You are on page 1of 2

07.08.

2022
CALIBRATION REPORT FOR COMPRESSION TESTING MACHINE

Details of Machine Calibrated Details of Standard Proving Ring Used

Make : SK Certificate No. : CL -2372


Model : ANALOG Certificate Date : 16.05.2019
Certified By. : NCCBM
Capacity : 1000Kn Date of Calibration : 19.01.2021
Used for : C.T.M Expiry Date : 18.01.2023
S.NO :06/30/22 Proving Ring Capacity : 2000 KN
Date of
Calibration : 06.08.2022 Proving Ring No. : 2000KN.207
Due for
Calibration : 05.07.2023 Dial Sl.No. & Make: : Mercer India No.87467
Environment Temp. : 28 Deg.
Client : SMD FLYASH BRICKS CHENNIMALAI.
Note: Error allowed is +/- 2% as per BIS 516.
This Certificate is valid for 12 Months from the date of Calibration.

Applied Proving Ring Actual Load Indicated In Average % Of

Load Deflection Load Gauge (KN) Readings Error

IN KN (Divisions) I II III In KN  

200 78.2 203 203 203 203 1.50

400 154.7 405 405 405 405 1.25

600 229.2 609 609 609 609 1.50

800 303.2 801 801 801 801 0.125

CALIBRATED BY: CERTIFIED BY: