Specifications Technical Specification
No.
01. Name of Item Portable Handheld XRF
02. Name of Manufacturer To be mentioned
03. Model To be mentioned
04. Country of origin USA, Canada, Japan, Germany
05. Country of Assembly Country of Origin and country of assembly should be same
06. Tube Must have Silver (Ag) anode 50 kV X-ray tube
Must have SDD (Silicon Drift Detector), In-line camera for
07. Detector
aiming, 3mm X-ray spot collimation
08. Excitation Source 4-Watt X-ray tube with application optimized anode material
Mg, Al, Si, P, S, Ti, V, Cr, Mn, Fe, Co, Ni, Cu, Zn, W, Hf, Ta, Re,
09. Element Detection Ability Pb, Bi, Zr, Nb, Mo, Pd, Ag, Cd, Sn, Sb.
10. Primary Beam Filtration 8-position auto selected filter per beam per mode
IP Rating Must be IP55: dust protected and protected against water jets
18. from all directions
20. Data Storage 4 GB embedded storage, micro-SD slot for expandable storage
MiniUSB cable, USB Thumb drive, Bluetooth, WiFi, removable
21. Data Transfer
micro-SD card
22. Security Password protected user security
Built-in barometer for automatic altitude and air density
23. Pressure Correction
correction
Mode
24. Alloy Modes: Metal Alloy
Touch-screen keyboard
25. Data Entry
User programmable pick lists
Locking shielded pelican carry case
Spare battery pack
18V power transformer 100-240 VAC, 50-60 Hz, 70 W
max.
PC connection cable (MiniUSB cable)
26. Standard Accessories
Vanta PC Software
Safety wrist strap
Docking Station with Check samples/standards
10 spare analysis windows
34. Compliance Statement Supplier should be provided a compliance statement with offer