PANEL DETAILS & SPECIFICATIONS SPECIFICATIONS
1199
‘The BEAR CUB 750rsv Infant Ventilator
Performance Characteristics and Specifications
Inspiratory Pressure
Ventilator Rate.
Inspiratory Time.
Base FLOW ssn
Inspiratory Flow...
Assist Sensitivity.
Volume Limit’
Audible Off (Volume Limit. az LED On/Off
Manual Breath... at so
Over Pressure Relief... 2 15 to 75 cmH20
% 02. ee 21 to 100%
Apnea Alarm. 5, 10, 20,30 sec
Alarm Loudness. soln. to Max, 60 to 75 db(A)
Alarms
Low PEEP/CPAP. a : swn5 to 30 cmH20
High Breath Rate. 1-3 t0 255 BPM
Low Minute Volume Off to 99 LPM
High Pressure Limit. 110 £0 75 emH20
Failed to Cycle. LED On/Off
Low Gas Supply . LED On/Off
Patient Circuit
Prolonged Inspiratory Presswe
Apnea,
Pressure Settings Incompatible
Flow Sensor..
Alarm Silence.
Visual Reset.
Line Power...
Low Battery.
Monitors
Breath Rate.
Breath Type (Patient initiated),
Inspiratory Time.. :
Expiratory Time...
LE Ratio .
Peak Inspiratory Pressure.
Mean Airway Pressure.
PEEP.
Air Pressure .
02 Pressure
Minute Volume
‘Tidal Volume (Exhaled),
Tidal Volume (Inspired).
YoLeak
LED On/Off
selaeaeeu Green/Red LED
- seenennnnanens LED On/Off
sa 0 255 BPM
Proximal Airway Pressure... vn=10 to 100 cmF120
Hourmeter. " von €0 99,999 hours.
Test - Push Button,
Battery nn [LED On/Off
“US, Patent No.5, 540,220
53SPECIFICATIONS PANEL DETAILS & SPECIFICATIONS
eee ees
The BEAR CUB 750rsv Infant Ventilator
Performance Characteristics and Specifications
Preset Values
Minimum Expiratory Time wn 150 milliseconds
Minimum Inspiratory Time.wnnnovuesnn ‘0. seconds|
Maximum Inspiratory Time. : ‘nnB.10 seconds
Maximum settable EE Ratio eee
Tolerances
Monitors:
Breath Rate.... : vtec 1 BPM of #20 msec, whichever is greater
Inspiratory Time. “002 sec
Expitatory Time <= 0.02 see
EE Ratio £0.1 or + 20 msec (on the calculation based on the monitored,
inspiratory and expiratory times), whichever is greater
Airway Pressure (continuous) . 2 LemFEO (-10 to 20 em:0)
2 emH.O (20 to 65 cmF:O)
+ 3.cmH.O (65 to 100 cmH.)
Peak Inspiratory Pressure.. : suc 2 EMELO or # 5%, whichever is greater
Mean Airway Pressure... +2emH.O or = 3%, whichever is greater
PEEP... = 2emELO or = 5%, whichever is greater
Air/O: Pressure. natant 5 PSI
Minute Volume . : "Accuracy is based on the Tidal Volume and Breath Rate monitors
Inspiratory Tidal Volume. “ont TmnL oF = 10%, whichever is greater
Expizatory Tidal Volume.
