You are on page 1of 4

Measuring Magnetoelectric and Magnetopiezoelectric

Effects
J.T. Evans, S.P. Chapman, S.T. Smith, B.C. Howard, A. Gallegos
Radiant Technologies, Inc.
Albuquerque, NM 87107 USA
radiant@ferrodevices.com

The authors have developed a test procedure and circuits for Ferroelectric hysteresis testers use virtual ground measurement
characterizing the charge generation of a magnetoelectric device circuits so they are naturally configured to determine the
when stimulated with a magnetic field. The magnetoelectric magnetoelectric coefficient according to (2). The relationship
effect comes in two physical forms. The first is from a ferroic between  and ME may be algebraically derived as
material such as Bismuth Iron Oxide (BFO) which exhibits
internal coupling between its magnetic properties and its ME =  / 0r (3)
ferroelectric properties. The second is from composite devices
which are constructed by mechanically coupling a magnetic Equation (2) is defined using fundamental thermodynamic
material to a piezoelectric material. A magnetic field applied to a units. The equation means that the capacitor has a unique
composite device creates stress in the magnetic material which polarization state for each magnetic field state. Changing the
transfers the stress to the piezoelectric material. The traditional magnetic field forces the sample to a new polarization state.
method for measuring magnetoelectric effect is to stimulate the The change in polarization will result in charge flowing out of
sample with a magnetic field while measuring the voltage or into the sample under test provided there is a low impedance
generated across the sample. These voltages can be so small that current path available between the plates of the capacitor. If
a lock-in amplifier is required to pull the signal out of ambient there is no current path between the plates, the magnetically
noise. Charge measurement under the same circumstances generated charge will remain in the capacitor, generating a
produces a much larger signal and can be related to the voltage voltage by its action across the dielectric constant of the
response using the dielectric constant of the material [1]. The material.
charge test is an exact analog to the traditional electrical
hysteresis test with the exception that the test instrument applies The goal of this experiment was to identify the issues
a magnetic field to the sample instead of an electric field. The associated with executing a charge based magnetoelectric
presenter will describe the test procedure as presently used and measurement, minimizing any negative effects, determining the
outline future objectives for improving the accuracy and accuracy of the measurement sensors, and executing the
sensitivity of the results. measurement on a composite magnetoelectric device.
I. INTRODUCTION
II. SAMPLE
A capacitor with a nonzero magnetoelectric coupling
coefficient will generate charge, voltage, or a combination of A. Multiferroic vs Composite Magnetoelectrics
both when exposed to a magnetic field. The ratio of charge to
Devices with magnetoelectric properties come in two
voltage generated by the capacitor as a result of a change in an
flavors. The first has a multiferroic material as the capacitor
applied magnetic field is determined by 1) the impedance
dielectric layer. A multiferroic material has multiple
across the terminals of the capacitor and 2) the frequency at
properties, in this case ferromagnetism together with remanent
which the magnetic field is changing. The capacitor will
polarization or piezoelectricity. If the two properties are
generate voltage if the meter across capacitor has a high-
coupled in the same device, a change in polarization state could
impedance input. This arrangement is the traditional method of
be forced by a magnetic field or a change in polarization could
characterizing the magnetoelectric coefficient using a lock-in
be detected by a magnetic sensor. The second type of
amplifier as the meter. The coefficient measured in this
magnetoelectric device is a composite made by laminating a
manner is ME, defined as pure ferromagnetic material with a pure piezoelectric or
ME = E / H (1) ferroelectric material. The coupling between the magnetic and
electric properties is through the mechanical interface between
If a virtual ground is connected across the capacitor, the device the layers. For instance, a magnetic field will distort the
will not be allowed to generate a voltage so the magnetoelectric ferromagnetic material. That distortion will in turn activate the
response will appear as charge. The coefficient measured at piezoelectric material to create charge or voltage.
zero electric field is , defined as
 = P/H (2) B. Sample Preparation
The sample used for this experiment was fabricated at
Virginia Tech by Su Chul Yang under the auspices of the
Center for Energy Harvesting Materials and Systems
(CEHMS). It consisted of KNNLS-NZF cut into disks with a
diameter of 0.785 cm2 with a thickness of 0.05 cm. The disks
were laminated to a nickel foil. The maximum response for
this sample occurred at 460 Oersted: 12.5mV/cm/Oe. At 43
Oersted, the sample generated an ME response of
2.1mV/cm/Oe [2] using the voltage measurement technique.

III. TEST CONFIGURATION


The test is executed by substituting a magnetic waveform
for the traditional voltage stimulus in a polarization hysteresis
test. The substitution is accomplished by placing a current
amplifier on the voltage output of the hysteresis tester to drive
a Helmholtz coil. Either a current sensor in line with the
Helmholtz coil or a Hall Effect sensor placed at the sample
location can capture the magnetic field during the stimulus
waveform.
Magnetoelectric measurement conducted in this manner is Figure 1. Magnetoelectric test configuration.
subject to a variety of limitations and parasitics.
D. Measuring the Magnetic Field
A. Frequency Limitations
Due to the presence of the electrical shield box, a Hall Effect
The inductance of the Helmholtz coil limits the maximum sensor could not be placed in the vicinity of the sample to
frequency that may be applied to the sample. collect magnetic field strength during stimulus. The authors
VCoil = L I/t (4) used a high speed current sensor in line with the Helmholtz coil
to determine the magnetic field at the sample while in the
The voltage limit of the current amplifier used by the authors in central cylinder of uniformity of the Helmholtz coil.
Fig. 1 becomes Vcoil in (4). Combined with inductance L of the Comparison of the current sensor to a calibrated gaussmeter
coil used by the authors, it limited the maximum frequency to indicated an accuracy of 97.5% for the current sensor at 1 Hz.
20 Hz. Distortion of the stimulus waveform sets in at
frequencies below the limit, the reason why 1 Hz was used as E. DC Field Bias
the test frequency for this experiment.
Many magnetoelectric devices exhibit higher coefficients
when exposed to a DC magnetic bias field. A uniform DC bias
B. Charge Loss field can be achieved by placing the sample and the Helmholtz
The most sensitive charge measurement requires an coil at the center of a much larger field coil driven by its own
electrometer for the RETURN channel in Fig. 1. Electrometers high power amplifier. For this experiment, the authors applied
leak and this leakage limits the minimum frequency at which a small DC field bias by placing a permanent magnet at a set
the test may be run. The Radiant Technologies Precision distance from the sample. The field was reversed by flipping
Multiferroic nonlinear materials tester used for the experiment the polarity of the magnet.
in Fig. 1 was characterized for charge loss by executing a
traditional voltage-driven hysteresis test on a calibrated 10pF F. Small Signal Capacitance
reference capacitor. At one volt, the capacitor should generate
10pC of charge plus or minus the precision of the reference The mathematical steps to convert the  coefficient to the
capacitor, in this case a NIST-traceable 0.1%. The tester ME coefficient requires the use of the small signal capacitance
retained 99% of the charge during a 10 Hz test and 98.1% of of the device under test. Multiplying the magnetoelectric
the charge during a 1 Hz test. charge measured under zero impedance by the sample’s small
signal capacitance should predict the voltage the sample would
generate under open circuit conditions. Most magnetoelectric
C. Ambient Electrical Noise
devices have nonlinear small signal capacitance as a function
The charge measured during the experiment was so small it of magnetic field. The composite device under test exhibited a
was easily swamped by ambient electric noise in the room. 1% variance in its small signal capacitance as a function of a
Polarization hysteresis systems measure absolute charge so magnetic field varied over ±50 gauss. The small signal
ambient noise cannot be eliminated by lock-in techniques. The capacitance as a function of magnetic field is measured by
ambient noise was eliminated by placing the sample inside an executing the classic small signal capacitance electrical test on
aluminum electrical shield box grounded to the tester. The box the sample as it rests inside the Helmholtz coil while the
fit inside the Helmholtz coil and coaxial cables were used to surrounding magnetic coil generates an adjustable DC
connect the sample inside the shield box to the tester. magnetic field.
Q/H

Figure 2. Magnetoelectric charge response of the CEHMS composite laminated sample.

