Rev1 (System Integration Guide - TDI)
Rev1 (System Integration Guide - TDI)
(T.D.I. interfaced)
Revision history
Contents
1 INTRODUCTION 1
1 INTRODUCTION
This Manual includes the interconnections list, the hardware and
software configuration for a typical radiological room using a generator
G 100 R/F-A.
For the use of this generator with different configurations of room and for
X-ray tube and AEC calibration, please make reference to the
G 100 R/F-A Service Manual and equipment you have to connect with.
Room configuration
• Apollo
– I.A.E. X-ray tube type
– Three field solid state AEC (Gilardoni or Comet type)
– I.I. H.R. – CCD N33.
• G 100 R/F-A
• Chest Stand VD3 35x43 with motorised foot rest (See Note)
– Three field solid state AEC (Gilardoni or Comet type).
Receptor configuration
NOTE:
When the Generator is supplied with "Falling load" feature, the GalKey
U29 # 73594213 is installed on CPU board.
The Falling Load option it is not available on the receptors where the
"SERIAL" exposure is enable (par. 2.3.4 - Image Receptor Programming
Worksheet: "SERIAL=YES"); whenever the Falling Load has to be
activated on the Spot Film Device receptor, the SERIAL parameter has to
be set to "NO".
In case Falling Load feature and SERIAL exposure are required will be
necessary to create for each a dedicated receptor.
NOTE:
Reference to the Chest Stand VD3 35x43 with motorized foot rest is made
in order to allows the installer to record the necessary information for the
hardware interface among G 100 R/F-A and Potter Bucky.
R R
S S
T T
N N
G 100 R/F-A
AUXILIARY CONTACTOR
R S T R S T N
* Made this connection only when a motorised foot rest chest stand is present.
Terminal
Description Wire n° Terminal Description
X0
Remote Exposure
38 TB6.9
(Room Interface Board)
Preparation radiography
39 TB6.7
(Room Interface Board)
Spare Output
Common 0VS 20 TB6.11
(Room Interface Board)
X-ray ON
X-ray ON
from 13 J6.21
(Table Digital Interface board)
Generator
Spot Film
Spare Output
Device 15 TB6.12
(Room Interface Board)
selection
Safety switch
Safety switch 9 J17.1
(Generator Interface PWB)
Safety switch
Safety switch 11 J17.2
(Generator Interface PWB)
NOTE:
During the auto calibration of the X-ray tube it is required to make the
jumper between J7.6 and J6.2 (force terminal J7.6 at 0Vdc).
It is important to remove the jumper at the end of the X-ray tube
auto calibration.
Terminal
Description Wire n° Terminal Description
X0
Tomo
42
Speed 1
Tomo
36
Speed 3
Tomo Angle 7°
Tomo 7° 44 J7.1
(Table Digital Interface board)
Signal Ground
Common 0V 20 J6.2
(Table Digital Interface board)
Signal Ground
J7.2
(Table Digital Interface board)
Chassis Ground
J7.7
(Table Digital Interface board)
Chassis Ground
J7.20
(Table Digital Interface board)
L.I.H.
Last Image Hold
(Table Digital Interface J6.17 X2.4
(see Note)
board)
Proportional dc
(Table Digital Interface J7.21 X2.1 Analog kV auto
board)
0Vdc
(Table Digital Interface J7.2 X2.2 0Vdc
board)
SYNC
(Table Digital Interface J7.24 X4.5 SYNC
board)
P.F. request
(Table Digital Interface J6.22 X4.6 Pulsed Fluoro
board)
Terminals Pcb A1
Description Wire n° Description
X0 4 fields 3 fields
Maximum
Minimum field 31 X4.3 X4.2
zoom
Minimum
Maximum field 29 X4.1 N.C.
zoom
Power supply N X0.2 Power supply
Terminal
Description Wire n° Pcb A2 Description
X0
54 X2.15 R.O.I.
R.O.I.
(see Note) X2.11 Recursive Filter X0
Common 0V 28
NOTE:
Depending by selection of the Apollo set-up, cell 246 - X0.54 output can
drive R.O.I. or Recursive Filter. In the first case X0.55 and X0.56 are
disabled, in the second case are enabled.
NOTE:
The following connections has to be used in case of "in-room monitor
with trolley" installation.
Terminal Terminals
Description Wire n° Description
X0 X4
0 Vs common 60 1 0 Vs common
Terminal
Description Wire n° Plug Description
X4
Shield 4 4 Shield
2.2.5 G 100 R/F-A – Three field solid state AEC device 1 (Gilardoni
or Comet type)
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
White
Start J1.8 Pink Common
Green
G100 R/F-A
AEC board AEC device 1
(solid state chamber) connecting cable
type 733347
G100 R/F-A
AEC board AEC device 1
(solid state chamber) connecting cable
type 737992
NOTE:
This hardware interface has to be made only if the Potter Bucky is
present.
WARNING:
The voltage selected on TB1.11 and TB1.12 is 24 Vdc. Refer to the Chest
Stand VD3 35x43 Technical and User's Manual for the Bucky start
voltage setting.
2.2.8 G 100 R/F-A – Three field solid state AEC device 2 (Gilardoni
or Comet type)
NOTE:
This hardware interface has to be made only if the Potter Bucky with
AEC device is present.
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
White
Start J2.8 Pink Common
Green
G 100 R/F-A
AEC board AEC device 2
(solid state chamber) connecting cable
type 733347
White
Anode J12.2 Pink Common
Green
G100 R/F-A
AEC board AEC device 2
(solid state chamber) connecting cable
type 737992
NOTE:
The configuration of the Jumpers not listed in the previous tables is not
relevant for this system integration.
Prior to beginning tube calibration, the tube used in the installation must
be properly selected, and generator limits should be programmed.
If the x-ray tube used in the installation is present in the library of the
generator refer to the G 100 R/F-A Service Manual, chapter 3C -
Programming the generator in order to record all the necessary
information to program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter-Programming the dual speed starter to record the necessary
information to complete the configuration.
If the X-ray tube used in the installation is not present in the library of
the generator should be present in the utility software. The data
communication with a computer it is allowed using the GenWare® utility
software. Refer to the G 100 R/F-A Service Manual, chapter 3C - DATA
LINK to record the necessary information.
2.3.1.3 X-ray tube not present neither in library nor in GenWare® utility
software
If the x-ray tube used in the installation is not present neither in the
generator library nor in the GenWare® utility software a manual setting
of the tube and generator parameters is required. Before changing these
parameters is strictly recommended to consult the data sheet of the x-ray
tube that is going to be installed. An X-ray tube present in the library or
in the GenWare® utility software, as similar as possible to the one is
going to be installed, can be selected. The tube and the generator
parameter required in the generator set-up must be set in according with
the information recorded from the data sheet of the x-ray tube. Refer to
the G 100 R/F-A Service Manual, chapter 3C - Tube selection and
Generator limits in order to record all the necessary information to
program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter - Programming the dual speed starter to record the necessary
information to complete the configuration.
The installation of the AEC devices and their calibration are covered in
the G 100 R/F-A Service Manual, chapter 3E - Automatic exposure
control (AEC) calibration.
NOTE:
The Fluoro set-up data must be set in according with the X-ray tube
technical specification.
The following tables meets the Fluoro set-up request in according with
most of the X-ray tube supplied by Villa Sistemi Medicali; during the
installation could be necessary to optimize the set-up.
The installation of the ABS device and his calibration are covered in the
G 100 R/F-A Service Manual, chapter 3 - Automatic brightness
stabilisation (ABS) calibration.
The next table show the standard parameters usually set.
