You are on page 1of 7

WEARABLE LACTATE THRESHOLD PREDICTING DEVICE

IS VALID AND RELIABLE IN RUNNERS


NATTAI R. BORGES1 AND MATTHEW W. DRILLER2
1
School of Medical and Applied Sciences, Central Queensland University, Rockhampton, Australia; and 2University of
Waikato, Hamilton, New Zealand

ABSTRACT INTRODUCTION

F
Borges, NR and Driller, MW. Wearable lactate threshold or decades, lactate threshold (LT) has been utilized
predicting device is valid and reliable in runners. J Strength by endurance athletes as a key performance pre-
Cond Res 30(8): 2212–2218, 2016—A commercially avail- dictor and to dictate training practices (29).
able device claiming to be the world’s first wearable lactate Alongside a high V_ O2max and efficiency, a high
threshold predicting device (WLT), using near-infrared LED
LT is essential for endurance performance as it allows an
athlete to perform at a higher work-rate for an extended
technology, has entered the market. The aim of this study
duration (19). Historically, to determine LT, invasive blood
was to determine the levels of agreement between the
measures and costly laboratory equipment are required to
WLT-derived lactate threshold workload and traditional
sample whole blood during incremental exercise tests. Fur-
methods of lactate threshold (LT) calculation and the thermore, skilled technicians are required to measure blood
interdevice and intradevice reliability of the WLT. Fourteen lactate concentration values and to interpret lactate profile
(7 male, 7 female; mean 6 SD; age: 18–45 years, height: data. Over the past 20 years, monitoring of blood lactate
169 6 9 cm, mass: 67 6 13 kg, V_ O2 max: 53 6 9 concentration has become less invasive and easier to acquire
ml$kg21 $min 21 ) subjects ranging from recreationally active because of advances in technology that require smaller
to highly trained athletes completed an incremental exercise amounts of whole blood and capillary blood sampling tech-
test to exhaustion on a treadmill. Blood lactate samples niques. However, new alternatives to quantify exercise
were taken at the end of each 3-minute stage during the threshold workloads via noninvasive and cost-effective
test to determine lactate threshold using 5 traditional meth- methods are constantly being introduced.
ods from blood lactate analysis which were then compared To circumvent the invasive nature of calculating exercise
threshold workloads from blood lactate concentration,
against the WLT predicted value. In a subset of the popu-
researchers have investigated the use of expired gas (8),
lation (n = 12), repeat trials were performed to determine
and to a lesser extent heart rate (7) inflection points and
both inter-reliability and intrareliability of the WLT device.
rating of perceived exertion scores (10) as means for deter-
Intraclass correlation coefficient (ICC) found high to very
mining exercise thresholds. Although no invasive measures
high agreement between the WLT and traditional methods are required, there are still high costs related to the labo-
(ICC . 0.80), with TEMs and mean differences ranging ratory equipment and skilled technicians needed to collect
between 3.9–10.2% and 1.3–9.4%. Both interdevice and and interpret the expired gas and heart rate data. More
intradevice reliability resulted in highly reproducible and recently, near infrared spectroscopy (NIRS) systems have
comparable results (CV , 1.2%, TEM ,0.2 km$h21, been used to continuously monitor muscle tissue oxygen-
ICC . 0.97). This study suggests that the WLT is a practi- ation via the absorption of the near infrared light by hemo-
cal, reliable, and noninvasive tool for use in predicting LT in globin and myoglobin as the near infrared light passes
runners. through the muscle tissue. Recent studies have been test-
ing the use of the muscle tissue oxygenation values derived
KEY WORDS lactate threshold, endurance performance, from NIRS to gain exercise threshold information via
performance testing, reliability, validity inflection points of muscle oxygenation parameters during
incremental exercise (2,3,13,27). Although the use of NIRS
is relatively new to sports science, it has proven to be
a viable noninvasive tool to determine peripheral muscle
Address correspondence to Nattai R. Borges, n.borges@cqu.edu.au. oxygenation during exercise (4,22).
30(8)/2212–2218 With further technological advancements, the use of
Journal of Strength and Conditioning Research portable NIRS systems to monitor peripheral muscle oxy-
Ó 2015 National Strength and Conditioning Association genation in field testing has become more common (6). A
the TM

