3 Calibration
3.1 Equipment Required
5520A
3334
or
3334-01
3458A
3.3 Requirement
Calibration requires the following ambient conditions:
x Temperature : 23qC r5qC
x Humidity : 80 rh or less
x Power supply : Power supply voltage specified on the instrument(100V AC to 240V AC)
x Warming up : 3 minutes or more
x Unless otherwise specified, input equals a sine wave of power factor(PF)=1, f=55 Hz.
x As for the calibration of GP-IB (Procedure No.11), carry out only on the 3334-01. (Carry out with the
address 1.)
x Carry out calibration of the analog output, and the waveform output in the adjustment mode.
Enter into to adjustment mode using one of the following procedures :
Manual : -1. During the self-test(all LEDs lights,model name, version display), press the keys in order as
follows. "SHIFT" Key o "SET key" o "a" key.
Via PC : -2. Enter the following command the self test has completed and the measurement value is
indicated.
" :ADJ:MODE ON "
3.4 Calibration Procedure
________________________________________________________________________________________
5
CONFIDENTIAL
________________________________________________________________________________________
6
CONFIDENTIAL
(Note) 1. There are two ways to judge for intrinsic error : " basic specifications - equipment accuracy" and "
basic specifications ×0.8".
Compare them, adopt the one that has a stricter judgment standard.
________________________________________________________________________________________
7
CONFIDENTIAL
x Baud rate
(Note) 2. The communication condition of RS-232C :
x Parity
x Stop bit
: None
: 1bit
Use a crossover cable for the communication. (Refer to an instruction manual 4.2.)
3. Confirm the GP-IB address in the setting screen. (Refer to an instruction manual 4.3.)
________________________________________________________________________________________