FLUORO SET-UP
FL TIMER: 5 Min
MIN FLUORO KV: 40
MAX FLUORO KV: 120
II MODES: 0
FL ABS: ON
FL ABS DEFAULT: NONE
FL/RAD KV DEF: NONE
FL/RAD KV CURVES Not Used
SID COMP Not Used
FL SETUP FL ABS SETUP ABS LEVEL 0: 1.00 V
ABS LEVEL 1: 4.00 V
ABS LEVEL 2: 1.10 V
ABS LEVEL 3: 1.60 V
SIGNAL GAIN 3
ABS GAIN 10
ABS DEADBAND 30%
ABD DELAY 150 mS
ABS START KV 75
ABS CURVE DEF: 1
PF SETUP PF: ON
PF DEFAULT: NONE
PPS DEFAULT: NONE
PF DEFAULT MA: 5=40mA
6=60mA
7=80mA
8=99mA
Note: Four options for the settings of the Pulsed Fluoro default mA are possible. It is
suggested to set the value 5 = 40mA as default parameter. The others settings are
possible on line by means the +/- mA keys. If, once set at 40 mA the default mA value
for the pulsed fluoro, is necessary to reduce the dose is possible to reduce the length
of the pulses by means the settings of the jumpers on the A2 board of the I.I. system
(make reference to the I.I. system service manual)
PF DEFAULT Ms: 2
PF SYNC FREQ: 50 HZ
PF ABS SETUP ABS LEVEL 0: 4.0 V
ABS LEVEL 1: 1.10 V
ABS LEVEL 2: 1.20 V
ABS LEVEL 3: 1.90 V
SIGNAL GAIN: 3
ABS GAIN: 5
ABS DEADBAND: 40%
ABS DELAY: 100 mS
ABS START KV: 75
PF INTEGRATOR: OFF
1PPS COMP: 0%
3PPS COMP: 0%
6/7.5PPS COMP: 0%
12/15PPS COMP: 0%
25/30PPS COMP: 0%
PF ABS CURVE DEF: OFF
PF ABS CURVES Not Used
PF DOSE LIMITS Not Used
PF II COMP Not Used
PF PPS SETUP 1 PPS: ON
3 PPS: ON
6/7.5 PPS: ON
12.5/15 PPS: ON
28/30 PPS: ON
HPF SETUP Not Enable
HLF SELECT: CONSOLE
PF SELECT: CONSOLE
NOTE:
Receptor 3 has to be programmed only if Chest Stand with Potter Bucky
and AEC device is present.
***INPUTS***
REMOTE EXP ----- OFF ----- ON -----
REMOTE PREP ----- ON ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- ON
CONSOLE EXP ----- OFF ----- OFF -----
CONSOLE PREP ----- OFF ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE ON ON ON ON ON
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON OFF
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- ON -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF ON OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
NOTE:
The configuration of the Receptor 3 I/O has to be made only if the Chest
Stand with Potter Bucky is present.
Room configuration
• Apollo
– I.A.E. X-ray tube type
– Three field solid state AEC (Gilardoni or Comet type)
– Diva-D.
• G 100 R/F-A
• Chest Stand VD3 35x43 with motorised foot rest (See Note)
– Three field solid state AEC (Gilardoni or Comet type).
Receptor configuration
NOTE:
When the Generator is supplied with "Falling load" feature, the GalKey
U29 # 73594213 is installed on CPU board.
The Falling Load option it is not available on the receptors where the
"SERIAL" exposure is enable (par. 2.3.4 - Image Receptor Programming
Worksheet: "SERIAL=YES"); whenever the Falling Load has to be
activated on the Spot Film Device receptor, the SERIAL parameter has to
be set to "NO".
In case Falling Load feature and SERIAL exposure are required will be
necessary to create for each a dedicated receptor.
NOTE:
Reference to the Chest Stand VD3 35x43 with motorized foot rest is made
in order to allows the installer to record the necessary information for the
hardware interface among G 100 R/F-A and Potter Bucky.
R R
S S
T T
N N
G 100 R/F-A
AUXILIARY CONTACTOR
R S T R S T N
* Made this connection only when a motorised foot rest chest stand is present.
Terminal
Description Wire n° Terminal Description
X0
Remote Exposure
38 TB6.9
(Room Interface Board)
Preparation radiography
39 TB6.7
(Room Interface Board)
Common 0V
20 J6.2
0Vdc (Table Digital Interface board)
0V analog
Common 0VS 58 J6.23
(Table Digital Interface board)
Fluoro kV
Man/Auto fluoro
manual/auto 49 J6.1
(Table Digital Interface board)
selection
Tomo
42
speed 1
Tomo
36
speed 3
Tomo 7°
Tomo 7° 44 J7.1
(Table Digital Interface board)
Tomo 20°
Tomo 20° 45 J7.3
(Table Digital Interface board)
Tomo 30°
Tomo 30° 46 J7.4
(Table Digital Interface board)
Terminal
Description Wire n° Terminal Description
X0
Tomo 45°
Tomo 45° 47 J7.5
(Table Digital Interface board)
Bucky 1 select
SFD selection 15 TB2.12
(Room Interface Board)
Bucky 2 select
DSI selection 14 TB2.1
(Room Interface Board)
X-ray ON
X-Ray on
from 13 J6.21
(Table Digital Interface board)
generator
Safety switch
Safety switch 9 J17.1
(Generator Interface PWB)
Safety switch
Safety switch 11 J17.2
(Generator Interface PWB)
NOTE:
During the auto calibration of the X-ray tube it is required to make the
jumper between J7.6 and J6.2 (force terminal J7.6 at 0Vdc).
It is important to remove the jumper at the end of the X-ray tube
auto calibration.
APOLLO DIVA-D
Terminal
Description Wire n° POD pcb Description
X0
37 20 Fluoroscopy request
38 15 Generator ready
40 21 Exposure request
D.S.I.
Common
20 11 0 V digital common
0Vdc
X-ray on
from 13 14 X-ray ON
generator
POD
Description TDI Wire n° Description
pcb
Terminals Pcb A1
Description Wire n° Description
X0 4 fields 3 fields
Maximum
Minimum field 31 X4.3 X4.2
zoom
Minimum
Maximum field 29 X4.1 N.C.
zoom
NOTE:
The following connections has to be used in case of "in-room monitor
with trolley" installation.
Terminal Terminals
Description Wire n° Description
X0 X4
0 Vs common 60 1 0 Vs common
Terminal
Description Wire n° Plug Description
X4
Shield 4 4 Shield
3.2.8 G 100 R/F-A – Three field solid state AEC device 1 (Gilardoni
or Comet type)
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
White
Start J1.8 Pink Common
Green
G100 R/F-A
AEC board AEC device 1
(solid state chamber) connecting cable
type 733347
G100 R/F-A
AEC board AEC device 1
(solid state chamber) connecting cable
type 737992
NOTE:
This hardware interface has to be made only if the Potter Bucky is
present.
WARNING:
The voltage selected on TB1.11 and TB1.12 is 24 Vdc. Refer to the Chest
Stand VD3 35x43 Technical and User's Manual for the Bucky start
voltage setting.
3.2.11 G 100 R/F-A – Three field solid state AEC device 2 (Gilardoni
or Comet type)
NOTE:
This hardware interface has to be made only if the Potter Bucky with
AEC device is present.
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
White
Start J2.8 Pink Common
Green
G 100 R/F-A
AEC board AEC device 2
(solid state chamber) connecting cable
type 733347
White
Anode J12.2 Pink Common
Green
G100 R/F-A
AEC board AEC device 2
(solid state chamber) connecting cable
type 737992
NOTE:
The configuration of the Jumpers not listed in the previous tables is not
relevant for this system integration.
Prior to beginning tube calibration, the tube used in the installation must
be properly selected, and generator limits should be programmed.
