QuadraSCAN - OPERATOR’S MANUAL

QuadraSCAN™
Scanner System For QuadraLASE System

OPERATOR’S MANUAL

MANUFACTURER:

CANDELA CORPORATION
530 Boston Post Road,
Wayland, MA 01778-1886, USA

Tel. 508-358-7400
Fax. 508-358-5569
www.candelalaser.com

REVISIONS
Number 0 1 2 3 4 5
Date 16.02.2009
Issue RAQ
Verification and Approval
DIR

Page 1 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 01 of 24
QuadraSCAN - OPERATOR’S MANUAL

CONTENTS

INTRODUCTION page 4

1.0 Introduction page 4


1.1 QuadraSCAN presentation page 4

SPECIFICATIONS page 5

2.0 Technical data page 5

SAFETY page 6

3.0 Classification page 6


3.1 Precautions page 6
3.2 Environmental considerations page 6
3.3 Accidental exposure page 7
3.4 Reflection hazards page 7
3.5 Explosion and fire risk page 7
3.6 Safety labels page 8

INSTALLATION page 9

4.0 Precautions page 9


4.1 QuadraSCAN installation page 9
4.2 Handpiece insert page 12

CONTROLS page 13

5.0 General description page 13


5.1 Buttons description page 13
5.2 Alarm messages page 14

OPERATION page 15

6.0 Operation page 15


6.1 Function selection and value setting page 16
6.2 SX Fractional page 17
6.3 Exposure control page 18
6.4 Scanning mode page 19

MAINTENANCE page 20

7.0 Service information page 20


7.1 Routine servicing page 20
7.2 External cleaning / disinfection page 21
7.3 Cleaning of optical components page 21
7.4 Sterilization page 22

TROUBLESHOOTING page 23

8.0 General page 23


8.1 Troubleshooting guide page 23

ANNEX "A" page 25

Installation check list

ANNEX "B" page 26

QUADRASCAN block wiring diagram

Page 2 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 02 of 24
QuadraSCAN - OPERATOR’S MANUAL

INTRODUCTION

1.0 - INTRODUCTION -

Congratulation for purchasing the QuadraLASE system and for the confidence placed in the Candela brand.

Candela Corporation manufactures and distributes innovative clinical solutions that enable
physicians, surgeons, and personal care practitioners to treat selected cosmetic and medical
conditions using lasers, aesthetic laser systems, and other advanced technologies. Founded near
Boston in 1970, Candela markets and services its products in 86 countries from offices and
distributors in the United States, Europe, Japan, China and other Asian locations. Candela
established the aesthetic laser market 20 years ago, and currently has an installed base of over
13,500 systems worldwide. Visit Candela on the Web at http://www.candelalaser.com

1.1 – QUADRASCAN PRESENTATION -

QuadraSCAN is a scanner accessory unique and proprietary to the QuadraLASE laser system.
QuadraSCAN delivers in Fractional mode a precise matrix of micro-spots that penetrate to the dermis and
stimulate the formation of new collagen. At the same time, immediate shrinkage of damaged tissue is
achieved. The flexibility of QuadraSCAN delivers unequalled results with an advanced method of Fractional
Skin Resurfacing and at the same time offering the option to use conventional ablation methods.
The QuadraSCAN accessory consists of an electronic component installed inside the appliance and a
scanning handpiece connected at the end of the articulated.

Page 3 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 03 of 24
QuadraSCAN - OPERATOR’S MANUAL

SPECIFICATIONS

2.0 - TECHNICAL DATA -

WAVELENGTH 10.6 micrometers


REFLECTIVITY > 98%
NOMINAL SCAN ANGLE 0.21 rad
GAIN ERROR < 5 mrad
ZERO OFFSET < 5 mrad
NONLINEARITY < 4.2 mrad
TRACKING ERROR AT 5 RAD/S < 1.5 mrad
REPEATABILITY < 50 urad
HANDPIECE Anodized aluminium handpiece
SCAN PATTERN Line, Rectangle, Square, Triangle, Circle,
Fractional Square (QL), Donut
SCAN PATTERN SIZE Spot 300um: from 6x6 to 20x20 mm increments of 2 mm
Spot 180um: from 5x5 to 12x12 mm increments of 1 mm
DWELL TIME CW: 0.1 to 0.8 msec, increments of 0.1 msec
Fractional mode: 8 step from 2.5 to 16 msec
SPOT DIAMETER Two handpieces : QS (300 um), QD (180um)
DENSITY Traditional mode: 100% of area coverage
Fractional “QL”: from 5 to 40% of area coverage
increments of 5%
OPERATING TEMPERATURE 25°C +/- 10°C
STORAGE TEMPERATURE -35°C to +60°C
HANDPIECE APPROX. WEIGHT 300g

Page 4 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 04 of 24
QuadraSCAN - OPERATOR’S MANUAL

SAFETY

3.0 - CLASSIFICATION -

QuadraSCAN is an accessory of the QuadraLASE laser system for surgical and therapeutically applications.
The QuadraSCAN is designed, manufactured and safeguarded specifically for use with the QuadraLASE
laser system.

WARNING !

AS THESE CONDITIONS ARE THE OUTCOME OF THE INITIAL EVALUATION OF THE MACHINE AND
CONSEQUENT RISKS, A DIFFERENT USE COULD CAUSE DAMAGES TO THE EQUIPMENT AND BE A
HAZARD FOR THE OPERATORS, PATIENTS AND THIRD PARTIES. THEREFORE THE NON-
OBSERVANCE OF THE TECHNICAL PARAMETERS AND THE INSTRUCTIONS GIVEN IN THIS
HANDBOOK COULD CAUSE ABNORMAL RUNNING CONDITIONS WHICH MAY JEOPARDISE THE
MACHINE SAFETY AND HEALTH OF THE OPERATOR AND PATIENT.

It is the operator’s responsibility to comply with the listed precautions in order to guarantee his own safety
and that of the patient.

QuadraLASE in combination with QuadraSCAN must be used by only medical personnel with
specific skill in procedures requiring the use of the CO2 laser system.

The QuadraLASE laser system and accessories comply with the ISO60601-1-1-2 for Medical Devices (Class
IIb).

3.1 - PRECAUTIONS -

CAUTION:
THE FOLLOWING MEASURES MUST BE OBSERVED TO GUARANTEE THE SAFETY OF PERSONS
WITHIN THE PHYSICAL OPERATIONAL LIMITS OF THE SYSTEM.

CAUTION:
THE SYSTEM SHOULD BE KEPT LOCKED. THE KEY SHOULD BE ENTRUSTED TO ONLY
AUTHORIZED PERSONNEL.

CAREFULLY READ THIS MANUAL

3.2 - ENVIRONMENTAL CONSIDERATIONS -

The equipment must be utilized in a room or in an area where the access door displays the sign: Laser
Controlled Area (Fig. 3.1).

WARNING!
ACCESS TO THE AREA, DURING OPERATION OF THE APPLIANCE, MUST BE RESTRICTED TO
ONLY AUTHORIZED PERSONNEL.

Page 5 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 05 of 24
QuadraSCAN - OPERATOR’S MANUAL

CAUTION:
IN THE EVENT WHERE MORE THAN ONE LASER SYSTEM IS WORKING SIMULTANEOUSLY IN THE
SAME AREA, LASER EMISSIONS SHOULD BE ISOLATED BY MOBILE OPAQUE BARRIERS THAT
BLOCK THE WAVELENGTH OF THE RELEVANT LASER SOURCES.

In view of the fact that light radiation entering the eye is attenuated in part by the response of the pupil, it is a
good policy to always operate lasers in a well-lit environment.

CAUTION:
THE ROOM OR AREA MUST BE WELL LIT TO CONSTRICT THE PUPILS TO A MINIMUM, THEREBY
REDUCING THE AMOUNT OF LIGHT ENTERING THE EYES.

In according with the EMC regulation CEI EN 60601-1-2, QuadraLASE is intended for use in the
electromagnetic environment as specified in the ANNEX “C” (Table 201, 206) of QuadraLASE Instruction
Manual.

3.3 - ACCIDENTAL EXPOSURE -

Because of the high energy emitted by the CO2 source, there is always a potential risk of damage to the
body and eyes.

WARNING!
NEVER LOOK STRAIGHT INTO THE APERTURE FROM WHICH THE LASER BEAM IS EMITTED.
OPERATOR AND PATIENT MUST ALWAYS WEAR PROTECTIVE GOGGLES WHICH PROTECT
AGAINST THE WAVELENGTH OF THE LASER SOURCE USED IN COMPLIANCE WITH DIRECTIVE
89/686/EEC (STANDARD EN207). THE CO2 LASER MUST ONLY BE DIRECTED AT THE
TREATMENT ZONE AFTER THE REQUISITE FUNCTIONS ON THE CONTROL PANEL HAVE BEEN
SELECTED / CONFIRMED, AND THE CORRECT FOCUSING HAS BEEN ACCURATELY DETERMINED.

