Professional Documents
Culture Documents
PTC 50 2002
PTC 50 2002
This Standard will be revised when the Society approves the issuance of a new
edition. There will be no addenda issued to this edition.
This code or standard was developed under procedures accredited as meeting the criteria for American
National Standards. The Standards Committee that approved the code or standard was balanced to assure
that individuals from competent and concerned interests have had an opportunity to participate. The
proposed code or standard was made available for public review and comment that provides an opportunity
for additional public input from industry, academia, regulatory agencies, and the public-at-large.
ASME does not “approve,” “rate,” or “endorse” any item, construction, proprietary device, or activity.
ASME does not take any position with respect to the validity of any patent rights asserted in connection
with any items mentioned in this document, and does not undertake to insure anyone utilizing a standard
against liability for infringement of any applicable letters patent, nor assume any such liability. Users of
a code or standard are expressly advised that determination of the validity of any such patent rights, and
the risk of infringement of such rights, is entirely their own responsibility.
Participation by federal agency representative(s) or person(s) affiliated with industry is not to be
interpreted as government or industry endorsement of this code or standard.
ASME accepts responsibility for only those interpretations of this document issued in accordance
with the established ASME procedures and policies, which precludes the issuance of interpretations by
individuals.
Copyright © 2002 by
THE AMERICAN SOCIETY OF MECHANICAL ENGINEERS
All rights reserved
Printed in U.S.A.
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
CONTENTS
Foreword . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Committee Roster . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Board Roster . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
INTRODUCTION
............................................................................ 1
3 Guiding Principles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2 Agreements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.3 Test Boundary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.4 Test Plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.5 Preparation for Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.6 Parameters to be Measured or Determined During the Test Period . . . . . . . . . 11
3.7 Operation of the Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.8 Calculation and Reporting of Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.9 Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
5 Computation of Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.2 Computation of Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.3 Computation of Electric Power Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
iii
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
5.4 Computation of Thermal and Mechanical Outputs . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.5 Computation of Average Net Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.6 Computation of Efficiencies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.7 Correction of Test Results to Reference Conditions . . . . . . . . . . . . . . . . . . . . . . . . 29
iv
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FOREWORD
During the mid 1990s the importance of developing fuel cell standards was recognized.
Fuel Cell power plants were in the early stages of commercialization. Potential
applications included vehicular power, on-site power generation, and larger scale
dispersal power generators. There was a growing demand to produce industry standards
that would keep pace with the commercialization of this new technology.
ASME had a very active Fuel Cell Power Systems technical committee within the
Advanced Energy Systems Division. Through its volunteer membership, it recommended
the formation of a standards committee to work on developing a fuel cell standard.
ASME Codes and Standard Directorate undertook this task. On October 14, 1996 the
Board on Performance Test Codes voted to approve the formation of a performance
test code committee, PTC 50.
This Committee had its first meeting on January 23-24, 1997. The membership
consisted of some 18 fuel cell experts from Government, academia, manufacturers,
and users of fuel cells. Ronald L. Bannister; Westinghouse Electric Corporation; retired,
chaired the first meeting. He had been appointed by the Board on PTC as the Board
Liaison member to the committee. He chaired and supervised the committee’s activities
until permanent officers were elected from the membership.
In the Fall 2001, the Committee issued a draft of the proposed Code to Industry
for review and comment. The comments were addressed in February 2002 and the
Committee by a letter ballot voted to approve the document on March 29, 2002. It
was then approved and adopted by the Council as a standard practice of the Society
by action of the Board on Performance Test Codes voted on May 6, 2002. It was
also approved as an American National Standard by the ANSI Board of Standards
Review on July 3, 2002.
v
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
NOTICE
All Performance Test Codes MUST adhere to the requirements of PTC 1, GENERAL
INSTRUCTIONS. The following information is based on that document and is included
here for emphasis and for the convenience of the user of this Supplement. It is expected
that the Code user if fully cognizant of Parts I and III of PTC 1 and has read them
prior to applying this Supplement.
ASME Performance Test Codes provide test procedures which yield results of the
highest level of accuracy consistent with the best engineering knowledge and practice
currently available. They were developed by balanced committees representing all
concerned interests. They specify procedures, instrumentation, equipment operating
requirements, calculation methods, and uncertainty analysis.
When tests are in accordance with a Code, the test results themselves, without
adjustment for uncertainty, yield the best available indication of the actual performance
of the tested equipement. ASME Performance Test Codes do not specify means to
compare those results to contractual guarantees. Therefore, it is recommended that the
parties to a commercial test agree before starting the test and preferably before signing
the contract on the method to be used for comparing the test results to the contractual
guarantees. It is beyond the scope of any Code to determine or interpret how such
comparisons shall be made.
vi
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
PERSONNEL OF PERFORMANCE TEST CODE
COMMITTEE 50
FUEL CELL POWER SYSTEMS PERFORMANCE
(The following is the roster of the Board at the time of approval of this Code.)
OFFICERS
A. J. Leo, Chair
K. Hecht, Vice Chair
J. H. Karian, Secretary
COMMITTEE PERSONNEL
D. H. Archer, Carnegie Mellon University
P. J. Buckley, Energy Alternatives
S. Comtois, H Power Enterprises of Canada, Inc.
J. S. Frick, SCANA Corp.
K. Hecht, UTC Fuel Cells
F. H. Holcomb, U.S. Army Corps of Engineers
J. H. Karian, The American Society of Mechanical Engineers
B. Knaggs, Ballard Generation Systems
M. Krumpelt, Argonne National Laboratory
A. J. Leo, Fuel Cell Energy
A. Skok, Alternate, Fuel Cell Energy
R. M. Privette, OMG Corp.
L. A. Shockling, Siemens-Westinghouse Power Corp.
R. P. Wichert, U.S. Fuel Cell Council
M. C. Williams, U.S. DOE, NETL
vii
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
BOARD ON PERFORMANCE TEST CODES
OFFICERS
S. J. Korellis, Chair
J. R. Friedman, Vice Chair
W. O. Hays, Secretary
COMMITTEE
PERSONNEL
viii
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002
Fuel cells convert the energy of a fuel directly Section 1 defines the objective and scope of this
into electricity, eliminating the combustion stage Code. Section 2 is dedicated to defining a fuel cell
that is characteristic of heat engines, and not requir- system and to definitions of terms. It also contains
ing any moving parts. Instead, the fuel molecules a brief discussion of the major types of fuel cells.
(usually hydrogen often derived from hydrocarbon In Section 3, methodology of establishing test proto-
fuels) interact with the surface of an anode material col is outlined. Instrumentation for measuring the
to form reaction products, liberating electrons. The energy of the feed stream as well as of the exiting
electrons flow through the electric load to the cath- gases and liquids is given in Section 4, as is the
ode where they react with an oxidant, typically instrumentation for measuring electric power. Sec-
oxygen from air. Ions migrate between the electrodes tion 5 describes how the efficiency of the systems
through the ionically conducting electrolyte to com- shall be calculated from the measurements, and
how corrections for nonstandard conditions shall be
plete the circuit. The product of this electrochemical
made.
energy conversion process is water, but unlike heat
Typically, this performance test code would be
engines, the process can take place at close to
used for an independent verification of the perform-
ambient temperature, or can also be conducted at ance of a particular fuel cell system by a customer
higher temperatures, depending on the types of or test agency. In the view of the members of the
anode, electrolyte, and cathode materials. Committee, the described procedures are rigorous,
Since fuel cells are not heat engines, the efficiency and the test will require committing significant re-
of a fuel cell system is not limited by the Carnot sources. For the casual user of fuel cells, it will
principle. It can, in fact, vary over a fairly wide suffice to determine the electric output of the system
range. When the current density of the fuel cell is under steady state conditions, and to measure the
very low, the energy conversion efficiency ap- fuel feed rate. As mentioned above, the efficiency
proaches the ratio of the Free Energy of Combustion of a fuel cell system varies significantly with power
of the fuel divided by the Enthalpy of Combustion. density. At power densities below the design point,
For methane this limit is 94%. However, such an the efficiency will usually increase, and it will de-
operating mode would require a very large fuel cell crease when the power output exceeds the design
and would be too expensive in most applications. point. One of the characteristics of fuel cells is the
ability to operate them over a wide power range,
In practice, fuel cell systems are designed to
even exceeding the design point by 50% for a few
operate at a power density reflecting the most eco-
minutes. Under dynamic operating conditions the
nomical trade-off of fuel and capital costs. At the efficiency of a fuel cell would be different than at
design point of the system the power output of the the design point, and would probably be higher,
system is specified by the manufacturer for certain since most loads contain significant segments of
standard conditions of fuel and air. It is the purpose low-power operation and normal system control
of this Code to define in a commonly acceptable (e.g., for fuel flow) responds fairly quickly to these
manner how the power output and the energy input load conditions. Measuring the efficiency under dy-
should be measured and how the efficiency should namic conditions goes beyond the scope of the
be calculated. document.
1
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
SECTION 1
OBJECT AND SCOPE
2
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
SECTION 2
DEFINITIONS AND DESCRIPTION OF TERMS
3
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
DC power Inverted AC
Power or otherwise
Conditioning regulated
power
Fuel Fuel
supply Anodes Heat Cogeneration
Processing
Recovery Heat Recovery
Cathodes
Exit gas
Fuel cell flows
Oxidant stack(s)
Supply
Molten Carbonate Fuel Cell (MCFC): the electrolyte run on hydrogen fuel or hydrocarbon fuels, in which
in this fuel cell is usually a combination of alkali case the fuel processing system includes provisions
carbonates retained in a ceramic wicking matrix. for hydrogen generation. Typical processes for hydro-
The fuel cell operates at 600°C to 700°C where the gen generation include steam reforming, partial oxi-
alkali carbonates form a highly conductive molten dation or auto thermal reforming and CO removal
salt. At the high operating temperatures in MCFCs, (shift reactors and/or selective oxidation reactors).
Ni (anode) and nickel oxide (cathode) are adequate PEMFC systems may be designed to run at atmo-
catalysts for the cell reactions. MCFC systems can spheric pressure or higher pressures, depending on
be run at ambient or elevated pressures. Hydrocar- application requirements.
bon fuels can be utilized with hydrogen generating
Solid Oxide Fuel Cell (SOFC): the electrolyte in
reforming subsystems in the balance-of-plant, or
this fuel cell is a solid, nonporous metal oxide,
utilizing the high temperature of the cells to promote
usually Y2O3 stabilized ZrO2. The cell operates at
reforming within the cell stacks.