If the x-ray tube used in the installation is present in the library of the
generator refer to the G 100 R/F-A Service Manual, chapter 3C -
Programming the generator in order to record all the necessary
information to program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter-Programming the dual speed starter to record the necessary
information to complete the configuration.
If the X-ray tube used in the installation is not present in the library of
the generator should be present in the utility software. The data
communication with a computer it is allowed using the GenWare® utility
software. Refer to the G 100 R/F-A Service Manual, chapter 3C - DATA
LINK to record the necessary information.
3.3.1.3 X-ray tube not present neither in library nor in GenWare® utility
software
If the x-ray tube used in the installation is not present neither in the
generator library nor in the GenWare® utility software a manual setting
of the tube and generator parameters is required. Before changing these
parameters is strictly recommended to consult the data sheet of the x-ray
tube that is going to be installed. An X-ray tube present in the library or
in the GenWare® utility software, as similar as possible to the one is
going to be installed, can be selected. The tube and the generator
parameter required in the generator set-up must be set in according with
the information recorded from the data sheet of the x-ray tube. Refer to
the G 100 R/F-A Service Manual, chapter 3C - Tube selection and
Generator limits in order to record all the necessary information to
program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter - Programming the dual speed starter to record the necessary
information to complete the configuration.
The installation of the AEC devices and their calibration are covered in
the G 100 R/F-A Service Manual, chapter 3E - Automatic exposure
control (AEC) calibration.
NOTE:
The Fluoro set-up data must be set in according with the X-ray tube
technical specification.
The following tables meets the Fluoro set-up request in according with
most of the X-ray tube supplied by Villa Sistemi Medicali; during the
installation could be necessary to optimize the set-up.
The installation of the ABS device and his calibration are covered in the
G 100 R/F-A Service Manual, chapter 3 - Automatic brightness
stabilisation (ABS) calibration.
The next table show the standard parameters usually set.
FLUORO SET-UP
FL TIMER: 5 Min
MIN FLUORO KV: 40
MAX FLUORO KV: 120
II MODES: 0
FL ABS: ON
FL ABS DEFAULT: NONE
FL/RAD KV DEF: NONE
FL/RAD KV CURVES Not Used
SID COMP Not Used
FL SETUP FL ABS SETUP ABS LEVEL 0: 5.00 V
ABS LEVEL 1: 5.00 V
ABS LEVEL 2: 5.00 V
ABS LEVEL 3: 5.00 V
SIGNAL GAIN 10
ABS GAIN 10
ABS DEADBAND 3%
ABD DELAY 150 mS
ABS START KV 75
ABS CURVE DEF: 1
PF SETUP PF: ON
PF DEFAULT: NONE
PPS DEFAULT: NONE
PF DEFAULT MA: 5=40mA
6=60mA
7=80mA
8=99mA
Note: Four options for the settings of the Pulsed Fluoro default mA are possible. It is
suggested to set the value 5 = 40mA as default parameter. The others settings are
possible on line by means the +/- mA keys. If, once set at 40 mA the default mA value
for the pulsed fluoro, is necessary to reduce the dose is possible to reduce the length
of the pulses by means the settings of the jumpers on the A2 board of the I.I. system
(make reference to the I.I. system service manual)
PF DEFAULT Ms: Not
Used
PF SYNC FREQ: 50 HZ
PF ABS SETUP ABS LEVEL 0: 5.0 V
ABS LEVEL 1: 5.0 V
ABS LEVEL 2: 5.0 V
ABS LEVEL 3: 5.0 V
SIGNAL GAIN: 10
ABS GAIN: 8
ABS DEADBAND: 10%
ABS DELAY: 150 mS
ABS START KV: 75
PF INTEGRATOR: OFF
1PPS COMP: 0%
3PPS COMP: 0%
6/7.5PPS COMP: 0%
12/15PPS COMP: 0%
25/30PPS COMP: 0%
PF ABS CURVE DEF: NONE
PF ABS CURVES Not Used
PF DOSE LIMITS Not Used
PF II COMP Not Used
PF PPS SETUP 1 PPS: ON
3 PPS: ON
6/7.5 PPS: ON
12.5/15 PPS: ON
28/30 PPS: ON
HPF SETUP Not Enable
HLF SELECT: CONSOLE
PF SELECT: CONSOLE
NOTE:
Receptor 3 has to be programmed only if Chest Stand with Potter Bucky
and AEC device is present.
***INPUTS***
REMOTE EXP ----- OFF ----- ON -----
REMOTE PREP ----- ON ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- ON
CONSOLE EXP ----- OFF ----- OFF -----
CONSOLE PREP ----- OFF ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT ON ON ON ON ON
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF ON ON
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
***INPUTS***
REMOTE EXP ----- OFF ----- ON -----
REMOTE PREP ----- ON ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- ON
CONSOLE EXP ----- OFF ----- OFF -----
CONSOLE PREP ----- OFF ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT ON ON ON ON ON
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF ON OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- ON -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF ON OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
NOTE:
The configuration of the Receptor 3 I/O has to be made only if the Chest
Stand with Potter Bucky is present.
Room configuration
• Apollo
– I.A.E. X-ray tube type
– Three field solid state AEC (Gilardoni or Comet type)
– Standard I.I. - CCD N23.
• G 100 R/F-A
• Chest Stand VD3 35x43 with motorised foot rest (See Note)
– Three field solid state AEC (Gilardoni or Comet type).
Receptor configuration
NOTE:
When the Generator is supplied with "Falling load" feature, the GalKey
U29 # 73594213 is installed on CPU board.
The Falling Load option it is not available on the receptors where the
"SERIAL" exposure is enable (par. 2.3.4 - Image Receptor Programming
Worksheet: "SERIAL=YES"); whenever the Falling Load has to be
activated on the Spot Film Device receptor, the SERIAL parameter has to
be set to "NO".
In case Falling Load feature and SERIAL exposure are required will be
necessary to create for each a dedicated receptor.
NOTE:
Reference to the chest stand VD3 35x43 with motorized foot rest is made
in order to allows the installer to record the necessary information for the
hardware interface among G 100 R/F-A and Potter Bucky.
R R
S S
T T
N N
G 100 R/F-A
AUXILIARY CONTACTOR
R S T R S T N
* Made this connection only when a motorised foot rest chest stand is present.
Terminal
Description Wire n° Terminal Description
X0
Remote Exposure
38 TB6.9
(Room Interface Board)
Preparation radiography
39 TB6.7
(Room Interface Board)
Common 0V
20 J6.2
0Vdc (Table Digital Interface board)
0V analog
Common 0VS 58 J6.23
(Table Digital Interface board)
Fluoro kV
Man/Auto fluoro
manual/auto 49 J6.1
(Table Digital Interface board)
selection
Tomo
42
speed 1
Tomo
36
speed 3
Tomo 7°
Tomo 7° 44 J7.1
(Table Digital Interface board)
Tomo 20°
Tomo 20° 45 J7.3
(Table Digital Interface board)
Tomo 30°
Tomo 30° 46 J7.4
(Table Digital Interface board)
Terminal
Description Wire n° Terminal Description
X0
Tomo 45°
Tomo 45° 47 J7.5
(Table Digital Interface board)
Alarm X-ray
X-ray off
on 48 J7.6
(Table Digital Interface board)
(see Note)
Bucky 1 select
SFD selection 15 TB2.12
(Room Interface Board)
X-ray ON
X-Ray on
from 13 J6.21
(Table Digital Interface board)
generator
Safety switch
Safety switch 9 J17.1
(Generator Interface PWB)
Safety switch
Safety switch 11 J17.2
(Generator Interface PWB)
NOTE:
During the auto calibration of the X-Ray tube it is required to make the
jumper between J7.6 and J6.2 (force terminal J7.6 at 0Vdc).
It is important to remove the jumper at the end of the X-Ray tube
auto calibration.
