Program Operation Manual

0
King Design / King Solution / King Service KING DESIGN INDUSTRIAL CO., LTD.
Index
1. Computer Hardware Requirement ............................... 1
2. Hardware function description ................................... 2
3. Installation Guidance of GM4 Measurement Software............. 3
4. Measurement Software Functions ................................ 4
4.1 Booting Procedure ......................................... 4
4.2 Interpretation of Main Program Page: ....................... 5
4.3 Software Functions Interpretation:.......................... 6
4.3.1 Elapsed numbers .................................... 6
4.3.2 Test buttons area .................................... 6
4.3.3 Fuction buttons area ................................. 6
Cpk ( Optional ) ....................................... 20
4.3.4 Hotkey and Cursors function ......................... 24
4.3.5 Graphical display area ............................... 27
Time domain ......................................... 27
FFT ( Optional ) ....................................... 28
Shock Response Spectrum_Analysis ( Optional ) ..... 29
Shock Response Spectrum_Pyro ( Optional ) ......... 30
Battery ( Optional ) ................................... 31
4.4 Interpretation of measurement procedures .................. 32
4.5 Remaining function description ............................. 38
4.5.1 Screen shortcut function ............................. 38
4.5.2 Serial number query ................................. 41
4.5.3 Keyboard shortcuts .................................. 42
4.6 Simplified Chinese Setting Instructions ..................... 43
4.7 License .................................................... 45
4.8 Linkage Setting ( Optional ) ............................... 46
4.9 Calibration system ......................................... 47
4.10 About .................................................... 49
5. Troubleshooting ................................................. 50
6. Service offices ................................................... 53

i
King Design / King Solution / King Service KING DESIGN INDUSTRIAL CO., LTD.
1. Computer Hardware Requirement
CPU i5 or above or PC of compatible model.

Basic spec:

 Main memory should be 4GB or more.

 Hard drive with 100GB capacity or more.

 Windows 10 Operating System or above.

 If you want to produce word report, the Office should be activated


( Version 2010 or above ).

 PC / Laptop is equipped with USB 2.0 or above interface.

1
King Design / King Solution / King Service KING DESIGN INDUSTRIAL CO., LTD.
2. Hardware function description
Option 1

Power Switch Low - Pass Filter ( 25 k Hz )

Circuit indicator light


USB

Input Channel Input Power

Option 2

Circuit indicator light Low - Pass Filter ( 25 k Hz )


Power Switch

Input Channel

Input Power

USB

King Design / King Solution / King Service 2 KING DESIGN INDUSTRIAL CO., LTD.
3. Installation Guidance of GM4 Measurement
Software
 Install the driver provided by King Design, there are 2 files in total, they are
DAQ-18.1 and GM4-21.0.x.

 Select default value and click next step for all steps till the installation is
complete.

 Reboot the computer (PC).

 Go to C:\Program Files(x86)\GM4-x.x.x and change the permission to “Full


Control”: right click the folder GM4-x.x.x, and select “Content”, click
“Security”, select the User Name, click “Edit” and check “Full Control”, then
click OK to exit.

King Design / King Solution / King Service 3 KING DESIGN INDUSTRIAL CO., LTD.
4. Measurement Software Functions
4.1 Booting Procedure
1. Boot the PC.

2. Go to C:\Program Files \GM4-21.0.x\ folder to run GM4-21.0.x.exe or

there is a shortcut icon on the desktop.

3. Left-click twice to access the main program.

King Design / King Solution / King Service 4 KING DESIGN INDUSTRIAL CO., LTD.
4.2 Interpretation of Main Program Page:
The GM4 is fully PC-operated. Operator only needs to own basic PC operation
skills for using the software. The graphic and humanized operation design allows
for either keyboard or mouse input, easy for operation.

After accessing the GM4 main page, system is under the main function menu
waiting for operation. The operation mode in the main program is illustrated in
below:

( 1 ) Elapsed numbers

( 2 ) Test buttons area

( 3 ) Fuction buttons area

( 4 ) Data display area

( 5 ) Graphical display area

(4)
(1)

(5)
(2)

(3)

King Design / King Solution / King Service 5 KING DESIGN INDUSTRIAL CO., LTD.
4.3 Software Functions Interpretation:
4.3.1 Elapsed numbers
You can switch to the waveform in the different numbers via clicking the
left/right button below, and you can also type the number of the waveform you
want to check directly in the current numbers

4.3.2 Test buttons area


Click the Start ( F2 ) to test / Stop ( F3 ), there will be a reminder of “waiting for
triggering”.

