Professional Documents
Culture Documents
Copyright (c) ASTM International. 100 Barr Harbor Drive, P.O. Box C700 West Conshohocken PA United States
FIG. 1 Overview of the Couplings Inside the GC Oven and the Couplings to the Solvent Vent Valve
6.10 Pressure Station— This shall ensure a sample in liquid 7.3 Mineral Oil or Local Hydrocarbon Fraction—Boiling
phase at a constant pressure. See Fig. 3 for a typical configu- point range approximately C10-C40. Alternatively, a well char-
ration. acterized local hydrocarbon fraction, within the range C10-C40,
6.11 Typical Column Overview—See Fig. 1. can be used to provide quantitative and qualitative comparison
6.12 Typical Operating Conditions—See Table 1. to the contaminant in the sample. Care should be taken to
ensure no significant fraction falls outside the C10-C40 range.
7. Reagents and Materials
7.4 Validation Standard, Mineral Oil in Pentane—Prepare a
7.1 Mineral Oil in LPG Calibration Mixture—Certified validation standard of mineral oil in pentane. Record the exact
calibration mixture with mineral oil in LPG. The concentration weighed value to the nearest milligram of mineral oil and
of the mineral oil shall be close to the expected concentration calculate the concentration in mg/kg. The concentration of the
of the contamination in the LPG sample. mineral oil shall be close to the expected concentration of the
7.2 Mineral Oil in Pentane Calibration Mixture—Prepare a contamination in the LPG sample.
calibration standard of mineral oil in pentane. Record the 7.5 N-alkane Retention Time Standard—Mixture containing
weighed value to the nearest milligram of mineral oil and at least C10 and C40 in a concentration of (nominally) 5 mg/L
calculate the concentration in mg/kg. The concentration of the each, dissolved in pentane or heptane.
mineral oil shall be close to the expected concentration of the
7.6 Solvent—GC grade pentane.
contamination in the LPG sample.
7.2.1 Standards that are prepared in pentane, normally
liquid at room temperature, should be stored in suitable 8. Hazards
containers under refrigeration and transferred to sample cylin- 8.1 There is a significant fire hazard from LPG, and since
ders prior to use. Alternatively, they may be stored in airtight the boiling point of LPG can be as low as -41°C, there is a risk
cylinders. of freezing “burns.” Take appropriate safety precautions to
A Sample cylinder
B Sample line in
C Injection device
D Cool on column inlet
E Gas chromatograph
F Sample line out
G Rotometer
H Vaporizer
I Waste system
P Pressure gauge
FIG. 3 Typical Configuration of a Pressure Station
prevent ignition or fire, and wear suitable protective equipment 9.3 High Pressure Liquefied Gas Injector—Install in accor-
to protect against skin contact with LPG. dance with the manufacturer‘s instructions.
8.2 An appropriate laboratory ventilation system shall be 9.4 Column Configuration—Install the columns as shown in
used. Fig. 1. Use low dead volume connections, and check for leaks.
8.3 An appropriate waste line shall be installed. The pres-
sure station and injector shall be connected to this line. The 10. Calibration
waste line should vent outside the building. 10.1 Perform a one point calibration at the startup of the
8.4 Pressure station, cylinder, injector, and controller shall instrument, when the result of the validation sample falls
be grounded appropriately. outside the acceptable SQC limits in accordance with Section
14 or after changes in the application hardware or gas supply,
9. Preparation of Apparatus or both.
9.1 Gas Chromatograph—Install and verify performance in 10.2 Run a blank run, without sample injection. Cycle the
accordance with the manufacturer‘s instructions. Typical oper- GC several times until the baseline is stable. A baseline is
ating conditions are shown in Table 1. stable when the start and end signal (in pA) of two consecutive
9.2 Pressure Station—Install in accordance with the manu- blank runs are within 5%. An unstable baseline can be caused
facturer‘s instructions. Purge sample and check carefully for by a leak, detector gases, or by high boiling point components
leaks. or materials that have not yet eluted from the column. The
12.4 Calculation of the response factor, using the calibration 13. Report
mixture: 13.1 Report the results to the nearest mg/kg oily residue in
Rf 5 Scg/Ac (1) LPG, referencing this test method.
where: 14. Quality Control
Rf = Response factor,
Scg = Mineral oil content in the LPG calibration standard 14.1 Confirm the performance of the instrument or the test
or in the standard in pentane in mg/kg weight, and procedure by analyzing validation samples (see 7.4) after each
Ac = Summed area of the peaks in the range of C10-C40 in calibration and at least each day of use thereafter.
the LPG calibration standard or in the pentane 14.2 When QC/quality assurance (QA) protocols are al-
standard. ready established in the testing facility, these may be used
12.5 Calculation of the sample residue concentration; when when they confirm the reliability of the test result.
the calibration mixture and the sample have the same density: 14.3 When there is no QC/QA protocol established in the
testing facility, Appendix X2 can be used as the QC/QA
S 5 Area * Rf (2)
system.
where:
S = Mineral oil content in the sample in mg/kg, 15. Precision and Bias 3
Area = Summed area of the peaks in the range of C10-C40 15.1 Precision—The temporary precision was determined
in the sample, and according to Section A21.2.3 of the Form and Style for ASTM
Rf = Response factor = mineral oil content in the cali- Standards. The full precision will be determined within five
bration standard in mg/kg divided by the area. years of the date of initial approval (Oct. 1, 2011).
