You are on page 1of 4

AUTOCLAVE - STEAM STERILIZER

Autoclaves are designed and fabricated for steam sterilization process to Kill
micro-organism through the application of saturated steam under pressure to
suit various I. S. standards application in the growing field of Medical,
Agricultural Institution, Bio-Chemical Industrial, Research laboratories and
Various Industries.
Features Steam Temp. 121.4 oC at 15 psi.
As per BS 2646-5:1993. Autoclaves for
sterilization in laboratories Methods of Test for
function and performance.

The Energy from the water is transferred into


Specification
high pressure steam inside the Stainless steel
chamber up to 15-psi pressure or 121.4oC
temperature. The Chamber (electrical) is fill with
clean water for sterilization. This heat process
generated chemical free steam for sterilization.
Microprocessor based LCD temp. controller (
Temp Controller
GMP model optional )
Pressure Mechanically operated pressure/steam release
Controller valve.
Pressure Gauge Analog Dial type Pressure Gauge
Temperature Standard coupling type PT 100 Sensor ( GMP
Sensor model)
The electrical heat exchanger transfer latent heat
Heat Exchanger to the clean fill water to generated steam under
pressure.
Low water level cut off electronic circuit
Water Level protects the electrical heat exchanger from low
Control water level inside the chamber, this Cut's off the
power supply to the heat exchanger.
Steam generated from the unit is raised and exits
Steam Outlet thought the electrically operated steam Purging
solonide valve from the outlet ( GMP model)
Specification :

CS48a CS48b CS48c CS48d CS48e


Models
Capacity 35 ltrs 52 ltrs 119 ltrs 178 ltrs 213 ltrs
30 dia X 35 dia X 40 dia X 55 dia X 55 dia X
Inner Size
50 h 55 h 75 h 75 h 90 h
Inner
Stainless Steel 304 quality
Chamber
External Stainless Steel 304 quality
Range Ambient to 15 psi pressure / 121.40C
Mechanical Spring loaded value / µ based temp.
Controller
controller (GMP)
Sensor PT 100 (GMP)
Accuracy +/-0.10C (GMP)
Uniformity +/-0.20C (GMP)
Max. Ramp 45 min. upto 121.40C without load with respect to
up time ambient (GMP)
Insulation Air Bank 50 mm
Top Lid With Dome Shape Die press Stainless Steel Plate
Optional
Printer Output Printer output port RS232
Validation Data logging facility for temperature validation
Facility report.
Capacity
2000 3000 4000 5000 6000
(Watts)
Power
10 amps 15 amps 20 amps 25 amps 30 amps
Consumption
Net Weight
15kgs 20kgs 30kgs 50kgs 55kgs
approx.
Electric requirements AC230 V & 50 Hz cycle.