0Vdc
(Table Digital Interface J7.2 X2.1 0Vdc
board)
Proportional DC
(Table Digital Interface J7.21 X2.2 Analog. kV auto
board)
Terminals Pcb A1
Description Wire n° Description
X0 4 fields 3 fields
Maximum
Minimum field 31 X4.3 X4.2
zoom
Minimum
Maximum field 29 X4.1 N.C.
zoom
Terminal
Description Wire n° Pcb A2 Description
X0
26 X3.4 Inv
27 X3.5 Inv
0 Vs common 60
Not
Shield 4 Shield
connected
4.2.5 G 100 R/F-A – Three field solid state AEC device 1 (Gilardoni
or Comet type)
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
White
Start J1.8 Pink Common
Green
G100 R/F-A
AEC board AEC device 1
(solid state chamber) connecting cable
type 733347
G100 R/F-A
AEC board AEC device 1
(solid state chamber) connecting cable
type 737992
NOTE:
This hardware interface has to be made only if the Potter Bucky is
present.
Terminal
Description Terminal Wire n° Description
(Room Int. Board)
Bucky 2
X-ray request X1.B1 TB2.2
select
Bucky 2
X-ray request X1.B2 TB2.1
select
Bucky
Potter reply X1.X TB2.5
contacts
Bucky
Potter reply X1.X TB2.4
contacts
WARNING:
The voltage selected on TB2.1 and TB2.2 is 24 Vdc. Refer to the Chest
Stand VD3 35x43 Technical and User's Manual for the Bucky start
voltage setting.
4.2.8 G 100 R/F-A – Three field solid state AEC device 2 (Gilardoni
or Comet type)
NOTE:
This hardware interface has to be made only if the Potter Bucky with
AEC device is present.
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
White
Start J2.8 Pink Common
Green
G 100 R/F-A
AEC board AEC device 2
(solid state chamber) connecting cable
type 733347
White
Anode J12.2 Pink Common
Green
G100 R/F-A
AEC board AEC device 2
(solid state chamber) connecting cable
type 737992
NOTE:
The configuration of the Jumpers not listed in the previous tables is not
relevant for this system integration.
Prior to beginning tube calibration, the tube used in the installation must
be properly selected, and generator limits should be programmed.
If the x-ray tube used in the installation is present in the library of the
generator refer to the G 100 R/F-A Service Manual, chapter 3C -
Programming the generator in order to record all the necessary
information to program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter-Programming the dual speed starter to record the necessary
information to complete the configuration.
If the X-ray tube used in the installation is not present in the library of
the generator should be present in the utility software. The data
communication with a computer it is allowed using the GenWare® utility
software. Refer to the G 100 R/F-A Service Manual, chapter 3C - DATA
LINK to record the necessary information.
4.3.1.3 X-ray tube not present neither in library nor in GenWare® utility
software
If the x-ray tube used in the installation is not present neither in the
generator library nor in the GenWare® utility software a manual setting
of the tube and generator parameters is required. Before changing these
parameters is strictly recommended to consult the data sheet of the x-ray
tube that is going to be installed. An X-ray tube present in the library or
in the GenWare® utility software, as similar as possible to the one is
going to be installed, can be selected. The tube and the generator
parameter required in the generator set-up must be set in according with
the information recorded from the data sheet of the x-ray tube. Refer to
the G 100 R/F-A Service Manual, chapter 3C - Tube selection and
Generator limits in order to record all the necessary information to
program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter - Programming the dual speed starter to record the necessary
information to complete the configuration.
The installation of the AEC devices and their calibration are covered in
the G 100 R/F-A Service Manual, chapter 3E - Automatic exposure
control (AEC) calibration.
NOTE:
The Fluoro set-up data must be set in according with the X-ray tube
technical specification.
The following tables meets the Fluoro set-up request in according with
most of the X-ray tube supplied by Villa Sistemi Medicali; during the
installation could be necessary to optimize the set-up.
The installation of the ABS device and his calibration are covered in the
G 100 R/F-A Service Manual, chapter 3 - Automatic brightness
stabilisation (ABS) calibration.
The next table show the standard parameters usually set.
FLUORO SET-UP
FL TIMER: 5 Min
MIN FLUORO KV: 40
MAX FLUORO KV: 120
II MODES: 0
FL ABS: ON
FL ABS DEFAULT: NONE
FL/RAD KV DEF: NONE
FL/RAD KV CURVES Not Used
SID COMP Not Used
FL SETUP FL ABS SETUP ABS LEVEL 0: 1.00 V
ABS LEVEL 1: 1.60 V
ABS LEVEL 2: 1.10 V
ABS LEVEL 3: 1.60 V
SIGNAL GAIN 10
ABS GAIN 20
ABS DEADBAND 3%
ABD DELAY 150 mS
ABS START KV 75
ABS CURVE DEF: 1
NOTE:
The Pulsed Fluoro set-up is not relevant for this system integration
because the function is not available with the N23 CCD camera.
NOTE:
Receptor 2 has to be programmed only if Chest Stand with Potter Bucky
and AEC device is present.
***INPUTS***
REMOTE EXP ----- OFF ----- ON -----
REMOTE PREP ----- ON ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- ON
CONSOLE EXP ----- OFF ----- OFF -----
CONSOLE PREP ----- OFF ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT ON ON ON ON ON
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- ON -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF ON OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
NOTE:
The configuration of the Receptor 2 I/O has to be made only if the Chest
Stand with Potter Bucky is present.
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
Room configuration
• Vision
– Under table X-ray (tube 1)
– Floor mounted tubestand X-ray (tube 2)
– Under table Potter Bucky with three field solid state AEC
– SFD with three field solid state AEC (Gilardoni or Comet type)
– Standard I.I. system – CCD N33.
• G 100 R/F-A
• Chest Stand VD3 35x43 with motorised foot rest (See Note)
– Three field solid state AEC (Gilardoni or Comet type).
Receptor configuration
NOTE:
When the Generator is supplied with "Falling load" feature, the GalKey
U29 # 73594213 is installed on CPU board.
The Falling Load option it is not available on the receptors where the
"SERIAL" exposure is enable (par. 2.3.4 - Image Receptor Programming
Worksheet: "SERIAL=YES"); whenever the Falling Load has to be
activated on the Spot Film Device receptor, the SERIAL parameter has to
be set to "NO".
In case Falling Load feature and SERIAL exposure are required will be
necessary to create for each a dedicated receptor.
NOTE:
Reference to the chest stand VD3 35X43 with motorized foot rest is made
in order to allows the installer to record the necessary information for the
hardware interface among G 100 R/F-A and Potter Bucky.
R R
S S
T T
N N
G 100 R/F-A
AUXILIARY CONTACTOR
R S T R S T N
* Made this connection only when a motorised foot rest chest stand is present.
Terminal
Description Wire n° Terminal Description
X0
Remote Exposure
8 TB6.9
(Room Interface Board)
Preparation radiography
5 TB6.7
(Room Interface Board)
Common 0V
3 J6.2
0Vdc (Table Digital Interface board)
0V analog
Common 0VS 11 J6.23
(Table Digital Interface board)
Fluoro Kv
Man/Auto fluoro
manual/auto 10 J6.1
(Table Digital Interface board)
selection
NOTE:
During the auto calibration of the X-Ray tube it is required to make the
jumper between J7.6 and J6.2 (force terminal J7.6 at 0Vdc).
It is important to remove the jumper at the end of the X-Ray tube
auto calibration.