12.6 Calculation of the sample residue concentration with 15.2 Repeatability—The difference between two test re-
correction for the density; to be used when the density of the sults, obtained by the same operator with the same apparatus
calibration mixture and the sample differ. under constant operating conditions on identical test material
S 5 Area * Rf * D / DC (3) would, in the long run, in the normal and the correct operation
of the test method, exceed the following values in only 1 case
where:
in 20, where X = the average of two test results:
S = residue content in the sample in mg/kg,
Area = Summed area of the peaks in the range of C10-C40 Repeatability Limit, mg/kg = 0.09(X + 12)
in the sample, Repeatability std dev = repeatability limit/2.77
Rf = Response factor = mineral oil content in the cali- NOTE 1—The applicable range for the interim precision equation is
bration standard in mg/kg divided by the area, nominally 11 to 592 mg/kg.
D = Density of the sample solution at measurement
temperature, g/mL, and
Dc = Density of calibration standard at measurement 3
Supporting data have been filed at ASTM International Headquarters and may
temperature, g/mL. be obtained by requesting Research Report RR:D02-1730.
APPENDIXES
(Nonmandatory Information)
X1. DENSITY CALCULATION OF THE LPG AND CORRECTION OF THE RESIDUE RESULT
X1.1 The mass of the sample injected into the GC using the X = all determined components in the sample.
liquefied injector depends on the pressure of the sample, the
time setting for the injection, and the density of the sample. X1.4 Component Properties for Common LPG
X1.1.1 An analytical method generally holds the pressure Components
and the time setting constant. A variation in injected mass X1.4.1 Excerpted from Practice D2421, Table 2 and D2598
occurs when there is a difference in composition (and therefore Table 1).
density) between the LPG sample and the calibration material Component Relative Density Factor Liquid Volume
(for example, calibration standards prepared in pentane). Conversion Factor
X1.1.2 In that case the final residue result in mg/kg needs to Methane 0.3 0.002261
Ethane 0.35639 0.003565
be corrected for the difference in the injected mass. Propane 0.50736 0.003672
X1.1.3 When the sample density has not been determined Iso butane 0.56293 0.004362
by direct measurement, it may be calculated according to N butane 0.58407 0.004205
Practice D2598 using the LPG composition analysis.
X1.5 Example Calculation
X1.1.4 The composition of an LPG is readily determined by
GC methods that generally conform to Test Method D2163. X1.5.1 A sample calculation using the following LPG
This test method suggests reporting results in liquid volume % Compositional analysis follows. Oily residue chromatogram
of the major sample components. Analytical results or calibra- area equivalent to 35 ppm mass/mass in pentane.
tion factors, or both, may alternatively be expressed in mass % Component Mass %
or mole % as needed by the user. Unit inter-conversion may be Methane 0.00
done using the procedures outlined in Practice D2421. Ethane 0.05
Propane 78.45
Isobutane 5.50
X1.2 Inter-conversion to liquid volume % from mass % and N butane 16.00
mole (gas vol) % according to Practice D2421:
X1.5.2 Step 1—Inter-conversion to Liquid Volume %:
Mass % x
Liquid volume % 5 SRelative density factor x (X1.1) Component Mass% Divide by Relative Quotient Multiply by Normalization
Density Component Factor (100/191.928)
Liquid volume % 5 S~mole % x · liquid volume factor x! Property
(X1.1) Methane 0.00 0.3 0.00 0.00
Ethane 0.05 0.35639 0.140 0.07
where: Propane 78.45 0.50736 154.624 80.57
Iso-Butane 5.50 0.56293 9.770 5.09
x = all determined components in the sample. N butane 16.00 0.58407 27.394 14.27
X1.2.1 The summed results are expressed in % through a
Total 100.00 ... 191.928 100.00
normalization step. Density factors and liquid volume factors
are taken from the tables provided in Practice D2421. X1.5.3 Step 2—Density Calculation of the Mixture:
X1.2.2 Factors for methane, ethane, propane, iso and nor-
mal butane, and pentane are provided in X1.4. density mixture 5 S S
0.07 * 0.356
100 DS 80.57 * 0.562
100 DS 5.09 * 0.562
100 D
X1.3 Density Calculation According Practice D2598 S14.27 * 0.584
100 D
X1.3.1 5 0.521 g/mL (X1.3)
X2.1 Confirm the performance of the instrument or the test X2.4 The frequency of QC testing is dependent on the
procedure by analyzing quality control (QC) sample(s). criticality of the quality being measured, the demonstrated
stability of the testing process, and customer requirements.
X2.2 Prior to monitoring the measurement process, the user Generally, a QC sample should be analyzed each testing day
of the test method needs to determine the average value and with routine samples. The QC frequency should be increased if
control limits of the QC sample. See Practice D6299 and a large number of samples are routinely analyzed. However,
MNL7.4 when it is demonstrated that the testing is under statistical
control, the QC testing frequency may be reduced. The QC
X2.3 Record the QC results and analyze by control charts sample testing precision should be periodically checked against
or other statistically equivalent techniques to ascertain the the ASTM method precision to ensure data quality. See
statistical control status of the total testing process. See PracticeD6299 and MNL7.4
Practice D6299 and MNL7.4 Investigate any out-of-control
data for root cause(s). The results of this investigation may, but X2.5 It is recommended that, if possible, the type of QC
not necessarily, result in instrument recalibration. sample that is regularly tested be representative of the material
routinely analyzed. An ample supply of QC sample material
NOTE X2.1—In the absence of explicit requirements given in the test should be available for the intended period of use, and must be
method, X2.4 provides guidance on QC testing frequency.
homogenous and stable under the anticipated storage condi-
tions.
4
MNL 7, Manual on Presentation of Data and Control Chart Analysis, ASTM X2.6 See Practice D6299 and MNL74 for further guidance
International. on QC and control charting techniques.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/
COPYRIGHT/).