You are on page 1of 1

Form No.

833, February 2012

Product:SmartGuard Basic 5100


Serial No:850 Image Version:2.1.168

Mainboard 3969691 Serial No 850

1. Function Test:
1.1 Mainboard tested with AADI Smartguard Test Program. 489 tests.
1.2 NA
1.3 NA

2. Visual and Mechanical tests:


2.1 Visual inspection of display image.
2.2 Test of keypad.
2.3 Connector housing to chassis connection test.
2.4 Chassis isolation from signal ground.

3. Current drains:
3.1 Current drain display on (max 120 mA)
(LAN disabled, recorder off, only power cable)
3.2 Current drain display off (max 30 mA).
(display off, LAN disabled, recorder off, only power cable)
3.3 Current drain configuration (max 35 mA).
(USB connected, display off)
3.4 Current drain in sleep before climate test (max 1.2 mA).
(recording on, display off, only power cable)

4. Climatic test of instrument:


4.1 Climatic test of instrument -40 to 60 degC. No recording errors.
4.2 Current drain in sleep after climate test (max 1.2 mA).
(recording on, display off, only power cable)

Date: 03 Feb 2021 Sign: