You are on page 1of 6

Downloaded from http://fn.bmj.com/ on December 30, 2016 - Published by group.bmj.

com
ADC-FNN Online First, published on September 19, 2016 as 10.1136/archdischild-2016-311164
Original article

Neopuff T-piece resuscitator: does device design


affect delivered ventilation?
Murray Hinder,1,2 Pranav Jani,1,3 Archana Priyadarshi,1,3 Alistair McEwan,2
Mark Tracy1,3
1
Neonatal Intensive Care, ABSTRACT
Westmead Hospital, Background The T-piece resuscitator (TPR) is in What is already known on this topic?
Westmead, New South Wales,
Australia common use worldwide to deliver positive pressure
2
Faculty of Engineering and ventilation during resuscitation of infants <10 kg. Ease
▸ Neopuff T-piece resuscitator performance has
Information Technologies, of use, ability to provide positive end-expiratory pressure
BMET Institute, Sydney been reported in preterm infants and low
(PEEP), availability of devices inbuilt into resuscitaires
University, Sydney, New South compliance lung models.
and cheaper disposable options have increased its
Wales, Australia ▸ Use of Neopuff in resuscitation of term infants
3
Department of Paediatrics and popularity as a first-line device for term infant
with normal lung compliance has increased.
Child Health, Sydney resuscitation. Research into its ventilation performance is
University, Westmead, New limited to preterm infant and animal studies. Efficacy of
South Wales, Australia providing PEEP and the use of TPR during term infant
resuscitation are not established.
Correspondence to What this study adds?
Dr Mark Tracy, Department of Aim The aim of this study is to determine if delivered
Paediatrics and Child Health, ventilation with the Neopuff brand TPR varied with
Sydney University, PO Box 533, differing ( preterm to term) test lung compliances (Crs) ▸ When using Neopuff T-piece resuscitator in a
Wentworthville, NSW 2145, lung model, as lung compliance increases, the
and set peak inspiratory pressures (PIP).
Australia; mark.tracy@sydney.
edu.au Design A single operator experienced in newborn delivered positive end-expiratory pressure
resuscitation provided positive pressure ventilation in a (PEEP) may increase from the set PEEP value
Received 5 May 2016 randomised sequence to three different Crs models (0.5, and peak inspiratory pressures may be less
Revised 18 August 2016 1 and 3 mL/cmH2O) at three different set PIP (20, 30 than intended.
Accepted 23 August 2016
and 40 cmH2O). Set PEEP (5 cmH2O), gas flow rate and ▸ Clinicians using T-piece devices to resuscitate
inflation rate were the same for each sequence. term infants should be aware of an unintended
Results A total of 1087 inflations were analysed. The rise in PEEP that may be mitigated by reducing
delivered mean PEEP was Crs dependent across set PIP inflation rate.
range, rising from 4.9 to 8.2 cmH2O. At set PIP
40 cmH2O and Crs 3 mL/cmH2O, the delivered mean PIP
was significantly lower at 35.3 cmH2O.
Conclusions As Crs increases, the Neopuff TPR can located at the distal end of the patient circuit. The
produce clinically significant levels of auto-PEEP and thus adjustable PEEP pressure range is 1–25 cmH2O.
may not be optimal for the resuscitation of term infants Setting delivery pressures must be carried out in
with healthy lungs. the correct sequence stated in manufacturer docu-
mentation; otherwise, delivery of adequate ventila-
tion will not be successful.1 2 Any alteration to the
gas inflow rate after the initial setting and start of
IPPV requires recalibration by the user, which
BACKGROUND requires disconnection from the patient.2 3 Hawkes
The T-piece resuscitator (TPR) is a device widely et al4 showed that adjustment of circuit gas inflow
used to provide positive pressure ventilation during from 5 to 15 LPM without recalibrating the deliv-
resuscitation of infants ≤10 kg.1 The device ery pressures resulted in an inadvertent increase in
requires an interface to the patient via face mask, PEEP (300%), PIP (40%) and safety over pressure
laryngeal mask airway or endotracheal tube (ETT). (33%) from the initial set value.
The Neopuff (Fisher & Paykel New Zealand) TPR IPPV is started by occluding the gas outlet located
is a gas flow-dependent resuscitator consisting of on the PEEP valve assembly. The operator deter-
three operator-adjusted valves. These are set with mines inflation rate and inspiratory:expiratory ratio
the aid of a built-in manometer to the desired level (I:E ratio). If IPPV is not initiated and circuit seal is
for peak inspiratory pressure (PIP), positive maintained, a continuous positive airway pressure is
end-expiratory pressure (PEEP) and safety pressure delivered to the patient at the same set PEEP value.
limit required to provide intermittent positive pres- The use of the Neopuff for preterm resuscitation
sure ventilation (IPPV). Driving gas of 5–15 L per is well published.5–7 Although it is widely used for
To cite: Hinder M, Jani P, minute (LPM) is provided by blended gas, air or resuscitation in term infants,8 limited data support
Priyadarshi A, et al. Arch Dis
Child Fetal Neonatal Ed
oxygen flowmeter. its use in infants with weight 3.5–10 kg. We tested
Published Online First: Two control dials located on the Neopuff the hypothesis that the delivered ventilation of the
[please include Day Month provide adjustable ranges for PIP of 2–75 cmH2O Neopuff TPR is not different between low and
Year] doi:10.1136/ and safety pressure limit of 40–80 cmH2O. PEEP is normal newborn lung compliance (Crs) (0.5–
archdischild-2016-311164 adjusted by changing a variable orifice flow resistor 3.0 mL/cmH2O).9
Hinder M, et al. Arch Dis Child Fetal Neonatal Ed 2016;0:F1–F5. doi:10.1136/archdischild-2016-311164 F1
Copyright Article author (or their employer) 2016. Produced by BMJ Publishing Group Ltd (& RCPCH) under licence.
Downloaded from http://fn.bmj.com/ on December 30, 2016 - Published by group.bmj.com

Original article

The aim of this study was to determine whether the delivered dwell of <2 mm above PEEP valve orifice was observed to
ventilation with the Neopuff TPR varied with differing test Crs impede gas output flow and increase PEEP during passive defla-
(0.5, 1 and 3 mL/cmH2O) and different set PIPs (20, 30 and tion of the test lung. Thus, the operator was instructed to
40 cmH2O). ensure finger distance from valve orifice during deflation was
>2 mm. PEEP valve set position was marked and checked after
METHODS AND DESIGN each sequence to ensure no operator induced change occurred
A single Neopuff TPR (Part Number: RD900AEU) and delivery in adjusted valve position during IPPV.11 12
circuit (Part Number RD 1300-10) with a measured compliance
of 0.4 mL/cmH2O and tubing flow resistance of 6 cmH2O/L/s at DATA ANALYSIS
30 LPM were used in this bench study. Analysis was conducted using Stata (V.13 MP, StataCorp,
Two different leak-free test lungs were used (1) a 50 mL College Station, Texas, USA). The measured parameters
Draeger test lung (Draeger, Lubeck, Germany) with measured included the mean, minimum and maximum PIP, PEEP and tidal
compliance of 0.5 mL/cmH2O; resistance 50 cmH2O/L/s and (2) volume (Vt). Analysis of variance (ANOVA) for repeated mea-
a 200 mL IMT newborn test lung (Smart Lung Infant, IMT sures was used to determine differences between test Crs at dif-
medical, Buchs, Switzerland) with adjustable measured com- ferent set PIP levels. Differences between means determined by
pliances of 1.0 and 3.0 mL/cmH2O; resistance 50 cmH2O/L/s. A ANOVA were reported with p values adjusted F test using Box’s
Florian respiratory function monitor (RFM) (Accutronics, conservative epsilon, p values of <0.05 were considered signifi-
Medical Systems AG, Zug, Switzerland) was connected via the cant. Table 1 provides measured mean PIP, PEEP and Vt with
hot wire pneumotach and pressure sensor line sited between the IQR and SD for each Crs and set PIP with p values calculated
Neopuff TPR and the test lung. The Florian monitor was cali- with ANOVA for repeated measures.
brated with an external syringe of known volume and pressure/
flow via a traceable reference ventilator analyser (PF300, IMT RESULTS
Medical, Buchs, Switzerland). The analogue signals output from A total of 1087 inflations were analysed. Inspiratory times were
the RFM were collected and digitised at 200 Hz with analysis statistically different across all sequences (mean=0.53,
software (Grove Medical, London, UK). The test lungs and SD=0.04 s p≤0.001) but not considered clinically significant.
monitoring system were pressurised to static pressure of The measured PIP as a percentage of set PIP ranged from
50 cmH2O and over 120 s, there was no fall in pressure indicat- 100% to 101% with Crs of 0.5 and 1 mL/cmH2O, which were
ing the system was leak free. not significantly different and was lowest at 88% with set PIP of
A single operator experienced in neonatal resuscitation was 40 cmH2O and Crs 3 mL/cmH2O ( p<0.001) (table 1, figure 1).
asked to deliver 2 min of IPPV in a randomised sequence to The measured PEEP as a percentage of set PEEP ranged from
each of the Crs models (0.5, 1, and 3 mL/H2O) at the prede- 98% to 106% for Crs of 0.5 mL/cmH2O, 106% to 110% for
fined PIP levels (20, 30 and 40 cmH2O). Cl of 1 mL/cmH2O and 122% to 164% for Crs of 3 mL/
The RFM pneumotach was rezeroed and the TPR settings for cmH2O across set PIP range (p<0.001), this was highest at set
circuit gas inflow 10 LPM air, PEEP 5 cmH2O and over pressure PIP of 40 cmH2O and Crs of 3 mL/cmH2O (table 1, figure 2).
50 cmH2O were reset and checked with ventilation analyser at The mean delivered Vt increased significantly with increasing
the start of each randomised sequence. An inflation rate of 60 Crs for each set PIP level, ranging from 8.5 mL (Cl 0.5 mL/
inflations per minute (IPM) guided by metronome was used cmH2O at set PIP 20 cmH2O) to 66 mL (Crs 3 mL/cmH2O at
across all Crs and inflation pressure combinations. Operator was set PIP 40 cmH2O) ( p<0.001) (table 1, figure 3).
blinded to RFM waveform display, only Neopuff manometer
was visible to operator. We have previously shown no difference DISCUSSION
in delivered mask ventilation with operators guided by either The results of this bench study show when using Neopuff TPR
the Neopuff pressure dial or the manikin chest rise.10 Finger to provide IPPV, there is a significant difference between set and

Table 1 Measured respiratory parameters with differing set peak inspiratory pressures (PIP) and test lung compliance
Test lung compliance

0.5 mL/H2O 1 mL/H2O 3 mL/H2O

Set pressure, cmH2O Mean IQR (SD) Mean IQR (SD) Mean IQR (SD) p Value

Measured PIP, cmH2O


20 20.1 0.06 (0.10) 20.3 0.09 (0.07) 20.0 0.06 (0.05) NS
30 30.0 0.06 (0.38) 30.2 0.09 (0.06) 29.3 0.21 (0.19) NS
40 40.4 0.06 (0.05) 40.3 0.09 (0.08) 35.3 0.99 (0.86) <0.001
Measured PEEP, cmH2O
20 4.9 0.03 (0.05) 5.3 0.06 (0.04) 6.1 0.18 (0.17) <0.001
30 5.1 0.06 (0.07) 5.4 0.09 (0.07) 6.9 0.37 (0.22) <0.001
40 5.3 0.12 (0.10) 5.5 0.12 (0.09) 8.2 0.77 (0.65) <0.001
Measured Vt, mL
20 8.5 0.21 (0.51) 10.1 0.22 (0.17) 25.6 0.45 (0.35) <0.001
30 12.7 0.14 (0.35) 15.8 0.18 (0.14) 46.4 0.83 (0.59) <0.001
40 16.3 0.14 (0.19) 21.8 0.21 (0.17) 66.6 3.70 (3.27) <0.001
PEEP, positive end-expiratory pressure; Vt, tidal volume.

F2 Hinder M, et al. Arch Dis Child Fetal Neonatal Ed 2016;0:F1–F5. doi:10.1136/archdischild-2016-311164


Downloaded from http://fn.bmj.com/ on December 30, 2016 - Published by group.bmj.com

Original article

Figure 1 Measured peak inspiratory pressure (PIP). Figure 3 Measured inflation volume.