Terminal
Description Wire n° Terminal Description
X0
Bucky 1 select
SFD selection 23 TB2.12
(Room Interface Board)
X-ray ON
X-Ray ON
from 22 J6.21
(Table Digital Interface board)
generator
Safety switch
Safety switch 35 J17.1
(Generator Interface PWB)
Safety switch
Safety switch 36 J17.2
(Generator Interface PWB)
0Vdc
(Table Digital Interface J7.2 X2.1 0Vdc
board)
Proportional dc
(Table Digital Interface J7.21 X2.2 Analog. Kv auto
board)
Pulsed fluoro
X-ray on (L.I.H.) Pulsed fluoro
J6.1 X2.4
(Table Digital Interface X-ray on (L.I.H.)
board)
Pulsed fluoro X-ray
request Pulsed fluoro x-ray
J7.24 X4.5
(Table Digital Interface request
board)
Pulsed fluoro request
(Table Digital Interface J6.22 X4.6 Pulsed fluoro request
board)
Terminals Pcb A1
Description Wire n° Description
X0 4 fields 3 fields
Maximum
Minimum field 16 X4.3 X4.2
zoom
Minimum
Maximum field 14 X4.1 N.C.
zoom
Terminal
Description Wire n° Pcb A2 Description
X0
19 X2.9 Inv
18 X2.8 Inv
Common 0 Vdc 3
Terminal Terminal
Description Wire n° Description
X0 (Room Int. Board)
Bucky 2
X-ray request 49 TB2.2
select
Bucky 2
X-ray request 50 TB2.1
select
Bucky
Potter reply 47 TB2.5
contacts
Bucky
Potter reply 48 TB2.4
contacts
WARNING:
The voltage selected on TB2.1 and TB2.2 is 24 Vdc. Refer to the Vision
service manual, under table Potter Bucky voltage setting chapter for
more information.
5.2.5 G 100 R/F-A – SFD three field solid state AEC device 1
(Gilardoni or Comet type)
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface has to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
G100 R/F-A
AEC device 1
Universal AEC board
connecting cable
Type 734614
White
Start J1.8 Pink Common
Green
G100 R/F-A
AEC board AEC device 1
(solid state chamber) connecting cable
type 733347
G100 R/F-A
AEC board AEC device 1
(solid state chamber) connecting cable
type 737992
5.2.6 G 100 R/F-A – Vision under table Potter Bucky three field
solid state AEC device 2 (Gilardoni or Comet type)
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface has to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
G100 R/F-A
AEC device 2
Universal AEC board
connecting cable
Type 734614
White
Start J2.8 Pink Common
Green
G100 R/F-A
AEC board AEC device 2
(solid state chamber) connecting cable
type 733347
G100 R/F-A
AEC board AEC device 2
(solid state chamber) connecting cable
type 737992
NOTE:
This hardware interface has to be made only if the Potter Bucky is
present.
Terminal
Description Terminal Wire n° Description
(Room Int. Board)
Bucky 3
X-ray request X1.B1 TB1.11
select
Bucky 3
X-ray request X1.B2 TB1.12
select
Bucky
Potter reply X1.X TB2.5
contacts
Bucky
Potter reply X1.X TB2.4
contacts
WARNING:
The voltage selected on TB1.11 and TB1.12 is 24 Vdc. Refer to the Chest
Stand VD3 35x43 Technical and User's manual for the Bucky start
voltage setting.
5.2.9 G 100 R/F-A – Three field solid state AEC device 3 (Gilardoni
or Comet type)
NOTE:
This hardware interface has to be made only if the Potter Bucky with
AEC device is present.
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
White
Start J3.8 Pink Common
Green
G 100 R/F-A
AEC board AEC device 3
(solid state chamber) connecting cable
type 733347
White
Anode J13.2 Pink Common
Green
G100 R/F-A
AEC board AEC device 3
(solid state chamber) connecting cable
type 737992
– TB8.1 – TB11.K3 (This wiring must be made only if the 24Vdc under
table Potter Bucky is present on Receptor 2)
– TB8.2 – TB11.K2 (This wiring must be made only if the 24Vdc Chest
stand Potter Bucky is present on Receptor 3)
– TB4.8 – TB4.9: Thermal switch
– TB4.4 – TB4.5: Room door interlock
– TB4.11 – TB4.12: Room light.
NOTE:
The configuration of the Jumpers not listed in the previous tables is not
relevant for this system integration.
Prior to beginning tube calibration, the tube used in the installation must
be properly selected, and generator limits should be programmed.
If the x-ray tube used in the installation is present in the library of the
generator refer to the G 100 R/F-A Service Manual, chapter 3C -
Programming the generator in order to record all the necessary
information to program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter-Programming the dual speed starter to record the necessary
information to complete the configuration.
NOTE:
The tube library, concerning the I.A.E. X-ray tubes, contains only high
speed tube types. When a low speed I.A.E. X-ray tube has to be installed
an high speed tube must be selected informing the generator about the
low speed tube. The generator and the tube limits must be set in
according with the tube type installed.
If the X-ray tube used in the installation is not present in the library of
the generator should be present in the utility software. The data
communication with a computer it is allowed using the GenWare® utility
software. Refer to the G 100 R/F-A Service Manual, chapter 3C - DATA
LINK to record the necessary information.
5.3.1.3 X-ray tube not present neither in library nor in GenWare® utility
software
If the x-ray tube used in the installation is not present neither in the
generator library nor in the GenWare® utility software a manual setting
of the tube and generator parameters is required. Before changing these
parameters is strictly recommended to consult the data sheet of the x-ray
tube that is going to be installed. An X-ray tube present in the library or
in the GenWare® utility software, as similar as possible to the one is
going to be installed, can be selected. The tube and the generator
parameter required in the generator set-up must be set in according with
the information recorded from the data sheet of the x-ray tube. Refer to
the G 100 R/F-A Service Manual, chapter 3C - Tube selection and
Generator limits in order to record all the necessary information to
program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter - Programming the dual speed starter to record the necessary
information to complete the configuration.
The installation of the AEC devices and their calibration are covered in
the G 100 R/F-A Service Manual, chapter 3E - Automatic exposure
control (AEC) calibration.
NOTE:
The Fluoro set-up data must be set in according with the X-ray tube
technical specification.
The following tables meets the Fluoro set-up request in according with
most of the X-ray tube supplied by Villa Sistemi Medicali; during the
installation could be necessary to optimize the set-up.
The installation of the ABS device and his calibration are covered in the
G 100 R/F-A Service Manual, chapter 3 - Automatic brightness
stabilisation (ABS) calibration.
The next table show the standard parameters usually set.
FLUORO SET-UP
FL TIMER: 5 Min
MIN FLUORO KV: 40
MAX FLUORO KV: 120
II MODES: 0
FL ABS: ON
FL ABS DEFAULT: NONE
FL/RAD KV DEF: NONE
FL/RAD KV CURVES Not Used
SID COMP Not Used
FL SETUP FL ABS SETUP ABS LEVEL 0: 5.00 V
ABS LEVEL 1: 5.00 V
ABS LEVEL 2: 5.00 V
ABS LEVEL 3: 5.00 V
SIGNAL GAIN 10
ABS GAIN 10
ABS DEADBAND 3%
ABD DELAY 150 mS
ABS START KV 75
ABS CURVE DEF: 1
PF SETUP PF: ON
PF DEFAULT: NONE
PPS DEFAULT: NONE
PF DEFAULT MA: 5=40mA
6=60mA
7=80mA
8=99mA
Note: Four options for the settings of the Pulsed Fluoro default mA are possible. It is
suggested to set the value 5 = 40mA as default parameter. The others settings are
possible on line by means the +/- mA keys. . If, once set at 40 mA the default mA
value for the pulsed fluoro, is necessary to reduce the dose is possible to
reduce the length of the pulses by means the settings of the jumpers on the A2
board of the I.I. system (make reference to the I.I. system service manual)
PF DEFAULT Ms: Not
Used
PF SYNC FREQ: 50 HZ
PF ABS SETUP ABS LEVEL 0: 5.0 V
ABS LEVEL 1: 5.0 V
ABS LEVEL 2: 5.0 V
ABS LEVEL 3: 5.0 V
SIGNAL GAIN: 10
ABS GAIN: 8
ABS DEADBAND: 10%
ABS DELAY: 150 mS
ABS START KV: 75
PF INTEGRATOR: OFF
1PPS COMP: 0%
3PPS COMP: 0%
6/7.5PPS COMP: 0%
12/15PPS COMP: 0%
25/30PPS COMP: 0%
PF ABS CURVE DEF: NONE
PF ABS CURVES Not Used
PF DOSE LIMITS Not Used
PF II COMP Not Used
PF PPS SETUP 1 PPS: ON
3 PPS: ON
6/7.5 PPS: ON
12.5/15 PPS: ON
28/30 PPS: ON
HPF SETUP Not Enable
HLF SELECT: CONSOLE
PF SELECT: CONSOLE
NOTE:
Receptor 3 has to be programmed only if Chest Stand with Potter Bucky
and AEC device is present.
