Professional Documents
Culture Documents
A111D7 2576.^
QC-
Jane E. Callanan, Sandra A. Sullivan,
100
and Dominic F. Vecchia
U57
260-99
1985
TMm he National Bureau of Standards was established by an act of Congress on March 3, 1901. The
Bureau's overall goal is to strengthen and advance the nation's science and technology and facilitate
their effective application for public benefit. To this end, the Bureau conducts research and provides: (1) a
basis for the nation's physical measurement system, (2) scientific and technological services for industry and
government, (3) a technical basis for equity in trade, and (4) technical services to promote public safety.
The Bureau's technical work is performed by the National Measurement Laboratory, the National
Engineering Laboratory, the Institute for Computer Sciences and Technology, and the Institute for Materials
Science and Engineering
Headquarters and Laboratories at Gaithersburg, MD, unless otherwise noted; mailing address
Gaithersburg, MD 20899.
2
Some divisions within the center are located at Boulder, CO 80303.
'Located at Boulder. CO. with some elements at Gaithershurf> MD
Standard Reference Materials:
Dominic F. Vecchia
Sponsored by:
Office of Standard Reference Materials
National Measurement Laboratory
National Bureau of Standards
Gaithersburg, MD 20899
For sale by the Superintendent of Documents, U.S. Government Printing Office, Washington, DC 20402
Preface
Standard Reference Materials (SRM's) as defined by the National Bureau of Standards (NBS) are
well-characterized materials, produced in quantity and certified for one or more physical or
chemical properties. They are used to assure the accuracy and compatibility of measurements
throughout the Nation. SRM's are widely used as primary standards in many diverse fields in
science, industry", and technology, both within the United States and throughout the world. They are
also used extensively in the fields of environmental and clinical analysis. In many applications,
traceability of quality control and measurement processes to the national measurement system is
carried out through the mechanism and use of SRM's. For many of the Nation's scientists and
technologists it is therefore of more than passing interest to know the details of the measurements
made at NBS in arriving at the certified values of the SRM's produced. An NBS series of papers, of
which this publication is a member, called the NBS Special Publication - 260 Series , is reserved for
this purpose.
The 260 Series is dedicated to the dissemination of information on different phases of the
preparation, measurement, certification and use of NBS SRM's. In general, much more detail will be
found in these papers than is generally allowed, or desirable, in scientific journal articles. This
enables the user to assess the validity and accuracy of the measurment processes employed, to judge
the statistical analysis, and to learn details of techniques and methods utilized for work entailing
the greatest care and accuracy. These papers also should provide sufficient additional information
not found on the certificate so that new applications in diverse fields not foreseen at the time the
SRM was originally issued will be sought and found.
Inquiries concerning the technical content of this paper should be directed to the author(s).
Other questions concered with the availability, delivery, price, and so forth, will receive prompt
attention from:
iii
OTHER NBS PUBLICATIONS IN THIS SERIES
Catalog of NBS Standard Reference Materials Spijkerman, J. L., Snediker, D. K., Ruegg, F. C,
(1984-85 edition), Catherine H. Hudson, ed , NBS and DeVoe, J. R., Standard Reference Mate-
Spec. Publ. 260 (February 1984). $5.50* SN003- rials: Mossbauer Spectroscopy Standard for the
Shultz, J. I., Standard Reference Materials: Marinenko, G., Sappenfield, K. M., and Shields,
Methods for the Chemical Analysis of White W. R., Standard Reference Materials: Boric
Cast Iron Standards, NBS Misc. Publ. 260-6 Acid; Isotopic and Assay Standard Reference
(July 1965).COM74-11068** Materials, NBS Spec. Publ. 260-17 (February
Bell, R. K., Standard Reference Materials: 1970). Out of Print.
Methods for the Chemical Analysis of NBS Geller, S. B., Mantek, P.A., and Cleveland, N.
Yakowitz, H., Vieth, D. L., and Michaelis, R. E., L., Standard Reference Materials: Comparison
Standard Reference Materials: Homogeneity of Redox Standards, NBS Spec. Publ. 260-24
Characterization of NBS Spectrometric Stand- (January 1972). COM72-50058**
ards III: White Cast Iron and Stainless Steel
Powder Compact, NBS Misc. Publ. 260-12
(September 1966). NBS Misc. Publ. 260-12**
iv
Hicho, G. E., Yakowitz, H., Rasberry, S. D., and Richmond, J. C, and Hsia, J. J., Standard Refer-
Michaelis, R. E., Standard Reference Materials: ence Materials: Preparation and Calibration of
A Standard Reference Material Containing Standards of Spectral Specular Reflectance,
Nominally Four Percent Austenite, NBS Spec. NBS Spec. Publ. 260-38 (May 1972). COM72-
Publ. 260-25 (February 1971). COM74-1 1356** 50528**
Martin, J. F., Standard Reference Materials: Clark, A. F., Denson, V.A., Hust, J. G., and
National Bureau of Standards-US Steel Corpora- Powell, R. L., Standard Reference Materials:
tion Joint Program for Determining Oxygen and The Eddy Current Decay Method for Resistivity
Nitrogen in Steel, NBS Spec. Publ. 260-26 Characterization of High-Purity Metals, NBS
(February 1971). 85 cents* PB 81176620 Spec. Publ. 260-39 (May 1972). COM72- 50529**
Garner, E.L., Machlan, L. A., and Shields, W. R.,
McAdie, H. G., Garn, P.D., and Menis, O., Stand-
Standard Reference Materials: Uranium ard Reference Materials: Selection of Thermal
Isotopic Standard Reference Materials, NBS Analysis Temperature Standards Through a
Spec. Publ. 260-27 (April 1971). COM74- Cooperative Study (SRM 758, 759, 760), NBS
11358**
Spec. Publ. 260-40 (August 1972.) COM72-
Heinrich, K. F. J., Myklebust, R. L., Rasberry, S. 50776**
D., and Michaelis, R. E., Standard Reference Wood, L. A., and Shouse, P. J., Standard
Materials: Preparation and Evaluation of
Reference Materials: Use of Standard Light-
SRM's 481 and 482 Gold-Silver and Gold- Sensitive Paper for Calibrating Carbon Arcs
Copper Alloys for Microanalysis, NBS Spec. Used in Testing Textiles for Colorfastness to
Publ. 260-28 (August 1971). COM71-50365** Light, NBS Spec. Publ. 260-41 (August 1972)
Standard Reference Materials: Cali-
Geller, S. B., COM72-50775**
bration of NBS Secondary Standard Magnetic Wagner, H. L. and Verdier, P. H., eds.. Standard
Tape (Computer Amplitude Reference) Using Reference Materials: The Characterization of
the Reference Tape Amplitude Measurement Linear Polyethylene, SRM 1475, NBS Spec.
