1. Function Test:
1.1 Mainboard tested with AADI Smartguard Test Program. 489 tests.
1.2 NA
1.3 NA
3. Current drains:
3.1 Current drain display on (max 120 mA)
(LAN disabled, recorder off, only power cable)
3.2 Current drain display off (max 30 mA).
(display off, LAN disabled, recorder off, only power cable)
3.3 Current drain configuration (max 35 mA).
(USB connected, display off)
3.4 Current drain in sleep before climate test (max 1.2 mA).
(recording on, display off, only power cable)