4.3.3 Fuction buttons area


Linkage control ( Optional )
Linked with the machine, the height (mm) and release action can be set.

Setting
There are functions such as Save set, Function, Report, Properties, Calculator...
etc.

King Design / King Solution / King Service 6 KING DESIGN INDUSTRIAL CO., LTD.
Save set

Test information
You can write down the test information, which will be displayed in the original
data and Word file of the time domain spectrum.

Save content
You can choose automatic or manual for the save mode. For automatic mode, the
test will be recorded everytime it is triggered, for manual mode, you have to cick
the “save” button by yourself, the save path can be switched to other paths, you
can fill in the information of test in the test context, and the information will be
saved in the Raw Data.
To select the archive path, click on the pre-archived folder, and then click to
select the folder.

King Design / King Solution / King Service 7 KING DESIGN INDUSTRIAL CO., LTD.
Function

Channel
The Function includes Channel, Test Target, Filter, SRS_Analysis, SRS.

Test Target
Number of tests : The maximum total number of times is 100,000, and the
maximum time is 5000 ms. The time can be adjusted
manually/automatically. The automatic adjustment time is 6
times the time domain in the pulse parameter. The delay time
is defined as the position of the main wave. The default is
50%. The definition is in the middle.
Trigger : You can select any channel to trigger or select a channel. The trigger
mode includes percentage or acceleration. The percentage is based on
the acceleration in the pulse parameter. If you select a channel that is
not open, a reminder will pop up when you press OK.
Pulse Parameters : You can choose half-sine wave, trapezoidal wave, trailing edge
sawtooth wave, acceleration and time domain setting test
conditions.

King Design / King Solution / King Service 8 KING DESIGN INDUSTRIAL CO., LTD.
Filter
Free switch function, selectable Auto/Manual filtering, the frequency of
automatic filtering is calculated according to the pulse parameters and multiplied
by the set magnification.

King Design / King Solution / King Service 9 KING DESIGN INDUSTRIAL CO., LTD.
SRS_Analysis ( Optional )
Standard line : You can choose to open/close.
Octave : 1~128 can be selected.
Type : Composite, Primary, and Residual can be selected.
Analyzed Range : The minimum value depends on the time in the test setting, and
the maximum value is 20 k Hz.
Tolerance : Percentage/dB can be selected, the minimum is 1, the maximum is
100.
Q & ζ : Q range of 1 ~ 500, ζ range of 0.001 ~ 0.5.

King Design / King Solution / King Service 10 KING DESIGN INDUSTRIAL CO., LTD.
SRS_Pyro ( Optional )
It can be edited directly on the table. Hz has editing restrictions. The frequency of
the previous stroke cannot be greater than the frequency of the next stroke, and a
prompt will pop up.
Octave: 1 ~ 128 can be selected.
Insert : To insert the default row in the previous stroke.
Delete : To delete the anti-yellow column.
Save : In order to save the file into the set up folder, you can enter the file name
yourself, with the extension .srs.
Read : To open the .srs file in set up.
Type : Composite, Primary, and Residual can be selected.
Q & ζ : Q range of 1 ~ 500, ζ range of 0.001 ~ 0.5.

King Design / King Solution / King Service 11 KING DESIGN INDUSTRIAL CO., LTD.
Save screen

Read screen

King Design / King Solution / King Service 12 KING DESIGN INDUSTRIAL CO., LTD.
Report

Word output is divided into 3 modes, One, Increment, and Customize.


․One
Select the Frame you want to output, and then Export the information on this
screen.

․Increment
Start : Start output from this pen.
Increment : Add this value to output data until the total number of tests at this
stage.

King Design / King Solution / King Service 13 KING DESIGN INDUSTRIAL CO., LTD.
․Customize
Separate by commas, for example: 1,3,5-7, the data of 1, 3, 5, 6, 7 will be output.