"Process A-Model 2," NBS Spec. Publ. 260-29 Publ. 260-42 (September 1972). COM72-
COM71-50282
(June 1971). 50944**
Gorozhanina, R. S., Freedman, A. Y., and Yakowitz, H., Ruff, A. W., and Michaelis, R. E.,
M. C. Selby),
Shaievitch, A. B. (translated by Standard Reference Materials: Preparation and
Standard Reference Materials: Standard Homogeneity Characterization of an Austenitic
Samples Issued in the USSR (A Translation Iron-Chromium-Nickel Alloy, NBS Spec. Publ.
from the Russian). NBS Spec. Publ. 260-30 (June 260-43 (November 1972). COM73-50760**
1971). COM71-50283** Schooley, J. F., Soulen, R. J., Jr., and Evans, G. A.,
Hust, J. and Sparks, L. L., Standard Reference
G., Jr.,Standard Reference Materials: Preparation
Materials: Thermal Conductivity of Electrolytic and Use of Superconductive Fixed Point
Iron SRM 734 from 4 to 300 K, NBS Spec. Publ. Devices, SRM 767, NBS Spec. Publ. 260-44
260-31 (November 1971). COM71-50563** (December 1972). COM73-50037**
Mavrodineanu, R., and Lazar, J. W., Standard Greifer, B., Maienthal, E. J., Rains, T. C, and
Reference Materials: Standard Quartz Cuvettes, Rasberry, S. D., Standard Reference Materials:
for High Accuracy Spectrophotometry, NBS Powdered Lead-Based Paint, SRM 1579, NBS
Spec. Publ. 260-32 (December 1973). 55 cents* Spec. Publ. 260-45 (March 1973). COM73-
SN003-003-01213-1 50226**
Wagner, H. L., Standard Reference Materials: Hust, J. G., and Giarratano, P. J., Standard Refer-
Comparison of Original and Supplemental ence Materials: Thermal Conductivity and
SRM 705, Narrow Molecular Weight Distri- Electrical Resistivity Standard Reference Mate-
bution Polystyrene, NBS Spec. Publ. 260-33 rials: Austenitic Stainless Steel, SRM's 735 and
(May 1972). COM72- 50526** 798, from 4 to 1200 K, NBS Spec. Publ. 260-46
Sparks, L. L., and Hust, J. G., Standard Reference (March 1975). SNO03-OO3-0I278-5*
Materials: Thermoelectric Voltage, NBS Spec. Hust, J. G., Standard Reference Materials: Elec-
Publ. 260-34, (April 1972). COM72-50371** trical Resistivity of Electrolytic Iron, SRM 797,
Sparks, L. and Hust, J. G., Standard Refer-
L., and Austenitic Stainless Steel, SRM 798, from 5
ence Materials: Thermal Conductivity of to 280 K, NBS Spec. Publ. 260-47 (February
Austenitic Stainless Steel, SRM 735 from 5 1974). COM74-50176**
to 280 K, NBS Spec. Publ. 260-35 (April 1972.) Mangum, B. W., and Wise, J. A., Standard Refer-
35 cents* COM72-50368** ence Materials: Description and Use of Precision
Mandel, J., Moore, L. J., and Young, D.
Cali, J. P., Thermometers for the Clinical Laboratory,
S., Standard Reference Materials: A Referee SRM 933 and SRM 934, NBS Spec. Publ. 260-48
Method for the Determination of Calcium in (May 1974). 60 cents* SN003-003-0 1278-5
Serum, NBS SRM 915, NBS Spec. Publ. 260-36 Carpenter, B. S., and Reimer, G. M., Standard
(May 1972). COM72-50527** Reference Materials: Calibrated Glass Stand-
Shultz, J. I. Bell., R. K. Rains, T. C, and Menis, ards for Fission Track Use, NBS Spec. Publ.
O., Standard Reference Materials: Methods of 260-49 (November 1974). COM74-51185
Analysis of NBS Clay Standards, NBS Spec.
Publ. 260-37 (June 1972). COM72-50692**
v
Hust, J. G., and Giarratano, P. J., Standard Refer- Soulen, R. J., and Dove, R. B., Standard Reference
ence Materials: Thermal Conductivity and Elec- Materials: Temperature Reference Standard
trical Resistivity Standard Reference Materials: for Use Below 0.5 K (SRM 768). NBS Spec.
Electrolytic Iron, SRM's 734 and 797 from 4 to Publ. 260-62 (April 1979). $2.30* SN003-003-
1000 K, NBS Spec. Publ. 260-50 (June 1975). 02047-8
$1.00* SN003-003-0 1425-7 Velapoldi, R. A., Paule, R. C, Schaffer, R.
Mavrodineanu, R., and Baldwin, J. R., Standard Mandel, J., Machlan, L. A., and Gramlich, J. W.,
Reference Materials: Glass Filters As a Standard Standard Reference Materials: A Reference
Reference Material for Spectrophotometry; Method for the Determination of Potassium in
Selection; Preparation; Certification; Use-SRM Serum. NBS Spec. Publ. 260-63 (May 1979).
930, NBS Spec. Publ. 260-51 (November 1975). $3.75* SN00 3-00 3-02068
$1.90* SN003-003-0I48I-8 Velapoldi, R. A., and Mielenz, K. D., Standard
Hust, J. G., and Giarratano, P. J., Standard Refer- Reference Materials: A Fluorescence Standard
ence Materials: Thermal Conductivity and Elec- Reference Material Quinine Sulfate Dihydrate
trical Resistivity Standard Reference Materials (SRM 936), NBS Spec. Publ. 260-64 (January
730 and 799, from 4 to 3000 K, NBS Spec. Publ. 1980). $4.25* SN003-003-02 148-2
260-52 (September 1975). $1.05* SN003-003- Marinenko, R. B., Heinrich, K. F. J., and Ruegg,
01464-8 F. C, Standard Reference Materials: Micro-
Durst, R. A., Standard Reference Materials: Homogeneity Studies of NBS Standard Reference
Standardization of pH Measurements, NBS Materials, NBS Research Materials, and Other
Spec. Publ. 260-53 (December 1975, Revised). Related Samples. NBS Spec. Publ. 260-65
$1.05 SN003-003-01551-2 (September 1979). $3.50* SN003-003-021 14-1
Burke, R. W., and Mavrodineanu, R., Standard Venable, W. H., Jr., and Eckerle, K. L., Standard
Reference Materials: Certification and Use of Reference Materials: Didymium Glass Filters for
Acidic Potassium Dichromate Solutions as an Calibrating the Wavelength Scale of Spectro-
Ultraviolet Absorbance Standard, NBS Spec. photometers (SRM 2009, 2010, 2013). NBS Spec.
