Sales in Each Country or Region Check the Operating Environment 1.

Specifications
User’s 91061 Authorized Representative in the EEA Model 91061: Hydraulic Pressure Pump
Manual Hydraulic Pressure Pump Yokogawa Europe B.V. is the authorized representative of WARNING Pressurized media Liquid
Yokogawa Test & Measurement Corporation for this product in Pressure generation range 0 to 70 MPa
the EEA. (EEA: European Economic Area) • This product is a pump for the pressure calibrator.
Do not use for any other purpose. M. W. P
To contact Yokogawa Europe B.V., see the separate list of 70 MPa
• Do not use the product if there is a problem with its physical (Maximum working pressure)
worldwide contacts, PIM 113-01Z2.
Thank you for purchasing the 91061 Hydraulic Pressure Pump. appearance.
This user’s manual primarily explains the handling precautions and Safety Precautions • This product is manufactured in accordance with Sound Operating temperature range: 0 to 50°C
basic operations of the 91061. Engineering Practice (SEP). Port: 1/4” FNPT, 1/4” FNPT
This product is designed to be used by a person with specialized
To ensure correct use, please read this manual thoroughly before • Be sure to check the measurement environment and conditions. Weight: Approx. 700 g
beginning operation. knowledge.
If you are using the product in an environment in which Dimensions: Approx. 258×153×62 mm
The general safety precautions described herein must be observed
supervision by a certified person is required by law or other
during all phases of operation. If the instrument is used in
Store this manual in an easily accessible ordinance, be sure to perform measurements in accordance
a manner not specified in this manual, the protection provided by with appropriate safety management standards. Name Model Note
place for quick reference. the instrument may be impaired. • Use tubing (connector, hose, etc.) that can withstand the Hydraulic pressure pump kit 91060 Hand pump (91061),
This manual is an essential part of the product; keep it a safe place pressure to be generated. Connector set (91062),
for future reference. • Make sure that air does not leak from the tubing Case (93053)
3rd Edition: October 2017 YOKOGAWA assumes no liability for the customer’s failure to (connector, hose, etc.) connections.
All Rights Reserved, Copyright © Hydraulic pressure pump 91061 0 to 70 MPa
2013, Yokogawa Meters & Instruments Corporation, comply with these requirements. If a connection comes loose or if air leaks under high pressure,
2017, Yokogawa Test & Measurement Corporation
The following symbols are used on this instrument and it can endanger the user or the surrounding instruments.
Printed in Japan Connector set 91062 Connector set for the 91061
in this manual. • Do not apply pressure exceeding the maximum allowable
(hydraulic pressure pump) (High-pressure hose *1,
IM 91061-01EN pressure of the device under test (DUT). Quick adapter *1,
3rd Edition: Oct. 2017 (YMI) Handle with Care. Do not generate pressure exceeding the maximum working Conversion adapter *1,
Refer to the user’s manual or service manual. pressure (M.W.P). Sealing tape, and Hex wrench)
This symbol appears on dangerous locations on • Removing tubing when there is remaining pressure is extremely
Case 93053 Case for the 91061 and 91062
the instrument which require special instructions for dangerous. (for 91056 and 91061) (same as the case for
proper handling or use. Safely release the pressure before removing tubing the pneumatic pressure pumps
(connector, hose, etc.) from the pressure pump. 91056 and 91057)
WARNING • This instrument is not explosion-proof.
Do not use the instrument in the presence of flammable gases
Calls attention to actions or conditions that could cause serious *1: The maximum working pressure (M.W.P) of
or vapors. Doing so is extremely dangerous.
or fatal injury to the user, and precautions that can be taken to the high-pressure hose, quick adapter, and
prevent such occurrences. conversion adapter is 70 MPa.