800°C to 1000°C where ionic conduction of oxygen
Phosphoric Acid Fuel Cell (PAFC): aqueous phos- ions across the electolyte takes place.
phoric acid is used for the electrolyte in this fuel
cell, which operates at 150°C to 220°C. The system
can be run at ambient or elevated pressures, and 2.3 FUEL CELL POWER SYSTEMS
on hydrogen or hydrocarbon fuels. On hydrocarbon A generic fuel cell power system block diagram
fuels the fuel processing system includes provisions is shown in Fig. 2.1. A fuel cell combines fuel
for hydrogen generation. Typical processes for hydro- (usually hydrogen derived from hydrocarbon fuels)
gen generation include steam reforming, partial oxi- and oxygen (usually from air) to produce DC power,
dation or auto thermal reforming. PAFC are signifi- water, and heat. In cases where CO and CH4 are
cantly less sensitive to CO than PEMFC or AFC, so reacted in the cell to produce hydrogen, CO2 is
CO removal subsystems are less elaborate. also a product. These reactions must be carried out
Polymer Electrolyte Membrane Fuel Cells: also re- at a suitable temperature and pressure for fuel cell
ferred to as Proton Exchange Membrane Fuel Cell operation. A system must be built around the fuel
(PEMFC); the electrolyte in this fuel cell is an ion cells to supply air and clean fuel, convert the power
exchange membrane (fluorinated sulfonic acid poly- to a more usable form such as grid quality AC
mer or other similar polymers) that is an excellent power, and remove the depleted reactants and heat
hydrogen ion conductor. PEMFC systems typically that are produced by the reactions in the cells. The
operate at less than 120°C, because of temperature components of that system are typically as follows:
limitations of the polymer electrolyte. The low op- (a) Cogeneration Heat Recovery. Consists of
erating temperature confers a degree of flexibility equipment to export thermal energy for use outside
on the system (e.g.,startup times will be shorter than the fuel cell system.
for higher temperature systems), but also means the (b) Fuel Supply. Can be as simple as a hookup
cell catalysts are more sensitive to CO poisoning than to a filtered, regulated natural gas source, or can
other systems. PEMFC systems can be configured to include a gas compressor.
4
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
(c) Fuel Cell Stack. Consists of the fuel cell stack components in a fuel cell power system besides the
or stacks that produce the DC power. electrochemical fuel cell stacks. The BOP can in-
(d) Fuel Processing. Involves a variety of pro- clude fuel processing equipment, heat recovery
cesses, depending on the fuel cell types. If the fuel equipment, power conversion equipment, control
cell is running on bottled hydrogen, fuel processing equipment, etc.
is minimal. Typically a fuel cell power system will
be run on common hydrocarbon fuels, such as 2.5 GENERAL DEFINITIONS
natural gas. Fuel processing usually involves cleaning
accuracy: the closeness of agreement between a
to remove possible fuel cell poisons (e.g., sulfur
measured value and the true value.
compounds used in natural gas odorants). It can
include reforming of the gas to produce hydrogen base reference conditions: the values of all the
(in “externally reformed” systems) or this function external parameters; i.e., parameters outside the test
can occur inside the stacks in some types of fuel boundary to which the test results are corrected.
cells. The fuel processor may also include other Also, the specified secondary heat inputs and outputs
equipment, such as shift reactors, CO oxidizer reac- are base reference conditions.
tors, higher hydrocarbon removal, etc.
bias error: see systematic error.
(e) Heat Recovery. Systems in fuel cell power
systems usually consist of fairly conventional heat calibration: the process of comparing the response
exchange equipment, which is used to extract waste of an instrument to a standard instrument over some
heat (typically from the system exhaust stream) to measurement range and adjusting the instrument to
provide reactant pre-heat and steam generation. match the standard, if appropriate.
(f) Oxidant Supply. Usually consists of a means calibration drift: see para. 4.1.3.6.
of providing fresh air to the system, typically with
a blower or compressor. calibration shift: see para. 4.1.3.6.
(g) Power Conditioning. Involves conversion of consistent liquid or gaseous fuels: fuels with a
the DC power output from the electrochemical fuel heating value that varies less than one percent over
cells to useful power. This usually involves inversion the course of a performance test.
to produce AC power suitable for the intended
purpose, but it can also include the production of electric efficiency: the ratio of the electrical energy
regulated DC or “chopped” DC. output to the energy supplied to the power system
expressed as a percentage. It is inversely related to
heat rate.
2.4 GENERAL FUEL CELL NOMENCLATURE
emissions: nuisance discharges from power plant
Some of the common terms used to describe fuel systems which are regulated by authorities having
cell power system components are as follows: jurisdiction, such as air pollutants, waste streams
(a) Fuel Cells. Individual electrochemical cells and noise.
that produce a voltage (typically 0.5 V to 1 V each)
when provided with a supply of fuel and oxidant. heat input: the flow of fuel(s) multiplied by the
Fuel cells consist of an anode (where fuel is con- high or low heating value of the fuel(s).
sumed and electrons are liberated), a cathode (where heat rate [kJ/kW·h or Btu/(kW·h)]: heat input per
oxidant is consumed and electrons are captured), unit of power output, based on either the low or
and an electrolyte that provides a path for transfer high heating value of the fuel.
of ions between the anode and cathode reaction sites.
(b) Fuel Cell Stacks. Consist of a group of individ- heat sink: the reservoir to which the heat rejected
ual fuel cells that are physically held together in a by the system is transferred. For a cooling pond,
stack or a bundle, are fed fuel and oxidant flows river, lake, or ocean cooling system, the reservoir
together, and whose electrical output is combined. is the body of water. For an evaporative or dry air-
Fuel cell stacks may be configured individually, cooled heat exchanger system, the reservoir is the
combined into multistack modules, or they may be ambient air.
configured as part of a fully self-contained power high heating value (HHV): the heat of combustion,
system. per unit quantity of fuel, when all combustion prod-
(c) Balance of Plant (BOP). Used to refer to all uct water is condensed to liquid water and all
5
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
associated heat is recovered. This is the highest Gaussian (normal) manner, from measurement to
possible heat of combustion of a particular fuel at measurement).
specified conditions and is typically the heating
rated power: the power output of the power system
value paid for by the fuel buyer.
when operating at specified control and ambient
influence coefficient: the ratio of the change in a conditions.
result to a unit change in a parameter. secondary energy inputs: energy streams, exclusive
instrument: a tool, or device used to measure physi- of the heating value of main fuel supply crossing
cal dimensions of length, thickness, width, weight into the power plant boundary to support fuel cell
or any other value of a variable. These variables systems.
include: size, weight, pressure, temperature, fluid secondary outputs: any useful nonelectrical energy
flow, voltage, electric current, density, viscosity, and output stream which is used by an external process.
power. Sensors are included which may not, by
themselves, incorporate a display but transmit signals secondary variables: variables that are measured
to remote computer type devices for display, pro- but do not enter into the calculation.
cessing or process control. Also, included are items secondary thermal energy inputs: the additional
of ancillary equipment directly affecting the display heat inputs to the test boundary which must be
of the primary instrument (e.g., ammeter shunt), and accounted, such as cycle makeup and process con-
tools or fixtures used as the basis for determining densate return.
part acceptability.
sensitivity: see influence coefficient.
low heating value (LHV): the heat of combustion,
serialize: means an instrument has been assigned
per unit quantity of fuel when all combustion product
a unique number and that number has been perma-
water is assumed to remain as vapor. This is the
nently inscribed on or to the instrument so that it
lowest possible heat of combustion of a particular
can be identified and tracked.
fuel at specified conditions and when used in calcu-
lations results in the highest values of power plant shaft work: mechanical energy crossing the power
efficiency. plant boundary for accomplishing useful work.
measurement error (␦): the true, unknown differ- specified corrected net power test: a test run at a
ence between the measured value and the true specified corrected net power that is near to the
value. design value of interest; for example, an acceptance
test where heat rate is guaranteed at a specific load
net power: the electrical power leaving the test and partial-load tests for development of heat-rate
boundary minus any electrical power entering the curve conditions.
test boundary.
specified net power test: a test run at a specified
parasitic power: energy produced by the power net power regardless of ambient or other external
plant and used within the fuel cell system. conditions. An example of this test goal is acceptance
test on a power system with an output guarantee
precision error: see random error.
over a range of ambient temperatures.
primary variables: those used in calculations of test
specific fuel consumption [ft3/(kW·h), or lbm/(kW·h),
results. They are further classified as:
or kg/kW·h]: fuel consumption rate per unit of
(a) Class 1. Primary variables are those that have power output.
a relative sensitivity coefficient of 0.2 or greater.
(b) Class 2. Primary variables are those that have standard atmospheric conditions: 14.696 psia
a relative sensitivity coefficient of less than 0.2. (101.325 kPa), 59°F (288.15 K), and relative humidity
Refer to ASME PTC 19.1 for calculations of relative of 60%.
sensitivity coefficients. standard temperature and pressure (STP): 101.325
kPa (14.696 psia), 59°F (288.15 K).
random error (): sometimes called precision error;
the true random error, which characterizes a member systematic error (): sometimes called bias; the true
of a set of measurements. (varies in a random, systematic or fixed error, which characterizes every
6
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
member of any set of measurements from the popula- these systems thermal efficiency is synonymous with
tion. The constant component of the total measure- electrical efficiency. Thermal efficiency differs from
ment error (␦). thermal effectiveness in that it does not include
utilized heat output. In mathematical terms:
test boundary: identifies the energy streams re-
quired to calculate corrected results. thermal efficiency p (En + W)/QI total
test reading: one recording of all required test
instrumentation. where
En p net electrical energy output
test run: a group of test readings taken while the QI total p total energy into the system
fuel cell power system is operating at steady state W p shaft work
at a specified operating condition.
thermal effectiveness: percentage of input energy traceable: means records are available to show that
captured for useful purpose which includes conver- the instrument can be traced through a series of
sion to electric energy, shaft work, and thermal calibrations to an appropriate ultimate reference such
energy recovery for external uses. as National Institute for Standards and Technology
(NIST).
thermal efficiency: the ratio of the energy output
(electrical and shaft) to the energy supplied to the uncertainty (U): ±U is the interval about the mea-
power system, expressed as a percentage. In most surement or result that contains the true value for
fuel cell systems shaft work is not produced. For a given confidence level.
7
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
SECTION 3
GUIDING PRINCIPLES
8
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
Test Boundary
Net electrical
energy
Fuel
Thermal
energy
Secondary
energy
input Waste thermal
energy output
Oxidant
Emission
input
GENERAL NOTES:
Required to be measured for test calculation
Not required to be measured for test calculation
Net electrical
Oxidant energy output
input
Fuel
Thermal to Mechanical
Secondary Thermal Mechanical to Electrical
thermal input
Energy Energy
Secondary Conversion Conversion
energy input
Thermal Energy
Recovery
Shaft work
input
FIG. 3.2 FUEL CELL SYSTEM TEST BOUNDARY ILLUSTRATING INTERNAL SUBSYSTEMS
9
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
(k) requisite facilities for maintaining constancy Regardless of the test goal, the results of a perform-
of load during test; ance test per this Code will be corrected net power,
(l) number of identical tests to be conducted to corrected electrical efficiency, corrected thermal ef-
assure statistically significant accuracy; fectiveness, and corrected thermal efficiency.
(m) duration of the test runs;
(n) acceptance of test run when disruptions occur; 3.4.2 Schedule of Test Activities. A test schedule
(o) test acceptance criteria for test completion; shall be prepared which should include the timing
(p) confidentiality of test results. of the test events, notification of parties, test prepara-
tion and conduct, and the preparation of the test
3.4.1 Test Goals. This Code recognizes that differ-
report.
ent types of conditions might require different types
of test goals. The following illustrates three different
test goals that are acceptable by this Code.
(a) The test can be run at a specified corrected 3.5 PREPARATION FOR TEST
net power that is near the design value of interest.
Examples of this test would be an acceptance test 3.5.1 Pretest Records. Nominal performance data
of a fuel cell power system where electric efficiency and operating limitations supplied by the manufac-
is guaranteed at a specific load, or a partial load, turer of the fuel cell system shall be available on
on a specific fuel composition, but the available the test site. This shall include the expected output
pipeline composition differs from the design value. parameter values, and the limitations on the op-
(b) The test can be run at a specified net power erating range. General physical conditions of the
regardless of ambient or other external conditions. fuel cell system prior to the test shall be documented
An example of this test goal is an acceptance test with photos and data recording. Information on
on a fuel cell power system with a rated power subsystems that may be temporarily nonfunctional
and electrical efficiency guarantee over a range of shall be recorded, including the reason for the non-
ambient temperatures. functionality and the expected effect on the test
(c) The test can be run at a specified maximum results. The fuel cell power system shall be fully
power overload state point. The test shall include characterized by identification of the design and
run times to demonstrate the ability to operate for construction features that can significantly affect
the stated limited time periods, and the test results the results. The system shall be identified by the
must be reported with operational time limitations. manufacturer’s serial number of the unit and of the
An example would be a short term (e.g., less than major components. When tested in the field, the date
a few hours) overload power excursion that would of delivery and other pertinent historical information
be useful for a peak power demand. This overload shall be recorded.
power operation might legitimately be accomplished
with an incomplete thermal transient in a subsystem.
While only steady-state operation is treated in the 3.5.2 Preliminary Operation and Adjustment. Be-
Code, it should be recognized that there might be fore starting the test, the fuel cell system should be
cases when operation and subsequent performance operated for a sufficient length of time to demonstrate
under non-steady-state conditions is important. Ex- steady-state operation, and to make any necessary
amples could include start up and shut down tran- adjustments to the electrical and thermal loads for
sients, load following operation, and idle standby. assuring conformity with the Code.
In general, system performance, especially average
efficiency, will be different under transient conditions 3.5.3 Site Applicability of Instrument Calibration.
as compared to corresponding steady-state condi- Current, applicable calibration certificates of the
tions. As an aid to achieving a more meaningful critical system instrumentation shall be available
characterization of fuel cell performance, including prior to the start of the test. Parties to the test shall
non-steady-state operation, it is recommended that agree that the ambient conditions at the test site do
the total energy input into the system be compared not have an adverse effect on the calibration of the
to the net useful energy output produced during the test instruments or they shall agree to the mitigation
time frame of interest using one or more representa- measure if ambient conditions do have an adverse
tive load cycle(s). effect.
10
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
3.6 PARAMETERS TO BE MEASURED OR MhtfR p mass flow rate of heat transfer fluid into
DETERMINED DURING THE TEST PERIOD and out of the fuel cell system from
the source of secondary thermal energy
(a) Data Periods
during the data period (may be used to
t p time (hr) of the test period
compute the total flow by integrating
tr p time (hr) of the data period
over the data period), kg/sec
Pstii p average pressure of the secondary ther-
(b) Ambient Conditions
mal input heat transfer fluid entering
ELEVsite p test site elevation
the fuel cell system at the test boundary,
Pamb p barometric pressure at the test site, kPa
kPa (absolute) (may be used to calculate
RHamb p relative humidity at the test site, per- the enthalpy of the heat transfer fluid)
cent R.H. Pstio p average pressure of the secondary ther-
Tamb p ambient Temperature at the test site ,°C mal input heat transfer fluid leaving the
fuel cell system at the test boundary,
(c) Fuel kPa (absolute) (may be used to calculate
Cfp p specific heat of fuel at the test bound- the enthalpy of the heat transfer fluid)
ary (at constant pressure), kJ/kg·K Tstii p average temperature of the secondary
HHVavg p fuel high-heating value at the test thermal input heat transfer fluid entering
boundary in kJ/kg for calculations the fuel cell system at the test boundary,
based on High Heating Value K, (may be used to calculate the en-
LHVavg p fuel low-heating value at the test thalpy of the heat transfer fluid)
boundary in kJ/kg for calculations Tstio p average temperature of the secondary
based on Low Heating Value thermal input heat transfer fluid leaving
Mf p total fuel flow to the fuel cell during the fuel cell system at the test boundary,
the test or data period, kg K, (may be used to calculate the en-
MfR p fuel flow into the test boundary in thalpy of the heat transfer fluid)
kg/s during the data period (may be
used to calculate the total fuel flow
(e) Oxidant
by integrating the flowrate over the
Mo p total oxidant flow to the fuel cell system
data period)
during the test period, kg
MWf p molecular weight of the gaseous fuel,
MoR p oxidant flow into the test boundary, kg/
kg/kmol
sec, during the data period (may be used
Pf p fuel pressure at the test boundary in
to calculate the total oxidant flow by integ-
kPa (absolute)
rating the flow over the data period)
RHf p relative humidity of the fuel, at the
MWo p molecular weight of the oxidant (28.9644
test boundary, percent R.H.
kg/kmol for standard dry air)
Tff p fuel temperature at the test bound-
Po p average oxidant pressure at the test bound-
ary, K
ary, kPa (absolute)
To p average oxidant temperature at the test
(d) Secondary Thermal Input boundary, K
Hi p average enthalpy of the secondary ther-
mal input heat transfer fluid entering
the fuel cell system at the test boundary, (f) Auxiliary Electrical Input
kJ/kg Ea p kilowatt-hours into the system crossing
Ho p average enthalpy of the secondary ther- the test boundary
mal input heat transfer fluid exiting the Ia p auxiliary electrical input current, A, at
fuel cell system at the test boundary, the test boundary
kJ/kg KWa p auxiliary electrical input power into the
Mhtf p mass of heat transfer fluid into and out test boundary, kWe
of the fuel cell system from the source Va p auxiliary electrical input voltage, at the
of secondary thermal energy during the test boundary
test period, kg
11
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
(g) Electrical Output (j) Shaft Work Into the System During the Test
Eg p total kilowatt-hours out of the system test Wsi p mechanical shaft work done on the system
boundary during the test period by an outside prime mover resulting in
Ig p electrical output current, A, at the test energy added to the system from across
boundary the system boundary, kJ
KVA p total real and reactive power out of the
test boundary, kVA (k) Shaft Work Out of the System During the Test
KWg p real electrical output power out of the Wso p mechanical shaft work done by the sys-
test boundary, kWe tem to an outside industrial or other
PFg p power factor of the output current and energy user system resulting in energy
voltage from the system at the test produced by the system being exported
boundary across the system boundary, kJ. This is
Vg p electrical output voltage at the test analogous to mechanical cogeneration.
boundary
3.6.1 Class I Primary Variables
(a) Data Periods
(h) Cooling System Heat Sink Temperature tdp p time, in hours and seconds, of the data
Tcooling p cooling system heat sink tempera- period
ture, K ttp p time, in hours and seconds, of the test
period
(i) Thermal Energy Capture (TEC)
Hin p enthalpy of TEC water or heat transfer (b) Fuel
fluid entering the fuel cell system at the HHVavg p fuel high heating value at the test
test boundary, kJ/kg boundary, kJ/kg, for calculations
Hout p enthalpy of TEC water or heat transfer based on High Heating Value
fluid leaving the fuel cell system at the LHVavg p fuel low heating value at the test
test boundary, kJ/kg boundary, kJ/kg, for calculations
Mw p total mass of TEC water or heat transfer based on Low Heating Value
fluid circulating through the test bound- Mf p total fuel flow to the fuel cell during
ary during the test period, kg the test or data period, kg
MwR p flow rate of TEC water or heat transfer Pf p fuel inlet pressure at the test boundary,
fluid circulating through the system kPa (absolute)
Pteci p pressure of TEC water or heat transfer Tff p fuel temperature at the test bound-
fluid entering the fuel cell system at the ary, K
test boundary, kPa (absolute) (may be (c) Secondary Heat Input
used to calculate enthalpy) Hi p average enthalpy of the secondary thermal
Pteco p pressure of TEC water or heat transfer input heat transfer fluid entering the fuel
fluid leaving the fuel cell system at the cell at the test boundary, kJ/kg
test boundary, kPa (absolute) (may be Ho p average enthalpy of the secondary thermal
used to calculate enthalpy) input heat transfer fluid exiting the fuel
Tdelivery p useful thermal energy delivery tempera- cell at the test boundary, kJ/kg
ture, K Mhtf p mass of heat transfer fluid into and out of
Tteci p temperature of TEC water or heat transfer the fuel cell from the source of secondary
fluid entering the fuel cell system at thermal energy during the test period, kg
the test boundary, K (may be used to Tstii p average temperature of the secondary
calculate enthalpy) thermal input heat transfer fluid entering
Tteco p temperature of TEC water or heat transfer the fuel cell at the test boundary, K (may
fluid leaving the fuel cell system at the be used to calculate the enthalpy of the
test boundary, K (may be used to calcu- heat transfer fluid)
late enthalpy) Tstio p average temperature of the secondary
thermal input heat transfer fluid leaving
the fuel cell at the test boundary, K (may
12
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
13
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
3.7 OPERATION OF THE TEST 3.7.4 Duration of Test Run and Frequency of
Readings. A test run shall not be less than 1 hr. A
3.7.1 Specified Conditions. Every effort shall be sufficient number of readings shall be spaced in
made to run the test under the agreed-upon base time to show the range of fluctuations and to provide
reference conditions, such as output, pressures, and a reliable average for the test run and to meet the
temperatures, or as close to specified conditions uncertainty requirements of this Code. The interval
as possible, in order to limit the application of between readings shall be not less than 1 min.
corrections.
3.7.2 Stabilization. Before starting the test, the 3.8 CALCULATION AND REPORTING OF
fuel cell power system shall be run until steady-state RESULTS
conditions have been established and maintained for
an agreed upon time period. Steady state will be 3.8.1 Validity of Results. If, during the conduct
achieved when continuous monitoring indicates that of a test or during the subsequent analysis or interpre-
readings are within the maximum permissible varia- tation of the observed data, an inconsistency is
tions. found which affects the validity of the results, the
14
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
parties should make every reasonable effort to adjust of this Code. No specific individual instrumentation
or eliminate the inconsistency by mutual agreement. uncertainties are given in this Code; but the overall
Failure to reach such agreement will constitute rejec- requirements to have an absolute uncertainty on the
tion of the run or test. result of less than 2% shall be met by appropriate
instrumentation selected by the user.
3.8.2 Reporting of Results. In all cases, the test
results shall be reported:
(a) as calculated from the test observations, with 3.9 RECORDS
instrument calibrations only having been applied; 3.9.1 Test Observations. Test observations shall
and be recorded on a data acquisition system or entered
(b) as corrected for deviations of the operating on log sheets and authenticated by the observers’
conditions from the specified conditions. signatures. For acceptance tests, a complete set of
unaltered data sheets and recorded charts, or facsim-
3.8.3 Objectives of Uncertainty Analysis. The
iles thereof, shall be available to the buyer and
application of uncertainty analysis to a Code test
manufacturer and distributed as specified by any
procedure has two objectives:
contractual agreements. The observations shall in-
(a) determine compliance of the test procedure
clude the date and time of day. They shall be the
with the uncertainty requirements of the Code; and
actual readings without application of any correc-
(b) reduce the risk of making an erroneous deci-
tions. The log sheets and all recorded charts consti-
sion when evaluating the results.
tute a complete record.
3.8.4 Uncertainty. Test uncertainty and contract
3.9.2 Direct Observations. Where direct observa-
tolerance are not interchangeable terms. This Code
tions of instrument readings are to be recorded at
does not address contract tolerance, which is a
frequent intervals during a test, it is not always
commercial term.
necessary to observe simultaneously all readings,
3.8.5 Uncertainty Calculations. Reference should which are made at the same intervals. In cases
be made to ASME PTC 19.1 for definitions and where the average of a series of readings is used
theory behind uncertainty analyses, as well as for in calculating results, uniform time periods, adapted
the mathematical derivations of the formulas used. to conditions of the test and nature of data required,
may be employed.
3.8.5.1 This Code provides a test procedure
that produces results with the lowest practicable 3.9.3 Data Reliability. During acceptance tests,
uncertainties. Since no measurement is error-free, it is recommended that data considered to be espe-
the uncertainty of each test result should be evaluated cially important be recorded electronically or manu-
by the parties. ally taken by at least two observers. A comparison
3.8.5.2 To assist the parties in developing an of these observations should be made as soon as
uncertainty analysis, Mandatory Appendix I contains possible and any discrepancies reconciled before
an outline of the procedure, guidance on the applica- the end of the test.
tion of the analysis, and a sample calculation. 3.9.4 Test Log. All events connected with the
3.8.5.3 All uncertainty values that have been progress of a test should be recorded on the test
determined and agreed upon by the parties to a log sheets, together with the time of occurrence and
test shall be included in the report (see Section 6). name of the observer. Particular care should be
taken to record any adjustments made to any equip-
3.8.6 Uncertainty Limits. As illustrated in Manda-
ment under test, whether made during a run or
tory Appendix I, it is not possible to define a single
between runs. The reasons for each adjustment shall
value of uncertainty in order to be designated a
be stated in the test records.
Code test. The Code defines the uncertainty limits
of the combined measurements. The limits for each 3.9.5 Test Recording Errors. In case of an error
of the measured parameters will be determined by in a recorded observation, a line shall be drawn
the pretest uncertainty analysis and agreed to by through the incorrect entry; the correct reading is
the parties to the test. Trade-offs may be made to be recorded above the incorrect entry and ini-
between instrument uncertainties as long as the tialed, and an explanation entered in the proper
overall uncertainty value is within the requirements place in the test records.
15
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
SECTION 4
INSTRUMENTS AND METHODS OF MEASUREMENT
Calibration equipment shall be calibrated using and auxiliary or parasitic power consumed by the
standards that are traceable to National Institute of fuel cell power system (auxiliary or parasitic power
Standards and Technology or equivalent source. is a primary variable only if gross and auxiliary
Purchased calibration services are acceptable. Rec- power are measured separately; otherwise, net power
ords shall be maintained for each instrument and output is the measured variable). Some examples
its calibration interval. of primary Class 2 variables include fuel cell power
system output voltage, output current, and output
frequency if they are only used in minor correction
4.1 GENERAL REQUIREMENTS factors. The precise calculation of relative sensitivity
4.1.1 Introduction. This Code represents the man- coefficients is required in order to classify primary
datory requirements for instrumentation employed variables. Refer to ASME PTC 19.1 for example
and the use of such devices. The instrumentation calculations.
recommended herein may be replaced by new tech- 4.1.2.2 Secondary Variables. Variables that are
nology, as it becomes available. The Instruments measured but do not enter into the calculation of
and Apparatus supplements (ASME PTC 19 Series) the results are secondary variables. These variables
outline the governing requirements for all ASME are measured throughout the test period to ensure
performance testing. that the required test condition was not violated.
SI units, followed by U.S. customary units in Some examples of these variables are fuel cell re-
parentheses, are shown in all equations in this formate inlet temperature, barometric pressure at
Section. However, any other consistent set of units site, ambient temperature, and absolute exhaust pres-
may be used. sure if they are not used in calculations or in
correction factors.
4.1.2 Instrumentation Classification. The instru- This Code does not require high accuracy instru-
mentation employed to measure a variable will have ments for secondary variables.
different required type, accuracy, redundancy, and The instruments that measure secondary variables
handling depending upon the use of the measured may be permanently installed plant instrumentation.
variable and depending upon how the measured The Code does, however, require verification of
variable affects the final result. For purposes of instrument output prior to the test period. This verifi-
this discussion, variables are temperature, pressure, cation can be accomplished by calibration or by
differential pressure, flow, velocity, voltage, current, comparison against two or more independent mea-
frequency, stream constituency, and humidity. Mea- surements of the variable referenced to the same
surements are classified as either primary or second- location.
ary variables.
4.1.2.1 Primary Variables. Variables that are 4.1.3 Instrument Calibration
used in calculation of test results are considered 4.1.3.1 Calibration. Calibration of an instru-
primary variables. Primary variables are further clas- ment is the act of applying process conditions to
sified as Class 1 or Class 2. Class 1 primary variables the candidate instrument and the reference standard
are those which have a relative sensitivity coefficient and adjusting the instrument to match the standard,
of 0.2 or greater. These variables will require higher if appropriate. During calibration, readings are taken
accuracy instruments with more redundancy than from the candidate instrument and the referenced
Class 2 primary variables, which have a relative standard. The output of the candidate instrument
sensitivity coefficient of less than 0.2. Some examples may be adjusted to the standard reading. As an
of primary Class 1 variables include input fuel flow alternative, the difference between the instrument
measurement, fuel heating value, gross power output, and the referenced standard may be recorded and
16
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
applied to the instrument reading. This alternative calibrated such that as the anticipated test measuring
method is recommended in the case of the thermo- value or range is approached, an increasing number
couples or Resistance Temperature Devices (RTDs) of calibration points are measured below, above
because their output cannot be easily altered. and through the anticipated test value or range to
4.1.3.2 Referenced Standards. In general, all minimize the hysteresis effects.
test instruments used to measure primary variables Some instruments are built with a mechanism to
(Class 1 and Class 2) shall be calibrated against alter the range once the instrument is installed. In
referenced standards traceable to the National Insti- this case, the instrument must be calibrated at each
tute of Standards and Technology (NIST) in the range to be used during the test period.
United States, other recognized international stan- (b) Class 2 Primary Variables. Instruments measur-
dards organization, or recognized physical constants. ing Class 2 primary variables shall be calibrated
All referenced standards shall be calibrated at an with at least one point more than the order of the
interval specified by the instrument manufacturer. calibration curve fit. If the instrument can be shown
The referenced standard shall have an uncertainty to typically have a sufficiently low hysteresis of less
of less than 1⁄4 of the required uncertainty of the than the required accuracy, the expected measuring
test instrument to be calibrated. point or range need only be approached from one
Instrumentation used to measure secondary vari- direction (either increasing or decreasing to the
ables need not be calibrated against a referenced point).
standard. These instruments may be calibrated (c) Secondary Variables. The Code does require
against a calibrated instrument that is traceable to verification of instrument output prior to the test
the NIST standards. period. Instruments used to measure secondary vari-
ables can be checked in place with at least two or
4.1.3.3 Ambient Conditions. Calibration of in-
more instruments measuring the variable with respect
struments used to measure primary variables (Class
to the same location or can be calibrated against
1 or Class 2) shall be performed in a manner which
a previously calibrated instrument. Should the instru-
replicates the condition that the instrument will be
ment be calibrated, it need only be calibrated at
used to make the test measurements. Consideration
one point in the expected operating range.
shall be given to all process and ambient conditions
that may affect the measurement, including tempera- 4.1.3.5 Timing of Calibration. All test instru-
ture, pressure, humidity, static electricity, electro- mentation used to measure primary variables (Class
magnetic interference, etc. 1 or Class 2) will be calibrated prior to the test and
have calibration checked following the tests. The
4.1.3.4 Instrument Ranges and Calibration
initial instrument calibration prior to tests and cali-
Points. The number of calibration points depends
bration check after the tests should be kept to a
on the classification of the variable the instrument
minimum length of time to obtain an acceptable
will measure. All Class 1 or 2 instruments shall be
calibration drift. Instruments that are within their
serialized and carry a label indicating each instru-
approved calibration intervals shall be acceptable
ment’s calibration status and date of calibration.
for use. The following exceptions to the prior and
Instruments used continuously shall be calibrated
post-calibration requirement are allowed. Flow mea-
per a defined interval and procedure as required
surement devices, current transformers, and potential
per the instrument manufacturer’s instructions. Instru-
transformers by nature are not conducive to post
ments shall be calibrated by qualified personnel,
test calibration checks. In the case of flow measuring
authorized suppliers, or third party calibration ser-
devices used to measure Class 1 primary variables,
vices, using written procedures. Records or instru-
the instrument shall be inspected following the test
ment calibration data sheets shall be maintained for
each instrument used during the test along with for physical integrity rather than recalibrating the
each instrument’s calibration interval. The calibration device. Flow elements for Class 2 primary variables
should closely bracket the expected measurement need not be inspected if steam blow or chemical
range anticipated for the test. cleaning has not occurred.
(a) Class 1 Primary Variables. The instruments Post-test calibration of current and potential trans-
measuring Class 1 primary variables shall be cali- formers is not required.
brated with at least two points more than the degree 4.1.3.6 Calibration Drift and Shift. Calibration
of the calibration curve fit. Each instrument shall be drift is defined as the difference in the calibration
17
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
correction as a percent of reading. When the post- program. This program is a method of documentation
test calibration check indicates that the drift is less where the following information can be found:
than the instrument bias uncertainty, the drift is (a) calibration procedures
considered acceptable and the pretest calibration is (b) calibration technician training
used as the bias for determining the test results. (c) standard calibration records
Should the calibration drift, combined with the refer- (d) standard calibration schedule
ence standard accuracy as the square root of the (e) instrument calibration histories with most re-
sum of the squares, exceed the accuracy of the cent individual instrument calibration sheets
instrument, it is not acceptable. The quality assurance program shall be designed
Calibration shift can result from instrument mal- to ensure that the laboratory standards are calibrated
function, transportation, installation, or removal of as required. The program also ensures that the
the test instrumentation. Should this occur, engi- properly trained technicians calibrate the equipment
neering judgment must be used to determine whether in the correct manner.
the initial test calibration or the final calibration All parties to the test should be allowed the
check is correct. Below are some good practices: opportunity to witness the calibration of instruments
(a) When instrumentation is transported to the in the calibration laboratory and at the site. In
test site between the calibration and the test period, addition, all test parties should have the right to
a single point check prior to and following the test review the quality assurance plan prior to or during
can isolate when the shift may have occurred. An the instrument calibration.
example of this check is a vented pressure transmit-
ter, no load on wattmeter, and an ice point tempera- 4.1.4 Plant Instrumentation. It is acceptable to
ture check on a thermocouple or RTD with or use plant instrumentation for primary variables only
without transmitter. if the plant instrumentation (including signal-condi-
(b) Redundant instrumentation is two or more tioning equipment) can be demonstrated to meet
devices measuring the same parameter with respect the overall uncertainty requirements. Many times
to the same location. In locations where redundant this is not the case. In the case of flow measurements,
instrumentation is employed, calibration drift should all measurements (process pressure, temperature,
be analyzed to determine which calibration data differential pressure, or pulses from a metering de-
the initial test calibration or the post-test check vice) must be made available for correcting the
produces the better agreement between the redun- primary instrument data, as plant conversions to
dant instruments. flow are often not rigorous enough for the required
accuracy.
