Symbols
Symbol Definition
WARNING. HAZARDOUS VOLTAGE. Risk of electric
X shock.
T Double Insulated
Earth
1
°C/°F
MIN MAX RANGE AutoHOLD
3
Peak MIN MAX
REL Hz %
4½ DIGITS
1 Second
mV
mA
A
2
V
LO
V A
OFF
A mA A COM V
400mA
10A MAX FUSED
FUSED
Probe tip
Probe body
Multimeter or voltmeter
Plug
GND tab
Protective earth ground lead clip
Probe Tip
RA HI 10 MΩ
RB (Voltmeter)
LO
Protective
Earth Lead (Dual Banana Plug)
Resistor
Probe
RA RB
Voltmeter Compatibility
Accuracy of the meter is not included in the accuracy of the probe and
must be added to the probe accuracy to determine system accuracy.
The probe is mechanically compatible with any ac or dc voltmeter or
multimeter that accepts a standard double banana connector (spaced
0.75" [19 mm]) with standard plugs (0.160" [4 mm]).
The probe is electrically compatible with any ac or dc voltmeter or
multimeter that has an input impedance of:
• 10 MΩ ±1 % for 80K-40
• 10 MΩ ±10 % for 80K-6
Voltmeters or multimeters with other input impedances require the use of
an external shunt or a correction factor to obtain an accurate
measurement. Higher impedance voltmeters or multimeters should be
equipped with a shunt, and lower impedance voltmeters or multimeters
should be assigned correction factors.
Use this formula to determine the value of an external shunt resistor
(meter impedance >10 MΩ):
Rm x 10
Rs =
Rm − 10
Where: Rs = Shunt resistance in MΩ
Rm = Voltmeter input impedance in ΜΩ (>10 MΩ)
Example: If Rm = 20 MΩ,
20 × 10 200
Rs = ------------------- = ---------- = 20 MΩ
20 – 10 10
Use this formula to calculate a correction factor (meter impedance
<10 MΩ):
1.11 + Rm
Cf =
1.11 x Rm
Where: Cf = Correction factor (multiplier for meter reading)
Rm = Voltmeter input impedance in MΩ
Example: If Rm = 1 MΩ,
1.11 + 1 2.11
Cf = = = 1.901
1.11 x 1 1.11
A meter reading of 0.526 volts represents an input of:
0.526 x 1.901 = 1 or 1 kV
Circuit Loading
The probe represents a load to the circuit:
• 1000 MΩ , or 1 μA/1 kV, for 80K-40
• 75 MΩ, or 13 μA/1 kV, for 80K-6
Maintenance
The probe has no user-serviceable parts.
Use a soft cloth dampened with distilled water to clean the probe. Never
use solvents of abrasive cleaners. Make sure the probes are dry before
use.
Performance Test
Verify the accuracy of the probe with a voltage source measurement:
• 25 kV dc ±0.25 % for 80K-40
• 5 kV dc ±0.25 % for 80K-6
When used with a compatible dc voltmeter, the probe should measure the
source with ±1 % accuracy. No calibration adjustments are available on
the probe.
Specifications
Voltage Range
80K-40: 0 kV to ±40 kV dc or peak ac, 28 kV rms ac
80K-6: 0 kV to ±6 kV dc or peak ac, 4.24 kV rms ac
Input Resistance
80K-40: 1000 MΩ
80K-6: 75 MΩ nominal
Division Ratio: 1000:1 (1000 x attenuation)
Accuracy dc
80K-40: 0 kV to 20 kV: 2 %
20 kV to 35 kV: ±1 % @ 20 °C to 30 °C, add
1 % @ 10 °C to <20 °C and >30 °C to 45 °C)
35 kV to 40 kV: 2 %
80K-6: 0 kV to 6 kV: 1 %
Accuracy ac:
80K-40: 5 % at 60 Hz
80K-6: 1 % dc to 500 Hz
2 % 500 Hz to 1 kHz
-30 % typ. @ 10 kHz
Safety: IEC 61010-031, Type B, Pollution Degree 2
Temperature
Operating: 0 °C to 50 °C
Storage: -20 °C to +60 °C
Relative Humidity: 80 % @ 30 °C
50 % @ 40 °C
Altitude
80K-40: 2000 m, 2000 m to 5000 m: derate linearly the
working voltage from 40 kV peak to 28 kV
peak, and derate linearly the transient
overvoltage from 60 kV peak to 44 kV peak
80K-6: 2000 m
Dimensions (L x W)
80K-40: 36 cm x 7.5 cm (14 in x 3 in)
80K-6: 24.9 cm x 5.10 cm (9.8 in x 2.0 in)
Weight
80K-40: 204 g (7.2 oz)
80K-6: 136 g (4.8 oz)