CAUTION:
LASER PLUME MAY CONTAIN VIABLE TISSUE PARTICULATES. THE USE OF A SMOKE EVACUATION
SYSTEM IS RECOMMENDED.

3.4 - REFLECTION HAZARDS -

CO2 laser emission can be reflected by many surfaces other than those normally considered reflective. It is
recommend that all metal objects such as watches, rings, necklaces etc. be removed from the active laser
field. Only surgical instruments designed for use with CO2 lasers should be used in an active laser field.
These instruments should be matt or with a burnished finish.

CAUTION:
THERE ARE MANY METALLIC SURFACES WHICH APPEAR MATT TO THE NAKED EYE, BUT ARE
STILL ABLE TO REFLECT INFRARED RADIATION.

3.5 - EXPLOSION AND FIRE RISK -

Many non-metallic materials will ignite when exposed to CO2 laser emission. Plastic, paper and wood
objects should be removed (as far as possible) from the active laser field.

WARNING!
NEVER USE THE CO2 LASER SYSTEM NEAR FLAMMABLE ANAESTHETICS AND/OR FLAMMABLE
DISINFECTANTS.

Never use plastic instruments for laser surgery, or if unavoidable, shield them with aluminium foil.

Page 6 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 06 of 24
QuadraSCAN - OPERATOR’S MANUAL

If the patient's head or face is to be treated, protect the patient's eyes with water soaked gauze, or with
protective goggles. Water highly absorbs CO2 laser emission and provides an extremely effective shield.
Whenever it is necessary to shield the body from laser radiation, it is recommended to cover the area with
water soaked gauze.

CAUTION:
FOR ANY AND ALL MATTERS RELATING TO SAFETY, PLEASE FOLLOW THE RECOMMENDATIONS
AND REGULATIONS ISSUED BY SAFETY AUTHORITIES RESPONSIBLE FOR THE CORRECT USE OF
LASER SYSTEMS.

Fig. 3.1

3.6 - SAFETY LABELS -

LABELS AND SAFETY WARNINGS ON THE EQUIPMENT ARE INDICATED IN THE QuadraLASE
INSTRUCTION MANUAL CHAPTER 3.
LABELS AND SAFETY WARNINGS MUST BE KEPT PERFECTLY LEGIBLE, AND REPLACED
IMMEDIATELY IF DETACHED OR FADED.

LASER APERTURE

A) This label identifies the exit point of the laser radiation beam (to be placed on the Scanner handpiece).

Page 7 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 07 of 24
QuadraSCAN - OPERATOR’S MANUAL

INSTALLATION

4.0 - PRECAUTIONS -

WARNING!
ONLY PERSONNEL EXPRESSLY AUTHORIZED BY CANDELA CORP. SHOULD INSTALL THE
SYSTEM.

A) Before switching on QuadraLASE it is recommended to wait about 3 hours at a temperature higher than
15 degrees (Celsius).This time guarantees the optimum working conditions for the appliance.

B) For the same reason, QuadraLASE must be used with an indoor temperature higher than 15
degrees (Celsius) and relative humidity 45 - 75 %.

C) Do not obstruct the air intakes of the QuadraLASE. Intake distance from obstacles (such as a wall)
should be at least 30 cm (12").

4.1 – QuadraSCAN INSTALLATION -

WARNING!
DURING QuadraSCAN INSTALLATION QuadraLASE MUST BE SWITCHED-OFF.
- Prepare the articulated arm:

1) Release the articulated arm from its clip and hold the arm with your hands.
2) Rotate the articulated arm 180° around the main bearing axis until the balancing spring is
coupled.
3) Place the arm in the "rest" position.

180°

Stowing
clip

Fig. 4.1

• Detach the plastic tube (3 Fig. 4.2) from its fitting on the QuadraLASE head (4 Fig. 4.2)
• Unscrew the articulated arm (1 Fig. 4.2) from the QuadraLASE (2 Fig. 4.2)

Page 8 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 08 of 24
QuadraSCAN - OPERATOR’S MANUAL

4
2

Fig. 4.2

• Remove the QuadraSCAN Fractional accessory from the storage case and screw it onto the
QuadraLASE head (2 Fig. 4.3)
• Connect the self-aligning coupling side of the articulated arm to the QuadraSCAN Fractional accessory
(1 Fig. 4.3).
• Lock the metal ring pushing the safety switch until the articulated arm is attached to the equipment.
• Reconnect the plastic tube to its fitting on the laser head (3-4 Fig. 4.3).