Publ. 260-54 (August 1977). $3.00* SN003-003- Publ. 260-66 (October 1979). $3.50* SN003-003-
01828-7 02127-0
Ditmars, D. A., Cezairliyan, A., Ishihara, S., and Velapoldi, R. A., Paule, R. C, Schaffer, R., Mandel,
Douglas, T. Standard Reference Materials:
B., J., Murphy, T. J., and Gramlich,
J. W., Standard
Enthalpy and Heat Capacity; Molybdenum Reference Materials:A Reference Method forthe
SRM 781, from 273 to 2800 K, NBS Spec. Publ. Determination of Chloride in Serum, NBS Spec.
260-55 (September 1977). $2.20* SN003-003- Publ. 260-67 (November 1979). $3.75* SN003-
01836-8 003-02136-9
Powell, R. L., Sparks, L. L., and Hust, J. G., Mavrodineanu, R. and Baldwin, J. R., Standard
Standard Reference Materials: Standard Reference Materials: Metal-On-Quartz Filters as a
Thermocouple Materials, Pt.67: SRM 1967, NBS Standard Reference Material for Spectrophoto-
Spec. Publ. 260-56 (February 1978). $2.20* metry-SRM 2031, NBS Spec. Publ. 260-68
SN003-003-0 18864 (April 1980). $4.25* SN003-003-02 167-9
Cali, J. P. and Plebanski, T., Guide to United Velapoldi, R. A., Paule, R. C, Schaffer, R.,
States Reference Materials, NBS Spec. Publ. Mandel, J., Machlan, L. A., Garner, E. L., and
260-57 (February 1978). $2.20* PB 277173 Rains, T. C, Standard Reference Materials: A
Barnes, and Martin, G. M., Standard Refer-
J. D., Reference Method for the Determination of
ence Materials: Polyester Film for Oxygen Gas Lithium in Serum, NBS Spec. Publ. 260-69 (July)
Transmission Measurements SRM 1470, NBS 1980). $4.25* SN003-003-02214-4
Spec. Publ. 260-58 (June 1979) $2.00* SN003- Marinenko, R. B., Biancaniello, F., Boyer, P. A.,
003-02077 Ruff, A. W„ DeRobertis, L., Standard Reference
Chang, T., and Kahn, A. H. Standard Reference Materials: Preparation and Characterization of
Materials: Electron Paramagnetic Resonance an Iron-Chromium-Nickel Alloy for Micro-
Intensity Standard; SRM 2601, NBS Spec. analysis, NBS Spec. Publ. 260-70 (May 1981).
Publ. 260-59 (August 1978) $2.30* SN003- $2.50* SN003-003-02328-1
003-01975-5 Seward, R. W., and Mavrodineanu, R., Standard
Velapoldi, R. A., Paule, R. C, Schaffer, R., Reference Materials: Summary of the Clinical
Mandel, J., and Moody, J. R., Standard Refer- Laboratory Standards Issued by the National
ence Materials: A Reference Method for the Bureau of Standards, NBS Spec. Publ. 260-71
Determination of Sodium in Serum, NBS Spec. (November 1981). $6.50* SN00 3-003-0238 1-7
Publ. 260-60 (August 1978). $3.00* SN003-003 Reeder, D.J., Coxon, B., Enagonio, D., Christensen,
01978-0 R. G., Schaffer, R., Howell, B. F., Paule, R. C,
Verdier, P. H., and Wagner. H. L., Standard Refer- Mandel, Standard Reference Materials: SRM
J.,
ence Materials: The Characterization of Linear 900, Drug Level Assay Standard,
Antiepilepsy
Polyethylene (SRM 1482, 1483, 1484), NBS NBS Spec. Publ. 260-72 (June 1981). $4.25*
Spec. Publ. 260-61 (December 1978). $1.70* SN00 3-003-02329-9
SN003-003-02006-I
vi
G.,and Hicho, G. E., Standard Refer-
Interrante, C. Swyt, D. A., Standard Reference Materials: A Look
ence Materials: A Standard Reference Material at Techniques for the Dimensional Calibration of
Containing Nominally Fifteen Percent Austen te :
Standard Microscopic Particles, NBS Spec. Publ.
(SRM 486), NBS Spec. Publ. 260-73 (January 260-85 (September 1983). $5.50* SN003-003-
Materials: A Standard Reference Material Roelandts, I., and Gills, T. E., Standard Reference
Containing Nominally Five Percent Austenite Materials: 1982 Compilation of Elemental
(SRM 485a), NBS Spec. Publ. 260-76 (August Concentration Data for NBS Biological, Geo-
1982). $3.50 SN003-003-024-33-3 logical, and Environmental Standard Reference
Furukawa, G. T., Riddle, J. L., Bigge, W. G., and Materials. Spec. Publ. 260-88 (March 1984).
Pfieffer, E. R., Standard Reference Materials: $7.00* SN003-003-02565-8
Application of Some Metal SRM's as Thermo- Hust, J. G., Standard Reference Materials: A Fine-
metric Fixed Points, NBS Spec. Publ. 260-77 Grained, Isotropic Graphite for Use as NBS
Wilmering, D. B., Standard Reference Materials: NBS Spec. Publ. 260-90 (September 1984). $3.00*
Second Surface Mirror Standards of Specular SN003-003-02609-3
Spectral Reflectance (SRM's 2023, 2024, 2025), Goodrich, L. F., Vecchia, D. F., Pittman, E. S.,
NBS Spec. Publ. 260-79 (October 1982). $4.50* Ekin, J. W., and Clark, A. F., Standard Reference
SN003-003-024-47-3 Materials: Critical Current Measurements on an
Schaffer, R., Mandel, Sun, T., Cohen, A., and
J., NbTi Superconducting Wire Standard Reference
Hertz, H. S., Standard Reference Materials: Material, NBS Spec. Publ. 260-91 (September
Evaluation by an ID/ MS Method of the AACC 1984) . $2.75* SN003-00 3-026 14-0
Reference Method for Serum Glucose, NBS Carpenter, B. S., Standard Reference Materials:
Spec. Publ. 260-80 (October 1982). $4.75* SN003- Calibrated Glass Standards for Fission Track Use
003-024-43-1 (Supplement to NBS Spec. 260-49). NBS
Burke, R. W., and Mavrodineanu, R. (NBS retired). Spec. Publ.. 260-92 (September 1984). $1.50*
Standard Reference Materials: Accuracy in SN003-003-02610-7
Analytical Spectrophotometry, NBS Spec. Publ. Ehrstein, J., Preparation and Certification of
260-81 (April 1983). $6.00* SN003-003-024-8 Standard Reference Materials for Calibration of
Weidner, V. R., Standard Reference Materials: Spreading Resistance Probes, NBS Spec. Publ.
