3 SV11-2 Tube 1# 20 /
38 T46 J3 SV12-2 Tube 1# 45 /
39 T47 Vacuum SV10-2 Tube 1# 130 /
chamber
40 T48 SV10-1 J2 Tube 1# 20 /
41 T49 J2 J3 Tube 1# 25 /
42 T52 SV2-1 J1 Tube 1# 20 /
6-5
List of Spare Parts
43 T53 J1 J2 Tube 1# 25 /
44 T54 J1 Waste Tube 5# 400 /
pump inlet
45 T55 J19 Waste Tube 5# 300 /
pump
outlet
46 T56 J20 SV3-NO Tube 1# 70 /
47 T61 Lower J27 Tube 1# 40 /
interface
of RBC
isolated
chamber
48 T62 Upper J28 Tube 1# 50 /
interface
of WBC
isolated
chamber
49 T63 Pressure Pressure Tube 1# 620 /
chamber sensor
50 T64 Vacuum Vacuum Tube 1# 420 /
chamber sensor
51 T67 J20 J17 Tube 5# 1050 /
52 T72 J21 F2 Tube 1# 20 /
53 T73 J21 F4 Tube 1# 20 /
54 T89 J27 SV11-1 Tube 1# 100 /
6-6
List of Spare Parts
Tube
2 T59 J8 SV7-1 60
2#
Tube
3 T21 J9 SV18-1 50
2#
Tube
4 T60 J9 SV17-1 60
2#
Tube
5 T26 J6 J14 15
1#
Tube
6 T24 J6 J23 30
1#
Tube
7 T25 J6 J24 30
1# RBC assembly
Top of the RBC Tube
8 T80 J23 15
back bath 1#
Bottom interface
Tube
9 T81 J24 of the RBC back 15
1#
bath
Tube
10 T29 J7 J13 15
1#
Tube
11 T27 J7 J25 30
1#
Tube
12 T28 J7 J26 30 WBC assembly
1#
Top of the WBC Tube
13 T82 J25 15
back bath 1#
Bottom of the Tube
14 T83 J26 15
WBC back bath 1#
Bottom of the Tube
15 T65 T33 50
RBC front bath 1#
Tube
16 T33 T87 T65 120 RBC assembly
4#
Top of RBC Tube
17 T87 T33 25
isolated chamber 1#
Below the WBC Tube
18 T66 T34 50
front bath 1#
Tube
19 T34 T88 T66 120 WBC assembly
4#
Top of WBC Tube
20 T88 T34 25
isolated chamber 1#
Lower interface of Tube Sample probe
21 T50 SV1-NC 650
the probe wipe 8# assembly
Upper interface Tube
22 T51 SV2-2 450
of the probe wipe 8#
6-7
List of Spare Parts
6-8
List of Spare Parts
assembly
6-9
List of Spare Parts
6.5 Cable
Figure “No.” in the Table 6-9 is corresponding to the “No.” in Figure B-1 of Appendix B.
“MINDRAY code” in the table indicates the cable type. The quantity of No. listed in the “Figure
No.” column of the table represents the quantity of cables used.
