Professional Documents
Culture Documents
11
Siemens Medical Solutions USA, Inc.
TABLE OF CONTENTS
1. INTRODUCTION................................................................................................. 3
1.1 Purpose of the Document............................................................................................... 3
1.2 Scope................................................................................................................................ 3
2. PREPARATION .................................................................................................. 4
2.1 Camera Quality Control .................................................................................................. 4
2.3 Supplies............................................................................................................................ 6
3. WORKFLOW TEMPLATES................................................................................ 7
3.1 Equipment Performance Workflow Templates............................................................. 7
Page 2 of 40
Siemens Medical Solutions USA, Inc.
1. INTRODUCTION
1.1 Purpose of the Document
The American College of Radiology (ACR) has created a “Nuclear Medicine Accreditation
Program” to evaluate the quality of personnel, equipment performance, effectiveness of
quality control measures and the quality of the clinical images of Nuclear Medicine facilities.
Please refer to the ACR documents for Nuclear Medicine Accreditation for a detailed
explanation of the specific requirements for the accreditation program.
ACR testing can be divided into two categories:
• Equipment Performance
o Flood Field Uniformity
o Spatial Resolution
o SPECT Phantom
• Clinical Images
o Module 1- Planar: A whole body or spot bone and another whole body bone,
spot bone, hepatobiliary, perfusion lung or MUGA
o Module 2 - SPECT: Bone SPECT and another Bone SPECT, Brain SPECT,
Hepatic blood pool, Liver SPECT and SPECT Myocardial Perfusion
o Module 3 - Nuclear Cardiology: SPECT Myocardial Perfusion and a MUGA or
Gated SPECT
1.2 Scope
This document contains instructions for performing Nuclear Medicine ACR Accreditation on
the Symbia and e.cam Imaging Systems with software version of syngo MI Applications
2007A or higher.
Page 3 of 40
Siemens Medical Solutions USA, Inc.
2. PREPARATION
Since the ACR equipment performance tests are performed with Tc-99m, Co-57, Tl-201 and
Ga-67. It is important for the quality control be completed for the low and medium energy
collimators that you will be using for these tests and your clinical images. In most cases,
these collimators would be the LEHR (low energy high resolution) and the ME or MELP
(medium energy) collimators.
Please refer to Nuclear Medicine /PET Accreditation Program Requirements for more details
and your ACR accredited physicist about the report that must be submitted with the ACR
Accreditation
intrinsic calibration flood should be performed again. The integral and differential
uniformity numbers should be recorded at the time the calibration is performed.
2.1.2.2 Extrinsic Calibration Flood Requirements (Only needed If collimator has been
damaged since installation.)
The camera should be peaked and tuned immediately preceding the extrinsic calibration
flood acquisition. A sheet source of up to 15 mCi Co-57 is sufficient for this calibration,
which should be performed according to recommendations from the user’s manual.
Please note that an intrinsic calibration flood needs to be acquired within 10 days of an
acquisition of an extrinsic calibration flood. To verify that the sheet source is not too hot,
the dead-time reported in the analyzer page should be below 7% for both detectors. A
fillable sheet source using 15 mCi Tc-99m in water can also be used but extreme care
should be used to ensure uniform mixing of the source.
A 120 million count extrinsic calibration flood is recommended for optimum Jaszczak
imaging. If the existing calibration flood has fewer counts, or if the number of counts in
the existing calibration is not known, the extrinsic calibration flood should be performed
again
NOTES
Use 5 point sources for the LEHR collimator.
Page 5 of 40
Siemens Medical Solutions USA, Inc.
NOTES
Sufficient time should also be allocated for image selection,
annotation, and filming. Phantom positioning and data acquisition,
including measurements described in this and the ACR
Accreditation documentation may be very time consuming.
Prior to performing all ACR tests, ensure that all of the camera quality control is current.
2.3 Supplies
• ACR / Jaszczak flangeless phantom
• Phantom Holder
• Point source vials
• Sheet source
• ACR Documentation
• Nuclear Medicine ACR Accreditation Hint’s and Tips: Symbia S, Symbia E and
e.cam systems with syngo MI Workplaces.
Page 6 of 40
Siemens Medical Solutions USA, Inc.
3. WORKFLOW TEMPLATES
3.1 Equipment Performance Workflow Templates
Table 1 contains the name of the workflow templates that have been created to acquire,
process, and display ACR test data. These workflows are available on workstations that are
running syngo MI Applications 2007B or higher.
These templates can be found in the ACR Category on the Command Module. If the
templates have not been loaded, they can be found on the Optional Workflow Template
CD.for Symbia systems and the service pack cd for e.cam systems.
SPECT Images
Acquisition Sec. 4.3.4 ACR Phantom SPECT ACQ
Reconstruction Sec. 4.3.5 ACR Phantom Reconstruction
The clinical measurements are based upon the site’s daily acquisition and processing
workflows.
Page 7 of 40
Siemens Medical Solutions USA, Inc.
4.1.2 Configuration:
Intrinsic Flood Field Uniformity
A point source in a scatter free environment should be used. Source activity should be
targeted to achieve between 5 kcps and 25 kcps in window with no collimator on the detector.
The activity should be ~15-25 μCi for acquiring an intrinsic flood with the source between the
detectors
Acquisition with source between the detectors:
• Move the camera to the 180 degree configuration
• Rotate detector 1 to 0 degrees and fully retract the detectors
• Place the source in the source holder and center the source between the detectors
• Acquire using the appropriate Flood Field Uniformity workflow template
Alternatively, the point source can be placed 5 fields of view away (approximately 9 feet) and
images for each detector can be acquired separately. This can be accomplished with the
system in the 90 degree or outer room configurations.
Page 8 of 40
Siemens Medical Solutions USA, Inc.
• Simultaneously acquire the images from both detectors using the ACR Co-57 Extrinsic
Flood Uniformity workflow template.
A list of the field uniformity workflows can be found in Section 3.1. These workflow templates
can be found in the ACR template category. An example of the activities and parameters of
those workflows can be found in Figure 1. An example of intrinsic flood field uniformity can be
found in Figure 2.
Figure 1: Workflow Template: ACR Tc99m Field Uniformity for source centered between the
detectors
NOTES
The raw data is a pair of 8 million count floods. After passing through the Flood
Calculation activity, they are transformed into 5 million count floods in a 256 x 256
matrix.
Page 9 of 40
Siemens Medical Solutions USA, Inc.
Figure 2: Example of an intrinsic flood field uniformity acquired with a Tc-99m point source
4.2.1 Overview:
ACR has two protocols for measuring spatial resolution. For cameras that only acquire planar
images, the Planar Only Protocol should be followed. All others should follow the SPECT and
Planar Protocol
Page 10 of 40
Siemens Medical Solutions USA, Inc.
A list of the spatial resolution workflows can be found in Section 3.1. These workflow
templates can be found in the ACR template category. Parameters for the static acquisition in
the ACR Tc-99m Spatial Resolution Bars Workflow are shown in Figure 3.. Typical results with
a bar phantom with 2.0-3.5 mm bars shown in Figure 4.
Page 11 of 40
Siemens Medical Solutions USA, Inc.
The spatial resolution for the Planar and SPECT protocol is acquired extrinsically using the
filled Jaszczak phantom. For details on source preparation and phantom filling please refer to
section 4.3.3
Page 13 of 40
Siemens Medical Solutions USA, Inc.
The rod and sphere sizes for flangeless Jaszczak SPECT Phantom that is included in the
Siemens ACR Phantom Kit are:
• Cold Rod Dimensions: 4.8 mm, 6.4 mm, 7.9 mm, 9.5 mm, 11.1 mm, 12.7 mm
• Solid Sphere Diameters: 9.5 mm, 12.7 mm, 15.9 mm, 19.1 mm, 25.4 mm, 31.8 mm
A list of the spatial resolution workflows can be found in Section 3.1. These workflow
templates can be found in the ACR template category. Parameters for the static acquisition in
the ACR Tc-99m Spatial Resolution Phantom Workflow are shown in Figure 6. Typical results
are shown in Figure 7.
Page 14 of 40
Siemens Medical Solutions USA, Inc.
Page 15 of 40
Siemens Medical Solutions USA, Inc.
Page 16 of 40
Siemens Medical Solutions USA, Inc.
• Rotate the detectors to ensure that they clear the phantom and the pallet. Adjust
radius if necessary. Note: The head with the smaller of the two detector radii should be
moved out so that both detectors have the same radius (approximately 21.5 cm or
closer
• Confirm phantom alignment, both axially and transaxilly, via the PPM
4.3.3 Acquisition
By default the template will be set up to acquire Tc99m. If using the template for Tl201, change
the series name to differentiate the study. If using Ga67, use the Gallium two peak preset (92
and 185 keV peaks) for optimal phantom imaging and change the series name. The Jaszczak
SPECT image should be set up with acquisition parameters of 128 views over 360°. ACR
recommends acquiring approximately 24M counts in the SPECT study. The acquisition time
per view field will need to be adjusted accordingly. The image size should be set to 128x128
and the image zoom should be set to 1.45. Uniformity corrections should be enabled. A
patient orientation of Head First is recommended but not required.
Select the ACR Phantom patient and launch the ACR Phantom SPECT ACQ workflow
template (Figure 8). The following information should be noted on the ACR Data Form:
Series Name:
Activity (mCi)
Collimator:
Matrix Size: 128 x128 (default)
Radius of Rotation
Orbit: Step and Shoot or Continuous (Step and Shoot is the default)
Orbit Shape: Circular or Non-Circular (Circular is recommended)
Acquisition Zoom: 1.45 (default)
Time/View: See ACR calculation below.
Slice Thickness: 3.37 (if using the zoom and matrix of 1.45 and 128 x 128)
Analyzer Setting: Refer to Analyzer Tab in Tomo Acquisition Activity
Page 17 of 40
Siemens Medical Solutions USA, Inc.
NOTES
ACR Time/View Calculation
The time per view for the phantom acquisition is equal to:
In the ACR Phantom SPECT ACQ, the number of views is defaulted to 128)
Page 18 of 40
Siemens Medical Solutions USA, Inc.
4.3.4 Processing
Select the ACR Phantom patient and launch the ACR Phantom Recon workflow
template(Figure 9). The following information should be noted on the ACR Data Form:
Page 19 of 40
Siemens Medical Solutions USA, Inc.
Figure 10: Workflow Template: ACR Phantom Recon: Chang’s Attenuation Correction
NOTES
Cine the raw data. If there is a tilt in the cine, use the 3D Orientation activity
within this workflow to properly align the corrected transverse images.
Select the reconstructed Jaszczak series and launch the ACR SPECT Phantom Analysis
workflow template to generate the following information:
• Uniformity Composite Image
• Resolution Composite Image
• Contrast Composite Image
• Contrast Composite Analysis (Regions of Interest and Curves)
Page 20 of 40
Siemens Medical Solutions USA, Inc.
In the first Series Arithmetic Activity, select the Arithmetic tab. Shift-left-click to select 6 slices
from the uniform section of the phantom just under the lid, between sphere and end of
phantom (Figure 13)
. With these slices highlighted, click on the Composite tool in the toolbar (Figure 12):
The composite image will then be displayed in the Operation Output: Composite Frames
series display at the bottom of the Arithmetic tab (Figure 13). The Rename Output Series
from the Series Arithmetic menu has been set so that the output series ‘Uniformity Composite
Image’ to distinguish it from other composite images. Additionally this output is connected to
Flexible Display so that the proper image size and annotations required by ACR can be edited
in this activity. Ensure that the image is clearly labeled with the image size, zoom, isotope and
total counts.
Page 21 of 40
Siemens Medical Solutions USA, Inc.
In the second Series Arithmetic Activity, select the Arithmetic tab. Shift-left-click to select 16
slices from the cold rod section at the end of the phantom (Figure 14). With these slices
highlighted, click on the Composite tool in the toolbar
The composite image will then be displayed in the Operation Output: Composite Frames
series display at the bottom of the Arithmetic tab. The Rename Output Series from the Series
Arithmetic menu has been set so that the output series ‘Resolution Composite Image’ to
distinguish it from other composite images. Additionally this output is connected to Flexible
Display so that the proper image size and annotations required by ACR can be edited in this
activity. Ensure that the image is clearly labeled with the image size, zoom, isotope and total
counts.
Page 22 of 40
Siemens Medical Solutions USA, Inc.
In the third Series Arithmetic Activity, select the Arithmetic tab. Shift-left-click to select 2 slices
from the cold sphere section in the upper third of the phantom (Figure 15). With these slices
highlighted, click on the Composite tool in the toolbar.
The composite image will then be displayed in the Operation Output: Composite Frames
series display at the bottom of the Arithmetic tab. The Rename Output Series from the Series
Arithmetic menu has been set so that the output series ‘Contrast Composite Image’ to
distinguish it from other composite images. Additionally this output is connected to Flexible
Display so that the proper image size and annotations required by ACR can be edited in this
activity. Ensure that the image is clearly labeled with the image size, zoom, isotope and total
counts.
Page 23 of 40
Siemens Medical Solutions USA, Inc.
The cold sphere contrast can be found by generating a line profile through the contrast
composite image. This must be done in the Series ROI and Curve activity.
On the Processing tab of this acitivity, select the image frame, and click on the ROI tool .
The ROI Tool window (Figure 17) opens. Select Sphere from the ROI menu and the Line
Profile ROI. . Draw a single pixel profile through the center of the largest sphere in the
contrast image. Select No Sphere from the ROI menu and draw another single pixel line
Page 24 of 40
Siemens Medical Solutions USA, Inc.
across the center of the Contrast image without intersecting any sphere if possible. The
profile should extend well outside the phantom on both sides.
Figure 17: Series ROI And Curve Activity: ROI Tool Window
After the lines are created, select the red line and generate the line plot by selecting the Line
Curve tool. . Repeat for the green line. Curves will appear in the Line Profile
Graph.
Page 25 of 40
Siemens Medical Solutions USA, Inc.
Page 26 of 40
Siemens Medical Solutions USA, Inc.
4.3.5.5 Displays
The Flexible Display Activity contains the following tabs that can be printed out onto individual
film or saved as jpegs: Uniformity, Resolution, Contrast, Contrast Analysis and Transverse (2).
There is also a print tab that combines all of the others into a 4:1 display that is suitable for
printing on a sheet of 11 x 14 film.
Page 27 of 40
Siemens Medical Solutions USA, Inc.
Page 28 of 40
Siemens Medical Solutions USA, Inc.
• Note: This image does not have to be submitted. Only the numbers are needed.
Page 29 of 40
Siemens Medical Solutions USA, Inc.
Page 30 of 40
Siemens Medical Solutions USA, Inc.
Page 31 of 40
Siemens Medical Solutions USA, Inc.
4.3.7.1 Camera QC
To ensure maximum image quality, the Monthly Intrinsic Flood Calibration should be updated
within the month preceding the study. The camera should be Peaked and Fine-Tuned before
the flood is acquired. The Intrinsic Flood Field Uniformity (uncorrected) should fall within
accepted levels.
Rings in the transverse reconstruction may be caused by inadequate counts in the flood
corrections. If the statistics of the flood corrections do not exceed the total counts in the
SPECT image (both the extrinsic and the intrinsic flood should be considered), the noise in the
flood image will dominate those in the image data. The flood noise is reinforced at every view,
and this results in rings in the reconstruction, especially near or at the center of the FOV. The
minimum flood counts required for this ACR acquisition are 200 million Intrinsic flood counts
and 120 million Extrinsic flood counts. If any uncertainty in the flood statistics exists, they
should be repeated. Flood corrections should be used in the SPECT images.
Poor resolution will result if the 180° MHR/COR/Head Alignment corrections are out of date.
Since this calibration takes a modest amount of time, it is recommended that the calibration be
performed just prior to the ACR acquisitions. The camera user manual discusses acceptable
limits for COR/MHR/Head Alignment values. In general, these calibration results should be
less than the following:
Symbia 5 Point Head Alignment Values (valid for low energy collimators)
given to the final agitation of the phantom. A small amount of food coloring may be added
before the last fill to help show when the phantom is adequately mixed.
Page 33 of 40
Siemens Medical Solutions USA, Inc.
4.3.7.8 Chang’s AC
Chang’s attenuation correction is a multiplicative correction that is applied after reconstruction,
once the object outlines have been determined. Because of the ideal shape of the Jaszczak
phantom, it is most efficient to determine the boundary outlines manually. The circular outline
should be adjusted to closely match the outline of the phantom. The attenuation coefficient is
assumed to be constant within the boundaries. An attenuation coefficient of approximately
80% of the ideal value is typically assumed (0.12 cm-1).
This attenuation correction method calculates attenuation seen by a point of emission in the
image matrix, and averages over emission view angles. Since the correction is applied post-
Page 34 of 40
Siemens Medical Solutions USA, Inc.
Figure 24 Cylindrical Phantom Count Profile, for Ideal (shown in blue) and Attenuation-
Corrected (Changs’) Images. Results at left have used a coefficient of 0.15 cm-1 for
correction. Results at right have used a coefficient of 0.12 cm-1 for correction.
Page 35 of 40
Siemens Medical Solutions USA, Inc.
5. CLINICAL MEASUREMENTS
5.1 FAQ’s
5.1.1 Determining whole body bone count density:
To determine the count density in a wholebody image, simply divide the number of counts in
the image by the area. The area of the image can be found activated by selecting Size within
the Properties dialog of the image. This dialog box can be displayed by right clicking on the
bottom window pane of the image.
The Size is displayed in mm2. This will need to be converted to cm2. Divide the number of
counts in the image by the converted size to get the count density of the view.
Example: In the following dataset, the Size of the image is: 1506930.37 mm2. To convert to
cm2, simply divide by 100. (1506930.73/100) = 15069.3 cm2 Now to calculate the counts
density, divide the number of counts (910000) by area in cm2 (15069.3). In this example the
count density would equal 60.37 cts/ cm2
Counts in image
Size
(Area)
Page 36 of 40
Siemens Medical Solutions USA, Inc.
Page 37 of 40
Siemens Medical Solutions USA, Inc.
Page 38 of 40
Siemens Medical Solutions USA, Inc.
Page 39 of 40
Siemens Medical Solutions USA, Inc.
Page 40 of 40