Do use the product in:


CAUTION • In direct sunlight or near heat sources Note
Contact information of Yokogawa offices worldwide is • In an environment that is subject to large levels of mechanical
provided on the following sheet. Calls attention to actions or conditions that could cause light injury If you require high airtightness or high withstand pressure,
vibration
to the user, or cause damage to the instrument or user’s data, • Near noise sources, such as high-voltage equipment or use a connector with a ferrule or sleeve, not the supplied
PIM 113-01Z2: Inquiries List of worldwide contacts and precautions that can be taken to prevent such occurrences. power source the high-pressure hose or quick adapter.
• Near strong magnetic field sources Use a hose that can withstand the pressure that you will generate.
Note • In an environment with excessive amounts of soot, grease,
dust, or corrosive gas
Calls attention to information that is important for the proper • In an environment where ignition or explosion may occur,
operation of the instrument. such as where flammable gas is present

Note
Be Sure to Read the Pressure Meter User’s Manual
After use, store the pressure pump in the hand pump case or
If you are going to use this product in combination with
in a safe place.
a pressure meter (standard pressure meter/monitor),
be sure to read the pressure meter user’s manual,
Notice Regarding This Manual and use the instruments correctly.
• The contents of this manual are subject to change without prior
notice as a result of continuing improvements to the instrument's
performance and functionality.
The figures given in this manual may differ from the actual
indication.
• Every effect has been made in the preparation of this manual to
ensure the accuracy of its contents. However, should you have
any questions or find any errors, please contact your nearest
YOKOGAWA dealer.
• Copying or reproducing all or any part of the contents of
this manual without the permission of YOKOGAWA is
strictly prohibited.
2. Components 3. Operating Instructions
CAUTION Note
Name Description 3.1 Handling Precautions Priming may require you to squeeze the handle numerous times
Be sure to check the operating environment (safety). • For the liquid (pressure medium), use distilled water or clean oil
1. Media reservoir Fill the reservoir with media depending on the connected load capacity (media capacity).
(up to hydraulic oil 100 (30W)).
(pressure media such as distilled water). Before use, be sure to check that connectors are secure, If the volume is large, prime many times.
Select a media type that is suitable for the device under test
that there are no intrusion of foreign substances,
2. Fill port A port for filling (draining) media. (DUT) and the pressure pump.
and that there are no damages.
It has a cap. • The reservoir seal is made of fluorocarbon and nonmetallic 3.4 Releasing Pressure
In addition, use the same thread specifications for all connectors.
plastic. Do not use highly viscous oil (higher than 30W).
3. Fill line Do not fill above this line. Perform preliminary pressure tests on tubing and other parts of
(MAXIMUM FILL LEVEL) • Hydrocarbon fluids should never be used in oxygen service WARNING
the system. test device.
4. Vernier adjustment Used to “fine tune” the pressure to the If you are applying pressure using a hand pump, • High pressure gases is extremely dangerous. To prevent accidents, safely return the pressure to zero
target value. be sure to monitor the pressure with a standard pressure meter (release the pressure) before removing the connector and hose.
“Prime” the test system to minimize air introduced into the DUT.
5. Needle bleed valve Close to generate pressure. (monitor). • Use connectors and hoses appropriate for the source pressure
Open to release pressure. value. After generating pressure, carefully open the needle valve to
6. Handle Operate (squeeze) the handle to 3.2 Connecting Connectors release the pressure. (The media will turn to the reservoir.)
generate pressure. Make sure that the media does not overflow from the reservoir.
Note
CAUTION When you are finished, drain the media from the reservoir.
7. Stroke limiter Controls the handle stroke.
We recommend that the user keep a record of the kind of media
Push (release) the limiter while
• Use connectors appropriate for the thread specifications of being used (was used previously). (records of use, cleaning, etc.)
squeezing the handle. Note
Prime mode:
the device under test (DUT) and tubing.
Allows the maximum amount of Otherwise, leakage and damage to threads may result. After use, store the pressure pump in the hand pump case or
handle movement • Apply appropriate seal tape to the connecting surface of [ Procedure ] in a safe place.
Pump mode: the connector. (1) Connect the standard pressure meter. We recommend that you clean the reservoir before storing it.
Limits the amount of handle movement Open the hose that will be connected to the DUT (for “priming”).
8. Pressure output port Port for outputting pressure. Standard pressure meter (2) Remove the cap from the fill port.
(1/4” NPT) Connect a connector. Fill the media reservoir 1/2 (50%) to 2/3 full, and close the cap.
(for connecting the standard pressure • Be sure to fill the reservoir at least half way full to prevent 4. Cleaning
meter and the calibration target) 1/4” NPT intrusion of air.
After use (calibration), clean the media reservoir.
• Do not exceed the fill line (maximum fill level).
2. Fill port (with cap) Clean it in way that is appropriate for the liquid (pressure medium)
Hose 1/4” NPT • Minimize load capacity.
(DUT) Connector that you used.
(3) Adjust the vernier adjustment to the mid position.
MAXIMUM FILL LEVEL
(4) Close the needle bleed valve.
VERNIER
(5) While squeezing the handle, push the stroke meter into 5. Troubleshooting
8. Pressure output port prime mode.
(1/4” NPT) <Top of the media reservoir> BLEED