King Design / King Solution / King Service 14 KING DESIGN INDUSTRIAL CO., LTD.
Properties

Time Domain_Scale
The X/Y axis can select the manual/automatic display range, the maximum
magnification is the magnification product of the maximum time domain
spectrum, and the minimum is the magnification product of the minimum time
domain spectrum.

Units
You can choose acceleration units-g, m/s^2, and speed change rate units-m/s,
in/s, ft/s.

King Design / King Solution / King Service 15 KING DESIGN INDUSTRIAL CO., LTD.
Plots
You can choose the background, grid, standard line, upper limit tolerance, lower
limit tolerance color and the thickness of each line.

FFT_Scale ( Optional )
X / Y axis-linear / logarithmic mode can be selected, manual / automatic display
range can be selected.

King Design / King Solution / King Service 16 KING DESIGN INDUSTRIAL CO., LTD.
SRS_Analysis_Scale ( Optional )
X / Y axis-linear / logarithmic mode can be selected, manual / automatic display
range can be selected.

SRS_Pyro_Scale ( Optional )
X / Y axis-linear / logarithmic mode can be selected, manual / automatic display
range can be selected.

King Design / King Solution / King Service 17 KING DESIGN INDUSTRIAL CO., LTD.
Calculator

Shock, take the following picture as an example, waveform mode: half-sine, click
the velocity change rate column, select the unit, enter acceleration and duration,
click the calculation button, and the velocity change rate will be calculated, this
value is just for your reference.

King Design / King Solution / King Service 18 KING DESIGN INDUSTRIAL CO., LTD.
File
Read data : You can read all the log files in the folder, or test the log files
individually to review the measurement.
Save data : Press “ Save data ” at any time to save the measurement data ( the
storage path is the archive information setting path).

King Design / King Solution / King Service 19 KING DESIGN INDUSTRIAL CO., LTD.
Cpk ( Optional )

Complex Process Capability index, there are two calculation methods.


Method 1 : Cpk = ( 1 - Ck ) * Cp
Ck = ( M – X ) / ( T / 2 )
Cp = T / ( 6 * σ )
M : Standard value.
X : Average of measured data.
T : Tolerance.
σ : Standard deviation.
Example : acceleration: 70.00 g, time domain: 2.00 ms.

Acceleration algorithm
Ck = ( 70 – 65.97909 ) / ( 14 / 2 ) = 0.574
Cp = 14 / ( 6 * 9.417636 ) = 0.248
Cpk = ( 1 – 0.574 ) * 0.248 = 0.105

Time domain algorithm


Ck = ( 2 – 1.6052 ) / ( 0.4 / 2 ) = 1.974
Cp = 0.4 / ( 6 * 1.116771 ) = 0.06
Cpk = ( 1 – 1.974 ) * 0.06 = 0.058
( If there is a negative value, the absolute value needs to be taken )

King Design / King Solution / King Service 20 KING DESIGN INDUSTRIAL CO., LTD.
King Design / King Solution / King Service 21 KING DESIGN INDUSTRIAL CO., LTD.
Method 2 : Cpk = min ( Cpl , Cpu )
Cpu : ( USL - X ) / ( 3 * σ )
Cpl : ( X - LSL ) / ( 3 * σ )
USL : Specification upper limit.
LSL : Specification lower limit.
X : Average of measured data.
σ : Standard deviation.
Example : acceleration: 70.00 g, time domain: 2.00 ms.

Acceleration algorithm
Cpu = ( 77.00 – 65.9790917 ) / ( 3 * 9.417636478 )
Cpl = ( 65.9790917 - 63.00 ) / ( 3 * 9.417636478 )
Cpk = min ( 0.105 , 0.39 )
Time domain algorithm
Cpu = ( 2.20 – 1.6052 ) / ( 3 * 1.116770607 )
Cpl = ( 1.6052 – 1.80 ) / ( 3 * 1.116770607 )
Cpk = min ( 0.058 , 0.178 )
( If there is a negative value, the absolute value needs to be taken )

King Design / King Solution / King Service 22 KING DESIGN INDUSTRIAL CO., LTD.
King Design / King Solution / King Service 23 KING DESIGN INDUSTRIAL CO., LTD.
4.3.4 Hotkey and Cursors function
Test Target
Cannot be adjusted during measurement.

Cursors

Respectively cursor, with Band cursor, mark, use the keyboard up and down keys
to change the displayed channel position.
Cursor : Display current curve name, X/Y axis value.

King Design / King Solution / King Service 24 KING DESIGN INDUSTRIAL CO., LTD.
Band : Calculate the difference between two cursor points and move it at will.

Mark: This function can only be used when the cursor is turned on. Press the
mark to display the current value independently, and multiple marks can be used.

King Design / King Solution / King Service 25 KING DESIGN INDUSTRIAL CO., LTD.
SRS_Analysis
Cannot be changed during measurement.

SRS_Pyro
Cannot be changed during measurement.

King Design / King Solution / King Service 26 KING DESIGN INDUSTRIAL CO., LTD.
4.3.5 Graphical display area
Time domain
Display acceleration, time domain, Velocity change rate, filtering, maximum value,
minimum value, status... and other information. Among them, the status is
judged to be within the upper and lower tolerance lines when there is an open
standard line, that is, In Band, otherwise Out band.

King Design / King Solution / King Service 27 KING DESIGN INDUSTRIAL CO., LTD.
FFT ( Optional )

Display the highest 3-point acceleration and correspond to its frequency.

King Design / King Solution / King Service 28 KING DESIGN INDUSTRIAL CO., LTD.
Shock Response Spectrum_Analysis ( Optional )

Display maximum g value, maximum frequency, minimum g value, minimum


frequency, Q, 1/n Octave, tolerance, status. When the status is that there is an
open standard line, it is judged to be within the upper and lower tolerance lines,
which is In Band, otherwise Out band.

King Design / King Solution / King Service 29 KING DESIGN INDUSTRIAL CO., LTD.
Shock Response Spectrum_Pyro ( Optional )

SRA : The percentage of the measured value higher than the standard value.
Upper limit : The percentage that exceeds the upper limit.
Lower limit : The percentage below the lower limit.

King Design / King Solution / King Service 30 KING DESIGN INDUSTRIAL CO., LTD.
Battery ( Optional )

Peak : Starting from the set percentage of the measured peak value, the g value at
the next 3 ms is displayed.

Average : Starting from the set percentage of the measured peak value, the
average g value from the beginning to the next 3 ms is displayed.

King Design / King Solution / King Service 31 KING DESIGN INDUSTRIAL CO., LTD.
4.4 Interpretation of measurement procedures
1. Turn on the power of the machine.
2. Boot the PC.
3. Connect the accelerometer with the machine without the interference with
other items, and also connect the accelerometer with the measurement system
GM4, the buttons in the right upper row are switch input buttons for the front
and rear panel, the buttons in the lower row are hardware filter, the following
picture are input in the front, you should press the button, after connecting
the accelerometer correctly, the indicator for accelerometer inspection on the
the top will be lit on.

Option 2

Option 1

4. After the PC is connected with GM4 via USB, the connection will be shown on
the PC graphic, this indicates that the connection is normal, and confirm that
the device is Dev1, you can enter NI MAX to modify the device position.

King Design / King Solution / King Service 32 KING DESIGN INDUSTRIAL CO., LTD.
5. Open the software

6. Click channel setting, check to open the corresponding channel where you
install the accelerometer, enter the sensitivity, select Auto Invert for biasing,
choose IEPE coupling method.

7. Click the test target and enter the total number of times you want to test. The
time can be first filled in 50 ms ( adjusted according to the test time domain ),
the trigger setting first selects any channel, and the trigger channel first
selects the percentage mode ( the magnitude of the trigger depends on the
requirement ). The pulse parameter is switched on and off according to the
demand, and the acceleration and time domain are filled in according to the
demand.

King Design / King Solution / King Service 33 KING DESIGN INDUSTRIAL CO., LTD.
8. Click filter, open it according to the needs, you can enter the value
automatically or munually.

King Design / King Solution / King Service 34 KING DESIGN INDUSTRIAL CO., LTD.
9. Click SRS_Analysis ( optional ), open the standard line as required, enter the
minimum/maximum frequency value, Octave value, Q, ζ ( Zeta ), and tolerance
range.

10. Click SRS_Pyro ( optional ), directly edit the standard line and the upper and
lower tolerance, you can add and delete, enter the Octave value, Q, ζ ( Zeta ).