4.1.3.7 Loop Calibration. All instruments used
to measure primary variables (Class 1 or Class 2)
should be loop calibrated. Loop calibration involves 4.2 CHECKLIST OF INSTRUMENTS AND
the calibration of the instrument through all of the APPARATUS
signal-conditioning equipment utilized up to and The following devices and instruments are typi-
including final indication. This may be accomplished cally used to measure the performance of a fuel
by calibrating instrumentation employing the test cell power system:
signal conditioning equipment either in a laboratory (a) For fuel input measurement
or on site during the test setup before the instrument (1) fuel composition instruments, e.g., gas chro-
is connected to the process. Alternatively, the signal- matograph
conditioning device may be calibrated separately (2) fuel input rate measurement, e.g., flow me-
from the instrument by applying a known signal to ters, pressure sensors, temperature sensors
each channel using a precision signal generator. (b) For ambient condition measurement
Where loop calibration is not practical, an uncer- (1) barometer, hygrometer, thermometer, etc.
tainty analysis must be performed to ensure the (c) For electrical output
combined uncertainty of the measurement and final (1) load; e.g., grid or load bank
indication system meets the accuracy requirements (2) output measurements, voltage and current
described herein. transducers, real and reactive power measurement
4.1.3.8 Quality Assurance Program. Each cali- (d) For thermal output
bration laboratory must have a quality assurance (1) flow and temperature measurement
18
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
19
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
report power measurements, in the presence of har- fuel heating value at constant fuel flow with some
monics, is referred to as a “true RMS” reading associated lag time depending on the internal storage
device. A standard RMS reading device might not of fuel and fuel utilization within the fuel cell
account for harmonic distortion and thus may intro- power system. Fuel cell system performance can be
duce measurement errors. estimated (although not in compliance with this
Electrical test measurement procedures are pro- code) if fuel heating value and fuel flow are recorded
vided in IEEE Std. 120, IEEE Master Test Guide for on frequent intervals, and the lag time associated
Electrical Measurements in Power Circuits, which with processing stored high heating value or low
should be consulted before applying this Code. heating value fuel can be ascertained from observa-
4.3.2.1 Net Electrical Power Output. Net elec- tion and agreement of the test parties.
trical power output of a fuel cell power system is 4.4.3 Consistency of Fuel Flow. Fuel flows to the
equal to the electrical output directly produced by fuel cell power system must be stable throughout
the fuel cell power system at the rated voltage minus the test. For gaseous fuels, fluctuations in flow may
auxiliary loads supplied by an external power source. be induced by reciprocating compressors or by other
4.3.2.2 Instrumentation. Calibrated true RMS sources of pulsation. The difference between the
reading watt and VAR meters shall be used. The indicated maximum and minimum flow shall be less
calibration and accuracy should include associated than 2% of the average flow for gases for an accept-
current and voltage transformers. able test. Introduction of a cushion chamber, surge
chamber, pulsation dampener, or other means of
4.3.2.3 Current and Potential Transformers.
absorbing the pulsations between the source of pulsa-
Potential transformers are typically required when
tion and the metering device before measurement
the measured three-phase voltage rating is greater
are considered acceptable for gaseous fuel usage.
than 600 V rms. Current and potential transformers
Leakage of liquid fuels after the point of measurement
must be operated within stated ratings to prevent
shall be measured and included in the calculation
saturation or similar sources of measurement error.
of liquid fuel mass flow.
4.4.4 Determination of Fuel Heating Value. The
4.4 DETERMINATION OF FUEL INPUT heating value of the fuel can be measured by an
4.4.1 Fuel Types Considered. Either gaseous and on-line chromatograph or by sampling periodically
liquid fuels types may be used when testing fuel (at a minimum of three samples per test) and analyz-
cell power systems. Fuel heating values must be ing each sample individually for heating value. Sam-
consistent throughout the test period. pling of fuels shall be done in accordance with
para. 4.6.9. The analysis of gaseous fuels, either by
4.4.2 Consistent Gaseous or Liquid Fuels. Consist- on-line chromatography or from laboratory samples
ent liquid or gaseous fuels are those with heating shall be done in accordance with ASTM D 1826-94
values, which vary less than 1% over the course (1998), Standard Test Method for Calorific (Heating)
of the performance test. Since liquid and gas flows Value of Gases in Natural Gas Range by Continuous
and heating values can be determined with high Recording Calorimeter; ASTM D 1945-96, Standard
accuracy, the heat input from these type fuels is Test Method for Analysis of Natural Gas by Gas
usually determined by the direct measurement of Chromatography; or ASTM 588-98, Standard Practice
fuel flow and laboratory or on-line chromatography- for Calculating Heat Value, Compressibility Factor,
determined heating value. Consistent liquid or gas- and Relative Density of Gaseous Fuels. ASTM D
eous fuels heat input can also be determined by 1945 gives the results in the type and amount of
calculation using the fuel constituents percentage gaseous constituents from which the heating value
multiplied by the constituent heating value. is calculated.
Homogenous gas and liquid fuel flows are usually ASTM 3588-98, Standard Practice for Calculating
measured directly for fuel cell power systems. Heat Value, Compressibility Factor, and Relative
Some types of fuel (landfill gas or wastewater Density of Gaseous Fuels. ASTM D 1945 gives the
treatment digester gas) may vary enough to fall results in the type and amount of gaseous constituents
outside the 1% heating value range (specified above) from which the heating value is calculated.
over the test period. Variations in fuel cell power Liquid fuel heating value shall be determined by
system power output can be expected to follow the calorimeter in accordance with ASTM D 4809-95,
20
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
Standard Test Method for Heat of Combustion of to correctly calculate the flow. Other flowmeter
Liquid Hydrocarbon Fuels by Bomb Calorimeter types are acceptable, provided a measurement error
(Precision Method). (ASTM D 240 shall not be used consistent with the required accuracy for the test is
because it results in lower precision.) achievable. Where an oil return system from the
For liquid fuels both the lower and high heating fuel cell power system is used, both supply and
value and the specific gravity shall be determined return flows shall be measured by calibrated meters.
from the fuel analyses. The specific gravity of the Positive displacement type oil flowmeters may be
fuel should be determined at three temperatures used without temperature compensation. The effects
covering the range of temperatures measured. of temperature on fluid density can be determined
When analyzing the fuel heating value, the bias by calculating the mass flow based on specific
or systematic error associated with the measurement gravity at the flowing temperature based on the
of such shall be considered. Table 4.1 contains the formula below:
bias limits for various fuel heating values. If a fuel
is used that is not included on the table, a bias Customary Units
limit for heating value that is consistent with the
accuracy of this code must be agreed upon by the Qmh p(8.337)(60)QvSg
parties to the test.
4.4.5 Determination of Liquid Fuel Specific Grav- where
ity. The liquid fuel specific gravity shall be deter- Qmh p mass flow, lbm/hr
mined from each fuel sample taken. Each sample Qv p volumetric flow, gal/min
shall have the specific gravity evaluated at three Sg p specific gravity of flowing temperature,
temperatures covering the range of temperatures dimensionless
measured during testing. The specific gravity at flow- 8.337 p density of water at 60°F, lbm/gal
ing temperatures shall then be determined by interpo- 60 p minutes per hour
lating between the measured values to the correct
temperature. Specific gravity determinations shall be SI Units
completed in accordance with ASTM D 1217.
Numerous methods are employed in industry to Qmh (from above)
Qmh (kg/hr) p
determine the flow of solid, liquid, or gaseous 2.20462
streams. ASME PTC 19.5 is the primary reference
for flow measurements. ASME MFC-3M, PTC 6, where
and ISO 5167 provide further information on flow 2.20462 p lbm per kg
measurement techniques. These sources include de-
sign, construction, location, and installation of
4.4.7 Measurement of Gaseous Fuel Flow. Gas-
flowmeters, the connecting piping, and computations
eous fuel flows may be measured using orifices
of flow.
or turbine-type flowmeters. Measurements used to
4.4.6 Measurement of Liquid Fuel Flow. Liquid determine the mass flow, such as fuel analysis to
fuel flows shall be measured using flowmeters that determine density, the static and differential pres-
are calibrated per para. 4.1.3 throughout their Reyn- sures, temperature, and frequency (if a turbine meter),
olds number range expected during the test using must be within an uncertainty range to meet the
the actual flow. For volume flowmeters the tempera- uncertainty requirement of this Code. Other flow-
ture of the fuel also must be accurately measured meters are permitted if it can be demonstrated that
21
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
the total uncertainty of the mass flow is consistent of the instruments within 2 sec to 3 sec to obtain
with the required accuracy for the test. all necessary data with the plant at the same condi-
The use of turbine meters is one alternative to tion. The system should be able to collect and store
orifice gas flow measurement. The turbine meter data and results within 1 min. The system should
measures actual volume flow. Uncertainty of turbine also have the ability to plot the test data and each
meters is by statement of the manufacturer and instrument reading over time to look for trends and
calibrated in atmospheric air or water, with formula- outlying data.
tions for calculating the increased uncertainty when The data collection system shall be calibrated for
used in gas flow at higher temperatures and pres- primary variables per para. 4.1.3.7.
sures. Sometimes a turbine meter is calibrated in 4.5.2 Data Management
pressurized air. The turbine meter calibration report
4.5.2.1 Storage of Data. Signal inputs from the
must be examined to confirm the uncertainty as
instruments should be stored to permit post-test
calibrated in the calibration medium.
data correction for application of new calibration
Mass flow as shown by computer printout or
corrections. The engineering units of each instrument
flow computer is not acceptable without showing
along with the calculated results should be stored
intermediate results and the raw data used for calcu-
in removable medium to secure against equipment
lations.
damage during transport.
EXCEPTION: If the computer printout or flow computer has been
loop calibrated in series with the flow device, intermediate results
4.5.2.2 Manually Collected Data. Most test pro-
are not required. grams will require some data to be taken manually.
The data sheets should each identify the data point,
When required, intermediate results for an orifice
test site location, date, time, data collector, and data
would include discharge coefficient, corrected diam-
collected.
eter for thermal expansion, expansion factor, etc.