4
2

Fig. 4.3

• Screw the scanner handpiece (1 Fig. 4.4) onto the end of the articulated arm (2 Fig. 4.4) and attach the
scanner handpiece connector (4 Fig. 4.4) to the scanner cable connector (3 Fig. 4.4)

Page 9 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 09 of 24
QuadraSCAN - OPERATOR’S MANUAL

Fig. 4.4

• Attach the Scanner handpiece cable and the QuadraSCAN cable to the rear panel receptacles (Fig. 4.5)

Scanner cable
Fractional cable

Scanner
cable

Fractional
cable

Fig. 4.5

CAUTION:
THERE IS A SAFETY INTERLOCK IN THE LASER HEAD TO PREVENT LASER EMISSION WHEN THE
ARTICULATED ARM IS REMOVED.

Page 10 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 10 of 24
QuadraSCAN - OPERATOR’S MANUAL

WARNING!
HANDLE THE ARTICULATED ARM WITH EXTREME CARE AND AVOID ANY KIND OF MECHANICAL
SHOCK THAT COULD CAUSE OPTICAL MISALIGNMENT. OPTICAL MISALIGNMENT WILL
ADVERSLEY COMPROMISE THE PERFORMANCE OF THE LASER

4.2 – HANDPIECE INSERT -

The handpiece insert is mounted with pressure on the handpiece body having care to push well the insert
untill the end race. For remove the insert is sufficient to pull it toward the outside (Fig. 4.6).

300um 180um
QS QD

Fig. 4.6

WARNING!
MOUNTING THE HANDPIECE INSERT CHECK WELL TO HAVE IT FIRMLY PUSHED UNTIL SEATED.

CONTROLS
Page 11 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194
8085-50-9194 A Candela Corp Proprietary Sheet 11 of 24
QuadraSCAN - OPERATOR’S MANUAL

5.0 - GENERAL DESCRIPTION -

On the QuadraLase laser system, the controls are grouped together on a single panel located on the top of
the console. On the control panel, there are six pushbuttons and a LCD monitor displaying the function
parameters.

WARNING!
USE OF CONTROLS, ADJUSTMENTS OR PERFORMANCE OF PROCEDURES OTHER THAN THOSE
SPECIFIED IN THIS MANUAL MAY RESULT IN HAZARDOUS RADIATION EXPOSURE.

Laser Messages
Emission

E mi s s i o n
Of f
S ca n n e r Sel 0
Operative CW O p e r a te Con t .

Section Sup e r p S ta n d b y S i n g l e
Pu l se r Re p e a t Sel 1

A i mi n g
30
P o we r.
1000
T i me
B e a m ( W ) Cal ( ms ) Sel 2

Down Up Enter

Fig. 5.1 Control panel

NOTE: Refer to the QuadraLASE Instruction Manual for a detailed description of all controls.

5.1 - BUTTON FUNCTIONS -

“SEL 0” BUTTON

When in the QuadraLASE standard screen mode, “Scanner” label is displayed above the SEL 0 button on
the LCD monitor. When the button is pressed, the LCD monitor screen changes to display the control
parameters for the scanner accessory.
When the button is pressed again, in the scanner screen mode, the black indicator bar under functions “S R
B” blinks. Continue pressing the button until the desired parameter is selected - Single “S”, Repeat “R” and
Beam “B”.

“SEL 1” BUTTON

Page 12 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 12 of 24
QuadraSCAN - OPERATOR’S MANUAL

When the SEL 1 button is pressed, the indicator rectangle located on the shape display area blinks.
Continue pressing the button to move the blinking indicator rectangle to the desired shape.

“SEL 2” BUTTON

The SEL 2 button is pressed to select either the power or time function (Fig. 5.1).

“ENTER” BUTTON

The ENTER button is pressed to confirm selected function “SRB” or the selected shape and size. After
confirmation the black indicator bars stop blinking. After all parameter settings are confirmed, the Enter
button is used to select either “Stand by” or “Operate” mode.
When the POWER function is set to 0.0 (CO2 laser source is OFF) it is not possible to select "OPERATE".

“UP” BUTTON

The UP button increases the shape size (size indicator bars enlarge), increases the power setting and
increases the time setting (Fig. 5.1).

“DOWN” BUTTON

The DOWN button decreases the shape size (size indicator bars reduce), decreases the power setting and
decreases the time setting (Fig. 5.1).

5.2 - ALARM MESSAGES -

IF ANY OF THE ALARMS ARE TRIGGERED, THE APPLIANCE AUTOMATICALLY SWITCHES INTO
"STAND BY" MODE. OPERATIONAL FUNCTIONS CANNOT BE SET UNTIL THE CAUSE OF THE
ALARM HAS BEEN REMOVED.

For Alarm Messages refer to the QuadraLASE Instruction Manual.

OPERATION

Page 13 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 13 of 24
QuadraSCAN - OPERATOR’S MANUAL

6.0 - OPERATION -

WARNING!
FOR ALL OPERATIVE PROCEDURES REFER TO QuadraLASE INSTRUCTION MANUAL.

Refer to the QuadraLASE Instruction Manual for system switch-on and operation.

After the QuadraLASE initial check procedure is complete, the LCD monitor shows the graphic display (Fig.
6.1)

Fig. 6.1

WARNING!
IF THE FRACTIONAL SWITCH IS IN THE “ON” POSITION, THE “INTERLOCK” ALARM WILL BE
ACTIVATED. TO REMOVE THE ALARM, MOVE THE FRACTIONAL SWITCH TO THE “OFF” POSITION.

Fractional
switch

Fig. 6.2

Press SEL 0 to go into the scanner screen mode and to start the scanner function test.

Page 14 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 14 of 24
QuadraSCAN - OPERATOR’S MANUAL

QL

Fig. 6.3

The check procedure lasts about 10 seconds. During this time, the message line on the LCD displays
“Check - please wait….” and the scanner handpiece output is a visible blinking red square.
During the check procedure, the control panel is deactivated and operation is not possible.
If the check is not passed, the LCD monitor displays the message “Scanner Alarm” and the system must be
switched off. If the problem persists, contact CANDELA CORP. or authorized CANDELA CORP. personnel.

6.1 - FUNCTION SELECTION AND VALUE SETTING -

All operations are followed step by step with messages displayed on the LCD monitor.

Function selection: “S” Single, “R” Repeat, “B” Beam

1. Press button SEL 0 to move the blinking black indicator bar to select the desired function – “SRB”. The
message line displays “ Enter to confirm”.
2. Press the ENTER button to confirm the selection. The black indicator bar stops blinking.

SHAPE and SIZE setting

1. When the SEL 1 button is pressed, the rectangle indicator in the shape area blinks. Continue pressing
the button to move the blinking rectangle indicator to the desired shape.
2. When the blinking rectangle indicator is positioned on the desired shape, press the UP or DOWN button
to increase or decrease the Shape size. The Size black indicator bars increase with the Up button and
decrease with the down button.
3. When all indicator bars are white and the DOWN button is pressed, the label “spot” is displayed on the
left side of the LCD. In this condition, the scanner is deactivated and a single spot output from the
scanner handpiece is set. In Fractional “QL” mode Beam function is not activated.
4. Press the button ENTER to confirm the Shape and Size selection. The shape indicator rectangle stops
blinking.

WARNING!
THE SELECTION WILL BE ACTIVATED ONLY IF CONFIRMED WITH THE ENTER BUTTON.

WARNING!
IF THE FRACTIONAL SWITCH IS IN THE “ON” POSITION AFTER SELECTING A STANDARD SHAPE,
THE “INTERLOCK” ALARM WILL BE ACTIVATED. TO REMOVE THE ALARM, MOVE THE
FRACTIONAL SWITCH TO THE “OFF” POSITION.

Page 15 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 15 of 24
QuadraSCAN - OPERATOR’S MANUAL

POWER and TIME value setting

1. Press button SEL 2 to select either the Power or Time function. The message line will display which
function has been selected.
2. Press the button UP or DOWN to increase or decrease the value of the selected function.

6.2 – QL FRACTIONAL -

When the Fractional “QL” function is selected, the “TIME” parameter and also the Density “D%” parameter
appear. In Fractional “QL” mode also the handpiece insert 180um can be used.

WARNING!
WHEN THE “QL” FRACTIONAL FUNCTION IS SELECTED, THE
FRACTIONAL SWITCH NEEDS TO BE IN THE “ON” POSITION.
.

QL

16
TIME
(msec)

Fig. 6.4

QL TIME value setting

1. Press button SEL 2 to move the double rectangle over TIME label. The message line will display “TIME –
Up / Down”.
2. Press the UP or DOWN button to increase or decrease the value of the TIME.