11. Click confirm button to go back to main program page.

King Design / King Solution / King Service 35 KING DESIGN INDUSTRIAL CO., LTD.
12. Click save set button to fill in the information such as Test information, Save
content.

13.Click the Start button to wait for the singnal source

King Design / King Solution / King Service 36 KING DESIGN INDUSTRIAL CO., LTD.
14. When you click to exit the program, a prompt will pop up.

King Design / King Solution / King Service 37 KING DESIGN INDUSTRIAL CO., LTD.
4.5 Remaining function description
4.5.1 Screen shortcut function

Vernier : Press the cursor to move the Cursor line on the screen. When the
Cursor line is moved, the g value and ms value on the screen will be
displayed immediately (freely move).
Panning : Press to move the screen, you can move the whole screen freely.
Zoom in : Press the selection zoom button and then drag and select the area you
want to zoom in.
Unzoom : Pressing this key will immediately adjust the waveform to the full
screen (this function is also available when you double-click the left
mouse button on the screen).
Delete All Mark : Remove marked lines.

King Design / King Solution / King Service 38 KING DESIGN INDUSTRIAL CO., LTD.
Vernier : Press the cursor to move the Cursor line on the screen. When the
Cursor line is moved, the g value and Hz value on the screen will be
displayed immediately (freely move).
Panning : Press to move the screen, you can move the whole screen freely.
Zoom in : Press the selection zoom button and then drag and select the area you
want to zoom in.
Unzoom : Pressing this key will immediately adjust the waveform to the full
screen (this function is also available when you double-click the left
mouse button on the screen).
X-axis scale : linear and logarithmic.
Y-axis scale : linear and logarithmic.
Delete All Mark : Remove marked lines.

King Design / King Solution / King Service 39 KING DESIGN INDUSTRIAL CO., LTD.
The position of the red frame can be adjusted at will. Press the right mouse
button on the line twice to return to the original screen.

King Design / King Solution / King Service 40 KING DESIGN INDUSTRIAL CO., LTD.
4.5.2 Serial number query

1. Click NI MAX or look for NI MAX.


2. Select NI USB-6210, Serial Number.

3. Click the drop-down menu, Function  License, and check GM401 Serial
Number.

King Design / King Solution / King Service 41 KING DESIGN INDUSTRIAL CO., LTD.
4.5.3 Keyboard shortcuts
F2 : Start measurement.
F3 : Stop measurement.
Up : When the cursor is turned on, adjust the previous channel, which is only
valid for the open channel.
Down : When the cursor is turned on, adjust the next channel, which is only
valid for the open channel.
Left : Move the cursor one space to the left.
Right : Move the cursor one space to the right.

King Design / King Solution / King Service 42 KING DESIGN INDUSTRIAL CO., LTD.
4.6 Simplified Chinese Setting Instructions
1. Enter the GM4 folder, and open the SC-x.x text file ( first installation ).

2. Click Save As.

King Design / King Solution / King Service 43 KING DESIGN INDUSTRIAL CO., LTD.
3. Select ANSI as the encoding and directly overwrite the current file.

King Design / King Solution / King Service 44 KING DESIGN INDUSTRIAL CO., LTD.
4.7 License
Click Function  License.

1. Click Import to import the provided file ( .lsn ) , and the screen will display the
currently activated items.
2. Manually type in the Licensed.
There are 3 states: Licensed ( permanent ), Evaluation ( trial ), Unlicensed
( expired ).

King Design / King Solution / King Service 45 KING DESIGN INDUSTRIAL CO., LTD.
4.8 Linkage Setting ( Optional )
After the linkage control authorization code is entered correctly, close the
software and go to the device manager to check the COM location of the CP210x
USB. When the software is opened, the connection setting will pop up, and the
Compor setting is consistent with the COM of the device manager CP210x USB.
Baud Rate: 9600
Parity Bit: None
Data Bit: 8
Stop Bit: 1.0

When the main screen will display the linkage control options, you can change
the height, replace the rubber block, reset the platform... and other functions.
The total number of times in the parameter setting will be written during
execution.

King Design / King Solution / King Service 46 KING DESIGN INDUSTRIAL CO., LTD.
4.9 Calibration system
Click Function  Import calibration.

Fill in the Serial Number and License to enter the calibration system.

There are two types of calibration file display : Not expired and Expired.