White Opal Glass Diffuse Spectral Reflectance 260-93 (January 1985).
Standards for the Visible Spectrum (SRM's 2015 Gills, T. E., Koch, W. F. Stolz, J. W., Kelly, W. R.,
and 2016). NBS Spec. Publ. 260-82 (April 1983). Paulsen, P. J., Colbert, J. C. Kirklin, D. R.,
$3.75* SNO03-003-024-89-9** Pei, P.T.S., Weeks, S., Lindstrom, R. M.
Bowers, G. N., Jr., Alvarez, R., Cali, J. P. (NBS Fleming, R. F., Greenberg, R. R., and
retired), Eberhardt, K. R., Reeder, D. J., Paule, R. C, Methods and Procedures Used at the
Schaffer, R., Uriano, G. A., Standard Reference National Bureau of Standards to Certify Sulfur
Materials: The Measurement of the Catalytic in Coal SRM's for Sulfur Content, Calorific
(Activity) Concentration of Seven Enzymes in Value, Ash Content, NBS Spec. Publ. 260-94
NBS Human Serum SRM 909, NBS Spec. (December 1984).
Publ. 260-83 (June 1983). $4.50* SNOO3-OO3-024- Mulholland, G. W., Hartman, A. W., Hembree,
99-6 G. G., Marx, E., and Lettieri, T. R., Standard
Gills, T. E., Seward, R. W., Collins, R. J., and Reference Materials: Development of a 1 urn
Webster, W. C, Standard Reference Materials: Diameter Particle Size Standard, SRM 1690,
Sampling, Materials Handling, Processing, and NBS Spec. Publ. 260-95 (May 1985).
Packaging of NBS Sulfur in Coal Standard Carpenter, B. S., Gramlich, J. W., Greenberg, R. R.,
Reference Materials, 2682, 2683, 2684, and 2685, and Machlan, L. A., Standard Reference
NBS Spec. Publ. 260-84 (August 1983). $4.50* Materials: Uranium-235 Isotopic Abundance
SN0O3-O03-O25-20-8 Standard Reference Materials for Gamma
Spectrometry Measurements, NBS Spec. Publ.
260-96 (In Press).
vii
Mavrodineanu, R. and Gills, T. E., Standard
Reference Materials: Summary of the Coal, Ore,
Mineral, Rock, and Refractory Standards Issued
by the National Bureau of Standards, NBS Spec.
Publ. 260-97 (In Press).
Hust, J. Standard Reference Materials: Glass
G.,
Fiberboard SRM for Thermal Resistance, NBS
Spec. Publ. 260-98 (In Press).
Callanan, J. E., Sullivan, S. A., and Vecchia, D. F.,
Page
List of Tables x
List of Figures xi
A. Procedures 6
C. Summary 16
A. Procedures 17
1. General 17
2. Temperature calibration 18
3. Enthalpy calibration 20
B. Results and discussion 21
C. Summary 26
References 33
Appendix A-l
ix
List of Tables
Page
Table 9. Results for repeat runs with specimens remounted (112) .... 14
x
List of Figures
Page
xi
Feasibility Study for the Development of Standards Using Differential Scanning
Cal orimetry
precision calorimetry, has created a need for more and better thermal
standards for use with scanning calorimeters and other thermal instru-
required, the associated expense, and the lack of facilities and per-
factors which affected it; Part II applied American Society for Testing
and Materials (ASTM) procedures for the temperature and heat flow cali-
bration. The study shows that fusion standards can be developed with a
The last decade has seen a tremendous increase in the number of differential
scanning calorimeters (DSC) 1 use. During this same decade, the number of
in
*The acronym DSC refers to an instrument; d.s.c. refers to the calorimetric tech-
nique.
adiabatic calorimeters available has declined alarmingly. Because of this de-
cline many investigators are constrained to use DSCs, but require the maximum ac-
curacy and precision possible. A paper to be published shortly in the open lit-
erature outlines the prescriptions for use found satisfactory in our labora-
tory [1].
current, the latter two obtained by use of a calibrated standard cell and cali-
eliminate heat transfer to the environment. The quantities required for the de-
cups, and a differential circuit that either keeps both cups at the same tempera-
ture or evaluates the temperature difference between them. It does not have an
electrical calibration heater nor the means to measure (and/or design to mini-
As the heat leak is large and somewhat dependent on the environment, standards
calibrants each time measurements are made. The validity of the results obtained
the correct value for the cali brant or the data may be corrected as required by
ability and the time that may be required for instrument adjustment, the most
values close to those in the literature and then to correct data as indicated by
imeters (DSC) are the Special Reference Materials (GM) distributed by the Office
of Standard Reference Materials (OSRM) as GM 754 and 757-60 [2]. These materials
late 1 960 s
in the strict sense. The temperature variations noted in the ICTA certificate
When the DSC is used in the scanning mode, as it is generally, standards be-
sample being tested, i.e., that will have similar thermal lag. Metallic stand-
ards are preferred for studies of metals, organic materials for organics, pow-
for Testing and Materials (ASTM) work best when calibrants available are about
50 K apart [7,8].
rials (SRM's) for d.s.c. Instruments commercially available today readily give
an order of magnitude improved precision over the instruments used for the ICTA
standards. There has been an explosion in the use of differential scanning calo-
rimetry; thus needs for standards are greater. Present day energy costs require
improved thermal property data. Facilities for obtaining very accurate and
-3-
• • • • •
i:
T
CM 2 a
3 *S338930
-4-
precise calorimetric data are few and far between, and becoming more so [9]. In
addition, the cost associated with these measurements is high. For many purposes
industry does not need data of the accuracy provided by methods comparable to
adiabatic calorimetry. For all these reasons it seemed worthwhile to attempt the
The first part of the study was intended to determine whether a differential
standards the usual practice is to develop the standard with a different method,
only one or two standards were required, precision calorimetry would be feasible,
but expensive and difficult. The facilities for making such measurements in the
temperature range needed most (300 - 1000 K) are virtually nonexistent or un-
to evaluate the variability inherent in the DSC and our procedures for working
with it, and thus the suitability of a DSC for the development of standards. The
experimental results; this analysis would consider both total variability and the
The second part of the study, which will be described in more detail later,
stances 1 and 3 were used, as bracketing substances, to calibrate the DSC and the
for bracketing and the properties of substance 3 obtained. This procedure, fol-
lowed through the selected list, would allow us to determine 1) whether the DSC
A. Procedures
the Statistical Engineering Division (714) of the Center for Applied Mathematics,
Gaithersburg [10].
Three forms of indium which were readily available (rod, granules, foil)
and enthalpy of transition. Information about the three forms of indium used for
The sixty replicate runs were made in random order. Since it was possible
that variations in mains voltage could affect the results, some samples were run
during the normal working day, some during start up and shut down periods, some
replicate measurements were made on six specimens without any operator interfer-
ence (leaving the specimens in place and rerunning them.) These allowed for the
results after completion of this work, two short studies were added. One of
these was designed to help us separate remount variations from specimen varia-
tions; the other allowed for comparison between two experienced operators.
-6-
The accuracy and precision of the weighing procedure were estimated from ten
was chosen as it is close to the mass of the specimens used. All mass determina-
The usual weighing procedure in this laboratory always involves duplicate weigh-
unusual variation, which may result from environmental factors, repeated measure-
gration routine for determining the area under the peak, i.e., the enthalpy of
fusion. Before the integration routine is called for in the data reduction
scheme, the operator must decide, from the scan, where the transition begins and
ends, and set cursors accordingly. Cursor judgement variations were determined
as follows. The same scan was used by three observers, each of whom made three
vised, it became obvious that we had no way to separate the variance due to re-
mounting the specimen in the DSC from that due to the specimen itself. For this
reason a single specimen was remounted five times. This operation was repeated
For the temperature calibration the instrument was first linearized over the
the observed onset temperature of fusion was matched to the literature value for
-7-
that standard. The enthalpy calibration of the instrument was based on an en-
thalpic heat capacity determination with sapphire. Temperature and enthalpy mea-
surements for the specimens were corrected from day to day as indicated by the
fusion standards, which were checked several times during each day of the study.
Our instrument was calibrated with an existing indium rod and a new tin
standard. On only one day were significant variations in the standard noted;
corrections for that day were made on the basis of the time of the calibration
and the experimental run. Because the corrected means obtained for the indium
specimens do not agree with the accepted values, it is obvious that the indium
fusion standard needed replacement. But, since the purpose of this work was to
The analysis in this report, coupled with information generated in our lab-
oratory, allowed for the determination of the significant variables and the mag-
nitude of their effects. Results of the analysis are discussed below. An ex-
planation of the methods used to account for specimen variability can be found in
Table 2 shows the estimated temperature and enthalpy of fusion for each form
from each of five specimens). The standard error of the mean^ is also included
the methods that were used to obtain the standard errors and their associated de-
grees of freedom can be found in the appendix. For reference the simple standard
2
Standard error of the mean is synonymous with standard deviation of the mean.
-8-
Table 2. Means obtained in indium variability study
Mean N Standard Error of Degrees of Standard Deviation
Mean* Freedom
Temperature
K K K
*See appendix
Though the DSC reads only to two decimals, a third is retained in reporting
derived results because of information that can sometimes be obtained from it.
Tables 3 and 4 list the original temperatures and enthalpies from which table 2
variables for comparisons made later in this paper, was selected at random from
values given for each form in tables 3 and 4. The mean, standard deviation and
The variations given in table 2 show the present data to be more precise, by
1-2 orders of magnitude, than the data shown in figure 1. Thus, these results
indicate that we can establish the temperature and enthalpy of fusion satisfac-
deviation considerably smaller than that noted with operator interference. The
results of these replicate runs are shown in table 5. In this table, three deci-
mals are retained, as earlier, for statistical purposes. The original data ap-
-9-
Table 3. Fusion temperatures for indium specimens (K)
Five-Specimen Results
-10-
Table 4. Enthalpies of fusion for indium specimens (J/g)
28 . 70 28.41* 28 . 62
Five-Specimen Results
Mean + Standard Deviation 28.504 + 0.257 28.726 + 0.424 28.970 + 0.204
Variance 0.066 0.180 0.042
-11-
Table 5. Mean and standard deviation for replicate runs, in succession, with no
disturbance of the specimen
Temperature Enthal py
Specimen Mean Standard Deviation Mean Standard Deviation Degrees i
-12-
The results of the evaluation of the weighing procedures, table 7, show that
b Excludes two measurements outside of our acceptable range. This result fol-
lows from our normal weighing procedure.
NOTE: The difference between the calibration mass of the 50 mg weight and
the average value obtained here is insignificant when the combined uncer-
tainties in the calibration weights and the standard deviation of the in-
strument are considered. For a weighing-by-difference procedure such as
used in this work, it is customary to use tne combined uncertainties
(0.0022 ng) plus three times the standard deviation.
lytical variance in table 10, was evaluated by having three operators evaluate a
single experimental run three times. These analyses were done relatively far
apart in time to reduce operator bias introduced by memory. There was no varia-
value, 429.87 K, was obtained in all nine evaluations. For the enthalpy evalua-
tion, the average value for all nine trials was 28.414 + 0.046 J/g, which corre-
sponds to a variability of 0.16%. Two of the operators always obtained the same
value. The third obtained different values which average to 28.380 + 0.063 J/g;
tine used in this laboratory cancels out systematic error involved in the inte-
gration. The random error associated with the integration arises from the same
sources as the random error associated with the instrument variation in this
-13-
study. The resulting contribution to the error of the measurements, therefore,
The results for the remount and operator study are shown in table 8. The
Operator 2
429.79 28.19
429.01 28.41
428.65 28.16
428.88 28.15
429.89 28.23
Mean 428.844 28.228
Standard Deviation 0.134 0.106
Mean corrected for 429.088 + 0.134 28.608 + 0.106
daily cal ibration
-14-
In order to determine whether time of run or mass of specimen affected the
The time at which the runs were made, either day of the week or time of day,
had no correlation with the results. Thus variations in mains voltage do not
3.49 mg. No correlation between specimen mass and either temperature or enthalpy
The statistical analysis indicated that somewhat "tighter" results were ob-
tained with the foil. In addition, auxilliary information showed that the most
serious variations appeared to exist among specimens of the same form, the
within-form variations.
was carried out. The range for each variable is also given. For the evaluation
of the analytical variability an existing indium scan was used. All the other
variables given in table 10 refer to foil specimens. Specimen 112 was used for
remount and operator variability tests. Five results each, both temperature and
enthalpy, were selected at random from the results for foil shown in tables 11
and 12 and used to obtain the estimate of total variance for a population of five
in table 10. (The value given in table 2 is for a population of 20.) The (1)
and (2) associated with remount variance refer to two different operators.
