RAPIDPOINT® 500e Blood Analysis System Specification

Name : RAPIDPOINT® 500e Blood Gas System


SMN : 11416755
AKL : AKL 20101026127
Nama Pendaftar : PT. SIEMENS HEALTHINEERS INDONESIA
Date of Expiry : 14 April 2025

System Specifications
System Description : Point of care blood gas, electrolytes,metabolites and hematocrit analyzer
Sample Types : Whole blood (arterial, venous, mixed venous and capillary); pleural fluid; dialysate
Sample Size : 100 uL
Time to Result : Approximately 60 seconds
: Cycle time: approximately 120 seconds
Calibration : 1-point calibration every 30 minutes; 2-point calibration every 2 hours
: full calibration every 8 hours
Quality Control : Automatic Quality Control (AQC) cartridge : three levels of independent
: Quality control solutions; customizable QC schedule; ampule QC
Integrated Bar-code Scanner : For patient ID, operator ID, and ampuled QC
: 1D barcode symbologies: Code 128, Codabar, Code 39, Interleaved 2 of 5
: 2D barcode symbologies: PDF47, MicroPDF417, Datamatrix,
: QR Code, Micro QR Code, Aztec and MaxiCode
External Interfaces : USB (3 ports); RS232; 10BASE-T Ethernet; barcode scanner
Power Requirements : Rating: 150 VA
: Voltage: 100–240 VAC
: Frequency: 48–62 Hz
Safety : TUV-listed, CSA, EN/IEC 61010-1, JIS
Operating System : MICROSOFT WINDOWS 10 IoT Enterprise (1809)
Communication : Wireless
: LIS
: Dual-port transmission via Ethernet and serial port
: POCcelerator

System Dimensions
Depth : 42.0 cm
Width : 30.0 cm
Height : 55.0 cm
Weight : 16.55 kg
Touchscreen : 21.1 cm x 15.8 cm

Environmental Requirements
Operating Temperature : 15°C–30°C
Humidity : 5–85% noncondensing
Barometric Pressure : 523–800 mmHg

Parameters Measured Parameters Calculated Parameters


Measurement Range Unit of Measure pH (T) BO2
pH : 6.5 – 7.8 pH units pCO2 (T) pO2 (A-a)(T)
pH Pleural : 7.0 - 7.5 pH units pO2 (T) pO2 (a/A)(T)
pCO2 : 5 – 200 mmHg HCO3 -act p50
pO2 : 20 – 700 mmHg HCO3-std Qsp/Qt(T)
Na+ : 100 - 200 mmol/L BE(B) Qsp/Qt(est)
K+ : 0.50 - 15.00 mmol/L BE(ecf) RI(T)
Ca++ (*) : 0.20 - 5.00 mmol/L ctCO2 pO2/FIO2
- (*) ++
Cl : 65.0 –140.0 mmol/L Ca (7.4) ctO2(a),ctO2(⊽)
Glu : 1.1 - 41.6 mmol/L AnGap ctO2([a-]/a)
Lac : 0.18 - 30.00 mmol/L Hct ctO2(v), ctO2(a-⊽)
CO-oximetry Parameters sO2 ctO2(Hb )
tHb : 2.0–25.0 g/dL O2SAT(est) VO2
nBili : 2.0–30.0 mg/dL mOsm DO2
sO2 : 15.0–100.0 %
FO2Hb : 0.0–100.0 %
FHHb : 0.0–100.0 %
FCOHb : 0.0–100.0 %
FMetHb : 0.0–100.0 %

Inspired by patients.Empowered by technology.


Integri-sense™ technology is the guardian of patient results.
It is a comprehensive series of analyzer functional checks and flagging mechanisms designed to deliver accurate test results.