time for gas to exit the lung.15 Imposed resistance due to venti-
lator device design has been shown to increase patient work of
breathing16–18 and change the overall respiratory system time
constant.18 19 Wald et al17 showed that the expiratory resistance
of the Neopuff TPR varied with gas in flow rate from
40.1 cmH2O/L/s at 15 LPM to 104.8 cmH2O/L/s at 6 LPM.
Finer et al detailed eight occurrences in a 12-month period
(n=120) of inadvertent increase in PEEP using TPR device
from set 5 cmH2O (min 6.7; max 15.8 cmH2O), IPM <60
during resuscitation of infants <1000 g which he attributed to
possible movement of the PEEP control knob. It is notable that
both the recordings of actual resuscitations illustrated have a
starting PIP of 40 cmH2O. Circuit gas inflow rate was not
reported, Finer concluded ‘the Neopuff has the potential to
cause an inadvertent and potentially toxic increase of PEEP
which might not be noticed by the operator’.11 We have
excluded PEEP valve positional changes contributing to rise in
Figure 2 Measured positive end-expiratory pressure (PEEP). PEEP in our study by checking there was no change in a marker
point on the PEEP valve during the experiment. Bennett et al
delivered pressures as Crs increases. The unintended rise in examined increasing PIPs with the Neopuff TPR in a manikin
delivered PEEP (auto-PEEP) at compliance of 3.0 mL/cmH2O model to 40 cmH2O as escalation to these levels may be
was clinically important. required with diseased or immature lungs.20 21
Auto-PEEP can impair the reduction in pulmonary vascular Limitations of our study are shared with other manikin and
resistance important to circulatory adaptation immediately after test lung studies of the ability to generalise to actual human
birth13 and increase risk of air leaks in term infants with meco- resuscitations at birth. Changes in inflation rate and circuit gas
nium aspiration syndrome. We suspect the auto-PEEP observed flow rate were not examined in this study and may contribute to
in our study is due to circuit-imposed expiratory resistance of the level of device imposed auto-PEEP. The performance of
the Neopuff PEEP valve and delivery circuit compliance, other brands of TPR’s may be different.
increasing the system time constant, thus increasing lung defla- Preliminary data by our group in a piglet study22 (n=10)
tion time. Similarly, the observed inability to attain set PIP of comparing self-inflating bag (SIB) and Neopuff TPR delivered
40 cmH2O at Crs of 3 mL/cmH2O during inspiration may be ventilation has confirmed our bench test results that the TPR
due to insufficient fill time at the circuit inflow rate of 10 LPM device can contribute to the production of clinically significant
and inflation rate of 60 IPM (figure 4). The compliance setting auto PEEP during IPPV compared SIB with PEEP valve.
that we chose for our term lung model of 3 mL/cmH2O may be Adaptive changes in pulmonary and circulatory physiology
typical of a term infant of 3500 g birth weight,9 a recent human and establishment of a functional residual capacity of the lung
study by McEvoy et al14 suggests higher term infant values of during birth are complex. Detecting and adjusting for Crs
4–5 mL/cmH2O. changes during resuscitation with either SIB or TPR devices is
The development of auto-PEEP is tied to the time constant of difficult.23–25 The presence of leak during mask resuscitation is
the lung and airway. The lung time constant is proportional to common and can also influence Neopuff performance.26 Mask
the product of airway resistance (RAW) and Crs. leak that may be variable might obscure TPR generated
Increasing RAW for a given Crs and volume results in a auto-PEEP by providing a path of least resistance for expired
longer time for gas to exit the lung. Similarly, for a given RAW gas. Mask ventilation by more experienced clinicians,27 using
and airways pressure increasing Crs will also result in longer improved mask techniques28–30 or the presence of ETT may

Hinder M, et al. Arch Dis Child Fetal Neonatal Ed 2016;0:F1–F5. doi:10.1136/archdischild-2016-311164 F3


Downloaded from http://fn.bmj.com/ on December 30, 2016 - Published by group.bmj.com

Original article

Figure 4 Pressure waveforms for lung compliance 3 mL/cmH2O at set peak inspiratory pressures 20 cmH2O (A), 30 cmH2O (B), 40 cmH2O (C) and
positive end-expiratory pressure 5 cmH2O.

provide a patient/device interface that is closer to leak free, 7 Finer NN, Rich W, Craft A, et al. Comparison of methods of bag and mask
increasing device imposed auto PEEP from TPR devices. ventilation for neonatal resuscitation. Resuscitation 2001;49:299–305.
8 Murthy V, Rao N, Fox GF, et al. Survey of UK newborn resuscitation practices. Arch
The efficacy of the T-piece device for term resuscitation is not Dis Child Fetal Neonatal Ed 2012;97:F154–F55.
established, and terms describing the device as the ‘gold- 9 Stocks J, Sly PD, Tepper RS, et al. Infant respiratory function testing. John Wiley and
standard’31 should be viewed with caution. Our data suggest Sons Inc, 1996.
that in contrast to previous studies using Neopuff TPR in 10 Tracy MB, Klimek J, Shingde V, et al. Neopuff T-piece mask resuscitator: is mask
leak related to watching the pressure dial? Acta Paediatr 2010;99:1314–18.
preterm lung models,32 delivered pressures are not consistent
11 Finer NN, Rich WD. Unintentional variation in positive end expiratory pressure
and vary from those preset as Crs increases. during resuscitation with a T-piece resuscitator. Resuscitation 2011;82:717–19.
12 Hawkes CP, Dempsey EM, Ryan CA. The Neopuff’s PEEP valve is flow sensitive.
CONCLUSION Acta Paediatr 2011;100:360–3.
We have shown in a test lung with compliance similar to that of 13 Polglase GR, Morley CJ, Crossley KJ, et al. Positive end-expiratory pressure
a term infant that use of Neopuff TPR may result in increasing differentially alters pulmonary hemodynamics and oxygenation in ventilated, very
premature lambs. J Appl Physiol (1985) 2005;99:1453–61.
auto PEEP and decreasing PIP values. This is likely to be greater 14 McEvoy C, Venigalla S, Schilling D, et al. Respiratory function in healthy
with higher Crs and higher inflation rates. Lower inflation rates late preterm infants delivered at 33–36 weeks of gestation. J Pediatr
may mitigate this effect. Operator finger position over PEEP 2013;162:464–9.
valve orifice during lung deflation may also contribute to unin- 15 Woda RP, Dzwonczyk R, Bernacki BL, et al. The ventilatory effects of auto-positive
end-expiratory pressure development during cardiopulmonary resuscitation. Crit Care
tended PEEP. Clinicians using Neopuff to resuscitate term
Med 1999;27:2212–17.
infants should be alert to these potential consequences. 16 Banner MJ, Downs JB, Kirby RR, et al. Effects of expiratory flow resistance on
inspiratory work of breathing. Chest 1988;93:795–9.
Contributors MH is the primary researcher responsible for conceiving, designing,
17 Wald M, Kribs A, Jeitler V, et al. Variety of expiratory resistance between different
data collection, statistical analysis and writing manuscript. PJ contributed to data
continuous positive airway pressure devices for preterm infants. Artif Organs
collection, interpretation, manuscript construction and review. AP and AM
2011;35:22–8.
contributed to interpretation, manuscript construction and review. MT contributed by
18 DiBlasi RM, Salyer JW, Zignego JC, et al. The impact of imposed expiratory
assisting design, statistical analysis, manuscript writing and review.
resistance in neonatal mechanical ventilation: a laboratory evaluation. Respir Care
Competing interests None declared. 2008;53:1450–60.
Ethics approval This study was approved by the Western Sydney Local Health 19 Goldsmith JP, Karotkin EH. Assisted ventilation of the neonate. 3rd edn. W.B.
District Human Ethics and Scientific committee approval number SAC2014/5/6.9 Saunders, 1996.
(3999)QA. 20 Bennett S, Finer NN, Rich W, et al. A comparison of three neonatal resuscitation
devices. Resuscitation 2005;67:113–18.
Provenance and peer review Not commissioned; externally peer reviewed. 21 Kattwinkel J. Textbook of neonatal resuscitation. 5th edn. American Heart
Association. American Academy of Pediatrics, 2006.
REFERENCES 22 ResearchGate.net. Changes in PEEP and airway resistance with the use of different
1 Healthcare FP. Neopuff infant T-piece resuscitator 900 series operating instructions devices in a swine model (PDF). Shah DM, Tracy MB, Hinder MK. (updated 9/9/
(PDF). Fisher & Paykel Healthcare NZ, 2016 (updated 5/5/2016. Ref 185041726 2016. Ref Conference: Pediatric Academic Societies, Volume: E-PAS2014:3843.539).
Rev I EN2011-10). https://www.fphcare.com.au/CMSPages/GetFile.aspx? https://www.researchgate.net/publication/282330122
guid=b7a71160-cdef-4858-83e9-2a859a8ef36e 23 Kattwinkel J, Stewart C, Walsh B, et al. Responding to compliance changes in a
2 Hawkes CP, Oni OA, Dempsey EM, et al. Should the Neopuff T-piece resuscitator be lung model during manual ventilation: perhaps volume, rather than pressure, should
restricted to frequent users? Acta Paediatr 2010;99:452–3. be displayed. Pediatrics 2009;123:e465–e70.
3 Schilleman K, Schmölzer GM, Kamlin OC, et al. Changing gas flow during neonatal 24 Hartung JC, Dold SK, Thio M, et al. Time to adjust to changes in ventilation
resuscitation: a manikin study. Resuscitation 2011;82:920–4. settings varies significantly between different T-piece resuscitators, self-inflating
4 Hawkes CP, Oni OA, Dempsey EM, et al. Potential hazard of the Neopuff T-piece bags, and manometer equipped self-inflating bags. Am J Perinatol
resuscitator in the absence of flow limitation. Arch Dis Child Fetal Neonatal Ed 2014;31:505–12.
2009;94:F461–3. 25 Boldingh AM, Solevåg AL, Benth JŠ, et al. Newborn manikin study shows that
5 Thio M, Dawson JA, Moss TJ, et al. Self-inflating bags versus T-piece resuscitator to physicians often fail to detect correct lung compliance when using a self-inflating
deliver sustained inflations in a preterm lamb model. Arch Dis Child Fetal Neonatal bag. Acta Paediatr 2016;105:172–7.
Ed 2014;99:F274–7. 26 Hartung JC, Te Pas AB, Fischer H, et al. Leak during manual neonatal ventilation
6 Szyld E, Aguilar A, Musante GA, et al. Comparison of devices for newborn and its effect on the delivered pressures and volumes: an in vitro study.
ventilation in the delivery room. J Pediatr 2014;165:234–9.e3. Neonatology 2012;102:190–5.