Prime mode: Mode that allows the maximum amount of If you cannot increase the pressure with the hand pump
handle movement (if the pressure decreases), check the following.
Connector
1. Media reservoir (6) Open the hose that will be connected to the DUT,
4. Vernier adjustment MAXIMUM FILL LEVEL
and squeeze the handle to beginning the priming process. • Check that the media reservoir cap is closed.
Priming: The act of squeezing the handle until media • Check that the needle valve is closed.
VERNIER

BLEED fills the hose all the way to the end to prevent • Check the thread specifications of the connection port.
5. Needle bleed
valve 3. Fill line the intrusion of air Check that the connection is secure (use seal tape or other
(maximum fill level)
(Using full stroke on the handle removes air in the pump and hose.) measures).
7. Stroke limiter • Check that the stroke limiter is being used adequately.
(7) After checking that the media has filled the hose all the way
6. Handle to the end, connect the hose to the DUT.
(8) Apply pressure (squeeze the handle) just enough so that
you can verify a sharp rise in pressure on the standard
CAUTION
<Connection example>
pressure meter. If you cannot determine the cause, continuing with the procedure
■ Using Two Wrenches (You can feel the pressure changing through the handle that
is dangerous. Stop immediately.
you are squeezing.)
To prevent leakage, you must tighten the connector or If the product does not work properly, contact your nearest
CAUTION While squeezing the handle, push the stroke limiter
quick adapter firmly. YOKOGAWA dealer.
to switch to pump mode.
Using a wrench only on the connector side may break
Do not remove the side sealing cap. Pump mode: Mode in which the amount of
the pressure pump.
handle movement is limited
Use two wrenches to prevent force from being applied to
Sealing cap
If you do not apply the stroke limiter at the appropriate timing,
the pressure pump. you may not be able to generate pressure up to
Do not remove
(The same applies when you connect a connector to the specifications.
the standard pressure meter.) After applying sudden pressure, if you remove your hand from
3.3 Generating Pressure the handle before applying the stroke limiter, it will be difficult
to operate the handle the next time.
WARNING If this happens, loosen the needle bleed valve to release some
<Side view > of the pressure, close the needle bleed valve,
• Remove tubing and wiring from the calibration target before and apply pressure (operate the handle) again.
carrying out the procedure. (9) Check that the needle valve is closed.
• Only connect to this pressure pump. (10) Squeeze the handle to apply pressure near the target
Do not connect to another high pressure source. pressure.
• Be sure to observe all safety precautions to avoid injury and (11) Adjust the vernier adjustment to “fine tune” the pressure to
damage to the device under test (DUT). the targetvalue.

IM 91061-01EN (P2)