2212 Journal of Strength and Conditioning Research

Copyright © National Strength and Conditioning Association Unauthorized reproduction of this article is prohibited.
the TM

Journal of Strength and Conditioning Research | www.nsca.com

commercially available wearable lactate threshold predicting give written consent. This study was given ethical clearance
device (WLT) (BSXinsight multisport edition; BSX Athletics, by the University of Waikato Human Research Ethics Com-
Austin, TX, USA) is a portable NIRS LED device, which is mittee in accordance with the Helsinki declaration.
marketed as the first wearable lactate threshold predictor
Procedures
claiming to have the ability to determine the workload at
LT during self-administered maximal cycling or running The GXT consisted of 3-minute stages at increasing work-
exercise tests (5). The noninvasive and portable nature of loads (km$h21) determined by the WLT software. The WLT
the device allows athletes and coaches to easily monitor software predetermined the protocol stage workloads via
their LT, prescribe exercise-training zones based on LT val- information about the subjects’ current fitness levels. The soft-
ues, and monitor training adaptations in a cost-effective ware required information on current “conversational pace”
manner. Because of the patented design and algorithm of explained to participants as the maximal pace that subjects
the device, it is difficult to evaluate the exact methodology would still be able to comfortably maintain a conversation
used to predict LT workload by the WLT. However, no and “10 km pace” explained as the maximal pace subjects
studies have investigated the validity nor the reliability of would be able to maintain over a 10-km distance. The GXT
the WLT devices’ ability to predict workload at LT. There- protocols started with a “warm up” walking stage at 4.8
fore, the aims of this investigation were twofold: (a) to km$h21 followed by a sharp increase in workload for the
determine the levels of agreement between the WLT and second stage (9.3–11.7 km$h21). Further stage increments
a number of traditional blood lactate methods including the ranged from 0.3 to 1.1 km$h21 until the subject was unable
linear spline fitting method, Dmax method, modified Dmax to maintain the stage workload. The exercise protocol and
method, fixed blood lactate concentration of 4 mmol$L21 workload increments set by the WLT had to be followed and
method, and first rise of blood lactate greater than 1 a minimum of 4 stages had to be completed for a successful
mmol$L21 method to determine workload (km$h21) at prediction of LT workload by the WLT. Expired gas variables
LT, and (b) to establish the reliability of the WLT device were collected and analyzed at 15 seconds intervals through-
during a repeated trial (intradevice reliability) and when out the GXT with an indirect calorimetry system (TrueOne
using 2 devices during a single test (interdevice reliability). 2400; Parvo Medics, Inc.; Sandy, UT, USA), which was cali-
brated as per the manufacturer’s instructions. V_ O2max was
taken as the highest V_ O2 value (ml$kg21$min21) recorded
METHODS
over a 1-minute period during the GXT.
Experimental Approach to the Problem
The LT workloads calculated by the WLT were compared BSX Insight
with traditional LT workloads after an incremental exercise The WLT (BSXinsight multi-sport edition; BSX Athletics) is
test to exhaustion (GXT) on a treadmill (HP Cosmos, a commercially available portable NIRS LED device that is
Traunstein, Germany). All testing was performed in con- housed in a compression sleeve and fitted over the gastroc-
trolled laboratory conditions (21 6 18 C and 60% relative nemius muscle of the user (Figure 1). Compression sleeves of
humidity). Subjects were asked to refrain from strenuous different sizes were utilized to ensure that the compression
exercise (,12 hours) and to arrive at each session in a fully sleeve fit firmly and comfortably on all subjects regardless of
rested and hydrated state. For the test–retest reliability trials, body size. During inter-reliability trials, 2 devices were used,
subjects were to report for testing at the same time of day with 1 device attached to each leg. The WLT monitors
(61 hour), separated by 24 hours. change in muscle oxygenation (total oxygenation index
[TOI]) and has been suggested by the manufacturers to be
Subjects able to predict LT at 97% accuracy via their patented algo-
Seven male (mean 6 SD; age: 19–45 years, height: 172.2 6 rithm that detects inflection points in the muscle oxygena-
8.5 cm, mass: 74.5 6 12.7 kg, V_ O2max: 58.6 6 7.0 tion curve at increasing workloads (5). The WLT was aligned
ml$kg21$min21) and 7 female (age: 18–41 years, height: with the thickest section of the gastrocnemius as per the
164.4 6 7.1 cm, mass: 59.8 6 7.6 kg, V_ O2max: 47.1 6 6.5 manufacturer’s instructions and LT workload was calculated
ml$kg21$min21) subjects from a range of recreationally in pace (min$km21), which was converted to speed
active to highly trained athletes (V_ O2max range = 39.7– (km$h21) for further analysis.
67.0 ml$kg21$min21) volunteered to participate in the cur-
rent study. Both male and female runners from a range of Blood Lactate Sampling and Analysis
abilities and demographics were recruited to test the WLT Blood lactate concentration was collected and analyzed
across a large span of lactate threshold speeds (10.9–16.2 (Lactate Pro 2; Arkray Global Business Inc., Shiga, Japan) in
km$h21). To be eligible for the study, all participants were the final 10 seconds of each stage and 1-minute postexercise via
required to be free from injury and medication that may have finger-prick capillary blood samples using universal procedures.
affected their ability to perform maximal exercise. Before Traditional method LTs were calculated using a previously
inclusion, all participants were informed about the study validated Excel spreadsheet (14,23). Traditional LT workloads
including potential risks and benefits and were required to were determined using a range of methods including:

VOLUME 30 | NUMBER 8 | AUGUST 2016 | 2213

Copyright © National Strength and Conditioning Association Unauthorized reproduction of this article is prohibited.
Wearable Lactate Threshold Predictor

Fixed blood lactate concentration of 4 mmol$L21


(4mmoL): Calculated using an inverse prediction by finding
the work rate/intensity corresponding to a lactate value
equal to a fixed blood lactate concentration of 4 mmol$L21.
First rise of blood lactate greater than 1 mmol$L21
(1mmoL): Defined as the workload preceding an increase
in lactate concentration of a fixed rise of 1 mmol$L21 after
baseline.
All LT workload calculations have also been previously
described (23).

Reliability Testing
In a subset of the sample used in the current study, further
trials were performed to evaluate the interdevice and
intradevice reliability of the WLT (n = 12). Subjects per-
formed 2 GXT tests separated by 24 6 1 hour to measure
the intradevice reliability of a single WLT. Given the aim
was to test the reliability of the measurement device (not
Figure 1. A) The wearable lactate threshold predicting device (WLT) the test itself ), the test was discontinued at exactly the
and (B) the compression calf sleeve that the WLT it is housed in, same point of the GXT in the second trial. The same in-
positioned over the gastrocnemius calf muscle. 1 = 3 near infrared dividuals also wore 2 WLT devices (one on each leg) dur-
LED’s; 2 = charging connector pins.
ing a single GXT to determine the interdevice reliability of
2 WLTs.

Linear spline fitting (LSP): The point of intersection Statistical Analyses


between 2 linear splines on a lactate curve. The location of Descriptive statistics (mean 6 SD) were calculated for all
this intersection is estimated by minimizing the sum of the data. A 1-way ANOVA was implemented to measure differ-
squared differences between the observed lactate values and ences between LT workload values derived from the WLT
the fitted values. and traditional methods. Between-method agreements for
Dmax: Defined as the point that yields the maximal LT were examined using intraclass correlation coefficients
distance from the lactate curve (using a third-order poly- (ICC) with 95% confidence intervals and interpreted as
nomial) to the line formed by the lowest and highest lactate 0.90–1.00 = very high correlation, 0.70–0.89 = high correla-
values of the curve. tion, 0.50–0.69 = moderate correlation, 0.26–0.49 = low cor-
Modified Dmax (mDmax): Defined as the point that relation and 0.00–0.25 = little, if any correlation (18,21). The
yields the maximal distance from the lactate curve (using mean differences and upper and lower levels of agreement (2
a third order polynomial) to the line formed between 0.4 standard deviations or 95% of a normally distributed popu-
mmol$L21 above the lowest lactate value and the highest lation) between methods were determined in absolute values
lactate value on the curve. (km$h21) and as a percentage (%). Between-method typical