The TIME steps are: 2.5, 3, 3.5, 4, 5, 8, 12, 16 msec.

DENSITY D% value setting

1. Press button SEL 2 to move the double rectangle over D% label. The message line will display
“Density – Up / Down”.
2. Press the UP or DOWN button to increase or decrease the value of the Density

D% indicate the percentage of area covered during the scanning. This parameter can be set from 5 to 40%
with step of 5%.

SPOT Diameter setting

Page 16 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 16 of 24
QuadraSCAN - OPERATOR’S MANUAL

1. Press button SEL 1 to move the rectangle indicator over “QL” for select Fractional QL mode and
press ENTER pushbutton to confirm.
2. The “SPOT” label blinking and the message line will display “ Spot – Up / Down”.
3. Press the UP or DOWN button to set the spot diameter.
4. Press ENTER pushbutton to confirm and SPOT label will come fix.

Every time that the pattern dimension is changed at the end of the operation the additional confirm of the
SPOT dimension is required.

WARNING !
The QD 180um handpiece insert can be used and selected only in
Fractional QL mode. The moving of selection from Fractional QL to any
other traditional pattern it will generate automatic request to mount QS
300um handpiece insert in case QD 180um handpiece insert had been
selected before.

6.3 - EXPOSURE CONTROL -

Refer to the QuadraLASE Instruction Manual for the procedure to activate laser emission towards the
patient.

“S” SINGLE (Single scanning)


Depressing the footswitch (if "OPERATE" is selected), activates a single CO2 laser scan emission towards
the patient.
Emission will stop automatically upon completion of the scan or when the footswitch is released. To activate
another scan press the footswitch.

“R” REPEAT (Repeat scanning)


Depressing the footswitch (if "OPERATE" is selected) activates a series of CO2 laser scan emissions
towards the patient. There is a fixed pause of 1000 msec between each scan. Emission stops when the
footswitch is released.

“B” BEAM
When Function “B” is selected and confirmed with the Enter button, the software commands the laser to
display the standard operative screen.

CO2 LASER EMISSION ACTIVATES A CONTINUOUS ACOUSTIC SIGNAL AND THE MESSAGE LINE ON
THE LCD MONITOR DISPLAYS “EMISSION ON”.

CAUTION:
AFTER FIVE MINUTES IN THE “OPERATE” MODE WITHOUT ANY ACTION ON THE CONTROL PANEL
OR FOOTSWITCH, THE SYSTEM AUTOMATICALLY GOES INTO "STAND BY" MODE.

Note: Laser exposure can be terminated at any time by turning the laser off or activating the Emergency
switch.

WARNING!
DURING CO2 LASER RADIATION EMISSION, THE CONTROL PANEL IS COMPLETELY LOCKED AND
ONLY THE LASER OFF FUNCTION IS ENABLED.

Page 17 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 17 of 24
QuadraSCAN - OPERATOR’S MANUAL

When the QL Fractional function is select with the Fractional switch in the OFF position, the Scanner Alarm
will engage when the footswitch is depressed (if "OPERATE" is selected). To remove the alarm, move the
Fractional switch to the “ON” position.

WARNING!
WHEN THE “QL” FRACTIONAL FUNCTION IS SELECTED, THE FRACTIONAL SWITCH NEEDS TO BE
IN THE “ON” POSITION BEFORE SELECTING “OPERATE” MODE.

6.4 – SCANNING MODE -

The flexibility of QuadraSCAN delivers unequalled results with an advanced method of Fractional Skin
Resurfacing and at the same time offering the option to use conventional ablation methods.

Traditional scanning procedure


For all shapes (except QL during the scanning procedure, the CO2 laser beam fills the entire area in a
sequential mode. Emission will stop automatically upon completion of the scan or when the footswitch is
released. In scanning Repeat, the emission stops when the footswitch is released.

QL Fractional scanning procedure


When the “QL” Fractional function is selected, a proprietary algorithm controls the scan of a precise matrix of
Microspots. In Fractional mode, the Microspots penetrate to the dermis and stimulate the formation of new
collagen. At the same time, immediate shrinkage of damaged tissue is achieved.

MAINTENANCE

Page 18 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 18 of 24
QuadraSCAN - OPERATOR’S MANUAL

7.0 - SERVICE INFORMATION -

Refer to the QuadraLASE laser unit maintenance instructions in the QuadraLASE Instruction Manual.
Hospital/clinic staff may carry out only specified maintenance procedures. All remaining procedures must be
performed only by CANDELA CORP. or authorized CANDELA CORP. personnel.