The total variability had contributions from the instrument itself, S-j ,
2 2 2
s = s + s + (1)
t s r
-15-
Table 10. Contributions to variance; range (foil)
Degrees
Temperature (K) Enthalpy (J/g) of Freedom
Range Variance Variance Range
0.00 (0.00) 2 (0.00) 2 0.00
0.00 (0.00) 2 analytical (0.055) 2
0.03
procedure
.03 (0.015) 2 instrument (0.088) z 0.17
2
0.33 (1)(0.144)
(0.106) 2 0.25
z
(2)(0.134)
0.01 (0.00) 2 operator (0.00) 2
0.58 (0.196) 2 TOTAL (0.204) 2 0.27
2 2
(0.133) specimen (0.172)
2
runs in place. Results for remounting the same specimen give Sr + S^ 2 ,
2 2
L 2 2
s/
t
- (S
r
+ S.
x
) = S
s
.
For this estimate the remount results for the primary operator were used and
the operator variance omitted. The contribution to the variance from the ana-
lytical procedure is included in the other contributions and was not subtracted
separately.
The estimate of specimen variability for enthalpy using this approach dif-
fers from that given in table 3 of the appendix, which was obtained using an al-
C. Summary
specimen results for foil given in tables 3 and 4, was 0.7%; in temperature,
-16-
0.05%. Thus, unquestionably, satisfactory values for SRM's for d.s.c. and other
thermal analysis techniques can be obtained using a DSC in its most precise modes
of operation. Also, metallic samples in foil form gave slightly better uni-
formity than rod; foil appears to be significantly better than granular speci-
mens. If, then, materials of equal purity are available for standards develop-
A. Procedures
1. General
The second part of this study called for an evaluation of the ASTM
The test materials selected were from the group of melting point standards
certified by the National Physical Laboratory (NPL) and marketed in the United
States through OSRM. The materials used and their reference temperatures are
The certificates for these materials indicate that these reference tempera-
tures refer to a specific heating regime, heating at 2 K/min with the specimen
contained in a glass capillary tube. Where other definitive work has been done,
-17-
We chose to use five specimens of each of these materials and run a four-
replicate analysis. This choice was based on the results of the indium study.
The specimen sets were randomly selected; within these sets, however, the
specimens were run in the order of increasing melting points. The two-point tem-
The protocol for temperature calibration calls for bracketing the tempera-
ture of interest with two known standards and then following the calculations
outlined below.
after the unknown runs, as well as at intervals in between. The actual number of
indium checks depended on the number of sets of samples run on any specific day.
In fact, the calibration checks were not necessary as the differences between ob-
served and literature values are incorporated into the equations. However, as
the indium runs are part of our normal procedures, they were made. Enthalpic
measurements of sapphire over the pertinent temperature ranges were made so that
2. Temperature calibration
T = (T x S) + I, (2)
Q
is the slope and I, the intercept. The slope and intercept are calculated from
I = [(TOj x TS )
- (TS x T0 )]/(T0 - T0 ). (4)
2 1 2 1 2
-18-
A preliminary evaluation of the method was based on results for metallic
well as those from a preliminary study of some organic NPL standards, are shown
in table 12.
Calculated
In,Pb Sn 505. 13
13
Indium 429. 784 430.07
11
Anisic Acid 456.45+0. 456.24
11
2-Chloroanthraquinone 482. 75+0. 481.83
14
Tin 505.118 505. 74
Calculated
In,Sn AA,2-CL 455.86, 481.83
In,2-CL AA 456.30
AA,Sn 2-CL 482.34
*The values cited as calculated in this table have been obtained using the two
substances in the first column as bracketing substances. The materials for which
calculated values were obtained are shown in the second column. The calculated
temperatures are shown in the third column.
-19-
3. Enthalpy calibration
protocol which was evaluated in the indium study discussed earlier in this re-
port. Corrections to the enthalpy were made as recommended in the ASTM Heat Flow
Protocol, E968 [8]. This protocol calls for determi nation of the melting iso-
comparison of the observed enthalpy of fusion with the literature value. This
A few changes were made in the recommended protocol. It was not necessary
rather than 10 K/min as, in our experience, accuracy is improved at lower scan
rates. The enthalpic method was used, rattier than a scanning method, to deter-
mine the heat capacities for extension of the calibration to other temperatures.
All specimens were run under identical conditions. Argon was used as purge gas,
set at 137.9 kPa gauge (20 psig). The specimen holders were hermetically sealed
better than 2 pg. For our purpose the peak melting temperature was not neces-
sary; for the equipment and program used, the onset temperature is considered to
Indium foil (99.999% pure) was used as the fusion standard and Calorimetry
Conference sapphire for the specific heat capacity measurements. The latter
measurements were made by the enthalpic method over a temperature interval chosen
-20-
so that the temperature of interest was at the midpoint of the run [4-6]. Since
our measurement system is automated, the equations specified in the ASTM protocol
The calibration coefficient (E) is obtained from the rat to of tire literature
value of the enthalpy of transition UH^) to- t*»* observed valu* t^^)*
E = AH(lit)/AH(obsJ, (5)
Then the true enthalpy of fusion UHact ) of measured specimens is obtained from
where AH(meas) is the measured enthalpy of fusion for that specimen. For exten-
results:
The AH(act) from eq (6) is multiplied by the ratio of the F-factor at the tem-
normally changes somewhat with temperature; for that reason an F-factor specific
Results for the temperature calibration study are shown in table 13 for the
initial heating only. Results for repeat meaurements are given in table 14. The
results for initial runs show the calculated temperatures in line 1. As the pro-
di phenylacetic acid were used to determine the values for acetanilide, which has
an intermediate melting temperature. Then acetanilide and anisic acid were used
to bracket di phenyl acetic acid, etc. The raw data used in the calculations are
shown in table 13. In all cases the results of this study for initial runs agree
with the certificate values. Though the standard deviations are greater than
-21-
Table 13a. Transition temperatures (K) - acetanilide (initial run)
Table 13b. Transition temperatures (K) - di phenyl acetic acid (initial run)
Di phenyl aceti c Acid 420.156 420.106 420.050 420.383 420.394 420.218 0.160
Anisic Acid 456.644 456.092 455.694 456. 552 455.910 456. 180 0.409
Diphenyl acetic Acid* 419.99 419.89 419. 79 420.07 418.37 419.62 0.708
2-chloroanthraquinone* 482.36 482.51 483.24 481.91 481.68 482.34 0.604
Anisic Acid* 456.30 455. 79 455. 76 455.98 454.48 455.66 0.695
Uncorrected data from which corrected values were obtained utilizing the procedure out-
lined in the text.
-22-
Table 14a. Transition temperatures (K) - acetanilide (repeat runs)
Table 14b. Transition temperatures (K) - di phenyl acetic acid (repeat runs)
-23-
those shown on the certificate, they are in line with those obtained in the
indium study. The calculated values for repeat runs are shown in table 14.
The results, shown in table 13, indicate that the ASTM two-point calibration
d.s.c. They also indicate that the substances used in this study are useful cal-
ibrants.
The results for the enthalpy measurements are shown in table 15. Columns 1-
5 refer to the replicate specimens. Rows 1-4 contain the data for the initial
and three subsequent runs. The columns under full correction refer to the ASTM
method using the results for sapphire in addition to those for indium, as de-
scribed earlier in the paper. The columns headed fusion correction refer to the
more commonly used method of correcting enthalpy by a factor obtained from an ex-
perimental indium scan compared to its accepted value without any extension to
for the fusion only method are listed both because this method represents the
most widely used method of making corrections and because a comparison of the two
methods is desirable.
The enthalpies obtained with the full correction differ significantly from
those with only the indium fusion correction for naphthalene, acetanilide and 2-
chl oroanthraqui none . The results using only the indium correction generally
agree with literature values within experimental error; results using the full
correction do not. Attempts are being made to resolve this difficulty through
communication with ASTM Committee E-37 (Thermal Methods) and by continued in-
possible to reuse a specimen for repeated calibrations? The shape of the scans
obtained with the five test materials sometimes varied with repeated use.
-24-
O **
O r-- cr> cm
>=i-
o LD O 00 LO MCOOOl
i— r— o 00 LO LO 0O
in (ji ro *
ld cm r-» lo i oo
oo lo oo CO LO oo lo CO LO OO P""- OO CNJ — i
LO LO 00
LO LO LO LO
Ol
LO
O LO OO
t-tLO
OO i~H
LO i—
O"*
C\J
—
o
i CM CM OO cn o
oo oo
LO lo oo
00 LO o O LO
I
<T> CM CM LO LO LO M LO
r-~.
r- oo iNJ cfi
oo r-i oo .—I r-l r-l O
LO OO LO LO
vf ro (\J lo
CM 00 LO o
LO LO 00 CXl
CM ^J- CM 0O
cm r- oo
Ol CM
OO OO
LO 0O
rH LO CM
CM 0O LO CO
0O LO OO
CT> LO 00 ^t"
CXt — 0O 0O
O LO
i l
^j-
CM LO LO —
o i
LO LO LO LO
LO LC LO LO
o-i lo oo cri
LO LC LO LO
O CT^ LO 0O
cm oo cn o 00 ^1" r— o
LO LO CM Ol
25-
Examples of original and subsequent fusion scans for the test materials are shown
in figures 2-6. Anisic acid was remarkable for stability. 2 Though the en-
thalpy of many of the substances does not differ significantly in repeat runs
from that of the initial runs the shapes of the curves were often distorted. The
automated program could not deal properly with temperature onset for these dis-
torted curves. This fault is not in the procedure itself but in the logic of the
commercial program used. In such instances, the onset temperature was satisfac-
when needed.
Table 16 summarizes the results obtained in the calibration study and gives
C. Summary
can be improved by more than an order of magnitude by use of the ASTM Recommended
Practice [7,8], The best results are obtained when like substances are used for
the calibrations and when the temperature difference between calibrants is about
curves are not distorted, the specimen may be used again; if the curves are mis-
2 Recentwork with a new lot of anisic acid has shown that the acid contains im-
purities which make it unsuitable for use as a standard. This fact reinforces
the need to use certified materials as calibrants.
-26-
NARWTHALE>C FIRST RU*
TEMPERATURE OO
TEMPERATURE <K)
-27-
Figure 3. Initial and Final Runs for Acetanil ide.
-28-
TEMPTRATUKE OO
TE>#TKATUKE OO
-29-
anisic acid n*rr run
—Sum
J
TLm 2a ** A. «T«
TEHPCTATURE OO
-30-
Table 16. Summary of transition temperatures and enthalpies
INITIAL RUN
Naphthalene 149.35 + 0.85
Acetani 1 ide 387.35 + 0.28 163.65 + 1.26
Di phenyl acetic Acid 420.22 + 0.16 148.67 + 1.32
Anisic Acid 456.18 + 0.41 195.49 + 1.11
2-Chl oroanthraqui none 148.15 + 0.76
ALL RUNS
Naphthal ene 149.01 + 1.18
Acetani 1 ide 387.38 + 0.37 163.40 + 1.343
Di phenyl acetic Acid 420.23 + 0.39 148.16 + 1.28
Anisic Acid 456.50 + 0.43 195.31 + 1.53
NOTE: Naphthalene and 2-chl oroanthraqui none have no temperatures listed since
they were the outer bracketing substances.
*The enthalpies of transition given are those that have been corrected by the
fusion correction only.
-32-
References
2. Hudson, C. H., editor. NBS Standard Reference Materials Catalog 1984-5, Nat.
Bur. Stand. (U.S.) Spec. Publ . 260; 1984 February. 174 pp.
11. NPL Certificate of measurement CRM No. M14-11. Set of ten melting point
-33-
12. Pel la, E.; Nebuloni, M. Temperature measurements with a differential calo-
indium; certification of a sample for use as a CRM. NPL Report Chem. 101;
1979 July.
-34-
Appendix
could arise from the DSC, from the procedures and laboratory conditions for
working with the DSC, or from i nhomogenei ty among specimens of the reference
material. This appendix describes results of a study that was designed to iden-
tify and quantify some of the sources of variation in the DSC procedure and the
five specimens from three forms of indium (granules, rod, foil). Each of the
should be noted that the latter protocol means that these data cannot provide
distinct estimates of remount variability and variation due to the instrument it-
dium. Evidence of inhomogeneity in melting points was found only for granular
i ndi urn.
A-1
Sample Data
The data and summary statistics from this study are displayed in table 1 and
wi thin-specimen) variability that were observed in the data for temperature and
i ndi urn.