TABLE 1. Intraclass correlation coefficients, typical error of measurement, and levels of agreement for lactate
threshold workload (km$h21) predicted by the WLT and traditional methods.*

WLT
compared ICC TEM (95% CI) TEM Mean difference Mean
with (95% CI) (km$h21) (95% CI) (%) (62 SD) (km$h21) difference (%)

LSF 0.91 (0.22–0.98) 0.74 (0.53–1.23) 6.1 (4.3–10.2) 0.86 (20.61 to 2.32) 6.6
Dmax 0.80 (20.17 to 0.92) 0.98 (0.74–1.48) 8.6 (6.4–13.3) 1.21 (20.82 to 3.23) 9.4
mDmax 0.89 (20.10 to 0.95) 1.21 (0.92–1.84) 10.2 (7.6–15.8) 0.59 (21.75 to 2.93) 4.5
4mmoL 0.98 (0.94–0.99) 0.48 (0.36–0.72) 3.9 (2.9–6.0) 0.18 (20.82 to 1.18) 1.3
1mmoL 0.92 (0.41–0.98) 0.72 (0.54–1.09) 6.3 (4.8–9.8) 0.84 (20.84 to 2.52) 6.4

*WLT = wearable lactate threshold predicting device; LSF = linear spline fitting; mDmax = Modified Dmax; 4mmoL = fixed blood
lactate concentration of 4 mmol$L21; 1mmoL = first rise of blood lactate greater than 1 mmol$L21; ICC = intraclass correlation
coefficient; TEM = typical error of measurement.

the TM

2214 Journal of Strength and Conditioning Research

Copyright © National Strength and Conditioning Association Unauthorized reproduction of this article is prohibited.
the TM

Journal of Strength and Conditioning Research | www.nsca.com

Figure 2. Bland Altman plots depicting levels of agreement of workload (km$h21) at lactate threshold for (A) linear spline fitting (LSF) and the wearable lactate
threshold predicting device (WLT) (B) Dmax and the WLT (C) mDmax and the WLT (D) fixed blood lactate of 4 mmoL21 (4mmoL) and the WLT (E) first increase
in blood lactate greater than 1 mmoL21 (1mmoL) and the WLT during incremental exercise tests to exhaustion on a treadmill.

error of measurement (TEM) was determined in absolute RESULTS


values (km$h21) and as a percentage (%) using an excel One-way ANOVA showed no statistically significant differ-
spreadsheet (15,17). Bland Altman plots were also con- ences between the LT workloads determined by the different
structed to visualize levels of agreements. Interdevice and
methods (p = 0.60). Intraclass correlation coefficients with
intradevice reliability data were log-transformed and ana-
95% CI ranged from 0.80 to 0.98 (20.10 to 0.99) for work-
lyzed using an Excel spreadsheet for reliability (16). Typical
error of measurement and overall reliability of the WLT is load at LT calculated by the WLT and traditional methods,
presented as a coefficient of variation percentage (CV%) and and are shown in Table 1.
as an absolute value (km$h21) along with ICCs and upper Table 1 also shows the TEM and levels of agreement
and lower 95% CI. All statistical analyses were performed in between the workloads at LT calculated by the WLT and
SPSS V22.2 (IBM Corporation, Armonk, NY, USA) unless traditional methods in absolute values and as percentages.
stated otherwise. Statistical significance was accepted at the TEM and levels of agreement and 95% CI ranged from 0.48
p # 0.05 level. to 1.21 km$h21 (3.9–10.2) and 0.18 to 0.86 km$h21 (1.3–9.4),

TABLE 2. The inter-reliability and intrareliability of the wearable lactate threshold predicting device (WLT) including
intraclass correlation coefficients, typical error of measurement (km$h21), and coefficient of variation for each
comparison.*†

Trial 1 Trial 2 ICC TEM (95% CI) CV


(km$h21) (km$h21) (95% CI) (km$h21) (95% CI) (%)