Raw data includes static and differential pressures, 4.5.3 Construction of Data Collection Systems
and temperatures. For a turbine meter, intermediate 4.5.3.1 Design of Data Collection System
results include the turbine meter constant(s) used in Hardware. With advances in computer technology,
the calculation, and how it is determined from the data collection system configurations have a great
calibration curve of the meter. Raw data also includes deal of flexibility. They can consist of a centralized
frequency, temperature, and pressure. For turbine processing unit or distributed processing to multiple
meters and orifices, fuel analysis and the intermediate locations in the plant.
results used in the calculation of density is required. Each measurement loop must be designed with
Ensure that the actual gas composition is properly the ability to be loop calibrated separately, if required
considered relative to the expected gas composition (see para. 4.1.3.7). Each measurement loop should
utilized in the flowmeter calibration. be designed so that they can be individually checked
for continuity and power supply if applicable to
4.4.8 Calculation of Fuel Input. The calculation
locate problems during equipment setup.
of fuel input is the product of the fuel mass flow
Each instrument signal cable should be designed
at test conditions multiplied by the fuel heating
to reject any stray induced currents.
value at test conditions.
4.5.3.2 Usage of Existing Plant Data Collection
4.4.9 Sampling of Fuels. Automatic sampling of and Control System. The code does not prohibit
gaseous fuels shall be done in accordance with the use of the plant measurement and control system
ASTM D 5287. Manual sampling of liquid fuels for code testing. However, this system must meet
shall be done in accordance with ASTM D 4057. the requirements of this Section.
Automatic sampling of liquid fuels shall be done in Some plant systems do not allow the instrument
accordance with ASTM D 4177-95 (2000), Standard signal prior to conditioning to be displayed or stored.
Practice for Automatic Sampling of Petroleum and The signal must be available to check the signal
Petroleum Products. conditioning calculation for error.
Distributed control systems typically only report
4.5 DATA COLLECTION AND HANDLING
changes in a variable which exceed a set threshold
4.5.1 Data Collection Systems. A data collection value. The threshold value must be low enough so
system should be designed to accept multiple instru- that all data signals sent to the distributed control
ment inputs and be able to sample data from all system during the test are reported and stored.
22
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
SECTION 5
COMPUTATION OF RESULTS
23
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
consider only the heating value of the fuel when consumption taken as a net loss against output (see
computing heat rate or thermal efficiency. However, test boundary diagram). If a fuel compressor is used
it is important for all interested parties to realize with the fuel cell system, it shall be included within
that, as stated above, fuel variables other than heating the test boundary, and any energy used to run it
value can affect computed values of heat rate and shall be either supplied from within the test boundary
thermal efficiency. or shall be deducted from the net electrical en-
For some fuel cell systems, the composition of ergy out.
the fuel may affect performance sufficiently to require If a pressure-reducing regulator is used to reduce
accounting for this fact, and some correction to pressure without energy recovery, the test boundary
fuel input, bringing it back to design composition can be defined such that the pressure-reducing regu-
specifications may be necessary. For these systems lator is outside of the test boundary.
the manufacturer is responsible for furnishing correc-
tion factors and curves to be applied using manufac- 5.2.4 State of Input Oxidant. Oxidant must be
turer supplied correction methods. These correction supplied to the fuel cell systems at rates which are
factors may be additive, multiplicative, or both. likely to greatly exceed the stoichiometric air/fuel
ratio which is about 15:1 for most hydrocarbon
5.2.3 State of Input Fuel. For a pressurized stack fuels. Thus, the power to compress the air will be
(utilizing gaseous fuel) the power to compress the much greater than that to compress the fuel and
gas is not negligible and thus must be taken into thus can be a significant fraction of the stack output
account. In addition, pressurized fuel may be used power in the case of pressurized systems. Conse-
to drive a pressure-reducing turbine to recover some quently, for computational purposes, it must be
mechanical energy for use in the system or to power assumed that the oxidant (air) is supplied at ambient
other auxiliaries. Specifically, if part of the gas
temperature and pressure, and some method of
pressurization is provided by the external supply
accounting for the power associated with com-
system, the power required for pressurization shall
pressed air from an air supply system is needed.
be charged to the fuel cell system. Thus, for computa-
The composition of the supply air will also have an
tional purposes, it should be assumed that all fuel
effect on fuel cell system performance. In particular,
is provided to the system boundary at local ambient
water vapor in the air will add to the power required
barometric pressure. Fuel delivery to the test bound-
to compress the air and subtract from the oxygen
ary at above atmospheric pressure shall result in an
available for the oxidation of fuel. Therefore, as it
additive correction factor of the form:
is already customary with heat engines, corrections
may be made to adjust performance to standard
absolute fuel inlet
pressure at test boundary, kPa dry air.
Ecfp(Mf RTstd ) ln Oxidant (air or other) delivery to the test boundary
ambient barometric pressure, kPa
at above local ambient barometric pressure shall
where require the use of an additive correction factor of
Ecf p ideal Pressure Energy Content of the the form:
Fuel, kJ
Mf p mass of fuel into the system during the absolute oxidant (air) inlet
test period, kg pressure at test boundary, kPa
Ecop(Mo RTstd ) ln
ambient barometric pressure, kPa
MWf p molecular weight of the gaseous fuel
(16.043 kg/kmol for methane)
R p Ru /MWf where
Ru p 8.314 kJ/kmol K (universal gas constant) Eco p ideal pressure energy content of the
Tstd p standard temperature (15°C) p 288.15 K oxidant, kJ
Mo p mass of oxidant into the system during
This formula is based on an ideal isothermal the test period, kg
compression of the fuel at standard temperature MWo p molecular weight of the oxidant
and does not take into account the compression (28.9644 kg/kmol for standard dry air)
efficiency. If a fuel compressor is used, it must be R p Ru /MWo
treated as within the test boundary with its energy Ru p 8.314 kJ/kmol K (universal gas constant)
24
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
25
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
Qst p total thermal energy input to the fuel cell Mo p mass of oxidant into the system during
during the test period the test period, kg
Qst shall be calculated by measuring the heat MWo p molecular weight of the oxidant (28.9644
content of the heat transfer fluid, and multiplying kg/kmol for standard dry air)
by the total flow of the heat transfer fluid into the R p Ru /MWo
test boundary and correcting to a base value enthalpy Ru p 8.314 kJ/kmol K (universal gas constant)
of saturated water at 15°Cp62.99 kJ/kg. Tstd p standard temperature (15°C) p 288.15°K
If an oxidant (air or other) compressor or blower
Qst p Mhtf ⴛ (Hi − 62.99 kJ/kg) is used with the fuel cell system, it shall be included
within the test boundary and the power to run the
where compressor shall be supplied from either inside the
Hi p average enthalpy of the steam or heat test boundary or shall be deducted from the net
transfer fluid entering the fuel cell during electrical energy out.
the test period, kJ/kg 5.2.5.6 Oxidant Thermal Energy Input, TQo
Mhtf p mass of steam or heat transfer fluid into
the fuel cell from the source of secondary TQop Mo (kg) ⴛ Cop [(kJ/(kg·K)] ⴛ Tof − 288.15(°K)
thermal energy during the test period, kg
(c) Or, for gaseous thermal inputs that do not where
exit the system boundary, the total heat input is Cop p average specific heat at constant pressure
determined by: of oxidant delivered during the test pe-
riod, kJ/kg·K
Qst p the total thermal energy input to the fuel Mo p the total oxidant flow to the fuel cell
cell during the test period during the test period, kg
Tof p average oxidant temperature at the test
Qst shall be calculated by measuring the heat boundary, K
content of the input gas stream, and multiplying by TQo p the total heat energy content required to
the total flow of the heat input gas stream into the heat the oxidant from 15°C (288.15 K)
test boundary and correcting to a base value enthalpy to the average oxidant temperature at
of dry air at atmospheric pressure and 15°Cp the test boundary during the test duration
33.0291 kJ/kg.
Standard Temperaturep15°C (288.15 K)
Qst p Mhtf ⴛ (Hi − 33.0291 kJ/kg) 5.2.5.7 Auxiliary Electrical Input, Ea
26
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
constraints, the input fuel, steam, or heat transfer with power factor setting. In most inverters now on
fluid or fluid drive state points shall be determined the market, the efficiency of the inverter will be at
to within 1% accuracy and the energy input by shaft its maximum at unity power factor.
work shall be computed. No penalty or correction
for turbine or prime mover efficiency is permitted. 5.3.3 Electrical Total Energy. The total electrical
Wherever possible the prime mover shall be moved energy (kilowatt-hours) measured over the test dura-
inside the test boundary and energy inputs to the tion may be obtained by integrating the power
prime mover shall be considered as fuel, auxiliary measurements over the test run period:
thermal input, or auxiliary electrical loads.
Eg p ⌺ Kw·he (kWhr)
5.2.5.9 Total Energy Into the System for
where
Calculation of Input Energy. This is expressed by:
Eg p total sum of all electrical output from
QItotal p Qf + Ecf + TQf + Qst + Eco + TQo + Wsi the fuel cell across the test boundary
for the duration of the test period
where Kw·he p kilowatt-hours out of the system during
QItotal p total energy into the system for calcula- the test period (measured)
tion of input energy and electrical con-
5.3.4 Net Electrical Energy. The total electrical
version efficiency
energy out of the system, integrated over the test
duration, must be reduced to account for any auxil-
5.3 COMPUTATION OF ELECTRIC POWER iary electrical loads supplied from external electrical
OUTPUT power sources. For computation of electrical conver-
5.3.1 Averaging of Test Data. The current, voltage sion efficiency, the net electrical energy produced
and power measurements shall be averaged over is given by:
the test run period. The calculation method for
average power or total energy should be performed En p Eg − Ea
in accordance with ANSI/IEEE Standard 120 for the
specific type of measuring system used. where
Ea p total auxiliary electrical energy input over
5.3.2 Real Power. Digital power measurement the test duration from para. 5.2.5.7
instrumentation may provide direct readings of real Eg p total (gross) electrical energy measured over
power in addition to voltage and current. When the test duration from para. 5.3.3
direct readings of power are not available, the follow- En p net electrical energy for the test duration,
ing equation may be used to determine real power kW·h
when the phase angle between voltage and current
is known.
27
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
5.4.1.3 Computation of Thermal Energy Cap- Carnot Cycle Units. Electrical Conversion Efficiency
tured (TEC) for useful service (space/water heating, is calculated by dividing the net electrical energy
etc.) is calculated by: produced by the total energy input, per unit of time,
as follows:
TEC p Mw (Hout − Hin)
el p 兵 (En ⴛ 3600 ] / QItotal 其 ⴛ 100
where
Hin p average enthalpy of water or heat transfer where
fluid entering the test boundary during 100 p to convert to percent
the test period, kJ/kg 3600 p kJ/kW·h conversion factor
Hout p average enthalpy of water or heat transfer En p net electrical energy for the test duration
fluid leaving the test boundary during the from para. 5.3.4
test period, kJ/kg el p Electrical Conversion Efficiency, %
Mw p total mass of water or heat transfer fluid
circulating through the test boundary dur- QItotal p total energy into the system for calcula-
ing the test period, kg tion of input energy and electrical con-
TEC p thermal energy captured for the duration version efficiency from para. 5.2.5.9
of the test period, kJ
The heat rate (H.R.) in U.S. Customary units can
5.4.2 Computation of Shaft Work Out Of System
be computed as:
Wso p mechanical shaft work done by the system
on an outside industrial process or prime
H.R. (in Btu/kW·h) p (el / 100) 3412.14
mover resulting in energy recovered from
the system and delivered across the sys-
The result should be clearly identified as to the
tem boundary, kJ
type of fuel used and either low or high heating
Wso shall be computed using torque and RPM
value (LHV or HHV). See para. 3.6.
measurements consistent with results in kilojoules.