***INPUTS***
REMOTE EXP ----- OFF ----- ON -----
REMOTE PREP ----- ON ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- ON
CONSOLE EXP ----- OFF ----- OFF -----
CONSOLE PREP ----- OFF ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT ON ON ON ON ON
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- ON -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 OFF OFF ----- OFF OFF
THERMAL SW 2 ON ON ----- ON ON
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF ON OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
NOTE:
The configuration of the Receptor 2 I/O has to be made only if the Vision
under table Potter Bucky is present or an external Potter Bucky has to be
connected on Receptor 2, in example, the Chest Stand.
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- ON -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 OFF OFF ----- OFF OFF
THERMAL SW 2 ON ON ----- ON ON
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF ON OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
NOTE:
The configuration of the Receptor 3 I/O has to be made only if the Chest
Stand with Potter Bucky is present.
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 OFF OFF ----- OFF OFF
THERMAL SW 2 ON ON ----- ON ON
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
Room configuration
• Vision
– X-ray tube type
– Three field solid state AEC (Gilardoni or Comet type)
– Diva-D.
• G 100 R/F-A
• Chest Stand VD3 35x43 with motorised foot rest (See Note)
– Three field solid state AEC (Gilardoni or Comet type).
Receptor configuration
NOTE:
When the Generator is supplied with "Falling load" feature, the GalKey
U29 # 73594213 is installed on CPU board.
The Falling Load option it is not available on the receptors where the
"SERIAL" exposure is enable (par. 6.3.4 - Image Receptor Programming
Worksheet: "SERIAL=YES"); whenever the Falling Load has to be
activated on the Spot Film Device receptor, the SERIAL parameter has to
be set to "NO".
In case Falling Load feature and SERIAL exposure are required will be
necessary to create for each a dedicated receptor.
NOTE:
Reference to the Chest Stand VD3 35x43 with motorized foot rest is made
in order to allows the installer to record the necessary information for the
hardware interface among G 100 R/F-A and Potter Bucky.
R R
S S
T T
N N
G 100 R/F-A
AUXILIARY CONTACTOR
R S T R S T N
* Made this connection only when a motorised foot rest chest stand is present.
Terminal
Description Wire n° Terminal Description
X0
Remote Exposure
8 TB6.9
(Room Interface Board)
Preparation radiography
5 TB6.7
(Room Interface Board)
Common 0V
3 J6.2
0Vdc (Table Digital Interface board)
Common 0V
11 J6.23
0Vdc (Table Digital Interface board)
Fluoro kV
Man/Auto fluoro
manual/auto 10 J6.1
(Table Digital Interface board)
selection
Terminal
Description Wire n° Terminal Description
X0
Bucky 1 select
SFD selection 23 TB2.12
(Room Interface Board)
Bucky 2 select
DSI selection 24 TB2.1
(Room Interface Board)
X-ray ON
X-Ray on
from 22 J6.21
(Table Digital Interface board)
generator
Safety switch
Safety switch 35 J17.1
(Generator Interface PWB)
Safety switch
Safety switch 36 J17.2
(Generator Interface PWB)
NOTE:
During the auto calibration of the X-ray tube it is required to make the
jumper between J7.6 and J6.2 (force terminal J7.6 at 0Vdc).
It is important to remove the jumper at the end of the X-ray tube
auto calibration.
VISION DIVA-D
POD
Description Terminal Wire n° Description
pcb
Common
X0.3 11 0 V digital common
0Vdc
X-ray on
from X0.22 14 X-ray ON
generator
POD
Description TDI Wire n° Description
pcb
Terminals Pcb A1
Description Wire n° Description
X0 4 fields 3 fields
Maximum
Minimum field 16 X4.3 X4.2
zoom
Minimum
Maximum field 14 X4.1 N.C.
zoom
6.2.7 G 100 R/F-A – Three field solid state AEC device 1 (Gilardoni
or Comet type)
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
White
Start J1.8 Pink Common
Green
G100 R/F-A
AEC board AEC device 1
(solid state chamber) connecting cable
type 733347
G100 R/F-A
AEC board AEC device 1
(solid state chamber) connecting cable
type 737992
NOTE:
This hardware interface has to be made only if the Potter Bucky is
present.
WARNING:
The voltage selected on TB1.11 and TB1.12 is 24 Vdc. Refer to the Chest
Stand VD3 35x43 Technical and User's Manual for the Bucky start
voltage setting.
6.2.10 G 100 R/F-A – Three field solid state AEC device 2 (Gilardoni
or Comet type)
NOTE:
This hardware interface has to be made only if the Potter Bucky with
AEC device is present.
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
White
Start J2.8 Pink Common
Green
G 100 R/F-A
AEC board AEC device 2
(solid state chamber) connecting cable
type 733347
White
Anode J12.2 Pink Common
Green
G100 R/F-A
AEC board AEC device 2
(solid state chamber) connecting cable
type 737992
NOTE:
The configuration of the Jumpers not listed in the previous tables is not
relevant for this system integration.
Prior to beginning tube calibration, the tube used in the installation must
be properly selected, and generator limits should be programmed.
If the x-ray tube used in the installation is present in the library of the
generator refer to the G 100 R/F-A Service Manual, chapter 3C -
Programming the generator in order to record all the necessary
information to program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter-Programming the dual speed starter to record the necessary
information to complete the configuration.
If the X-ray tube used in the installation is not present in the library of
the generator should be present in the utility software. The data
communication with a computer it is allowed using the GenWare® utility
software. Refer to the G 100 R/F-A Service Manual, chapter 3C - DATA
LINK to record the necessary information.
6.3.1.3 X-ray tube not present neither in library nor in GenWare® utility
software
If the x-ray tube used in the installation is not present neither in the
generator library nor in the GenWare® utility software a manual setting
of the tube and generator parameters is required. Before changing these
parameters is strictly recommended to consult the data sheet of the x-ray
tube that is going to be installed. An X-ray tube present in the library or
in the GenWare® utility software, as similar as possible to the one is
going to be installed, can be selected. The tube and the generator
parameter required in the generator set-up must be set in according with
the information recorded from the data sheet of the x-ray tube. Refer to
the G 100 R/F-A Service Manual, chapter 3C - Tube selection and
Generator limits in order to record all the necessary information to
program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter - Programming the dual speed starter to record the necessary
information to complete the configuration.
The installation of the AEC devices and their calibration are covered in
the G 100 R/F-A Service Manual, chapter 3E - Automatic exposure
control (AEC) calibration.
NOTE:
The Fluoro set-up data must be set in according with the X-ray tube
technical specification.
The following tables meets the Fluoro set-up request in according with
most of the X-ray tube supplied by Villa Sistemi Medicali; during the
installation could be necessary to optimize the set-up.