F4 Hinder M, et al. Arch Dis Child Fetal Neonatal Ed 2016;0:F1–F5. doi:10.1136/archdischild-2016-311164


Downloaded from http://fn.bmj.com/ on December 30, 2016 - Published by group.bmj.com

Original article
27 Schilleman K, Witlox RS, Lopriore E, et al. Leak and obstruction with mask 30 Joffe AM, Hetzel S, Liew EC. A two-handed jaw-thrust technique is superior to the
ventilation during simulated neonatal resuscitation. Arch Dis Child Fetal Neonatal Ed one-handed “EC-clamp” technique for mask ventilation in the apneic unconscious
2010;95:F398–402. person. Anesthesiology 2010;113:873–9.
28 Wood FE, Morley CJ, Dawson JA, et al. Improved techniques reduce face mask leak 31 Rafferty AR, Johnson L, Maxfield D, et al. The accuracy of delivery of target
during simulated neonatal resuscitation: study 2. Arch Dis Child Fetal Neonatal Ed pressures using self-inflating bag manometers in a benchtop study. Acta Paediatr
2008;93:F230–4. 2016;105:e247–51.
29 Tracy MB, Klimek J, Coughtrey H, et al. Mask leak in one-person mask ventilation 32 Hawkes CP, Ryan CA, Dempsey EM. Comparison of the T-piece resuscitator with
compared to two-person in newborn infant manikin study. Arch Dis Child Fetal other neonatal manual ventilation devices: a qualitative review. Resuscitation
Neonatal Ed 2011;96:F195–200. 2012;83:797–802.

Hinder M, et al. Arch Dis Child Fetal Neonatal Ed 2016;0:F1–F5. doi:10.1136/archdischild-2016-311164 F5


Downloaded from http://fn.bmj.com/ on December 30, 2016 - Published by group.bmj.com

Neopuff T-piece resuscitator: does device


design affect delivered ventilation?
Murray Hinder, Pranav Jani, Archana Priyadarshi, Alistair McEwan and
Mark Tracy

Arch Dis Child Fetal Neonatal Ed published online September 19, 2016

Updated information and services can be found at:


http://fn.bmj.com/content/early/2016/09/19/archdischild-2016-311164

These include:

References This article cites 27 articles, 9 of which you can access for free at:
http://fn.bmj.com/content/early/2016/09/19/archdischild-2016-311164
#BIBL
Email alerting Receive free email alerts when new articles cite this article. Sign up in the
service box at the top right corner of the online article.

Topic Articles on similar topics can be found in the following collections


Collections Resuscitation (145)
Child health (1515)
Infant health (857)
Neonatal health (928)

Notes

To request permissions go to:


http://group.bmj.com/group/rights-licensing/permissions

To order reprints go to:


http://journals.bmj.com/cgi/reprintform

To subscribe to BMJ go to:


http://group.bmj.com/subscribe/

You might also like