C122-E059A

Specification Sheet

AA-7000 Series
Atomic Absorption Spectrophotometers
The AA-7000 series of Atomic Absorption Spectrophotometers
features high-sensitivity analysis, flexible system configuration,
and a compact footprint for user-friendly operation. In addition,
the AA-7000 systems are the first AAS to employ a vibration
sensor as standard. The AA-7000 series ensures efficient and
comfortable operation in your laboratory.

Main Unit
Basics
Wavelength range 185.0 to 900.0 nm Choose from 100, 120, 220, or 230 VAC, 230
Aberration-corrected Czerny–Turner mounting, Number Power requirements VA, 50/60 Hz (Power is required separately for
Monochromator of grating grooves: 1800 lines/mm, Focal length: 300 mm the personal computer.)
• AA-7000F/AAC: W 700 × D 588 × H 714 mm, 76 kg
Bandwidth 0.2, 0.7, 1.3, 2.0 L nm (4-step automatic switching)
• AA-7000F: W 700 × D 588 × H 714 mm, 73 kg
Detector Photomultiplier tube Dimensions and • AA-7000G: W 700 × D 580 × H 538 mm, 66 kg

Optics Optical double-beam weight (Protruding parts and optional equipment are
Optics
• BGC-SR (high-speed self-reversal method) not included.)
Background (185.0 to 900.0 nm)
correction method Ambient temperature/ 10 to 35 °C, 20 to 80% (less than 70% when
• BGC-D2 (D2 lamp method) (185.0 to 430.0 nm) humidity temperature is higher than 30 °C)
Number of 6-lamp turret, 2 lamps simultaneously lit (1 for
HC lamps measurement, 1 warming up for next measurement)
Lamp mode EMISSION, NON-BGC, BGC-D2, BGC-SR
Software requirements Microsoft Windows 7 Professional / Vista Business Required Software Operation Environment
Parameter setting Wizard method The personal computer and monitor specifications are as follows.
Measurement Flame continuous method, flame micro sampling
mode method, furnace method, flame emission method Operating system Microsoft Windows 7 Professional (32/64 bit)
• Calibration curve method Intel® Celeron 420 (1.60 GHz or higher)
(select primary, secondary, tertiary) CPU To use GFA-TV, Intel® Pentium DualCore
Calibration mode • Standard addition method, simple standard E2180 (2 GHz) or higher is required.
addition method (primary expression) RAM 1 GB or higher
Up to 20 repetitions. Mean value, standard deviation Monitor XGA (1024 × 768 dots) or higher
(SD) and coefficient of variation (RSD) display
Repeat analysis Automatic exclusion of deviant values by setting SD One CD-ROM drive (for installing a software)
and %RSD Storage device Requires 60 MB min. hard disk space for
AA installation.
Automatic correction of baseline drift by offset
Baseline correction correction in peak height / peak area modes. I/O port One serial port (for AA control)
USB 2.0 port (for GFA-TV)
Signal processing Signal processing segments can be changed in
segment setting peak height / peak area modes. Peripheral device Monitor, keyboard, mouse, printer
Sensitivity Automatic calibration curve correction function * Notes
correction using sensitivity monitoring • The PC, monitor and printer are not included in the standard
Data 2 channels configuration. Prepare them separately.
processing (atomic absorption/energy signal, background signal) • The ASC-7000, ASK-7000 and GFA-7000A are optional accessories.
Analog output Output range: 5.0, 2.5, 1.25, 0.625 Abs./V
(each settable in 4 stages) • The separately available micro sampling kit is required for the flame
Fixed at 1 V F.S. in EMISSION mode. micro sampling method.
Final concentration calculations based on sampled • Microsoft Windows 7 Professional and Vista Business are registered
Calculation factor volume, dilution rate, fixed volume, and factor inputs trademarks of Microsoft Corp., U.S.A.
Method file Template functions available • The AA-7000 Series does not conform to IEC60601 –– Medical
Electrical Equipment.
Procedure/result display MRT (Measurement Results Table) worksheet
Report generation Summary report
Select whether to continue or discontinue measurements
QA/QC based on results of evaluation on coefficient of correlation,
%RSD, ICV·ICB, CCV·CCB, PB, LCS, SPK, PDS, and DUP.
• Select whether or not to conduct re-analysis.
• Automatic dilution and re-analysis of
Re-analysis unknown samples via autosampler (flame
micro sampling method, furnace method)
• Management by login ID and password
Digital recording • Control user access authority by user level
• Log record • Audit trail • Electronic signatures
Main Unit Autosampler
Flame Common Specifications
Type Air-cooled pre-mix type Control RS-232C communication control from AA unit
Titanium 10 cm slot (5 cm titanium slot for Maximum reagent / Reagents: 8 positions Samples: 60 positions
Burner head
N2O–C2H2 flame available as an option) sample positions (random access available with reagents or samples)
Nebulizer • Pt-lr capillary • PTFE orifice • ceramic impact Rinse water bottle 2L
bead (capable of handling hydrofluoric acid) Choose from 100, 120, 220, or 230 VAC,
Power requirements
Chamber Engineering plastics 50 VA, 50/60 Hz
• AA-7000F
Burner • Lateral/vertical manual adjustment
unit
Positioning • AA-7000F/AAC
• Automatic flame/furnace switching by motor Flame Continuous Method
• Automatic search of optimum burner height Zero-point detection, auto diagnosis, auto
0 to 90° (Angle adjustment is not possible Functions
Angle rinse, random access
when the optional GFA-7000A is installed on Sample volume Sample vials: 15 mL, reagent vials: 53 mL
adjustment
the AA-7000F/AAC.)
Nozzle rinse Solvent rinse aspiration method
Type Air–C2H2 flame, N2O–C2H2 flame
Sampler: W 340 × D 280 × H 270 mm, 8 kg
• Automatic fuel gas flow rate setting
Flow rate Dimensions and weight Controller (including stand):
(0.1 L/min step)
control W 260 × D 200 × H 320 mm, 7 kg
• Automatic search of optimum gas flow rate
• Automatic gas leak check
• Automatic switching between Air–C2H2 and
N2O–C2H2
• Flame monitor Furnace Method / Flame Micro Sampling Method
Gas Zero-point detection, auto diagnosis, auto
• Prevention of wrong burner head use Functions
control rinse, auto mixing, random access
Safety • Gas pressure monitor
measures • Drain tank level monitor Sample volume Sample vials: 2 mL, reagent vials: 20 mL
• Automatic flame extinction upon power Solvent rinse discharge method from overflow
Nozzle rinse
outage or sudden power interruption rinse-port
• Automatic flame extinction via flame Sampling functions Dilution function, reagent addition function
vibration sensor Syringe 250 μL
• Internal fan stop sensor Injection volume 2 to 90 μL
Repeatability 1%RSD (20 μL)
Rinse port: Less than 0.0001
Carryover
Mixing port: Less than 0.00001
Solvent rinse discharge method, solvent
Mixing port rinse
Furnace (GFA-7000A) discharge and rinse with next sample
Heating temperature rangeAmbient to 3,000 °C Performed in mixing port. Maximum mixture
• Mixing function
volume: 0.6 mL
Drying: Digital current control with automatic

temperature calibration function Up to 4 solutions
Heating control system Possible for mixing
• Ashing, Atomization: Digital temperature
control via optical sensor Maximum number of • Calibration curve method: 5 solutions max.
• Maximum 20 stages mixing reagents added (sample + 4 types of reagents)
• Heating mode: RAMP/STEP • Standard addition method: 6 solutions max.
(sample + standard solution + 4 types of reagents)
• Inner gas type: Dual automatic switching type
For measurement result on unknown samples
Setting heating • High-sensitivity mode possible
which exceed the calibration curve:
conditions • Enrichment in furnace: Maximum 20 times
• If extrapolation is possible: automatic
• Automatic optimum temperature search Auto dilution / calculation of dilution rate and dilution to
function for furnace program re-analysis bring concentration within calibration curve
• Inner gas flow rate: 0 to 1.50 L/min range
• Cooling water flow rate monitor • If extrapolation is not possible: dilution rate
• Gas pressure monitor fixed at 10×
Safety measures • Overcurrent protection unit (double check by Sampler (including sliding parts):
circuit protector and optical sensor) W 340 × D 280 × H 400 mm, 11 kg
Dimensions and weight
• Furnace block cooling check Controller: W 235 × D 110 × H 240 mm
(built into GFA-7000A), 6 kg
• AA-7000G
• Lateral/vertical manual adjustment
Positioning
• AA-7000F/AAC
• Automatic flame/furnace switching by motor
200, 220, 230, or 240 VAC ±5%, 7400 VA,
Power requirements
50/60 Hz
Dimensions and weight W 260 × D 560 × H 510 mm, 47 kg

For Research Use Only. Not for use in diagnostic procedure.


This publication may contain references to products that are not available in your country. Please contact us to check the availability of these
products in your country.
Company names, product/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation or its
affiliates, whether or not they are used with trademark symbol “TM” or “®”. Third-party trademarks and trade names may be used in this
publication to refer to either the entities or their products/services. Shimadzu disclaims any proprietary interest in trademarks and trade names
other than its own.

www.shimadzu.com/an/ Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication.
© Shimadzu Corporation, 2016