King Design / King Solution / King Service 47 KING DESIGN INDUSTRIAL CO., LTD.
30 days before the expiration, a prompt message will pop up when the software is
opened.

A reminder message will pop up when it expires.

King Design / King Solution / King Service 48 KING DESIGN INDUSTRIAL CO., LTD.
4.10 About
Show software version and company.

King Design / King Solution / King Service 49 KING DESIGN INDUSTRIAL CO., LTD.
5. Troubleshooting
 File not found:

Solution:
Ensure that there are 5588C1378.bin 、 0043616C.crtn 、 C159AABB.bin 、
Comport.bin、Cpk.bin、Function.bin、GM4-EN-21.x.x.txt、GM4-SC-21.x.x.txt、
GM4-TC-21.x.x.txt 、 Hot key.bin 、 Language.inf 、 Properties.bin 、 Report
setting.bin、sample.doc、SRS.bin, 15 files in total in C:\Program Files \GM4-
x.x.x. If not, please go to D drive or run the installation disk provided by King
Design, run setup.exe in the folder Volume of the folder GM4 installer.

 You do not have the correct permissions for the file:

Solution:
If the Operating System is Win 7, after the installation process is complete,
change the permission of C:\Program Files(x86)\GM4-x.x.x folder to fully
accessable. (right-click this folder, select "Content" and click "Security", select
user name, press "Edit" and check "Full Control",after that, click OK to exit.)

King Design / King Solution / King Service 50 KING DESIGN INDUSTRIAL CO., LTD.
 Device cannot be accessed:

Solution:
1. Ensure that the wiring connection of computer devices is correct.

2. Click the icon on desktop, and then click «Device and Interfaces», ensure
that NI USB-6210 is Dev1, if not, right-click NI USB-6210, select Rename and
rename it to Dev1.
3. Contact King Design maintenance staff.

 There is no error message during the measuring process, but no


measuring result is displayed on screen:
Solution:
1. Ensure if the accelerometer is connected and if the connecting cable is loose.

2. Is the accelerometer sensitivity set in accordance with the aforesaid standard?

3. Is the trigger level of the accelerometer higher than the maximum of the actual
signal
4. Is the display graphic switched to manual and the setting range doesn’t
include signal display reange.

King Design / King Solution / King Service 51 KING DESIGN INDUSTRIAL CO., LTD.
 There are situations such as graphic flicker, continuous measurement
cycles, early stopping for the program during the measuring process:
Solution:

1. Is the accelerometer sensitivity set in accordance with the aforesaid standard ?


2. Is the trigger level of the accelerometer lower than the ripple maximum of
default collision signal ?

 G value and time of the waveform signal is different from the ideal value:
Solution:

1. Ensure if the accelerometer is connected and if the connecting cable is loose.

2. Is the filter frequency 5 times bigger than signal frequency ?

3. Is the test height setting correct ?

4. Is the rubber block correct ?

King Design / King Solution / King Service 52 KING DESIGN INDUSTRIAL CO., LTD.
6. Service offices
Head Office: Taipei, Taiwan
Address: 5F, No.3, Lane 270, Sec. 3, Beishen Road, Shenkeng District,
New Taipei City 222, Taiwan
Tel: 886-2-2662-5100
Fax: 886-02-2662-3094
E-mail: service@kdi.tw
https://www.kdi.tw/en/

Taichung Service Center


Address: No.25, Lane 59, Gongxue Road, South District, Taichung City
402, Taiwan
Tel: 886-4-2262-0971
Fax: 886-4-2262-7130

Hsinchu PV R&D Center


Address: No.635, Sec. 4, Zhongxing Road, Zhudong Township, Hsinchu
County 310, Taiwan (R.O.C.)
Tel: 886-3-583-5033

China Service Offices


 Kunshan Factory, Jiangsu
Address: No.66, Jinpu Road, Zhoushi Town, Kunshan, Jiangsu 215314,
China
Tel: 86-512-5786-8288; Fax: 86-512-5786-8299

 Dongguan Factory
Address: No.5, Xiangnan 4th Street, Huangjiang Township, Dongguan,
Guangdong 523750,China
Tel: 86-769-8351-6866; Fax: 86-769-8351-6000

King Design / King Solution / King Service 53 KING DESIGN INDUSTRIAL CO., LTD.

You might also like