% Leak
“i UmL or = 10%, whichever is greater
£ LmL or = 2%, whichever is greater
(when compared to the Inspired and Expired Tidal Volume monitors)
+= 2% of reading
Hour Meter -n
Alarms!Limits:
Low PEEP/CPAP. 2 2emH.0
High Breath Rate. +1BPM
Low Minute Volume : 210 mL/min (0 to 99 mL/min)
£0. L/min (0.1 to 9.9 L/min)
High Pressure Limit + 4cmH.O
Prolonged Inspiratory Pressure. a 2cmH.O
‘Volume Limit. sn £ 2mL or + 10% of setting, whichever is greater
Apnea... = 1sec
Maximum (Pop-off) Pressure =4cmH.O
Controls:
Ventilator Rate “#1 BPM or = 20 msec (applied to the breath interval), whicheveris greater
Inspiratory Time eno 0.020 see (0.10 to 0.50 sec)
0.025 sec (0.50 to 3.00 sec)
Assist Sensitivity smn Not a calibrated scale
PEEP/CPAP and Ipiratory Pressure
Repeatabil
LomH.O or # 5%, whichever is greater (breath to breath)
Base/Inspiratory Flow. 0.5 LPM or + 10% of setting, whichever is greater (0 to 60 cmH:O)
LLPM or +10/-15% of setting, whichever is greater (60 to 72 cmH:O}
Oxygen %. 23%
Emissions/Susceptibility
‘The ventilator has been tested to conform to the following specifications: MIL-STD-461D:1983, MIL-STD-462D:1993,
ENSSOLL:1991, IEC 1000-4-2:1994, IEC 1000-4-3:1994, IEC 1000-4-8:1994, IEC 1000-4-5:1994, QUAS-STATIC:1993,
‘This ventilator is designed and manufactured to comply with the safety requirements of IEC 601-1, IEC 601-2
CAN/CSA-C22.2 No. 601:1-M90, and UL 2601-1.
5a nisPANEL DETAILS & SPECIFICATIONS SPECIFICATIONS
The BEAR CUB 750:sv Infant Ventilator
Performance Characteristics and Specifications
RS-232 Bi-Directional
‘Analog: (see page 5-29)
Proximal Pressure... : 10 to 100 em#20, 1 emH20/25 mv
Proximal Inspiratory Expiry Flow. 40 (expiratory) to 40 (inspiratory) LPM, 1 LPM/S0 mw
Machine Delivered Flow. “0 to 30 LPM, 1 LPM/50-mv
Breath Phase 0 sn “Logic” signal; 0,5V
Remote Nurse Callnennuonimnaninnaninnseaninnnainn(S ADS MAX at 24-vde
Preumatic:
‘Auxiliary Blended Gas Outlet 2 7 t017 psig, 0 to8 LPM
KPa (48 to 117)
Inputs
Electrical:
Supply Ratings
Voltage: 100V 80 to 110 VAC
i20V 95 to 135 VAC
2300 176 to 264 VAC
240v 192 to 264 VAC
Current: 100V 1.0.4 maximum
s20v 1.0 A maximurn
2300 05.A maximum
2400 05 Amaximum
Frequency: 100V 47 065 Hz
soy 55 to 65H
230 47 to 65 Hz
240V 47 10 65 Hz
Fuses: 100/120V T05A,5x20 mm
230/240 70.25 A,5x 20 mm
* For 220V operation, set the power entry module selector switch to 230V setting.
Pneumai
(Oxygen and Air. eee -nninnn80 to 80 psig, 50 LPM Supply Ratings
KPa (206 to 551)
11/99 55SPECIFICATIONS PANEL DETAILS & SPECIFICATIONS
The BEAR CUB 750rsv Infant Ventilator
Performance Characteristics and Specifications
Physical Dimension and Shipping Information
Ventilator Weight... 30 lbs, 13.6 kg
Graphics Display Weight. 5.5 Ibs,
Pedestal Stand Weight 9 Ibs,
Compressor Weight... 110 Ibs.
Ventilator Dimensions eae
Graphics Display Dimensions ...
Pedestal Stand Dimensions
‘Compressor Dimensions...
Veto
Wx10"D x11"
3" Wx2.5" Dx9.5"
x 40.5" H
Wx215'Dx36"H
wwAO Ibs., 18.2 kg
Graphics Display. “a1 Ibs.
Pedestal Stand acm 2 Ths.
Compressor “122 Ibs.
Ventilator 8° Wx 18" Dx 19"H
__ Graphics Display. eee “16 Wx 13" Dx 1"
| Pedestal Stand... z 25° Wx 5" Dx45"H
Compressor. Wx22"Dx0"H
Environmental Specifications
‘Temperature:
Storage and Shipping eno 31 to 158° F (-40 1070" C)
Checkout and Operating sr “30 to 104° F (10 to 40°C)
Altitude
‘Checkout, Operating, sna t0 9,000 feet (14.7 to 10.5 PSLA/760 to 543 mim Hg)
kPa (101 to 73)
Humidity:
Storage and Shipping... i : sweat) t0 99% Relative Humidity — Non-condensing
Checkout and Operating... nnnn