Inter-reliabilityz 12.99 6 0.84 13.05 6 0.81 0.97 (0.90–0.99) 0.16 (0.12–0.28) 1.2 (0.9–2.1)
(n = 12)
Intrareliability§ 13.02 6 0.78 13.03 6 0.87 0.97 (0.90–0.99) 0.16 (0.11–0.27) 1.2 (0.9–2.1)
(n = 12)

*ICC = intraclass correlation coefficient; TEM = typical error of measurement; CV = coefficient of variation.
†Data are shown as means 6 SD unless stated otherwise.
zInter-reliability of the WLT device as determined during a single test where subjects wore a device on each leg.
§Intrareliability of WLT determined using the same device in a repeated trial separated by ,24 hours.

VOLUME 30 | NUMBER 8 | AUGUST 2016 | 2215

Copyright © National Strength and Conditioning Association Unauthorized reproduction of this article is prohibited.
Wearable Lactate Threshold Predictor

Figure 3. Comparison of lactate threshold workloads (km$h21) for the WLT (wearable lactate threshold predicting device), linear spline fitting (LSF), Dmax,
modified Dmax (mDmax), fixed blood lactate accumulation of 4 mmol$L21 (4mmoL), and first rise of blood lactate greater than 1 mmol$L21 (1mmoL).

respectively, with the highest level of agreement between the 4mmoL at 0.48 km$h21 and 3.9%, and 0.18 km$h21 and
WLT and 4mmoL method. Bland Altman plots showing the 1.3%, respectively. Additionally, the current study found high
mean difference and limits of agreement for the WLT and reliability when comparing 2 WLT devices during the same
traditional methods are shown in Figure 2. test (inter-reliability; CV% = 1.2, TEM = 0.16 km$h21,
Both interdevice and intradevice reliability resulted in ICC = 0.97), which was almost identical to the comparison
highly reproducible and comparable results (Table 2). When between the same device during a repeated test (intrareli-
comparing the interdevice reliability during a single test in 12 ability; CV% = 1.2, TEM = 0.16 km$h21, ICC = 0.97). These
separate trials, the WLT resulted in a CV and 95% CI of 1.2% results suggest that the WLT is both valid and reliable in
(0.9–2.1), a TEM of 0.16 km$h21 (0.12–0.28) and an ICC of determining workload at LT through the threshold range
0.97 (0.90–0.99). Similarly, the intradevice reliability deter- of 10.9–16.2 km$h21 and promote the suitability of the
mined by the same device during 2 tests separated by ;24 WLT in a practical setting.
hours, resulted in a CV and 95% CI of 1.2% (0.9–2.1), a TEM As this is the first study to compare the WLT and
of 0.16 km$h21 (0.11–0.27), and an ICC of 0.97 (0.90–0.99). traditional methods, it is difficult to make direct comparisons
with existing research. However, previous studies investigat-
DISCUSSION ing the suitability of muscle oxygenation breakpoints using
The results for the current study support both the validity other NIRS systems to determine exercise thresholds have
and reliability of the WLT during an incremental exercise test shown similar results to those of the present study (2,3,27). It
to exhaustion on a treadmill. This study showed no is important to note, however, that when comparing the
significant differences in the workloads at LT determined results of the present study to existing research, the hetero-
by the WLT when compared with 5 traditional methods of geneous methodology such as the NIRS device used, site of
measuring LT. Correlation analysis of LT speeds showed measurement (vastus lateralis, gastrocnemius) and which
high to very high levels of agreement (ICC = 0.80–0.98) parameter was used in determining the NIRS breakpoint
between the WLT and traditional methods, with 4mmoL (TOI, change in oxyhemoglobin concentration, change in
threshold resulting in the highest correlation (ICC = 0.98). deoxyhemoglobin concentration) may have an effect on
The typical error of measurement and levels of agreement the compatibility of the results. Furthermore, the majority
(absolute and %) were also lowest between WLT and of existing literature has compared NIRS inflection points to
the TM

2216 Journal of Strength and Conditioning Research

Copyright © National Strength and Conditioning Association Unauthorized reproduction of this article is prohibited.
the TM

Journal of Strength and Conditioning Research | www.nsca.com

ventilatory thresholds (VT) (3,20,24,30) with only a handful and 95% limits of agreement with visual inspection of Bland
of studies comparing NIRS inflection points to blood LTs Altman plots suggest that magnitude of agreements can
(2,13,28). Additionally, there is a high prevalence of cycle occasionally show variation between the WLT and LT tra-
ergometer studies in existing research which is most likely ditional methods.
a result of the possible increase in signal noise owing to Given the range of LT values that were produced in this
movement artifact of the NIRS leads while running (25). study (12.3–13.3 km$h21) by the different traditional meth-
The levels of agreement in our study are supported by ods, it makes it difficult to conclude whether the WLT pro-
existing literature showing high correlations between NIRS vides a valid prediction of LT. Therefore, it could be argued
breakpoints and LTs (2,12). Similar to the results of this study, that the reliability of the WLT is perhaps the most important
Belloti et al. (2) reported no significant differences, high cor- factor when it comes to using the device to monitor running
relations (r = 0.81; r = 0.76), and mean differences of 0.26 performance. Owing to the logistics associated with testing
L$min21 and 8 bpm for V_ O2 and heart rate values derived multiple athletes at 1 time, scientists commonly interchange
from change in deoxyhemoglobin concentration from the multiple units of the same brand device between repeat tests.
vastus lateralis muscle and from maximal lactate steady state As such, it is important that the brand of device employed
during cycle ergometry. Wang et al. (28) compared threshold demonstrates good intradevice and interdevice reliability.
workload (W), V_ O2 and heart rate values from multiple NIRS This way, measurement errors are minimized, and a greater
parameters (TOI, change in oxyhemoglobin concentration, level of confidence is achieved when comparing and inter-
change in deoxyhemoglobin concentration) also from the preting repeat tests on the same athlete. Both the intradevice
vastus lateralis muscle with those derived from VT and LT and interdevice reliability of the WLT device resulted in
during incremental cycling exercise. Threshold for all param- a CV% of ,1.2 (TEM , 0.16 km$h21, ICC . 0.97), suggest-
eters was determined as the inflection point of 2 regression ing that the WLT can produce reliable and reproducible
lines. The authors report the NIRS parameter with the stron- results both when using the same device and 2 separate
gest correlations to be between change in deoxyhemoglobin devices. Given, it has been suggested that a CV of ,10% is
concentration with VT and LT (28). However, contrasting acceptable for reliability in other sport science measurement
the results from our study, the authors reported a significant tools (1), such as metabolic analyzers (9), and the intradevice
correlation between TOI measures and VT (r = 0.95) but not and interdevice CV of the one of the most commonly used
LT (r = 0.68) (28). The contrasting results between the pres- blood lactate analyzers (Lactate Pro; Arkray Global Business
ent study and that of Wang et al. (28) could be attributed to Inc.) has been reported as 5.7 and 5.2% (26), an intradevice
heterogeneous methodology. and interdevice CV of ;1.2% suggests that the WLT appears
Although showing high levels of agreement through to be highly reliable.
ICCs, mean difference values reveal a tendency of the
WLT to overestimate LT workload when compared with PRACTICAL APPLICATIONS
traditional methods (Figure 3). However, higher workloads This study provides the first data on the validity and
at NIRS inflection points compared with VT and LT has reliability of the WLT’s ability to predict lactate threshold
been previously reported in existing literature (11) and there in runners. The results highlight that the WLT is a reliable
is also evidence to suggest that NIRS inflection points at the tool that shows marginal variance in predicting workload at
gastrocnemius muscle occur at higher workloads compared LT when compared with some of the common LT methods
with those at the vastus lateralis muscle (27). Taken together, derived through blood-sampling techniques. These observa-
these past findings may explain why the WLT showed tions suggest that coaches and athletes could use this device
a small overestimation of LT workload in our study. The to monitor workload at LT and prescribe training parameters
WLT showed the highest levels of agreement with the based on the LT workload predictions through the noninva-
4mmoL method (ICC = 0.98; TEM = 0.48 km$h21, 3.9%; sive and self-administered exercise test. The nature of the
mean difference = 0.18 km$h21, 1.3%) with narrow 95% exercise test prescribed by the WLT would allow coaches
limits of agreement (20.82 to 1.18 km$h21). However, this to remotely monitor their athletes and could allow coaches
magnitude of agreement was not apparent with all tradi- and athletes to monitor training adaptations through
tional methods. Visual inspection of Bland Altman plots changes in LT workload.
demonstrates that outliers in are evident when comparing
the WLT to Dmax and mDmax (Figures 1B,C) with data CONCLUSIONS
points falling outside the 95% limits of agreement. Further- The findings of our study suggest that the WLT is a practical
more, inspection of the scatter points for the WLT and tool for estimating lactate threshold workloads for runners.
1mmoL (Figure 2E) also shows a downward trend with We found acceptable levels of agreement between the WLT
increasing workload, suggesting greater agreement between and traditional blood lactate threshold methods and also
the 2 measures at higher threshold workloads. Therefore, that the device was highly reliable in both a test–retest set-
although no significant statistical differences between the ting and when comparing 2 devices during the same test.
WLT and traditional methods were found, overall means The WLT could be implemented by coaches and athletes as

