Professional Documents
Culture Documents
TMS 561 Turbidimeter
TMS 561 Turbidimeter
FEATURES
The Wallace & Tiernan® TMS 561 turbidimeter provides an economical, compact
system for turbidity measurement in most water treatment applications. Available
with either a white light or IR light source, it is ideally suited for reporting results
under EPA and other international standards. Each TMS 561 turbidimeter offers
these standard and optional features:
PERIFERICO SUR ORIENTE #3520, COL. COLONIAL TUXTLA GUTIERREZ, CHIAPAS TEL Y FAX. (01961) 604 1121 (01961) 614 1631 VENTAS@BONAINSA.COM
DESIGN
Vapor Purge
To eliminate condensation on the flow-through cuvette, a replaceable desiccant
pack dries the air in the area of the cuvette. A circulation fan and heat from the
system’s electronics further maintain a dry environment around the cuvette to
eliminate any interference from transient moisture. An alarm on the control display
indicates when the desiccant pack requires replacement.
Light Source
White light is recommended for use in turbidimeters reporting results under USEPA
(US standard) jurisdiction. The TMS utilizes an advanced krypton-filled white light
technology with a lamp life expectancy up to seven years.
Infrared light is recommended for use in turbidimeters reporting results under ISO
7027 (European standard) jurisdiction. Lamp life expectancy is up to 11 years.
Infrared light is also recommended for wastewater final effluent and industrial
applications where color is present in the sample stream.
Automatic Self-Cleaning
For applications with poorer water quality, automatic self-cleaning is available as
an option. This unique system works by sending ultrasonic frequency into a piezo
transducer bonded to the bottom of the flow-through cuvette. This dramatically
increases the time between manual cleaning to enable turbidity measurements in
difficult applications. This extends flow-through measurement technology into the
area once only capable with surface scatter instrumentation, but at a lower cost
and improved reliability.
Menu-Driven Electronics
An integral LCD display provides the operator with the turbidity levels and
messages for guidance in setup, calibration, and troubleshooting. There are three
operational modes of the instrument accessible through the three-button touchpad:
AUTO (measurement), CAL, and CONFIG. Through these three menu selections,
the instrument can be calibrated, customized (displays, alarm functions, outputs,
etc.), and setup for limited sample calibration. A security access option can be
selected to prevent unauthorized tampering with the instrument operating
parameters.
PERIFERICO SUR ORIENTE #3520, COL. COLONIAL TUXTLA GUTIERREZ, CHIAPAS TEL Y FAX. (01961) 604 1121 (01961) 614 1631 VENTAS@BONAINSA.COM
Simple, Economical Calibration
Calibration can be completed using EPA-approved primary standards available as
an accessory from Siemens’ range of Wallace & Tiernan® Products. Unlike
competitive units, these calibration standards are sealed cuvettes and require no
mixing or dilution of expensive bulk standards. This dry method of calibration is
fast, clean, and reusable. One set of standards can be used on multiple
turbidimeters. Onscreen menu prompts guide you through the calibration
procedure quickly and easily. It is not necessary to disconnect any piping or turn off
the sample flow when performing the calibration.
PERIFERICO SUR ORIENTE #3520, COL. COLONIAL TUXTLA GUTIERREZ, CHIAPAS TEL Y FAX. (01961) 604 1121 (01961) 614 1631 VENTAS@BONAINSA.COM
OPERATION
Turbidity of a fluid is caused by undissolved substances, such as inorganic or
organic particles, colloids, or gas bubbles. If light strikes such suspended particles,
the light beam is reflected or refracted and a small amount of light is scattered
throughout the fluid. The intensity of the light scattered is determined by the
amount and the properties (size, shape, color) of these particles. Measurement of
the turbidity is based on the evaluation of the intensity of the scattered light.
The measurement value depends on the wave length of the light used and the
geometry of the measuring instruments or the angle of measurement. In order to
establish an international standard of turbidity, these parameters have been
defined by the EPA (for U.S. applications) and ISO (for international applications).
PERIFERICO SUR ORIENTE #3520, COL. COLONIAL TUXTLA GUTIERREZ, CHIAPAS TEL Y FAX. (01961) 604 1121 (01961) 614 1631 VENTAS@BONAINSA.COM
TECHNICAL DATA
Model Selection: All models provided fully calibrated and include 4-20 mA output,
desiccant, spare measuring cuvette with light shield, power supply, and instruction
manual. Also, an optional flow alarm is available to provide indication of a loss of
sample flow.
Measurement Range: 0-1000 NTU (Models BW, BR, DW & DR)0-100 NTU
(Models CW & CR)
PERIFERICO SUR ORIENTE #3520, COL. COLONIAL TUXTLA GUTIERREZ, CHIAPAS TEL Y FAX. (01961) 604 1121 (01961) 614 1631 VENTAS@BONAINSA.COM
Sample Flow Rate: 100 ml/min to 1 liter/min (0.026 to .26 gal/min)
Sample Flow Alarm: Optional, factory installed. Provides alarm indication in the
event of a loss of sample flow.
PERIFERICO SUR ORIENTE #3520, COL. COLONIAL TUXTLA GUTIERREZ, CHIAPAS TEL Y FAX. (01961) 604 1121 (01961) 614 1631 VENTAS@BONAINSA.COM