WARNING!
ANY UNAUTHORIZED SERVICE OR MODIFICATION TO THIS UNIT CAN CAUSE SERIOUS INJURY
AND MAY INVALIDATE THE CANDELA CORP. SERVICE WARRANTY AGREEMENT.

The QuadraLASE laser unit generates high voltages within the main cabinet. Maintenance and
repairs inside the main cabinet are to be done only by CANDELA CORP. or authorized CANDELA
CORP. personnel.

All correspondence with your CANDELA CORP. representative, regarding the unit, should include the model
and serial numbers appearing on the identification plate located on the rear panel.

7.1 - ROUTINE SERVICING -

The QuadraSCAN accessory should be periodically inspected and serviced to maintain optimum
performance.

CHECK FOR PROPER BEAM ALIGNMENT

For this inspection, refer to the QuadraLASE Instruction Manual.

FREQUENCY Before every operation


PERFORMED BY Operator or Staff

GENERAL INSPECTION OF THE SCANNER HANDPIECE

Inspect the general condition of the scanner handpiece - check the optics for dust, smudges and scratches.
Except for dust and smudges on the optics, refer alignment and operational problems to CANDELA CORP.
or authorized CANDELA CORP. personnel.

FREQUENCY Before each operation


PERFORMED BY Operator or Staff

EXTERIOR SYSTEM INSPECTION

Inspect the system exterior to ensure that electrical connections between QuadraSCAN and QuadraLASE
are not loosened or damaged.

FREQUENCY Before each operation


PERFORMED BY Operator or Staff

WARNING !
BEFORE RESTORING ELECTRICAL CONNECTIONS, MAKE SURE THAT QuadraLASE IS SWITCHED
OFF.

POWER LOSS

Refer to the QuadraLASE Instruction Manual for this inspection.


Check and confirm the maximum power output.

Page 19 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 19 of 24
QuadraSCAN - OPERATOR’S MANUAL

FREQUENCY Every 6 months


PERFORMED BY CANDELA CORP. or authorized CANDELA CORP. personnel.

A power loss of more than 20% (standard IEC 601.2.22) indicates the need to:

- Re-alignment of optics
- Recalibration of the integrated power meter.

Should problems of any kind become persistent contact CANDELA CORP..


Installation, maintenance, technical assistance etc. must only be carried out by CANDELA CORP. or
authorized CANDELA CORP. personnel.

7.2 - EXTERNAL CLEANING / DISINFECTION -

The QuadraSCAN scanning handpiece surface should be disinfected when the unit is initially received, and
afterwards as required by clinical / hospital protocol. The surfaces may be wiped down with a damp cloth or
cotton dipped in an alcohol solution or disinfectant liquid.

CAUTION:
DISCONNECT THE POWER CABLE FROM THE AC OUTLET BEFORE CLEANING / DISINFECTING
WITH LIQUIDS.

ONLY THE EXTERNAL SURFACES OF THE UNIT MAY BE CLEANED BY OPERATOR OR STAFF.

Lenses and mirrors within the unit are to be cleaned only by CANDELA CORP. or authorized CANDELA
CORP. personnel.

WARNING!
DO NOT EXPOSE QuadraLASE AND QuadraSCAN ACCESSORY TO WATER SPRAY.

7.3 - CLEANING OF OPTICAL COMPONENTS -


Cleaning of CO2 optics is an important part of laser maintenance. By following the suggested cleaning
procedures, optical problems can be prevented. Optical elements can be damaged by fingerprints, oil, heat
sink grease, dust, and other contaminants.

CAUTION: DO NOT USE SOAP AND WATER TO CLEAN OPTICS.

When not in use, the Scanner handpiece should be protected from dust. Storage in a plastic bag is
recommended.

If dust accumulates on the surface, use a jet of purified dry air for cleaning.

The scanner handpiece contains a focusing lens which may become contaminated during use.

Cleaning Instructions:
- Unscrew the distal conical protection tube from the end of the Scanner handpiece.
- Clean the lens surface with lens paper soaked in Ethyl alcohol or distilled water. Acetone solvent may also
be used. If gloves are used, make sure they will not dissolve on contact with such fluids.
- Place a sheet of lens paper on the surface to be cleaned and apply a few drops of solvent.
- Gently wipe in a circular motion to clean the surface of the optics.

Page 20 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 20 of 24
QuadraSCAN - OPERATOR’S MANUAL

TROUBLESHOOTING

8.0 - GENERAL -

WARNING !
IF QuadraLASE MALFUNCTIONS, CONSULT THE TROUBLESHOOTING CHAPTER IN THE
QuadraLASE INSTRUCTION MANUAL

Clinic / hospital staff may perform the troubleshooting procedure in this chapter, except where specifically
stated that troubleshooting must be carried out by CANDELA CORP. or authorized CANDELA CORP.
personnel.

CAUTION: QuadraLASE GENERATES HIGH VOLTAGES AND EMITS HIGH-INTENSITY LASER


RADIATION. IMPROPER HANDLING OF THE UNIT CAN CAUSE SERIOUS INJURY.

Improper use or adjustment of the QuadraSCAN may invalidate the CANDELA CORP. Service Warranty
Agreement. Contact your CANDELA CORP. representative before attempting to troubleshoot the unit in any
manner other than as directed in this manual.

8.1 – TROUBLESHOOTING GUIDE -

The equipment is designed with a number of fail safe circuits that shut off all systems in the event of
malfunction.

PROBLEMS TRIGGER OF EVENT/ACTION


Interlock Alarm An external switch system activates a remote interlock system which
disconnects the unit from operation.
- Arm safety switch
- Box safety switch
- Fractional switch in the ON position when the standard screen is
displayed on the LCD
- Fractional switch not in ON or OFF position

Action:
Reconnect external switch contacts, or install short circuit on remote
interlock connection point.
Check that the Arm is correctly connected to the QuadraLASE unit.
Check that the box is correctly installed.
Check the Fractional switch position.
Switch on again; if trouble persists, contact CANDELA CORP. or
authorized CANDELA CORP. personnel.

Scanner Alarm - The scanning handpiece is not connected to the extension cable
- The extension cable is not connected to the receptacle on the rear
panel of the QuadraLASE
- The fractional group cable is not connected to the receptacle on the
rear panel of the QuadraLASE
- Malfunction of scanner control electronics
- With QL select in “Operate” the footswitch is pressed with the
Fractional switch in OFF position

Make sure the cable connections have been made. If not, switch off the
appliance. Check the Fractional switch position during operation.

Page 21 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 21 of 24
QuadraSCAN - OPERATOR’S MANUAL

If the problem is due to a malfunction of the scanner control electronics


or if the alarm continues, contact CANDELA CORP. or authorized
CANDELA CORP. personnel.

There is no pilot beam from - Problem with the pilot beam power supply or with the control circuit
the articulated arm - Misalignment of the articulated arm

If trouble persists, contact CANDELA CORP. or authorized CANDELA


CORP. personnel.

The pilot beam intensity is: - Articulated arm out of alignment


- very low
If trouble persists, contact CANDELA CORP. or authorized CANDELA
- diffuse CORP. personnel.

ANNEX “A”

Page 22 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 22 of 24
QuadraSCAN - OPERATOR’S MANUAL

INSTALLATION CHECK OF QUADRASCAN ACCESSORY

- Open the shipping container and check the contents against the packing list.
- Make sure that all safety labels are applied on the equipment as described in this Manual.

UNIT WITHOUT MAINS POWER CONNECTION

- Mains power supply cable of QuadraLASE not connected to the AC power receptacle.
- Key switch in "O" position.

MECHANICAL TESTS

- Visually check all QuadraSCAN parts to ensure that there are no cosmetic defects such as scratches,
dents, paint finish or silk screen imperfections etc.
- Install QuadraSCAN as directed in this Manual.

UNIT CONNECTED TO THE MAINS POWER SUPPLY.

- Mains power supply cable of QuadraLASE connected to the AC power receptacle.


- Key switch in "I “ position.
- Select “Scanner” from the standard screen and go into the scanner functions screen.

- Test all control panel operations as described in this Manual.


- Check all Shapes at the output of the Scanner handpiece
- Check all Size of shapes
- Check the Fractional switch in the ON and OFF position
- Check the scanning procedure:

a) Traditional scanning procedure for each shape for three different Size and Time selection
including 100 usec.
b) QL Fractional scanning procedure for three different Sizes and Time including “2.5 msec”.

ANNEX “B”

Page 23 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 23 of 24
QuadraSCAN - OPERATOR’S MANUAL

Page 24 Rev. 0 of 16.02.2009 (AI0551) 8085-50-9194


8085-50-9194 A Candela Corp Proprietary Sheet 24 of 24