Where torque and RPM cannot be obtained due to
Constants for other frequently used units are given
system constraints, the output process parameters
below.
shall be determined to within 1% accuracy and the
energy output by shaft work shall be computed. Power Energy
No penalty or correction for process efficiency is Outage Input Constant
permitted. hp Btu/hr 2544.43
kW kJ/sec 1
5.5 COMPUTATION OF AVERAGE NET POWER
Net Power, KWe, can be computed from watt- 5.6.2 Computation of Thermal Effectiveness.
hour meters on the output terminals divided by the Thermal effectiveness is a measure of energy utility;
duration of the test. Average net power over the that is, the percentage of total energy input, which
test period will document the unit capacity. is captured and put to useful service by conversion
to electrical energy or as thermal energy:
KWe p En /ttp
where
冦
eff p [(En ⴛ 3600) + TEC ] / QItotal 冧 ⴛ 100
En p net electrical energy delivered across the
where
test boundary for the test duration from
100 p convert to percent
para. 5.3.4., kW·h
3600 p kJ/kW·h conversion factor
ttp p duration of the test, hr
En p net electrical energy for the test duration
5.6 COMPUTATION OF EFFICIENCIES from para. 5.3.4
eff p thermal effectiveness, %
5.6.1 Computation of Electrical Efficiency. The QItotal p total energy Into the system for calcula-
efficiency of thermal energy conversion to electrical tion of input energy and electrical con-
energy is computed in the same manner as for version efficiency from para. 5.2.5.9
28
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
TEC p thermal energy captured for useful ser- heat (assuming 100% heat exchanger effectiveness).
vice (space/water heating, etc.), kJ from FCP is expressed in BTU/(kW·h).
para. 5.4.1
For systems that export shaft work to a process FCP p [(QItotal − TEC ) ⴛ 1.05435] / En
outside the test boundary, the thermal effectiveness
may be modified to include credit for the shaft work where
done on the external system. The modified equation 1.05435 p conversion factor for Btu/kJ
will take the form of: En p net electrical energy for the test dura-
tion from para. 5.3.4, kW·h
冦
eff p [(En ⴛ 3600) + (TEC + Wso )] / QItotal 冧 ⴛ 100 QItotal p total energy into the system for calcu-
lation of input energy and electrical
conversion efficiency from para.
where 5.2.5.9, kJ
En p net electrical energy for the test duration TEC p thermal energy captured, kJ, for the
from para. 5.3.4, kW·h duration of the test period from
eff p thermal effectiveness, % para. 5.4.1
QItotal p total energy into the system for calcula-
tion of input energy and electrical con-
5.7 CORRECTION OF TEST RESULTS TO
version efficiency from para. 5.2.5.9
REFERENCE CONDITIONS
TEC p thermal energy captured for useful ser-
vice (space/water heating, etc.), kJ, from The baseline state of fuel, air, and water inputs
para. 5.4.1 will be standard temperature and pressure and 60%
Wso p mechanical shaft work done by the sys- relative humidity for the air and zero humidity for
tem to an outside industrial or other the fuel. This may be agreed to or otherwise modified
energy user system resulting in energy by the parties, but absent an agreement to the
produced by the system being exported contrary, performance shall be based on units op-
across the system boundary, kJ (see erating at sea level, at 15°C, with a relative humidity
para. 5.4.2) of 60% and dry fuel. All inputs that vary from
3600 p kJ/kW·h conversion factor these baseline parameters shall be corrected using
100 p to convert to percent manufacturer supplied correction curves prior to
computing the results of the test. All corrections to
5.6.3 Computation of Heat Rate. Heat Rate is test data shall be made and documented before the
electrical conversion efficiency, expressed as Btu/ calculations in para. 5.5 are made. All values in
(kW·h). This can be calculated as follows: para. 5.5 refer to corrected values only.
The heat rate (H.R.) in U.S. customary units can The procedure for correction of test results to
be computed as: specified conditions depends on the type of fuel
cell system and its inverter and heat load. It is
3421.14 necessary to have the test conditions within limits
H.R. [Btu/(kW·h)] p
el /100 agreed to by the parties to the test, to avoid operation
at extreme conditions far from its design or specified
where condition, which could make the determination of
100 p convert from percent accurate results impossible.
3412.14 p BTU/(kW·h) conversion factor The off-design characteristics of each fuel cell
el p Electrical Conversion Efficiency, % system type are unique. Hence, the manufacturer’s
from para. 5.6.1 published performance curves for the particular fuel
cell system must be used to correct that actual test
5.6.4 Computation of Fuel Chargeable to Power data to rated or standard conditions. Unless other-
Heat Rate. Fuel chargeable to power can be used wise agreed by the parties to the test, these correction
to compute the cost of electrical generation, giving curves (or data) are applied without any uncertainty.
credit for cogeneration heat as a credit towards the A step-by-step method of correcting test data will
fuel used. Credit is given equivalent to the fuel that be prepared by the manufacturer. The user will
would be used to produce the thermal cogeneration review this procedure and any discrepancies must
29
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
be resolved prior to the start of the test. 5.7.4 Ambient Pressure. If the fuel cell system is
to be tested above sea level, the decreased ambient
5.7.1 Humidity. High ambient humidity may af- pressure may affect output or efficiency due to
fect performance by limiting oxygen content and increased blower or compressor power requirements,
increasing blower or compressor power require- or decreased oxygen content. If the test is to be
ments. If ambient humidity is expected to be different conducted above design elevations, the manufacturer
from design parameters, the manufacturer shall pro- shall provide correction factors and methods prior
vide correction factors and methods prior to the test. to the test.
5.7.2 Fuel Composition. Fuel composition may 5.7.5 Useful Heat Load. If the heat supplied to
affect output or efficiency. If fuel composition is the useful heat load is provided at a temperature
different from the design of the heat exchange loop,
expected to be different from design parameters, the
insufficient heat exchanger duty may limit heat re-
manufacturer shall provide correction factors and
covery. If heat is to be supplied at temperatures
methods prior to the test.
outside design parameters, the manufacturer shall
5.7.3 Ambient Temperature. Ambient tempera- provide correction factors and methods prior to
ture may affect output or performance by limiting the test.
heat exchanger duty, overloading components, or 5.7.6 Power Factor. If the power conversion and
increasing power conversion losses. If ambient tem- power conditioning equipment is sensitive to power
perature is expected to be different from design factor, power factor shall be adjusted to be within
parameters, the manufacturer shall provide correc- design parameters during the test. No correction
tion factors and methods prior to the test. factor for power factor is allowed.
30
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FUEL CELL POWER SYSTEMS PERFORMANCE ASME PTC 50-2002
SECTION 6
TEST REPORT REQUIREMENTS
6.1 GENERAL REQUIREMENTS to allow any major deviations from the test require-
ments;
The test report shall clearly and concisely docu-
(c) date and time of the test;
ment all data generated by the test as well as all
ensuing computations. Definitive statements of the (d) summary of the test results including uncer-
purpose of the test and attainment of the objectives tainty;
should be provided. (e) conclusions.
The test report for a performance test should
incorporate the following: 6.3 INTRODUCTION
(a) title page
(b) executive summary The introduction shall include the following:
(c) introduction (a) test objective and all agreements among the
(d) instrumentation parties to the test;
(e) results (b) a listing of the representatives of the parties
(f) conclusions to the test;
(g) appendices (c) a process flow diagram showing the test
This outline is a recommended report format. boundary;
Other formats are acceptable; however, the test (d) a brief historical background.
report should contain all the information described
in the following sections as a minimum. 6.4 INSTRUMENTATION
6.1.1 Title Page. The title page shall include the The following should be included in the instrumen-
following: tation section:
(a) report number (a) tabulation of instrumentation used for primary
(b) date and time of test and secondary measurements, including make,
(c) location of test model number, etc.;
(d) title of test (b) description of the instrument location;
(e) equipment owner (c) means of data collection for each data point,
(f) equipment identification such as temporary data acquisition system print-out,
(g) parties conducting test plant control computer print-out, or manual data
(h) parties responsible for test report sheet, and any identifying tag number and/or address
(i) date of report of each;
(j) report acceptance (d) identification of any instrument which was
used as a back-up;
(e) description of data acquisition system used;
6.2 EXECUTIVE SUMMARY (f) summary of pretest and post-test calibration
The executive summary shall concisely address of each instrument used in the test.
the following:
(a) general information about the fuel cell power
6.5 RESULTS
system and the performance test, such as the plant
type and operating configuration, and the test ob- The following should be included in this Section:
jective; (a) the performance equations used and applica-
(b) any agreements among the parties to the test ble correction factors;
31
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 FUEL CELL POWER SYSTEMS PERFORMANCE
(b) tabulation of reduced data necessary to calcu- (a) the results of the test versus the test objectives/
late the performance results; performance guarantees;
(c) summary of operating conditions correspond- (b) the recommended changes to future test proce-
ing to the reduced data; dures due to lessons learned;
(d) step-by-step calculation of the performance
results from the reduced data including uncertainty
6.7 APPENDICES
analysis;
(e) detailed calculation of flow; Appendices to the test report should include:
(f) detailed calculation of heat input; (a) the signed test requirements contract;
(g) detailed calculations of fuel properties — heat- (b) copies of original data sheets and/or data
ing value, density, etc.; acquisitions system printouts;
(h) any calculations showing the elimination of (c) copies of operator logs or other recording of
data for outlier reason, or for any other reason. operating activity during each test;
(d) results of laboratory fuel analysis;
(e) instrumentation calibration results from labora-
6.6 CONCLUSIONS
tories;
The conclusion section should address: (f) certification from manufacturers.
32
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002
MANDATORY APPENDIX I
UNCERTAINTY ANALYSIS AND SAMPLE CALCULATION
33
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 MANDATORY APPENDIX I
(e) These are then root-sum-squared to obtain a variations of the system being tested, the test proce-
95% confidence uncertainty as follows: dure, and the performance of the test.
sensitivity: sometimes called influence coefficient;
冤 冥
1
+ U95 p ±2 (B/2)2 + (S−x )2 ⁄2
the ratio of the change in a result to a unit change
in a parameter.
which is equivalent to systematic error: sometimes called bias; the true
systematic or fixed error that characterizes every
34
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
MANDATORY APPENDIX I ASME PTC 50-2002
(e) Calculate the Uncertainty. Calculation of un- el p 100 [(3600 En) / QItotal ]
certainty is done in accordance with para. 7.5 of
ASME PTC 19.1, combining the systematic and ran- where
dom uncertainties to get the total uncertainty. 100 p to convert to percent
(f) Prepare the Report in Accordance With ASME 3600 p kJ/kW·h conversion factor
PTC 19.1-1998. En p net electrical energy for the test duration
ASME PTC 19.1 provides a thorough treatment from para. 5.3.4, kW·h
of the subject and includes specific requirements el p electrical conversion efficiency, %
for report format, general guidance, details of special QItotal p total energy into the system for calcula-
calculations that may be required in some circum- tion of input energy and electrical con-
stances, and background information. version efficiency from para. 5.2.5.9
35
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 MANDATORY APPENDIX I
TABLE I-2 VARIOUS MEASURED PARAMETERS Mf p total fuel flow to the fuel cell during
AND THE NOMINAL CALCULATED RESULTS the test period, kg
Calculated Results MWf p molecular weight of the gaseous fuel
(16.043 kg/kmol for methane)
Nominal Pf p fuel pressure at the test boundary, kPa
Results Description Units Value
(absolute)
Qf Energy content of kJ 7.2000E+05 R p Ru /MWf
fuel, kJ Ru p 8.314 kJ/kmol K (universal gas con-
Ecf Compression energy kJ 6.9522E+02 stant)
content of fuel
RHf p relative humidity of the fuel, at the
TQf Temperature energy kJ 5.9618E+01
content of fuel test boundary, percent R.H.
Tff p fuel temperature at the test bound-
ary, K
QItotal p total energy into the system for calcula- Tstd p standard temperature (15°C) p
tion of input energy and electrical con- 288.15 K
version efficiency, kJ
Table I-2 lists the measured parameters and their
For this example, Qst , Eco , TQo , and Wsi are nominal calculated values.
all taken to be zero so the total input energy equation
reduces to: I-6.1.2 List Elemental Error Sources. Elemental
error sources can be estimated based on judgment,
QItotal p Qf + Ecf + TQf calculated using prior test data, or developed using
calibration laboratory calculations. Elemental error
3600En sources for the various parameters in the example
el p ⴛ 100 procedure are given in Table I-3.
Qf + Ecf + TQf
I-6.1.3 Calculate or Assign the Systematic
And the efficiency equation can be expanded as: Uncertainty and Random Standard Deviation for
Each Parameter
(a) Systematic Uncertainty. Calibration errors will
Qf p (Mf ) (LHVavg ) (from para. 5.2.5.1)
account for the majority of the systematic uncertainty
and and so they are used with judgment as the absolute
systematic uncertainty for those instruments where
absolute fuel inlet the calibration uncertainty is known. For instruments
pressure at test boundary, kPa or instrument loops where calibration is ±1% of
Ecf p(Mf RTstd ) ln full scale, the absolute uncertainty is calculated by
ambient barometric pressure, kPa
multiplying 0.01 by the full-scale reading of the
instrument. For instruments where the calibration
and uncertainty is ±1% of reading, the absolute uncer-
tainty is calculated by multiplying 0.01 by the nomi-
TQf p Mf Cfp (Tff − 288.15 K) nal reading. Different calibration uncertainties will
require different numerical inputs.
where All uncertainties are assumed to be at 95% confi-
Cfp p specific heat at constant pressure of dence.
fuel delivered during the test period, For the fuel mass flowmeter, a revenue quality
kJ/(kg·K) meter with an uncertainty of ±0.75% (of reading)
Ecf p ideal pressure energy content of the is available.
fuel, kJ For the fuel heating value, laboratory procedures
En p net electrical energy for the test dura- are adequate to achieve and uncertainty of ±0.5%
tion from para. 5.3.4, kW·h of reading.
LHVavg p fuel low heating value at the test The molecular weight of the fuel can be calculated
boundary for calculations based on by the laboratory with an uncertainty of ±0.5% of
Low Heating Value, kJ/kg the value.
36
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
MANDATORY APPENDIX I ASME PTC 50-2002
LHVavg Fuel lower heating value kJ/kg 50 000 Sampling errors, laboratory analysis errors,
errors in tabular data, mass chromatograph
calibration errors, random errors
MWf Molecular weight of fuel kg/kmol 16.043 Sampling errors, laboratory analysis errors,
Delivered errors in tabular data, mass chromatograph
calibration errors, random errors
Pf Fuel pressure at the test kPa 140 Pressure gauge calibration errors, pressure
Boundary transducer calibration errors, loop calibration
errors, random errors
Tff Fuel temperature at the test K 290 Temperature gauge calibration errors, pressure
boundary transducer calibration errors, loop calibration
errors, random errors
Cfp Fuel specific heat kJ/kg 2.2379 Sampling errors, laboratory analysis errors,
errors in tabular data, mass chromatograph
calibration errors, random errors
En Net kW·h delivered ... 100 Revenue meter calibration errors, loop
calibration errors, random errors
BP Ambient barometric pressure kPa 101.325 Pressure gauge calibration errors, loop
calibration errors, random error
37
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 MANDATORY APPENDIX I
LHVavg Fuel lower heating kJ/kg 50 000 0.005 4.9948E+01 4.9948E+01 −9.9791E-04
value
Mf Mass of fuel delivered kg 14.4 0.00000144 4.9948E+01 4.9948E+01 −3.4686E+00
MWf Molecular weight of kg/kmol 16.043 16.043E-06 4.9948E+01 4.9948E+01 3.0031E-03
fuel delivered
Pf Fuel pressure at the kPa 140 0.000014 4.9948E+01 4.9948E+01 −1.0644E-03
test boundary
Tff Fuel temperature at K 290 0.000029 4.9948E+01 4.9948E+01 −2.2332E-03
the test boundary
Cfp Fuel specific heat kJ/kg 2.2379 2.23793E-07 4.9948E+01 4.9948E+01 −1.8461E-03
En Net kW·h delivered ... 100 0.00001 4.9948E+01 4.9948E+01 4.9948E-01
3 600 Conversion factor ... 3 600 ... ... ... ...
100 Conversion factor ... 100 ... ... ... ...
Ru Universal gas kJ/ 8.314 ... ... ... ...
constant kmol K
288.15 Standard temperature K 288.15 ... ... ... ...
BP Ambient barometric kPa 101.325 0.101325 4.9948E+01 4.9948E+01 1.4699E-03
pressure
The fuel pressure gauge is able to achieve an method (see Appendix C of ASME PTC 19.1-1998).
uncertainty of ±5% of full scale. This requires a calculation of the sensitivity factors,
The fuel temperature gauge is able to achieve an either by differentiation or by computer perturbation.
uncertainty of ±2 K. The systematic uncertainty and random standard
Fuel specific heat is assumed to be measured by deviations for each parameter must be multiplied
the laboratory within an uncertainty of ±0.5% of by the proper sensitivity in accordance with the
measured value. following equation.
Net kW·h delivered can be measured within an
uncertainty of ±0.5% of reading. i p sensitivity coefficient for the parameter i
Ambient barometric pressure can be measured
within an uncertainty of ±5 kPa of reading. In order to propagate the various uncertainties
properly through the equation, the various sensitivi-
(b) Random Uncertainty. Since this is a pretest ties of each parameter must be calculated. The
uncertainty estimate, the random uncertainty contri- sensitivity of a particular parameter is calculated by
bution must be estimated from previous data. Prior
either taking the partial differential of the parameter
test data indicate that the absolute standard deviation
with respect to the result (the efficiency in this case)
of the mean of one-hour tests with one-minute data
or doing a computer perturbation of the data fields
sets is as shown in Table I-4. For a post-test analysis,
using small changes in each parameter indepen-
the actual standard deviation must be used. If the
dently to ascertain the change in the result for a
random uncertainty is too high, additional test dura-
small change in the parameter. See Section 7 of
tion and additional data points should lower the
standard deviation. ASME PTC 19.1-1998 for the details of this process.
For this example, a computer spreadsheet perturba-
I-6.1.4 Propagate the Systematic and Random tion was done and the results are shown as Table
Standard Deviation for Each Parameter. The system- I-5. Table 1-5 gives the various parameters, the
atic and random (sample) standard deviations of the perturbation amount used in the computer program,
independent parameters are propagated separately and the resulting sensitivities.
all the way to the final result.
Propagation of the standard deviations is done, I-6.1.5 Calculate the Uncertainty. Calculation of
according to the functional relationship defined in uncertainty is done in accordance with para. 7.5
section I-5, step (a)(4), by using the Taylor series and Section 9 of ASME PTC 19.1, combining the
38
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
MANDATORY APPENDIX I ASME PTC 50-2002
LHVavg Fuel lower heating kJ/kg 50 000 2.5000E+02 8.5000E+01 −9.9791E-04 1.5560E-02 7.1948E-03
value
Mf Mass of fuel delivered kg 14.4 1.0800E-01 3.6000E+02 −3.4686E+00 3.5083E+00 1.5592E-02
MWf Molecular weight of kg/kmol 16.043 8.0215E-02 2.7273E-02 3.0031E-03 1.4507E-08 6.7081E-09
fuel delivered
Pf Fuel pressure at the kPa 140 7.0000E+00 7.0000E+00 1.0644E-03 1.3879E-05 5.5514E-05
test boundary
Tff Fuel temperature at K 290 2.0000E+00 1.0000E+00 −2.2332E-03 4.9874E-06 4.9874E-06
the test boundary
Cfp Fuel specific heat kJ/kg 2.2379 1.1190E-02 4.4759E-03 1.8461E-03 1.0668E-10 6.8278E-11
En Net kW·h delivered ... 100 5.0000E-01 2.0000E-01 4.9948E-01 1.5592E-02 9.979E-03
3 600 Conversion factor ... 3600 0 0 ... ... ...
100 Conversion factor ... 100 0 0 ... ... ...
Ru Universal gas kJ/kmolK 8.314 0 0 ... ... ...
constant
288.15 Standard temperature K 288.15 0 0 ... ... ...
BP Ambient barometric kPa 101.325 5.0000E+00 5.0000E-01 1.4699E-03 1.3505E-05 5.4018E-07
pressure
systematic and random uncertainties to get the total Table I-6 gives the parameters, the nominal values
uncertainty. expected, the calculated sensitivities, the absolute
The absolute systematic uncertainty contribution systematic uncertainty contribution, and the absolute
is calculated for each parameter separately by multi- random uncertainty contribution for each parameter,
plying one half the absolute systematic incertainty calculated as described above. All of these calcula-
by the Sensitivity for that parameter and squaring tions are based on ASME PTC 19.1-1998, Section 9.
the result: The Total Absolute Uncertainty of the Result is
computed by combining the absolute systematic
2 uncertainty with the absolute random uncertainty in
冢 冣
Bi
the following manner.
2 i
URp冪B2R + 2SR2
where
i p sensitivity coefficient for the parameter i where
Bi p absolute systematic uncertainty for the pa- BR p the absolute systematic uncertainty
rameter i
冪冢
2
冣
Bi
The absolute random uncertainty contribution for p2 ⌺
2 i
parameter i is calculated similarly:
冪冢
2
where p2 ⌺ S
x,i
i 冣
p sensitivity coefficient for the parameter i
S−
x,i p absolute standard deviation of the mean
for thei parameter.
39
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME PTC 50-2002 MANDATORY APPENDIX I
Table I-7 shows the final results with the absolute lute uncertainty of the result. This table also shows
systematic uncertainty of the result, the absolute the percent uncertainty, expressing the absolute un-
random uncertainty of the result, and the total abso- certainty as a percent of the result.
40
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w