MINDRAY
No. Description Connection
Code
Power cord of CPU 2800-20-28753 2800-30-28670<->3005-30-44651
1
board and drive board 2800-30-28670<->3003-30-34910
Analog board power 2800-20-28765 2800-30-28670<->3003-30-34903
2
cord
Temperature sensor 3006-20-74763 3006-20-74763<->3003-30-34910
3
connecting line
Patch cord of LCD 2800-20-28768 Screen <-> inverter
4
backlight board
5 LCD connecting line 3006-20-74764 3003-30-34912<->3003-30-34910
6 Filter power cord 2800-20-28820 Filter <->3003-30-28670
Safety earth 2800-20-28821 Filter <-> Power supply mount board
7
connection
2800-20-28822 3003-30-34910<->3003-30-34903
8 34PIN connecting line
3003-30-34910<->3005-30-44651
Inverter connecting 2800-20-28849 Inverter <->3003-30-34905
9
line
BNC head connecting BNC head<->3003-30-34899
10 3001-21-07090
line
Meniscus sensor 3006-20-74765 3003-30-34910<->3001-21-07090
11
connecting line
12 50µL motor line 3006-20-74766 50µL motor <->3005-30-44651
13 10mL motor line 3006-20-74767 10mL motor <->3005-30-44651
X motor connecting 3003-20-34936 X motor<->3005-30-44651
14
line
Y motor connecting 3003-20-34937 Y motor<->3005-30-44651
15
line
Valve/Pump control 3006-20-74768 Valve, pump <->3005-30-44651
16
line
Volumetric unit 3003-20-34896 3003-30-34909<->3003-30-34910
17
connecting line
18 Keypad and recorder 3003-20-34897 3003-30-34910<->3003-30-34905
6-10
List of Spare Parts
6-11
7 Appendices
A Fluidic Diagram
A-1
B Electric Connection Diagram
B-1
Appendices
B-2
Appendices
B-3
Appendices
B-4
Appendices
B-5
Appendices
B-6
Appendices
B-7
Appendices
B-8
Appendices
B-9
Appendices
B-10
C Test Point of Hardware Board
C-1
Appendices
TP22 V+12 +12.8V power Not used for this board, isolated from
supply other power supply of this board and
supplied to the recorder and keypad,
affecting the recorder, buzzer and
backlight of the LCD
TP23 G+12 Power ground Ground of the +12.8 power supply
TP24 HGB HGB analog Input to the A/D of this board, marked “H”
signal on the PCB
TP25 RBC RBC analog Input to the A/D of this board, marked “R”
signal on the PCB
TP26 WBC WBC analog Input to the A/D of this board, marked “W”
signal on the PCB
TP27 PLT PLT analog signal Input to the A/D of this board, marked “P”
on the PCB
TP28 VREF A/D reference 2.5V, ensuring the A/D works normally
voltage
C-2
Appendices
C-3
Appendices
C-4
D Valve Test
Name Description
V1 Together with V3 and the 10mL syringe, it supplies cleanser for the probe wipe.
Moving simultaneously with the fluidic pump, it washes and dries the outer
V2
surface of the sample probe.
It dispenses diluent aspirated by the 10mL syringe into the corresponding
V3
tubing of the analyzer by switching.
Moving simultaneously with V3 and the 10mL syringe, it aspirates sample by
V4
the sample probe or washing the inner surface of the sample probe.
Together with V3 and the 10mL syringe, it dispenses diluent into the WBC bath
V5
or RBC bath by switching.
It dispenses lyse aspirated by the 2.5mL syringe into the WBC bath by
V6
switching.
When the vacuum chamber is in the negative pressure status, it opens to drain
V7
WBC metering tube.
When the vacuum chamber is in the negative pressure status, it opens solely
V8 for WBC counting; Opening with the V14 simultaneously, it aspirates rinse to
wash the WBC back bath.
Its open connects the vacuum chamber with pressure chamber to make a
V9
positive pressure for the vacuum chamber, and this helps to flush the aperture.
Moving simultaneously with waste pump, it drains the vacuum chamber or
V10
establish vacuum in the vacuum chamber.
V11 It moves simultaneously with the waste pump to drain the RBC bath.
V12 It moves simultaneously with the waste pump to drain the WBC bath.
When the vacuum chamber is in the negative pressure status, Its open with
V13
V18 simultaneously can aspirate rinse to wash the RBC bath.
When the vacuum chamber is in the negative pressure status, Its open with V8
V14
simultaneously can aspirate rinse to wash the WBC bath.
V15 It opens to bubble-mix the RBC bath.
When the vacuum chamber is in the negative pressure status, it opens to drain
V17
the RBC metering tube.
When the vacuum chamber is in the negative pressure status, it opens solely
V18 for RBC counting; Opening with the V13 simultaneously, it aspirates rinse to
wash the RBC back bath.
D-1
P/N: 3006-20-74778 (V1.0)