The installation of the ABS device and his calibration are covered in the
G 100 R/F-A Service Manual, chapter 3 - Automatic brightness
stabilisation (ABS) calibration.
The next table show the standard parameters usually set.
FLUORO SET-UP
FL TIMER: 5 Min
MIN FLUORO KV: 40
MAX FLUORO KV: 120
II MODES: 0
FL ABS: ON
FL ABS DEFAULT: NONE
FL/RAD KV DEF: NONE
FL/RAD KV CURVES Not Used
SID COMP Not Used
FL SETUP FL ABS SETUP ABS LEVEL 0: 5.00 V
ABS LEVEL 1: 5.00 V
ABS LEVEL 2: 5.00 V
ABS LEVEL 3: 5.00 V
SIGNAL GAIN 10
ABS GAIN 10
ABS DEADBAND 3%
ABD DELAY 150 mS
ABS START KV 75
ABS CURVE DEF: 1
PF SETUP PF: ON
PF DEFAULT: NONE
PPS DEFAULT: NONE
PF DEFAULT MA: 5=40mA
6=60mA
7=80mA
8=99mA
Note: Four options for the settings of the Pulsed Fluoro default mA are possible.
In order to meet the code of federal regulations 21 paragraph 1020 it is suggested to
set the value 5 = 40mA as default parameter. If, once set at 40 mA the default mA
value for the pulsed fluoro, is necessary to reduce the dose is possible to reduce the
length of the pulses by means the settings of the jumpers on the A2 board of the I.I.
system (make reference to the I.I. system service manual)
PF DEFAULT Ms: Not
Used
PF SYNC FREQ: 50 HZ
PF ABS SETUP ABS LEVEL 0: 5.0 V
ABS LEVEL 1: 5.0 V
ABS LEVEL 2: 5.0 V
ABS LEVEL 3: 5.0 V
SIGNAL GAIN: 10
ABS GAIN: 8
ABS DEADBAND: 10%
ABS DELAY: 150 mS
ABS START KV: 75
PF INTEGRATOR: OFF
1PPS COMP: 0%
3PPS COMP: 0%
6/7.5PPS COMP: 0%
12/15PPS COMP: 0%
25/30PPS COMP: 0%
PF ABS CURVE DEF: NONE
PF ABS CURVES Not Used
PF DOSE LIMITS Not Used
PF II COMP Not Used
PF PPS SETUP 1 PPS: ON
3 PPS: ON
6/7.5 PPS: ON
12.5/15 PPS: ON
28/30 PPS: ON
HPF SETUP Not Enable
HLF SELECT: CONSOLE
PF SELECT: CONSOLE
NOTE:
Receptor 3 has to be programmed only if Chest Stand with Potter Bucky
and AEC device is present.
***INPUTS***
REMOTE EXP ----- OFF ----- ON -----
REMOTE PREP ----- ON ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- ON
CONSOLE EXP ----- OFF ----- OFF -----
CONSOLE PREP ----- OFF ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT ON ON ON ON ON
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF ON ON
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
***INPUTS***
REMOTE EXP ----- OFF ----- ON -----
REMOTE PREP ----- ON ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- ON
CONSOLE EXP ----- OFF ----- OFF -----
CONSOLE PREP ----- OFF ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT ON ON ON ON ON
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF ON OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- ON -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF ON OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
NOTE:
The configuration of the Receptor 3 I/O has to be made only if the Chest
Stand with Potter Bucky is present.
Room configuration
• Vision
– Under table X-ray tube (1)
– Floor mounted tubestand X-ray tube (2)
– Under table Potter Bucky with three field solid state AEC
– SFD with three field solid state AEC (Gilardoni or Comet type)
– Standard I.I. system with N23 CCD Camera.
• G 100 R/F-A
• Chest Stand VD3 35x43 with motorised foot rest (See Note)
– Three field solid state AEC (Gilardoni or Comet type).
Receptor configuration
NOTE:
When the Generator is supplied with "Falling load" feature, the GalKey
U29 # 73594213 is installed on CPU board.
The Falling Load option it is not available on the receptors where the
"SERIAL" exposure is enable (par. 6.3.4 - Image Receptor Programming
Worksheet: "SERIAL=YES"); whenever the Falling Load has to be
activated on the Spot Film Device receptor, the SERIAL parameter has to
be set to "NO".
In case Falling Load feature and SERIAL exposure are required will be
necessary to create for each a dedicated receptor.
NOTE:
Reference to the chest stand VD3 35X43 with motorized foot rest is made
in order to allows the installer to record the necessary information for the
hardware interface among G 100 R/F-A and Potter Bucky.
R R
S S
T T
N N
G 100 R/F-A
AUXILIARY CONTACTOR
R S T R S T N
* Made this connection only when a motorised foot rest chest stand is present.
Terminal
Description Wire n° Terminal Description
X0
Remote Exposure
8 TB6.9
(Room Interface Board)
Preparation radiography
5 TB6.7
(Room Interface Board)
Common 0V
3 J6.2
0Vdc (Table Digital Interface board)
Common 0V analog
11 J6.23
0VS (Table Digital Interface board)
Fluoro Kv
Man/Auto fluoro
manual/auto 10 J6.1
(Table Digital Interface board)
selection
NOTE:
During the auto calibration of the X-Ray tube it is required to make the
jumper between J7.6 and J6.2 (force terminal J7.6 at 0Vdc).
It is important to remove the jumper at the end of the X-Ray tube
auto calibration.
Terminal
Description Wire n° Terminal Description
X0
Bucky 1 select
SFD selection 23 TB2.12
(Room Interface Board)
X-ray ON
X-Ray on
from 22 J6.21
(Table Digital Interface board)
generator
Safety switch
Safety switch 35 J17.1
(Generator Interface PWB)
Safety switch
Safety switch 36 J17.2
(Generator Interface PWB)
0Vdc
(Table Digital Interface J7.2 X2.1 0Vdc
board)
Proportional DC
(Table Digital Interface J7.21 X2.2 Analog. Kv auto
board)
kV/mA auto/man
selection kV/mA auto/man
J6.1 X2.4
(Table Digital Interface selection
board)
Terminals Pcb A1
Description Wire n° Description
X0 4 fields 3 fields
Maximum
Minimum field 16 X4.3 X4.2
zoom
Minimum
Maximum field 14 X4.1 N.C.
zoom
Terminal
Description Wire n° Pcb A2 Description
X0
19 X3.4 Inv
18 X3.5 Inv
Terminal Terminal
Description Wire n° Description
X0 (Room Int. Board)
Bucky 2
X-ray request 49 TB2.2
select
Bucky 2
X-ray request 50 TB2.1
select
Bucky
Potter reply 47 TB2.5
contacts
Bucky
Potter reply 48 TB2.4
contacts
WARNING:
The voltage selected on TB2.1 and TB2.2 is 24 Vdc. Refer to the Vision
service manual, under table Potter Bucky voltage setting chapter for
more information.
7.2.5 G 100 R/F-A – SFD three field solid state AEC device 1
(Gilardoni or Comet type)
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface has to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
G100 R/F-A
AEC device 1
Universal AEC board
connecting cable
Type 734614
White
Start J1.8 Pink Common
Green
G100 R/F-A
AEC board AEC device 1
(solid state chamber) connecting cable
type 733347
G100 R/F-A
AEC board AEC device 1
(solid state chamber) connecting cable
type 737992
7.2.6 G 100 R/F-A – Vision under table Potter Bucky Three field
solid state AEC device 2 (Gilardoni or Comet type)
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface has to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
G100 R/F-A
AEC device 2
Universal AEC board
connecting cable
Type 734614
White
Start J2.8 Pink Common
Green
G100 R/F-A
AEC board AEC device 2
(solid state chamber) connecting cable
type 733347
G100 R/F-A
AEC board AEC device 2
(solid state chamber) connecting cable
type 737992
NOTE:
This hardware interface has to be made only if the Potter Bucky is
present.