A-2
Table 2. Enthalpy of Fusion of Indium
Within-
Specimen Specimen
Form Specimen Enthalpy , J/g Mean Std. Dev,
Statistical Analysis
x
ij «» + 6
i
+
V M
where
A-
4
Granular
430.2 r r
4292
2 3 4
Specimen
Figure la. Fusion temperature measurements for indium granules
( indicates specimen mean)
Rod
4302 I
430.0
g 429.8
!
& 429.6
I
429.4 h
4292
1 2 3 4 5
Specimen
Figure lb. Fusion temperature measurements for indium rod.
( indicates specimen mean)
A-
Fbil
4302 1 1 1
1
,
+
- + -
+ +
+ +
+
+
t
1 2 3 4 5
Specimen
Figure lc. Fusion temperature measurements for indium fc
( indicates specimen mean)
Granular
1 2 3 4 5
Specimen
Figure 2a. Enthalpy measurements for indium granules,
( indicates specimen mean)
Rod
29.50 i f i 1
2925 - -
+
+
+
29.00 -
+
+ + +
28.75 a -
+
28.50
+
on pc +
••
28.00
27.75 L_ I
1 2 3 4 5
Specimen
Figure 2b. Enthalpy measurements for indium rod.
( indicates specimen mean)
Foil
Speci
Figure 2c. Enthalpy measurements for indium foil,
( indicates specimen mean)
X.jj = measured temperature (enthalpy) for repeat run j on specimen i
instrument error).
For each fonn of indium, the index i runs from 1 to 5 = the number of specimens,
The inhomogenei ty , <5^ , and measurement error, ^j, are assumed to have
probability distributions that are approximately Gaussian with zero means and
standard deviations a. and a, respectively. Thus, for the indium data, the model
0
represents variability in the data arising in two stages. First, all four meas-
wi thin-specimen variance (a
2 comprises other relevant contributions to overall
)
(separately) on both temperature and enthalpy data for each form of indium. Es-
were obtained from the analysis are shown in table 3. Significance levels for
The significance levels that were attained show a strong indication of variation
significant variation for either rod or foil. However, all three forms of indium
variations may depend on the form of indium, a more general analysis (combining data
Temperature Enthal py
Granul ar (0.148) 2
(0.171) 2 .0034 (0.256) 2 (0.153) 2 .0937
2
Rod (0.168) 2 0 .5419 (0.270) (0.193) 2 .0502
variance. Statistical theory [1] gives the following formula for the true
standard error of the grand mean of 20 form measurements under model (1):
1/2
2
SE(X) = (
20
The usual formula for the standard error of the mean, based on the simple (un-
forms are given in table 4, along with standard errors that allow for
the data were used to compute uncertainties of rod and foil melting points (based
A-8
Table 4. Estimated Fusion Temperatures, Enthalpies, and Standard Errors
Mean
Temper- Standard Degrees of Mean Standard Degrees of
ature Error of Mean Freedom Enthal py Error of Mean Freedom
(K) (K) (J/g) (J/g)
Reference
A-9
NBS-114A (REV. 2-8C)
U.S. DEPT. OF COMM. 1. PUBLICATION OR 2. Performing Organ. Report No. 3. Publication Date
REPORT NO.
BIBLIOGRAPHIC DATA
SHEET (See instructions) NBS/SP-260/99 August 1985
4. title and subtitle standard Reference Materials:
Feasibility Study for the Development of Standards Using Differential Scanning
Calorimetry
5. AUTHOR(S)
J. E. Callanan, S. A. Sullivan, and D . F. Vecchia
6. PERFORMING ORGANIZATION (If joint or other than NBS, see instructions) 7. Contract/Grant No.
Same as item 6.
I |
Document describes a computer program; SF-185, FlPS Software Summary, is attached.
11. ABSTRACT (A 200-word or less factual summary of most significant information. If document includes a significant
bibliography or literature survey, mention it here)
12.KEY WORDS (Six to twelve entries; alphabetical order; capitalize only proper names; and separate key words by semicolons)
differential scanning calorimetry; enthalpy; fusion; pilot study; standards; tempera-
ture; thermal analysis
fj^] Order From National Technical Information Service (NTIS), Springfield, VA. 22161
USCOMM-DC 6O43-P80
Technical Publications
Periodical
Journal of Research —The Journal of Research of the National Bureau of Standards reports NBS research
ana development in those disciplines of the physical and engineering sciences in which the Bureau is active.
These include physics, chemistry, engineering, mathematics, and computer sciences. Papers cover a broad
range of subjects, with major emphasis on measurement methodology and the basic technology underlying
standardization. Also included from time to time are survey articles on topics closely related to the Bureau's
technical and scientific programs. Issued six times a year.
Nonperiodicals
Monographs— Major contributions to the technical literature on various subjects related to the Bureau's scien-
tific and technical activities.
—
Handbooks Recommended codes of engineering and industrial practice (including safety codes) developed in
cooperation with interested industries, professional organizations, and regulatory bodies.
Special Publications— Include proceedings of conferences sponsored by NBS, NBS annual reports, and other
special publications appropriate to this grouping such as wall charts, pocket cards, and bibliographies.
Applied Mathematics Series —Mathematical tables, manuals, and studies of special interest to physicists,
engineers, chemists, biologists, mathematicians, computer programmers, and others engaged in scientific and
technical work.
—
National Standard Reference Data Series Provides quantitative data on the physical and chemical properties
of materials, compiled from the world's literature and critically evaluated. Developed under a worldwide pro-
gram coordinated by NBS under the authority of the National Standard Data Act (Public Law 90-3%).
NOTE: The Journal of Physical and Chemical Reference Data (JPCRD) is published quarterly for NBS by
the American Chemical Society (ACS) and the American Institute of Physics (AIP). Subscriptions, reprints,
and supplements are available from ACS, 1155 Sixteenth St., NW, Washington, DC 20056.
Building Science Series —
Disseminates technical information developed at the Bureau on building materials,
components, systems, and whole structures. The series presents research results, test methods, and perfor-
mance criteria related to the structural and environmental functions and the durability and safety
characteristics of building elements and systems.
Technical Notes —
Studies or reports which are complete in themselves but restrictive in their treatment of a
subject. Analogous to monographs but not so comprehensive in scope or definitive in treatment of the subject
area.Often serve as a vehicle for final reports of work performed at NBS under the sponsorship of other
government agencies.
—
Voluntary Product Standards Developed under procedures published by the Department of Commerce in
Part 10, Title 15, of the Code of Federal Regulations. The standards establish nationally recognized re-
quirements for products, and provide all concerned interests with a basis for common understanding of the
characteristics of the products. NBS administers this program as a supplement to the activities of the private
sector standardizing organizations.
Official Business
Penalty for Private Use $300