Professional Documents
Culture Documents
1
D 4764
6. Reagents titanium dioxide, and each test paint using the following
6.1 Purity of Reagents—Reagent grade chemicals shall be procedure. Into a 20-mL vial weigh approximately 7.00 g of
used in all tests, unless otherwise specified. It is intended that paint to 1 mg. Pipet 2 mL of TEAB internal standard solution
all reagents conform to the specifications of the Committee on (see 6.4) into the paint and thoroughly mix.
Analytical Reagents of the American Chemical Societywhere
such specifications are available.7 Other grades may be used, TABLE 1 Instrument Conditions
provided it is first ascertained that the reagent is of sufficiently Ti Ka Br Ka
high purity to permit its use without lessening the accuracy of Analyzer crystal LiF LiF
the determination. Counter Flow or Scintillation Scintillation
Collimator Fine Coarse
6.2 Purity of Water—Unless otherwise indicated, references Order First First
to water shall be understood to mean reagent water that Tube KV/mA 10/5 60/24
conforms to the requirements of Type II of Specification Time, s 20 20
Peak, 2u,° 86.09 29.97
D 1193. Background, 2u,° 85.00 29.00
6.3 2-ethoxyethanol, purified grade.
6.4 Tetraethylammonium Bromide, internal standard
solution—Into a 100-mL volumetric flask weigh approxi- 8.2 Place a plastic sheet on the perforated suction plate. On
mately 10.80 g of tetraethylammonium bromide (TEAB) to 1 the plastic sheet make a 3.0-mil (75-µm) thick drawdown of the
mg. Dissolve the TEAB in 2-ethoxyethanol, fill the flask to the liquid paint. Air dry the drawdown film overnight. Cut from the
mark with 2-ethoxyethanol, and mix thoroughly. (TEAB dis- drawdown on the plastic sheet a disk of a size appropriate for
solved in 2-ethoxyethanol can be used as an internal standard the specimen holder of the spectrometer.
for both water-reducible and solvent-reducible paint. However, 8.3 Determine the percent nonvolatile content of each paint
water may be substituted for 2-ethoxyethanol, if only water- at 110°C in accordance with Test Method D 2369 if the
reducible paint is to be analyzed.) information is not already available.
9. Procedure
7. Hazards
9.1 Turn on the spectrometer and allow it to stabilize
7.1 As exposure to excessive quantities of x-radiation is
thoroughly before beginning collection of results. With the aid
injurious to health, X-ray producing equipment can be danger-
of the manufacturer’s literature, select spectrometer settings to
ous to both the operator and persons in the immediate vicinity
permit measurement of the intensity of the Ka peaks of
unless safety precautions are strictly observed. Therefore, users
titanium and bromine. For instruments equipped with an X-ray
should avoid exposing any parts of their bodies, not only to the
tube with a tungsten or chromium target, the settings pre-
direct beam, but also to secondary or scattered radiation that
scribed in Table 1 may be used as guidelines (Note 1).
occurs when an X-ray beam strikes or passes through any
9.2 Place the disk in the specimen holder in such a way that
material. It is strongly recommended that users check the
the coated side of the sheet is turned toward the incident
degree of exposure by film carried on them or by the use of
X-rays. Before measuring the specimens, record the back-
dosimeters and that blood counts be made periodically. Before
ground spectrum in the ranges of the Ka peaks of titanium and
utilizing the equipment, all persons designated or authorized to
bromine, including at least 10° on each side. Check the spectra
operate X-ray instrumentation or supervise its operation,
for the presence of interfering peaks and select for each peak an
should have a full understanding of its nature and should also
angle for background measurement where interference is at a
become familiar with established safe exposure factors by a
minimum.
careful study of the National Bureau of Standards Handbook
9.3 At the selected angles measure the background intensi-
“X-ray Recommendations of the International Roentgen Ray
ties and the intensities of the Ka peaks of titanium and bromine
Committee on X-ray Protection,”8 the manufacturer’s instruc-
in counts per second (c/s). Determine the net intensity, c/s, of
tion manual, and other standard publications on the subject.
each peak by subtracting the background count from the peak
Inquiries should be made of state agencies as to existing
count.
requirements.
10. Calculation
8. Specimen Preparation
10.1 Use the values of the standard paint of known TiO2
8.1 Thoroughly mix and sample each paint in accordance
content to calculate Kt, the reference intensity ratio of titanium
with Practice D 3925. Prepare duplicate specimens of the
relative to bromine in the film, defined as:9,10
standard paint, which contains a known concentration of
B
B 1 N I t B It
Kt 5 T 3 I 5 T 3 I (1)
7 b b
Reagent Chemicals, American Chemical Society Specifications, American
Chemical Society, Washington, DC. For suggestions on the testing of reagents not B1N
listed by the American Chemical Society, see Analar Standards for Laboratory
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
9
and National Formulary, U.S. Pharmaceutical Convention, Inc. (USPC), Rockville, Chung, F. H., “A New Approach to Quantitative Multi-element X-ray Fluores-
MD. cence Analysis,” Advances in X-ray Analysis, 19, 1976, p. 81.
8 10
Available from National Bureau of Standards, National Center for Standards, Chung, F. H., Lentz, A. J., and Scott, R. W., “A Versatile Thin Film Method for
Gaithersburg, MD 20899. Quantitative X-ray Emission Analysis,” X-ray Spectrometry, 3(4), 1974, p. 172.
2
D 4764
The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection
with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such
patent rights, and the risk of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsible
technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your
views known to the ASTM Committee on Standards, 100 Barr Harbor Drive, West Conshohocken, PA 19428.