direction on both the center conductor and the metal shield


G. Parasitic Cable Capacitance braid. The parasitic contribution can be measured by running
A major advantage of the charge-based measurement approach the test without the sample present and then subtracting that
is the use of the virtual ground across the sample. With zero measurement from the one with the sample present. That
volts across the coaxial cables connecting the sample to the parasitic subtraction was not performed on the data below as
tester, the parasitic capacitance of the cable cannot affect the the test was executed at only 1 Hz.
test. The same is not true for the volt-based measurement
using a lock-in amplifier. The voltage generated by the sample I. Voltage Bias
must also charge the cables and the input capacitance of the
lock-in amplifier, reducing the value of the measured voltage It is possible with the test configuration used for this
from the value the sample would generate if it were truly experiment to apply a voltage bias to the sample during
isolated. This parasitic effect can be subtracted out by magnetic field AC actuation. That was not done for this
characterizing the cables and test fixture for capacitance but the experiment. The sample electrodes were shorted with zero
compensation must be adjusted for every sample and every set volts on one plate and the virtual ground on the other.
of cables.
IV. RESULTS
H. Induced Charge Three tests were executed on the sample at different DC
The cables connecting the sample to the test must intrude field biases:
inside the AC magnetic field generated by the Helmholtz coil.  ±45G with 0G bias
The AC magnetic field will induce current in the cables as it
varies amplitude. This charge will add to the charge generated  ±45G with +56G bias
by the sample in the charge-based measurement or the voltage
generated by the sample in the voltage-based measurement.  ±45G with -56G bias
The induced charge is linearly related to the test frequency, The bias fields were created by a nearby permanent magnet,
meaning that a 1Hz test will have 0.1% of the induced charge meaning that the DC bias field was not uniform across the
as a 1 kHz test. The geometric relationship of the cables to the diameter of the sample. The ±56G value used for plotting the
AC magnetic field will also have a dramatic effect on the results was measured with a gaussmeter at the location of the
magnitude of the induced current. In this case, the DRIVE and center of the sample. The three measurements are plotted in
RETURN cables were parallel to the AC H-field axis and Fig. 2.
displaced from the axis by approximately the diameter of the
coils. Shielded coax connected to a differential input on the Since the charge plotted in Fig. 2 is an absolute response of
test equipment should further reduce the impact of induced the sample to the applied magnetic field, its  coefficient at
charge since the induced charge will move in the same each magnetic field value is the slope of the polarization
generated at that field. The zero bias charge generation result its electrometer circuit, measured at 1.9%. Thus, the charge
in Fig. 2 was fit with a polynomial resulting in (5) having an R2 measurement error for this experiment extends from -2.4% to
value of 0.9664. +0.5%. The magnetic field measurement error was measured
to be ±2.5%. The  coefficient is found by dividing the charge
Q = 0.0012x2 + 0.006x -0.2126 pC(Oe) (5)
by the magnetic field. Using the maximum charge uncertainty
The derivative of (5) gives the equation that calculates the against the minimum magnetic uncertainty and vice versa
slope of the zero bias charge response in Fig. 2 at a specified yields a maximum error range of +3.1% to -4.8%, ignoring
magnetic field. induced current. Therefore there is a reasonable expectation
that the  of 0.139pC/cm2/Oe and the ME of 2.87mV/cm/Oe
Q′ = 0.0024x + 0.006 pC/Oe (6) calculated for this sample at 43 Oersted are accurate to within
Dividing (6) by the area of the sample results in the equation ±5%. The next step in our research will be to correlate these
that calculates the  coefficient of the sample at the specified results with those produced by the traditional voltage-based
magnetic field. measurement technique for the same sample.
 = 0.00306x + 0.0076 pC/cm2/Oe (7)
ACKNOWLEDGMENT
2
At 43G,  equals 0.139pC/cm /Oe. The authors wish to acknowledge the invaluable assistance
The ME coefficient is calculated from (6) by dividing that of Dr. Shashank Priya, Su Chul Yang, and Shashaank Gupta
charge value by the small signal capacitance of the sample at from CEHMS at Virginia Tech.
the magnetic field strength of interest (761pF) to convert to
units of volts and then again by the sample thickness (0.05cm) REFERENCES
to convert to electric field. [1] C. S. Park, J. Evans, and S. Priya, “Quantitative understanding of the
elastic coupling in magnetoelectric laminate composites through the
ME = 6.31x10-5 x + 1.58x10-4 mV/cm/Oe (8) nonlinear polarization-magnetic (P-H) response,” Smart Mater. Struct.,
vol. 20, 6pp., 2011.
At 43G, ME equals 2.87mV/cm/Oe.
[2] Sample synthesized by Su Chul Yang of the Center for Energy
Harvesting Material and Systems. All data rights are retained by
ANALYSIS CEHMS.

The error in the charge measurement consists of the 0.5%


specification provided by the manufacturer for the Precision
Multiferroic tester used in the test plus the charge leakage from

You might also like