VOLUME 30 | NUMBER 8 | AUGUST 2016 | 2217

Copyright © National Strength and Conditioning Association Unauthorized reproduction of this article is prohibited.
Wearable Lactate Threshold Predictor

a portable and noninvasive method to monitor lactate 14. Higgins, D. Lactate-e for Microsoft Excel. 2007. Available at: http://
threshold workload changes in runners. www.uiginn.com/lactate/lactate-e.html. Accessed March 1, 2015.
15. Hopkins, WG. Analysis of validity by linear regression. Available at:
REFERENCES sportsci.org/resource/stats/xvalid.xls. Accessed January 10, 2015.
1. Atkinson, G, Nevill, A, and Edwards, B. What is an acceptable 16. Hopkins, WG. Analysis of reliability with a spreadsheet. Available
amount of measurement error? The application of meaningful at: http://www.sportsci.org/resource/stats/xrely.xls. Accessed
“analytical goals” to the reliability of sports science measurements January 10, 2015.
made on a ratio scale. J Sports Sci 17: 18, 1999. 17. Hopkins, WG. A new view of statistics. Available at: http://www.
2. Bellotti, C, Calabria, E, Capelli, C, and Pogliaghi, S. Determination sportsci.org/resource/stats/index.html. Accessed January 10, 2015.
of maximal lactate steady state in healthy adults: Can NIRS help? 18. Impellizzeri, FM and Marcora, SM. Test validation in sport
Med Sci Sports Exerc 45: 1208–1216, 2013. physiology: Lessons learned from clinimetrics. Int J Sports Physiol
3. Bhambhani, YN, Buckley, SM, and Susaki, T. Detection of Perform 4: 269–277, 2009.
ventilatory threshold using near infrared spectroscopy in men and 19. Joyner, MJ and Coyle, EF. Endurance exercise performance: The
women. Med Sci Sports Exerc 29: 402–409, 1997. physiology of champions. J Physiol 586: 35–44, 2008.
4. Boushel, R, Langberg, H, Olesen, J, Gonzales-Alonzo, J, Bulow, J, 20. Karatzanos, E, Paradisis, G, Zacharogiannis, E, Tziortzis, S, and
and Kjaer, M. Monitoring tissue oxygen availability with near Nanas, S. Assessment of ventilatory threshold using near-infrared
infrared spectroscopy (NIRS) in health and disease. Scand J Med Sci spectroscopy on the gastrocnemius muscle during treadmill
Sports 11: 213–222, 2001. running. Int J Ind Ergon 40: 206–211, 2010.
5. BSX Athletics. A deeper look through BSXinsight—Muscle 21. Munro, B. Statistical Methods for Health Care Research. Philadelphia,
oxygenation. Available at: http://blog.bsxathletics.com/2015/02/ PA: JB lippincott, 1997.
20/muscle-oxygenation-through-bsxinsight/. Accessed May 1,
2015. 22. Neary, JP. Application of near infrared spectroscopy to exercise
sports science. Can J Appl Physiol 29: 488–503, 2004.
6. Buchheit, M, Bishop, D, Haydar, B, Nakamura, FY, and Ahmaidi, S.
Physiological responses to shuttle repeated-sprint running. Int J 23. Newell, J, Higgins, D, Madden, N, Cruickshank, J, Einbeck, J,
Sports Med 31: 402–409, 2010. McMillan, K, and McDonald, R. Software for calculating blood
lactate endurance markers. J Sports Sci 25: 1403–1409, 2007.
7. Bunc, V, Hofmann, P, Leitner, H, and Gaisl, G. Verification of the
heart rate threshold. Eur J Appl Physiol 70: 263–269, 1995. 24. Rao, R, Danduran, M, Loomba, R, Dixon, J, and Hoffman, G. Near-
infrared spectroscopic monitoring during cardiopulmonary exercise
8. Cheng, B, Kuipers, H, Snyder, A, Keizer, H, Jeukendrup, A, and testing detects anaerobic threshold. Pediatr Cardiol 33: 791–796, 2012.
Hesselink, M. A new approach for the determination of ventilatory
and lactate thresholds. Int J Sports Med 13: 518–522, 1992. 25. Robertson, FC, Douglas, TS, and Meintjes, EM. Motion artifact
removal for functional near infrared spectroscopy: A comparison of
9. Crouter, S, Antczak, A, Hudak, J, DellaValle, D, and Haas, J. methods. IEEE Trans Biomed Eng 57: 1377–1387, 2010.
Accuracy and reliability of the ParvoMedics TrueOne 2400 and
MedGraphics VO2000 metabolic systems. Eur J Appl Physiol 98: 26. Tanner, R, Fuller, K, and Ross, MR. Evaluation of three portable
139–151, 2006. blood lactate analysers: Lactate Pro, lactate scout and lactate plus.
Eur J Appl Physiol 109: 551–559, 2010.
10. Fabre, N, Mourot, L, Zerbini, L, Pellegrini, B, Bortolan, L, and
Schena, F. A novel approach for lactate threshold assessment based 27. Wang, B, Xu, G, Tian, Q, Sun, J, Sun, B, Zhang, L, Luo, Q, and
on rating of perceived exertion. Int J Sports Physiol Perform 8: 263– Gong, H. Differences between the vastus lateralis and gastrocnemius
270, 2013. lateralis in the assessment ability of breakpoints of muscle
oxygenation for aerobic capacity indices during an incremental
11. Fujimoto, S, Yoshikawa, T, Tateishi, Y, Lixin, W, Hara, T,
cycling exercise. J Sports Sci Med 11: 606–613, 2012.
Mimura, T, Nakao, H, and Hirata, K. Evaluation of muscle
oxygenation during exercise by NIRS in normal subjects - 28. Wang, L, Yoshikawa, T, Hara, T, Nakao, H, Suzuki, T, and
Significance of the NIRS threshold. J Jpn Col Angiol 47: 21–27, 2007. Fujimoto, S. Which common NIRS variable reflects muscle
estimated lactate threshold most closely? Appl Physiol Nutr Metab
12. Grassi, B, Pogliaghi, S, Rampichini, S, Quaresima, V, Ferrari, M,
31: 612–620, 2006.
Marconi, C, and Cerretelli, P. Muscle oxygenation and pulmonary
gas exchange kinetics during cycling exercise on-transitions in 29. Yoshida, T, Chida, M, Ichioka, M, and Suda, Y. Blood lactate
humans. J Appl Physiol (1985) 95: 149–158, 2003. parameters related to aerobic capacity and endurance performance.
13. Grassi, B, Quaresima, V, Marconi, C, Ferrari, M, and Eur J Appl Physiol 56: 7–11, 1987.
Cerretelli, P. Blood lactate accumulation and muscle 30. Zhang, Z, Wang, B, Nie, Q, Luo, Q, and Gong, H. Portable muscle
deoxygenation during incremental exercise. J Appl Physiol oxygenation monitor based on near infrared spectroscopy. Front
(1985) 87: 348–355, 1999. Optoelectron China 2: 248–252, 2009.

the TM

2218 Journal of Strength and Conditioning Research

Copyright © National Strength and Conditioning Association Unauthorized reproduction of this article is prohibited.

You might also like