Terminal
Description Terminal Wire n° Description
(Room Int. Board)
Bucky 3
X-ray request X1.B1 TB1.11
select
Bucky 3
X-ray request X1.B2 TB1.12
select
Bucky
Potter reply X1.X TB2.5
contacts
Bucky
Potter reply X1.X TB2.4
contacts
WARNING:
The voltage selected on TB1.11 and TB1.12 is 24 Vdc. Refer to the Chest
Stand VD3 35x43 Technical and User's manual for the Bucky start
voltage setting.
7.2.9 G 100 R/F-A – Three field solid state AEC device 3 (Gilardoni
or Comet type)
NOTE:
This hardware interface has to be made only if the Potter Bucky with
AEC device is present.
NOTE:
If the AEC interface PWB it is not supplied with LEMO connectors the
following hardware interface have to be done. The installer should cut the
AEC cable by the PWB side connector and connect the colored wires to
the terminals on the Universal AEC board making reference to the next
table.
G100 R/F-A
AEC device 3
Universal AEC board
connecting cable
Type 734614
White
Start J3.8 Pink Common
Green
G 100 R/F-A
AEC board AEC device 3
(solid state chamber) connecting cable
type 733347
White
Anode J13.2 Pink Common
Green
G100 R/F-A
AEC board AEC device 3
(solid state chamber) connecting cable
type 737992
– TB8.1 – TB11.K3 (This wiring must be made only if the 24Vdc under
table Potter Bucky is present on Receptor 2)
– TB8.2 – TB11.K2 (This wiring must be made only if the 24Vdc Chest
stand Potter Bucky is present on Receptor 3)
– TB4.8 – TB4.9: Thermal switch
– TB4.4 – TB4.5: Room door interlock
– TB4.11 – TB4.12: Room light.
NOTE:
The configuration of the Jumpers not listed in the previous tables is not
relevant for this system integration.
Prior to beginning tube calibration, the tube used in the installation must
be properly selected, and generator limits should be programmed.
If the x-ray tube used in the installation is present in the library of the
generator refer to the G 100 R/F-A Service Manual, chapter 3C -
Programming the generator in order to record all the necessary
information to program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter-Programming the dual speed starter to record the necessary
information to complete the configuration.
NOTE:
The tube library, concerning the I.A.E. X-ray tubes, contains only high
speed tube types. When a low speed I.A.E. X-ray tube has to be installed
an high speed tube must be selected informing the generator about the
low speed tube. The generator and the tube limits must be set in
according with the tube type installed.
If the X-ray tube used in the installation is not present in the library of
the generator should be present in the utility software. The data
communication with a computer it is allowed using the GenWare® utility
software. Refer to the G 100 R/F-A Service Manual, chapter 3C - DATA
LINK to record the necessary information.
7.3.1.3 X-ray tube not present neither in library nor in GenWare® utility
software
If the x-ray tube used in the installation is not present neither in the
generator library nor in the GenWare® utility software a manual setting
of the tube and generator parameters is required. Before changing these
parameters is strictly recommended to consult the data sheet of the x-ray
tube that is going to be installed. An X-ray tube present in the library or
in the GenWare® utility software, as similar as possible to the one is
going to be installed, can be selected. The tube and the generator
parameter required in the generator set-up must be set in according with
the information recorded from the data sheet of the x-ray tube. Refer to
the G 100 R/F-A Service Manual, chapter 3C - Tube selection and
Generator limits in order to record all the necessary information to
program the generator set-up.
The configuration of the starter board it is requested too. Refer to the
G 100 R/F-A Service Manual, chapter 2 - Programming the low speed
starter - Programming the dual speed starter to record the necessary
information to complete the configuration.
The installation of the AEC devices and their calibration are covered in
the G 100 R/F-A Service Manual, chapter 3E - Automatic exposure
control (AEC) calibration.
NOTE:
The Fluoro set-up data must be set in according with the X-ray tube
technical specification.
The following tables meets the Fluoro set-up request in according with
most of the X-ray tube supplied by Villa Sistemi Medicali; during the
installation could be necessary to optimize the set-up.
The installation of the ABS device and his calibration are covered in the
G 100 R/F-A Service Manual, chapter 3 - Automatic brightness
stabilisation (ABS) calibration.
The next table show the standard parameters usually set.
FLUORO SET-UP
FL TIMER: 5 Min
MIN FLUORO KV: 40
MAX FLUORO KV: 120
II MODES: 0
FL ABS: ON
FL ABS DEFAULT: NONE
FL/RAD KV DEF: NONE
FL/RAD KV CURVES Not Used
SID COMP Not Used
FL SETUP FL ABS SETUP ABS LEVEL 0: 1.00 V
ABS LEVEL 1: 1.60 V
ABS LEVEL 2: 1.10 V
ABS LEVEL 3: 1.60 V
SIGNAL GAIN 10
ABS GAIN 20
ABS DEADBAND 3%
ABD DELAY 150 mS
ABS START KV 75
ABS CURVE DEF: 1
NOTE:
The Pulsed Fluoro set-up is not relevant for this system integration
because the function is not available with the N23 CCD camera.
***INPUTS***
REMOTE EXP ----- OFF ----- ON -----
REMOTE PREP ----- ON ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- ON
CONSOLE EXP ----- OFF ----- OFF -----
CONSOLE PREP ----- OFF ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 ON ON ----- ON ON
THERMAL SW 2 OFF OFF ----- OFF OFF
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT ON ON ON ON ON
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- ON -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 OFF OFF ----- OFF OFF
THERMAL SW 2 ON ON ----- ON ON
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF ON OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
NOTE:
The configuration of the Receptor 2 I/O has to be made only if the Vision
under table Potter Bucky is present or an external Potter Bucky has to be
connected on Receptor 2, in example, the Chest Stand.
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- ON -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 OFF OFF ----- OFF OFF
THERMAL SW 2 ON ON ----- ON ON
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF ON OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF
NOTE:
The configuration of the Receptor 3 I/O has to be made only if the Chest
Stand with Potter Bucky is present.
***INPUTS***
REMOTE EXP ----- OFF ----- OFF -----
REMOTE PREP ----- OFF ----- ----- -----
REMOTE FL. EXP ----- ----- ----- ----- OFF
CONSOLE EXP ----- OFF ----- ON -----
CONSOLE PREP ----- ON ----- ----- -----
TOMO EXP ----- ----- ----- OFF -----
REM. TOMO SEL OFF ----- ----- ----- -----
I/I SAFETY ----- OFF ----- OFF OFF
COLL. ITLK ----- OFF ----- OFF OFF
BUCKY CONTACTS ----- ----- ----- OFF -----
SPARE OFF OFF ----- OFF OFF
THERMAL SW 1 OFF OFF ----- OFF OFF
THERMAL SW 2 ON ON ----- ON ON
DOOR ITLK OFF ON ----- ON ON
MULTI SPOT EXP OFF ----- ----- ----- -----
***OUTPUTS***
BKY 1 SELECT OFF OFF OFF OFF OFF
BKY 2 SELECT OFF OFF OFF OFF OFF
BKY 3 SELECT OFF OFF OFF OFF OFF
TOMO/BKY 4 SEL OFF OFF OFF OFF OFF
TOMO/BKY START OFF OFF OFF OFF OFF
ALE OFF OFF OFF OFF OFF
COLL. BYPASS OFF OFF OFF OFF OFF
ROOM LIGHT OFF ON ON ON ON
SPARE OFF OFF OFF OFF OFF